931-932 - User Manual
931-932 - User Manual
931-932 - User Manual
User Manual
TQM
COMMITTED TO TOTAL QUALITY
Process Instruments
PN 903-8600, Rev. D Western Research
2876 Sunridge Way N.E.
Calgary, Alberta, Canada T1Y 7H9
Offices
CHINA
CANADA Room 406, Metro Tower
2876 Sunridge Way N.E. No. 30 Tian Yao Qiao Road
Calgary, Alberta T1Y 7H9 Shanghai, 200030
Phone: 403-235-8400 Phone: 8621 6426 7049
Fax: 403-248-3550 Fax: 8621 6426 7054
CHAPTER 1 OVERVIEW
Theory of the Model 931 and Model 932 UV Analyzers ............................ 1-1
Model 932 Multi-Gas UV Analyzers ....................................................... 1-2
Model 931 Single-Gas UV Analyzers ...................................................... 1-2
Features ............................................................................................................... 1-3
Application ......................................................................................................... 1-4
Implementation ................................................................................................. 1-4
Microcontroller Board and Host Controller Board.............................. 1-5
Discrete Wavelengths ................................................................................ 1-6
About the Analyzer Sample System ............................................................... 1-7
Relays ................................................................................................................... 1-9
Status Relays ................................................................................................ 1-9
Alarm Relays .............................................................................................. 1-10
Solid State Relays ...................................................................................... 1-10
CHAPTER 2 SPECIFICATIONS
Methodology ...................................................................................................... 2-1
Standard Range Measuring Cell .................................................................... 2-1
Response Time ................................................................................................... 2-2
Measurement Accuracy ................................................................................... 2-2
Repeatability ....................................................................................................... 2-2
Calibration .......................................................................................................... 2-2
Linearity .............................................................................................................. 2-2
Stability ................................................................................................................ 2-2
Cross-Talk ............................................................................................................ 2-2
Temperature Drift ............................................................................................. 2-2
24-Hour Zero Drift ............................................................................................ 2-3
Analog Outputs ................................................................................................. 2-3
Digital Communication ................................................................................... 2-3
Contents | iii
Electrical Requirements ................................................................................... 2-3
Maximum Start-Up Power ........................................................................ 2-3
Maximum Current / Supply Voltage....................................................... 2-4
Status Relay Contacts ................................................................................ 2-4
Typical Flow ........................................................................................................ 2-4
Ambient Limits .................................................................................................. 2-4
Temperature ................................................................................................. 2-4
Humidity ...................................................................................................... 2-4
Maximum Altitude ..................................................................................... 2-4
Zero Gas .............................................................................................................. 2-5
Sample Transport............................................................................................... 2-5
Sample Pressure Requirement (Optional) .................................................... 2-5
Instrument Air Requirements ......................................................................... 2-5
Pressure Compensation (Optional) ............................................................... 2-5
Temperature Control ........................................................................................ 2-5
Physical Dimensions (on Backpan) ............................................................... 2-6
Approvals and Certifications .......................................................................... 2-6
Hydrogen Sensor (Optional)........................................................................... 2-6
Accuracy ....................................................................................................... 2-6
Cross-Talk on Hydrogen ............................................................................ 2-6
ATEX (CENELEC) Certificate and Analyzer Marking ............................... 2-7
CSA Certificate and Analyzer Marking ...................................................... 2-10
Contents | v
Matrix ................................................................................................. 4-41
Analytical Parameters ........................................................................ 4-42
Analog Inputs ..................................................................................... 4-43
Setting Gas Calibration Parameters .................................................. 4-45
Alarm Setting and Stream Switch Parameters ................................. 4-51
Analyzer Information ......................................................................... 4-55
Status Tab ................................................................................................... 4-57
Analyzer Data Tab ..................................................................................... 4-62
Optical Bench Data Tab ........................................................................... 4-65
Scanning Data .................................................................................................. 4-67
Display Modes ........................................................................................... 4-68
Defining Scan Settings ............................................................................ 4-68
Starting a Scan ........................................................................................... 4-70
Stopping a Scan ........................................................................................ 4-70
Scheduling Scans ...................................................................................... 4-70
Working With Data Files ................................................................................ 4-72
Saving Data Files ....................................................................................... 4-72
Opening Data Files ................................................................................... 4-73
Scrolling Through Data Records ........................................................... 4-74
Converting Data Files .............................................................................. 4-74
Printing Data ............................................................................................. 4-74
CHAPTER 5 CALIBRATION
Manual Span Gas Calibration ......................................................................... 5-1
Analog Output Calibration .............................................................................. 5-5
Current Calibration .................................................................................... 5-5
Voltage Calibration ..................................................................................... 5-7
Contents | vii
APPENDIX B OPTIONS
Environmental Purge Option (Field Upgrade Available) ......................... B-1
Connection and Use ................................................................................. B-1
Pressure Compensation Option (Field Upgrade Available) ..................... B-3
Setting Up the Pressure Compensation Option.................................. B-3
Using the Pressure Compensation Option .......................................... B-5
Diagnostics Feature ............................................................................. B-5
Measuring Cells ................................................................................................ B-6
Loop-Powered Current Outputs (Field Upgrade Available) ................... B-6
Using the Loop-Powered Current Option ........................................... B-6
Gas Sensor Interface Options ........................................................................ B-7
Digital Input (Contact Interface) Option .................................................... B-7
APPENDIX D DRAWINGS
Upper Enclosure Layout ............................................................................... D-2
Lower Enclosure Layout ................................................................................ D-3
Backpan Component Layout ........................................................................ D-4
AC Distribution PWB (P/N 100-1771) .......................................................... D-5
Customer I/O Board (P/N 100-1758) ............................................................. D-6
Customer Connections, I/O Board (120 VAC) ........................................... D-7
Host Controller Board (P/N 100-1757) ......................................................... D-8
Customer Connections, Host Controller Board (North America/
Europe) .............................................................................................................. D-9
RS-485 Port Terminations and Bias Options (Host Controller Board) ......D-10
Microcontroller Board (P/N 100-1781) ........................................................D-11
Micro-Interface Board (P/N 100-1898) ........................................................D-12
Optical Bench Service Electronics Board (P/N 100-1662) ....................... D-13
Conduit Entries and AC Power Connection Locations (North
America) ........................................................................................................... D-14
RS-232 Cable Communications Connections ........................................... D-15
Analyzer Wiring Diagram .............................................................................D-16
Inlet and Vent Flame Arrestor Locations for Environmental Purge
Option .............................................................................................................. D-17
PMT Buffer Board (P/N 100-0140-2) ............................................................D-18
Optical Bench Assembly (Model 932) .........................................................D-19
Solenoid Block (North American), 100-1892 ..............................................D-20
Solenoid Block (European), 100-1893 ..........................................................D-21
SUPPLEMENTAL INFORMATION
An operating procedure which, if not strictly observed, may result in personal injury or
! environmental contamination.
WARNING
An operating procedure which, if not strictly observed, may result in damage to the equipment.
!
CAUTION
Electrical Safety
High voltages are present in the analyzer housings. Always shut down power source(s) before
performing maintenance or troubleshooting. Only a qualified electrician should make electrical
connections and ground checks.
Any use of the equipment in a manner not specified by the manufacturer may impair the safety
protection originally provided by the equipment.
Grounding
Instrument grounding is mandatory. Performance specifications and safety protection are void if
instrument is operated from an improperly grounded power source.
Contents | ix
Personnel and Equipment Safety Information
This section describes important safety information to avoid personal injury and damage to the
equipment while installing, operating, maintaining, or servicing the equipment. All safety
regulations, standards, and procedures at the analyzer location must be followed.
All personnel involved with the installation, start-up, operation, maintenance, service, or trou-
bleshooting of the analyzer must review and follow these Warnings and Cautions.
Warnings
Review and follow these Warnings to avoid personal injury or environmental contamination.
Always disconnect main AC power and/or alternate power sources (if used, i.e., for relay
! contacts) to the analyzer before opening any covers on the analyzer.
WARNING
If it is necessary to open the Upper/Lower Enclosures while the circuits are alive, test the area
for hazardous gases and proceed only when it is found to be safe. When the EEx d Electronics
Enclosures are open, take appropriate precautions to avoid electrical shock. Hazardous voltages
are present inside.
Any electrical connections, adjustments, or servicing of the analyzer should be performed only
! by properly trained and qualified personnel.
WARNING
All electrical connections, materials, and methods (plus all safety policies and procedures) must
be made in compliance with local wiring regulations and electrical code for the hazardous
location, and be approved by the Owner Company.
Before working on the sample system, purge the sample system with zero gas and confirm
! it is isolated (blocked in) from the process.
WARNING
Because ultraviolet radiation can harm your eyes, avoid direct viewing of the light emanating
! through the end window of the source lamp. If the source lamp must be viewed while energized,
WARNING
wear safety glasses that block ultraviolet radiation.
If the analyzer uses the optional Measuring Cell heater, take precautions to avoid burning
! yourself; the Measuring Cell and Heater Assembly are hot.
WARNING
The electronic circuit boards and other static-sensitive components should be stored and
! transported in static-shielding carriers or packages.
CAUTION
For electrical-shock protection, the analyzer must be operated from a grounded power source
! that has a securely connected protective-ground contact.
CAUTION
If it becomes necessary to handle any of the electronic circuit boards, do not subject the boards
! to static discharge. The ideal solution is a static-safe work area. Since such areas typically are
CAUTION
not available at field installation sites, the use of a wrist strap connected directly to a ground is
recommended. If a wrist strap is not available, you should at the very least touch the metal
chassis (to ground yourself) before handling or touching the boards.
When handling the source lamps, it is very important not to touch the lamp windows because
! residual oils from the fingers will absorb ultraviolet light. The window is the flat surface at the
CAUTION
end of the narrow glass tube. The lamp assembly is fragile and should be handled with care.
Do not exceed the maximum working pressure rating of the sample system – 21 000 KPAG
! (3000 PSIG) without optional Filterblock Assembly – under any circumstances.
CAUTION
For analyzers using the optional Filterblock Assembly, do not exceed the maximum working
pressure rating of the sample system – 14 000 KPAG (2000 PSIG) with optional Filterblock
Assembly – under any circumstances.
If the Analyzer is equipped with the Environmental Purge option, clean air or nitrogen can be
! supplied to the inlet flame arrester at a pressure of 7–70 KPAG (1–10 PSIG).
CAUTION
Contents | xi
Warning Labels
These symbols may appear on the instrument in order to alert you of existing conditions.
Schutzerde
Read and follow the recommendations in this section to avoid performance variations or
! damage to the internal circuits of this equipment when installed in harsh electrical environ-
CAUTION ments.
The various configurations of the Model 931 and Model 932 UV Analyzers should not produce,
or fall victim to, electromagnetic disturbances as specified in the European Union’s EMC Direc-
tive. Strict compliance to the EMC Directive requires that certain installation techniques and
wiring practices are used to prevent or minimize erratic behavior of the Analyzer or its electronic
neighbors. Below are examples of the techniques and wiring practices to be followed.
In meeting the EMC requirements, the various Analyzer configurations described in this manual
rely heavily on the use of metallic shielded cables used to connect to the customer’s equipment
and power. Foil and braid shielded I/O and DC power cables are recommended for use in
otherwise unprotected situations. In addition, hard conduit, flexible conduit, and armor around
non-shielded wiring also provides excellent control of radio frequency disturbances. However,
use of these shielding techniques is effective only when the shielding element is connected to
the equipment chassis/earth ground at both ends of the cable run. This may cause ground loop
problems in some cases. These should be treated on a case-by-case basis. Disconnecting one
shield ground may not provide sufficient protection depending on the electronic environment.
Connecting one shield ground via a 0.1 microfarad ceramic capacitor is a technique allowing
high frequency shield bonding while avoiding the AC-ground metal connection. In the case of
shielded cables the drain wire or braid connection must be kept short. A two-inch connection
distance between the shield’s end and the nearest grounded chassis point, ground bar or termi-
nal is highly recommended. An even greater degree of shield performance can be achieved by
using metallic glands for shielded cable entry into metal enclosures. Expose enough of the braid/
foil/drain where it passes through the gland so that the shield materials can be wrapped back-
wards onto the cable jacket and captured inside the gland, and tightened up against the metal
interior.
Inductive loads connected to the low voltage “Alarm Contacts” are not recommended. However,
if this becomes a necessity, adhere to proper techniques and wiring practices. Install an appropri-
ate transient voltage suppression device (low voltage MOV, “Transzorb,” or R/C) as close as
possible to the inductive device to reduce the generation of transients. Do not run this type of
signal wiring along with other I/O or DC in the same shielded cable. Inductive load wiring must
be separated from other circuits in conduit by using an additional cable shield on the offending
cable.
In general, for optimum protection against high frequency transients and other disturbances, do
not allow installation of this Analyzer where its unshielded I/O and DC circuits are physically
mixed with AC mains or any other circuit that could induce transients into the Analyzer or the
overall system. Examples of electrical events and devices known for the generation of harmful
electromagnetic disturbances include motors, capacitor bank switching, storm related transients,
RF welding equipment, static, and walkie-talkies.
Contents | xiii
SPECIAL WARNINGS AND INFORMATION
EQUIPMENT USED IN HAZARDOUS LOCATIONS
This Equipment is Suitable for Use in Hazardous Locations as follows:
ATEX (KEMA)
II 2 G EEx d IIB T3.
Explosion Hazard – Substitution of Components May Impair Suitability for Class I, Division 2.
!
WARNING
Explosion Hazard – Do Not Disconnect Equipment Unless Power Has Been Switched Off or
! the Area is Known to be Non-Hazardous.
WARNING
All input and output wiring must be in accordance with Class I, Division 2 wiring methods (NEC
Sec 501.4(b) or CEC 18-152) and in accordance with the authority having jurisdiction.
Contents | xv
WARRANTY AND CLAIMS
We warrant that any equipment of our own manufacture or manufactured for us pursuant to our specifica-
tions which shall not be, at the time of shipment thereof by or for us, free from defects in material or workman-
ship under normal use and service will be repaired or replaced (at our option) by us free of charge, provided
that written notice of such defect is received by us within twelve (12) months from date of shipment of
portable analyzers or within eighteen (18) months from date of shipment or twelve (12) months from date of
installation of permanent equipment, whichever period is shorter. All equipment requiring repair or replace-
ment under the warranty shall be returned to us at our factory, or at such other location as we may designate,
transportation prepaid. Such returned equipment shall be examined by us and if it is found to be defective as a
result of defective materials or workmanship, it shall be repaired or replaced as aforesaid. Our obligation does
not include the cost of furnishing any labor in connection with the installation of such repaired or replaced
equipment or parts thereof, nor does it include the responsibility or cost of transportation. In addition, instead
of repairing or replacing the equipment returned to us as aforesaid, we may, at our option, take back the
defective equipment, and refund in full settlement the purchase price thereof paid by Buyer.
Process photometric analyzers, process moisture analyzers, and sampling systems are warranted to perform the
intended measurement, only in the event that the customer has supplied, and AMETEK has accepted, valid
sample stream composition data, process conditions, and electrical area classification prior to order acknowl-
edgment. The photometric light sources are warranted for ninety (90) days from date of shipment. Resale items
warranty is limited to the transferable portion of the original equipment manufacturer’s warranty to AMETEK.
If you are returning equipment from outside Canada, a statement should appear on the documentation
accompanying the equipment being returned declaring that the goods being returned for repair are Canadian
goods, the name of the firm who purchased the goods, and the shipment date.
The warranty shall not apply to any equipment (or part thereof) which has been tampered with or altered after
leaving our control or which has been replaced by anyone except us, or which has been subject to misuse,
neglect, abuse or improper use. Misuse or abuse of the equipment, or any part thereof, shall be construed to
include, but shall not be limited to, damage by negligence, accident, fire or force of the elements. Improper use
or misapplications shall be construed to include improper or inadequate protection against shock, vibration,
high or low temperature, overpressure, excess voltage and the like, or operating the equipment with or in a
corrosive, explosive or combustible medium, unless the equipment is specifically designed for such service, or
exposure to any other service or environment of greater severity than that for which the equipment was
designed.
The warranty does not apply to used or secondhand equipment nor extend to anyone other than the original
purchaser from us. Should the Buyer’s technical staff require the on-site assistance of AMETEK’s agents or
employees for service calls covered by this warranty clause, the Buyer shall pay travel time plus actual travel
and living expenses.
THIS WARRANTY IS GIVEN AND ACCEPTED IN LIEU OF ALL OTHER WARRANTIES, WHETHER
EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION AND WARRANTIES OF FITNESS OR OF
MERCHANTABILITY OTHER THAN AS EXPRESSLY SET FORTH HEREIN, AND OF ALL OTHER
OBLIGATIONS OR LIABILITIES ON OUR PART. IN NO EVENT SHALL WE BE LIABLE UNDER THIS
WARRANTY OR ANY OTHER PROVISION OF THIS AGREEMENT FOR ANY ANTICIPATED OR LOST
PROFITS, INCIDENTAL DAMAGES, CONSEQUENTIAL DAMAGES, TIME CHANGES OR ANY OTHER
LOSSES INCURRED BY THE ORIGINAL PURCHASER OR ANY THIRD PARTY IN CONNECTION
WITH THE PURCHASE, INSTALLATION, REPAIR OR OPERATION OF EQUIPMENT, OR ANY PART
THEREOF COVERED BY THIS WARRANTY OR OTHERWISE. WE MAKE NO WARRANTY, EXPRESS
OR IMPLIED, INCLUDING WITHOUT LIMITATION ANY WARRANTIES OF FITNESS OR OF MER-
CHANTABILITY, AS TO ANY OTHER MANUFACTURER’S EQUIPMENT, WHETHER SOLD SEPA-
RATELY OR IN CONJUNCTION WITH EQUIPMENT OF OUR MANUFACTURE. WE DO NOT AUTHO-
RIZE ANY REPRESENTATIVE OR OTHER PERSON TO ASSUME FOR US ANY LIABILITY IN CON-
NECTION WITH EQUIPMENT, OR ANY PART THEREOF, COVERED BY THIS WARRANTY.
The AMETEK Model 931 and Model 932 UV Analyzers use photometric
analysis to monitor H2S levels in acid gas (and other) applications or other
high dewpoint applications. The microprocessor-based Model 931 and
Model 932 Analyzers minimize the complexity of these tasks and simpli-
fies operation of the analyzer system.
The Beer-Lambert law, in combination with the ideal gas situation, states
that:
Overview | 1-1
Model 932 Multi-Gas UV Analyzers
In a situation where several species are absorbing light at several measur-
ing wavelengths, the total absorbance at any particular wavelength is the
sum of the absorbances for each of the species present. Using Beer’s law
for each measuring wavelength, a system of linear equations relating the
unknown concentration to the measured absorbance at each wavelength
is obtained. The total absorbance at each measuring wavelength is equal
to the proportionality constant multiplied by the molar absorptivity for
the first species times its concentration, plus the (molar absorptivity) factor
for the second species, times its concentration, and so on for all species
present in the Measuring Cell. Provided that there are more measuring
wavelengths than unknown concentrations, this system of linear equa-
tions is solved using standard methods of linear algebra.
[ x ] kλ l P
A= ,
RT
where l is the path length of the light in the gas sample, [x] is the concen-
tration (mole percent [see Note]) of absorber x, kλ is the molar absorptivity
of absorber x at the measuring wavelength, λ, T and P are the temperature
and pressure of the gas sample, respectively, and R is the universal gas
constant.
For any given system at constant temperature and pressure, the equation
can be simplified to:
A= C kλ [ x ] ,
Auto Zero
Automatically updates the analyzer zero. You can initiate the Auto-
Zero function on command using the Configurator Software, at user-
specified time intervals, or by a remote (potential free) contact clo-
sure.
Serial Communication
One RS-485 serial communications port is dedicated to customer data
acquisition. The port uses Modicon Modbus® protocol (optional).
Analog Outputs
There are eight independent output-signal channels. Four channels
consist of voltage outputs (1–5 V full-scale) while the other four are
isolated 4–20 mA outputs.
Temperature Control
The analyzer controls the temperature of up to three temperature-
controlled zones:
• Optical Bench
• Measuring Cell
• Heated Sample Probe (optional)
Independent over-temperature circuits protect the three zones.
Status Relays
There are five independent sets of SPDT (Form C) relay contacts (2 A,
250 V).
Overview | 1-3
Low Maintenance
The analyzer and sample system design reduces maintenance re-
quirements.
Application
The principal application of the Model 931 and Model 932 Analyzers is
the analysis of hydrogen sulfide (H2S) in acid gas and other process
streams. It can also be used to measure other UV-absorbing compounds.
Implementation
The Model 931 and Model 932 Analyzers are comprised of (see Figure 1-1):
Overview | 1-5
Discrete Wavelengths
Model 932 Multi-Species Analyzers
To achieve measurements at discrete wavelengths, the locations of the
filters in the filter wheel and the placement of the source lamps are chosen
so that, at any given time, the photodetectors are exposed to light from
only one lamp and filter combination (i.e., a discrete wavelength).
For practical reasons related to timing and lamp control, two revolutions
of the filter wheel are required to collect a complete set of wavelength
data. During one revolution, consecutive pulses of light from Source
Lamp 1 arrive at the beam splitter. During the next revolution, consecutive
pulses of light arrive at the beam splitter from Source Lamp 2. Since the
nominal speed of the filter wheel is 400 RPM, pulses for any given wave-
length arrive at the beam splitter at a rate of 200 pulses per minute.
For the Model 931 Analyzers measuring H2S in acid gas (and other)
applications, 228 nm is used to measure the H2S absorbance and 326 nm
is used for reference.
When absorbers are present in the gas sample, the amount of light received
by the Measure photodetector is reduced further. The magnitude of the
reduction is dependent upon the concentrations of the absorbers. Thus, for
each wavelength, the difference in the amounts of light received by the two
photodetectors (corrected for the optical-path factors) is used to determine
the absorbance of the gas sample. Then, the concentrations of the individual
absorbers (for which the analyzer is configured) are calculated.
The major components of the sample system and their functions are
detailed below. The components listed below are for a variety of analyzer
configuration possibilities. Depending on your application, only some of
these components will be used. See Figure 1-2 for an example of a
Plumbing (Flow) Schematic configuration.
Component Function
Sample Probe Extends into the process stream to obtain a representative
sample. Can be heated or unheated. Heated probes return
used gas sample to the process stream.
Sample Shut-Off Valve Isolates the sample path from the process stream.
Heated Sample/Vent Lines
Optional – Used to keep the sample gas above the dewpoint
temperature (application-specific). Some applications vent
the used gas sample to a low pressure flare header (<0.5
PSIG).
Measuring Cell The Measuring Cell is a chamber of known length, with
optical-grade quartz windows, through which the radiation
beam passes through the gas sample. The windows are
sealed by means of o-rings and removable end caps.
Overview | 1-7
LEGEND
Figure 1-2.
Plumbing schematic
(European).
Status Relays
The Customer I/O board contains three independent relays that indicate
the operational status of the analyzer. Each relay provides a set of SPDT
(form C) dry (potential free) contacts. The relays are configured for fail-
safe operation (i.e., energized for the non-alarm condition). Each relay
also has an associated green LED on the Customer I/O board, which
remains on steady to indicate a non-alarm (normal operating) condition.
• The Fault relay indicates that an error has been detected by the built-
in diagnostics system. If this relay is de-energized, the green LED
(D404) will turn off to indicate the analyzer requires service. If the
relay is configured to latch (default) it will not clear unless the user
manually resets the alarm; otherwise, it will reset automatically upon
correction of the fault.
• The Alarm Valid relay indicates whether information from the Con-
centration 1 and Concentration 2 relays is valid or whether the
analyzer is capable of alarming. The contacts automatically change
state upon correction of the condition which caused the alarm. If this
relay is de-energized, the green LED (D406) will turn off to indicate
there is a problem.
Overview | 1-9
Alarm Relays
There are two independent relays, which indicate that the gas concentra-
tion has exceeded the alarm On-Threshold or Off-Threshold setpoint.
Each relay provides a set of SPDT (Form C), dry (potential free) contacts.
Each relay has an associated red LED (D407 and D408 on the Customer
I/O board), which lights to indicate an alarm condition.
If values are assigned to these alarm relays, they are used for concentra-
tion alarms; if there are no values assigned to either of these relays, they
become stream status relays. The relay contacts for each alarm will remain
in the alarm state after the concentration returns to normal if the relay
configuration is set to be latching (default). It is possible to configure the
concentration alarm to be non-latching if desired. Refer to “Alarm Setting
and Stream Switching Parameters” in Chapter 4 for more information.
The Specifications listed in this chapter apply to both Model 931 and
Model 932 Analyzers, except where noted.
Methodology
Model 931 Analyzers:
Dual wavelength, high resolution, non-dispersive UV.
Specifications | 2-1
Response Time
90 % response in 30 seconds or less (excluding sample system)
Measurement Accuracy
± 1.0 % full-scale of standard ranges (excluding drift and temperature
effects). Accuracy is determined by comparing analyzer response to a
known standard gas after a calibration has been performed.
Repeatability
<± 0.5 % full-scale of standard ranges
Calibration
Calibration gases in nitrogen
Linearity
Typically <± 1.0 % of reading for H2S
Stability
Noise: <0.2 % full-scale of standard ranges
Cross-Talk
Typically better than 1 % cross-interference between species used
(application-specific).
Temperature Drift
H2S @ 228 nm = (40 PPM / Cell length) / °C at atmospheric pressure
Analog Outputs
Up to four isolated 4–20 mA DC current outputs, loop-powered or self-
powered.
Four non-isolated voltage outputs, 1–5 VDC full-scale.
Digital Communication
One RS-485 port (for Modicon Modbus® protocol).
One RS-232 port (or optional RS-485) for service diagnostics.
Explosion-Proof Serial Communication Port on Lower Enclosure.
Electrical Requirements
With optional Heated Acid Gas Probe and Measuring Cell heater:
<500 W
Specifications | 2-3
Maximum Current / Supply Voltage
ATEX/Hazardous Locations:
120 ± 10 % VAC, 50–60 Hz, 3 A Maximum
240 ± 10 % VAC, 50–60 Hz, 2 A Maximum
CSA:
104–132 VAC, 47–63 Hz, 3 A Maximum
207–264 VAC, 47–63 Hz, 2 A Maximum
Output Signal
4–20 mA DC, 30 VDC Maximum
Typical Flow
2.5 L/minute (5 SCFH)
Ambient Limits
Temperature
From 0–50 °C (32–122 °F).
For applications using unheated Measuring Cells, the surrounding ambi-
ent temperature of the analyzer must be at least 5 °C above the sample
dew point temperature. In high dewpoint applications, a heated Measur-
ing Cell and heat-traced Sample/Vent Lines must be used to prevent
condensation of liquids in the sample system.
Humidity
0–95 % RH
Maximum Altitude
2000 m
Sample Transport
Application dependent.
Temperature Control
Independent control of up to three temperature-controlled zones:
• Optical Bench
• Measuring Cell
Specifications | 2-5
Physical Dimensions (on Backpan)
Height: 1185 mm (46.65")
Accuracy
± 5.0 % full-scale of standard ranges. Accuracy is based on a full-scale of
10 % H2 in a Scot Contactor Overhead application, where the sensor is in
a temperature-controlled oven.
Cross-Talk on Hydrogen
± 1.3 % of H2 full-scale. This is the RMS error in the reading due to CO2,
H2O, and pressure effects.
Specifications | 2-7
2-8 | Model 931 and Model 932 UV Analyzers
ATEX-certified Model 931 Analyzers are marked with the label shown
below.
0344
PROCESS INSTRUMENTS II 2 G
Caution
Do not open when an explosive gas atmosphere is present.
ATEX-certified Model 932 Analyzers are marked with the label shown
below.
0344
PROCESS INSTRUMENTS II 2 G
Caution
Do not open when an explosive gas atmosphere is present.
Specifications | 2-9
CSA Certificate and Analyzer Marking
For installation sites that require CSA certification, AMETEK’s CSA certifi-
cate for the Model 931 and Model 932 Analyzers is included in the follow-
ing pages.
PROCESS INSTRUMENTS
Caution
Open circuit before removing cover.
CSA-certified Model 932 Analyzers are marked with the label shown
below.
PROCESS INSTRUMENTS
Caution
Open circuit before removing cover.
Specifications | 2-13
This page intentionally left blank.
This chapter shows you how to install and start up the Model 931 and
Model 932 UV Analyzers. It includes:
• Pre-installation requirements
• Electrical installation
• Normal operation
• Analyzer configuration
The analyzer and its backpan weighs approximately 146 kg (320 lb).
! Use caution when lifting it from its crate.
WARNING
After the inspection, replace and tighten at least one hex screw for
! each enclosure. This will keep the enclosures secure and will prevent
CAUTION damage to the joining surfaces (flamepaths) during installation.
When handling the source lamps, it is very important not to touch the
! lamp windows because residual oils from the fingers will absorb
CAUTION
ultraviolet light. The window is the flat surface at the end of the
narrow glass tube. The lamp assembly is fragile and should be handled
with care.
Do not mount the Optical Bench in the EEx d Upper Enclosure until
! after the analyzer has been mounted in its designated location.
CAUTION
The Model 931 and Model 932 Analyzers are designed for indoor opera-
tion (Pollution Degree 2) so they must be shielded from the elements.
The entire analyzer system (and its backpan) can be mounted directly on
a wall inside a building, in a specially designed cabinet, or in a custom-
built shelter. In all cases, the analyzer system must be installed indoors to
ensure it is shielded from harsh environmental elements.
Figure 3-1.
Analyzer backpan
dimensions (North
American).
All electrical connections to the Optical Bench Assembly are made via
pre-wired connector plugs. No hard wiring is required.
2. Locate the cam pins, which are part of the Optical Bench Mount
(Figure 3-2). Confirm that the grooved lines in the cam pins are
facing inward.
The cam pins must stay aligned while the Optical Bench Assembly
is being mounted in the Upper Enclosure. The lined groove on the
NOTE
head of the cam pin must stay parallel to the Measuring Cell until
the Optical Bench Assembly is in place.
4. Rotate the lower cam pin counter-clockwise and the upper cam pin
clockwise until they are tight.
Figure 3-2.
Upper Enclosure
(Model 932) layout and
cam pin locations.
11 10
1 2 3 4
SOLENOID BLOCK DETAIL
Figure 3-3.
Overall component
layout (North American).
For installation procedures, refer to the Heated Acid Gas Sample Probe
Installation & Maintenance Guide.
The Sample and Vent Lines can be supplied by AMETEK or by the cus-
tomer. The tubing must be 1/4" 316 stainless steel tubing with appropriate
1/4" NPT connectors. Applications that used AMETEK’s Heated Sample
Probe will return the used sample gas to the process stream through the
probe. Alternate configurations are to disperse the used sample gas to a
low pressure flare header or vent pipe to atmosphere.
Refer to the Analyzer Plumbing Schematic and other related drawings for
important installation information. These drawings are located in the
“Supplemental Information” tab of this manual, and in the Documenta-
tion Package shipped with the analyzer.
The vent line should vent to a low pressure (<0.5 PSIG) flare header.
NOTE
The Zero/Aspirator gas line is supplied by the customer. The tubing must
be 1/4" 316 stainless steel tubing with appropriate 1/4" NPT connectors.
Refer to system drawings for Zero/Aspirator Drive Gas pressure settings
for your application.
2. Blow down (purge) the line to remove any liquids or particulate that
may be present in it line before connecting it to the analyzer.
3. Connect the line to the Zero/Aspirator Drive Gas Inlet on the ana-
lyzer.
4. Connect the Zero Gas air line to the Zero Gas cylinder or the Instru-
ment Air line to the Instrument Air supply.
Where required, certified EEx d cable glands must be used and each
! unused cable entry port must be plugged with a certified EEx d plug.
WARNING
OR
OR
Figure 3-4.
Customer Connections,
I/O board (120 VAC
applications), example
only.
Hazardous Locations:
! Before attempting to access the Digital Communications Port, test the area for
WARNING
hazardous gases and proceed only when the area is found to be safe.
There are also two other serial port connections available for permanent
connection: Remote Service Port and Customer Data Acquisition Port
(Modbus communications), both RS-485.
JP300 = 1-2
Local Service Port RS-232 or Modem JP301 = 1-2
JP302 = 1-2
JP300 = 2-3
Remote Service Port RS-485 JP301 = 2-3
(* See Note.) JP302 = 2-3
OR
OR
Figure 3-5.
Customer Connections,
Host Controller board.
If using the Remote Service Port (RS-485), you must configure it for Full-
Duplex (4-Wire) or Half-Duplex (2-Wire), as shown below in the “Remote
Service Port RS-485, Full/Half Duplex” configuration, and Figures 3-5/3-6.
The RS-485 Remote Service Port can also be used for customer
service, but not both at the same time as the RS-232 Digital Commu-
NOTE
nications Port.
Hazardous Locations:
! Do not power up the laptop computer if there is an explosive gas
WARNING
atmosphere present.
2. If using the RS-485 port, select either a 2- or 4-wire link for both
Customer Data Acquisition and Remote Service ports.
3. For termination and biasing of RS-485 links, refer to Figure 3-7. Refer to
Figures 3-5 and 3-6 for communication/termination options and
jumper (JP) locations/settings.
The termination and biasing options will depend on the analyzer role
and location in the network (master or slave, located at the end of the
transmission line or at a drop point). It will also depend on whether a
2- or 4-wire line is used, and the transmission line conditions (e.g.,
length of cable, or termination and biasing on the other end of the
cable).
A
RX+
R J302 (Customer Data Acquisition) J303 (Remote Service Port)
130
RT
B
RX- JP305 (RX RT) 1-2 JP307 (RX RT) 1-2
JP306 (TX RT) 1-2 JP308 (TX RT) 1-2
130
RT
Z
TX- JP314 (TX RB2) 2-3 JP316 (TX RB2) 2-3
+5V
R B1
619
A
RX+
R J302 (Customer Data Acquisition) J303 (Remote Service Port)
B
RX- JP305 (RX RT) 2-3 JP307 (RX RT) 2-3
R B2
2-3 2-3
(c) +5V JP309 (RX RB1) 1-2 JP311 (RX RB1) 1-2
JP310 (TX RB1) 1-2 JP312 (TX RB1) 1-2
R B1
619
619
+5V
R B1
619
A
RX+
R J302 (Customer Data Acquisition) J303 (Remote Service Port)
130
RT
B
RX- JP305 (RX RT) 1-2 JP307 (RX RT) 1-2
R B2
1-2 1-2
(d) +5V JP309 (RX RB1) 1-2 JP311 (RX RB1) 1-2
JP310 (TX RB1) 1-2 JP312 (TX RB1) 1-2
R B1
619
Z
TX-
NOTES:
1. RB - BIAS RESISTOR
R B2
619
2. RT - TERMINATION RESISTOR
3. RESISTANCE VALUES IN OHMS
Figure 3-7.
Various RS-485 port
terminations and bias
resistor configurations
(Host Controller board
Rev. B and newer).
For safety reasons during maintenance, this switch allows the main
AC power to be disconnected from the analyzer prior to performing
service on the analyzer.
DETAIL A
Figure 3-8.
Conduit entries and
AC power connection
locations, Lower
Enclosure (North
American).
Hazardous Locations:
! Before proceeding, test the area around the analyzer for hazardous
WARNING
gases and proceed only when the area is found to be safe.
Figure 3-9.
Destination Folder
screen.
You can also start the software by clicking 931 & 932 Configurator
from the Start\Programs\Ametek menu.
The software will create a shortcut icon on the desktop to start the
software (called 931 & 932 Configurator).
NOTE
Figure 3-10.
InstallShield Wizard
Completed screen.
1. From the Windows Start menu, select Settings, and then Control
Panel.
Before powering up the analyzer, test the area for flammable gases. If
! an explosive gas atmosphere is present, do not apply power to the
WARNING
analyzer or any alternate power sources that supply power to the
analyzer components. Proceed only when the area is found to be safe.
Before operating the analyzer for the first time, following a power-up
! or reset, or after maintenance, you must manually block in the sample
WARNING gas flow.
• The state of the Fault (Fail) Status Relay and the Warning
Status Relay depends on the current status of the analyzer.
However, it is most likely that both relays will be de-energized
upon initial power-up as the temperature of the Optical
Bench has not yet reached its setpoint, and is generating an
alarm error fault (“f: Probe temperature too low” – if the
Heated Sample Probe is used) and warning (“w: Bench tem-
perature out of range”). Verify this in the HC Scode and MC
Scode alarm lists on the Status tab of the 931 & 932 Properties
dialog box. These green LEDs D404 (Fail) and D405 (Warning)
will be off in this case.
c. The Probe heater (if the Heated Sample Probe is used) will be
turned on to heat up the Probe. Verify this by viewing the red
LED D401 on the Customer I/O board (Figure 3-11). It should be
on steady or pulsing.
You can verify these heaters are on by observing that the red LED
(LED201, Optical Bench heater) and the LED (LED200, Measur-
ing Cell heater) on the Micro-Interface board (see Figure 3-2 for
location of board) are on continuously or pulsing.
First, select the Analyzer Data tab to view the current tempera-
ture readings, under Temperature Zones. Record the current
values for each temperature-controlled zone. You can use these
values as a reference later while comparing the temperature value
(after the analyzer has warmed up). This will help you to deter-
mine if the signal wiring is properly terminated and if all of the
circuitry is operating properly.
Next, click the Setup tab and then click the Temp Control button
to view the Temperature Control dialog box (Figure 3-13).
(Optical) Bench
Verify that the (Optical) Bench temperature Setpoint and Range
are correct. The Bench temperature setpoint should be at least
8 °C to 10 °C above the maximum ambient temperature around
the analyzer. For applications with maximum ambient tempera-
ture of 30 °C and below, a default setpoint temperature of 40 °C is
normally used. The temperature setpoint is set at the factory
using the ambient temperature data that is specific to the applica-
tion and location. Refer to Chapter 4 for additional information
about how to alter the setpoint or range.
Verify that the Too Hot setpoint for the Bench is correct. It
should be 5 °C to 10 °C above the Bench Setpoint temperature.
The Too Hot setpoint that is used should not exceed 65 °C. The
Too Hot setpoint is set at the factory using the ambient tempera-
ture data that is specific to the application and location.
Verify that the Too Cold setpoint for the Bench is correct. It
should be 5 °C to 10 °C below the Bench Setpoint temperature.
The Too Cold setpoint is set at the factory using the ambient
temperature data that is specific to the application and location.
Verify that the Too Hot setpoint for the Cell is correct. It should
be 5 °C to 10 °C above the Cell Setpoint temperature. The Too Hot
setpoint is set at the factory using the ambient temperature data
that is specific to the application and location.
Verify that the Too Cold setpoint for the Cell is correct. It should
be 5 °C to 10 °C below the Cell Setpoint temperature. The Too
Cold setpoint is set at the factory using the ambient temperature
data that is specific to the application and location.
Verify that the Too Hot setpoint for the Probe is correct. It should
be 5 °C to 10 °C above the Probe Setpoint temperature. The Too
Hot setpoint is set at the factory using the ambient temperature
data that is specific to the application and location.
Verify that the Too Cold setpoint for the Probe is correct. It
should be 5 °C to 10 °C below the Probe Setpoint temperature.
The Too Cold setpoint is set at the factory using the ambient
temperature data that is specific to the application and location.
Verify the correct range for the current and voltage output
channels:
From the Setup tab click the Output button to view the Output
Setup dialog box (Figure 3-14).
• For each configured channel, confirm that the Low Scale and
High Scale for Current Outputs and Voltage Outputs are the
correct minimum and maximum ranges required. Compare
these values to the factory settings listed on the Analyzer
Programming Parameters sheet included in the “Supplemen-
tal Information” tab of this manual, or the Documentation
Package shipped with the analyzer.
Figure 3-14.
Output Setup dialog box
(with two streams).
• The alarm setpoint ranges are set at the factory using the data
that is specific to the application. Compare these values to the
factory settings listed on the Analyzer Programming Param-
eters sheet included in the “Supplemental Information” tab of
this manual, or the Documentation Package shipped with the
analyzer.
Figure 3-15.
Alarm Setting and
Stream Switch
Parameters dialog box.
From the Setup tab, click Modbus to view the Modbus dialog
box (Figure 3-16).
Figure 3-16.
Modbus dialog box.
To perform this test, follow the procedures listed under the section
titled “Sample System Leak Check,” later in this chapter. Then,
return to this section and complete the remaining steps.
5. After all of these checks have been made and verified, close the Upper
and Lower Enclosures. The sample gas should remain blocked in.
Hazardous Locations:
! Before proceeding, test the area around the analyzer for hazardous
WARNING
gases and proceed only when the area is found to be safe.
b. After the area has been tested for hazardous gases (and has been
found to be safe), loosen the hex screw on the cover of the Explo-
sion-Proof Digital Communications Port (RS-232) on the Lower
Enclosure enough to allow the cover to slide to the right and
reveal the Digital Communications Port.
a. Click the 931 & 932 Configurator shortcut icon on the desktop or
from the Start\Programs\Ametek menu. The software will start
and the Devices & Log window will appear in the AMETEK 931
& 932 Configurator window. From here, select Device Properties
from the Edit menu to view the 931 & 932 Properties dialog box
(General tab, Figure 3-17).
The 931 & 932 Properties dialog box will be your primary interface
when working on the analyzer or viewing live data.
NOTE
Figure 3-17.
General tab
(with Modbus Serial
communication
settings).
b. By default, the Live Data and Device Enabled check boxes are
enabled. When Live Data is selected, the Configurator Software
automatically attempts to establish communication with the
analyzer. If the communication attempt fails, the communication
settings may be incorrect and/or the communication cable may
not be connected properly (or it may be damaged), or you may be
attempting to communicate with the wrong model of analyzer.
The 931 & 932 Properties dialog box is your primary interface
with the analyzer when viewing live data, or when modifying or
viewing the analyzer’s configuration or calibration parameters.
Refer to “Starting the Configurator Software” in Chapter 4 for
information about the functions on the General tab, and all other
tabs.
If at any time you close the 931 & 932 Properties dialog box (by
clicking OK or Cancel), the software will revert to the AMETEK 931
NOTE
& 932 Configurator screen. Return to the 931 & 932 Properties
dialog box by clicking Edit and then Device Properties.
Hazardous Locations:
! If you will be working with the computer powered up for prolonged
WARNING
periods, periodically test the area around the analyzer for hazard-
ous gases, according to your company’s safety policies.
3. After you have completed communicating with the analyzer, exit the
Configurator Software, power down the computer, and disconnect
the serial communication cable from the Explosion-Proof Digital
Communications Port on the Lower Enclosure.
Problem Encountered
The “Wrong Device” message is displayed on the Devices & Log window.
Corrective Action
Check the analyzer markings to ensure it is a Model 931 or Model 932 Analyzer
(the 931 & 932 Configurator Software cannot communicate with other models of
analyzers, such as a Model 900 ADA Analyzer).
Problem Encountered
The “Connected” message is displayed on the Devices & Log window, but the
software is not communicating with the analyzer.
Corrective Action
Select the Live Data check box to establish communication.
Problem Encountered
The “DEMO Online” message is displayed in the Status field on the General
tab. This message will be displayed if the Live Data check box is not selected (but
the Device Enabled check box is).
Corrective Action
Select the Live Data check box to establish communication.
Corrective Action
Observe the Device Enabled and Live Data check boxes on the General tab. One
or both of these check boxes may not be selected. Select both of them to establish
communication.
Check the Serial Communication Cable for proper connections and for damage
such as cuts, nicks, or burn marks. Also, check it to ensure the internal wiring is
still terminated at the connectors.
Problem Encountered
The “DISABLED” message is displayed in the Status field on the General tab.
This message will be displayed if the computer is connected to the analyzer but
is not communicating with it.
Corrective Action
Select the Device Enabled check box, if necessary, to establish communication.
Check the analyzer markings to ensure it is a Model 931 or Model 932 analyzer
(the 931 & 932 Configurator Software cannot communicate with other models of
analyzers).
Problem Encountered
The Device Properties command under the Edit menu is grayed out (access to
the 931 & 932 Properties dialog box is not available). If this occurs, the analyzer
configuration and calibration parameters cannot be viewed or modified. This
problem suggests that the “s93x.dys” file has become corrupt.
Corrective Action
Delete the “s93x.dys” file from the C:\Ametek\93x folder. The software will
restore this file (and correct the problem) the next time it is started.
Problem Encountered
The LEDs for various analyzer operation are not operating properly, as described
in the procedures under “Powering Up the Analyzer.”
Corrective Action
Take appropriate safety precautions and check for proper I/O wiring and power
connections.
If the LEDs are still not operating properly, reset the analyzer. To do this, cycle off
the AC power to the analyzer for a minimum of 15 seconds before cycling it on
again.
Problem Encountered
One or both source lamps are not firing.
Corrective Action
Take appropriate safety precautions and check for proper wiring connections.
If the wiring is okay, check the connections of the lamps in their sockets. Review
the section titled “Source Lamp Replacement” in Chapter 6 to make adjustments
to the lamps. Perform an Auto-Setup if necessary to ensure the lamps are operat-
ing at peak efficiency.
View the Temperature Control dialog box to check the Range. Compare these
values to the Cell Temp RTD Range and/or Optical Bench Temp Range listed on
the Analyzer Programming Parameters sheet included in the “Supplemental
Information” tab of this manual, or the Documentation Package shipped with
the analyzer.
Corrective Action
Take appropriate safety precautions and check the related RTD for a short or an
open circuit. If necessary, replace the RTD with an AMETEK-approved replace-
ment part. Contact AMETEK for assistance.
Problem Encountered
Temperature-controlled zone values do not increase.
Corrective Action
Take appropriate safety precautions and:
• Check the fuses for the temperature-controlled zones to ensure they have
not blown. These fuses are located on the AC Distribution PWB in the
Lower Enclosure.
• Check the Thermo-Switch (Over-Temp) on the Optical Bench Assembly. If
the plunger has popped out, press it to reset it.
• Check for proper wiring terminations for the appropriate temperature-
controlled zone.
Problem Encountered
Alarms remain active after the analyzer has warmed up to normal operating
temperature, as displayed in the HC Scode and MC Scode alarm lists on the
Status tab.
The “f: Probe temperature too low” (if a Heated Sample Probe is used),
“w: Bench temperature out of range”, and “w: Cell temperature out of range”
alarm error messages will typically be displayed upon start-up due to low
temperatures in these temperature-controlled zones.
Corrective Action
If these or other alarms remain active, refer to the alarm conditions and correc-
tive action in Chapter 6.
Before proceeding, test the area around the analyzer for hazardous
! gases and proceed only when the area is found to be safe.
WARNING
1. Isolate the sample gas from the Sample Inlet by closing the isolation
valve on the probe. This will prevent sample gas from flowing
through the sample system.
2. Disconnect the Sample Line from the Sample Inlet and connect a
zero gas line to the Sample Inlet. This is in addition to the zero gas
line already connected to the Zero/Aspirator Drive Gas Inlet on the
analyzer.
Allow zero gas to flow into the Sample Inlet at a pressure approxi-
mately the same as the recommenced sample pressure for your
application to backpurge the entire sample system. Allow the purge
to continue for 5 minutes. Refer to system drawings in the “Supple-
mental Information” tab or the Documentation Package shipped with
the analyzer for sample pressure requirements.
3. Disconnect the vent tubing from the analyzer and plug it with a cap
fitting.
Open the Metering Valve and the Sample Flow Pressure Regulator
(regulator closest to the Sample Shut-Off Ball Valve on the Sample
Inlet line).
a. From the Setup tab of the 931 & 932 Properties dialog box (Figure
3-18), click the Gas Calibration button to view the Gas Calibra-
tion dialog box (Figure 3-19).
b. Enter “2” (zero) in the Flow Control text field. Click OK to accept
the change but do not save the parameter changes to EEPROM (it
is a temporary change only).
Do not set the pressure higher than its upper pressure rating if the
! optional pressure transducer is installed. Refer to the system draw-
CAUTION
ings (in the “Supplemental Information” tab or the Documentation
Package shipped with the analyzer) for details about the pressure
transducer rating.
Figure 3-18.
Setup (Model 932).
Figure 3-19.
Gas Calibration dialog
box (Model 932).
The Sample Pressure Gauge and the Zero Gas Pressure Gauge
must be set at the same pressure to avoid one leaking into the
NOTE
other. Adjust the Zero Gas Pressure Regulator until these pressure
readings are the same.
6. Close the Sample and Vent Shut-Off Ball Valves and allow the sample
system to remain pressurized for 10 minutes. Observe the pressure
gauges during this time. The pressure should not drop at all.
During this time, use a suitable leak detection fluid (such as Snoop®)
on all of the fittings except those that are hot, such as those on the
heated Measuring Cell (if used).
7. If leaks are found, shut off the zero gas flow at the cylinder, bleed
down the sample system from the vent port of the analyzer (by
removing the cap), check and repair all suspect fittings, and repeat
the procedures of testing/repairing/testing until all leaks have been
eliminated.
8. After all leaks have been eliminated, close the pressure regulators and
then remove the cap from the Vent line to allow the sample system
pressure to bleed down (by removing the cap).
9. Reconnect the vent line to the Sample Vent outlet on the analyzer
backpan.
Disconnect the zero gas line from the Sample Inlet and reconnect the
Sample Line to the Sample Inlet.
11. Close the Metering Valve and the Sample Flow Pressure Regulator
(regulator closest to the Sample Shut-Off Ball Valve on the Sample
Inlet line).
12. Slowly open the isolation valve on the probe. Open the Vent Shut-
Off Ball Valve and then slowly open the Sample Shut-Off Ball Valves
to allow sample gas into the sample system.
The Metering Valve (or Aspirator Drive Air Regulator, if used) is used to
control the sample gas flow rate through the system. A good sample gas
flow rate (2.5 L/min (5 SCFH)) will ensure a measurement response time is
approximately 15 seconds. To obtain a response time within this range,
adjust the Metering Valve (or Aspirator Drive Air Regulator, if used) to the
appropriate sample pressure listed in the system drawings.
To check the sample response time, click the Gas Calibration button from
the Setup tab to view the Gas Calibration dialog box. Observe the output
concentration changes. If the sample response time is adequate (first
response after a Zero function is approximately 15 seconds), no further
adjustment is necessary for a Sample Line of less than 10 metres.
The sample response time may vary, depending on Sample Line length.
NOTE
You can also use the rotameter to set the sample gas flow rate
! (arrows on the three-way ball valve handle facing downward).
CAUTION
After the flow rate has been set, switch the Bypass Valve to Bypass
by turning it 180° (arrow on valve handle facing upward).
If the zero flow is set too high, then the zero gas cylinder will need to be
charged more often. It is important to note that in the simplest analyzer
configuration with only one solenoid, if the analyzer is powered down or
has a fault, the analyzer will flow zero gas. A very high or prolonged zero
gas flow will empty a zero gas cylinder fairly quickly when there is a
fault. This is less critical if plant nitrogen or instrument air is being used
as the zero gas.
Keep the log book in a safe location until you need to review it for pat-
terns of diminishing PMT signals or a longer sample response time.
2. While the zero is in progress, observe the PMT Measure and PMT
Reference signals (on the Optical Bench Data tab) for the optical
filters and record the readings for both signals.
The analyzer’s sample response time can be used for two functions. One
function is to help you set the sample gas flow rate (as discussed earlier in
this chapter, under “Setting Sample Response Time”). The other function
is to help you determine if the typical response time is increasing, which
may indicate a plug in the sample system.
2. After the zero is complete, view the Status tab and observe the
output concentration changes. Observe and record the time it takes
the analyzer to display the first reading after the analyzer’s Flow
Control function switches from “zero” to “sample”.
The AMETEK 931 & 932 System 200 Configurator Software program
provides a graphical user interface to configure and control the functions
of the analyzer. Running under Windows 98, ME, NT4.0, 2000, or XP, it
provides powerful features and flexible operating modes.
While working from the Configurator Software you can view real-time
and historical data, and modify configuration and calibration parameters.
The software is also necessary while performing maintenance and while
troubleshooting the analyzer. The majority of these functions are per-
formed from the 931 & 932 Properties dialog box of the software. From
this window, all current operating and calibration parameter settings can
be viewed from the various dialog boxes that can be selected from the
Setup tab. Live data for different aspects of the analyzer can be viewed
from the Status, Analyzer Data, and Optical Bench Data tabs.
Hazardous Locations:
! Before proceeding, test the area around the analyzer for hazardous
WARNING
gases and proceed only when the area is found to be safe.
2. The software will start and the Devices & Log window (Figure 4-5)
will appear in the AMETEK 931 & 932 Configurator window. From
here, select Device Properties from the Edit menu to view the 931 &
932 Properties dialog box (General tab, Figure 4-1).
The 931 & 932 Properties dialog box will be your main interface
when working on the analyzer or viewing live data.
NOTE
Figure 4-1.
General tab
(with Modbus Serial
communication
settings).
The 931 & 932 Properties dialog box is your primary interface with the
analyzer when viewing live data, or when modifying or viewing the
analyzer’s configuration or calibration parameters.
If at any time you close the 931 & 932 Properties dialog box (by
clicking OK or Cancel), the software will revert to the AMETEK 931
NOTE
& 932 Configurator Software screen. Return to the 931 & 932
Properties dialog box by clicking Edit and then Device Properties.
Hazardous Locations:
! Before proceeding, test the area around the analyzer for hazardous
WARNING
gases and proceed only when the area is found to be safe.
After you have completed communicating with the analyzer, exit the
Configurator Software and disconnect the computer from the analyzer.
1. From the 931 & 932 Properties dialog box, click OK or Cancel at the
bottom of the screen.
2. From the AMETEK 931 & 932 Configurator window, select Exit from
the File menu.
PC Communications
Displays information that has been configured from the Communica-
tion Settings dialog box.
Setup
Allows you to choose the communication option required to
establish communication with the analyzer, either Modbus Serial
or Modbus TCP and enter or select the appropriate information.
Detailed descriptions about these communication options are
discussed in the following pages under “Modbus Serial / Modbus
TCP Communications Setup.”
Device
Name
Enter a tag number or tag name.
Description
Enter a description to further define the analyzer, such as a loca-
tion or area where it is installed in the plant.
Live Data
Status
Displays the status of the analyzer (device) you are communicating
with. Refer to “Configurator Software Troubleshooting” in Chapter 3
for information about how to correct the problem.
Before accessing the local service port on the front of the Lower
! Enclosure, test the area around the analyzer for hazardous gases and
WARNING
proceed only when the area is found to be safe.
1. Start the Configurator Software. The 931 & 932 Properties dialog box
(General tab) will appear.
2. Click the Setup button (or click the Live Data check box) to view the
Communication Settings dialog box (Figure 4-2). This dialog box
defaults to Modbus Serial. If you require Modbus TCP, select it. The
screen will change to display appropriate parameters (Figure 4-3).
Figure 4-2.
Communication
Settings dialog box
(Modbus Serial
communication).
Figure 4-3.
Communication
Settings dialog box
(Modbus TCP
communication).
Slave ID
This is the Modicon Modbus® slave address of the analyzer. If
the RS-232 port is selected, the address setting is not appli-
cable (it is grayed out).
5. View the Status tab. Live data should begin updating within 10 sec-
onds.
Standard Toolbar
Custom Mode Toolbar
Data File Toolbar
Main Window
Figure 4-4.
Configurator Software
main window (Trend
mode display active). Status Bar
• Standard Toolbar
Contains icons that are used most often including the main display
modes.
• Status Bar
Located at the bottom of the window, the Status Bar displays system
messages, including information such as if the system is currently
scanning.
The settings for your analyzer have been configured at the factory to
! meet specified customer requirements. Changing the factory-set
CAUTION configuration could cause the analyzer to operate incorrectly.
Figure 4-5.
Devices & Log
window.
1. From the View menu, select Devices & Log or click . A log is
displayed showing all active devices and their status.
3. To remove the display, select View and then Devices & Log again or
click .
1. View and record the current version of the Host Controller and
Microcontroller software to ensure they are an older version than that
being replaced.
From the General tab of the 931 & 932 Properties dialog box, view
and record the current version of the Host Controller (HC ver) and
Microcontroller (MC ver) software, located to the right of PC Commu-
nications on this tab.
2. From the General tab, click the Save Configuration button to view
the Save As dialog box. To store the file, navigate to C:\Ametek\93x
on the computer’s hard drive (AMETEK does not recommend saving
this file to another path).
3. Name the file and click Save. The file will be saved with a “.dev”
extension. This file can be used only by the Configurator Software.
1. Click the Restore Configuration button from the General tab of the
931 & 932 Properties dialog box to view the Open dialog box.
3. View the Devices & Log window. If the file restoration is successful,
the message “This new CFG has been restored successfully!” will be
displayed.
4. After the file restoration is complete, check the analyzer serial number
(on the General tab) to ensure it matches the analyzer serial number
(on the analyzer).
Descriptions for the functions on the General tab can be found earlier
in this chapter, under “Starting the Configurator Software.”
NOTE
The settings for your analyzer have been configured at the factory to
! meet specified customer requirements. Changing the factory-set
CAUTION
configuration could cause the analyzer to operate incorrectly.
Figure 4-7.
Setup tab (Model 932).
For all first-time analyzer setups, set up the analyzer from the 931
& 932 Properties dialog box. From here, you must check the
NOTE
Device Enabled and Live Data text boxes to initiate communica-
tion. Then, you must set up communication parameters by clicking
the Setup button from the General tab.
• Information and/or functions that are grayed out are not applicable
or currently unavailable for the analyzer you are currently working
with.
• After changing any of the parameters or their values from any setup
screen, you must click the OK or Apply button on that screen (or
within the group) to activate the new settings. These new settings will
be sent to the analyzer.
• If you want the parameter changes to be valid only for the current
session, do not click Save CFG to EEPROM. The changes will revert to
the previously saved changes after the system is powered down and
then powered up again, or if the analyzer is reset.
Message box
The message box (to the right of the Modify buttons) displays infor-
mation about the software and analyzer it is currently communicating
with.
When you select the Print to File button from the Setup tab, the Save
As dialog box appears (Figure 4-8). Select the path you wish to save the
file to, give the file a name, and then click Save. The software saves the
current analyzer configuration parameters to an ASCII file. Note that the
file will be saved with a “.dmp” extension.
Figure 4-8.
Save As dialog box
(Print to File button).
When you select the Print button from the Setup tab, the Print dialog box
appears (Figure 4-9). Make sure a printer is connected to the computer or
available through a network before printing these settings. Click OK to
print a hard copy of the current analyzer configuration parameters.
Figure 4-9.
Print dialog box
(Print button).
After changing any of the parameters or their values from any setup
screen, click OK to exit the screen and return to the Setup tab (see Figure
4-7). Parameter changes are evident when the Save CFG to EEPROM
button is active. Click this button to activate the new settings and to
permanently save the new value(s) to the analyzer’s EEPROM.
The Optical Bench dialog box (Figure 4-10), accessed by clicking the
Optical Bench button, allows you to modify Optical Bench control and
data acquisition parameters that can be viewed and modified. To access
this dialog box, click Optical Bench from the Setup tab.
Figure 4-10.
Optical Bench dialog
box (Model 932).
Tzero**
This is the PMT signal transmittance value of each filter, with Zero Gas
in the sample system. It is adjusted automatically whenever the
analyzer is zeroed.
ALC On button
The Automatic Lamp Control automatically turns on at the completion
of an Auto-Setup function, and should be on during normal opera-
tion. However, this function should be turned off while performing a
lamp alignment.
Bench Type**
Defines the Optical Bench orientation and the filter sequence assigned
to each source lamp. This is user-adjustable but should not be
changed.
Processing Filtration
MaxT90 (sec) This is the time (default is 60 seconds)
required for the output signal to reach 90
percent of the final value after a step
change in the concentration. Larger values
reduce the noise level on the output signal
but, at the same time, increase the sample
response time. This is user-adjustable but
should not be changed.
Sigma of Adaptive Filtering
This displays the expected noise level for
the lamps (default is 0.00025). This is user-
adjustable but should not be changed.
Samples per Peak**
For Model 931 Analyzers, this displays the number of times the PMT
signal is measured for each filter.
For Model 932 Analyzers, this displays the number of times the PMT
signal is measured while a filter is completely within the light path.
For Model 932 Analyzers, this displays the time interval between when
the filter becomes completely within the light path and the first PMT
signal measurement. The interval is shown as multiples of 1.6 micro-
seconds (µs).
K Msr**
This is the molar absorptivity of the gas species at the wavelengths
used, for the Measure PMT signal. Model 931 Analyzers only.
K Ref**
This is the molar absorptivity of the gas species at the wavelengths
used, for the Reference PMT signal. Model 931 Analyzers only.
Key Filter**
This displays the position of the key filter, which gives the weakest
signal for a given source lamp current.
Auto-Setup button
Initiates the Automatic-Setup function, which optimizes the PMT
gains and source lamp currents. This function combines automatic
signal adjustment with manual jumper positioning. The Auto-Setup
process takes approximately three minutes. Always perform an Auto-
Setup after any lamps, optical filters, or PMTs have been replaced.
Figure 4-11.
Cell/Miscellaneous
(compensations)
Parameters dialog box
(Model 932).
Cell Length**
Displays the axial length of the Measuring Cell gas space (centimetres
or inches).
If the 100 PSIA pressure transducer is being used, the jumpers (JP301/
JP302) on the Customer I/O board allow you to re-scale the pressure
range. Jumper settings for the different ranges for this pressure
transducer are listed under “100 PSIA Pressure Transducer: Jumper
Settings, Customer I/O Board”.
<Conc>corr. = <Conc>uncorr. / Z
Output Format
Allows you to choose a format to output the Algorithm Setting,
either Concentration, Absorbance, or Transmittance.
For Model 931 Analyzers, this is done by entering ‘1’ in the K Msr
field on the Optical Bench dialog box.
For Model 932 Analyzers, this is done from the Matrix dialog box,
by entering ‘1’ in the top-left text field and each field going
diagonally to the lower-right text field. In addition, all other fields
on the Matrix dialog box must be set to zero (0).
Also, all Algorithm Setting selections on the Cell/Misc Param-
eters dialog box must be disabled (unchecked).
IMPORTANT
Before changing these settings, be sure to record the values and
NOTE
selected options. These settings must be returned to their original
state when changing the output display to its original format.
Figure 4-12.
Temperature Control
dialog box (shown
with unheated
Measuring Cell).
Bench
The measure range (°C) for the Optical Bench temperature-controlled
zone. For more information, refer to “Optical Bench” following these
descriptions.
Cell
The measure range (°C) for the Measuring Cell temperature-controlled
zone. For more information, refer to “Measuring Cell” following these
descriptions.
Probe
The parameters for the optional Heated Acid Gas Probe temperature-
controlled zone. For more information, refer to “Heated Acid Gas
Probe Temperature Control” following these descriptions. The Probe
temperature-controlled zone is optional (AMETEK Heated Acid Gas
Probe required).
Kp (band)**
Displays the proportional term of the PID temperature control algo-
rithm for each temperature-controlled zone.
Ti (sec)**
Displays the integral term of the PID temperature control algorithm
for each temperature-controlled zone (seconds).
Startup (%)
This is the default duty cycle percentages used upon start-up for each
temperature-controlled zone during steady-state operation at the set
point.
Too Cold
This is an alarm temperature setpoint for each temperature-controlled
zone. If the current temperature of the zone falls below its Too Cold
setpoint, it will cause the analyzer to default to an alarm state. For
more information, refer to “Optical Bench,” “Measuring Cell,” or
“Heated Acid Gas Probe” following these descriptions.
Too Hot
This is an alarm temperature setpoint for each temperature-controlled
zone. If the current temperature of the zone exceeds its Too Hot
setpoint, it will cause the analyzer to default to an alarm state. For
more information, refer to “Optical Bench,” “Measuring Cell,” or
“Heated Acid Gas Probe” following these descriptions.
Temperature
Displays the current temperature for each temperature-controlled
zone.
Alarm Indicator
The alarm display box, at the bottom of this dialog box, displays all
current temperature-related alarms. If any alarms are active, refer to
the “Diagnostics and Troubleshooting” section in Chapter 6 for details
about the alarm, and corrective action to take to reset the alarm.
If the analyzer is controlling the Measuring Cell (as is the case with a
heated Measuring Cell), it will generate a Fault alarm (“f: Cell tempera-
ture out of range” will be displayed in the MC Scode alarm list on the
Status tab). If the analyzer is not controlling the Measuring Cell (as is the
case with an unheated Measuring Cell), it will generate a Warning alarm
(“w: Cell temperature out of range” will be displayed in the MC Scode
alarm list on the Status tab).
If the Heated Acid Gas Probe operating temperature falls below its Too
Cold setpoint, the analyzer will generate a “f: Probe temperature too low”
Fault alarm. If its operating temperature exceeds its Too Hot setpoint, the
analyzer will generate a “f: Probe temperature too high” Fault alarm. Both
of these Fault alarms will be displayed in the HC Scode alarm list on the
Status tab when they are set.
The Output Setup dialog box (Figure 4-13) contains Current and Voltage
output assignments. Refer also to the “Customer Connections, I/O Board”
drawing (North American or European) in Chapter 3. To access this dialog
box, click Output from the Setup tab.
There are also four 1000–5000 VDC outputs. The default signal assign-
ment is configured to meet the specifications of the end user. If either the
current or voltage output assignments need to be changed, click the
drop-down arrow beside the output and select a different assignment.
In addition to the usual capability of Zeroing and Spanning the V/I and
Voltage Outputs, you can also set them to a mid-scale level for diagnostic
purposes. The low end of each analog output range can be a non-zero
value.
Figure 4-13.
Output Setup dialog
box, with two streams.
Figure 4-13 shows species under Species Iout and Species Vout with
a two-stream configuration. The “A:” and “B:” prefixes indicate
NOTE
Stream A and Stream B, respectively. If the system is set up for a
single-stream configuration, these prefixes are not shown here.
Channel# (1, 2, 3, 4) lists various parameters for each of the four analog
output channels.
Species Iout is the gas type of interest for the current output.
You can assign a different species for each of the four channels.
Depending on the analyzer, certain options may not be available.
Options include Live H2S(%), Live Sensor 1(%), Live H2(%), Live
Sensor 3(%), T&H H2S(%), T&H Sensor 1(%), T&H H2(%), T&H
Sensor 3(%), and T&H H2S(%). N/A is selected if no output is
assigned to the channel.
Species Vout is the gas type of interest for the voltage output.
You can assign a different species for each of the four channels.
Depending on the analyzer, certain options may not be available.
Options include Live H2S(%), Live Sensor 1(%), Live H2(%), Live
Sensor 3(%), T&H H2S(%), T&H Sensor 1(%), T&H H2(%) and
T&H Sensor 3(%) T&H H2S(%). N/A is selected if no output is
assigned to the channel.
Current Outputs
Low Scale
Enter the current low scale set point corresponding to 4 mA DC,
for each of the analog outputs. If an output is not used, set this
parameter to zero (0).
Zero (mA)
Enter the measured current (mA) in the current output loop when
the output is set to zero scale during the Zero calibration proce-
dure, for each analog output. The actual mA value is displayed for
each corresponding output channel when 4 mA (Zero function) is
activated. This is factory-calibrated.
Span (mA)
Enter the measured full-scale current (mA) in the current output
loop when the output is set to full-scale during the Span calibra-
tion procedure, for each of the analog outputs. The actual mA
value is displayed next to the corresponding output channel when
20 mA (Span function) is activated. This is factory-calibrated.
Result
Displays the real time output value for each output.
Low Scale
Enter the voltage low scale set point corresponding to 1000 mV
DC, for each of the analog outputs. If an output is not used, set
this parameter to zero (0).
Zero (mV)
Enter the measured voltage output when the analog output is set
to zero scale during the Zero calibration procedure for a voltage
output.
Span (mV)
Enter the measured voltage output when the analog output is set
to full-scale during the Span calibration procedure for a voltage
output.
Result
Displays the real time output value for each output.
Normal button
Current and Voltage outputs track the analyzer displayed output. The
status of this function is ON/OFF.
Mid-Scale button
Sets the current outputs to 12 mA and the voltage outputs to 3000
mV for calibration verification. The status of this function is ON/OFF.
Apply button
Saves the changes you made to the Assignment parameters and the
values you entered in any of the Current Outputs and Voltage Out-
puts text boxes.
Refresh button
Reverts the changes you made on this screen back to their original
settings, but only before you click OK to exit this screen.
Hazardous Locations:
! To comply with the requirements of the hazardous location certifica-
WARNING
tion for the sensors, do not operate these sensors in a hazardous
location at a pressure greater than 825 mmHg (1.1 BAR or 16 PSIA).
Figure 4-14.
Sensors dialog box
(Model 932).
Sensor
Displays the names of available, factory-set sensors – Sensor 1, 2, 3
(e.g., H2 for hydrogen).
Span Conc
This is the concentration of the component (PPM or %) in the calibra-
tion gas mixture, for each sensor. The concentration is entered in
decimal notation. The maximum concentration which can be entered
is 999 999. If a calibration gas is not used, this value must be set to
zero (0).
SFactor
The Span (calibration) Factor for the calculated result (component
concentration), for each sensor. The analyzer’s Span Factor is adjusted
automatically at the completion of a span.
ManSpan buttons
Allow the user to manually span the analyzer for each component
being measured by its corresponding sensor (must be enabled under
Enable Sensor Result Cal). The analyzer calibration is automatically
adjusted based on the average reading during the Integration Timer
period after clicking the corresponding ManSpan button.
The Span1 CAL, Span2 CAL, and Zero CAL solenoids are config-
ured (at the factory) on the Alarm Setting and Stream Switch
Parameters dialog box (under Flow Operation).
UV Result Compensation
Display
During normal operation, these message lines display the real-time
status of the species being measured. The species is displayed on the
top line and the concentration value is displayed on the bottom line.
Figure 4-15.
Modbus dialog box.
Address
Displays and allows you to select the Modbus slave address assigned
to the analyzer.
Baud Rate
Displays and allows you to select the baud rate.
Stop Bits
Displays and allows you to select the number of stop bits.
Parity
Displays and allows you to select the parity of the Modbus network.
The field at the bottom of this dialog box (i.e., RTU 9600 N 8) represents
the current settings.
The Matrix dialog box is used only for the Model 932 Multi-Gas
Analyzer. For the Model 931 Single-Gas Analyzer; the Matrix button
NOTE
is grayed out, this function is not available.
The Matrix dialog box (Figure 4-16) displays information that is respon-
sible for partially calculating the concentration output and negating the
effects of cross-talk by interfering gas species. To access this dialog box,
click Matrix from the Setup tab.
Figure 4-16.
Matrix dialog box
(Model 932).
Species**
This designates which gas species are currently configured for each
row (up to six).
Row/Filter**
This displays the value for each species of gas (Row 1–6) and each
column (Filter 1–6) in the calculation matrix.
The Analytical Parameters dialog box is used only for the Model 932
Multi-Gas Analyzer. For the Model 931 Single-Gas Analyzer, the
NOTE
Analytical Parameters button is grayed out; this function is not
available.
Figure 4-17.
Analytical Parameters
dialog box (Model 932).
Linear Coefficients
This is the calibration parameters used for factory calibration. The
displayed results may vary, depending on the application.
Linearization c0–c6**
This displays the factory-set linearization coefficients that are used to
provide polynomial correction in the data analysis.
The Analog Inputs dialog box (Figure 4-18) allows you to configure
various parameters for the analog inputs associated with the analyzer. To
access this dialog box, click Analog Inputs from the Setup tab.
Figure 4-18.
Analog Inputs dialog
box.
Refresh button
Reverts the changes you made on this screen back to their original
settings, but only before you click OK to exit this screen.
The Gas Calibration dialog box (Figure 4-19) contains gas species infor-
mation and sample system control parameters. To access this dialog box,
click Gas Calibration from the Setup tab.
Figure 4-19.
Gas Calibration dialog
box (Model 932).
Result Code
Each of these factory-set Result Codes corresponds to a specific
species. Result Codes are used to define outputs.
Species
Displays the names of available, factory-set gas species/output set-
tings.
Span Conc
This is the concentration of the component (PPM or %) in the calibra-
tion gas mixture. The concentration is entered in decimal notation.
The maximum concentration which can be entered is 999 999. If a
calibration gas is not used, this value must be set to zero (0).
SFactor
The Span (calibration) Factor for the calculated result (component
concentration) for each configured gas species. The analyzer’s Span
Factor is adjusted automatically at the completion of a span.
ZCalDriftSpec
At the end of every Auto-Zero or Manual Zero, the zero drift result is
calculated and displayed (PPM).
Auto CAL
Sample Delay (min) Allows you to enter a time value (in min-
utes) that sets the time delay for the AUTO/
MANUAL relay to continue indicating that a
calibration (Zero) is in progress after
completion of the Zero function. This
function also sets the time delay that the
Normal/Fault relay will wait before indicat-
ing the fault alarm has been cleared. This
allows for a smooth transition from the
Zero/Span Gas back to the sample gas. The
delay can be set from 0–255 minutes. Setting
the delay to zero (0) turns the function off.
Display
During normal operation, these message lines display the real-time
status of the species being measured. The species is displayed on the
top line and the concentration value is displayed on the bottom line.
The Flow Control modes, listed in the Flow Control Legend, vary for
different analyzers. The analyzer will operate in the corresponding
Flow Control mode code that is entered in the Flow Control Mode
text field below the Flow Control Legend. After entering a Flow
Control mode code, click the Apply button (next to the Flow Con-
trol text field) to save the change and set the analyzer into the new
Flow Control mode.
0 = auto
Enter this code to enable the analyzer to automatically determine and
set the mode of operation (either sample or shut-in) by monitoring
the state of the Fault alarm relay contacts. Upon analyzer start-up
and during Fault alarm condition, the analyzer will automatically
shut down the flow.
1 = shut-in/off
Enter this code to manually shut off the flow of sample gas through
the sample system. In this mode, the analyzer is under user control.
2 = zero
Enter this code to manually force the analyzer to continuously zero
the analyzer sample system with zero gas.
3 = span1
Enter this code to manually force the analyzer to continuously
backpurge its sample system with Span Gas1.
4 = span2
Enter this code to manually force the analyzer to continuously
backpurge its sample system with Span Gas2.
5 = stream A
Enter this code to manually force the analyzer to switch to sampling
from Stream A.
6 = stream B
Enter this code to manually force the analyzer to switch to sampling
from Stream B.
7 = sample
Enter this code to manually force the analyzer to continuously
sample.
The Alarm Setting and Stream Switch Parameters dialog box (Figure
4-20) contains gas species information and sample system control param-
eters. To access this dialog box, click Alarm/Stream Switch from the
Setup tab.
Figure 4-20.
Alarm Setting and
Stream Switch
Parameters dialog box
(Model 932).
The Stream Switching settings are made at the factory and should
not be changed.
NOTE
Before proceeding, test the area around the analyzer for hazardous
! gases and proceed only when the area is found to be safe.
WARNING
Do not remove the cover of the Explosion-Proof Digital Communi-
cations Port on the front of the Lower Enclosure and do not power
up the laptop computer if there is a hazardous gas atmosphere
present.
Figure 4-21.
Analyzer Information
dialog box (Model 932).
The Info dialog box (Figure 4-21) displays information about the analyzer
and allows you to reset the analyzer’s Host Controller board. To access
this dialog box, click Analyzer Info from the Setup tab.
Host Version
Displays Host Controller board software version.
Micro Version
Displays Microcontroller board software version.
Serial #
Displays analyzer serial number.
EEPROM Writes
Displays number of times that new configuration data has been
saved to the analyzer’s EEPROM. The maximum allowable number
of saves is 20,000.
Model ID
Displays a code (see Analyzer Legend, below) that the software uses
to define the AMETEK analyzer model the software is currently
communicating with (not user-adjustable). Analyzer models include:
151 = 0x97 = M931 (Model 931 Analyzer)
152 = 0x98 = M932 (Model 932 Analyzer)
Display
During normal operation, these message lines display the real-time
status of the species being measured. The species is displayed on the
top line and the concentration value is displayed on the bottom line.
Refresh button
Reverts any values that you have modified back to their original
settings, but only before you select an Update button or OK. When
you click this button, any values that you have changed but do not
wish to set permanently will revert to the settings that were last saved
to the system’s EEPROM.
HC Reset button
Resets the analyzer’s Host Controller board.
Figure 4-22.
Status tab (Model 932).
System Status Messages along the top portion of this screen display
information about the System 200 Configurator Software and the
analyzer. These include:
Analyzer Status
Displays the communication status between the System 200
Configurator Software and the analyzer. Messages that can appear
here include: Online, OFFLINE, DISABLED, Wrong Device (if the
analyzer cannot be recognized), or DEMO.
Host Runtime
This is the amount of time the Host Controller board has been
performing controls since the analyzer was last powered up or
reset.
Micro Runtime
This is the amount of time the Microcontroller board has been
performing controls since the analyzer was last powered up or
reset.
SN#
Analyzer’s serial number. This information will be displayed only
if the System 200 Configurator Software is communicating with
the analyzer.
Calibration/Solenoid Status
Located in the upper-right corner of this dialog box, this displays
the current status of the calibration functions and the current
status of the solenoids.
Calibration Status
The status of the calibration functions is displayed on the top
line of this display. Depending on which calibration function is
in progress, one of the following messages is displayed: Auto-
CAL, Auto-Zero, Auto-span1, Auto-span2 (Model 932 only),
Manual-zero, Manual-span1, Manual-span2 (Model 932 only).
Solenoid Status
The status of the solenoids is displayed on the second line of
this display. Solenoids Z, B, and A are always displayed, but
their ON/OFF status will change, depending on their opera-
tion.
HC Scode
Displays active alarm conditions that originate from the Host Control-
ler board. Each alarm condition is preceded by its status number,
followed by the type of alarm (Fault or Warning), and the alarm
description. The sum of their respective primary codes is listed in the
‘f’ault and ‘w’arning fields next to HC Scode.
Scode History
These readings are history buffers for the Host Controller (HC Scode)
and Microcontroller (MC Scode) boards, for the last five measuring
cycles. Each history buffer displays a status code (if any alarms have
been set), which is a sum of the primary error codes, where ‘1’ is the
most recent status code and ‘5’ is the oldest. To resolve a non-primary
status code into its primary error codes, click on the history buffer
fault or warning status code. The alarms that make up the status code
will be displayed under the HC Scode or MC Scode alarm lists.
Display
During normal operation, these message lines display the real-time
status of the species being measured. The species is displayed on the
top line and the concentration value is displayed on the bottom line.
Calculated Results
Located to the right of the HC Scode and MC Scode History buffers,
the existing results and values are displayed. The value for each
result will be displayed and updated at one-second intervals.
Conc Displays the current concentration for each compo-
nent displayed to the left of this column.
Remote ACal
The current status of the remote, dry (potential free) contact, either
OFF (closed) or ON (open). When start is displayed, the remote auto-
calibration function has been initiated from a remote control room,
such as a Data Acquisition System (DAS).
Fault Relay
When the analyzer is operating in its normal Run state, this relay is
energized. This relay is de-energized if a fault alarm condition exists
within the analyzer’s built-in diagnostic system (the analyzer requires
service).
If the relay is configured to latch (default) it will not clear unless the
user manually resets the alarm; if not, it will reset automatically upon
correction of the fault alarm.
Figure 4-23.
Analyzer Data tab
(Model 932).
Results
Displays the real-time calculated result of each species of gas being
measured (one for Model 931 Analyzers; up to six for Model 932
Analyzers). The results, which will vary from analyzer to analyzer, are
updated at one second intervals. The results for the analyzer you are
communicating with will be displayed on this screen.
Temperature Zones
Displays the real time Temperature (°C) and Duty Cycle (%) of each
temperature-controlled zone:
Countdown timers
Scaled/Input Name
Displays the scaled, real-time Measuring Cell pressure
reading.
For Model 931 Analyzers, only Filter 1 and Filter 2 information will
be displayed; for Model 932 Analyzers, information for all six filters
NOTE
will be displayed.
Figure 4-24.
Optical Bench Data
tab (Model 932).
Absorbance
Displays the real time absorbance values for each filter location.
Tzero
Displays the real time PMT signal transmittance value of each filter,
with Zero Gas in the sample system. It is adjusted automatically
whenever the analyzer is zeroed.
Before you can scan data, you must create a mode based on one of the
four predefined mode types, or create a custom mode that is made up of a
combination of the predefined modes (see “Display Modes” in this sec-
tion). For each mode you create or add to custom mode, you must assign
inputs to channels. Assigning inputs is described under “Defining Scan
Settings” in the following pages.
1. From the File menu, select New. The Custom Mode toolbar is acti-
vated, allowing you to use the toolbar’s icons to create a custom setup.
• Track-and-hold concentration
• Analog inputs
2. Click Save Configuration from the File menu to save the file. The
Custom Mode file will be saved in the same directory in which the
Configurator Software has been installed.
Figure 4-25.
Scan tab (Trend mode
dialog box).
1. Right-click on the mode window and then click Scan from the
context menu to view the Trend Properties dialog box. The
Scan tab will be active.
For Trend and Tabular modes, you must first select a Channel
before you can define scan settings.
NOTE
3. Next, click Device under Data Source to select the device from
which to scan data. If only one device (analyzer) is being used,
only Dev1 will be displayed. When Device is selected, the Input
Channel list box is also displayed and enabled. From this list,
select a signal to assign to the channel.
5. Data can also be linked from one window into another. This can
save scanning time since the system only needs to scan the data
one time. To link data from one window into another, use the Link
from function. To do this:
b. Select the display window that will receive the linked data
from another data source. The yellow border around the
window indicates it is selected.
c. Right-click over the display window and then click Scan from
the context menu that appears. The Link from function will
become enabled.
To start a scan, click or select Scan from the Control menu. The
message “Scanning” and the scanning progress will appear in the Status
Bar at the bottom of the window (if it has been enabled).
If you have scheduled scans at timed intervals and the system has fin-
ished a scan and is waiting to start the next scan, the time remaining
before the start of the next scan is displayed.
“Idle” is displayed in the Status Bar when the system is not scanning.
Stopping a Scan
Click or select Scan from the Control menu to toggle off the scan.
Scheduling Scans
Figure 4-26.
Schedule scans dialog
box.
1. Select the display window or windows for which data will be saved.
• Hold down the Shift key and click on each window to be in-
cluded when selecting multiple display windows.
2. From the File menu, select Save Data or click on the Data File
Toolbar to view the Save As dialog box (Figure 4-27). From here, you
can specify data file attributes.
Figure 4-27.
Save As dialog box.
4. To stop saving data, select Stop Saving from the File menu or click
1. From the File menu, select Open Data or click on the Data File
Toolbar.
3. Select Open. The data file appears on the screen. From here it can
either be viewed or printed.
1. From the File menu, select Convert Data File and select one or more
data files.
2. Click Convert to convert and save the file with a CSV extension.
Printing Data
The Configurator Software will print scanning information in a format
similar to how it is displayed on the screen using the display parameters
created for that mode. You can view a preview of what will be printed by
selecting Print Preview from the File menu.
In Custom mode, the system prints the currently selected display win-
dow. If you do not select a display window, the system prints information
for each display window using the required number of pages.
To print data:
4. If you select Print from the File menu, a standard Windows Print
dialog box appears. Set the additional print information.
Figure 5-1.
Gas Calibration dialog
box (Model 932).
Calibration | 5-1
To perform a span gas calibration:
2. Once the Auto Zero process is completed, record the zero gas pres-
sure setting and then turn off the zero gas.
PROCESS INSTRUMENTS
11 10
1 2 3 4
SOLENOID BLOCK DETAIL
Figure 5-2.
Span gas and zero gas
inlet locations.
4. Adjust the Cal Gas Regulator until the Cal Gas Pressure Gauge on the
analyzer backpan is at the recommended pressure listed on the
system drawings.
5. In the Flow Control text field, enter “2” (zero) and then click Apply
(next to the Flow Control text field). Click OK to return to the Setup
tab but do not click the Save CFG to EEPROM button (it is a tempo-
rary change only).
6. Adjust the Span Gas Regulator until the Span Gas Pressure Gauge on
the analyzer backpan is at the recommended pressure listed on the
system drawings.
8. View the Analyzer Data tab (Figure 5-3) and observe the concentra-
tion readings.
Once the span concentration readings have stabilized, return to the Gas
Calibration dialog box and click the ManSpan button below the gas
species being calibrated. To check the time remaining for the span
function, observe the Time to end of cal(sec) timer under Countdown
timers on the Analyzer Data tab. You can also verify that the span
calibration is in progress by viewing Calibration/Solenoid Status field
in the upper-right corner of the Status tab. It should display “Manual-
span1” (can also display “Manual-span2” for Model 932 Analyzers
only).
At the completion of the span, confirm that the span factor reading
(SFactor, on the Gas Calibration dialog box) for the species is up-
dated. The default value for span factor is 1.0.
Calibration | 5-3
Figure 5-3.
Analyzer Data tab
(Model 932).
9. Bleed down the calibration gas and disconnect it from the span gas
connection point.
10. Replace and tighten the Swagelok cap on the span gas connection
point.
11. Turn on the zero gas and adjust the pressure to the level it was at
before performing procedure. Confirm that the pressure is at the
recommended pressure listed on the system drawings (located in the
“Supplemental Information” tab of this manual, and in the Docu-
mentation Package shipped with the analyzer).
12. Return the analyzer Flow Control mode to Analyzer (auto) Control
by entering “0” (auto) in the Flow Control text field, and then click-
ing Apply (next to the Flow Control text field). Click OK to return to
the Setup tab and then click the Save CFG to EEPROM button to
permanently save the changes to the analyzer’s EEPROM.
If the Flow Control mode (Gas Calibration dialog box, Figure 5-1)
has been in a mode other than auto (0) for longer than the AZero
NOTE Interval (hr) duration, and you attempt to return it to Auto (by
entering 0 and then clicking the Apply button next to the Flow
Control text field), the mode will not switch until after the Auto-
Zero Timer (A-Z 0) duration, under Auto Calibration Timers,
elapses.
Current Calibration
The current outputs are calibrated at the factory. If a current output
module is replaced or added, calibrate that output. Calibration of a
current output is performed by entering the measured low-scale (zero)
and full-scale (span) signals for each output (see Figure 5-4). These values
are used to offset the output to the correct values.
Figure 5-4.
Output Setup dialog
box, with two streams
(Model 932).
a. Click the Zero Cal button to change its status to ON. The status of
Normal, Mid-Scale, and Span Cal should be OFF.
c. Repeat this step for Output 2 (Pins 3 and 4), Output 3 (Pins 5 and
6), and Output 4 (Pins 7 and 8), recording the value for each out-
put in the corresponding field.
Calibration | 5-5
2. Measure the Span (full-scale) signal of each output:
a. Click the Span Cal button to change its status to ON. The status of
Normal, Mid-Scale, and Zero Cal should be OFF.
c. Repeat this step for Output 2 (Pins 3 and 4), Output 3 (Pins 5 and
6), and Output 4 (Pins 7 and 8), recording the value for each out-
put in the corresponding field.
3. Click the Normal button to change its status to ON, which activates
the outputs. The outputs will now track the concentration readings.
4. Click OK to return to the Setup tab and then click the Save CFG to
EEPROM button to permanently save the changes to the analyzer’s
EEPROM.
a. Click the Zero Cal button to change its status to ON. The status of
Normal, Mid-Scale, and Span Cal should be OFF.
c. Repeat this step for Output 2 (Pins 3 and 4), Output 3 (Pins 5 and
6), and Output 4 (Pins 7 and 8), recording the value for each out-
put in the corresponding field.
a. Click the Span Cal button to change its status to ON. The status of
Normal, Mid-Scale, and Zero Cal should be OFF.
Calibration | 5-7
c. Repeat this step for Output 2 (Pins 3 and 4), Output 3 (Pins 5 and
6), and Output 4 (Pins 7 and 8), recording the value for each out-
put in the corresponding field.
3. Click the Normal button to change its status to ON, which activates
the outputs. The outputs will now track the concentration readings.
4. Click OK to return to the Setup tab and then click the Save CFG to
EEPROM button to permanently save the changes to the analyzer’s
EEPROM.
Safety Considerations
To prevent an explosion, test the area around the analyzer for hazard-
! ous gases and proceed with maintenance only when the area is found
WARNING
to be safe (nonhazardous).
Clean the Measuring Cell and the rest of the analyzer’s sample
system. This schedule is a minimum requirement. If other
conditions are found to be present, as outlined under the section
titled “Disassembling and Cleaning the Measuring Cell,” cleaning
may be required more frequently.
If the sample response time is longer than normal, it is most likely due to
plugging somewhere in the sample system. The next step is to locate and
remove the plug. The portion of the sample system with the highest risk
of plugging is the sample inlet and the section of the sample inlet most
prone to plugging is the Sample Probe Valve. One way to confirm that a
plug is in the sample inlet is to cap the Vent Line (or close the Vent
Valve’s isolation valve if available) and Zero the analyzer.
After confirming that the plug is somewhere in the sample inlet, take
all appropriate safety precautions and dismantle the sample system,
NOTE starting at the Sample Probe.
Plugs in the sample outlet (or vent side) of the sample system are rare. If
an aspirator is used to transport the sample gas (i.e., part of the Heated
Acid Gas Probe, if used), a plug in the vent system can cause the aspira-
tor drive air to flow back into the Measuring Cell and, essentially, simulate
a Zero Gas Flow or Backpurge condition.
To do this, click Gas Calibration from the Setup tab to view the Gas
Calibration dialog box. Enter “2” (zero) in the Flow Control text field.
Click Apply (next to Flow Control) to set the analyzer into zero Flow
Control mode and then click OK. Do not save this parameter change
to EEPROM (it is a temporary change only).
3. Close the isolation valve on the Sample Line or Sample Probe (if
used). Observe the flow on the rotameter. If the flow reduces to a
stop, a large plug has occurred somewhere in the sample system.
Open the isolation valve on the Sample Line or Sample Probe (if
used).
4. Close the isolation valve on the Vent Line or Sample Probe (if used),
or temporarily install a cap.
Observe the flow on the Zero Gas rotameter. If the flow reduces to a
stop, a large plug has occurred somewhere in the sample system.
6. Return the isolation valve on the Vent Line or Sample Probe (if
available) to the open position (or remove the cap and reconnect the
Vent Pipe to the Vent port on the Measuring Cell). Both isolation
valves should be open.
To do this, click the Gas Calibration button from the Setup tab to
view the Gas Calibration dialog box. Enter “0” (auto) in the Flow
Control text field. Click Apply (next to Flow Control) to set the
analyzer into auto Flow Control mode and then click OK.
8. Check and record the sample response time to use later as a refer-
ence while testing response time.
Plugs in the sample system are less common in low dewpoint tempera-
ture applications. If a plug occurs on a system with an automatic shut-off
valve installed in the sample system, the plug may be caused by a faulty
valve operation.
The o-rings in the Measuring Cell should be replaced every year to avoid
a leak in the sample system. While the Measuring Cell is disassembled,
clean its chambers and parts, including the windows, and replace the
o-rings.
Refer to Figure 6-1 and Figure 6-2 (or the Measuring Cell drawing in the
“Supplemental Information” tab of this manual) for assistance in locating
and identifying the parts that make up the Measuring Cell and its connec-
tion point on the Optical bench.
• If the analyzer responds slowly to zero gas when the sample system
is free of restrictions, or
LOOSE CLAMPED
Figure 6-1.
Upper Enclosure layout,
Cell Window Housing.
Before proceeding, test the area around the analyzer for hazardous
! gases and proceed only when the area is found to be safe.
WARNING
Do not remove the cover of the Explosion-Proof Digital Communica-
tions Port on the front of the Lower Enclosure and do not power up
the laptop computer if there is a hazardous gas atmosphere present.
From the Setup tab click Gas Calibration to view the Gas Calibra-
tion dialog box and enter “2” (zero) in the Flow Control text field.
Click the Apply button (next to Flow Control) to set the analyzer
into the zero Flow Control mode and then click OK. Do not save this
parameter change to EEPROM (it is a temporary change only).
Close the isolation valve on the Sample Line or Sample Probe (if
used). Allow the analyzer sample system to Zero for 2–3 minutes.
3. Return to the Gas Calibration dialog box and enter “1” (shut-in/off)
in the Flow Control text field. Click Apply (next to Flow Control)
and then click OK.
If the sample system does not have an isolation valve, Zero gas
may continue to flow. Shut off the Zero gas.
NOTE
Grasp the edge of the Window and carefully remove it. Remove the
two o-rings.
9. Inspect the interior of the Measuring Cell for foreign material. If you
need to clean the interior, use a nonabrasive detergent and water
solution, isopropanol, or reagent-grade acetone followed by a rinse
with pharmaceutical-grade distilled water.
1. Hold the Measuring Cell vertically with the Reflector Block end up
and install two new o-rings and the Window.
2. Place the Reflector Block on the Measuring Cell, with the “peak-line”
of the Reflector Block at 90 degrees to a line through the fittings.
Ensure that the Measuring Cell Window does not slide out of posi-
tion when the Measuring Cell is initially placed against the Reflector
Block.
Using the three screws removed earlier, secure the Reflector Block to
the Measuring Cell. Tighten the screws evenly.
3. Invert the Measuring Cell and install a new o-ring and the window.
Before proceeding, test the area around the analyzer for hazardous
! gases and proceed only when the area is found to be safe.
WARNING
Do not power up the analyzer or computer if there is a hazardous
gas atmosphere present.
8. Open the valves isolating the sample system from the process stream
and/or remove the cap from the Vent Line outlet.
10. Click Optical Bench from the Setup tab and perform an Auto-Setup
function by clicking Auto-Setup.
12. When the zero calibration is complete, return the Flow Control
setting in the Gas Calibration dialog box to “auto” by entering “0” in
the Flow Control text field. Click OK, and then click Apply (next to
Flow Control) to set the analyzer into auto Flow Control mode and
then click OK. Save this parameter change to EEPROM by clicking
the Save CFG to EEPROM button on the Setup tab.
The typical life span of the source lamps is approximately 9–12 months of
continuous operation. However, a source lamp can exhibit signs of degra-
dation after 5–6 months.
Figure 6-3.
Optical Bench Data
tab (Model 932).
Figure 6-4.
HC Scode and
MC Scode alarm lists
on Status tab
(Model 932).
• The cathode of the source lamp must be centered on the optical axis
of the beam splitter for optimal operation of the analyzer. The cath-
ode and the outer envelope of the lamp are not necessarily concen-
tric or consistent from lamp to lamp.
Refer to Figure 6-5 for assistance in locating and identifying the hard-
ware components encountered while replacing the source lamps.
Before proceeding, test the area around the analyzer for hazardous
! gases and proceed only when the area is found to be safe.
WARNING
2. After the area has been tested for hazardous gases (and found to be
safe), remove the 24 M10 screws from the EEx d Upper Enclosure
door and open it.
4. Loosen the Lamp Retaining Screw on the bottom of the lamp socket
and rotate the Lamp Adjustment Plate 90 degrees. Remove the Lamp
Socket Assembly.
Do not rotate or pull on the base of the source lamp because this
! may cause it to separate from the glass envelope. Grasp the glass
CAUTION
envelope when rotating or pulling the source lamp.
6. Install the new lamps. Make sure that the narrow end of each lamp is
pushed completely into the holder. Do not touch the (flat) window at
the end of the lamp.
Before proceeding, test the area around the analyzer for hazardous
gases and proceed only when the area is found to be safe.
9. After the area has been tested for hazardous gases (and found to be
safe), restore power to the analyzer. Allow the analyzer to warm up
before continuing.
a. From the Status tab, click the Conc. Alarm Soft Enable On/Off
button to temporarily disable the Concentration Alarms.
b. From the Gas Calibration dialog box (Figure 6-6), enter “2” (zero)
in the Flow Control text field. Click Apply (next to Flow Con-
trol) to set the analyzer into zero Flow Control mode and then
click OK. Do not save this parameter change to EEPROM (it is a
temporary change only). Allow the zero gas to flow for 2–3
minutes.
Figure 6-6.
Gas Calibration dialog
box (Model 932).
Figure 6-7.
Optical Bench dialog
box (Model 932).
12. Click the Open Configuration icon on the Standard Toolbar and
double-click the “LampSetup.dyc” file to open it.
Figure 6-8.
Trend graph
(LampSetup.dyc file),
Model 931 Analyzers.
a. Right-click on the Trend graph window and then click Scan from
the context menu to view the Trend Properties dialog box (Figure
6-9).
Figure 6-9.
Trend Properties dialog
box (as shown for Model
931 Analyzers).
c. Next, click Device under Data Source to enable the Input Chan-
nel list box. From this list, select “PMT Ref1” to assign this signal
to the first channel.
d. Repeat Step b and Step c to assign “PMT Ref 2”, “PMT Msr1”, and
“PMT Msr2” from the Input Channel list to subsequent Channels.
Figure 6-10.
Trend graphs – Measure
and Reference PMT
(LampSetup.dyc file),
Model 932 Analyzers.
c. Next, click Device under Data Source to enable the Input Chan-
nel list box. From this list, select “PMT Ref1” to assign this signal
to the first channel.
d. Repeat Step b and Step c to assign “PMT Msr2”, “PMT Msr 3”,
“PMT Msr4”, “PMT Msr5”, and “PMT Msr6” from the Input
Channel list to subsequent Channels.
h. Next, click Device under Data Source to enable the Input Chan-
nel list box. From this list, select “PMT Msr1” to assign this signal
to the first channel.
i. Repeat Step g and Step h to assign “PMT Ref 2”, “PMT Ref 3”,
“PMT Ref 4”, “PMT Ref 5”, and “PMT Ref 6” from the Input
Channel list to subsequent Channels.
To change the Y-axis voltage scale, move the mouse pointer to the
graph and right-click. Click Display and change the Upper limit and
NOTE
Lower limit.
14. Align the Source Lamps (described under “Source Lamp Alignment,”
later in this chapter) to ensure optimal operation and then perform
another Auto-Setup. Return to these procedures after the Auto-
Setup is complete.
15. Return to the Status tab and click the Conc. Alarm Soft Enable On/
Off button to enable the Concentration Alarms and then click OK.
Return to the Gas Calibration dialog box and change the Flow
Control setting back to “auto” by entering “0” in the Flow Control
text field. Click the Apply button (next to Flow Control) to set the
analyzer into auto Flow Control mode and then click OK.
Close the cover of the Digital Communications Port and tighten the
hex screw to secure it in place.
17. Using a soft, nonabrasive cloth, gently clean the mating areas
(flamepath) of the Upper Enclosure and its door and replace the 24
M10 screws. Tighten the M10 screws to 9.0 Nm, ±1.0 Nm (80 in.-lb,
±9 in.-lb).
Auto-Setup
PMT Level and PMT Balance are adjusted only during the Auto-Setup
sequence. The range of operation for both these parameters is 0–10000 mV.
After a successful Auto-Setup, the PMT Level should be between
4500 mV and 7500 mV, while the PMT Balance should be between
1000 mV and 9000 mV. The analyzer will function with values outside
these ranges, but it is a sign that a change has likely occurred in the
transmission of light within the Optical Bench.
At the end of every Auto-Setup, always check the PMT Level and PMT
Balance. Take note of severe changes in PMT Level and PMT Balance
because they are as important as the absolute value.
The PMT Level signal adjusts the gain of both PMTs simultaneously in
the same direction. PMT Balance adjusts the gain of the two PMTs in
opposite directions by controlling a virtual ground circuit on the Optical
Bench Service Electronics board.
If the PMT signals are outside their normal range of operation after the
Auto-Setup function is complete, check for the following:
• Was the Auto-Setup done with zero gas flowing through the sample
system?
• Are the PMTs in their original sockets (i.e., is the higher gain PMT in
the Measure path)?
• Do the source lamps need replacement? If not, are the source lamps
aligned?
• Have there been any leaks from the Measuring Cell? If so, check the
condition of the mirrors, windows, and o-rings in the Measuring Cell
(including Reflector Block). Contact AMETEK for procedures not
discussed in this manual.
If during the Auto-Setup the warning error message “w: Bench auto-
setup problem” is displayed (check the MC Scode alarm list on the Status
tab), it may indicate a problem with the PMT Level or PMT Balance signal.
Refer to “Microcontroller Board Error Conditions and Corrective Action”
later in this chapter for corrective action to take to reset this alarm. Restart
the Auto-Setup function after this alarm has been corrected.
To perform an Auto-Setup:
1. From the Gas Calibration dialog box, enter “2” (zero) in the Flow
Control text field. Click Apply (next to Flow Control) to set the
analyzer into zero Flow Control mode and then click OK. Do not
save this parameter change to EEPROM (it is a temporary change
only). Allow the zero gas to flow for 2–3 minutes.
2. From the Status tab, click the Conc. Alarm Soft Enable On/Off
button to temporarily disable the Concentration Alarms and then
click OK.
3. From the Optical Bench dialog box, check and record the PMT Level
and PMT Balance values.
Observe the top line of the Display readout. The following messages
are displayed to indicate which step of the Auto-Setup process is
being performed:
6. From the Optical Bench dialog box, check the PMT Level and PMT
Balance values again. The PMT Level should be between
4500 mV and 7500 mV, while the PMT Balance should be between
1000 mV and 9000 mV.
Cancel Auto-Setup
There are six jumper positions (four in P300 and two in P301) that can
accommodate one jumper each. At least one jumper is required in P300
and P301 at all times. Adding jumpers increases the supply voltage avail-
able to the Reference path PMT. This increases the PMT Balance and, to a
small degree, decreases the PMT Level. The opposite occurs if jumpers
are removed.
Figure 6-11 describes the effect on the PMT Balance in relation to different
jumper positions.
2. After an acceptable PMT Balance and PMT Level has been achieved,
initiate a manual zero.
To do this, click Gas Calibration from the Setup tab to view the Gas
Calibration dialog box and enter “2” (zero) in the Flow Control text
field.
3. Return to the Gas Calibration dialog box and change the Flow
Control setting back to “auto” by entering “0” in the Flow Control
text field (to return the analyzer to Analyzer Control Mode). Click
Apply (next to Flow Control) to set the analyzer into auto Flow
Control mode and then click OK. Click the Save CFG to EEPROM
button on the Setup tab to save this parameter change to EEPROM.
4. From the Status tab, click the Conc. Alarm Soft Enable On/Off
button to enable the Concentration Alarms and then click OK.
3. Loosen the Locking Screw located between the two source lamps.
4. To obtain the maximum signal from the Measure PMT voltage, adjust
the Lamp Adjusting Screw for the lamp that is being adjusted.
6. Tighten the Locking Screw located between the two source lamps.
8. Tighten the Lamp Retaining Screw to secure the lamp socket. Do not
overtighten; the spring should not be collapsed.
To do this, click Gas Calibration from the Setup tab to view the Gas
Calibration dialog box and enter “2” (zero) in the Flow Control text
field.
Click the ManZero button to initiate the manual zero calibration and
then click OK.
To check the time remaining for the zero function, observe the Time
to end of cal(sec) timer under Countdown timers on the Analyzer
Data tab. You can also verify that the zero calibration is in progress
by viewing the Calibration/Solenoid Status field in the upper-right
corner of the Status tab. It should display “Manual-zero”.
12. When the zero calibration is complete return to the Status tab, click
the Conc. Alarm Soft Enable On/Off button to enable the Concentra-
tion Alarms, and then click OK.
Return to the Gas Calibration dialog box and change the Flow
Control setting back to “auto” by entering “0” in the Flow Control
text field (to return the analyzer to Analyzer Control Mode). Click
Apply (next to Flow Control) to set the analyzer into auto Flow
Control mode and then click OK. Click the Save CFG to EEPROM
button on the Setup tab to save this parameter change to EEPROM.
The lamp replacement and alignment process is complete and the ana-
lyzer is ready to operate.
The EEPROM must be replaced when the Host Controller “w: EEPROM
full” warning alarm error is displayed by the Configurator Software in the
HC Scode alarm list on the Status tab (see Figure 6-13).
Before proceeding, test the area around the analyzer for hazardous
! gases and proceed only when the area is found to be safe.
WARNING
Do not remove the cover of the Explosion-Proof Digital Communica-
tions Port on the front of the Lower Enclosure and do not power up
the computer if there is a hazardous gas atmosphere present.
2. From the General tab, click Save Configuration to save the current
analyzer configuration parameter values to a file on the computer.
Record the location and name of this file; it will be necessary to use
this file to restore the analyzer configuration parameter values after
replacing the EEPROM.
4. Exit the Configurator Software. Power down the analyzer and the
computer. Remove the 24 M10 screws from the Lower Enclosure
door and open it.
Close the Lower Enclosure door and replace the 24 M10 screws.
Tighten the M10 screws to 9.0 Nm, ±1.0 Nm (80 in.-lb, ±9 in.-lb).
Before proceeding, test the area around the analyzer for hazardous
! gases and proceed only when the area is found to be safe.
WARNING
Do not power up the analyzer or laptop computer if there is a hazard-
ous gas atmosphere present.
9. From the General tab, click Restore Configuration. Locate and open
the file saved earlier. Refer to “Restoring Analyzer Configuration
Parameters” in Chapter 4 for more information, and for information
about checking and re-entering the analyzer serial number (if neces-
sary).
The Model 931 and Model 932 UV Analyzers are designed with
flamepaths that will prevent flame propagation from within the
analyzer’s EEx d Electronics Enclosures to the outside, should an internal
explosion occur.
The flamepaths on the Model 931 and Model 932 Analyzers consist of:
If at any time the enclosure door is open and service personnel are not
working on the EEx d Upper/Lower Enclosure, close the door and
secure it with at least one screw. This will reduce the risk of inadvert-
ently scratching or damaging the flamepath.
Before closing the door, gently clean the flamepath with a soft, non-
abrasive cloth and make sure the surfaces are free of debris.
Before proceeding, test the area around the analyzer for hazardous
! gases and proceed only when the area is found to be safe.
WARNING
Do not remove the cover of the Digital Communications Port on the
front of the Lower Enclosure and do not power up the laptop computer
if there is a hazardous gas atmosphere present.
The Model 931 and Model 932 Analyzers have a built-in error detection
system that continuously monitors the operation of key parameters. An
error can be detected by either the Host Controller or the Microcontroller
and can be of two types: Fault and Warning. An error is annunciated by
the Fault Status relay or the Warning Status relay. If the analyzer is com-
municating with the Configurator Software, the particular condition that
causes an error and the type of error are shown in the HC Scode and MC
Scode alarm lists on the Status tab (Figure 6-13).
Figure 6-13.
Status tab (Model 932).
The Warning Status relay returns to the normal energized state automati-
cally when the offending error condition is corrected and/or the analyzer
completes one of the above operations.
A Fault indicates that a failure has been detected and the analyzer data is
invalid. Under the Fault situation, the analyzer de-energizes the Fault
Status relay and switches the sample system into shut-in/off mode, if the
Flow Control setting is “auto” (set to “0”). If the Fault Status relay is
configured as non-latching (Enable latching of fail alarm on the Alarm
Setting and Stream Switch Parameters dialog box is unchecked), the relay
returns to its normal energized state and the sample system is switched
back to sampling state when the offending error condition is corrected.
However, if the Fault Status relay is configured as latching (Enable latch-
ing of fail alarm on the Alarm Setting and Stream Switch Parameters
dialog box is checked), the operator must unlatch the relay in one of the
following ways:
Figure 6-14.
Host Controller error
conditions/diagnostics.
Figure 6-14.
Host Controller error
conditions/diagnostics.
Figure 6-14.
Host Controller error
conditions/diagnostics.
Figure 6-14.
Host Controller error
conditions/diagnostics.
Figure 6-14.
Host Controller error
conditions/diagnostics.
Figure 6-14.
Host Controller error
conditions/diagnostics.
Figure 6-14.
Host Controller error
conditions/diagnostics.
Figure 6-14.
Host Controller error
conditions/diagnostics.
Figure 6-17.
Microcontroller error
conditions/diagnostics.
Figure 6-17.
Microcontroller error
conditions/diagnostics.
Figure 6-17.
Microcontroller error
conditions/diagnostics.
Figure 6-17.
Microcontroller error
conditions/diagnostics.
Analyzer Reset
Should a situation arise where the Host Controller board or the
Microcontroller board must be reset, take appropriate safety precautions
and cycle the AC power to the analyzer off for a minimum of 15 seconds
before cycling it on again.
If this doesn’t correct the error, view the Info dialog box and click the HC
Reset button.
If this still doesn’t correct the error, take appropriate safety precautions,
open the Electronics Enclosure covers, and:
• Reset the Host Controller board (press SW100, Lower Enclosure – see
Figure 6-12).
The Service & Parts chapter discusses what to do if you need technical
support from AMETEK, or if you are returning parts for service. This
chapter also lists the recommended spare parts to have on hand to ensure
all consumable and replaceable parts are available for service according to
the “Preventive Maintenance Schedule” in Chapter 6.
Technical Support
AMETEK Western Research is committed to providing you the best
technical support in the industry. If you need service or application
assistance, contact your local or nearest AMETEK Service Centre or the
AMETEK factory AMETEK at (403) 235-8400 or 1-800-661-9198, or contact
your local AMETEK Western Research representative.
• AMETEK part number for the specific component you are enquiring
about, if known.
PROCESS INSTRUMENTS
Canada: 1-800-661-9198 U.S.A.: 1-800-537-6044
Because the Model 931 and Model 932 Analyzers are designed to accom-
modate a wide variety of applications (and therefore can accommodate a
variety of sample system components), there is no typical spare parts list
available for the Optical Bench and sample system.
Before ordering spare parts, refer to the Custom Spare Parts list (or
Measuring Cell assembly drawings) in the “Supplemental Informa-
tion” tab of this manual or the Documentation Package included with
the analyzer. Use the Part Numbers in this list.
Parts that require scheduled replacement (or due to damage) includes, but
is not limited to, source lamps, Measuring Cell o-rings and windows,
Measuring Cells (heated or unheated), and Chopper Bearing Wheel
(Model 932 Multi-Gas Analyzers only). If optional equipment, such as
AMETEK’s Heated Acid Gas Probe, is used with your system, additional
spare parts for the probe will also be required (listed in the Heated Acid
Gas Probe Installation & Maintenance Guide).
Fuses, Optical Bench Service Electronics Board – Recommended Parts to Have On-Site
120 V 240 V
Description (Location*) Part No. Fuse Type Part No. Fuse Type
Main Board (F200) 300-8778 125 mA 300-8777 63 mA
Photomultiplier Tube (F300) 300-3214 0.2 A 300-3214 0.2 A
Lamp (F201) 300-9524 32 mA 300-9524 32 mA
* These fuses are located on the Optical Bench Service Electronics PWB in the Upper Enclosure.
1. Select a mode in which to display the parameters that you will ob-
serve. This can be done by either selecting a mode from the Mode
menu or by clicking on the mode icon on the Standard Toolbar.
For Trend and Tabular modes, you must first select a Channel before
you can define scan settings. Refer to “Defining Scan Settings” in
NOTE
Chapter 4 for detailed information about how to assign inputs to
channels for scanning data.
3. Click Device under Data Source and select the device from which to
set up. If only one device (analyzer) is being used, only Dev1 will be
displayed.
When Device is selected, the Input Channel list box is also displayed
and enabled. From this list, select a signal that you want to display in
the selected area on your mode display window by clicking on it.
(View this list to see the parameters that can be monitored.)
Any display or scan parameters that you define for a mode will not be
transferred to a new mode when you switch modes. However, the system
will continue to scan when you change modes.
Meter Mode
Meter mode allows you to view information in a meter-type format
(Figure A-1). You can assign the parameters in the Input Channel list box
(Scan tab) to any one of 12 meters. Alarms can be set up and displayed on
the meter to view the current reading in relation to the defined High and
Low alarm limits. You can also define unique scan, display, and alarm
output settings.
Figure A-1.
Meter mode display.
• Each meter has a bar representing a measured value. Text under the
meter displays the numeric value of the measured parameter.
• Each meter can be set to display alarm limits. As long as the measured
value is within the High and Low alarm limits, the meter bar appears
in green.
1. Select Meter from the Mode menu or click from the Standard
Toolbar.
3. Right-click the active meter and define the Scan, Display, and/or
Alarm settings for the meter.
Figure A-2.
Annunciator mode
display.
• For each channel, a green display indicates the value is within accept-
able limits.
• Before changing the display properties, you must select the display by
clicking on it. The display will be highlighted with a yellow border
when selected.
3. Right-click the active display and define the Scan, Display, and/or
Alarm settings for the display.
Figure A-3.
Trend mode (graph)
display.
• You can view the exact value of a trend line by holding down the left
mouse button in the trend graphical display area. Display the Data
File Toolbar to view the time of each data point.
3. Right-click the graph and define the Scan, Display, and/or Alarm
settings for the graph.
Figure A-4.
Tabular mode display.
3. Right-click the active display and define the Scan, Display, and/or
Alarm settings for the display.
For example, you can customize the window to show one Meter mode
display window, one Trend mode window, and two Annunciator win-
dows. Or, you could set up two Trend mode display windows, or two
Tabular mode windows.
Three default custom mode setups are included with the Configurator
Software and more can be added as needed. For each display window
added in Custom mode the scanning, displays, and alarm output options
are set as if in that mode.
For example, if adding a Meter mode window within the Custom mode,
all settings for that window are defined as if in Meter mode.
Figure A-5.
Set Custom Mode
Menu dialog box.
2. From the Edit menu, select Add Display and select the type of display
you want to add from the list of standard mode types. Or, use the
Custom mode icons on the Custom Mode Toolbar to add different
displays.
3. Specify the data source. Select the applicable source of data from the
Data Source group by clicking Device and then selecting the device
you want to enable.
5. From the File menu, select Save As to save the Custom mode. Or click
7. To retrieve the Custom mode file at a later time, select Open Configu-
ration from the File menu. If you used one of the predefined icons,
double-click one of the files to retrieve it.
• To move the window, click and drag the display window to the new
location and then release the mouse button (just an outline of the
window is visible while it is being moved).
1. From the Edit menu, select Add Display. Choose the type of display
window to add.
2. From the Edit menu, select Delete Display, press the Del key on the
1. From the File menu, select Save As. You can save the Custom mode
setup under a new name or use one of the predefined icons.
3. By default, the system places these files in the folder where the
Configurator Software is installed and gives the file a .dyc extension.
If you save a Custom mode file in the install folder with the name
“startup.dyc” it will automatically be opened when the program starts.
1. From the Mode menu, select Set Menu to view the Set Custom Mode
Menu dialog box (Figure A-6). You can associate a Custom mode file to
appear at the bottom of the Mode menu.
You can also provide a menu/button name to appear instead of the file
name at the bottom of the menu.
2. Use the Browse button to navigate to the location of the Custom mode
file.
3. To enter a name that will appear at the bottom of the Mode menu,
enter a name in the Menu/Button Name text field.
Figure A-6.
Preferences dialog
box.
• Trend Colors
Set the default trend channel colors.
You can copy any of the columns from a Tabular display window to
the clipboard.
Copying Magnitudes
You can copy any of the columns from a Tabular display window to the
clipboard.
3. From the Edit menu, select Paste Special (or Paste Link, depending on
the version of Windows running) to create the DDE link and continu-
ously update the data in real time.
Copying Channels
3. From the Edit menu, select Paste Special (or Paste Link, depending on
the version of Windows running) to create the DDE link and continu-
ously update the channel names in real time.
If you are importing data into Microsoft Excel, select Preference from
the Edit menu. To place the numbers into separate columns, select
NOTE
Return from the DDE Item Separator group. To place the numbers
in rows, select Tab.
Figure A-7.
Display parameters
tab.
Title
Type the title for the display window. It will then appear in the title
bar for that window and in other areas of the software to identify the
display window.
Lower limit
Applies to Trend and Meter modes only. This value is used only when
the Autoscale check box is not checked. If the Logarithmic check box
is selected, use this text field to enter the number of logarithmic
decades to display on the Y-axis using the upper limit as the top of the
display range. If the Logarithmic is not checked, enter a value into this
text box for the lower limit.
Units
This is the value in which your results will be displayed.
• In the Units text field, type in the units on which you want the
scanned information displayed.
• Precision
Defines the number of digits after the decimal point to be shown.
For example, if the precision is set to 2, a reading might be 5.22.
Channels
Allows you to select and define attributes for the selected device.
Figure A-8.
Display parameters
tab, Trend mode.
Channel color
Applies to Trend mode only. Defines a different color for the displayed
channel. A custom color can also be chosen by clicking on the Custom
button and selecting a color.
Minutes
Applies to Trend mode only. Defines the time interval to include on
the display.
Display Values
If checked, displays channel values in Trend mode.
Figure A-9.
Alarm setup tab.
For Trend and Tabular modes, first select the channel as the source of the
alarms before you define the alarm values. Setting alarms includes defin-
ing alarm limits and defining the actions that should be taken when an
alarm limit is reached. Alarm limits also appear on the main display for
the current mode. High and Low limit alarms appear in red; High and
Low warning alarms appear in yellow.
Name Enter a name for the alarm that will be used to identify
the alarm to the system.
Value Enter the high (upper) limit value for the alarm.
High Warn
When checked, enables the High Warning alarm.
Name Enter a name for the alarm that will be used to identify
the alarm to the system.
Value Enter the high (upper) limit value for the alarm. Must
be lower than the high alarm value.
Low Warn
When checked, enables the Low Warning alarm.
Name Enter a name for the alarm that will be used to identify
the alarm to the system.
Value Enter the low limit value for the alarm. Must be higher
than the low alarm value. For analog mode, any mass
values that fall below this limit are counted even if they
are from the same scan.
Name Enter a name for the alarm that will be used to identify
the alarm to the system.
Action button
Choose the Action button to view the Alarm Action dialog box (Fig-
ure A-10), where you can specify what action the system should take
when the specified alarm is set. This can be used for all four alarms.
Dialog Box
Specify a message to display when the alarm is set.
Figure A-10.
Alarm Action dialog
box.
Info File Enter the name and path of a help file or use the
Browse button to navigate to the location of a help
file and select the file.
Sound
Enter the name and path of a sound file or use the Browse button to
navigate to the location of the sound files and select a file. This allows
you to activate a warning sound (in addition to text instructions) to
alert the operator when the alarm is set.
Digital Output
Define how to send the alarm information to a digital output. Select
the device to which you want to send the digital output by using the
Device list box. Select the digital output module using the Module list
box.
Figure A-11.
Alarm message dialog
box.
• Use a 1/8" NPT female thread connector on the inlet flame arrestors.
• Use stainless steel fittings and tubing to connect clean and dry air or
nitrogen to both inlet flame arrestors. Refer to system drawings for
recommended pressure setting and other requirements.
• The purge gas should have a water dew point temperature of less
than -30 °C and be contaminant free. Refer to system drawings in the
“Supplemental Information” tab of this manual, or the Documentation
Package shipped with the analyzer
UPPER ENCLOSURE
Figure B-1.
Inlet and vent flame
arrestor locations for LOWER ENCLOSURE
Environmental Purge
option.
Before proceeding, test the area around the analyzer for hazardous
! gases and proceed only when the area is found to be safe.
WARNING
Do not remove the cover of the Explosion-Proof Digital Communica-
tions Port on the front of the Lower Enclosure and do not power up
the computer if there is a hazardous gas atmosphere present.
2. From the Setup tab click the Cell/Misc button to view the Cell/Misc
(Compensation) Parameters dialog box (Figure B-2). Enter a value in
the Default Cell Pres field, as indicated in the instructions included
with the upgrade.
3. From the Analog Inputs dialog box (Figure B-3), enter Low scale and
High scale values for Cell Pres – as indicated on the instructions
shipped with the upgrade package – to activate the pressure compen-
sation measurement (refer to Chapter 4 and the instructions included
with the upgrade for more information).
Figure B-3.
Analog Inputs dialog
box.
Jumpers on the Customer I/O Board are set up for a default pressure
measurement range of 3413.2–5171.5 mmHg absolute (66–100
NOTE PSIA). Depending on the pressure transducer used, these jumpers
may have to be adjusted.
Diagnostics Feature
• The user must supply an external power supply (15–30 VDC) capable
of supplying 25 mA per loop.
• 4–20 mA inputs.
If your application uses any of these optional gas sensors, refer to the
“Customer Connections (Micro-Interface Board)” drawing for wiring
details, such as termination points, options, and locations. This drawing is
included in the “Supplemental Information” tab of this manual and in the
Documentation Package shipped with the analyzer.
If your application uses this digital input option, refer to the “Customer
Connections (Micro-Interface Board)” drawing for wiring details, such as
termination points, options, and locations. This drawing is included in the
“Supplemental Information” tab of this manual and in the Documentation
Package shipped with the analyzer.
Overview
This Appendix (M93x Modbus Register Layout, Revision E, 2004-03-19)
describes the customer serial communication interface on the Model 931
Single-Gas and Model 932 Multi-Gas Analyzers.
• The analyzer also responds to Modbus queries with the special node
address of 255 from the service ports (RS-232 and RS-485). This acts as
a back-door access to the analyzer when the assigned node address is
unknown. The special node address should only be used when di-
rectly connecting to the analyzer (i.e., not in a multi-drop environ-
ment).
The concentration result, the zero drift specification, the span gas
concentration, and the alarm thresholds of a gas species must be in the
NOTE
same unit of measurement.
b When the analyzer is used to output absorbance or transmittance, the scaling factor of
10000 is used instead of 100 which is for concentration output.
c Each pair of registers represents the HC Warning and Fault bits as defined in registers
153 and 154. The first pair is the newest while the last pair is the oldest.
d Each pair of registers represents the OBMC Warning and Fault bits as defined in
registers 155 and 156. The first pair is the newest while the last pair is the oldest.
0 EEPROM Full
7–15 Reserved
7–15 Reserved
5–15 Reserved
5–15 Reserved
Bit Definition
0 Species 1 (93x)
1 Species 2 (932)
2 Species 3 (932)
3 Species 4 (932)
4 Species 5 (932)
5 Species 6 (932)
6 Sensor 1 (93x)
7 Sensor 2 (93x)
8 Sensor 3 (93x)
9–15 Reserved
p Service host configuration parameter is used by the service host software application
to store the user-preference information (e.g., show temperature and pressure in
imperial units). The definition and meaning of the parameter is outside the scope of
this document and its value has no effect on analyzer operation.
Bit Definition
0 Bench Orientation:
0 = normal
1 = reserved
1–3 Optical Filter Pulsing Sequence
000 = 1-3-5, 4-6-2
001 = 3-5, 4-6-1-2
010 = 1-3-4-5, 6-2
011 = motorless (1-3-2)
100 = 1-3-5, 4-6-1(2)
4–15 Reserved
s Each pair of registers represents a 32-bit floating-point value. The order of the
registers holding one floating-point value is implemented in the de facto standard
manner (i.e., the first register holds the lower 16-bits while the second register holds
the upper 16-bits of a 32-bit floating-point value).
LOOSE CLAMPED
PROCESS INSTRUMENTS
11 10
1 2 3 4
SOLENOID BLOCK DETAIL
OR
OR
OR
OR
A
RX+
R J302 (Customer Data Acquisition) J303 (Remote Service Port)
B 130
RX- JP305 1-2 JP307 1-2
JP306 1-2 JP308 1-2
Z
TX-
+5V
619
A
RX+
R J302 (Customer Data Acquisition) J303 (Remote Service Port)
B
RX- JP305 2-3 JP307 2-3
JP306 JP308
619
2-3 2-3
(c) +5V JP309 1-2 JP311 1-2
JP310 1-2 JP312 1-2
619
+5V
619
A
RX+
R J302 (Customer Data Acquisition) J303 (Remote Service Port)
130
B
RX- JP305 1-2 JP307 1-2
JP306 JP308
619
1-2 1-2
(d) +5V JP309 1-2 JP311 1-2
JP310 1-2 JP312 1-2
619
Z
TX-
NOTE: RESISTANCE VALUES IN OHMS.
619
LOWER ENCLOSURE
DETAIL A
OR
UPPER ENCLOSURE
LOWER ENCLOSURE
Conversion Factors
PARAMETER TO CONVERT FROM: TO: MULTIPLY BY:
AMOUNT of SUBSTANCE lb mol kmol 0.453 592 4
g mol kmol 0.001
(for Ideal gas) m3 of gas (0 °C, 101.325 kPa) kmol 0.044 615 8
m of gas (15 °C, 101.325 kPa)
3
kmol 0.042 493 2
ft of gas (60 °F, 1.0 atm)
3
kmol 0.001 195 30
LENGTH ft m 0.304 8
in mm 25.4
yd m 0.914 40