Modelling Dose Assessment During Nuclear Medicine
Modelling Dose Assessment During Nuclear Medicine
Modelling Dose Assessment During Nuclear Medicine
55-63
Keywords: Radiation protection to staff member in nuclear medicine, Finger dose rate.
INTRODUCTION
A very important part of following ALARA (As Low As Reasonably
Achievable) principles is the establishment of a formal program for personnel
monitoring and bio-assay [1]. Medical use of radioactive materials falls broadly into
56 Khaled Abdelaziz and H.M. Diab
ο e − µr
d D = CΓ dV (1)
r2
e − µr
ο V
D = C Γ∫ 2
dV (2)
0 r
Eq. (1) can be rewritten as;
ο
D = C Γg (3)
e − µr
V
where g = ∫0 r 2 dV (4)
The factor, g, depends only on the geometry of the source, hence it is called the
geometry factor. The geometry factor in Eq. (4) applies to a given point within a
volume of source. The geometrical factor takes into account the effects of both distance
and energy absorption on the intensity of gamma photons as they penetrate the
medium.
58 Khaled Abdelaziz and H.M. Diab
The analytical solution of dose rate equation (2) in any volume source with any
shape except sphere is not found up till now and maybe impossible so the need of
modeling and found numerical solution becomes so important to calculate dose rate in
daily work or environmental studies.
The MC method is a mathematical technique for solving a problem that is
dependent upon probability in some manner. The technique is useful when exact
formulation describing a process may be too difficult, or even impossible, to derive and
solve by direct methods. The MC method constructs a stochastic model representing
the process of interest. A set of high quality random numbers is then used to sample the
probability distribution functions defined by the model. The result is an estimate of a
physical quantity characteristic of the process, specified with a measured degree of
confidence. Two of the main reasons why we use MC methods are because of their
anti-aliasing properties and their ability to approximate quickly an answer that would
be very time-consuming to find out the answer too if we were using methods to
determine the exact answer. Fig. 2 represents a 3D view of the calculated cylinder
volume.
A computer program (in Delphi language) was programmed and tested by the
first author to carry out gamma-ray dose rate outside volume sources or on the exterior
surface of volume containers fill of radioactive substance based on Monte Carlo
simulation techniques.
Radionuclides such as, 59Fe, 198Au, 113m In, 131I, 203Hg and 99mTc have γ-
emissions that readily penetrates through medical or laboratory containers including
syringes, vials and test tubes. The simulated sources are syringes with volume 1, 2.5
and 5 ml with dimensions tabulated in table (1), the properties of solutions and wall
materials for the containers in the present work for dominant photon energies are
tabulated in table (2).
MODELLING DOSE ASSESSMENT DURING NUCLEAR 59
Table 2. The properties of solutions and wall materials for the containers in the present work
for dominant photon energies are tabulated in.
Solutions and Density Mass attenuation coefficients (cm2 g-1) J.E. Pattison 1996
wall material (g cm-3) 59
Fe 198
Au 113m
In 131
I 203
Hg 99m
TC
Water 1.00 0.0659 0.1050 0.1071 0.1106 0.1225 0.1541
IM HCl 1.035 - 0.1048 - - - -
Polypropylene 0.90 0.0676 0.1077 0.1098 0.1135 0.1254 0.1567
Experimental work
In this study, the theoretical model was experimentally verified using TLD
(LiF-100) discs. Before each use, the discs were annealed for 1 h at 400oC followed
with 100oC for 2 hours [10]. Calibration of the discs was performed using 90Sr/90Y
gamma equivalent irradiator (Harshow 4000, USA).
A common type of polypropylene syringe was used to measure the dose rate
around. The syringes were filled with 99mTc in aqueous solution (supplied as
pertechnetate with initial activities of approximately 10 GBq). Dose rate was
measured by placing four discs around each of the syringe (as shown in Fig. 1) which
were then left to irradiate for about 36-48h. Four discs were used as controls (to
monitor background dose). The TLD discs were read on a Harshow 4000.
The dose received by each individual disc for each syringe was then
calculated taking into consideration the various corrections and calibration factors. The
testes with 99mTc were repeated three times providing twelve measurements of dose
rate of each syringe.
TLD
Radioactive solution
Figure 3. The syringe held with radioactive Solution and thermoluminescent dosimeter (TLD)
60 Khaled Abdelaziz and H.M. Diab
THEORETICAL RESULTS
Useful comparisons may be made with the measured dose rates of Takaku and
Kida (1972), and the calculated dose rates of Henson (1973), for a number of
radioisotopes and syringe sizes. The dimensions of the syringes used by Henson (1973)
are shown in table 1. Dose rates at points on the surface of partly filled syringes and
centered over the active volume.
The results are shown in table 3. In general, the results of this study are lower
than Henson’s results. The average difference between the two sets of results, for all
results, is – 6: 9%. Agreement is particularly poor for the 99mTc results, where the
average difference is −12:8%. However, there are very high correlations (r = 0:996)
between Henson’s results and the results of this study, for each of the radionuclides
listed, which suggests the presence of systematic errors. A possible cause of the
observed differences could be that Henson used different values of than were used in
this study. Henson did not indicate in his paper the values of that he used.
Table 3. Comparison of calculated gamma– ray dose rates with values adapted from Henson
[9], Pattison [3] and our study
Volume Dose rates (mGy/MBq min) per gram
Syringe 59
Fe 198
Au 113m
In
Injected
(ml)
(ml) 1 2 3 1 2 3 1 2 3
1.0 0.3 123.0 121.0 118.6 44.6 44.7 43.5 35.4 33.3 32.5
2.5 1.0 60.4 62.7 60.6 23.6 23.2 22.2
5.5 1.0 28.8 27.1 25.6 21.0 20.2 19.0
3.0 18.4 14.7 14.2 13.1 11.0 10.6
5.0 11.8 10.3 9.7 7.9 7.6 7.2
131 203 99m
I Hg Tc
1.0 0.3 43.3 42.4 41.4 26.2 24.3 23.7 14.4 13.7 13.4
2.5 1.0 23.6 22.0 21.1 13.1 12.6 12.1
5.0 1.0 9.2 8.3 7.7
5.0 3.0 5.2 4.5 4.3
5.0 5.0 4.0 3.1 2.9
EXPERIMENTAL RESULTS
For a 5-ml syringe containing 3 ml of liquid, the absorbed dose at the midpoint
of the liquid region was 5.1 mGy/MBqmin. This is in good agreement with
calculations and measurements reported in the literature. The syringe dimensions and
the volume of the liquid (source) region are variable input parameters. Table 4
illustrated the variation of volume used and the dose received due to preparation of
syringe compared with the calculated results. From the results obtained, it was found
MODELLING DOSE ASSESSMENT DURING NUCLEAR 61
that, there is a good correlation between the calculated dose around the syringe and the
measured dose rate using TLD discs.
DISCUSSION
Various factors can influence the dose rate at the surface of the syringe.
Principal among these factors are: (1) the volume of liquid required to contain the
desired amount of radioactivity; the larger the volume, the lower the dose rate at any
given point; (2) the composition, wall thickness, and diameter of the syringe; (3) the
amount of radioactivity being used and (4) the specific radioactive element involved. It
is apparent from the readings over the loaded and unloaded areas of the syringe that 2
important factors influence exposure to the skin: (1) the portion of syringe held during
loading or injecting; and (2) the time the syringe is held [11]. To minimize exposure of
the skin, the syringe should be held by the portion of the barrel farthest from the
radioactivity, and for the shortest time possible. In loading a syringe, one should hold
the barrel by the extreme rear-end, quickly aspirate the necessary volume, and lay the
syringe down promptly [12].
RECOMMENDATION
Reduction in occupational radiation dose levels is an important consideration
in the overall ALARA program. By using radiation attenuating gloves, a significant
reduction with a mean of 42% in radiation exposure to extremities was noted with
minimal effects on dose preparation time, dose handling, and technologist comfort. It is
recommended, when using syringes without guards, that, the fingers should never
approach the active volume closer than the rear end of the syringe barrel, and the
syringes should not be filled beyond 75 % of their capacity.
REFERENCES
1- Wandergouw H.F. "Guidelines to Radiopharmacy", Scientific Pub. Co., New
York, 136-137, 1982.
2- Jansen S.E., van Aswegen A., Lotter M.G., Herbst C.P., Otto A.C. "Staff radiation
doses during eight years in a nuclear medicine radiopharmacy", Nucl Med
Commun. 1994; 15:114–118.
62 Khaled Abdelaziz and H.M. Diab
ZeR اgaopRاد اrpR اd]stn `ءQvوى أrQR اTURات اWX وZ[ \]^_`aR` اbR ضdaef ZeR اg]h`aij اghdkR اl]]mn ln
ln W و.gaopR`دة اpR اZ^h gfretpR اgkd{R اU Z^h Z_`kRض اdaeRل اWa_ {`بX |fd} \h ~R وذfdpR` اb \mtf
gfdQR{``ت اtR` اe g`رm_ fd وا.{``تt^R rR آ`رr_ gmfd} Z^h Wpea_ ىd ذجrp امWe ghdkR ه اl]]mn
مr]{QeR _`دة اZ^h ىretn ZeRو^]\ واd ZRrR _\ اgaQpR اgkd{Rام اWe اWQh `b]^h لrtR اln ZeR اg]^paR` اeQR`
.opRا
_) `_` `تao_ Z^h ىretn gao_ _`دةZ^h gfret_ gkd امWe اWQh ` اW^ Z^h `تhdkRل اWapR g{QR`
ىd `ذج اpQR{``ت اX `e _ حdempRذج اrpQR{``ت اX `e رr (TهR ا،مr]{QeR ا،lfW وا،|R ا،دr]R ا،WfWtRا
Z`رR اU{Rاد اWe_ اZ^h g]_`kR اghdkRل اWa_ {``تX ZU¡]R _`dRر اr} W و،g]RوWR`ت اfورWR اZ[ رةroQpRا
`eQRح واdempRذج اrpQR{``ت اX `e \] W] |[اrn `QX§ و.م٩٩-مr]{QeR ا،١٩٨-gs[ d`QR` `ً] g£^ep_ `تkd{R
اد ذاتrpR اZ^h gfretpR`ت اkd{Rام _ ه اWe اWQh g_©زR ا_`ن اd]f`a_ « وZ[ `eQRم ه اWe{n و.g]^paRا
.gf`ام و ` وWe دون اg]_`kR`ت اv`ajا
gfretpR اgmUQpRاب _\ اdeم اWa gf` و ` وd]¬ \_ gkd{Rام اWe اWQh `رehj اZ[ ® اtR اZrf
. ±m[ `bQ_ %٧٥ ادWe_ p{n ZeR اT `QpR اlktR ذات اgkd{Rام اWe واgaopR`دة اpR اZ^h