0% found this document useful (0 votes)
143 views15 pages

Impedance Test

The document provides instructions for performing a dummy cell test using a Solartron 1260A frequency response analyzer (FRA) to ensure it is working properly. A 1.8kΩ resistor is connected between the input and output of the FRA. ZPlot software is used to sweep the frequency from 1MHz to 1Hz at 100mV amplitude. The impedance data obtained from the test matches the expected 1.8kΩ value, indicating the FRA is functioning correctly. The dummy cell test helps verify questionable data stems from external factors rather than issues with the FRA itself.

Uploaded by

Noor Raihan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
143 views15 pages

Impedance Test

The document provides instructions for performing a dummy cell test using a Solartron 1260A frequency response analyzer (FRA) to ensure it is working properly. A 1.8kΩ resistor is connected between the input and output of the FRA. ZPlot software is used to sweep the frequency from 1MHz to 1Hz at 100mV amplitude. The impedance data obtained from the test matches the expected 1.8kΩ value, indicating the FRA is functioning correctly. The dummy cell test helps verify questionable data stems from external factors rather than issues with the FRA itself.

Uploaded by

Noor Raihan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 15

Solartron 1260A

RESISTOR IMPEDANCE TEST-ZPLOT SOFTWARE


HTTP://WWW.AMETEKSI.COM/SUPPORT-CENTER/SOLARTRON-ANALYTICAL/USER-TESTS
Why Run a Dummy Cell Test?
Before beginning a dummy cell test, please ensure that the Solartron Analytical 1260A is connected and
communicating with your PC. If you are unsure or need assistance with this, please refer to AMETEK®/
Solartron at: http://www.ameteksi.com/support-center/solartron-analytical/user-tests

Additionally, it is imperative that the latest version of the ZPlot software corresponding to the Solartron
Analytical 1260A be installed onto the connected PC. This can be found at:
http://www.ameteksi.com/products/software/zplot-software.
When using ZPlot, it is also necessary to download ZView, in order to view the data derived from testing.
This can be found at:
http://www.ameteksi.com/products/software/zview-software.
When working with the Solartron Analytical 1260A, as with all Solartron Analytical frequency response
analyzers (FRA), it is important to perform a dummy cell test to ensure that the equipment is working
properly. Running a test with known parameters and expected results can help to determine whether an
FRA that is supplying questionable data requires repair or recalibration.
A properly working FRA that outputs questionable data can indicate any of the following:
• Faulty GPIB connection
• Noise interference (could be from power outlet, or from pumps, heaters, etc.)
• Bad connections to the experiment (i.e.-something may have disconnected mid-experiment)
• Broken or corroded cables
• AC voltage level is set too low
• Insufficient integration time during the experiment
• Measurements taken outside the specified range of the apparatus
Disqualifying the notion that the FRA is responsible is a crucial step to determining the cause of
erroneous data. Running a successful dummy cell test and obtaining experimental results that match
expected results indicates that the FRA is working correctly under that software configuration.
Running a dummy cell test with ZPlot and ZView will be addressed in this document. The following is a
guide which outlines the steps needed to perform this dummy cell impedance test.

1|Page
Experimental Setup-Physical Equipment

In order to run the impedance experiment, a metal wire must be used to connect the shield of the
“INPUT I” and the shield of the “GEN OUTPUT”. This is done to give the current a return path to
complete the circuit. The wire may be insulated, as seen in Figure 1, or can be bare. However, the ends
attached to the 1260A must be free of insulation so that the current can travel freely through the wire.

Not
Insulated

Figure 1

Two BNC coaxial connector adapters, and two BNC T adapters, as shown in Figure 2, respectively, will be
needed to connect the cables to the resistor.

Figure 2

2|Page
The wires are connected as shown in Figure 3. Connect “GEN OUTPUT” and “INPUT V1 HI” to one side of
a resistor, and connect “INPUT I” and “INPUT V1 LO to the other side of the resistor. The known value of
the resistor in this example is 1.8kΩ.

Figure 3

Ensure that the test resistor is connected to the pin side of the BNC adaptor, not the ground side. The
pin is the side without the tab labeled “GND” for “ground”. See Figure 4 below.

Pin: Connect
on this side-
without tab
Ground

Figure 4

3|Page
Experimental Setup & Run-ZPlot Software
Having installed the latest version of ZPlot, double click the ZPlot icon on your desktop to start the
software. The ZPlot software protection key will need to be entered into a USB port on the PC to run this
test. Additionally, a USB-GPIB adaptor will be needed to connect the PC to the instrument.
For more information regarding connections and communications, please refer to the 1260A
Connections/Communications document at http://ameteksi.com/support-center/solartron-
analytical/user-tests.
Select the “Z” icon at the top to open “Setup ZPlot” (Figure 5).

Figure 5

Ensure that Solartron 1260 is selected as the “Analyzer Type” and that 1260 Standalone is selected as
the “Pstat/Gstat or Interface” (Figure 6), and that the GPIB address is correct (agrees with the
instrument address setting). Click “Test GPIB”. If the window displays a “test pass” message, continue.
Otherwise, refer to the 1260A Connections/Communications document found at
http://ameteksi.com/support-center/solartron-analytical/user-tests for further instructions. Click “OK”.

Figure 6

4|Page
Select the “Ctrl E: Sweep Freq” tab (Figure 7).

Figure 7

Leave the applied “DC Potential” at “0”, and change the “AC Amplitude” to 100mV. Keep “Monitor” in
the “off” position. Set the “Initial Frequency” to 1MHz and the “Final Frequency” to 1Hz. Leave the
setting on “Logarithmic” and 10 steps/decade. (Figure 8).

Figure 8

5|Page
Select “Measure, then “Sweep” (Figure 9).

Figure 9

A screen similar to below will appear. This will update the progress of the experiment with regards to
points and time. It will also indicate that the cell is on and currently working while giving real-time
measurements (Figure 10).

Figure 10

6|Page
When the experiment is complete, a dialogue box will appear prompting the user to save the project.
Type a name for the data file, and click “OK” (Figure 11). In this example, the data file is named
“Impedance”.
The data can be viewed while it is being collected by clicking on the “Z-Z” Icon in Figure 11. Autoscale
may be needed to better view the data. Data can also be viewed by using ZView as outlined in the
following pages.

Figure 11

7|Page
Viewing the Data with ZView
Open ZView, and click “File”, then “Data Files” (Figure 12)

Figure 12

Under “Select Data Files”, select the “Impedance” file created as shown in Figure 13.

Figure 13

8|Page
Click the arrow pointing to the right, and select “OK” to add the file to “Files to Plot:” (Figure 14).

Figure 14

This dialogue box will disappear, and the screen will return to that shown in Figure 12.

9|Page
Use the drop-down box on the task bar to select the file created-“Impedance” (Figure 15)

Figure 15

The selected file should now show in the drop-down box as the first item (Figure 16).

Figure 16

Select “Graph”, then “New Bode” as shown in Figure 17 (for example-or choose whichever graph
preferred).

Figure 17

10 | P a g e
A graph such as the one in Figure 18 will appear.

Figure 18

11 | P a g e
In order to make the graph easier to view, click the magnifying glass with the “A” inside to automatically
adjust the scaling within the graphs. This will produce a graph similar to Figure 19.

Use this slide bar to move Exact Data


cursor within graphs Values

Roughly 1.8kΩ

Figure 19

The upper graph shows the impedance value from the experiment is roughly 1.8kΩ. This is what was
expected, as it is the known value of the resistor. Values can also be seen in the upper righthand corner
of the screen. Additionally, the slide bar at the top can be used to control the cursor on the graph, giving
data at each individual point. The data obtained from this experiment suggests that the FRA is working
properly. If other experiments are giving erroneous data, causes outside of an impaired FRA should be
explored.

12 | P a g e
Actual data values can also be obtained by clicking on the “Data Values” icon shown in Figure 20. This
will then display data in the format shown in Figure 21.

Figure 20

Figure 21

Like the graph shown in Figure 19, the data shown in the first column highlighted in Figure 21 indicates
an approximate impedance value of 1.8kΩ. This again confirms the expected resistor value indicating a
properly working FRA. It can also be noted that at low frequency, the phase angle approaches zero,
indicating the presence of the resistor. At high frequency, due to stray capacitance and inductance from
the component under test and from its connections, the phase angle will deviate from zero ohms and
the impedance magnitude will deviate from its low frequency value of 1.8kΩ.

13 | P a g e
141117RevA
|Page

You might also like