CHAPTER 7 INSTALLATION
7.4 Setting....................................................................................................................................... 23
7.4.1 Software settings........................................................................................................... 23
7.4.1.1 System configuration settings (System Configurator) ........................................... 23
7.4.1.2 Common procedure for using Windows to perform tasks...................................... 24
7.4.1.3 Keyboard setting.................................................................................................... 27
7.4.1.4 Touch panel settings [AJ] .................................................................................... 28
7.4.1.5 GP printer settings................................................................................................. 31
7.4.1.6 Installation procedure for online manual................................................................ 31
7.4.1.7 Default rules (not supported) ................................................................................. 32
7.4.1.8 Locale setting [AJ] [BM] [BN]................................................................................. 32
2015011
NOTE: USB keyboard and USB mouse are needed for installation work.
[AJ]
7.1.1 XN-530/550
(1) Remove tape from the front of the instrument.
[AJ] 2015056
(6) Remove the machine screws with captive washers (for thin plate t0.8) M3 x 16 SUS: x2 that fasten
fixture NO. 1657.
(10) Two plates (INTERRUPTER PLATE NO.395) are installed on the bottom of the two sampler
adapters (ADAPTER NO.328) respectively, then set the adapters to the sampler, the installation
positions shown as below for the plate are used for setting the testing tube.[AH]
*The following picture shows the normal tube position is installed by the plate (INTERRUPTER
PLATE NO.395).
7.1.2 XN-430/450
(1) Remove tape from the sample holder.
[AH]2015165
(3) Remove the indicated binder clip NO. 105 and CV-250 bundled wire shipping stopper.
[AN]2016144
Caution: [BM]
Be sure to connect the instrument waste fluid to a waste container in the facility (a
container that has an insertion nipple for the waste tube or other means of fixing the
waste tube so that there is no risk of waste fluid splattering) or an optional waste
tank.
When connected to a waste container in the facility, periodically verify that the
waste tube is securely connected and there is no risk of waste fluid splattering.
[BM] 2017117
04315817 96308015
WIRING CORD NO.3497 WIRING CORD NO.2188
[BA] 2017060
(3) The screen below appears. Set the destination and units, and press the "OK" button.
- Destination : Select EU, JP, AM, or CN.
- Units : Set "Use" or "Do not use" SI units.
Caution:
Set the destination and units settings with care.
If any changes become necessary, initialization of the settings will be required.
(5) Make sure that you have set the configuration as intended, and press the [Close] button.
(6) Splash screen will appear again, then touch the screen after confirming [Start up] and green dots
are shown in the right bottom of the screen.
(7) After beeps, confirm [Service Mode On] shown in the right bottom of the screen.
2015057
(13) Wait until the green part of the display in the lower left corner of the screen changes from blinking
to solidly lit, and "Starting up..." changes to "Enter a sample number".
[AJ] 2015056
(19) Touch the screen when beep sounds once and hold the touching the screen until beep sounds
twice.
(20) Touch “Power” icon at lower left in logon screen.
Check
Yes No
(22) The screen below appears. Enter “sysmex” in the [Login Name] field and enter “c9.0” in the
[Password] field, and press the [login] button.
(23) The dialog box below appears. Press the [Start] button.
Note:
1) This procedure can be used when the analyzer is started up in Japanese, after turning
the analyzer OFF and back ON when Initial setup wizard stays on the back of splash
screen with a destination setting “JP”.
2) This procedure describes initialization process of region settings and stored data by
executing “XNL_Data Initialize.exe.” The program “XNL_Data Initialize.exe” will
initialize only the region setting and stored data. It does not initialize NVRAM file.
Caution: [BP]
Software and Windows settings should not be changed except for certain items
described in this chapter. If changed, it may cause the device to malfunction.
Example: If changing the following item, it was confirmed that the IPU browser
screen program will not work properly.
(Control Panel -> Personalization -> Change the theme)
(2) The screen below appears. Set the destination and units, and press the "OK" button.
- Destination : Select EU, JP, AM, or CN.
- Units : Set "Use" or "Do not use" SI units.
Caution:
Set the destination and units settings with care.
If any changes become necessary, initialization of the settings will be required.
[BP] 2018024
(4) Make sure that you have set the configuration as intended, and press the [Close] button.
(7) The instrument automatically restarts and the screen below appears ("Finish" and "Shutdown"
are disabled).
Wait while the instrument automatically shuts down.
(3) Click the [Add] button in the "Text Services and Input Languages" screen.
(2) Press "Config" in the confirmation dialog box to close the IPU.
[AJ] 2015056
(5) When the "Touch Panel Settings" screen opens, select the "Calibration" tab.
(5) Set the date and time to the current date and time and click OK.
(7) Select "Control Panel" - "Clock, Language, and Region" - "Regional and Language Options".
(9) Select the system locale of the installation location, and click the [OK] button.
(12) The McAfee Solidifier command line (DOS prompt) screen opens.
(13) Input “sadmin eu” from the keyboard, then press the [Enter] key. Text to input:sadmin eu
(14) Enter the password that you set (red part), and hit the return key.
Password: McAfee password(e.g. c9.0) [BG]
[BG] 2018022
(18) The instrument automatically restarts. Wait while the instrument automatically shuts down.
[AJ] 2015056 [BM] 2017117
[BM] 2017117
(3) Enter “sysmex” in the [Logon Name] field and enter service login password in [Password] field,
then press [OK]. [BG]
(7) Wait until the green part of the display in the lower left corner of the screen changes from blinking
to solidly lit, and "Starting up..." changes to "Enter a sample number".
▪ Repeat this procedure to read in all the remaining diluents and hemolytic agents.
04315817
WIRING CORD NO.3497
(2) Connect Main connector on RU-20 rear panel (item 11 in figure below) and RU Supply Tank
connector on XN-L rear panel (item 13 in figure below) with Wiring Cord 8471.
Use either of following cord according to the installation situation.
AA050180 WIRING CORD_ASSY NO. 8471 (13m)
BN528860 WIRING CORD_ASSY NO. 8471 (5m)
[AJ] 2015056
Polyurethane tube
6X4 (1500)
XN‐L rear Panel