SOP for Temperature Mapping Study


Standard operating procedure of temperature mapping study in biological incubators, refrigerator and freezer using calibrated probes and data logger in pharmaceutical
industries.
1.0 OBJECTIVE
To lay down the procedure for temperature mapping.
2.0 SCOPE
This SOP shall be applicable for walk-in coolers, biological incubators, refrigerator and freezer.
3.0 RESPONSIBILITY
Officer/ Executive/ Assistant Manager

4.0 ACCOUNTABILITY
HOD of Engineering and QA
5.0 PROCEDURE
5.1 Temperature mapping is to be carried out by using the calibrated temperature probes and data logger.
5.2 Identify positions & number of locations for temperature mapping such that, it covers every corner & center of the chamber.
5.3 Prepare a protocol for temperature mapping indicating objective, scope, responsibilities, methodology, acceptance criteria, calibration status,
summary and conclusion.
5.4 Protocol to be prepared by officer/ executive of concerned department, checked by the head of concerned department and shall be approved by
head QA/QC.
5.5 Place temperature probes at the identified locations without touching the tip to any surface inside the chamber.
5.6 Select the number of channels and set the print interval of 30 minutes on the data logger.
5.7 Start the mapping after stabilization of the displayed temperature.
5.8 Record the temperature readings of each probe in temperature mapping record.
5.9 If results do not fall within the specified limit, take necessary corrective actions to rectify the failure & perform the mapping again.
5.10 Final report for temperature mapping to be prepared by officer/ executive of the concerned department based on results obtained indicating the
summary and conclusion.
5.11 Report shall be checked by the head of concerned department and shall be approved by head QA/QC.
5.12 After approval of the report daily monitoring shall be done at the identified hot spot.
5.13 For new equipment, the mapping is to be done for three consecutive days.
5.14 Remapping to be done for equipment once in the year for 24 hours.
Get Faster Cleavage of Peptide
5.15 Acceptance Criteria OctaGel™ saves you money and out-performs Open
ChemMatrix
Temperature shown by individual temperature probe should and than
not be more standard Rink Amide resins.
25±2°C.
6.0 ABBREVIATIONS
6.1 SOP: Standard Operating Procedure
6.2 QA: Quality Assurance
6.3 QC: Quality Control

Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of pharmaguideline.com, a widely-read pharmaceutical
blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question

Copyright © 2008-2022 Pharmaguideline


Setup ❘ Terms of Use ❘ FAQ

You might also like