0% found this document useful (0 votes)
80 views3 pages

XRay DR Philips

The document provides a recommended room layout and equipment configuration for an Essenta DR preferred room. It includes a diagram of the recommended layout showing the placement of key equipment pieces. It notes that site conditions and customer preferences may impact the final design. It also lists environmental and power requirements for the general radiology equipment.

Uploaded by

Wilson Azevedo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
80 views3 pages

XRay DR Philips

The document provides a recommended room layout and equipment configuration for an Essenta DR preferred room. It includes a diagram of the recommended layout showing the placement of key equipment pieces. It notes that site conditions and customer preferences may impact the final design. It also lists environmental and power requirements for the general radiology equipment.

Uploaded by

Wilson Azevedo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

Essenta DR

Preferred Room Layout


The layout shown below is based upon a typical equipment configuration and should be considered as a general design guideline. Site conditions,
application requirements, customer preferences, and/or equipment configuration may significantly impact suite design and equipment layout. It is
recommended to request site-specific drawings from a Philips representative early in the design process.

16'-0"
[4877mm]

Fixed Stand Column


7'-4" Centerline
[2235mm]
[811mm]
2'-8"
Rear Bolt Holes

ME
Centerline

MS
[4268mm]
14'-0"

FFM

4' [1219mm] Wide Patient


Entry (Recommended)
[610mm]

PBC AWS
2'-0"

Counter and shelf below


provided and installed by
customer.

Equipment Layout
Recommended Ceiling Height: 10'-10" [3302mm]
Absolute Minimum Ceiling Height: 9'-2 1/4" [2800mm]

Ceiling heights (from finished floor to finished ceiling) other than recommended may impact
equipment functionality; consult with Philips.

0 1' 2' 4' 8' 12'

0 .5m 1m 2m 4m
Equipment Legend

A Furnished and installed by Philips


B Furnished by customer/contractor and installed by customer/contractor
C Installed by customer/contractor
D Furnished by Philips and installed by contractor
E Existing
F Future
G Optional item furnished by Philips
Equipment Designation
Weight Heat Load
Description
lbs [kg] Btu/hr [W]

A ME Optimus 80 Control Cabinet (40E Rack) 462 [210] 1707 [500]

A FFM Floor Mounted Fixed Stand FF-M 1253 [569] 1853 [543]

A PBC Optimus Control Panel 9 [4] 137 [40]

A AWS Acquisition Work Spot with Touch Screen Monitor 154 [70] 680 [199]
(CPU and UPS on shelf under counter)
G MS Height Adjustble Trolley TA-M 293 [133] - [-]
(Shown in suggested "parked" location)
Environmental Requirements for General Equipment Locations

Power Requirements
Optimus 80 / Optimus CXA

Supply Configuration: 3 phase, 3 wire power and ground. Delta or Wye.


(with optional* transformer)

Nominal Line Voltage: 400, 440, 460, or 480 VAC, 60 Hz

Branch Power Requirement: 167 kVA

Circuit Breaker: 3 pole, 100 A(@480 V)

* transformer standard in U.S.

Remote Service Diagnostics


Medical Imaging equipment to be installed by Philips is equipped with a service diagnostic feature which allows for remote and
on-site service diagnostics. To establish this feature, a RJ45 type Ethernet 10/100/1000 Mbit network connector must be
installed. Access to customer's network via their remote access server is needed for Remote Service Network (RSN)
connectivity. All costs with this feature are the responsibility of the customer.

Koninklijke Philips Electronics N.V. 2011.