0% found this document useful (0 votes)
54 views41 pages

PHMeter PHM0033 Manual Eg v1.0

The document is a user manual for a PHM0033 pH/ORP meter. It includes sections on assembly, dimension drawings, mounting options including pipe, wall and panel mounting. It defines the terminal block wiring and provides keypad and interface descriptions. The interface displays real-time pH/ORP and temperature values, current output, and relay states. Error codes are displayed for issues like thermistor failures.

Uploaded by

Mauro César
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
54 views41 pages

PHMeter PHM0033 Manual Eg v1.0

The document is a user manual for a PHM0033 pH/ORP meter. It includes sections on assembly, dimension drawings, mounting options including pipe, wall and panel mounting. It defines the terminal block wiring and provides keypad and interface descriptions. The interface displays real-time pH/ORP and temperature values, current output, and relay states. Error codes are displayed for issues like thermistor failures.

Uploaded by

Mauro César
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 41

PHMeter_manual_eg

V1_rev0
sep23

USER MANUAL
PHM0033 - pH / ORP Meter
Contents

Overview 3
Assembly 4
Dimension Drawings 4
Pipe Mounting 5
Wall Mounting 6
Panel Mounting 6
Terminal Black Definitions 7
Keypad 9
Main Interface 9
System Information Setting 11
pH Probe Setting 12
pH Probe Calibration 19
ORP Probe Setting 23
ORP Probe Calibration 27
Analog Output 31
Relay Output 34

1
www.fertron.com.br
● Over view
PHM0033 series pH/ORP meter can work with analog pH or ORP probes to
meet versatile online pH or ORP monitoring applications.

◎ 1 probe input channel for analog pH or ORP probe.


◎ 1 thermistor input channel for NTC10k, Pt100 or Pt1000.
◎ Smart calibration design for pH probe. (Auto-recognition of Cal-points)
◎ Temperature correction available.
◎ Tag setting available.
◎ 2 current output channel for 4-20mA transmission.
◎ 3 SPDT relay output channel for Hi/Lo alarm.
◎ Operation password management authority provided.
◎ Modbus RS485 communication provided.
◎ English menu default, Chinese menu optional.
◎ IP67 protection level for whole meter housing achieved.

---------------------------- Main specification ---------------------------

Function Display/Transmission/Alarm Com. Port Modbus RS485

Screen 3.5’ monochrome (320*240) IP Level IP67

Probe Input 1 analog pH/ORP probe Dimension 144*152*113.3mm

Thermistor 1 NTC10k (2-wire), or Hi/Lo alarm (hysteresis &


Alarm
Input 1 Pt100 or Pt1000 (3-wire) delay configurable)

Analog 2 channel 4-20mA Power (100-240)VAC Default

Output ±0.2% (max load 500Ω) Supply (24VDC Optional)

Relay 3 SPDT for Hi/Lo alarm Panel/Pipe/Wall mounting


Mounting
Output 250Vac/30Vdc/3A(max) (cutout: 138*138mm)

2
www.fertron.com.br
● Assembly

Part ①:
M20 cable gland * 5
(with locknut)
Part ②:
M20 screw plug * 3
(with locknut)

● Dimension Drawings

3
www.fertron.com.br
● Pipe Mounting
Step 1: Fix the mounting bracket (optional) to the back cover with 4 M3*8
screws (with spring washer).
Step 2: Lock the bracket to the pipe (Ø40-60mm) with 2 metal clamps.

4
www.fertron.com.br
● Wall Mounting
Step 1: Fix the mounting bracket to the back cover with 4 M4*8 screws (with
spring washer).
Step 2: Fix the bracket to the drilled wall by 4 expansion bolts.

● Panel Mounting
Step 1: Insert the tail of the meter into the cutout of the panel (138*138mm).
Step 2: Assemble the 4 compression locks to the meter with 4 M4*10 screws
(with spring washer).
Step 3: Screw the M4*16 screw (already assembled in the compression locks)
in until it reaches the back of the panel.

5
www.fertron.com.br
● Terminal Block Definitions

Terminal blocks

inside the meter

AC Powered Version

DC Powered Version

6
www.fertron.com.br
TB1 TB2 TB3
Wiring definitions Wiring definitions Wiring definitions
1 Temp (Line) 1 / 1 PE
2 Temp signal 2 / 2 RS485A / -

(2 wire)

(3 wire)
3 Temp signal 3 / 3 RS485B / +
4 Reference 4 / 4 Ground
5 pH/ORP signal 5 / 5 4-20mA output -
6 4-20mA output +
7 4-20mA output -
8 4-20mA output +

TB4 TB5 TB6


Wiring definitions Wiring definitions

1 / 1 / 1 L

2 (washing)Normally-open contact 3 2 / 2 N 220Vac±10%

3 (washing)Common terminal 3 3 / 3 Earth

4 (washing)Normally-closed contact 3 4 / 1 24V+

5 (Lo)Normally-open contact 2 5 / 2 24V- 24Vdc±10%

6 (Lo)Common terminal 2 6 / 3 Earth

7 (Lo)Normally-closed contact 2

8 (Hi)Normally-open contact 1

9 (Hi)Common terminal 1

10 (Hi)Normally-closed contact 1

(Note: The power terminal mark for the 24VDC version is different from the AC version)
When using 2-wire NTC10k or 2.25K thermistor, please connect the 2 signal wires to the
TB1-2 and TB1-3 terminals. When using 2-wire Pt100 or Pt1000 thermistor, please connect
the 2 signal wires to the TB1-2 and TB1-3 terminals, meanwhile, short connect the TB1-1 and
TB1-2 terminals to ensure that the meter can work correctly. (A piece of metal bridge is
already included in the accessories of the meter)

7
www.fertron.com.br
● Keypad

Back Up

Confirm Down

Left Right

● Main Inter face



⑨ ⑩


④ ⑤ ⑥
⑧ ⑦
① : "pH" or "ORP" default, can also be tag for the current measuring point.
② : Normally the real-time measured value of pH or ORP is displayed. When
there are abnormal conditions, the meaning of the information displayed
in this area is as follow:

8
www.fertron.com.br
Display Indication

Transition state after enabling new configuration settings .

*.** Continuous display indicates the configuration is incorrectly set.

(Display according to the decimals in the range limit setting)

…… Invalid measured value caused by temperature measurement failure.

③ : Bar graph for pH/ORP measured value compared to range limit setting.
④ : Original voltage measured value for pH application, none for ORP.
⑤ : Source of temperature value:

Display Source of temperature value

Real-time measured by thermistor.

Manually set.

⑥ : Temperature value (unit: ℃ default, ℉ optional).


⑦ : Current output value (unit: mA). Once the current output is enabled, a
warning sign will indicate an open-circuit failure of the current loop:

Display Indication

Open-circuit failure of the current loop detected.

⑧ : The left and right square shape respectively indicates the state of relay 1
(high alarm) and relay 2 (low alarm):

Display Indication

The relay is activated. (Alarm on)

The relay is deactivated. (Alarm off )

The relay output is not enabled.

⑨ : An exclamation mark ("!") will indicate the thermistor failure.


⑩ : Following the above-mentioned "!", the error code will be displayed:

9
www.fertron.com.br
Error code Indication

E:001 Open-circuit in Pt100 or Pt1000 loop detected.

E:002 Open-circuit in NTC10k loop detected.

E:004 Too large lead resistance of Pt100 or Pt1000 detected.

E:008 Short-circuit in NTC10k loop detected.

E:010 Too small resistance of thermistor detected.

● System Information Setting


Press "Back" and "Confirm" simultaneously at the main interface to enter the
setting menu of the meter (the password verification interface will appear first,
press "Confirm" directly if the password is not set).

Select "System" option in the setting menu by "Up/Down" key, and press
"Confirm" or "Right" to enter the system information setting sub-menu:
10
www.fertron.com.br
All options are introduced as follow:

"Password": Select it and press "Confirm", you can enter the password setting
interface to set the 6-digit operation password of the meter (position the data
input cursor by "Left/Right" key, then use "Up/Down" key to select options
between number "0" to "9" in turn). When setting is done, press "Confirm" to save.
When pressing "Confirm" on the password setting or verification interface, the
system will uniformly identify the unset parts (reserved as "*") as "0" , e.g., if you
press "Confirm" in the state of displaying "**27**", the system will recognize
"002700" as the actual input value.

● pH Probe Setting
After connecting the signal wire of the analog pH probe to meter, the essential
settings for the probe should be done. Press "Back" and "Confirm" simultaneously
at the main interface to enter the setting menu of the meter (the password
verification interface will appear first, press "Confirm" directly if the password is
not set):

11
www.fertron.com.br
Sensor Channel Setting
In the setting menu, select "Sensor" option and press "Confirm" or "Right" to
enter the sensor channel setting sub-menu:

All options are introduced as follow:


"Type": Sensor type. When the cursor is located at "Type", you can use the
"Left/Right" or "Confirm" key to choose between "pH" and "ORP". Be sure to
pick "pH" when using pH probe.
"Tag": Tag of measuring point. Select it and press "Confirm" to enter the tag
input interface as below (the tag supports up to 12 characters):
12
www.fertron.com.br

The tag input interface is divided into two areas:


Area ①: Input result area for the tag. When entering the tag input interface,
the underline cursor is at the first character from the left by default, as shown in
the figure above.
Area ②: Input option area for the tag. When this area is not activated as
shown in the figure above, press "Confirm" to activate it. After activation, the
background highlight cursor selects "Save" option by default as below:

When area ② is activated, press "Back" to return to the inactive state.

13
www.fertron.com.br
When area ② is activated, The sub-options are as follow:

Option Result of conformation

<-- Cursor in area ① shifts to the left one character. (Selects the previous one)

--> Cursor in area ① shifts to the right one character. (Selects the next one)

Save Save the tag and return to the previous page

The underline cursor alternately selects "8", "A", and "a" inside the option, and

8/A/a the content of the rest of area ② alternates between numbers, capital letters

and lower case letters correspondingly.

Delete the selected character in area ①, and all characters on the right side will
Del
shift to the left by one position automatically.

Clear Clear all characters in area ①.

The rest Input the chosen option to the currently selected character in area ①.

There are two ways to set the tag:


No.1: In the case of area ② being inactive, use "left/Right" key to select the
desired character of the tag in area ①, then use "Up/Down" key to cycle between
lower case letters, capital letters, numbers and symbols to find the target option,
and select the next tag character to continue setting by "Right" key. After setting
all desired characters, press "Confirm" to activate the area ②. The background
highlight cursor selects "Save" button by default, then press "Confirm" again to
save the tag.
No.2: Activate area ② first, use "<--" or "-->" button to move the underline
cursor in area ① to a desired position, then use arrow keys to find the target
numbers, capital letters, lower case letters or symbols, and press "Confirm" to
update the selected content to Area ①. After setting all desired characters, select
"Save" button and press "Confirm" to save the tag.
14
www.fertron.com.br
"Temperature Source": Source of temperature value. Select it and you can use
the "Left/Right" or "Confirm" key to choose the temperature source between
"NTC10k", "Pt1000", "Pt100" and "Manual".
"Manual Input": This option becomes selectable when the above-mentioned
"Temperature Source" option is set to "Manual". After selecting "Manual input",
you can firstly switch the temperature unit between "℃" and "℉" by "Left/Right"
key, then press "Confirm" to further set the desired temperature value (position
the data input cursor by "Left/Right" key, then use "Up/Down" key to select
options between number "0" to "9" in turn). When setting is done, press "Confirm"
to save.
"Temperature Offset": This option becomes selectable when the above-
mentioned "Temperature Source" option is set to "NTC10k", "Pt1000" or "Pt100".
This offset can be used to fix the deviation between the measured temperature
and the actual temperature. After selecting "Temperature Offset", you can firstly
switch the temperature unit between "℃" and "℉" by "Left/Right" key, then press
"Conffrm" to further set the desired offset value (position the data input cursor
by "Left/Right" key, then use "Up/Down" key to select options between number
"0" to "9" in turn). When setting is done, press "Confirm" to save, and the value
of the temperature displayed in the main interface will be the sum of the
measured temperature and the offset value.

Database Channel Setting


Normally there is no need to set the database channel. But if you want to
modify the displayed decimals of the measured value, or a different type of probe
is replaced for the meter, the essential settings for the database channel should
be done. Press "Back" and "Confirm" simultaneously at the main interface to enter
the setting menu of the meter (the password verification interface will appear first,
press "Confirm" directly if the password is not set):

15
www.fertron.com.br
Select "Database" option with "up/down" key and press "Confirm", you can

enter the database channel setting sub-menu as shown below:

All options are introduced as follow:


"Channel": Database channel. When the cursor is located at "Channel", you
can select one of the four database channel by "Left/Right" or "Confirm" key. The
default assignment rule for database channel is (take pH meter application for
example):

16
www.fertron.com.br
When "Type" of "Sensor" is set "pH"
Unit
Database channel Default signal source
DB01 pH.pH pH measured value from pH probe /
Temperature measured value from pH probe
DB02 pH.Temperature ℃
(or temperature value from manual input)
DB03 pH.pH mV Original voltage measured value from pH probe mV
DB04 pH.Reserved Not in use

"Range Limit (Low)"/"Range Limit (High)": Select one of these options and
press "Confirm" to enter the range setting interface, then you can set the range
and display decimals of the chosen database channel. The range of DB01 channel
(pH value) will also be used for the bar graph display ratio calculation. The
recommended ranges and display decimals for each parameter are as follow (can
be changed according to your actual application requirements):

Sensor information Range and display decimals Recommended

Type Parameter Max range Range limit (Low) Range Limit (High)

pH value 0-14 0.00 14.00


pH
Temperature -20 to 120 ℃ -20.0 120.0
probe
Original voltage -1500 to 1500 mV -1500 1500

(Note: The change of the range limits setting will cause the linkage change of the alarm

value setting. If the alarm function is enabled, the alarm setting in the "relay output" option

should be confirmed immediately after changing the range limits setting)

In the range setting interface, you can position the data input cursor by
"Left/Right" key, then use "Up/Down" key to select options between "-" (minus
sign), "." (decimal point) and number "0" to "9" in turn. When setting is done,
press "Confirm" to save.
"Filter": Filtering time constant in seconds. Selecting the first row and press

17
www.fertron.com.br
"Confirm", you can set the first-order inertia filtering time of this parameter, the
setting range is 0~255. The value in the second row stands for the display
updating frequency for the parameter, and the setting range is 0~15, keeping
default recommended.
After setting the sensor channel and database channel, press "back" to return
to the main interface, then all setting takes effect.

● pH Probe Calibration
The meter adopts an intelligent calibration design that automatically
recognizes the calibration point, which allows the user to easily carry out the two-
point or three-point calibration for the pH probe. Press "Confirm" at the main
interface, a check information will appear first to ask whether you need to
perform the calibration:

If you accidentally press the "Confirm" at the main interface, causing the
above information to appear, you can press "Back" to return to the main
interface.
If you really need to calibrate the probe, please press "Confirm" again (the
password verification interface will appear first, press "Confirm" directly if the
password is not set) to enter the calibration interface of the pH probe:
18
www.fertron.com.br
The "Result" option row in the calibration interface displays the last calibration
result stored by the meter. The factory default value is "E0=0.0mV; S=100.0%".
Meanwhile, the "Reset to default" option is provided. Select "Reset to default"
option and press "Confirm" will restore the calibration results stored in the meter
to the default value. The operation process of the pH probe calibration is
described as follow:
Step 1: Setting the buffer type. When "Buffer Type" option is selected, you can
choose between "USA (7.0, 4.01, 10.01)" and "NIST (6.86, 4.0, 9.18)" options by
"Left/Right" or "Confirm" key.
Step 2: Performing the first point calibration. The cleaned probe should be
placed in a first point buffer solution (type "USA", pH7.0 for example) and stirred
slowly for about 20s. The centered reading of the "Calibrate" option row is the
real-time pH value measured by the probe. Then select "Calibrate" by "Up/Down"
key and press "Confirm" to start the calibration of the first point:

19
www.fertron.com.br
The cursor will automatically jump to the "Result" option, and the last
calibration result will no longer be displayed. When the message on the right
side changes from "In Progress" to "OK", the first point calibration is done and
the new calibration result will be displayed:
(Note: If the "Fault k, b" message appears, it means that the performance of the probe

may have significantly deteriorated)

Step 2: Performing the follow-up calibration. The cleaned probe should be


placed in the second point buffer solution (type "USA", pH4.01 for example) and
stirred slowly for about 20s. Then select "Calibrate" option and press "Confirm" to
20
www.fertron.com.br
start the calibration of the second point. The cursor will automatically jump to the
"Result" option, and the last calibration result will no longer be displayed. When
the message on the right side changes from "In Progress" to "OK", the second
point calibration is done and the new calibration result will be displayed:
(Note: If the "Fault k, b" message appears, it means that the performance of the probe

may have significantly deteriorated)

Till now, the two-point calibration of the pH probe is completed. If the third
point calibration is not required, select "Quit" option and press "Confirm" to return
to the main interface. If needed, the cleaned probe should be placed in the third
point buffer solution (type "USA", pH10.01 for example) and stirred slowly for
about 20s. Then select "Calibrate" option and press "Confirm" to start the
calibration of the third point. The cursor will automatically jump to the "Result"
option, and the last calibration result will no longer be displayed. When the
message on the right side changes from "In Progress" to "OK", the third point
calibration is done and the new calibration result will be displayed:
(Note: If the "Fault k, b" message appears, it means that the performance of the probe

may have significantly deteriorated)

21
www.fertron.com.br
Till now, the three-point calibration of the pH probe is completed, select "Quit"
option and press "Confirm" to return to the main interface.

● ORP Probe Setting


After connecting the signal wire of the analog ORP probe to meter, the
essential settings for the probe should be done. Press "Back" and "Confirm"
simultaneously at the main interface to enter the setting menu of the meter (the
password verification interface will appear first, press "Confirm" directly if the
password is not set):

22
www.fertron.com.br
Sensor Channel Setting
In the setting menu, select "Sensor" option and press "Confirm" or "Right" to
enter the sensor channel setting sub-menu:

All options are introduced as follow:


"Type": Sensor type. When the cursor is located at "Type", you can use the
"Left/Right" or "Confirm" key to choose between "pH" and "ORP". Be sure to
pick "ORP" when using ORP probe.
"Tag": Tag of measuring point. Please refer to the "Tag" function description
on page 14 for operating instructions.
"Temperature Source": Source of temperature value. Select it and you can use
the "Left/Right" or "Confirm" key to choose the temperature source between
"NTC10k", "Pt1000", "Pt100" and "Manual".
"Manual Input": This option becomes selectable when the above-mentioned
"Temperature Source" option is set to "Manual". After selecting "Manual input",
you can firstly switch the temperature unit between "℃" and "℉" by "Left/Right"
key, then press "Confirm" to further set the desired temperature value (position
the data input cursor by "Left/Right" key, then use "Up/Down" key to select
options between number "0" to "9" in turn). When setting is done, press "Confirm"

23
www.fertron.com.br
to save.
"Temperature Offset": This option becomes selectable when the above-
mentioned "Temperature Source" option is set to "NTC10k", "Pt1000" or "Pt100".
This offset can be used to fix the deviation between the measured temperature
and the actual temperature. After selecting "Temperature Offset", you can firstly
switch the temperature unit between "℃" and "℉" by "Left/Right" key, then press
"Confirm" to further set the desired offset value (position the data input cursor
by "Left/Right" key, then use "Up/Down" key to select options between number
"0" to "9" in turn). When setting is done, press "Confirm" to save, and the value
of the temperature displayed in the main interface will be the sum of the
measured temperature and the offset value.

Database Channel Setting


Normally there is no need to set the database channel. But if you want to
modify the displayed decimals of the measured value, or a different type of probe
is replaced for the meter, the essential settings for the database channel should
be done. Press "Back" and "Confirm" simultaneously at the main interface to enter
the setting menu of the meter (the password verification interface will appear first,
press "Confirm" directly if the password is not set):

24
www.fertron.com.br
Select "Database" option with "up/down" key and press "Confirm", you can
enter the database channel setting sub-menu as shown below:

All options are introduced as follow:


"Channel": Database channel. When the cursor is located at "Channel", you
can select one of the four database channel by "Left/Right" or "Confirm" key. The
default assignment rule for database channel is (take ORP meter application for
example):

When "Type" of "Sensor" is set "ORP"


Unit
Database channel Default signal source
DB01 ORP .ORP ORP measured value from ORP probe mV
Temperature measured value from ORP probe
DB02 ORP .Temperature ℃
(or temperature value from manual input)
DB03 ORP .Reserved Not in use
DB04 ORP .Reserved Not in use

"Range Limit (Low)"/"Range Limit (High)": Select one of these options and
press "Confirm" to enter the range setting interface, then you can set the range
and display decimals of the chosen database channel. The range of DB01 channel
(ORP value) will also be used for the bar graph display ratio calculation. The

25
www.fertron.com.br
recommended ranges and display decimals for each parameter are as follow (can
be changed according to your actual application requirements):

Sensor information Range and display decimals Recommended

Type Parameter Max range Range limit (Low) Range Limit (High)

ORP ORP value -1500 to 1500 mV -1500 1500

probe Temperature -20~120 ℃ -20.0 120.0

(Note: The change of the range limits setting will cause the linkage change of the alarm

value setting. If the alarm function is enabled, the alarm setting in the "relay output" option

should be confirmed immediately after changing the range limits setting)

In the range setting interface, you can position the data input cursor by
"Left/Right" key, then use "Up/Down" key to select options between "-" (minus
sign), "." (decimal point) and number "0" to "9" in turn. When setting is done,
press "Confirm" to save.
"Filter": Filtering time constant in seconds. Selecting the first row and press
"Confirm", you can set the first-order inertia filtering time of this parameter, the
setting range is 0~255. The value in the second row stands for the display
updating frequency for the parameter, and the setting range is 0~15, keeping
default recommended.
After setting the sensor channel and database channel, press "back" to return
to the main interface, then all setting takes effect.

● ORP Probe Calibration


The meter supports single-point or two-point calibration for the ORP probe.
Press "Confirm" at the main interface, a check information will appear first to ask
whether you need to perform the calibration:

26
www.fertron.com.br
If you accidentally press the "Confirm" at the main interface, causing the
above information to appear, you can press "Back" to return to the main
interface.
If you really need to calibrate the probe, please press "Confirm" again (the
password verification interface will appear first, press "Confirm" directly if the
password is not set) to enter the calibration interface of the ORP probe:

The "Result" option row in the calibration interface displays the last calibration
result stored by the meter. The factory default value is "k=1.000, b=0.0".
Meanwhile, the "Reset to default" option is provided. Select "Reset to default"
27
www.fertron.com.br
option and press "Confirm" will restore the calibration results stored in the meter
to the default value. The operation process of the ORP probe calibration is
described as follow:
Step 1: Setting the ORP value of the ORP buffer solution. When entering the
calibration interface, the cursor will be located at the buffer ORP value setting
option for point 1. Press "Confirm" to further set the desired value (position the
data input cursor by "Left/Right" key, then use "Up/Down" key to select options
between "-" (minus sign) and number "0" to "9" in turn). When setting is done,
press "Confirm" to save.
Step 2: Performing the first point calibration. The cleaned probe should be
placed in a first point buffer solution (470mV ORP buffer for example) and stirred
slowly for about 20s. The centered reading of the "Calibrate" option row is the
real-time ORP value measured by the probe. Then select "Calibrate" by
"Up/Down" key and press "Confirm" to start the calibration of the first point:

The cursor will automatically jump to the "Result" option, and the last
calibration result will no longer be displayed. When the message on the right
side changes from "In Progress" to "OK", the first point calibration is done and
the new calibration result will be displayed:

28
www.fertron.com.br
(Note: If the "Fault k, b" message appears, it means that the performance of the probe

may have significantly deteriorated)

Till now, the single-point calibration of the ORP probe is completed. If the
second point calibration is not required, select "Quit" option and press "Confirm"
to return to the main interface. If needed, the cleaned probe should be placed in
the second point buffer solution (220mV ORP buffer for example) and stirred
slowly for about 20s. Then select "Calibrate" option and press "Confirm" to start
the calibration of the second point. The cursor will automatically jump to the
"Result" option, and the last calibration result will no longer be displayed. When
the message on the right side changes from "In Progress" to "OK", the second
point calibration is done and the new calibration result will be displayed:
(Note: If the "Fault k, b" message appears, it means that the performance of the probe

may have significantly deteriorated)

29
www.fertron.com.br
Till now, the two-point calibration of the ORP probe is completed, select "Quit"
option and press "Confirm" to return to the main interface.

● Analog Output
Press "Back" and "Confirm" simultaneously at the main interface to enter the
setting menu of the meter (the password veriÞcation interface will appear first,
press "Confirm" directly if the password is not set):

Select "AO" option by "Up/Down" key and press "ConÞrm" or "Right" key to
enter the analog output setting sub-menu:
30
www.fertron.com.br
The meter provides two AO channels. When the cursor is located at "Channel",
you can use the "Left/Right" or "Confirm" key to choose between "AO 01" and
"AO 02".
The "Source" option of the AO channel represents whether the channel is
enabled and signal source of it. When "Source" option is selected, you can use
"Left/Right" or "Confirm" key to alternately select each option on the right side of
it, as explained below:

Status Source Description

Not enabled Closed Analog output channel closed.

Manual output Manual Output the manually preset value of current.

Transmit the measured value of a desired parameter to 4-20mA.


DB01
DB01 is the primary measured value.
Transmitter to
DB02 is usually temperature measured value.
DB03
DB03 is the other measured value from the digital probe.

When "Source" option is set to "Manual", the "Preset Value" option at the
bottom of the interface will become selectable as below:
31
www.fertron.com.br
Then you can select "Preset Value" option and press "Confirm" to further set
the preset value for manual output (position the data input cursor by "Left/Right"
key, then use "Up/Down" key to select options between number "0" to "9" in turn).
When setting is done, press "Confirm" to save.
If transmission output is needed, the "source" option should be set to the
database channel representing the desired parameter. The boundary for 4mA
and 20mA can be set according to the actual application requirements. The
interface is as below (taking pH application for example):

The "source" option is set to "DB01 pH.pH" in the figure above, indicating that
32
www.fertron.com.br
the "AO 01" channel is used to transmit the pH measured value according to the
application range of 0 to 14.
When performing transmission, if the system detects a thermistor failure, a
specific value of the output current can be used to make an indicating to the
monitoring system which the current loop is connected to. When "Sensor Fault
Indication" option is selected, you can choose the desired setting by "Left/Right"
or "Confirm" key, as described below:

Options for sensor fault indication Current output when thermistor failure detected

Preset Value The manually preset value

4mA 4.00mA

20mA 20.00mA

Hold Keep the last valid output value

After completing the AO setting and returning to the main interface, the new
configuration takes effect. You can check the real-time status of AO channel at
the bottom right of the main interface. Once the current output is enabled, a
warning sign will indicate an open-circuit failure of the current loop:

Display Indication

Open-circuit failure of the current loop detected.

● Relay Output (Including auto-clean control)


Press "Back" and "Confirm" simultaneously at the main interface to enter the
setting menu of the meter (the password verification interface will appear first,
press "Confirm" directly if the password is not set):

33
www.fertron.com.br
Select "Relay" option by "Up/Down" key and press "Confirm" or "Right" key to
enter the relay output setting sub-menu:

The meter provides three SPDT relays. When the cursor is located at "Channel"
option, you can select "R01", "R02" or "R03" by using "Left/Right" or "Confirm" key.
When the relay channel is selected, the options below are all for the chosen
channel.
The "Function" option of the relay channel represents whether the channel is
enabled and the designated function. When "Function" option is selected, you
can use "Left/Right" or "Confirm" key to alternately select each option on the right
34
www.fertron.com.br
side of it, as explained below:

Status Function Description

Not enabled Closed Relay output channel closed.

Manual output Manual Output the manually preset state of the relay.

R01 High Alarm Perform high alarm function for a desired parameter.
Alarm
R02 Low Alarm Perform low alarm function for a desired parameter.

Auto-clean R03 Auto-clean Perform auto-clean output according to configuration.

Relay for Alarm Output


If alarm output function is needed, please refer to the corresponding content
in the above table to set the "Function" option of the relay output channel:

The rest five options (scrolling to the bottom) are introduced as follows:
"Source": The signal source of the alarm (high alarm for "R01" and low alarm
for "R02"). When the cursor is located at "Source" option, you can select from
DB01 to DB03 (in pH application) by "Left/Right" or "Confirm" key. In the figure
above, the "DB01 pH.pH" is selected, indicating that the high alarm signal source
of "R01" is designated as the pH measured value from the pH probe.

35
www.fertron.com.br
"Alarm Value": The threshold value to trigger the alarm (high alarm for "R01"
and low alarm for "R02"). Select it and press "ConÞrm" to further set the desired
threshold value (position the data input cursor by "Left/Right" key, then use
"Up/Down" key to select options between "-" (minus sign) and number "0" to "9"
in turn). When setting is done, press "Confirm" to save.
"Hysteresis": The alarm hysteresis. Select it and press "Confirm" to further set
the hysteresis value (position the data input cursor by "Left/Right" key, then use
"Up/Down" key to select options between number "0" to "9" in turn). When setting
is done, press "Confirm" to save. A hysteresis value requires the measured value
to return within the threshold by a specified value before the relay is deactivated.
"Delay": Delay time for triggering an alarm in seconds. Select it and press
"Confirm" to further set the desired length of time (position the data input cursor
by "Left/Right" key, then use "Up/Down" key to select options between number
"0" to "9" in turn). When setting is done, press "Confirm" to save. A time delay
requires the threshold to be exceeded continuously for the specified length of
time before activating the relay. The relay will not be activated if the alarm
condition disappears before the delay period is over. The default value for "Delay"
option is "5s".
"Min. Changeover Interval": The minimum time interval between the two
different states of the relay output. That is, from the time when the relay switched
the output state, the state will be maintained for at least the specified length of
time before the new output state switching instruction is actually executed. This
function helps to avoid frequent actions of contacts on critical conditions in
specific scenarios. Select "Min. Changeover Interval" option and press "Confirm"
to further set the desired duration in seconds (position the data input cursor by
"Left/Right" key, then use "Up/Down" key to select options between number "0"
to "9" in turn). When setting is done, press "Confirm" to save. The default is "0s".
After completing relay output setting and returning to the main interface,
36
www.fertron.com.br
the new configuration takes effect. You can check the real-time state of relay
output channel at the bottom left of the main interface. The left and middle
square shape respectively indicates the state of relay 1 (high alarm) and relay 2
(low alarm):

Display Indication

The relay is activated. (Alarm on)

The relay is deactivated. (Alarm off )

The relay output is not enabled.

Refer to the table below for states of relay (taking relay 1 for example):

State of Actual state of contacts between

Relay 1 (R01) NO1 and COM1 NC1 and COM1

Alarm off Break Make

Alarm on Make Break

Relay for Manual Output


When "Function" option is set to "Manual" (taking relay 1 for example):

37
www.fertron.com.br
The last options "Preset Value" (scrolling to the bottom) will be enabled:

Select "Preset Value" option and you can use the "Left/Right" or "Confirm"
key to choose between "Off" and "On".
After completing relay output setting and returning to the main interface,
the new configuration takes effect. You can check the real-time state of relay
output channel at the bottom left of the main interface. The left and middle
square shape respectively indicates the state of relay 1 and relay 2:

Display Indication

The relay is activated. (Manual output on)

The relay is deactivated. (Manual output off)

The relay output is not enabled.

Refer to the table below for states of relay (taking relay 1 for example):

State of Actual state of contacts between

Relay 1 (R01) NO1 and COM1 NC1 and COM1

Manual off Break Make

Manual on Make Break

38
www.fertron.com.br
Relay for Auto-Clean Control
If auto-clean control is needed, please refer to the corresponding content in
the table on page 50 to choose channel R03 and set the "Function" option to
"Auto-clean":

The rest three options are introduced as follows:


"Cycle": The interval between two auto-cleaning process for the probe. Select
it and press "Confirm" to enter the interval setting interface. The time format is
"DD-HH:MM" (day-hour:minute). Press "Confirm" to save the setting.
"Time": Duration of each auto-clean process in seconds. Select it and press
"Confirm" to enter the duration setting interface. Press "Confirm" to save the
setting.
(Note: During auto-cleaning process, the word "hold" will be displayed in the main interface

above the probe reading, the probe reading and the analog output will remain as the last value

before the cleaning starts.)

"Resume Delay": The delay (in seconds) from the end of cleaning to returning
to normal measurement. Select it and press "Confirm" to enter the duration
setting interface. Press "Confirm" to save the setting.
(Note: During the delay, the word "hold" will be displayed in the main interface above the

probe reading, the probe reading and the analog output will remain as the last value before
39
www.fertron.com.br
the cleaning starts. when delay is completed, the word "hold" in the main interface will

disappear, the probe reading and the analog output will resume.)

After completing relay output setting and returning to the main interface, the
new configuration takes effect and the first cleaning is performed directly. You
can check the real-time state of relay 3 at the bottom left of the main interface.
The third square shape from left indicates the state of relay 3:

Display Indication

The relay is activated. (Cleaning being carried out)

The relay is deactivated. (No cleaning was carried out)

The relay output is not enabled.

Refer to the table below for states of auto-clean relay (R03) output:

State of Actual state of contacts between

Relay 3 (R03) NO3 and COM3 NC3 and COM3

Off Break Make

On Make Break

40
www.fertron.com.br

You might also like