Experiment 4 2000
Experiment 4 2000
Experiment 4 2000
Objective
1. Determine the various conductivity of serial sodium chloride solution and calcium
carbonate solution.
2. Observe the relationship between conductivity and concentration in a solution.
Introduction
The unit of measurement for conductivity is Siemens/cm. This is a very high conductivity level,
so in practice, millisiemens/cm (mS/cm) and microsiemens/cm (μS/cm) is commonly reported.
In theory, ideal conductivity cell measures the ability of current to flow through a trapped
volume of solution between two plates 1 cm square, and separated by 1 cm.
If conductivity cell has becomes fouled with residue from adhering samples this should be
removed using a suitable solvent; check it is compatible with the materials used in the probe
assembly or it should be then thoroughly rinsed with deionized water.
Chemicals
Sodium chloride serial solutions (0.1 mol/l; 0.5 mol/l; 1.0 mol/l; 2.0 mol/l)
Calcium carbonate serial solutions (0.1 mol/l; 0.5 mol/l; 1.0 mol/l; 2.0 mol/l)
Distilled water
Equipment
Sensor Cassy
Cassy Lab
Conductivity sensor
Crossed bosshead
Miniature clamp
Beakers
Volumetric flask, 50 ml
Pipette balls
Spatula
Weighing paper
Funnel
Balance
Watch glass
Procedure
CaCo3 standard solutions (0.1 mol/l; 0.5 mol/l; 1.0 mol/l; 2.0 mol/l) in 50 ml
volumetric flask. (Mw of CaCo3 is 100.0869 g/mol)
Calculate amount of CaCo3 powder needed to make the standard solutions!!
2. Experiment Setup
Plug the chemistry box with the conductivity sensor into input A of the sensor CASSY. After
thoroughly rinsing the conductivity sensor with distilled water, fix it in the clamp, so that its
height can be adjusted with the crossed bosshead.
Rinse the electrode and the beaker thoroughly with 20 ml of the solution.
Fill the rest of the solution into the beaker.
Adjust the height of the conductivity sensor so that its immersion depth in the
liquid is 2 cm and its distance from every wall at least 1 cm.
If necessary, change the measuring range in Setting Conductivity CA1. Always
select possible measuring range.
When the displayed value is stable, record the measured value by pressing F9 or
the clock symbol.
Enter the concentration and the substance associated with the measured value in
the diagram. To do this, press Alt + T, type the substance and the value of the concentration and
the position text at desired place in the diagram with the mouse.
In order to simplify the evaluation, select Append New Meas. Series in
Measuring Parameters before measuring the conductivity of a new substance.
Before measuring the conductivity of a new substance, thoroughly rinse the
beaker and conductivity sensor with distilled water.
Take the first reading of conductivity before the number goes down!!
Before measuring the conductivity, mix the solution by turning upside down several times!!
2. Using the above data, plot the Conductivity (mS/cm) in Y axis against
Concentration (in mol/L) in X axis!
Use manual plot with graph paper.
Draw separately: NaCl and CaCo3.
Add pure distilled water conductivity value as blank in each curve.
Find the equation (linearity by least square) of each line (NaCl calibration curve and
CaCo3 calibration curve).
3. Determine the concentration of NaCl and CaCo3 in the samples (in mol/l)
4. In the case of conductivity measurement; find three difference between NaCl and
CaCo3.
5. Explain the following terms:
a) Strong electrolytes
b) Weak electrolytes
c) Degree of dissociation
6 Write the conclusion of the experiment.