0% found this document useful (0 votes)
49 views14 pages

SLX-F2 SH Chapter 3 Block Diagrams

The document contains block diagrams and descriptions of the major components of an SLX-F2 medical device system. It includes block diagrams of the power unit, electronic rack, water circulatory system, safety circuits, and hydraulic unit. The first block diagram shows the overall SLX-F2 system and connections between the main components including the power supply, control panel, movement process, water/air process, and shockwave process.

Uploaded by

Michael Lafertte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
49 views14 pages

SLX-F2 SH Chapter 3 Block Diagrams

The document contains block diagrams and descriptions of the major components of an SLX-F2 medical device system. It includes block diagrams of the power unit, electronic rack, water circulatory system, safety circuits, and hydraulic unit. The first block diagram shows the overall SLX-F2 system and connections between the main components including the power supply, control panel, movement process, water/air process, and shockwave process.

Uploaded by

Michael Lafertte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 14

SERVICE -

DOCUMENTATION
®
MODULITH
SLX-F2

Chapter 3

Block Diagrams

10-004-02-1005 3-1
Table of contents

3.1 SLX-F2 System overview ..................................................................................................3


3.2 Power unit...........................................................................................................................4
3.3 Electronic rack ...................................................................................................................5
3.4 Water circulatory system ..................................................................................................6
3.5 BMC combinational logic ..................................................................................................7
3.6 BMC safety circuit..............................................................................................................8
3.7 SWC safety circuit .............................................................................................................9
3.8 WAC safety circuit .......................................................................................................... 10
3.9 PCCU block diagram ...................................................................................................... 11
3.10 PCCU timing diagram ..................................................................................................... 12
3.11 CTU100 GAL .................................................................................................................... 13
3.12 Hydraulic unit .................................................................................................................. 14

3-2 10-004-02-1005
Mains Voltage
Therapy Wave

3.1
Release Button
Power
Control Panel Main Process
Supply

10-004-02-1005
Therapy Wave
DC AC Release
and Unlocking
CAN Bus

Basic Movement Process Water Air Process Shockwave Process


SLX-F2 System overview

Safety Signals

Charge
And
Trigger

Unit
Unit

Movement
Airbag Unit
Water Circuit

X,Y,Z, Trend.
CTU
Switch Focus

Therapy Head
Heating/Cooling

Inline Movement

Table Movements
RC1416AL

C-arc Controller
C-arc Parallel Interface
ECG/Resp. Trigger

US Transducer Position

3-3
3-4
N L PE PA
3.2
Mains
Fuse

PCCU

Mains
control
Switch
T1
K1 bistable relay
Heating
240V 230V Element
Power unit

220V
200V
control control control
TH
Reset
TH
0 0

Therapy Head
Switchmode Motor Driver
Power Supply
control
100VAC to
24VDC Supply
240 VAC
Table Positioning
Controller
control
RC1417AL

Temperature
Control
Temperature

Sensor Safety
Heating

TPCEnable

24VMotEnable
PCCU On
Switch Safety

Water Pump Control Logic Sensor Supply BMC BMC SWC WAC WAC Temperature
Power Supply Switch Safety

10-004-02-1005
3.3 Electronic rack

Service Board

Stormlink Controller

10-004-02-1005 3-5
3.4 Water circulatory system

RC1418AL

3-6 10-004-02-1005
3.5 BMC combinational logic

XrayEmergHalt

Interface
Emergency Halt button

C-arc
MotionStop
SLX-F2
& EmergHalt
Therapy head
SCOUPS
mechanically
coupled

Patient table in MRLS VMotInline


therapy position &

Anti-tip beam MRPS


driven out

SWC MotionEnable
Safety Logic
24VMotEnable
&
to therapy head
Patient table X2FS
motor regulator
X2 footwards

Z1 lift in highest Z1THS


position

Therapy head in
park position >=1
SPS

SFS

Therapy head in
front position

>=1 TPCEnable
&
to control unit of
MX c-arc in hydraulic unit
park position

MXPS
Patient table TRD0S
Trendelenburg in >=1
horizontal position

Patient table Z2 lift Z2TRDS


within uncritical position
while Trendelenburg
&
MXTHS
MX c-arc in
therapy position
RC1409AL

10-004-02-1005 3-7
3-8
3.6
C-arc
Interface

SWC
Vmotor
Safety Logic
Safety Circuit 24VMotEnable
Patient Table Therapy
Therapy Head
Position Swivelling
Control
M

Safety Signals
Anti-tip Beam
Therapy Head
BMC safety circuit

Read
Status
Motor Driver Vmotor,
Emergency Vlogic
Halt
Button TPCEnable

Therapy Head
Positioning Patient Table
Sensors Control
RS232c Positioning
Microcontroller Controller
Vmotor
Patient Table
Positioning
Inline
Enable
Sensors
Inline

Logic Sensor Signals


Inline Control
M
Positioning
Sensors
Inline
CAN Bus Motor Driver
BMC RC1408AL

10-004-02-1005
3.7 SWC safety circuit

RC1413AL

MotionActive
BMC MotionEnable
Switch Focus
Status Switch
WAC
Safety Circuit Focus
OverTemp

Comparator
Charge And HV Actual
Strobe Value

Limit
Therapy Head Locked Trigger Enable
In Therapy Position Interlock
STHS Safety Logic HV

EOC
Shock Wave Charge On
Hand Switch
Local
Trigger Enable
HV Unit
Unlocking Power
ELMEN Local
Test Overvoltage

Trigger Switch 1
Charge On

Shock Wave
Sel Trig

Release
Trig1 Local
Trig2 Local

Switch Focus
Control
Shock Wave EOC
Hand Switch
Remote HV Set Value

Unlocking SWC
ELMEN Remote HV Set Control Charge and
Microcontroller
Trigger Unit
HV Actual
230VAC PCCU
Shock Wave Value
Release

Trig1 Remote PCCU On


Trig2 Remote
Select Trigger Mode

Trigger Switch 2
Internal Trigger

Sel Trig

SWC Power Trigger Unit


Unit
CAN Bus

Trigger
Trigger Safety Logic
ECG / (Hardware) XRESP
ECG Trig
Resp.
System Resp Trig X-ray Interface (Option)

10-004-02-1005 3-9
3-10
Temperature Overtemp. Power Unit Shockwave 3.8
Switch Safety
Controller
Bistable
SWC
230VAC
Relay

Heating
Element
control control
Reset
Temperature Heating
Water Tank Switch Safety Control
Safety Circuit
Overtemperature
WAC safety circuit

Therapy
Head
Temperature Comparator Overtemp.
R U
Sensor Safety
Reset
Bistable Safety Circuit
Relay
Water Air Controller Vcc
WAC
Switch Water
Tank Min.

Microcontroller

Temp.
R U
Sensor
RC1396AL

10-004-02-1005
3.9 PCCU block diagram

RC1419AL

10-004-02-1005 3-11
3.10 PCCU timing diagram

RC1420AL

3-12 10-004-02-1005
3.11 CTU100 GAL

10-004-02-1005 3-13
3.12 Hydraulic unit

3-14 10-004-02-1005

You might also like