EPU User Manual
EPU User Manual
User Manual
PN 1025707
Revision 2.14 • 25-JAN-2022
Contents
1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
1.1 Target audience for this user manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 System and software compatibility. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2 Getting Started. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.1 Prepare for an EPU session. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Start the EPU software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Verify that the microscope is ready for high quality data acquisition. . . . . . . . . . . . 7
2.4 Log in on Thermo Scientific Athena. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3 User Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1 User interface panels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Messages side panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
3.3 Status side panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3.4 Histogram side panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3.5 Image Information side panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 .
3.6 Image and plot display area. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 .
7 Atlas Tab. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
.
7.1 Setup Session task. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 .
7.2 Screening task. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
8 EPU Tab. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
.
8.1 Session Creation task. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95 .
8.2 Session Setup task. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
.
8.3 Square Selection task. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .107
..
8.4 Area Selection task for Lacey Carbon specimens. . . . . . . . . . . . . . . . . . . . . . . . .112 .
8.5 Hole Selection task for Quantifoil specimens. . . . . . . . . . . . . . . . . . . . . . . . . . . . 126 ..
8.6 Template Definition task for Quantifoil specimens. . . . . . . . . . . . . . . . . . . . . . . . .141 .
8.7 Template Execution task for Quantifoil specimens. . . . . . . . . . . . . . . . . . . . . . . . 151 ..
8.8 Automated Acquisition task. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .152
.
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .209
.
Chapter | Introduction
1 Introduction
Cryo-electron microscopy (Cryo-EM) for high resolution three-dimensional single particle
reconstruction (3D-SPR) has reached near atomic resolution for structures of biological complexes of
all kinds.
For the high resolution reconstruction of a particle, large amounts of well-aligned images must be
summed and averaged. Biological material is highly sensitive to electron radiation, so an extremely
low dose must be applied to prevent damage to the specimen. This can result in a low signal-to-noise
ratio in the recorded images. It would be a demanding and time consuming task for a microscopist to
acquire the images that the reconstruction software needs to complete its task with a satisfactory
result. Fortunately, automation software is available to drastically reduce the time and effort, and at
the same time improve the reliability of harvesting such large quantities of high quality data.
EPU is a Thermo Fisher Scientific software product for automated high quality data acquisition in a
Single Particle Analysis (SPA) workflow. SPA is an approach to 3D image creation, during which a
large number of vitrified, low-contrast complexes are imaged under low electron dose conditions.
After conformational classification and particle averaging, this results in a high resolution 3D
representation.
The acronym EPU is from the Latin "E Pluribus Unum" — Out of Many, One. This reflects the actual
pathway: a single 3D particle structure is extracted from numerous 2D images with multiple particles
per image. EPU automates the most time-consuming step in the SPA workflow: the acquisition of
useful images. The 3D reconstruction is done with various open source software applications.
EPU facilitates the following steps in the data acquisition procedure:
● Screening of multiple specimens to select the specimens and specimen areas with the highest
potential for high quality data acquisition.
● Automated or customized selection of the optimal acquisition areas on the specimen.
● High throughput, high resolution data collection.
For online training materials to help you improve your Cryo-EM skills, visit the EM-learning.com
website .
For questions, remarks and support regarding the EPU software, please contact the EPU support
desk at [email protected] .
2 Getting Started
For an extensive preconditions check, see Detailed Preconditions for Successful EPU Usage on
page 170 .
Detailed instructions for the individual alignments and calibrations are available in the online help of
the TEM User Interface.
2.3 Verify that the microscope is ready for high quality data acquisition
In the lower-left corner of the EPU user interface, the Microscope State, also called Traffic Light,
shows if the system is ready for high quality data acquisition. If a subsystem reports an error or is
otherwise not in an optimal status, then the status of that subsystem changes to red and the Recover
function becomes available to return the system to an all-green status.
When a subsystem is in a red state it is often still possible to use EPU, but the image quality may not
be optimal.
On Tundra systems, the automated Recover procedure includes the automated Daily Alignments
procedure.
● The Daily Alignments procedure is executed only when the preceding execution is expired. This
procedure can take up to 30 minutes. The Daily Alignments expire after 24 hours.
● For the best result, make sure that there is no specimen in the stage, or that the beam can pass
through a hole in the specimen.
1. In the lower-right corner of the EPU user interface, select Athena Login
The red dot indicates that there is no active connection to Athena yet.
The Thermo Scientific Athena login screen appears.
3 User Interface
● Tab Selection
The tabs are typically worked through from left to right. Each tab provides a set of tasks.
● Task Selection
Tasks are typically executed from top to bottom. The set of available tasks depends on the
selected tab.
● Task Execution
The content of the Task Execution pane depends on the active task. It can display an input
dialog, one or multiple acquired images, or progress information for an ongoing function.
● Ribbon Bar
The Ribbon Bar offers a set of controls that are necessary or helpful for completing the active
task.
● Side Panels
The Side Panels pane contains a set of collapsible panels. The set of available side panels
depends on the active task and/or the selected image in the Task Execution pane.
● Status Bar
The Status Bar displays the Athena login and status, and the Traffic Light. The Status Bar is only
visible when Athena and/or the Traffic Light functionality are available.
The Status panel displays various types of messages in chronological order, such as:
● Errors and Notifications.
● Progress messages for ongoing automated procedures.
● Intermediate and final results of automated procedures.
● Instructions and recommendations to the user.
When ticked, it is possible to adjust contrast, brightness and gamma for the current image. To reset
contrast, brightness and gamma to their default values, clear Auto Filter and tick it again.
Black Level:
Drag the red line at the left-side of the main histogram to adjust the black level. Pixels with an
intensity below the black level value are displayed with zero intensity (black).
White Level:
Drag the red line at the right-side of the main histogram to adjust the white level. Pixels with an
intensity above the white level value are displayed with maximum intensity (white).
Gamma:
Drag the diagonal black line up or down to adjust the Gamma curve.
If the Black Level and/or White Level are adjusted, then the Gamma curve is scaled proportionally in
the range between the Black Level and White Level values.
Zoom:
In the lower histogram, drag a range to zoom in on a section of the spectrum.
Click outside the zoom range in the lower histogram to reset the zoom level.
Drag the slider with the mouse to change to the zoom level.
Color Enhancement
Applies a color mapping to the intensity values in the image. This makes it easier to recognize
intensity gradients and areas with similar intensity.
The FFT filter optimizes the contrast and brightness of the FFT, so that for example Thon Rings are
shown clearer. The FFT filter does not change the acquired image data for which the FFT is
displayed.
Note This filter is not adjustable and has not been optimized for performance and low resource usage.
Activate the FFT filter only when necessary. Deactivate the FFT filter when there is no direct need to
display the FFT for an image.
Swap the images in the Inset and the Main windows. If the Inset window is hidden, then the Main
window will toggle between the FFT and the normal camera image.
Zoom to fit
Adjust the zoom level, so the entire image fits in the image display frame.
Zoom to fit is also available in the right-click context menu of the image display.
Zoom to 100%
Adjusts the zoom level to 100%, so the image is displayed in actual size.
Zoom 1:1 is also available in the right-click context menu of the image display.
The red crosshair is commonly referred to as the stage position. Although the red crosshair moves
when the stage moves, it does not mark the stage position itself. The red crosshair marks the center
of the field of view if a new image were acquired.
The options in the context menu depend on the active task and image type.
2. Select either:
● Export image
Create a file of the image. The image is saved with the original resolution.
● Export image with overlay
Create a file of the image with scales, markers and other visual aids. The resolution of the
image file is the same as as displayed in EPU. This may be less detailed than the original
image.
4 Home Tab
Note The Home tab and all functionalities of the Home tab are only available on Tundra systems.
The main purpose of the Home tab is to quickly set up an EPU Session, based on pre-determined
settings and values that are stored as a Preference.
● The Advised Preferences contain generic, best-practice settings and values that are proposed by
EPU.
● A Saved Preference contains custom settings and values that have been defined and stored in
an earlier EPU Session.
The Home tab also provides fast access to the EPU User Manual (this document), the Athena portal,
and the OffloadData folder on the Storage Server.
The loaded settings and values are generic, best-practice for the selected selected Experiment Type
and grid. Continue with the EPU workflow to verify and adjust the settings and values where
necessary.
Continue with the EPU workflow to verify and adjust the settings and values where necessary. It is
not possible to update the values in a stored Saved Preference. If adjustments are necessary in the
new experiment, and it is expected that there will be similar experiments in the future, then store the
adjusted values in a new Saved Preference.
5 Preparation Tab
The Preparations tab provides a set of tasks and functionalities to set up the microscope and the
EPU application for successful automated acquisition.
For the Thermo Scientific Falcon 3EC and Falcon 4(i) camera, the Exp. Time (s) is available in the
Exposure Settings.
Camera
Select the camera that is used for the selected Preset.
Binning
Select the sensor-to-image pixel grouping mode.
The options are:
● 1: each pixel in the acquired image corresponds to a single sensor pixel.
● 2: the signal of 4 sensor pixels (2x2) is integrated into a single image pixel.
● 4: the signal of 16 sensor pixels (4x4) is integrated into a single image pixel.
A higher Binning value:
● Does not affect the field of view.
● Decreases the image resolution.
● Increases the image acquisition speed for CCD cameras.
For CMOS cameras the image acquisition speed is practically independent of the Binning value.
CMOS cameras are:
● All FEI and Thermo Scientific Falcon and Ceta cameras.
● Gatan K2, K3 and OneView cameras.
● Increases the signal strength of the image pixels.
Confidential, limited rights EPU User Manual
PN 1025707 | Revision 2.14 | 25-JAN-2022 Page 18
Chapter | Preparation Tab
Readout
Select the area section of the camera sensor that is used for image acquisition.
The options are:
● Full: on a 4096x4096 sensor, all the signal of all pixels is used.
● Half: on a 4096x4096 sensor, the signal of a 2048x2048 area around the center is used.
● Quarter: on a 4096x4096 sensor, the signal of a 1024x1024 area around the center is used.
A smaller Readout value:
● Decreases the field of view.
● Does not affect the image resolution.
● Does not affect the signal strength of the image pixels.
EPU validates the specified value. If necessary, EPU adjusts the specified value to the nearest valid
value and shows a message.
5.1.1.2 Exposure Settings for Thermo Scientific Falcon 3EC and Falcon 4(i) cameras
Mode
Select the integration mode.
When not in the LM magnifications range, the following options are available:
● Linear: the normal integrating mode.
For the Falcon 4(i) camera, the Linear mode is not available in the Data Acquisition preset.
● Counted: Electron Counting mode.
The Counted mode gives better performance of the detector, but requires low dose rates. The
Counted option delivers sub-pixel accuracy without increasing the image size. It describes the
detected electrons by a normalized pattern that is positioned with sub pixel accuracy.
Fractions
Fraction settings are not applied when acquiring images for focusing and other preparatory actions.
Align
Select whether or not the camera frames are aligned before summing.
When Align option Yes is selected:
● The camera calculates the image shift between consecutive frames.
● The camera applies a matching correction to each frame before it is summed into an integrated
image or a Dose Fraction image.
The result of the Align function depends on the selected Fractions mode:
● Fractions is No: the frames are aligned before summing into an integrated image.
● Fractions is Manual or Auto: the frames are aligned before summing into a Dose Fraction image.
After that, the Dose Fractions are aligned before summing into the integrated image. An XML file
with the applied shifts is saved next to the Dose Fraction images.
● Fractions is Maximum: the Dose Fractions are aligned before summing into the integrated image,
and an XML file with the applied shifts is saved next to the Dose Fraction images.
Frames (Nr)
This is not a user setting, it is a calculated value. The number of frames depends on:
● The internal frame rate of the camera.
● The total Exposure time.
Fractions
● Enabled: acquire and store Dose Fractions.
● Disabled: acquire and store only integrated images.
Fractions can only be enabled for the Data Acquisition preset.
Frames (Nr)
This is not a user setting, it is a calculated value. The number of frames depends on:
● The internal frame rate of the camera.
● The total Exposure time.
If the Dose Rate is known, then EPU automatically calculates Dose or Exposure Time. EPU uses the
leading parameter as the basis to calculate the value of the other parameter. In the image below,
Dose is the leading parameter.
Measure
Select Measure to determine the Dose Rate that the camera sensor receives. The color bar
indicates if the measured dose rate is suitable for high quality data acquisition.
● Blue: the Dose Rate is too low. Images may not be usable for 3D reconstruction.
● Green: the Dose Rate is suitable for high quality data acquisition.
● Red: the Dose Rate is too high. The detector is over-exposed.
The measured Dose Rate is only valid for the current Optics Settings. When the Optics Settings are
changed, the Dose Rate must be measured again.
The acceptable Dose range in EPU can be different than the Dose range that is used in Velox. The
range in Velox is based on the technical range of the camera. The value in EPU is based on what a
typical life-science specimen can handle before it becomes severely damaged.
5.1.1.4.1 The automatic Dose Fractions calculation for Falcon 3EC and Falcon 4(i) cameras
The number of frames in each Dose Fraction image is determined as follows:
● Initially, the frames are distributed equally over the Dose Fractions.
● When Fractions is Auto, the average dose in each Dose Fraction must be at least 1 e/px.
This requires that the the Dose Rate is known.
● For the Falcon 3EC camera, when Align is Yes, the number of frames per Dose Fraction in
the initial distribution must be a multitude of 6.
For the Falcon 4(i) camera there is no additional condition when Align is Yes.
● Remaining frames are added to the last Dose Fraction.
The rules above may lead to unexpected fractionation schemes. Run a test acquisition to check the
distribution. For the best results, the Dose Fractions should be as equidistant as possible.
If necessary, adjust the Dose or Exposure Time, or specify a different number of Dose Fractions to
reach a more balanced fractionation scheme.
In the example below, 216 frames are distributed in 20 Dose Fractions. This example does not meet
the optimal equidistant goal:
● Falcon 4(i), or Falcon 3EC with Align is No
● 216 / 20 = 10.8
Each Dose Fraction contains 10 frames.
● The remainder is 216 - (20 ´ 10) = 16 frames.
These are added to the last Dose Fraction, so that the total number of frames in the last
Dose Fraction is 26 frames.
● 216 / 20 = 10.8
Each Dose Fraction contains 6 frames.
● The remainder is 216 - (20 ´ 6) = 96 frames.
These are added to the last Dose Fraction, so that the total number of frames in the last
Dose Fraction is 102 frames.
Noise reduction
Select Yes to decrease the readout noise at low dose. Enabling Noise reduction decreases the
maximum frame rate by up to 50%.
Noise reduction is only available when Frames Summed is 1.
Frames Summed
Specify the number of frames that is summed during acquisition. A higher number of frames
increases the dynamic range in the acquired image.
Mode
Select the integration mode.
When not in the LM magnifications range, the following options are available:
● Linear: the normal integrating mode.
The Linear mode is only used by the Gatan K2 camera.
The Gatan K3 camera always uses Counted mode or Counted/Super Resolution mode.
● Counted: use electron counting.
The Counted mode gives better performance of the detector, but requires low dose rates.
● Counted/Super Resolution:
This mode is only available in the Data Acquisition Preset, and when Fractions (Nr) is set to a
value higher than 1.
In Counted/Super Resolution mode, the position of an electron is assigned to a pixel to achieve
sub-pixel resolution. This information is saved only in the Dose Fraction images. Because of this
additional information, the Dose Fraction images will have double the width and height in pixels.
Number of frames
This is not a user setting, it is a calculated value. The number of frames depends on:
● The internal frame rate of the camera.
● The total Exposure time.
Fractions (Nr)
Specify the number of Dose Fraction images.
The specified value is only applied when acquiring images with the Data Acquisition Preset. When
acquiring a preview image, the Dose Fraction images are saved in the Preparation Preview
sub-folder of the camera root folder.
The beam diameter and magnification are used to calculate the illuminated area on the specimen.
The beam diameter is displayed as an Intensity value. On a microscope with two condenser lenses it
is not possible to calculate the beam diameter or the illuminated area on the specimen.
Note Insert Slit and Slit Width are only available in EFTEM mode on a system with an energy filter.
Get
Imports the Optics Settings parameter values from the microscope.
Set
Applies the Optics Settings parameter values to the microscope.
Probe mode
The microscope software remembers the last used defocus value for MicroProbe and for NanoProbe
mode. When returning to a probe mode, the last used defocus value for that probe mode is
automatically restored. Because of this behavior, some of the Presets must use the same Probe
mode.
To prevent that the result of the Autofocus function is lost, the following Presets and Autofunctions
must use the same Probe Mode as the Data Acquisition Preset:
● Presets: Autofocus, Drift Measurement.
● Autofunctions: Eucentric height by beam tilt, Eucentric height by stage tilt, Autofocus
Magnification
The Magnification determines the field of view, and therefore also the dimensions of the imaged area
per pixel (or pixel size).
Defocus
In most Presets it is useful to apply a small defocus for enhanced contrast. Too much defocus could
invalidate the alignments and calibrations.
EPU applies the specified defocus relative to the current focus, not relative to eucentric focus.
During the preparation of the Automated Acquisition run, the Defocus value in the Data Acquisition
Preset is used for all acquisitions with the Data Acquisition Preset. During the Automated Acquisition
run, EPU uses the Defocus values from the Defocus List.
Spot Size
The Spot Size determines the beam current. A higher Spot Size value corresponds to a lower beam
current, and therefore a lower Dose Rate.
Illuminated Area (Ill. Area)
Only available on microscopes with three condenser lenses.
If the beam diameter is larger than the camera field of view, then parts of the specimen that are not
imaged at that time are needlessly exposed to the electron beam. This may destroy valuable
specimen area, which is then lost for high quality data acquisition.
The ideal beam has the following properties:
● The beam is parallel.
● The beam diameter is just large enough to fully cover the camera field of view.
● The beam has no distortions or aberrations along the edges.
Intensity
Only on microscopes with two condenser lenses.
The Intensity value determines the spreading of the beam. On a system without a C3 lens, the beam
diameter can not be locked to a specific size. If the Probe, Magnification or Spotsize value changes,
then the Intensity value may require adjustment as well.
If the beam diameter is larger than the camera field of view, then parts of the specimen that are not
imaged at that time are needlessly exposed to the electron beam. This may destroy valuable
specimen area, which is then lost for high quality data acquisition.
The ideal beam has the following properties:
● The beam is parallel.
● The beam diameter is just large enough to fully cover the camera field of view.
● The beam has no distortions or aberrations along the edges.
Insert Slit
Only available in EFTEM mode on a system with an energy filter.
Select Yes to insert the slit.
5.1.2.1 Guidelines for the Optics Settings when using Phase Plates
A Volta Phase Plate is generated by exposing the Phase Plate to the beam. This means there are
limitations to the Optics Settings parameters of the Acquisition and Optics Presets when working with
a Phase Plate:
● The beam must be parallel, so that the unscattered beam focuses into a small spot on the Phase
Plate.
● The focus position on the phase plate has to stay within tight limits, which in turn puts tight limits
on the stability of the beam direction. For reference, the stability limits for imaging with a Phase
Plate are far stricter than those for coma free imaging.
In practice, the optics system of a microscope is not ideal. Changing the beam diameter may induce
a slight tilting of the beam, which may cause the beam to partially leave the Phase Plate. To prevent
this effect:
● Always use a parallel beam to illuminate the Phase Plate.
● For a selection of Presets, the Optics Settings must be partially or fully identical.
If a Phase Plate related restriction is applicable to a Preset, then this is described in the chapters
for that Preset.
It is possible to use a convergent beam when working on the microscope, as long as it does not hurt
the existing Phase Plate and does not activate a new one.
5.1.3 The recommended order to define the Acquisition and Optics Presets
The recommended order to define the Presets is not the same as their order in the Presets list:
1. Data Acquisition
2. Hole/EucentricHeight
3. Autofocus
4. Drift Measurement
5. Thon Ring
6. Zero Loss (only for EFTEM mode)
7. GridSquare
8. Atlas
In an Automated Acquisition run, the data acquisition step is the value-creating action. To get the
best quality images for 3D reconstruction, the Data Acquisition Preset must be optimized without
sacrifices to the other Presets.
Although the optics system is highly reproducible, it is always better to avoid changes that are not
strictly necessary. To achieve maximum stability, use the Data Acquisition Preset as the basis for all
other Presets that are used during the Automated Acquisition run.
Camera ● Binning: 1
Settings ● Readout: Full
● On a system without a C3 lens (typically Mid Range systems with Talos software),
verify that the specimen and the Objective aperture are both focused.
If no Objective aperture is inserted, then:
● Select a small Objective aperture.
● Verify that the aperture is focused.
● Return the Objective mechanism to its initial position.
9. Select Optics Settings > Get to import the current optical settings from the microscope.
10. In Camera Settings:
a. Select the Camera
b. Select Binning and Readout
c. If the camera is not a Thermo Scientific Falcon 3EC or Falcon 4(i),
then specify the Exp. Time (s).
For Thermo Scientific Falcon 3EC and Falcon 4(i) cameras, this parameter is specified in
Exposure Settings, after the Dose Rate has been measured.
11. If a Thermo Scientific Falcon 3EC or Falcon 4(i) camera is used,
then select Dose Rate > Measure
5.1.4.1 Guidelines for the Data Acquisition Preset when using Phase Plates on a microscope
with a C3 lens
Microscopes with a C3 lens are typically the High End systems with Titan software.
5.1.4.1.1 Perform the Phase Plate Microprobe (uP) Alignment
On systems with Titan software, perform the Phase Plate Microprobe (uP) alignment procedure:
2. Select Auto help to display detailed instructions for accurate execution of the alignments below.
After the Phase Plate Microprobe (uP) alignment procedure is completed, continue with the Phase
Plate Nanoprobe (nP) alignment procedure.
2. Select Auto help to display detailed instructions for accurate execution of the alignments below.
3. Select the Align PhasePlate > Phase Plate nP alignment.
In this alignment procedure:
a. Very accurately align the diffraction lens focus at the highest camera length.
This alignment ensures that the phase plate is exactly in focus.
b. Very accurately align the beam shift pivot points.
This alignment ensures a stable beam position relative to the phase plate, when an image-
beam shift is applied.
a. Select Active
b. Tick MF-Y Fine focus back-focal plane
8. With the Handpanels > Multifunction Y knob, focus the beam to the smallest possible spot on
the FluScreen.
9. Select Handpanels > Diffraction again to return to imaging mode.
Note During the Automated Acquisition run, make sure that the Phase plate control panel is in Active status.
Note If the spot size and/or illuminated area are changed, the on-plane condition needs to be verified again.
5.1.4.2 Guidelines for the Data Acquisition Preset when using Phase Plates on a microscope
without a C3 lens
Microscopes without a C3 lens are typically the Mid Range systems with Talos software.
5.1.4.2.1 Perform the Phase Plate Nanoprobe (nP) Alignment
Perform the Phase Plate Nanoprobe (nP) alignment procedure.
Note On systems with Titan software,
the Phase Plate nP Alignment must be preceded by the Phase Plate uP Alignment.
2. Select Auto help to display detailed instructions for accurate execution of the alignments below.
3. Select the Align PhasePlate > Phase Plate nP alignment.
In this alignment procedure:
a. Very accurately align the diffraction lens focus at the highest camera length.
This alignment ensures that the phase plate is exactly in focus.
b. Very accurately align the beam shift pivot points.
This alignment ensures a stable beam position relative to the phase plate, when an image-
beam shift is applied.
Optics Settings ● TEM Imaging Mode: the same as the Data Acquisition Preset.
● Illuminated Area: typically ±20 µm.
On Quantifoil samples: twice as large as the Foil Hole interspacing.
● Magnification: low SA range (±3800X),
The Illuminated Area must not be much larger than the field of view.
● Defocus: 10-20 mm.
Defocus helps to enhance contrast, so it is easier to detect the edge of the carbon foil
around a hole.
● Parallel beam.
● Dose must be smaller than 0.1 [e- / Å2sec].
When using Phase Plates, the Hole/EucentricHeight Preset must be defined with an off-
plane illumination, so that an activated Phase Plate is not harmed.
Camera ● Binning: 1
Settings ● Readout: Full
If the FluScreen is not fully illuminated with sufficient intensity, the Dose rate value in the
Microscope User InterfaceTEM User Interface is not accurate.
● On a system without a C3 lens (typically Mid Range systems with Talos software),
verify that the specimen and the Objective aperture are both focused.
If no Objective aperture is inserted, then:
● Select a small Objective aperture.
● Verify that the aperture is focused.
● Return the Objective mechanism to its initial position.
11. Acquire an image with the following camera settings:
● Bias/Gain correction: Bias/Gain
● The same Integration time, Binning and Readout area as used for the live image view.
12. Verify that the acquired image meets the requirements above.
If necessary:
a. Use the Handpanels to adjust Magnification, Intensity and/or Spot Size
b. Acquire a new image and verify it against the requirements above.
13. In EPU:
a. Select Optics Settings > Get to import the current optics values from the microscope.
b. In Camera Settings, select the same Camera, Binning, Readout and Exp. time (s) values
as used for the previously acquired image.
c. Select Acquisition > Preview
14. Verify that the image quality is identical to the previously acquired image.
5.1.5.1 Guidelines for the Hole/EucentricHeight Preset when using Phase Plates
When using Phase Plates, the Hole/EucentricHeight Preset must be defined with an off-plane
illumination, so the currently activated Phase Plate is not harmed.
On Titan systems:
Use NanoProbe and verify that TEM User Interface > Beam Settings control panel > Illumination
is Parallel.
Above a certain illuminated area (~8 μm for a 50 μm C2 aperture) the beam changes from parallel to
spreading.
On Talos systems, verification of the off-plane condition is done in Diffraction mode.
To verify the off-plane condition:
1. Switch to MicroProbe
2. Set sufficient Defocus to enhance contrast.
3. Select Handpanels > Diffraction
4. With the Handpanels > Magnification knob, set the highest camera length.
5. Verify that the diffraction spot size is at least as large as the 40 mm circle on the FluScreen or
Flucam Viewer.
Sample flatness and stability Optics Settings > Camera Settings >
Illuminated Area or Intensity Readout
Worse than average Slightly larger than the Data Acquisition Preset Full
4. Depending on the selected camera, select the same camera-specific values as used in the Data
Acquisition Preset:
● Thermo Scientific Falcon 3EC and Falcon 4:
Duplicate the parameter values in the Exposure Settings section.
● Thermo Scientific Ceta and Gatan K2 / K3:
Duplicate the parameter values in the Advanced Camera Settings section.
Settings that are not used in the Autofocus Preset can be duplicated without negative
consequences. When a value is not applicable to the Autofocus Preset, it will be ignored.
5. Select Acquisition > Preview
The Autofocus Preset cannot be finalized without executing the Autofocus calibration and/or
Autofocus function. Depending on the flatness and stability of the sample, the Camera Settings
and/or the Optics Settings parameters may need adjustment. If adjustments are necessary, this will
be done during Autofocus calibration and/or during Stand-alone execution of the Autofocus function.
5.1.6.1 Guidelines for the Autofocus Preset when using Phase Plates
When using Phase Plates, changes in the Condenser Lens System must be avoided. This makes it
even more important to use the same Optics Settings parameter values as the Data Acquisition
Preset.
Since the Autofocus function acquires images with a tilted beam, it will create satellite spots on the
phase plate film. These spots are sufficiently far removed from the actual Volta area in the optical
center to have a negative effect on the quality of the data acquisitions.
If Phase Plates are used in combination with a Direct Detection camera in electron counting mode,
the beam intensity can be very low and exposure time for data acquisition can be very long.
Nevertheless, Autofocus will work with an integration time that is significantly shorter than the time
that is specified in the Data Acquisition preset. For the Autofocus function a very small dose is
sufficient to reliably perform cross-correlation based shift measurements. The equivalent of a couple
of dose fractions should be sufficient.
The Drift Measurement function tracks the position of recognizable features and patterns in
consecutive images, and compares the detected position shift with a specified threshold value.
The Drift Measurement Preset must meet the following requirements:
● The Optics Settings for the Drift Measurement Preset must be as close to the Optics Settings of
the Data Acquisition as possible.
● The exposure time must be small enough to minimize blur in the images that are used for the
Drift Measurement.
In a blurry image, it is hard to accurately identify recognizable features and patterns.
● The time interval between consecutive images must be short enough to allow for fast
measurements.
● If the Drift Threshold is set very low, the resolution of the acquired images must be high enough
to enable detection of very small shifts.
● For Quantifoil samples: the beam diameter must be small enough, so that the Drift Measurement
Area can be placed close to the Data Acquisition Area(s) without risking multiple exposure of the
Data Acquisition Area(s).
Optics Settings If Phase Plates are used, then the Optics Settings for the Drift Measurement Preset must be
identical to the Data Acquisition Preset, including the on-plane illumination conditions.
If Phase Plates are not used, then the below recommendations are applicable.
● Probe Mode: if the AFIS functionality is present and calibrated, then use the same
Probe Mode as for the Data Acquisition Preset.
● TEM Imaging Mode: same as the Data Acquisition Preset.
● Magnification:
If the drift threshold is very low, it may be necessary to select a higher magnification
than is used for the Data Acquisition preset.
● Illuminated Area / Intensity for Quantifoil samples:
Decrease the beam diameter if possible, but make sure the entire field of view is
illuminated.
Camera ● Binning: 1
Settings Readout: Half.
This increases the readout speed.
The Drift Measurement Preset cannot be finalized without executing the Drift Measurement function.
Depending on the specified Drift Threshold and the measurement results, the Optics Settings and/or
Camera Settings parameters may need adjustment. If adjustments are necessary, this will be done
during Drift Measurement calibration and/or during Stand-alone execution of the Drift Measurement
function.
The recommendations below are specified relative to the Data Acquisition Preset.
Optics Settings ● When using Phase Plates, the Optics Settings for the Thon Ring Preset must be
identical to the Data Acquisition Preset to prevent accidental overexposure of an
activated Phase Plate.
● Spot Size:
● Do not use Spot Size 1 or 2.
● When using the survey camera: select Spot Size 3 or higher.
● When using the high sensitivity camera: check the dose rate to select a suitable
Spot Size.
● Illuminated Area / Intensity:
If desired, slightly decrease the beam diameter to increase the signal strength. When
the beam is too narrow, less Thon Rings will be visible.
● Defocus: typically between -1 mm and -3 mm.
Camera ● Binning: 2
Settings ● Readout: Full
● Binning: 2
● Readout area: Full
c. Start the live image acquisition
d. Move the specimen to an area with thin, uninterrupted, amorphous carbon foil that is not
close to a grid bar.
e. Display the FFT of the live image.
5. Assess the quality of the Thon Rings
If necessary adjust the optics parameters to improve the sharpness of the Thon Rings.
6. Verify that the beam is parallel.
Depending on the system type, either:
● On a system with a C3 lens (typically High End systems with Titan software),
verify that Beam Settings control panel > Illumination is Parallel.
● On a system without a C3 lens (typically Mid Range systems with Talos software),
verify that the specimen and the Objective aperture are both focused.
If no Objective aperture is inserted, then:
● Select a small Objective aperture.
● Verify that the aperture is focused.
● Return the Objective mechanism to its initial position.
7. Acquire an image with the following camera settings:
● Bias/Gain correction: Bias/Gain
● The same Integration time, Binning and Readout area as used for the live image view.
8. Verify that the acquired image meets the requirements above.
If necessary:
a. Use the Handpanels to adjust Magnification, Intensity and/or Spot Size
b. Acquire a new image and verify it against the requirements above.
9. In EPU:
a. Select Optics Settings > Get to import the current optics values from the microscope.
b. In Camera Settings, select the same Camera, Binning, Readout and Exp. time (s) values
as used for the previously acquired image.
c. Select Acquisition > Preview
10. Verify that the image quality is identical to the previously acquired image.
11. In the Image Display, select Show/Hide Inset and/or Swap Inset and Main to display the FFT
for the acquired image.
Optics Settings ● TEM Imaging Mode: same as the Data Acquisition Preset.
● Magnification: LM 400X - LM 600X.
● Illuminated Area: typically ±200 mm.
● Parallel beam.
When using Phase Plates, do not focus the beam to a spot to prevent high doses on the
Phase Plate.
5. Use the handpanels to adjust the Magnification and/or Intensity to meet the requirements for
contrast and brightness, and for the field of view.
This will typically be the case at a Magnification of LM 400X - LM 600X with an Illuminated Area
of 150-200 mm.
6. Verify that the beam is parallel.
Depending on the system type, either:
● On a system with a C3 lens (typically High End systems with Titan software),
verify that Beam Settings control panel > Illumination is Parallel.
● On a system without a C3 lens (typically Mid Range systems with Talos software),
verify that the specimen and the Objective aperture are both focused.
If no Objective aperture is inserted, then:
● Select a small Objective aperture.
● Verify that the aperture is focused.
● Return the Objective mechanism to its initial position.
7. Acquire an image with the following camera settings:
● Bias/Gain correction: Bias/Gain
● The same Integration time, Binning and Readout area as used for the live image view.
8. Verify that the acquired image meets the requirements above.
If necessary:
a. Use the Handpanels to adjust Magnification, Intensity and/or Spot Size
b. Acquire a new image and verify it against the requirements above.
9. In EPU:
a. Select Optics Settings > Get to import the current optics values from the microscope.
b. In Camera Settings, select the same Camera, Binning, Readout and Exp. time (s) values
as used for the previously acquired image.
c. Select Acquisition > Preview
10. Verify that the image quality is identical to the previously acquired image.
Camera ● Binning: 1
Settings ● Readout: Full
11. Verify that the image quality is identical to the previously acquired image.
Export
Writes the current parameter values for all Presets to a file (*.sxml) and save it to an appointed
drive and folder. The Presets file contains:
● The Optics Settings for all Presets.
● The Camera Settings for all Presets.
Depending on the selected camera the Advanced Camera Settings and/or Exposure Settings are
also included.
Import
Overwrite all current parameter values for all Presets with the values from the selected Presets file.
● There is no Undo function. It may be wise to export the current Presets to a file before importing
a different Presets file.
● Presets files that are created with previous software versions are supported with limitations.
When EPU can not import or convert a legacy Presets file it will display an error message.
Note A Presets file contains the parameter values for all Presets. It is not possible to export or import the
settings for a single Preset.
If the image does not meet the requirements for the Atlas Preset, then follow the instructions in
Define the Atlas Preset on page 50 . The requirement that no cut-offs must be visible does not
apply yet.
2. In the acquired image, check for cut-offs.
A Cryobox aperture cut-off is very similar to a C2 aperture or Objective aperture cut-off. It
appears as a dark area with a round edge.
The red dot represents the position of the system's optical axis relative to the center of the
Cryobox aperture.
3. If one or more cut-offs are present:
a. Estimate where the physical center of the Cryobox aperture is.
b. In the image, double-click on the estimated location.
Note The Atlas Optics Alignment calibration is based on the Optics Settings of the Atlas Preset.
If the Optics Settings of the Atlas Preset are changed, then the calculated beam shift value may not be
accurate anymore, and the calibration must be renewed.
1. Find a suitable feature and move it near the center of the field of view:
a. In the Preparation > Acquisition and Optics task,
select Preset Selection > Presets: Atlas
The calibration procedure acquires the first image and displays it on the left side of the Task
Execution panel.
The I0 Calibration is an optional task. If no I0 reference value is available, then the Ice Quality Filter
boundaries use fixed gray scale values. This is fine, as long as the source intensity is kept stable for
the duration of the experiment.
If the specimen is unloaded and reloaded after the Calibrate I0 task is performed, then the I0
reference value measurements during the Automated Acquisition run may be less reliable.
The Move stage here function uses a backlash correction routine to ensure stage position
reproducibility. During the Automated Acquisition run the same backlash correction routine is
used to revisit the selected location for I0 calibration updates.
6. Set the diameter of the I0 region:
a. Double-click on the center of the I0 region.
A circle appears.
b. Increase the diameter of the circle, but make sure it stays at least 250 nm away from the
surrounding carbon foil.
The I0 measurement determines the average signal inside the circular boundary. A larger
diameter increases the reliability of the measurement.
7. Select Get Stage Position and Diameter
8. Either:
● Select Calibrate Current to measure I0 for the GridSquare Preset.
● Select Calibrate All to measure I0 for the GridSquare and the Hole/Eucentric Height
Presets.
The duration of an Automated Acquisition run generally exceeds the lifespan of a Phase Plate area.
During an Automated Acquisition run procedure, EPU selects and activates a new Phase Plate at
regular intervals. A new Phase Plate must be activated to make sure that:
● The next data acquisition starts with a decent phase shift, so that the acquired image has
sufficient contrast for fast, reliable and accurate feature identification.
● The change in phase shift during the first acquisition stays within limits. At the beginning of the
Phase Plate development, phase shift changes rapidly. Activating the new Phase Plate before
acquiring an image stabilizes the phase shift.
To successfully activate new Phase Plates during an Automated Acquisition run, the following
parameters must be specified:
Activation time
For a successful EPU experiment, the Phase Plate should establish a phase shift range of 0.5π ±
0.3π. To achieve a phase shift of 0.5π, the Phase Plate typically needs to be irradiated with an
electron dose of 50 nC – 200 nC. The required activation time can be calculated from the electron
dose number:
t = dose / (beam current)
For example: with a 1 nA beam current, an activation time of 120 seconds is needed to reach 120nC.
Note The Phase Plate activation time is best determined with the Sherpa > AutoCTF function.
For regular users it is possible to check the activation time in the Preparations tab > Activate Phase
Plate task. To do so, follow these steps:
If the selected Objective aperture is not inserted yet, then the Optimize Beam procedure will
automatically insert the selected aperture.
4. Follow the instructions that are displayed.
The Optimize Beam procedure will request to navigate to an area of bare carbon foil. Then the
procedure executes various optimizations.
Confidential, limited rights EPU User Manual
PN 1025707 | Revision 2.14 | 25-JAN-2022 Page 64
Chapter | Auto Functions Tab
After the optimizations are completed, the Optimize Beam procedure updates the Optics Settings in
the Data Acquisition preset.
Note The Optimize Beam task does not update the Optics Settings in Presets that should have the same
Optics Settings as the Data Acquisition preset.
After the algorithm established the initial defocus, the Autofocus function applies the final defocus
value. Which final defocus value is used depends on where in the EPU workflow the Autofocus
function is executed:
● In the Auto Functions tab > Auto-Functions (TEM): Autofocus task,
EPU uses the Auto Function Settings > Desired Defocus value.
● To acquire an image from an Acquisition Area during the Automated Acquisition run,
EPU uses the defocus value from the Defocus list. The Defocus list is specified in the EPU >
Area Selection task for Lacey Carbon specimens, or in the EPU > Template Definition task for
Quantifoil specimens.
● At any other time, EPU uses the defocus value from the active Preset.
2. Select the Preset Selection > Preset that will be used for the stand-alone execution of the
Autofocus function.
3. In the Auto Function Settings section:
To run the Auto-eucentric by beam tilt Auto Function, follow the procedure below:
1. Select Auto Functions > Auto-Functions (TEM) > Auto-eucentric by beam tilt
2. Select the Preset Selection > Presets with which the Auto Function will be executed.
3. In the Auto Function Settings section:
For a description of the parameters, see Run the Autofocus auto-function on page 65 .
a. Specify the Desired Defocus
b. Specify the Iterate to defocus value
c. (Optional) Select Use Three Image Method: Yes to include drift correction. EPU will acquire
an extra image to measure drift.
4. Select Execution > Start
To run the Auto-eucentric by stage tilt auto-function, follow the procedure below:
1. Select Auto Functions > Auto-Functions (TEM) > Auto-eucentric by stage tilt
2. Select the Preset Selection > Presets that will be used to execute the auto-function.
The Auto Functions can be executed with any Preset.
3. In the Auto Function Settings ribbon:
To acquire images with the highest possible quality, data acquisition must be postponed until drift has
decreased to an acceptable level. The Drift stabilization task provides a good estimation for the
Delay after Stage Shift value that EPU uses during the Automated Acquisition run.
Confidential, limited rights EPU User Manual
PN 1025707 | Revision 2.14 | 25-JAN-2022 Page 70
Chapter | Auto Functions Tab
● Quantifoil: Delay after Stage Shift is typically used to postpone the next action after a long move,
for example to a new Grid Square.
● Lacey Carbon: Stage Shift Delay is typically used to postpone high quality data acquisition after
the specimen is moved the next Acquisition Area.
The Drift Measurement function acquires a series of images, and then measures the image shift
between consecutive images to calculate the drift speed.
● If the drift speed decreases below a specified threshold value before it times out, then the Drift
stabilization function reports OK.
● If the drift speed times out before it reaches the specified threshold value, then the Drift
stabilization function reports a failure.
To run the Drift stabilization Auto Function, follow the procedure below:
The Drift stabilization Auto Function will run until either the Time Out is exceeded or the specified
Max. Remaining Drift is reached. The drift measurement results are plotted in a graph. The Max.
Remaining Drift is indicated by the red line.
3. Select Start
The Autofocus calibration calibrates the Autofocus function, and measures and corrects astigmatism.
Note The Autofocus calibration result is stored in the Windows Registry. It can only be completed
successfully by Thermo Fisher Scientific engineers, or after logging in with the Supervisor account.
For the best result, the Autofocus calibration requires an area on the specimen:
● Where no ice is present.
● Where only carbon foil with a decent amount of gold particles is visible.
If the currently loaded specimen does not meet all the above requirements, then please exchange
the specimen. For example, use the Combined Test Specimen (Agar S142) that is delivered with the
microscope. This is a holey carbon foil specimen with gold particles and graphitized carbon.
To prepare for the Autofocus calibration, use the Handpanels and the TEM User Interface to perform
the procedure below:
The Autofocus calibration ignores the Defocus value of the selected Preset.
The Autofocus calibration procedure acquires images with negative beam tilt and positive beam
tilt, and calculates their cross-correlation. The encircled bright spot in the Cross-Correlation
image corresponds to the shift between the beam tilt images.
8. Inspect the last Cross-Correlation image. If necessary, zoom in for a more accurate look.
a. Verify that the green arrows form a symmetric cross with the arms at 90° angles and of
approximately the same length. This indicates that there is no significant astigmatism.
If the arms are not at 90°, or are of different lengths, then correct the astigmatism and
perform the Autofocus calibration again.
b. Verify that the red circle marks a clear and bright spot. This indicates that the shift
between the beam tilt images is measured without problems.
The Autofocus function is not only used in the Automated Acquisition run. It can also be executed as
a stand-alone Auto Function with any Preset.
The result of the Autofocus calibration is coupled to the Probe Mode in which it is executed. If both
Probe Modes are used in the Presets, then the Autofocus calibration must be performed with a
Preset that uses Nanoprobe mode and a Preset that uses Microprobe mode. The procedure is
exactly the same, except for the selected Preset.
If the green arrows in the final Cross-correlation image are not at 90 degree angles relative to each
other, the astigmatism has not been corrected completely.
Although the Autofocus function may work well enough for automated data acquisition, it may be
worthwhile to improve the stigmation and the accuracy of the calibration.
At higher magnifications, it can be difficult or even impossible to use the stage tilt method in an
automated routine, especially when the starting position is too far off eucentric height. Two reasons
are:
● The initial shift between images that are acquired at opposite Alpha tilt positions would be so
large that there is little or no overlap, so no meaningful cross-correlation can be calculated.
Without a meaningful cross-correlation, any adjustments to the CompuStage Z-position would be
guesswork.
Confidential, limited rights EPU User Manual
PN 1025707 | Revision 2.14 | 25-JAN-2022 Page 78
Chapter | Auto Functions Tab
● Due to mechanical tolerances, the CompuStage is not infinitely accurate. At low magnifications,
the field of view of the camera is large enough to absorb these mechanical tolerances, especially
after running a backlash correction routine. As the magnification increases, the field of view of
the camera approaches the same order of magnitude as the tolerances of the CompuStage. This
means it is no longer certain that the region of interest on the specimen is returned to the field of
view after using the Alpha tilt wobbler.
The beam tilt method is the preferred method at higher magnifications, because it can use very small
tilt angles to limit initial image shift and does not suffer from mechanical tolerances.
To get the same result from both methods:
● The beam tilt pivot points must be accurately aligned at eucentric focus height.
● Eucentric focus height and the mechanical eucentric height must be exactly the same.
On an accurately aligned system, the stage tilt method and the beam tilt method result in the same Z-
position. In reality, the result of the beam tilt method can have a small offset relative to the
mechanically defined eucentric height. The purpose of the Eucentric correction calibration is to
measure this offset, so that the Auto-eucentric by beam tilt Auto Function can compensate for it.
The Eucentric correction calibration is optional. If there is a noticeable offset between the results of
the stage tilt method and the beam tilt method, then it is also possible to renew the microscope's lens
series alignments.
1. Select Auto Functions > Auto-Functions (TEM) > Auto-eucentric by stage tilt
2. Select Preset Selection : Presets: Hole/EucentricHeight
3. Select Execution > Start
7 Atlas Tab
An Atlas is an overview of a specimen. An Atlas is a collage of large area images (tiles) to form a
map of the specimen. At the edges of the tiles, small offsets can be visible. This is acceptable.
4. Select the Image format that will be used to store the acquired images.
● MRC: Electron microscopy image format. The MRC format includes an extensive set of
metadata about the microscope and the microscope settings.
See The MRC2014 Image Format on page 197 .
● TIFF: Raster image format: TIFF file format .
5. At Output folder, select [...] and navigate to the target folder.
In the specified folder, EPU creates a sub-folder with the Name of the Atlas session.
Note Do not rename or move the Output folder.
Note If a specimen has already been (partly) processed in an Automated Acquisition run, then do not load an
existing Screening Session file for a new EPU session with that specimen.
The Screening Session file does not contain data about which Grid Squares have been processed
already in a preceding Automated Acquisition run. Processed areas may be too damaged to yield
new high quality data. Revisiting these areas is therefore not a productive use of system time.
1. In the Atlas > Session Setup task, select Session Management > Load Session
2. Navigate to the Output folder of the desired Screening Session and select the
ScreeningSession.dm file.
To acquire Atlases for specimens from multiple Slot Positions, follow the steps below:
● Occupied slot:
At the creation of a new Screening session, EPU copies the slot descriptions that are present
the TEM User Interface > Autoloader control panel. After the Screening session is created,
EPU and the TEM User Interface do not synchronize their slot descriptions.
3. In Atlas Settings:
a. If the currently loaded specimen is included in the set of selected Slot Positions,
then select the Start position. The selected Start Position only applies to the Slot Position of
the currently loaded specimen. For all other Slot Positions, Atlas acquisition starts at Close to
center. :
● Close to center
Atlas acquisition starts close to the center of the specimen.
● Close to current
Atlas acquisition starts close to the current stage position.
b. (Optional) Specify the Number of tiles to restrict the area that will be covered in the new
Atlas. The specified Number of tiles applies to all Slot Positions. It is not possible to specify a
different value per Slot Position.
The green LED shows that the specimen from this Slot Position is currently loaded on the
CompuStage.
● Acquired
● Failed
● Aborted
The Slot Position is highlighted and the Atlas for the specimen appears in the Image Display.
Viewing the Atlas for a specimen does not affect an ongoing acquisition for a different
specimen.
Sometimes the occupation status of the stage is not known. It is unknown if a specimen is currently
present on the stage, or it is unknown from which Slot Position the specimen on the stage was
loaded. This situation can occur when, for example, the cassette is undocked or when the Autoloader
is not initialized.
Under these circumstances, the Single Atlas functionality allows the acquisition of an Atlas from the
unknown specimen. If no specimen is present, then the Atlas will be blank.
The Single Atlas functionality is available in the Screening task as an additional (virtual) Slot Position,
below the regular Slot Positions. Unlike regular Slot Positions, the Single Atlas slot has its own
Acquire function. The Acquire function is only available when the occupation status of the stage is
unknown. An Atlas that is acquired with the Single Atlas functionality behaves the same and is
treated the same as an Atlas that is acquired for a regular Slot Position in a regular Screening
session.
To acquire an Atlas from an unknown specimen on the stage, follow the steps below:
If the Objective Aperture Mechanism is enabled and inserted, then EPU retracts the aperture
before the acquisition starts. After the acquisition is completed, the aperture is inserted
again.
d. Wait until the Single Atlas acquisition is completed.
After a Slot Position is processed by the Screening task, it is possible to reset the status back to the
initial value. This means that the following information is erased or returned to its initial value:
● Slot Position status
● Atlas
● Selected Grid Square categories
The Reset Selected function is not available for the Slot Position from which the specimen is
currently loaded on the stage.
The status, Atlas, name and selection Grid Square categories are erased, or are returned to their
initial values.
1. Select Atlas > Screening to display all Slot Positions in the Autoloader.
2. Select the Slot Position at which the desired specimen is located.
To select a Slot Position, click anywhere except for the checkbox or the description.
The Slot Position is highlighted and the Atlas for the specimen appears in the Image Display.
4. Wait until the loading procedure is completed and the LED is green
If an Atlas exists for the selected Slot Position, then EPU automatically aligns the existing Atlas with
the specimen on the stage. Because there may be small inaccuracies between the aligned Atlas and
the physical position of the specimen on the stage, EPU displays a warning in the upper-left corner of
the Image Display.
If the specimen is unloaded and reloaded after the Calibrate I0 task is performed, then the I0
reference value measurements during the Automated Acquisition run may be less reliable.
● Acquired
● Failed
● Aborted
If the image quality is not good enough, or if stitching the tiles does not result in a good Atlas:
a. Select Stop
This can be done at any time during the Atlas acquisition and may take a few seconds.
b. Select the Preparation > Acquisition and Optics Settings task,
then select Preset Selection > Presets: Atlas
c. Adjust the Optics Settings and or Camera Settings
d. Select Atlas > Screening > Start again to acquire a new Atlas.
All tiles that were acquired before adjusting the Atlas Preset are discarded.
The procedure to unload the current specimen, and to load a new specimen is as follows:
7.2.3 Acquire an Atlas for a single specimen on a system with removable holders
(side entry)
To acquire an Atlas on a system with a removable holder (a so-called side entry system), follow the
instructions below:
● Acquired
● Failed
● Aborted
If the image quality is not good enough, or if stitching the tiles does not result in a good Atlas:
a. Select Stop
This can be done at any time during the Atlas acquisition and may take a few seconds.
b. Select the Preparation > Acquisition and Optics Settings task,
then select Preset Selection > Presets: Atlas
c. Adjust the Optics Settings and or Camera Settings
d. Select Atlas > Screening > Start again to acquire a new Atlas.
All tiles that were acquired before adjusting the Atlas Preset are discarded.
Note If the Optics Settings parameters of the Atlas Preset are changed, and the system has a Cryobox, then
the Atlas Optics Alignment must be performed again.
7.2.4 Pre-select the Grid Squares for the Automated Acquisition run
The Screening functionality assesses the quality of the individual Grid Squares. Grid Squares with
similar characteristics are assigned to the same category. Above the Atlas, a checkbox is visible for
each category. Each checkbox shows the number of Grid Squares in that category.
If present, the Broken Grid Squares category is always red. The other Grid Square categories have
randomly assigned colors. These colors do not indicate the suitability of the Grid Square category for
high quality data acquisition. The colors are help to mark the Grid Squares in each category.
EPU assumes that less than 50% of the Grid Squares is broken. If more than 50% of the Grid
Squares would be marked as broken, than the Broken category is omitted and all Grid Squares are
assigned to the non-broken categories.
In the Screening task, it is possible to make a pre-selection of suitable Grid Squares for the
Automated Acquisition run. If the Atlas from the Screening Session is used for the Automated
Acquisition run, then the EPU > Square Selection task loads the Grid Squares in the selected Grid
Square categories as the initial selection.
To pre-select Grid Squares for the EPU > Square Selection task:
Confidential, limited rights EPU User Manual
PN 1025707 | Revision 2.14 | 25-JAN-2022 Page 92
Chapter | Atlas Tab
1. If the system has an Autoloader, then select a Slot Position with status Acquired.
To select a Slot Position, left-click anywhere, except for the checkbox or the description.
8 EPU Tab
The EPU tab provides a set of tasks to start, prepare and execute Automated Acquisition sessions.
This chapter describes the functions and actions for an EPU experiment with a single specimen. The
single specimen workflow is supported on all system configurations.
If the system has an Autoloader and the EPU Multigrid license is registered and activated, then the
Multigrid functionality is available. With the Multigrid functionality it is possible to first prepare multiple
sessions, and then start the Automated Acquisition for the prepared Sessions in the queue. The
preparation tasks for each specimen in a Multigrid workflow are mostly the same as for the single
specimen workflow.
For instructions for the Multigrid workflow, see: The EPU Multigrid Option on page 161 .
The set of available tasks depends on the Grid type: Holey carbon or Lacey carbon.
The actions in each task depend on the Session type: Automated or Manual target selection.
Automated Manual
Start Session The Session Creation task is generic for both Grid types
Creation and for both Session types.
Preparation Session Setup The Session Setup task is generic for both Grid types
and for both Session types.
Square The Square Selection task is generic for both Grid types
Selection and for both Session types.
Template How to define a Foil Hole Template does not depend on the Session type.
Definition
(Holey)
Automated Manual
Template How to validate the Foil Hole Template does not depend on the Session type.
Execution
(Holey)
Execution Automated The Automated Acquisition task is generic for both Grid types
Acquisition and for both Session types.
In a Manual session, the Area Selection or Hole Selection tasks offer Automated Preparation
functions to apply the manual selection settings from a prepared Grid Square to other Grid Squares.
This saves a large amount of time and effort that would otherwise be spent on manually repeating
the same selection settings for multiple Grid Squares.
3. If no Atlas is available for the currently loaded specimen, then acquire an Atlas.
4. Continue with the tasks in the EPU tab.
For some tasks, the settings are pre-loaded from the EPU Session Preferences file.
The EPU Session Preferences file can contain the following values:
Calibrations: None
Eucentric correction
Screening None
8.1.2 Create a new EPU Session without an EPU Session Preferences file
To setup a new Automated Acquisition session without using a Session Preferences file, follow the
procedure below:
● Yes: create a new EPU Session that uses mostly the same settings and values as the
current EPU Session. For an overview of the copied settings and values, see The EPU
Session Preferences file on page 95 .
● No: create a new EPU Session without pre-selected settings and values.
● Cancel: no new EPU Session is created. The current EPU Session stays as it is.
EPU has two methods to locate and center a position on the specimen:
● Optical centering with Image/Beam Shift:
The specimen remains stationary while the beam shifts to illuminate the feature of interest and
project it in the center of the camera field of view.
Although this is very fast, Image/Beam Shift may induce coma under sub-optimal conditions. The
amount of coma is typically not significant for small offsets, but for larger offsets this may reduce
the image quality.
● Mechanical centering with Stage Shift:
The beam remains stationary while the specimen moves, so that the feature of interest is located
in the center of the camera field of view.
Although the aberrations that are associated with Image/Beam Shift are absent, mechanical
centering is considerably slower than optical centering.
If certain conditions are met, then EPU can use optical centering also for relatively large offsets, and
thus replace most Stage Shifts by Image/Beam Shifts.
These conditions are:
● Aberration Free Image Shift (AFIS) is present, and is calibrated by a Thermo Fisher Scientific
engineer.
● A wide usable Image/Beam Shift range is available:
● The system does not have an Image Corrector.
An Image Corrector restricts the usable Image Shift range.
● Phase Plates are not used.
The use of Phase Plates is not compatible with the beam tilt angles that are needed for coma
correction.
For Lacey Carbon specimens, EPU automatically determines when to use Stage Shift and when to
use Image/Beam Shift to optimize accuracy and throughput.
For Quantifoil specimens, EPU > Session Setup offers a manual selection between two Acquisition
Mode options:
● Accurate Hole Centering
● Faster Acquisition
The Acquisition Modes use different methods to center and visit the Foil Holes for data acquisition.
The execution of the Foil Hole Template is the same for both Acquisition Modes.
Faster Acquisition
In Faster Acquisition mode, EPU processes the Foil Holes in groups.
The Faster Acquisition procedure uses Stage Shifts to center multiple Foil Holes at once as a group.
This reduces the number of Stage Shifts, magnification changes and lens normalizations.
● If the conditions for Faster Acquisition mode are fulfilled,
then EPU uses Image/Beam Shifts to visit the individual Foil Holes within the centered group.
This reduces the number of Stage Shifts even further.
● If the conditions for Faster Acquisition mode are not fulfilled,
then EPU uses Stage Shifts to visit the individual Foil Holes in the centered group.
See The Acquisition Mode for Quantifoil specimens on page 98 for the conditions for Faster
Acquisition mode.
If Image/Beam Shift is used to visit the individual Foil Holes, then the stage remains stationary. As a
result, the metadata of all acquired images in the centered group contains the same stage
coordinates. To separate and group the images per Foil Hole or per Acquisition Area, the file names
of the acquired images must be processed.
The format of the file names is:
FoilHole_[Hole ID]_Data_[Acquisition Area ID]_[date]_[time].mrc
For example: in FoilHole_31545690_Data_31547881_31547882_20210601_0819.mrc
● [Hole ID] is 31545690
● [Acquisition Area ID] is 31547881_31547882
● [date] is 20210601 in yyyymmdd format
● [time] is 0819 in 24-hour hhmm format
In Faster Acquisition mode, the Automated Acquisition run starts with an automatic calibration to
maximize centering accuracy. This calibration takes 2–3 minutes. In a longer automated run, the
calibration duration is easily recovered by the much faster Foil Hole processing method. For short
runs that acquire images from only a few Foil Holes, the Accurate Hole Centering is often faster.
The pixel format of the images that are acquired with the Gatan K2 and K3 cameras depends on the
Acquisition and Optics Presets > Advanced Camera Settings > Mode and the EPU > Session Setup
> Dose fraction output format. The pixel format has a direct impact on the image file sizes.
Counted 8-bit unsigned integer 8-bit unsigned integer 4-bit unsigned integer
+ gain reference image + gain reference image
per Grid Square per Grid Ssquare
Counted / 8-bit unsigned integer 8-bit unsigned integer 4-bit unsigned integer
Super Resolution + gain reference image + gain reference image
per Grid Square per Grid Square
The pixel format specification below applies to Gatan K2 cameras on a system with Titan 3.5 or
earlier, or Talos 2.5 or earlier.
All Gatan Dose Fraction TIFF images use lossless LZW file compression.
If the Microscope PC has a connection to the Athena software on the Data Management Platform
(DMP) Server, then EPU can store the acquired data in an Athena Dataset. Optionally, the Athena
Dataset can be related to a Grid, but this is not required for a successfull EPU session. To select the
Athena Dataset and to optionally link the Dataset to a Grid, the Session Setup > Athena Settings
must be defined.
The Athena Settings section is only visible when the Athena client software is installed on the
Microscope PC. If the Athena Settings are disabled, then first log in on Thermo Scientific Athena.
For instructions, see Log in on Thermo Scientific Athena on page 7 . For a successul Automated
Acquisition run it is not required to select an Athena Dataset.
Note It is not possible to change the Athena Settings during the Automated Acquisition run.
To select an Athena Dataset in the EPU, perform the actions below in the Session Setup task.
Note It is not possible to change the Dataset selection for data that is already acquired.
If the Athena Settings are changed after data has already been acquired, then the new Athena
Settings apply only to the data that is acquired from that moment onwards.
6. (Optionally) Create a link between the Dataset and a specimen:
a. Select Select Grid
b. Select the Sample > select the Grid for which a relation with the Dataset must be created.
It is not possible to add a new Grid in EPU. If the desired Grid is not available,
then select Portal to open the Athena web portal to add a Grid to the Dataset.
7. If the Dataset is incorrectly linked to a Grid,
then select Reset Grid to remove the link.
Removal of the link does not affect the execution of the data acquisition and does not erase any
data that already been acquired.
8. Close the Select Dataset dialog(s).
The Session Setup form displays:
● The Workflow to which the select Dataset belongs.
● The selected Dataset.
● If selected, the Grid to which the selected Dataset is linked.
9. (Optional) If available, tick Enable Quality Monitor to use the EPU Quality Monitor (EQM)
functionality.
If the EPU Session is created from the current EPU Session or from an EPU Session
Preferences file, then several values in the Sesion Setup form may be pre-loaded. All pre-loaded
values can be adjusted.
See The EPU Session Preferences file on page 95 for:
● An overview of the supported values and results.
● The changes in the EPU Workflow when using an EPU Session Preferences file.
● Manual: suitable areas must be hand-picked. Use this mode when the number of areas
from which high quality data must (or can) be acquired is very limited.
The Manual option is not available for EPU Sessions in an EPU Multigrid Queue.
It is not possible to use a mix of automated and manual area selection in a single Automated
Acquisition run. It is possible though to follow-up an automated session by a manual session
or vice versa, as long as the specimen remains loaded and the Atlas is not replaced or
renewed.
d. Select the Acquisition Mode
For a description of the available options, see The Acquisition Mode for Quantifoil specimens
on page 98 .
e. Tick or clear Use Phase Plates
Phase plates can only be used in a Session type: Manual session.
3. If the Athena Settings section is available, then optionally select a Workflow, Dataset and Grid.
See: The Athena settings on page 101 .
4. In the Output section:
a. Select the Image Format.
● MRC: Electron microscopy image format.
The MRC format includes an extensive set of meta-data about the image, the
microscope settings, and the used detector.
For a description of the Thermo Scientific meta-data, see The MRC2014 Image Format
on page 197 .
For a description of the generic meta-data, see MRC/CCP4 file format for images and
volumes .
● TIFF: Generic raster image format.
A description of the format can be found on the web, e.g. at TIFF file format .
Most 3D reconstruction software supports the MRC2014 format. If your reconstruction
software does not support the MRC format, select the TIFF format.
Note When using the Dose Fractions functionality that is available for Direct Detection cameras, the
TIFF image format cannot be used.
b. For systems with a Gatan K2 or K3 camera only: select the Dose fractions output format.
For a description of the options, see: The pixel format for Gatan Dose Fraction images on
page 100 .
c. At Output folder, select [...] and navigate to the target output folder.
All session data and acquired images are stored in the selected folder, except for Dose
Fraction images. Dose Fraction images are saved on the Storage Server (Falcon 3EC and
Falcon 4) or on the Gatan PC (Gatan K2 and K3).
Note Do not rename or move the EPU Session Output folder.
EPU loads the active EPU session at startup. If the folder path is changed, EPU cannot find
the session data. It is not possible to manually load an EPU session from a different location.
d. (Optional) Tick Default folder.
When ticked, the specified Output folder is the default location for succeeding sessions.
5. (Optional) In the Email section:
The Screening Session file does not contain data about which Grid Squares have been processed
already in a preceding Automated Acquisition run. Processed areas may be too damaged to yield
new high quality data. Revisiting these areas is therefore not a productive use of system time.
Once data acquisitions have taken place, the Automated Acquisition session relies on the underlying
Atlas to keep track of the processing progress and status and to maintain data consistency.
The Square Selection task is the same for lacey and holey carbon grids.
● If the specimen has been unloaded and reloaded after the Atlas was acquired, then EPU
automatically re-aligns the Atlas with the physical position and orientation of the specimen on
the stage. Because there may still be small inaccuracies, a warning is displayed in the upper-
left corner of the image display.
3. Customize the selection of Grid Squares from which data must be acquired.
● Use your own judgment to manually select or unselect individual Grid Squares.
Blue and orange Grid Squares are already (partly) processed in an Automated Acquisition
run. They cannot be selected or unselected anymore.
● Use the context menu:
Right-click on a Grid Square, then select Select or Unselect from the context menu.
The processing order is determined on Atlas level, not per individual Atlas Tile.
2. Select Processing Order > Change
3. In the image display, select the Grid Square that must be processed first.
The selected Grid Square is moved to the number 1 position in the processing order. The other
Grid Squares in the Atlas automatically move one position down in the processing order.
4. Select the Grid Square that must be processed next.
This Grid Square is moved to the number 2 position in the overall processing order.
5. Continue selecting Grid Squares in the desired processing order.
After the Grid Squares with the best chances of high quality data yield are re-assigned to the
start o fthe processing order, then it is not necessary to re-assign all remaining Grid Squares.
6. Select Change again to leave the processing order change mode.
EPU switches to the Area Selection task for the selected Grid Square.
d. If the specimen is already at eucentric height, or if there is a different reason to skip the Auto
Eucentric function, then select Acquire to only acquire a new Grid Square image.
If necessary, EPU first moves the specimen, so that the center of the selected Grid Square is
in the center of the field of view.
While the Auto Eucentric function runs, the Auto Eucentric button changes to Stop Auto
Eucentric. If selected, the Auto Eucentric function is aborted and the specimen returns to its
previous Z position.
The Auto Eucentric function includes the acquisition of a new Grid Square image.
2. If the particles all have a similar orientation in the ice layer:
a. Adjust the A-tilt angle of the specimen.
A different A-tilt axis can be applied for each Grid Square to acquire images from various
incident angles. EPU stores the A-tilt angle of the Grid Square during Hole Selection, and
applies the A-tilt angle during Automated Acquisition at the same Grid Square. This typically
results in a higher quality 3D reconstruction.
b. Select Acquire to update the Grid Square image.
On Tundra systems it is not possible to adjust the A-tilt angle of the stage.
3. (Optional) Modify the Defocus List
The Defocus List is applied to all Acquisition Areas on the specimen.
For instructions, see Define the Defocus List for a Lacey Carbon specimen on page 125 .
4. Generate a pattern of Acquisition Areas:
a. Specify the Spacing value.
This is the Center-to-Center distance of adjacent Acquisition Areas.
The Spacing value can not be smaller than the diameter of the Acquisition Area.
● On microscopes with a Three Lens Condenser system, the size of the Acquisition Area
is defined by the Illuminated Area which is specified in the Data Acquisition Preset.
● On microscopes with a Two Lens Condenser system, the Acquisition Area is defined as
the smallest circle that fully covers the camera field of view at the Data Acquisition
Preset magnification. The beam is typically slightly wider than the camera field of view.
b. Select Generate Pattern.
EPU draws hexagonal pattern of Acquisition Areas across the entire Grid Square image.
Depending on the microscope type and configuration, the pattern contains circles and
rectangles. Rectangles mark the field of view of the camera. Circles mark the beam
diameter.
● On microscopes with a Three Lens Condenser system, the size of the Acquisition Area
is defined by the Illuminated Area which is specified in the Data Acquisition Preset.
● On microscopes with a Two Lens Condenser system, the Acquisition Area is defined as
the smallest circle that fully covers the camera field of view at the Data Acquisition
Preset magnification.
c. If the Acquisition Areas are too close to each other or too far apart,
then adjust Spacing and select Generate Pattern again.
5. Set or adjust the Ice Quality Filter:
a. Select Preset Ice Filter to reset the Ice Quality Filter boundaries.
● Drag the red lower boundary marker to exclude Acquisition Areas that are too dark.
● Drag the red upper boundary marker to exclude Acquisition Areas that are too bright.
After every change, the Ice Quality filter updates the selection of suitable Acquisition Areas
immediately. Depending on the number of areas in the Grid Square image, this may take
some time.
6. (Optional) Select Remove Areas Close To Grid Bar to exclude Acquisition Areas near the edge
of the Grid Square.
If the result of the Remove Areas Close to Grid Bar filter is not satisfactory, then select the filter
button again to de-activate it. This will automatically generate a new pattern and apply the Ice
Quality filter.
After the suitable Grid Squares are selected in the Square Selection task, the procedure to handpick
the target Acquisition Areas is as follows:
1. Select the Grid Squares from which data will be acquired:
a. Select the Square Selection task.
b. Select Processing Order > Show
EPU switches to the Area Selection task for the selected Grid Square. The lower-left corner
of the Area Selection Image Display displays the processing order number of the Grid
Square and Z-position of the selected Grid Square.
2. Prepare the first selected Grid Square:
a. Select the Area Selection task.
If a Grid Square image is available, then the image display shows the selected Grid Square.
The lower-left corner of the image display shows the Stage Z-position for the displayed Grid
Square.
● White: the Eucentric Height value is not determined yet.
● Green: the Eucentric Height value is determined.
b. If the specimen is already at eucentric height, or if there is a different reason to skip the Auto
Eucentric function, then select Acquire to only acquire a new Grid Square image.
If necessary, EPU first moves the specimen, so that the center of the selected Grid Square is
in the center of the field of view.
While the Auto Eucentric function runs, the Auto Eucentric button changes to Stop Auto
Eucentric. If selected, the Auto Eucentric function is aborted and the specimen returns to its
previous Z position.
The Auto Eucentric function includes the acquisition of a new Grid Square image.
c. Define the Acquisition Area pattern and filter settings.
For instructions, see Define the Acquisition Area Pattern and Filter Settings for a Manual
Selection Session on page 118 .
d. (Optional) Customize the selection of target Acquisition Areas.
For instructions, see Customize the selection of Acquisition Areas on page 121 .
3. Prepare the other selected Grid Squares:
a. (Optional) Use the Automated Preparation functions to apply the Acquisition Area pattern
and filter settings of the current Grid Square to all other selected Grid Squares.
For instructions, see Automatically create a Manual Selection in all selected Grid Squares on
page 120, Automatically create a Manual Selection in all selected Grid Squares on page
135 .
b. Select Navigate > Next Square to move to the next Grid Square.
In the Square Selection image display, the white border moves to the next Grid Square. The
specimen does not move until Auto Eucentric or Acquire is selected.
c. If desired or necessary:
● Define or adjust the Acquisition Area pattern and filter settings.
For instructions, see Define the Acquisition Area Pattern and Filter Settings for a Manual
Selection Session on page 118 .
● Customize the selection of target Acquisition Areas.
For instructions, see Customize the selection of Acquisition Areas on page 121 .
4. (Optional) Inspect or adjust the Acquisition Area selection for a prepared Grid Square:
Either:
● Select Navigate > Previous Square until the desired Grid Square is reached.
In the Square Selection image display, the white border moves to the previous Grid Square.
The specimen does not move until Auto Eucentric or Acquire is selected.
● Select the Square Selection task
> Right-click on the desired Grid Square
> Select Select Areas
EPU displays the Grid Square image with the Acquisition Area selection for the selected Grid
Square.
Note Unless the current selection of Acquisition Areas must be cleared:
• Do not select Acquire
• Do not select Generate
• If Remove Areas Close to Grid Bar is active, then do not select Remove Areas Close to Grid Bar
8.4.2.1 Define the Acquisition Area Pattern and Filter Settings for a Manual Selection Session
The Spacing value can not be smaller than the diameter of the Acquisition Area.
● On microscopes with a Three Lens Condenser system, the size of the Acquisition Area
is defined by the Illuminated Area which is specified in the Data Acquisition Preset.
● On microscopes with a Two Lens Condenser system, the Acquisition Area is defined as
the smallest circle that fully covers the camera field of view at the Data Acquisition
Preset magnification. The beam is typically slightly wider than the camera field of view.
b. Select Generate Pattern.
EPU draws hexagonal pattern of Acquisition Areas across the entire Grid Square image.
Depending on the microscope type and configuration, the pattern contains circles and
rectangles. Rectangles mark the field of view of the camera. Circles mark the beam
diameter.
● On microscopes with a Three Lens Condenser system, the size of the Acquisition Area
is defined by the Illuminated Area which is specified in the Data Acquisition Preset.
● On microscopes with a Two Lens Condenser system, the Acquisition Area is defined as
the smallest circle that fully covers the camera field of view at the Data Acquisition
Preset magnification.
c. If the Acquisition Areas are too close to each other or too far apart,
then adjust Spacing and select Generate Pattern again.
3. Set or adjust the Ice Quality Filter:
a. Select Preset Ice Filter to reset the Ice Quality Filter boundaries.
● Drag the red lower boundary marker to exclude Acquisition Areas that are too dark.
● Drag the red upper boundary marker to exclude Acquisition Areas that are too bright.
After every change, the Ice Quality filter updates the selection of suitable Acquisition Areas
immediately. Depending on the number of areas in the Grid Square image, this may take
some time.
4. (Optional) Select Remove Areas Close To Grid Bar to exclude Acquisition Areas near the edge
of the Grid Square.
If the result of the Remove Areas Close to Grid Bar filter is not satisfactory, then select the filter
button again to de-activate it. This will automatically generate a new pattern and apply the Ice
Quality filter.
Note When the Generate Pattern function is executed, all Acquisition Areas that match with the Ice
Quality Filter are automatically included in the target area selection.
Any previous customization of the target area selection is cleared.
5. (Optional) Customize the initial selection of target Acquisition Areas as described in Customize
the selection of Acquisition Areas on page 121 .
Note The Filter Ice Quality boundaries and the Remove Areas Close to Grid Bar status are not stored for
each Grid Square.
The settings of the Filter Ice Quality and the status of the Remove Areas Close to Grid Bar remain in
their last adjusted values. When a Grid Square is displayed for which the selection is already
completed, then the displayed filter settings may be different than the filter settings that were used to
select the Acquisition Areas in the displayed Grid Square.
To apply the currently displayed filter settings to all Grid Squares, select Prepare All Squares or
Redo Preparation as described in Automatically create a Manual Selection in all selected Grid
Squares on page 120, Automatically create a Manual Selection in all selected Grid Squares on page
135 .
The Automated Preparation functions significantly decrease the time and effort that is needed to
manually select the acquisition targets (Foil Holes or Acquisition Areas) for an Automated Acquisition
run. After the suitable Grid Squares are selected and the first Grid Square is prepared, Automated
Preparation offers the following functions:
If Prepare All Squares has been executed at least once in the EPU session, then the following
functions can be used:
● Redo Preparation:
Use Redo Preparation when the existing target selection must be renewed after the parameters
and/or filters are adjusted.
● In already prepared Grid Squares, Redo Preparation clears the target selection, then creates
a new selection. The Grid Square image and Eucentric Height value are maintained.
● For not yet prepared Grid Squares the Redo Preparation function performs the same actions
as the initial Prepare All Squares execution.
● Select Prepare All Squares again to prepare newly selected Grid Squares, or to complete an
aborted preparation run. Prepare All Squares uses the current selection and filter values.
● Already prepared Grid Squares remain as they are. If the selection and filter values have
changed, then the existing selection is not changed.
● For not yet prepared Grid Squares the initial Prepare All Squares actions are performed.
Before starting the manual selection, check if the manual selection tools are available in the Ribbon
Bar.
Note The following steps must be repeated for each individual selected Grid Square.
1. Perform all the steps that are described in Define the Acquisition Area Pattern and Filter Settings
for an Automated Selection Session on page 112 .
Note Make sure the specimen is accurately set to eucentric height.
The positions of the selected areas are stored as (X,Y,Z) coordinates. If the specimen is not at
eucentric height during manual selection, the X and Y coordinates of the selected areas may
have an offset. This may cause failure of the Autofocus and/or Eucentric Correction functions
during the Automated Acquisition run, which may negatively impact the amount and quality of the
acquired data.
2. Create a selection of Acquisition Areas.
Either:
● Select the suitable areas in the generated pattern.
● Unselect the bad areas in the generated pattern.
● Select the bad areas in the generated pattern, then invert the selection.
● Create a selection of suitable areas without using a generated pattern.
Instructions for each of these selection procedures are described below.
If the Ice Quality Filter boundaries are changed after a manual selection has been made, then the
selected set is not updated. Selected areas that fall outside the new boundaries will not be removed
from the selection. Likewise, areas that are not part of the selection will not be added to selection if
their average gray scale value now falls inside the new boundaries.
● To add a single area to the selected set, right-click on an area, then select Add from the
context menu.
● To add one or more separate areas to the selected set, hold the Control key and select the
individual suitable areas.
● To add multiple adjacent areas, select Selection Brush, hold the Control key and swipe
across the region of suitable areas.
8.4.2.3.3 Select the bad areas in the generated pattern, then invert the selection
When the bad areas in the generated pattern are easier to identify than the suitable areas:
1. Select Unselect All to clear the selected set.
2. Select the bad areas as if they were added to the selection:
● To add a single area to the selected set, right-click on an area and select Add from the
context menu.
● To add one or more separate areas to the selected set, hold the Control key and select the
individual areas.
● To add multiple adjacent areas, select Selection Brush, hold the Control key and swipe
across the region of bad areas.
3. Select Invert
8.4.2.3.4 Add and remove acquisition areas with the Selection Brush
The instruction in Add and remove Foil Holes with the Selection Brush on page 137 are applicable for
a Quantifoil specimen. These instructions are also applicable for the generated pattern of Acquisition
Areas on a Lacey Carbon specimen.
To decide whether or not to include an Acquisition Area in the selection, a closer look may be needed
to assess particle content, ice thickness, charging, etcetera.
1. In the Grid Square image, right-click on the desired Acquisition Area
and either select:
● Move stage to location to center the Acquisition Area in which the right mouse button was
clicked.
● Move stage here to center the exact position at which the right mouse button was clicked.
2. Select TEM User Interface > CCD/TV Camera control panel > Acquire to acquire a single
image.
To avoid unnecessary exposure:
● Do not select Search.
● Do not use the FluScreen to view the specimen.
3. If necessary, change magnification and/or illumination, and acquire a new image.
Note Every exposure causes damage to the specimen.
To maintain the highest quality, try to keep the accumulated dosage to a minimum.
a. Select Recurrence:
● Never: Do not perform the Autofocus function at all.
● Always: Autofocus is performed at every Foil Hole.
● After Distance: Autofocus is performed only when the distance between the current Foil
Hole and the most recent Autofocus location is larger than a specified value.
● After Centering: Autofocus is performed after a cluster of Foil Holes is centered.
Only available when Session Setup > Acquisition Mode is Faster Acquisition.
b. If Recurrence is set to After Distance, then specify the Distance.
c. Select Focus using:
● Objective Lens: EPU adjusts the focal plane to the Z-position of the specimen.
● Stage Z axis: EPU adjusts the Z-position of the specimen to the focal plane.
2. In the Data Acquisition Settings section of the Ribbon Bar,
specify the Delay after Image Shift and Delay after Stage Shift:
It is not possible to modify the Defocus list while an Automated Acquisition run is ongoing or paused.
To modify the Defocus list, the run must first be stopped.
EPU switches to the Holes Selection task for the selected Grid Square.
d. If the specimen is already at eucentric height, or if there is a different reason to skip the Auto
Eucentric function, then select Acquire to only acquire a new Grid Square image.
If necessary, EPU first moves the specimen, so that the center of the selected Grid Square is
in the center of the field of view.
While the Auto Eucentric function runs, the Auto Eucentric button changes to Stop Auto
Eucentric. If selected, the Auto Eucentric function is aborted and the specimen returns to its
previous Z position.
The Auto Eucentric function includes the acquisition of a new Grid Square image.
2. If the particles all have a similar orientation in the ice layer:
a. Adjust the A-tilt angle of the specimen.
A different A-tilt axis can be applied for each Grid Square to acquire images from various
incident angles. EPU stores the A-tilt angle of the Grid Square during Hole Selection, and
applies the A-tilt angle during Automated Acquisition at the same Grid Square. This typically
results in a higher quality 3D reconstruction.
b. Select Acquire to update the Grid Square image.
On Tundra systems it is not possible to adjust the A-tilt angle of the stage.
3. Specify the Foil Hole diameter and center-to-center interspacing:
a. Select Measure Hole Size.
a. If the Foil Holes appear darker than the surrounding carbon foil, then first select Allow Dark
Foil Holes.
By default, the Hole Detection algorithm searches for bright circles in a darker environment.
If the Allow Dark Foil Holes toggle in active, then the algorithm searches for dark circles in a
brighter environment.
Allow Dark Foil Holes is not active: Allow Dark Foil Holes is active:
EPU does not recognize the real Foil Holes. EPU finds the real Foil Holes.
The Hole Detection algorithm searches for Foil Holes that match with the specified diameter
and interspacing, and generates a pattern of Foil Hole outlines in the Grid Square image.
c. Verify that the generated pattern matches with the physical Foil Holes in the Grid Square
image.
If necessary, adjust the diameter and interspacing and select Find Holes again.
5. Set or adjust the Ice Quality filter:
a. Select Preset Ice Filter to reset the Ice Quality filter boundaries.
● Drag the red lower boundary marker to exclude Foil Holes that are too dark.
● Drag the red upper boundary marker to exclude Foil Holes that are too bright.
After every change, the Ice Quality filter updates the target Foil Holes selection.
6. (Optional) Select Remove Holes Close To Grid Bar to exclude Foil Holes near the edge of the
Grid Square.
Foil Holes that are near the Grid Bars are excluded from the selection.
If the result of the Remove Holes Close to Grid Bar filter is not satisfactory, then select the filter
button again to de-activate it. This will automatically run the Find Holes function and apply the Ice
Quality filter.
After the suitable Grid Squares are selected in the Square Selection task, the procedure to handpick
the target Foil Holes is as follows:
1. Select the Grid Squares from which data will be acquired:
a. Select the Square Selection task.
b. Select Processing Order > Show
EPU switches to the Hole Selection task for the selected Grid Square. The lower-left corner
of the Hole Selection Image Display displays the processing order number of the Grid
Square and Z-position of the selected Grid Square.
2. Prepare the first selected Grid Square:
If necessary, EPU first moves the specimen, so that the center of the selected Grid Square is
in the center of the field of view.
While the Auto Eucentric function runs, the Auto Eucentric button changes to Stop Auto
Eucentric. If selected, the Auto Eucentric function is aborted and the specimen returns to its
previous Z position.
The Auto Eucentric function includes the acquisition of a new Grid Square image.
c. Define the Foil Hole dimensions and the filter settings.
For instructions, see Define the Foil Hole dimensions and Filter Settings for a Manual
Selection Session on page 132 .
d. Customize the selection of target Foil Holes.
For instructions, see Customize the selection of Foil Holes on page 136 .
3. Prepare the other selected Grid Squares:
a. (Optional) Use the Automated Preparation functions to apply the Acquisition Area pattern
and filter settings of the current Grid Square to all other selected Grid Squares.
For instructions, see Automatically create a Manual Selection in all selected Grid Squares on
page 120, Automatically create a Manual Selection in all selected Grid Squares on page
135 .
b. Select Navigate > Next Square to move to the next Grid Square.
In the Square Selection image display, the white border moves to the next Grid Square. The
specimen does not move until Auto Eucentric or Acquire is selected.
c. Repeat steps a – e above until all Grid Squares are prepared.
4. (Optional) Inspect or adjust the Acquisition Area selection for a prepared Grid Square:
Either:
● Select Navigate > Previous Square until the desired Grid Square is reached.
In the Square Selection image display, the white border moves to the previous Grid Square.
The specimen does not move until Auto Eucentric or Acquire is selected.
● Select the Square Selection task
> Right-click on the desired Grid Square
> Select Select Holes
EPU displays the Grid Square image for the selected Grid Square.
Note Unless the current selection of target Foil Holes must be cleared:
• Do not select Acquire
• Do not select Find Holes
• If Remove Holes Close to Grid Bar is active, then do not select Remove Holes Close to Grid Bar
8.5.2.1 Define the Foil Hole dimensions and Filter Settings for a Manual Selection Session
In a Manual Selection Session, the actions below must be repeated for each selected Grid Square.
a. If the Foil Holes appear darker than the surrounding carbon foil, then first select Allow Dark
Foil Holes.
By default, the Hole Detection algorithm searches for bright circles in a darker environment.
If the Allow Dark Foil Holes toggle in active, then the algorithm searches for dark circles in a
brighter environment.
Allow Dark Foil Holes is not active: Allow Dark Foil Holes is active:
EPU does not recognize the real Foil Holes. EPU finds the real Foil Holes.
The Hole Detection algorithm searches for Foil Holes that match with the specified diameter
and interspacing, and generates a pattern of Foil Hole outlines in the Grid Square image.
c. Verify that the generated pattern matches with the physical Foil Holes in the Grid Square
image.
If necessary, adjust the diameter and interspacing and select Find Holes again.
3. Set or adjust the Ice Quality filter:
a. Select Preset Ice Filter to reset the Ice Quality filter boundaries.
● Drag the red lower boundary marker to exclude Foil Holes that are too dark.
● Drag the red upper boundary marker to exclude Foil Holes that are too bright.
After every change, the Ice Quality filter updates the target Foil Holes selection.
4. (Optional) Select Remove Holes Close To Grid Bar to exclude Foil Holes near the edge of the
Grid Square.
Foil Holes that are near the Grid Bars are excluded from the selection.
If the result of the Remove Holes Close to Grid Bar filter is not satisfactory, then select the filter
button again to de-activate it. This will automatically run the Find Holes function and apply the Ice
Quality filter.
Note When the Find Holes function is performed, all Foil Holes that match with the Ice Quality Filter are
automatically included in the target Foil Hole selection.
Any previous customization of the target Foil Hole selection is cleared.
5. (Optional) Customize the initial selection of target Foil Holes as described in Customize the
selection of Foil Holes on page 136 .
Note The Filter Ice Quality boundaries and the Remove Holes Close to Grid Bar status are not stored for
each Grid Square.
The settings of the Filter Ice Quality and the status of the Remove Holes Close to Grid Bar remain in
their last adjusted values. When a Grid Square is displayed for which the selection is already
completed, then the displayed filter settings may be different than the filter settings that were used to
select the Foil Holes in the displayed Grid Square.
To apply the currently displayed filter settings to all Grid Squares, select Prepare All Squares or
Redo Preparation as described in Automatically create a Manual Selection in all selected Grid
Squares on page 120, Automatically create a Manual Selection in all selected Grid Squares on page
135 .
The Automated Preparation functions significantly decrease the time and effort that is needed to
manually select the acquisition targets (Foil Holes or Acquisition Areas) for an Automated Acquisition
run. After the suitable Grid Squares are selected and the first Grid Square is prepared, Automated
Preparation offers the following functions:
If Prepare All Squares has been executed at least once in the EPU session, then the following
functions can be used:
● Redo Preparation:
Use Redo Preparation when the existing target selection must be renewed after the parameters
and/or filters are adjusted.
● In already prepared Grid Squares, Redo Preparation clears the target selection, then creates
a new selection. The Grid Square image and Eucentric Height value are maintained.
● For not yet prepared Grid Squares the Redo Preparation function performs the same actions
as the initial Prepare All Squares execution.
● Select Prepare All Squares again to prepare newly selected Grid Squares, or to complete an
aborted preparation run. Prepare All Squares uses the current selection and filter values.
● Already prepared Grid Squares remain as they are. If the selection and filter values have
changed, then the existing selection is not changed.
● For not yet prepared Grid Squares the initial Prepare All Squares actions are performed.
Before starting the manual selection, check if the manual selection tools are available in the Ribbon
Bar.
Note The following steps must be repeated for each individual selected Grid Square.
1. Perform all the steps that are described in Define the Acquisition Area Pattern and Filter Settings
for an Automated Selection Session on page 112 .
Note Make sure the specimen is accurately set to eucentric height.
The positions of the selected areas are stored as (X,Y,Z) coordinates. If the specimen is not at
eucentric height during manual selection, the X and Y coordinates of the selected areas may
have an offset. This may cause failure of the Autofocus and/or Eucentric Correction functions
during the Automated Acquisition run, which may negatively impact the amount and quality of the
acquired data.
2. Create a selection of Foil Holes.
Either:
● Select the suitable Foil Holes in the generated pattern.
● Unselect the bad Foil Holes in the generated pattern.
● Select the bad Foil Holes in the generated pattern, then invert the selection.
● Create a selection of suitable Foil Holes without using a generated pattern.
Instructions for each of these selection procedures are described below.
If the Ice Quality Filter boundaries are changed after a manual selection has been made, then the
selected set is not updated. Selected areas that fall outside the new boundaries will not be removed
from the selection. Likewise, areas that are not part of the selection will not be added to selection if
their average gray scale value now falls inside the new boundaries.
8.5.2.3.1 Select the suitable Foil Holes in the generated pattern
When the number of bad Foil Holes is larger than the number of suitable Foil Holes:
1. Select Unselect All to clear the selected set.
2. Add the suitable Foil Holes to the selection:
● To add a single Foil Hole to the selected set, right-click on the Foil Hole and select Add from
the context menu.
● To add one or more separate Foil Holes to the selected set, hold the Control key and select
the individual suitable Foil Holes.
● To add multiple adjacent Foil Holes, select Selection Brush, hold the Control key and swipe
across the region of suitable Foil Holes.
8.5.2.3.3 Select the bad Foil Holes in the generated pattern, then invert the selection
When the bad Foil Holes in the generated pattern are easier to identify than the suitable Foil Holes:
1. Select Unselect All to clear the selected set.
2. Select the bad Foil Holes as if they were added to the selection:
● To add a single Foil Hole to the selected set, right-click on an Foil Hole and select Add from
the context menu.
● To add one or more separate Foil Holes to the selected set, hold the Control key and select
the individual Foil Holes.
● To add multiple adjacent Foil Holes, select Selection Brush , hold the Control key and
swipe across the region of bad foil holeFoil Holes.
3. Select Invert.
8.5.2.3.4 Add and remove Foil Holes with the Selection Brush
1. From the Selection section of the the Ribbon Bar, select Selection Brush.
Foil Holes that are touched by the brush have a purple highlight.
c. Release the left mouse button
The purple Foil Holes are removed from the selection.
5. To add Foil Holes to the selection:
a. Hold down the Control key.
b. Hold down the left mouse button
c. Drag the Selection Brush across the image.
d. Release the left mouse button
The purple Foil Holes are added to the selection.
To decide whether or not to include a Foil Hole in the selection, a closer look may be needed to
assess particle content, ice thickness, charging, etcetera.
1. In the Grid Square image:
a. Right-click on the Foil Hole you wish to inspect.
b. Select one of the Move stage... options:
● Move stage to location: centers the Foil Hole in which the right mouse button was
clicked.
● Move stage here: centers the exact position at which the right mouse button was
clicked.
● Move stage here and acquire: centers the exact position at which the right mouse
button was clicked, and acquires a new Grid Square image.
To prevent clearing the current selection, this option is only available when no Foil Holes
have been selected.
2. Select TEM User Interface > CCD/TV Camera control panel > Acquire to acquire a single
image.
To avoid unnecessary exposure:
● Do not select Search.
● Do not use the FluScreen to view the specimen.
3. If necessary, change magnification and/or illumination, and acquire a new image.
Note Every exposure causes damage to the specimen.
To maintain the highest quality, try to keep the accumulated dosage to a minimum.
8.5.2.5 The impact of Hole Position Refinement on the 'Move stage to location' function
In a Session type: Manual session, the Hole Selection > Find Holes function identifies Foil Holes in a
Grid Square image, and stores the XYZ coordinates of each Foil Hole center. When the Move stage
to location function is used during preparation for the Automated Acquisition run, the red crosshair
in the Image Display will end up in the center of the Foil Hole.
Between the acquisition of a Grid Square image in the Hole Selection task and the time at which the
Automated Acquisition run reaches that same Grid Square, a small positioning offset may have
accumulated. To prevent skipped Foil Holes due to such offsets, the Automated Acquisition
procedure features an automatic Hole Position Refinement function. This function acquires a new
just-in-time Grid Square image and updates the stored coordinates for each Foil Hole, so that the Foil
Holes can be centered with better accuracy for execution of the Foil Hole Template.
During the Automated Acquisition run it is possible to fine-tune the Foil Hole Template. To do so,
return to the Hole Selection task, use the Move stage to location function to center a Foil Hole,
and then select the Template Definition task to revise the Foil Hole Template.
In the Hole Selection task, the image display uses the initial Grid Square image, but the Move stage
to location function uses the updated Foil Hole coordinates. As a result, the crosshair in the image
display may end up at a small offset from the uncorrected Foil Hole center in the Grid Square image.
5. Use the Template Definition functions to define the Foil Hole Template.
as described in Specify the Maximum Image Shift and Delay Timers on page 148 .
When an Acquisition, Autofocus or Drift Measurement Area is selected in the image display, an
additional Ribbon Bar section appears to specify the values for the selected area. The Template
Definition section may change to a more compact layout to make room for the additional Ribbon Bar
sections.
Instead of creating a new Foil Hole Template for every new session, it is also possible to import the
Template Areas and their settings from an earlier session. See Import and export a Foil Hole
Template on page 149 .
EPU can automatically generate a pattern of Acquisition Areas in the Foil Hole Template.
EPU adds as many Acquisition Areas as possible according to the selected pattern.
2. Define the Defocus List as described in Define the Defocus List for one or more Acquisition
Areas on page 144 .
● On microscopes with a Three Lens Condenser system, the size of the Acquisition Area is
defined by the Illuminated Area which is specified in the Data Acquisition Preset.
● On microscopes with a Two Lens Condenser system, the Acquisition Area is defined as the
smallest circle that fully covers the camera field of view at the Data Acquisition Preset
magnification.
3. In the image display, select the Acquisition Area.
The Data Acquisition Area section appears in the Ribbon Bar.
4. Specify a Defocus List as explained in Define the Defocus List for one or more Acquisition
Areas on page 144 .
5. (Optional) Add more Acquisition Areas in the Foil Hole Template:
a. If the new Acquisition Area must use the same Defocus list as the first Acquisition Area,
then select the first Acquisition Area.
The Defocus List of the selected Acquisition Area will be copied to the new Acquisition
Area(s). Each Acquisition Area has its own Defocus List. Modifications to the Defocus List of
one Acquisition Area are not shared with the other Acquisition Areas.
b. Select Add Acquisition Area and select the location for the new Acquisition Area
6. (Optional) Adjust the location of an Acquisition Area by dragging it with the mouse.
7. (Optional) Modify the Defocus List for one or more individual Acquisition Areas.
8.6.3 Define the Defocus List for one or more Acquisition Areas
EPU acquires one image per Acquisition Area. By default, this image is acquired with the Defocus
value that is specified in the Data Acquisition Preset. To use a different defocus, specify one or more
defocus values in the Defocus List. During the Automated Acquisition run, EPU cycles through the
Defocus List to acquire images with different defocus values. On Quantifoil specimens, the Foil Hole
Template can contain multiple Acquisition Areas. Each Acquisition Area has its own Defocus List.
The example below describes a Foil Hole Template with two Acquisition Areas:
● Acquisition Area A has a Defocus List with three values.
● Acquisition Area B has a Defocus List with two values.
As the Foil Holes are processed one after the other, EPU steps through the Defocus Lists for each of
the Acquisition Areas.
If a Foil Hole is skipped by EPU or by the user, then the defocus values that were planned for the
acquisitions on that Foil Hole are also skipped.
To add or remove Defocus values for the currently selected Acquisition Area, follow the instructions
below:
After the Defocus List is applied to all Acquisition Areas, it is still possible to modify the Defocus
List for an individual Acquisition Area.
It is not possible to modify the Defocus List while an Automated Acquisition run is ongoing or paused.
To modify the Defocus List, the run must first be stopped.
● On microscopes with a Three Lens Condenser system, the size of the Autofocus Area is
defined by the Illuminated Area which is specified in the Autofocus Preset.
● On microscopes with a Two Lens Condenser system, the Autofocus Area is defined as
the smallest circle that fully covers the camera field of view at the Autofocus Preset
magnification.
2. In the image display, select the Autofocus Area
3. Select or specify the Autofocus Area settings:
a. Select Recurrence:
● Never: Do not perform the Autofocus function at all.
● Always: Autofocus is performed at every Foil Hole.
● After Distance: Autofocus is performed only when the distance between the current Foil
Hole and the most recent Autofocus location is larger than a specified value.
● After Centering: Autofocus is performed after a cluster of Foil Holes is centered.
Only available when Session Setup > Acquisition Mode is Faster Acquisition.
b. If Recurrence is set to After Distance, then specify the Distance.
c. Select Focus using:
● Objective Lens: EPU adjusts the focal plane to the Z-position of the specimen.
● Stage Z axis: EPU adjusts the Z-position of the specimen to the focal plane.
Confidential, limited rights EPU User Manual
PN 1025707 | Revision 2.14 | 25-JAN-2022 Page 146
Chapter | EPU Tab
The Drift Measurement Area is best placed outside the Foil Hole perimeter, on the carbon
foil. This way no valuable specimen material is exposed.
● On microscopes with a Three Lens Condenser system, the size of the Drift
Measurement Area is defined by the Illuminated Area which is specified in the Drift
Measurement Preset.
● On microscopes with a Two Lens Condenser system, the Drift Measurement Area is
defined as the smallest circle that fully covers the camera field of view at the Drift
Measurement Preset magnification.
2. In the image display, select the Drift Measurement Area.
3. In the Drift Measurement section of the Ribbon bar:
movements, the Delay after Stage Shift timer is used. If the Maximum Image Shift is set
smaller than the distance between the Acquisition Areas, then the Once per
GridSquare option is the only valid choice.
b. Specify the Drift Threshold.
This is the maximum allowed drift speed for data acquisition. If the specified threshold is not
achieved within 600 seconds, then the Foil Hole will be skipped and the Automated
Acquisition run continues with the next Foil Hole.
Follow the steps below to specify the Image Shift and delay values.
The values for the delay timers can be determined with the Auto Functions > Drift stabilization
autofunction.
8.6.6.1 Guidelines for the Drift Measurement Function and the Delay After Stage Shift Time
There are circumstances in which Drift Measurement function may not always be as accurate and
fast as desired. The appearance of the specimen can change due to the accumulating damage that
is caused by repetitive exposure. This may make it difficult to accurately determine the image shift
between consecutive images. In extreme cases, this could even cause the Drift Measurement
function to time out at 10 minutes.
Under these circumstances:
● Adding a Drift Measurement Area and selecting Recurrence: Always would significantly slow
down the Automated Acquisition run, and make the predictions in the Progress panel less
reliable. It may also result in a relatively large number of skipped Foil Holes due to failed Drift
Measurements.
● Adding a Drift Measurement Area and selecting Recurrence: Once per Grid Square decreases
the chance of occurrence and impact of a time out. At the same, the Automated Acquisition run
is not paused longer than necessary after moving on to the next Grid Square.
● If a predictable throughput is most important, it can be better to not add a Drift Measurement
Area to the Foil Hole Template, and use the Delay after Stage Shift timer for all moves.
Run the Auto Functions > Auto-Functions (TEM) > Drift stabilization autofunction to determine a
usable value for Delay after Stage Shift. For instructions, see Run the Drift Stabilization auto-function
on page 70 .
For the Template Definition, the following Smart Plugins are available:
● Stage Settling Time Predictions provides a stage settling time that is derived from the acquired
data. When a stage move is executed, EPU can request a stage settling time from Smart EPU.
The stage settling time that is provided by this Smart Plugin overrides the Template Definition >
Minimum stage settling time value.
This Smart Plugin can only be enabled when Session Setup > Acquisition mode is Accurate.
● Smart Focus Predictions provides a focus value that is derived from the acquired data. Instead of
executing the Autofocus auto function, EPU can request a focus value from Smart EPU. If the
Smart Plugin can provide a focus value, then EPU skips the execution of the Autofocus function
and uses the provided value.
Export
Select Export to write the current Foil Hole Template to a file (*.epuTemplate).
The Template file contains:
● The type, location and all settings of all Template Areas.
The Template file does not contain the size of the Template Areas. The size is determined by the
pertaining Presets.
● The Maximum Image Shift, Delay after Image Shift and Delay after Stage Shift values.
● The Defocus list for the Acquisition Area(s).
The Template file also contains the Foil Hole diameter. The Foil Hole diameter is only used as a
reference value for the Import functionality.
Import
The Foil Hole diameter of the loaded specimen does not have to be the same as the Foil Hole
diameter in the Template file. EPU calculates the ratio of the Foil Hole diameters to proportionally
adjust the positions of the Template Areas. If the ratio is too large, then EPU shows a warning. In that
case, the Focus and/or Drift Area may be located outside the field of view of the Hole/Eucentric
Preset.
1. Select Import and select a previously stored Template file.
2. If necessary or desirable, reposition the Template Areas and/or adjust their settings.
The time it takes to complete a Preview run of the Foil Hole Template is a good estimate for the
Session Information panel > Exposures per hour. Keep in mind that the Foil Hole Template can
contain multiple Acquisition Areas.
8.8.2 Verify that there is sufficient free disk space and sufficient liquid nitrogen (LN2)
Before starting the Automated Acquisition run, check the following:
1. Make sure there is sufficient available disk space.
● On the Microscope PC.
Session related data and images are stored in the Session Setup > Output folder. In case
no more data can be stored in this location, EPU pauses the Automated Acquisition run and
requests another storage location. If this happens, select a new Output folder and resume
the Automated Acquisition run.
● On the Storage Server or Gatan PC.
Dose Fractions are stored on the Storage Server or Gatan PC. Data volumes of several
hundreds of Gigabytes per day are no exception. If there is no more space to store Dose
Fraction images, the Automated Acquisition run will stop.
2. Make sure the Liquid Nitrogen (LN2) dewars are full.
To avoid failure of an unattended run, make sure that the LN2 supply is sufficient to finish the
run. EPU frequently checks if a refill of the dewars is required and will request a refill if needed.
EPU cannot detect if the main LN2 supply runs out. If the LN2 is not refilled, the Automated
Acquisition run keeps going while the sample temperature slowly increases.
a. Select Auto Zero Loss: Yes to enable periodic execution of the Auto Zero Loss function
during the Automated Acquisition run.
b. Specify the Periodicity
3. If the system has a Gatan filter with Gatan K3 camera,
then specify the Dark Reference Settings:
a. Select Dark reference acquisition: Yes to enable periodic acquisition of Dark Reference
images during the Automated Acquisition run.
b. Specify the Periodicity
4. (Optional) If available, enable the Smart Plugins > Skip Gridsquare Prediction plugin.
This Smart Plugin uses the data that is acquired during the Automated Acquisition, and which is
evaluated by EPU Quality Monitor (EQM). If the quality of the acquired data is not good, then the
Skip Gridsquare Prediction plugin sends a recommendation to EPU to skip all remaining
acquisitions on the current Grid Square. The Skip Gridsquare Prediction plugin will never
recommend to skip the first Grid Square.
5. Select Start Run
On a Lacey Carbon specimen, the counters apply to the number of defined and processed
Acquisition Areas.
Exposures
● Number of completed data acquisitions.
If too many acquisitions are skipped, the number of exposures will appear yellow or red.
● Number of planned data acquisitions (set).
● Number of remaining data acquisitions (left).
Grid Square
● Number of processed Grid Squares.
The processing status of each Grid Square is also indicated by color coding in the Atlas.
● Number of planned Grid Squares.
Holes
Holey carbon only
● The number of processed Foil Holes.
● The number of planned holes.
● The number of skipped holes.
When too many holes are skipped, the counter will turn red.
Exposure rate
The actual number of data acquisitions per hour, based on the statistics of the current session,
including the skipped areas.
Time remaining
The estimated remaining duration, based on the actual Exposure rate and the number of remaining
data acquisitions.
End time
The estimated end time of the run.
The color indicates whether or not the available tool time is fully used, assuming that no Grid
Squares or holes are skipped:
● Red: poor utilization.
When the Total time is spent, there is still a significant amount of tool time left.
● Yellow: adequate utilization, there is room for improvement.
When the Total time is spent, there is a small amount of tool time left. If the Session type is
Manual, add a few more holes to the selection so all available tool time is used for data
acquisition.
● Green: maximum utilization.
When the Total time is spent, there is no tool time left.
A progress bar at the bottom side of the Progress panel indicates the overall progress.
1. Select Pause
When the Session Setup > Acquisition Mode is Faster acquisition, then EPU recalculates the
Foil Hole groups before data acquisition starts again.
● If the quality of the current Grid Square results in too many skipped Foil Holes or Acquisition
Areas, select Skip Gridsquare to move on to the next Grid Square.
● If the quality of the current Foil Hole or Acquisition Area causes a delay, or will surely result in
failure of an automated function, select Skip Foilhole or Skip Area.
When the Session Setup > Acquisition Mode is Faster acquisition then EPU automatically
recalculates the Foil Hole grouping.
The Stop button changes into a Start Run button, and the processing status of the Grid Squares
and the selected areas are stored in the EPU Session file.
2. Depending on the type of failures, adjust the Preparations tab > Acquisition and Optics
Presets values.
Note Be very careful! Some adjustments may invalidate certain calibrations. See the descriptions of the
Presets, Calibrations and Auto Functions to check if additional updates are necessary.
● In a Session type: Manual session, the Foil Hole center positions are already defined by
their XYZ-coordinates during the Hole Selection task, so an adjustment of the Foil Hole
spacing will have no effect.
To update the Foil Hole positions, return to the Square Selection task and Hole Selection
task to create a new selection of Foil Holes with more accurate coordinates.
● In a Session type: Automated session, an adjustment of the Foil Hole diameter and/or
spacing will have an impact on the Find And Center Hole function, but it may also have
an impact on the Find Holes detection algorithm. These functions use the Hole/
EucentricHeight Preset and the GridSquare Preset. If these Presets do not have a
parallel beam and/or do not use little to no defocus, then the hole diameters in the
acquired images may deviate significantly from the physical specimen.
4. (Optional) Change the processing status of one or more Grid Squares from Processed (blue)
back to Open (green).
In EPU > Square Selection, right-click on a processed Grid Square and select Reopen
EPU reads the EPU Session file and starts the Automated Acquisition run from where it was
stopped earlier. Even if the EPU software is closed and opened again, the Automated Acquisition
run will start again from where it was stopped.
On a system with an Autoloader, the EPU Multigrid option makes it possible to prepare a queue of
individual EPU Sessions, and then start data acquisition for the entire prepared queue. It is possible
to prepare one EPU Session per specimen, but is is also possible to prepare multiple EPU Sessions
for a single specimen. If the Multigrid queue involves specimens from multiple Autoloader slot
positions, then EPU will automatically exchange the specimen between consecutive EPU Sessions.
The sections below explain how to create, edit and execute an EPU Multigrid experiment.
The Atlas Alignment function does not correct the coordinates of the I0 Calibration position.
Because the specimens are unloaded and reloaded after the I0 Calibration is performed, the I 0
reference value measurements during the Automated Acquisition run may be less reliable.
2. In the Preparation > Calibrate I0 task, select Remove I0 Measurements
If there is no calibrated value, then the Remove I0 Measurements button is not enabled.
3. In Atlas > Screening, acquire an Atlas for each specimen that will be included in the Multigrid
experiment.
4. Select New Queue
5. If a EPU Multigrid Session is currently active, then choose if the new EPU Multigrid Session uses
the preferences from the current EPU Multigrid Session.
● Yes: create a new EPU Multigrid queue and create an EPU Multigrid Session that uses
mostly the same settings and values as the currently active EPU Session. For an overview of
the copied settings and values, see The EPU Session Preferences file on page 95 .
● No: create a new EPU Multigrid queue and EPU Multigrid Session without pre-selected
settings and values.
● Cancel: no new EPU Multigrid queue is created. The current EPU (Multigrid) Session stays
as it is.
EPU creates an EPU Multigrid Session for the specimen that is currently loaded on the stage,
and moves on to the Session Setup task.
6. Perform the Preparation tasks for the first EPU Session in the Multigrid queue.
If necessary, the settings and calibrations in the Preparation tab and/or Auto Functions tab can
be updated.
a. Perform the Session Queue task.
For instructions, see: Session Setup task on page 98 .
For an EPU Multigrid Session, the Session type is always Automated. Manual target
selection is not supported.
b. Perform the Square Selection task.
For instructions, see: Square Selection task on page 107 .
c. For a Holey carbon grid,
perform the Hole Selection, Template Definition and Template Execution tasks.
For instructions, see:
● Hole Selection task for Quantifoil specimens on page 126 .
● Template Definition task for Quantifoil specimens on page 141 .
● Template Execution task for Quantifoil specimens on page 151 .
d. For a Lacey Carbon grid,
perform the Area Selection task.
For instructions, see: Area Selection task for Lacey Carbon specimens on page 112 .
7. (Optional) Select Save preferences to store an *.epuSessionPreferences file.
Specify a filename that is easily recognizable, so that the stored preferences can be selected
later for a similar specimen. See The EPU Session Preferences file on page 95 for an overview
of the stored settings and calibrations.
8. (Optional) Specify Queue > Max Exposures to limit the number of acquisitions in this EPU
Multigrid Session.
The Max Exposures value can changed at any time during Preparation and Execution, also when
a different EPU Session is active.
9. Add one or more new EPU Multigrid Sessions to the queue.
For each new EPU Multigrid Session:
a. (Optional) Load a different specimen on the stage.
For instructions, see: Load a specimen on the stage on page 87 .
It is also possible to create multiple EPU Multigrid Sessions for a single specimen, for
example for different sets of Grid Squares.
b. Select Add Session or Add from Preferences to create a new EPU Multigrid Session for
the specimen that is currently loaded on the stage.
1. Select the Queue > EPU Multigrid Session that needs to be edited.
To select a session in the Queue, double-click on it.
2. (Optional) Specify Queue > Max Exposures to limit the number of acquisitions in this EPU
Multigrid Session.
The Max Exposures value can changed at any time during Preparation and Execution, also when
a different EPU Session is active.
3. (Optional) In the Preparation tasks, change the desired settings and values.
If necessary, the settings and calibrations in the Preparation tab and/or Auto Functions tab can
be updated.
Note If the specimen of the selected EPU Multigrid Session not currently loaded on the stage,
then all actions that require a stage move or that require an image acquisition are not available.
The Slot number of the currently loaded specimen has a green highlight.
4. (Optional) Select Save preferences to store a new *.epuSessionPreferences file,
or to update an existing file that contains the preferences for this EPU Multigrid Session.
To exclude an EPU Multigird Session from data acquisition, set Queue > Max Exposures to 0
(zero).
9.4 Change the order of the EPU Multigrid Sessions in the Queue
To change the order in which the EPU Multigrid Sessions will be processed during the Automated
Acquisition, drag and drop the EPU Multigrid Sessions to their desired position in the Queue.
It is not possible to change the order in the Queue while data acquisition is ongoing or paused. It is
only possible to change the order in the Queue before data acquisition starts, or after data
acquisition is stopped.
9.5.1 Perform Automated Acquisition for a single EPU Multigrid Session from the
Queue
To acquire data for only one of the prepared EPU Multigrid Sessions from the Queue, follow the
steps below:
a. Select Auto Zero Loss: Yes to enable periodic execution of the Auto Zero Loss function
during the Automated Acquisition run.
b. Specify the Periodicity
a. Select Dark reference acquisition: Yes to enable periodic acquisition of Dark Reference
images during the Automated Acquisition run.
b. Specify the Periodicity
4. (Optional) If available, enable the Smart Plugins > Skip Gridsquare Prediction plugin.
This Smart Plugin uses the data that is acquired during the Automated Acquisition, and which is
evaluated by EPU Quality Monitor (EQM). If the quality of the acquired data is not good, then the
Skip Gridsquare Prediction plugin sends a recommendation to EPU to skip all remaining
acquisitions on the current Grid Square. The Skip Gridsquare Prediction plugin will never
recommend to skip the first Grid Square.
5. Select Start Run
The actions and functions to pause, resume, stop and troubleshoot the Automated Acquisition run
are the same as in a single specimen experiment. For descriptions and instrucions, see: Automated
Acquisition task on page 152 .
9.5.2 Perform Automated Acquisition for all EPU Multigrid Sessions in the Queue
To successfully complete an Automated Acquisition run in an EPU Multigrid experiment, the following
prerequisites must be fulfilled for each specimen in the queue:
● The tasks in the Preparations tab and the Auto Functions tab must be completed.
● An Atlas must be available.
To perform an Automated Acquisition run for the entire Multigrid Queue, follow th esteps below:
a. Select Auto Zero Loss: Yes to enable periodic execution of the Auto Zero Loss function
during the Automated Acquisition run.
b. Specify the Periodicity
2. (Optional) If available, enable the Smart Plugins > Skip Gridsquare Prediction plugin.
This Smart Plugin uses the data that is acquired during the Automated Acquisition, and which is
evaluated by EPU Quality Monitor (EQM). If the quality of the acquired data is not good, then the
Skip Gridsquare Prediction plugin sends a recommendation to EPU to skip all remaining
acquisitions on the current Grid Square. The Skip Gridsquare Prediction plugin will never
recommend to skip the first Grid Square.
3. (Optional) In the Queue, drag the EPU Multigrid Session for the currently loaded specimen to
the top of the list.
Although it is not strictly necessary, this will save time because the Automated Acquisition
doesn't have to start with a specimen exchange action.
4. Select Start Queue
● If issues are found, then EPU displays these in the Messages side panel, and the automated
acquisition run is not started.
5. Monitor the Automated Acquisition run during the first few data acquisitions.
6. If there are no irregularities, select the desired Options:
10.3 View the MRC, EMI, DMX and other microscope images
The high resolution MRC files can be viewed and post-processed with most electron microscopy
post-processing applications.
It is also possible to view the acquired images with the built-in Windows Photo Viewer:
10.4 View and post-process MRC images with Thermo Scientific Velox
software
MRC images can be viewed and post-processed:
● On the Microscope PC with the Thermo Scientific Velox Online Processing software.
● On any other computer with the Thermo Scientific Velox Offline software.
See the Velox User Manual for detailed descriptions and instructions.
The Velox software does not provide 3D reconstruction functionalities.
The instructions and verifications in this chapter typically use the TEM User Interface > CCD/TV
Camera control panel and TIA for image acquisition and inspection. Depending on the camera type,
other applications may be more suitable to acquire and inspect images:
● Thermo Scientific Falcon 4(i): use Velox.
● Stand-alone Gatan cameras: use Gatan Digital Micrograph.
● Direct alignments should be checked and, if necessary, adjusted in the modes that are used for
EPU.
These include, but may not be limited to the LM and SA range, Nanoprobe and Microprobe
modes.
● Bias/Dark and Gain Reference images are available and well averaged for all cameras.
● The camera is cooled and at a stable temperature.
● After any actions that may have introduced exceptionally strong drift, such as inserting a holder,
enough time should be allowed for settling.
11.2.1 Clear the Apertures > Options > React on Optical Mode Changes option
Note Do not use the React on optical mode changes function to change the apertures automatically.
Among others, the React on optical mode changes function automatically returns to the C2 aperture
that was selected the previous time that the optical mode was used. This may conflict with the C2
Aperture value in the Acquisition and Optics Presets.
5. When Spot Size 11 is checked, verify the Gun Alignment in reverse sequence:
a. Decrease the Spot Size number by one step.
b. Verify that the beam stays centered.
If the verification in step 5 fails:
● Perform all Alignments control panel > Gun alignments.
● Repeat steps 1 to 5.
5. Decrease the Intensity to the smallest value at which the Beam Settings control panel >
Illumination is Parallel.
6. On the FluScreen, check that the diffraction pattern circles for gold are sharply defined.
7. Take note of the Area value.
8. Slowly increase the Intensity until the Beam Settings control panel > Illumination is no longer
Parallel.
9. Take note of the Area value.
10. Decrease the Intensity, so that the Area value is the average of the lower and upper Area
values that are noted in steps 7 and 9.
11. On the FluScreen, check that the diffraction pattern circles for gold are sharply defined.
If the diffraction pattern circles are not sharply defined at step 11, perform the Align HM-TEM >
Basic SA alignment.
If the check at step 10 fails, perform the Direct Alignments control panel > Condenser Zoom
alignment and repeat steps 1 to 10.
1. With the Handpanels > Intensity knob, condense the beam to the smallest achievable spot.
2. Verify that the Beam Settings control panel >Area, is 0±100 nm
If this check 2 fails:
a. Perform the Condenser > Condenser Preparation alignment.
b. Depending the used optical mode, either:
● Perform the Focus + calib nP alignment.
● Perform the Focus + calib nP alignment.
11.4.2 Deflectors
The image/beam shift calibration must have been done carefully:
● To keep the illumination centered when some image shift is added.
● To shift the image when you want to move the beam to another area of the specimen.
Both Image and Beam Shift pivot points must be aligned accurately so the illumination conditions
(coma, rotation center) and crossover shifts (GIF energy selection) do not change when applying
image (beam) shift.
If the check at step 5 and/or step 10 fails, perform the Calibrate HR-TEM > Image/Beam
Calibration alignment.
11.4.3 Projector
The LM and HM lens series must be aligned accurately to ensure that a centered feature stays
centered when switching to another magnification. The alignment must be performed accurately on
the camera that is used for data acquisition.
1. With the Handpanels > Magnification knob, select the highest Acquisition and Optics
Presets magnification
2. In the TEM User Interface > Image Settings control panel, verify that no image shift is applied.
3. In the TEM User Interface > CCD/TV Camera control panel:
a. Select the Camera that is used for data acquisition.
b. Select Search
4. Move the specimen, so that a feature is visible in the image center that is also recognizable on
the camera at the Atlas Preset magnification.
5. Place a marker on the recognizable feature.
6. With the Handpanels > Magnification knob:
a. Step down through all Acquisition and Optics Presets magnifications and check if the feature
stays aligned with the marker.
b. Step up again through all Acquisition and Optics Presets magnifications and check if the
feature stays aligned with the marker.
This alignment is often done on the FluScreen, which might have a different magnification center.
Perform the Preparation tab > Calibrate Image Shifts task to compensate for magnification center
offsets.
The HM eucentric focus preset should not deviate more than a few microns from the true focus at
stage eucentric height.
1. If not visible yet, add the Defocus value to the TEM User Interface > Information Panels.
2. Accurately bring the specimen to eucentric height by using the stage wobbler.
3. Select a higher SA magnification (approx. 150kX).
4. Select Handpanels > Eucentric Focus.
5. Reset Defocus.
Either:
● Select the Handpanels > User button that is assigned to Reset Defocus.
This is often R2.
● Select TEM User Interface > Image Settings control panel > Reset Def.
6. Use the Handpanels > Focus knob to accurately focus the image.
7. Verify that the Information Panel > Defocus value is 0±500 nm.
The lens series alignment should be parfocal as much as possible, so switching magnification does
not significantly affect (de-)focus. For the LM range and the lower magnifications of the HM series,
this alignment is often not performed with maximum achievable accuracy.
11.4.3.4 Test the Magnification-Dependent Beam and Image Shift - TEM mode
If the check at step 18 fails, perform the nP calibrations and/or HM uP calibrations, and repeat
steps 1 to 18.
11.4.3.5 Test the Magnification-Dependent Beam and Image Shift - TEM mode
The image shift and the beam shift from highest SA to lowest HM magnification must be less
than 50 nm
If the check at step 18 fails, perform the nP calibrations and/or HM uP calibrations, and repeat
steps 1 to 18.
Among others, the React on optical mode changes function automatically returns to the C2 aperture
that was selected the previous time that the optical mode was used. This may conflict with the C2
Aperture value in the Acquisition and Optics Presets.
● Astigmatism
In a simple FFT it is very difficult to distinguish astigmatism from beam tilt induced coma. Make
sure that the system is nearly coma-free before correcting astigmatism.
Use the objective stigmator to adjust astigmatism. This procedure can only be done properly:
● On a thin carbon foil.
● With the camera in Search mode.
● With live FFT.
11.7 Calibrations
11.7.1 Magnification Calibrations
The Magnification Calibrations do not only contain information about the magnification, but also
about image rotation and how to transform shifts seen in the image to corresponding shift with stage
and image/beam deflectors. The Magnification Calibrations are generally very stable. They only need
to be renewed when the system alignments have changed significantly.
The calibrations must be present:
● For all cameras.
● For LM, SA and, if used, Mi ranges.
Use the TEM User Interface > Calibrations or Magnification Calibration control panel to perform
the Magnification Calibrations and follow the instructions in the control panel.
Atlas does not look consistent: The specimen may not be in focus. Refocus.
Large deviations in defocus will
● Tiles do not match.
cause the images to rotate and,
● Grid bars do not continue across since illumination is not parallel in
neighboring tiles. this step, the apparent
● Atlas contains a regular pattern of black magnification will also change.
areas.
If the specimen is focused, the Recalibrate.
calibrations are probably not valid
anymore and must be redone. (Has
there been a big change in
alignments or was another
alignment file loaded?)
The system's beam and image shift Re–align the beam and
are not well–aligned, causing loss image shifts for LM mode
of illumination. using the TEM User
Interface (Alignments
Control, Calibrate LM–
Image/Beam calibration).
“Move to” command does not center feature or The specimen may not be in focus. Refocus.
grid square. Large deviations in defocus will cause
the images to rotate and, since
Check this at the magnification at which the
illumination is not parallel in this step,
Atlas is acquired. Use the camera, not the
the apparent magnification will also
Flucam Viewer.
change.
“Move to” command (in Atlas) does not Some image shift is applied (check in the Reset it.
center the Grid Square. 'Image Settings' control panel in the TEM
User Interface).
Atlas looks OK, but navigation to the grid
square is inaccurate.
If you see the same problem at Atlas Focus and/or redo the
Grid square is not centered when
magnification (see above), specimen was calibrations.
acquiring an image in the Location / Area
not focused when taking the atlas (see
Selection view.
above) or calibrations are not valid
anymore.
“Move Stage to” command does not center Foil Hole or Too much defocus Make sure to work with no
feature. can change image or just a few mm defocus.
rotation. A non- (Focus specimen correctly
If you see this while staying at the magnification used for
parallel beam will by using the wobbler and
imaging the grid square, check the causes and solutions in
lead to additional specify the wanted defocus
this table.
change in in the optics settings of
If you see this problem only in the Template Definition view magnification. EPU.) If possible, use a
(at higher magnification), see “Move Stage to …” parallel beam.
command in Location Selection / Area Selection view does
not center foil holeFoil Hole or feature in Template Mechanical play in Make sure that before
Definition view.” stage. taking the grid square
image, the grid was
centered with a Move
Stage… command in the
atlas (compensates for any
backlash problems).
Note Since the feature positions are calculated and not all characteristics of the stage are known exactly, a
feature will not be centered 100% accurately.
A micron deviation is not uncommon.
“Move Stage to …” command in Location Some image shift is applied (check this in Reset it.
Selection / Area Selection view does not 'Image Settings' control panel in the TEM
center Foil Hole or feature in Template User Interface).
Definition view.
If you see this problem also at the Bad alignment of LM lens series against HM Repeat EPU
magnification at which the grid square is lens series and bad EPU image shift image shift
imaged, see information under “Grid Square calibrations. calibration or
(Location Selection/ Area Selection View):”. when the shifts
Test:
are large, adjust
When switching between Location Selection LM lens series
view and Template Definition view alignment.
magnifications, a feature should stay
centered in the acquired images.
● Select the Preparation tab in EPU.
● Select the Acquisition and Optics
Settings task.
● Use the Acquire buttons for easy
switching with correct lens normalizations
and image shift corrections.
“Move Stage here...” command in Mechanical play in stage. Try at least twice to make sure
Template Definition view does that a backlash correction was
not center feature. performed.
Working far from eucentric focus will Make sure the sample is at
render the magnification calibrations eucentric height, focused, rotation
invalid. Large defocus will slightly center is well corrected, and the
rotate the image and (with a defocus chosen in the Optics
nonparallel beam) magnification settings of EPU should not
changes. exceed 20 um.
Yellow circle does not Correct Quantifoil type was not selected at Set the size correctly in this view.
match Foil Hole size. start of session or the size of the holes
deviates from specification.
Size of the holes deviates from specification. Set the size correctly in this view.
Measure Foil Holes/Find Foil Holes in the Return to Location Selection and
Location Selection view was not performed. do it.
Too large defocus is used and beam is not Select a defocus selected in the
parallel. (The apparent magnification of the Template Optics settings of EPU
image may change under these of less than 20 μm. If possible,
circumstances.) choose a parallel beam.
The Measure Foil Hole step was inaccurate. Repeat this step (Location
Selection view) with a slight
change in the size of the yellow
glasses.
“Find Foil Hole” Does the yellow circle match the hole size? -
does not succeed.
If not, see above, Yellow circle does not
match Foil Hole size.
Measure Foil Holes/Find Foil Holes in the Switch to the Location Selection view and
Location Selection view not done. (The perform Measure Foil Holes and Find Foil
“find Foil Hole” algorithm uses information HoleFoil Holes.
about hole size and spacing.)
The circle size is wrong. (The algorithm of Repeat the Measure Foil Holes step
finding holes may be quite sensitive to the (Location Selection view). Change the
exact hole size.) size of the yellow glasses slightly.
Magnification is too low. (The current Increase the magnification such that only
algorithm does not work with too many Foil the Foil Hole and its nearest neighbors
Holes visible.) are visible on the acquired image.
The current specimen location contains too Find a cleaner sample area.
many features (crystalline ice, etc.) that
cause the algorithm to fail.
Image quality too bad (noise). The human Take image with higher dose.
eye is very good at integrating over
features, so noise may easily be
underestimated at this magnification
Foil Holes appear dark when the Allow If the Foil Hole appears darker than the
Dark Foil Holes option is not selected. foil itself, select the option Allow Dark Foil
HoleFoil Holes (in ribbon of Location
Selection view).
Foil Hole centering fails (or hole is not correctly The Foil Hole was not fully Reduce the
centered). visible initially and the fitted magnification.
circle degenerated to an
Tracing this problem is a combination of tracing problems
ellipse. In this case the
in:
centering will not be perfect.
● “Find Foil Hole”: see “'Find Foil HoleFoil Hole' does
not succeed”.
Can be tested separately.
● “Move stage here”: see “'Move Stage here...'
command in Template Definition view does not
center feature”.
Can be tested separately.
Pattern acquisition Foil Hole cannot be recentered. (Pattern See above, Foil Hole centering fails
fails. acquisition requires the Foil Hole to be centered (or hole is not correctly centered).
within 300 nm.)
Test:
Click Find and Center Foil Hole button.
Pattern Specimen is too far off from eucentric Adjust eucentric height
acquisition: height and focus. Usually during an
Auto Focus fails. automatic run of EPU, after every move to
a new grid square, the eucentric height is
adjusted. During the setup phase, you
moved to a grid square, but may not have
adjusted eucentric height
Cause: Auto focus does not converge Use the Direct Alignments panel to align the
because the beam is visible in the images. beam shift at the auto focus or data
There maybe a couple of reasons for this. acquisition magnification.
Beam tilt pivot point misaligned. Use Direct Alignments panel to align the
beam tilt pivot point.
Tip: Adjust eucentric height and focus first.
There is a little problem in that the pivot point
is well aligned only at a certain focus setting.
You may have to allow for some inaccuracy
by choosing a slightly larger beam diameter
in the focus optics setting since in the
automatic data acquisition you will not always
start close to focus.
The image/beam shift calibration is not Perform the Calibrate HM-TEM > Image/
done. Beam calibration (use the 'Alignments' control
panel in the TEM User Interface).
Test:
When you add some image shift in the
Image Settings control cluster, do you then
see the beam edge moving into the field of
view?
Auto focus does not converge, although the Redo focus calibration
beam edge is not visible; invalid focus
calibration.
Test with the standalone Auto Focus
routine.
Pattern acquisition: Acquire a single image with the Data Use the Direct Alignments panel to
beam edge visible in Acquisition presets (Preparation tab). align the beam shift data acquisition
data acquisition magnification.
If the beam edge is visible on this image, this
means Beam shift was not aligned correctly in Note: If the beam moves when you
the first place. change intensity of illuminated area,
align apertures or condenser system
as well.
When you add some image shift in the Image Perform HM-TEM > Image/Beam
Settings control cluster, do you then see the calibration ('Alignments' control panel
beam edge moving into the field of view? If so of the TEM User Interface).
Image/beam shift calibration is not done.
Pattern acquisition: you use Auto Focus by Stage Z Adjustment, maybe the Try again. This time the Z
acquisition areas are first time Z had to be adjusted a lot, which leads to adjustment should be
not placed correctly. displacements in X and Y. small.
If you use Auto Focus by Change of Objective Lens Use the Direct Alignments
Current, focusing may shift the feature (hole) when the panel to align the rotation
rotation center is not well aligned. center.
EPU image shift calibration no longer valid or not done. Redo the EPU image shift
calibrations. If the shift is
Use the Preparation tab to acquire images with the Hole/
very large, redo the
EucentricHeight preset and Data Acquisition preset.
microscope HM image shift
Does a given feature stay centered? If yes, then the lens
alignment first.
series alignment is not well done and the deviations are
not captured by the EPU image shift calibration.
Image beam shift calibration not valid. (If you use image/ Perform HM-TEM > Image
beam shift only to place the acquisition areas, placement Beam calibration
should generally be quite accurate.) ('Alignments' control panel
of TEM User Interface).
Grid squares The eucentric height procedure is failing. Increasing the number of counts per
are all skipped. exposure may help if the images are
The performance of the auto-eucentric height
very noisy. The focus calibration may
procedure can be tested standalone in the Auto
have to be redone.
Functions tab. During automated data acquisition,
the procedure is performed with the Hole/
EucentricHeight optical preset (a different defocus
may be applied). The procedure used is the one that
is based on a focus measurement and, therefore, a
valid focus calibration is needed.
No holes are found. A grid square image is acquired, Redo the “Measure Foil Holes”
and auto eucentric height succeeds, but then no Foil procedure that should have been
Holes were found. Measure Foil Holes procedure done during the setup phase
was not set up correctly. (Location Selection task).
No Foil Holes are selected. A grid square image is Redefine the filters (go to the
acquired, holes are found, but none is selected by Location/Area Selection). If the
the ice thickness filters. intensity changes due to an instability
of the gun, use the I0 calibration in
EPU in addition. It will periodically
measure the intensity in a hole and
correct the ice filters accordingly.
Too many Foil They are not found in the grid Repeat setup steps
Holes: square image in the first place
target areas are because step(s) were omitted in
skipped. the setup phase.
They do not pass the ice filtering. Go to the Location/Area Selection view and adjust the
Either the illumination conditions ice filter settings
or acquisition settings have
changed after the filters have been
selected or the filters were tuned
on a grid square with too thin/thick
ice.
Foil Holes were selected in the For troubleshooting, see “'Find Foil Hole' does not
grid square image, but could not succeed.” and “Foil Hole centering fails (or hole is not
be found or centered at the Foil correctly centered)”.
Hole magnification. A Foil Hole is
skipped if it could not be centered
within certain accuracy
Auto focus fails. A Foil Hole is For trouble shooting, see “Pattern acquisition: Auto
skipped if the focus could not be Focus fails.”.
determined.
During the automatic data acquisition, auto focus may
also fail if the lens series focus presets are (very) ill-
defined. The routine is based on the measurement of
beam tilt-induced image shifts that are proportional to
the amount of defocus (it works like the wobbler on the
hand panel).
For valid measurements, the images must overlap by
at least 60%, which means that the range that the auto
focus can cover is limited, especially when run at high
magnification. Initially, eucentric height and focus are
adjusted at the magnification used for centering the
Foil Hole (before an image of the grid square is
acquired), but later the auto focus routine is run at a
much higher magnification typical for data acquisition.
If the magnification switch results in a large focus
change, auto focus may fail. In this case, correct the
objective lens presets of your lens series (part of the
lens series alignment).
Acquisition areas are See “Pattern acquisition: acquisition areas are not placed correctly"
not placed correctly.
When you have chosen Z adjustment as the method of Little to nothing can be
auto focus, Z adjustments will always induce some X and done to fix this.
Y movements of the specimen. Big adjustments can be a
consequence of a non-flat or tilted specimen ().
Beam is
shifted in
Autofocus
routine.
For high magnifications (> ~75 kX) and small beam sizes as shown in the image above, it can
happen during the Autofocus routine that the beam seems slightly shifted, resulting in a not
completely illuminated detector.
This is usually due to one or both of the following causes:
● Beam tilt pivot points not set correctly
● A too large value in Autofunctions > Autofocus > Iterate to:.
To resolve:
● Make sure you are at eucentric height (run Autofunctions > Eucentric Height or set
manually).
● Select Eucentric focus on the hand panel; a well aligned TEM will now be in focus.
If not, carefully focus manually or use Autofunctions > Autofocus at a lower magnification.
● Select Reset defocus.
Defocus value in the TEM User Interface is now '0'.
● Perform Direct Alignments > beam tilt pivot points.
● Go to the desired optical settings for Autofocus.
● Check the number in Iterate to: in Autofunctions > Autofocus.
● Manually change defocus to this number and check the movement of the beam.
● If it is too large (as show in the image above), try the following (in the order shown below):
● Find the largest defocus value for which the beam stays on the detector and reduce
Iterate to: to the number found,
● Reduce the camera area to Half or Quarter,
● Increase the beam diameter.
The tables below specify the content of the FEI1 and FEI2 Extended Headers for the MRC2014 file
format. In these tables, the Format and 'IsPresent' flag columns have to the following values:
● Format:
Timestamp 12 0x000C Float64 Bitmask 1 – #0 Time when the image was taken. The used
format is the DATE data type that is used
in OLE automation by Microsoft:
Microsoft OLE DATE data type
specification
Gun
Name Offset (dec) Offset (hex) Format 'Is Present' flag Description
Stage
Alpha tilt 100 0x0064 Float64 Bitmask 1 – #7 Holder Alpha tilt along axis [degr.]
Beta tilt 108 0x006C Float64 Bitmask 1 – #8 Holder Beta tilt along axis [degr.]
Tilt axis angle 140 0x008C Float64 Bitmask 1 – #12 Angle of tilt axis in image [degr.]
Dual axis 148 0x0094 Float64 Bitmask 1 – #13 Measured rotation angle after b flip
rotation [degr.]
(Tomography only)
Pixel Size
Name Offset (dec) Offset (hex) Format 'Is Present' flag Description
Pixel size X 156 0x009C Float64 Bitmask 1 – #14 Pixel size X [m]
Pixel size Y 164 0x00A4 Float64 Bitmask 1 – #15 Pixel size Y [m]
Optics
STEM Defocus 228 0x00E4 Float64 Bitmask 1 – #23 STEM defocus [m]
Applied defocus 236 0x00EC Float64 Bitmask 1 – #24 Relative defocus applied by
application [m]
EFTEM On 288 0x0120 Bool Bitmask 1 – #30 TRUE when the magnifications are
adapted to the energy filter
Camera length 301 0x012D Float64 Bitmask 2 – #0 Nominal camera length [m]
EFTEM Imaging
Name Offset (dec) Offset (hex) Format 'Is Present' flag Description
Image Shifts
Shift offset 387 0x0183 Float64 Bitmask 2 – #12 Corrective image or beam shift relative to
X exposure preset (in logical units)
● TEM: pure image shift
Shift offset 395 0x018B Float64 Bitmask 2 – #13
Y ● STEM: image-beamshift-
Shift X 403 0x0193 Float64 Bitmask 2 – #14 Applied shift due to optimized position and
tracking (in logical units)
Shift Y 411 0x019B Float64 Bitmask 2 – #15 ● TEM: image beam shift
Camera
Integration time 419 0x01A3 Float64 Bitmask 2 – #16 Camera or dose fraction
exposure time
Camera name 435 0x01B3 16 chars Bitmask 2 – #19 Name of the camera
Ceta frames summed 468 0x01D4 Int32 Bitmask 2 – #25 Number of frames
summed for dynamic
range
STEM
Name Offset (dec) Offset (hex) Format 'Is Present' flag Description
Scan settings
Name Offset (dec) Offset (hex) Format 'Is Present' flag Description
Dwell time 571 0x023B Float64 Bitmask 3 – #15 Dwell time per pixel [sec]
Frame time 579 0x0243 Float64 Bitmask 3 – #16 Frame time [sec]
(currently it will not be used)
Full scan FOV X 603 0x025B Float64 Bitmask 3 – #21 Field of view [m]
Name Offset (dec) Offset (hex) Format 'Is Present' flag Description
Dose fractions
Name Offset (dec) Offset (hex) Format 'Is Present' flag Description
Reconstruction
Name Offset (dec) Offset (hex) Format 'Is Present' flag Description
Scan rotation 768 0x0300 Float64 Bitmask 4 – #2 Rotation of the scan pattern in STEM
mode [radians]
Diffraction pattern 776 0x0308 Float64 Bitmask 4 – #3 Rotation of the diffraction pattern in
rotation diffraction mode [radians]
Image rotation 784 0x0310 Float64 Bitmask 4 – #4 Rotation of the image in imaging mode
[radians]
Start tilt angle 820 0x0334 Float64 Bitmask 4 – #8 Start tilt angle of a tomography series
[degr.]
End tilt angle 828 0x033C Float64 Bitmask 4 – #9 End tilt angle of a tomography series
[degr.]
Tilt per image 836 0x0344 Float64 Bitmask 4 – #10 Tilt increment per image in a
tomography series [degr.]
Tilt speed 844 0x034C Float64 Bitmask 4 – #11 Tilt speed in a tomography series
[degr./sec]
Beam center X 852 0x0354 Int32 Bitmask 4 – #12 Beam center X on image [pixels]
pixel
Beam center Y 856 0x0358 Int32 Bitmask 4 – #13 Beam center Y on image [pixels]
pixel
Phase plate 868 0x0364 Int32 Bitmask 4 – #15 Position index of the phase plate
position index aperture
Objective aperture 872 0x0368 16 chars Bitmask 4 – #16 Name of the inserted objective aperture
name
Most image viewers and image processing applications use the same pixel position sequence as the
MRC file. Some image viewing and processing applications such as IMOD and Fiji/ImageJ use a
pixel position sequence. In these applications, the image display may be mirrored and/or rotated.
A.3.1 The MRC image pixel data encoding for Thermo Scientific Ceta cameras
If the image is acquired with a Ceta camera, then the MRC image pixel data encoding depends on
the presence of the Ceta Speed Enhancement (Ceta-2).
Limited Rights
Contractor Name: FEI Company (part of Thermo Fisher Scientific)
Contractor Address: 5350 NE Dawson Creek Drive, Hillsboro OR 97124
The Government's rights to use, modify, reproduce, release, perform, display, or disclose these
technical data are restricted to those rights specified in DFARS 252.227-7015(b)(2), FAR 52.227-
14(g)(2)(Alternate II) and FAR 12.211. Any reproduction of technical data or portions thereof marked
with this legend must also reproduce the markings. Any person, other than the Government, who has
been provided access to such data, must promptly notify the above named Contractor.
To provide feedback on this document, please submit via thermofisher.com/EM-Sales
Revision Table
Index
Change the processing order of the Grid Squares
A
110
Acquire Atlases for multiple specimens 83
Check the C2 Apertures center positions on a system
Acquire a single Atlas from an unknown specimen on without a C3 lens 181
the stage 85
Check the C2 Apertures center positions on systems
Acquire an Atlas for a single specimen on a system with a C3 lens 180
with removable holders (side entry) 91
Check the Objective Apertures center positions
Acquire an Atlas on a Tundra system 89 181
Acquisition and Optics Settings task 18 Clear the Apertures > Options > React on Optical
Activate Phase Plate task 61 Mode Changes option 171
Activate a new Phase Plate before an Automated Copyright, Limited Rights and Revision History 207
Acquisition run 154 Create a new Atlas Session 81
Add and remove Foil Holes with the Selection Brush Create a new EPU Session without an EPU Session
137 Preferences file 97
Add and remove acquisition areas with the Selection Create a new Session from an EPU Session
Brush 122 Preferences file 97
Advanced Camera Settings for Gatan K2 and K3 Create a queue with EPU Multigrid Sessions 161
cameras 24
Create a selection of Foil Holes without a generated
Advanced Camera Settings for Thermo Scientific pattern 138
Ceta cameras 23
Customize the selection of Acquisition Areas 121
Alignments tasks 64
Customize the selection of Foil Holes 136
Aperture Alignments 180
Area Selection task for Lacey Carbon specimens D
112
Define a selection of Acquisition Areas without a
Atlas Optics Alignment task 53 generated pattern 122
Atlas Tab 81 Define one or more Acquisition Areas 142
Auto Functions Tab 64 Define the Acquisition Area Pattern and Filter
Auto-Functions (TEM) tasks 65 Settings for a Manual Selection Session 118
Automated Acquisition task 152 Define the Acquisition Area Pattern and Filter
Settings for an Automated Selection Session 112
Automatically create a Manual Selection in all
selected Grid Squares 120, 135 Define the Atlas Preset 50
Automatically generate a pattern of Acquisition Areas Define the Autofocus Area 146
142 Define the Autofocus Preset 40
Define the Data Acquisition Preset 28
C
Define the Defocus List for a Lacey Carbon specimen
Calibrate I0 task 58 125
Calibrate Image Shifts task 55 Define the Defocus List for one or more Acquisition
Calibration tasks 75 Areas 144
Calibrations 183 Define the Drift Measurement Area 147
Camera Settings for all camera types 18 Define the Drift Measurement Preset 42
Change the order of the EPU Multigrid Sessions in Define the Foil Hole dimensions and Filter Settings
the Queue 164 for a Manual Selection Session 132
Define the Foil Hole dimensions and Filter Settings
for an Automated Selection Session 126
Confidential, limited rights EPU User Manual
PN 1025707 | Revision 2.14 | 25-JAN-2022 Page 209
Chapter | Index
Define the GridSquare Preset 48 Guidelines for the Drift Measurement Function and
Define the Hole/EucentricHeight Preset 36 the Delay After Stage Shift Time 149
Define the Thon Ring Preset 44 Guidelines for the Hole/EucentricHeight Preset when
using Phase Plates 39
Define the Zero Loss Preset 47
Guidelines for the Optics Settings when using Phase
Deflectors 175 Plates 27
Description of the Camera Settings 18 Gun and Condenser 172
Description of the Optics Settings 25
Description of the Session Setup options 98 H
Detailed Preconditions for Successful EPU Usage Handpick the Acquisition Areas for a Manual
170 Selection Session 115
Determine the Phase Plate Activation Time 62 Handpick the Foil Holes for a Manual Selection
Direct Alignments and Astigmatism 182 Session 130
Disable Beam Settings > Intensity Zoom 171 Histogram side panel 11, 11
Dose Rate for Thermo Scientific cameras 22 Hole Selection task for Quantifoil specimens 126
Home Tab 16
E How to improve the Autofocus Calibration result 77
EPU Tab 94 How to plan for high microscope utilization 152
Edit an EPU Multigrid Session in the Queue 163
Enable the Smart Plugins in the Template Definition I
149 Image Information side panel 12
Eucentric Correction Calibration (Optional) 183 Image and plot display area 13
Exchange the specimen on a Tundra system 90 Import and export a Foil Hole Template 149
Export an image to file 15 Import and export the Acquisition and Optics Presets
Exposure Settings for Thermo Scientific Falcon 3EC 52
and Falcon 4(i) cameras 19 Inspect a Foil Hole to assess suitability 139
Exposure Settings for the Thermo Scientific Ceta-F Inspect an Acquisition Area to assess suitability
camera 21 123
Inspect the Acquired Images 168
F Introduction 4
FEI1 Extended Header specification 198
FEI2 Version 2 Extension to the Extended Header L
specification 204 Limitations to loading or acquiring an Atlas after
Focus Calibration 183 Session Setup 106
Load a specimen on the stage 87
G Load an existing Atlas Session 82
Getting Started 6 Log in on Thermo Scientific Athena 7
Guidelines for the Autofocus Preset when using
Phase Plates 41 M
Guidelines for the Data Acquisition Preset when MRC image pixel format for Gatan K2 and K3
using Phase Plates on a microscope with a C3 lens cameras 100
31
Magnification Calibrations 183
Guidelines for the Data Acquisition Preset when
using Phase Plates on a microscope without a C3 Manually define one or more Acquisition Areas
lens 34 143
Messages side panel 10
System and software compatibility 5 The pixel format for Gatan Dose Fraction images
100
T The recommended order to define the Acquisition
TIFF image pixel format for Gatan K2 and K3 and Optics Presets 27
cameras 101 The red crosshair 15
Target audience for this user manual 5 Troubleshoot an ongoing Automated Acquisition run
Template Definition task for Quantifoil specimens 159
141 Troubleshooting: Atlas 184
Template Execution task for Quantifoil specimens Troubleshooting: Automated Acquisition 194
151 Troubleshooting: GridSquare 186
Test the Condenser Alignment 173 Troubleshooting: Symptoms and Solutions 184
Test the Condenser Focus Alignment 175 Troubleshooting: Template Definition 188
Test the Condenser Zoom on Titan Systems 174
Test the Eucentric Focus 176 U
Test the Gun Alignment 172 Unselect the bad Foil Holes in the generated pattern
Test the Image/Beam Shift Calibration 175 137
Test the Lens Series Magnification Center Alignment Unselect the bad areas in the generated pattern
176 122
Test the Magnification-Dependent Beam and Image Use the Sherpa APM function to maintain the
Shift - TEM mode 178, 179 microscope alignments 172
Test the Parallel Beam range on Titan Systems User interface panels 9
173
The Acquisition Mode for Quantifoil specimens 98 V
The Athena settings 101 Validate the Foil Hole Template 151
The Autofocus Calibration task 75 Verify that the microscope is ready for high quality
data acquisition 7
The Automated Acquisition task in the EPU Multigrid
workflow 164 Verify that there is sufficient free disk space and
sufficient liquid nitrogen (LN2) 153
The EPU Multigrid Option 161
Verify the on-plane illumination on a microscope with
The EPU Session Preferences file 95 a C3 lens 33
The Eucentric Correction Calibration task 78 Verify the on-plane illumination on a microscope
The Extended Header specification 197 without a C3 lens 35
The MRC image pixel data encoding for Thermo View and post-process MRC images with Thermo
Scientific Ceta cameras 206 Scientific Velox software 169
The MRC2014 Image Format 197 View the JPEG images 168
The Main Header and Extended Headers in an MRC View the MRC, EMI, DMX and other microscope
file 197 images 168
The Screening task for Tundra systems 89 View the images in Windows Explorer and Photo
The Screening task for systems with an Autoloader Viewer 168
83
The automatic Dose Fractions calculation for Falcon W
3EC and Falcon 4(i) cameras 22 When to reset and renew the Image Shift Calibration
The impact of Hole Position Refinement on the 'Move 55
stage to location' function 139
Z
Zoom in on an individual Atlas Tile image 109
Zoom in/out 14
For current certifications, visit FEI.com/certifications. © 2022 Thermo Fisher Scientific Inc. All rights reserved.
All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.