Cusa Excel User Manual Operation Manual
Cusa Excel User Manual Operation Manual
Distributed By:
Integra LifeSciences Corporation
Plainsboro, NJ 08536 USA
888.980.7742 (Fax)
integralife.com
Integra, the Integra logo, CUSA and CUSA Excel are registered trademarks of Integra
LifeSciences Corporation or its subsidiaries in the United States and/or other countries.
© 2019 Integra LifeSciences Corporation. All Rights Reserved.
60903711, Rev E 0868211-2
CUSA® Excel/CUSA® Excel+
Ultrasonic Surgical Aspirator
System
User’s Guide
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator
System User’s Guide
This User’s Guide and the equipment it describes are for use only by qualified
medical professionals trained in the particular technique and surgical procedure to
be performed. It is intended as a guide for using the CUSA® Excel and CUSA®
Excel+ Ultrasonic Surgical Aspirator System only.ystem.stem only.
Equipment Covered in this Manual
The CUSA Excel refers to the following product models: CUSA Excel and
CUSA Excel-8. Both product models utilize 23 kHz and 36 kHz handpieces.
The CUSA Excel+ refers to the product models: CUSA Excel 2 and CUSA Excel-9.
The system incorporates graphic modifications of the logo design and color of the
top cover of the system. The CUSA Excel+ System utilizes the same CUSA Excel
family of handpieces.
Caution
Y Federal (USA) law restricts this device to sale by or on the order of a physician.
Trademark Acknowledgements
Integra, the Integra logo, CUSA, CUSA Excel and Tissue Select are registered
trademarks of Integra LifeSciences Corporation or its subsidiaries in the United
States and/or other countries.
MacroTip, MicroTip, PrecisionTip, SaberTip, CEM and ShearTip are trademarks of
Integra LifeSciences Corporation or its subsidiaries.
Ajax is a registered trademark of Colgate-Palmolive Company.
Comet is a registered trademark of the Spic and Span Company.
Klenzyme is a registered trademark of Steris Corporation.
Enzol is a registered trademark of Johnson & Johnson.
All other trademarks and trade names are the property of their respective owners.
Manufacturer:
Integra LifeSciences (Ireland) Limited
IDA Business & Technology Park
Sragh
Tullamore, County Offaly, Ireland
Distributed By:
Integra LifeSciences Corporation
Plainsboro, NJ 08536 USA; 1-800-997-4868
Notice
A technical description of the CUSA Excel/CUSA Excel+ System is provided in
the CUSA Excel/CUSA Excel+ System Ultrasonic Surgical Aspirator System
Service Manual. The Service Manual is not supplied with the system as it is
intended to be used only by Integra service personnel and/or their agents.
System Features
The CUSA Excel/CUSA Excel+ System includes several important features:
• 23 kHz (straight and angled) and 36 kHz (straight) handpieces
• A large variety of 23 kHz and 36 kHz surgical tips that attach to the
handpieces:
Sterile, single-use tips
Nonsterile, extended-life tips
• Tissue Select® feature, which increases the selectivity of the surgical tip,
allowing greater control and precision.
Organization
This User’s Guide is organized into these sections:
• Section 1: Patient and Operating Room Safety presents the Warnings,
Cautions and Notices that you need to read and understand to operate the
CUSA Excel/CUSA Excel+ System with maximum safety.
• Section 2: Introduction to the CUSA Excel/CUSA Excel+ System gives an
overview of the console and its functions, and an overview of handpieces and
related accessories.
• Section 3: Console Components describes the console, its subsystems and its
components.
• Section 4: Control Panel Display and Functions describes the control panel
display, the signs and symbols on the control panel and the control panel
behavior during system startup and operation.
• Section 5: Setting Up the CUSA Excel/CUSA Excel+ System explains how to
prepare the console and its subsystems for use in surgery. It includes the
handpiece assembly options; how to prepare the console for startup, connect
and test the handpiece, connect the irrigation and suction systems; how to
prime the system, adjust settings, and switch to the Run mode.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide iii
• Section 6: Using CUSA Excel/CUSA Excel+ Console Controls describes the
control panel buttons and functions. It also includes guidelines on using the
Tissue Select feature.
• Section 7: Handpiece Components presents the items that compose an
assembled handpiece, their physical characteristics, and the function of each
item. It also presents items that, although not a part of the handpiece itself, are
essential in assembling a handpiece or preparing it for sterilization.
• Section 8: Assembling the Handpiece in a Nonsterile Area provides step-by-
step instructions for assembling the handpiece and preparing it for sterilization.
• Section 9: Sterilizing Handpieces and Accessories provides the sterilization
parameters you need to ensure that the handpiece is ready for use in the
sterile field.
• Section 10: Completing Handpiece Setup in the Sterile Field describes how to
complete the handpiece assembly in the sterile field.
• Section 11: Assembling or Changing Tips in the Sterile Field describes how to
attach or change sterile tips in the sterile field.
• Section 12: Shutting Down the CUSA Excel/CUSA Excel System explains
how to turn off the system; disconnect suction tubing, irrigation tubing, and the
handpiece; and how to clean the console.
• Section 13: Disassembling Handpieces describes how to disassemble and
clean the handpiece, and how to clean the tip torquing set.
• Section 14: Troubleshooting the CUSA Excel/CUSA Excel+System offers
suggestions for problem solving before or during surgery.
• Section 15: Maintaining the CUSA Excel/CUSA Excel+System describes the
maintenance tasks that help to keep the console and handpieces operating as
intended.
• Appendix A: Technical Specifications provides detailed technical information.
• Appendix B: Sterilization Validation provides information on the sterilization
validation procedure.
• Appendix C: Warranty provides warranty information.
Intended Uses
When you receive your CUSA Excel/CUSA Excel+ System, we recommend that
you read and understand all of this User’s Guide before using the system. Also,
use the guide for:
• Reference - When you need specific information on a task. Once you are
familiar with the system, use the “Quick Reference” cards (located behind the
Control Panel).
• Training - When training new personnel to use the system.Conventions Used
in this Guide
To draw immediate attention to matters of importance, this guide presents
Warnings, Cautions, Notices, and Important information.
Warning
Indicates a potentially hazardous situation that, if not avoided, could result in
serious injury or death, or product damage.
Notice
Indicates a hazard that may result in product damage.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide vii
Adjusting Setpoint Values 6-1
Changing Functions 6-3
Using the Tissue Select Feature 6-4
Using Other Features 6-6
viii CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Section 13. Disassembling and Cleaning Handpieces 13-1
For Your Information 13-1
Disassembling the Handpiece 13-2
Cleaning the Handpiece 13-4
Permitted Number of Reprocessing Cycles for Handpieces when Cleaned with an
Automatic Washer 13-6
Cleaning the Tip Torquing Set 13-7
xii CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 1
Patient and Operating Room
1
Safety
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 1-1
Intended Users
Warning
The CUSA Excel/CUSA Excel+ System cannot be used in an MRI (Magnetic
Resonance Imaging) environment.
Warning
No modification of this equipment is allowed.
Notice
When you receive the CUSA Excel/CUSA Excel+ System and accessories, if
any component is damaged, contact your Integra service representative for
assistance. If the packaging for a sterile accessory is damaged, do not use the
sterile accessory.
Intended Users
The intended users of this guide and the equipment it describes are
qualified medical professionals who are trained in the particular surgical
technique and surgical procedure to be performed, and trained in the use
of this equipment. The CUSA Excel/CUSA Excel+ System should only be
used in a surgical environment by qualified medical professionals.
Warning
It is the responsibility of the Healthcare Facility to ensure that intended users of
CUSA Excel/CUSA Excel+ System are appropriately trained in the use of this
equipment.
Safety Information
The safe and effective use of ultrasonic surgery depends to a large degree
on factors solely under the control of the operator. Only medical
professionals that are properly trained in the use of ultrasonic equipment
should operate the CUSA Excel/CUSA Excel+ System. It is important that
medical professionals read, understand, and follow the operating
instructions supplied with this equipment.
Before starting any surgical procedure, medical professionals should be
familiar with the medical literature, complications, and hazards of using
ultrasonic surgery in that procedure.
1-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Warnings, Cautions, and Notices
Notice
When you receive the CUSA Excel/CUSA Excel+ System and/or accessories, if
any component is damaged, contact your Integra service representative for
assistance.
Single Use devices are for single patient use only. Do not reprocess or re-use.
Devices (s) is (are) intended to be used for one procedure only. If reprocessed or
re-used this may result in the infection of patient (or patient specimen) through
cross-contamination, as well as would incur the risk of modifying the properties
and performance of the device, and of increasing the likelihood of complications
and/or undesirable effects. Once used, devices must be disposed of in
accordance with hospital policies.
Caution
Read the instructions, warnings, cautions, and notices provided with the
CUSA Excel/CUSA Excel+ System before use. Otherwise injury to the patient or
user or equipment damage may result.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 1-3
Warnings, Cautions, and Notices
Console Components
Warning
Ignoring alarms on the CUSA Excel/CUSA Excel+ System while continuing to use
the system may result in injury to the patient and/or surgical personnel, or
equipment damage.
To avoid injury to surgical personnel, keep fingers away from the suction pinch
valve.
The power cord on this product contains lead, a chemical known to the State of
California, USA, to cause cancer, and birth defects or other reproductive harm.
Wash hands after handling. The power cord complies with the DIRECTIVE
2011/65/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 8
June 2011on the restriction of the use of certain hazardous substances in
electrical and electronic equipment (recast).
When you connect the handpiece to the console, the handpiece becomes a
functional surgical device.
Inadvertent contact between handpiece accessories and the patient may result in
burns.
To avoid injury to surgical personnel, keep fingers away from the suction pinch
valve.
When you connect the handpiece to the console, the handpiece becomes a
functional surgical device.
The handpiece and handpiece accessories must be sterile before surgical use.
1-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Warnings, Cautions, and Notices
Warning
Touching of the tip of the handpiece by the operator, while the handpiece is
powered on, can result in personal injury.
When the handpiece is powered on, contact of the tip with a hard surface (e.g. a
metal instrument, tray, staples, clips, instruments, etc) may damage the tip of the
handpiece and require replacement before use.
CUSA Excel tips utilize silicone flues. Compressing the flue against the side of
the vibrating surface along the length of the tip can cause excessive heating and
potential hazard to adjacent tissue, such as burns.
Excessive loading of CUSA Excel tips at the surgical site can induce heating due
to vibration and acoustic power transmissions. Thermal management of the
surgical site with the aid of the appropriate irrigation and aspiration settings is
essential.
Avoid contacting bone with the CUSA Excel tips (excluding SaberTip™).
Caution
When you test the handpiece, do not allow the handpiece tip to contact anyone or
anything during tip activation. Contact may result in patient injury, user injury, or
handpiece tip damage.
Make sure that the irrigation tubing centers between the V-shaped tubing
retainers before you close the pump latch. Otherwise, the pump latch will pinch
the tubing, preventing the flow of irrigation fluid.
Before surgery, apply the brakes locks to all wheels on the console to stop the
wheels from rolling.
Notice
To prevent fluid flowing into the vacuum line, only use a canister that has a non-
return valve.
During surgery, do not allow the handpiece tip to touch metal objects such as
staples, clips, instruments, etc. Handpiece tip damage will result.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 1-5
Warnings, Cautions, and Notices
Handpiece Components
Warning
Do not use the sterile wrench for more than one surgical procedure.
Caution
Do not assemble the nonecone to the handpiece until you have sterilized the
handpiece.
To avoid product damage, NEVER hold the handpiece in your hand while using
the torque wrench to tighten or loosen the tip.
Notice
Do not sterilize the C5600 tip torquing base or the associated the torque wrench
because it destroys the lubrication in the torquing mechanism, resulting in
product damage.
The handpiece and handpiece accessories must be sterile before surgical use.
Before use, sterilize the sterilizable torque base in the sterilizer tray.
Turning the torque wrench further clockwise will damage the handpiece.
Caution
To avoid product damage, NEVER hold the handpiece in your hand while
using the torque wrench to tighten or loosen the tip.
1-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Warnings, Cautions, and Notices
Notice
Do not disconnect the handpiece until the control panel goes completely blank.
Otherwise, product damage may result.
Do not rub, press, or touch any panels with solvents; caustic, corrosive, or
abrasive cleaning or disinfectant compounds, or other materials that could
scratch the panels. Do not use a betadine-based solution because it will cause
discoloration.
Disassembling Handpieces
Caution
To avoid product damage, NEVER hold the handpiece in your hand while using
the torque wrench to tighten or loosen the tip.
Product damage will result if you do not follow these notices when cleaning the
handpiece:
Do not immerse the handpiece cable electrical connector in liquid
Do not use ultrasonic washers
Do not use chlorinated substances such as bleach solution
Do not clean the handpiece with abrasives such as Ajax®, Comet®, or steel wool
Product damage will result if you do not follow these notices when cleaning the
Tip Torquing Set:
Do not use ultrasonic washers
Do not autoclave
Do not use chlorinated substances such as bleach solution
Do not clean with abrasives such as Ajax, Comet, or steel wool
Do not clean the sterilizer case with abrasives. Product damage will result.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 1-7
Warnings, Cautions, and Notices
Explosion Hazard – Do not use the CUSA Excel/CUSA Excel+ System in the
presence of flammable anesthetics or any potentially explosive or flammable
atmosphere.
To avoid injury to surgical personnel, keep fingers away from the suction pinch
valve while powering the unit on or off, activating vibration, or using fast flush.
The console must be earthed and therefore it MUST only be fitted with a 3-pin
plug, or a 2-pin plug that has an integral earth grounding connection. Mains plug
type and construction MUST comply with Legal requirements within country of
installation. Only Integra Service personnel or Integra authorized representatives
or agents can change the mains plug on the console.
1-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Classification and Console Symbols
Symbol Description
Console Symbols
Warning
To avoid injury to surgical personnel, keep fingers
away from the suction pinch valve while powering the
unit on or off, activating vibration, or using fast flush.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 1-9
Classification and Console Symbols
Symbol Description
ON (power)
Applies power at the AC Main Switch and the System Power
Switch.
OFF (power)
Fuse Replacement
Indicates the location of the fuse drawer on the rear panel.
Volume Control
VOLUME
Footswitch Connector
Manufacturer
1-10 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Classification and Console Symbols
Symbol Description
Catalogue Number
REF
Serial Number
SN
Please dispose of in accordance with local regulations for the
collection or disposal of waste electrical and electronic
equipment.
Recyclable material
Footswitch Symbols
Activate vibration
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 1-11
Classification and Console Symbols
Symbol Description
NOT FOR Accessories marked with the ‘NOT FOR HUMAN USE’ label are
HUMAN supplied for demonstration purposes only.
USE
Do not push, lean or rest against the control panel, or against the
front or sides of the console. When moving the console, push the
console using the handle only.
1-12 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 2
Introduction to the System
2
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 2-1
About the CUSA® Excel/CUSA® Excel+ System
Fragmentation
Electromechanical Operation
The console provides alternating current at 23 or 36 thousand cycles per
second (kHz) to the handpiece (the frequency depends on which
handpiece you connect to the console). In the handpiece, the current
passes through a coil, which induces a magnetic field. The magnetic field
excites a transducer of nickel alloy laminations, resulting in an oscillating
motion in the transducer laminated structure – vibration – along its long
axis (refer to the figure at left). The transducer vibrates at 23 or 36 kHz.
The transducer transmits vibrations through a metal connecting body to an
attached surgical tip. The frequency of vibration remains the same at the
tip (23 or 36 kHz), but the amount of motion (amplitude) at the tip varies:
lower frequency, greater amplitude; higher frequency, smaller amplitude.
Amplitude also varies with the transducer/connecting body/tip
configuration: handpiece angles reduce tip amplitude.
When the vibrating tip contacts tissue, it breaks cells apart (fragments
them).
2-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
About the CUSA® Excel/CUSA® Excel+ System
Cooling
The high frequency vibration generates heat. To reduce the heat, the
system includes a closed, recirculating cooling water system. This system
pumps water from a cooling water reservoir, through a tube in the
handpiece cable, through the handpiece, and through a return tube in the
handpiece cable to the cooling water reservoir. Cooling water flows at
35 to 50 ml/min.
As it passes through the handpiece, the water removes heat. Normal
handpiece temperature, in sustained heavy use, remains at less than
40°C/104°F.
Use distilled water for the cooling water system. Do not use tap water
because it contains impurities (natural minerals, chemical additives, or
organic materials) that can cause problems within the cooling water
system.
Irrigation
Sterile irrigation fluid flows from an IV set (bottle or bag and IV
administration tubing) to a variable speed peristaltic pump. The pump:
• Moves fluid at 1 to 10 ml/min; default flow is 3 ml/min. Use the
adjustment buttons (blue up/down arrows on black buttons at the
bottom of the irrigation display column on the control panel) to increase
or decrease the irrigation flow.
• Accelerates to a Fast Flush speed, pumping at greater than 25 ml/min.
The Fast Flush pedal on the system footswitch activates the Fast Flush
feature.
The accuracy of the irrigation flow is +/- 20% of setting or +/-1 ml/min,
whichever is greater.
The pump pushes the fluid through the manifold irrigation tubing to a flue,
a sleeve surrounding the vibrating tip. As the irrigation fluid passes
through the flue, it cools the tip.
When the fluid reaches the distal end of the tip, as much as 99% of it
passes through two pre-aspiration holes in the tip, eliminating fluid pooling
in the sterile field and continually clearing the suction system. Fluid that
does not pass through the pre-aspiration holes irrigates the surgical site
and suspends fragmented tissue.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 2-3
About the Handpieces
Aspiration (Suction)
A vacuum pump in the console body provides up to 26 inHg/660 mmHg
maximum vacuum at sea level. Use the adjustment buttons (green
up/down arrows on black buttons at the bottom of the aspiration display
column on the control panel) to increase or decrease the suction from 10
to 100% in 10% increments.
Threads
The suction, which produces an air stream moving toward the vacuum
pump, pulls irrigation fluid, fragmented tissue, and other materials through
the distal end of the surgical tip. From the tip, the aspirated materials pass
through the manifold suction tubing into the suction canister. From the
suction canister, the air stream continues to flow through a contamination
guard that filters any remaining particulate matter or moisture, preventing
them from entering the vacuum pump.
The accuracy of the vacuum level between contamination guard and
vacuum pump inlet is ± 15% of scale setting or ± 2.6 inHg/ 66 mmHg,
whichever is greater.
A suction pinch valve on the front of the console opens when the system is
on, and closes to stop suction when:
• Priming the irrigation system
• Pressing the Fast Flush pedal
• Releasing the Vibration pedal in Run Status (in this case, the pinch
valve closes for approximately one second, then re-opens).
• Releasing the Vibration pedal in Lap Mode (this suction stoppage
prevents depletion of the pneumoperitoneum).
When the System Power Switch is off, the suction pump remains off and
the suction pinch valve remains closed. Use the button on the front of the
suction pinch valve to open the valve manually.
Warning
Only use Integra® handpieces and accessories with the CUSA
Excel/CUSA Excel+ System. Non-Integra handpieces and accessories are not
supported.
2-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
About the Handpieces
Handpiece Functions
Together, a handpiece and CUSA Excel/CUSA Excel+ console form an
ultrasonic surgical aspirator system. This system has three functions:
• Fragmentation: As the vibrating tip contacts tissue, the vibrations break
the tissue cells into fragments.
• Irrigation: Sterile irrigation fluid, pumped from the console through the
manifold tubing, suspends the fragmented cells so that suction can
easily remove them and prevent tip blocking. The irrigation fluid also
cools the vibrating tip.
• Aspiration (Suction): Suction through the hollow tip removes the
fragmented tissue and irrigation fluid from the surgical site.
All three functions can occur at the same time.
Handpiece Configurations
Notice
The 36kHz handpiece operates within a frequency range and 36kHz is a
representative value. Refer to the Technical Specification section for the
frequency range.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 2-5
About the Tissue Select® Feature
Handpiece Tips
A variety of handpiece tips are available.Tips vary in inside diameter,
length, and shape (straight or curved). For information on the tips
available for each handpiece, contact your Integra representative.
2-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
About the Tissue Select® Feature
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 2-7
About the Tissue Select® Feature
2-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Sterilization of Handpieces and Accessories
For information on using the Tissue Select feature, see Using the Tissue
Select Feature, page 6-4.
Table 2-1 describes the sterilization requirements for the handpiece and
accessories. For information on the sterilization parameters and
sterilization procedure for the handpieces and accessories, see
Sterilization Parameters, page 9-1.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 2-9
Sterilization of Handpieces and Accessories
2-10 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Sterilization of Handpieces and Accessories
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 2-11
Sterilization of Handpieces and Accessories
Warning
Single Use devices are for single patient use only. Do not reprocess or re-use.
Devices (s) is (are) intended to be used for one procedure only. If reprocessed
or re-used this may result in the infection of patient (or patient specimen)
through cross-contamination, as well as would incur the risk of modifying the
properties and performance of the device, and of increasing the likelihood of
complications and/or undesirable effects. Once used, devices must be disposed
of in accordance with hospital policies.
2-12 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 3
Console Components
3
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 3-1
About the Console and Structural Features
Control Panel
IV pole
®
Arm
Housing for
Irrigation Pump
Head and Suction
Pinch Valve
Arm Housing
Console Body
Cooling Water
Reservoir
Casters
Component Description
IV Pole Supports the sterile irrigation fluid container. You can raise or
lower the pole, and rotate it 90° in the lower position. The
safe working load for the I.V. pole is 2 kilograms.
Control Panel Allows the user to control all system functions. See Section
4: Control Panel Display and Functions.
3-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
About the Console and Structural Features
Component Description
Arm Housing Covers the arm attachment along the console body side
panel.
Housing for Prominent black area at the top of the console body:
Irrigation Pump
• The irrigation pump head and irrigation tube pathway
Head and
(marked with blue lines)
Suction Pinch
Valve • The suction pinch valve and suction tube pathway (marked
with a light green line)
Casters Unlocked, both front and rear casters roll easily and rotate
freely. When locked, each caster has a brake lock that stops
the wheels from rolling.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 3-3
Console Body – Front Panel
Handpiece
Connector
Receptacle ®
Irrigation Pump
Head
Suction Pinch
Valve
System Power
Switch
Contamination Guard
3-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Console Body – Front Panel
Component Description
Warning
When you connect the handpiece to the console, the
handpiece becomes a functional surgical device.
Suction Pinch Closed, the valve pinches off suction flow to the handpiece
Valve when:
• Priming the handpiece with irrigation fluid
• Pressing the Fast Flush pedal
• Releasing the Vibration pedal (in LAP Mode)
When the System Power Switch is off, the suction pump
remains off and the suction pinch valve remains closed.
Use the button on the front of the suction pinch valve to
open the valve.
Warning
To avoid injury to surgical personnel, keep fingers
away from the suction pinch valve.
Irrigation Pump Rotate the latch from right to left to open the pump head,
Head and from left to right to close. When open, insert irrigation
tubing into the peristaltic pump. When closed, the pump
head holds irrigation tubing against the peristaltic pump.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 3-5
Console Body – Front Panel
Component Description
Retaining Bail for Holds the suction canister in place on the shelf.
Suction Canister
3-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Console Body – Side Panel
IV Pole
Control Panel
Cooling Water
Reservoir Release
Button
Cooling Water
Reservoir
Component Description
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 3-7
Console Body – Side Panel
Component Description
Rotation Lock Releases the control panel, allowing it to swivel and lock
Handle into one of four positions:
• Straight forward
• 45° right
• 45° left
• 90° left
IV Pole
Arm Release Tab Releases a lock that holds the arm in the vertical position.
3-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Console Body – Rear Panel
Speaker
Volume Control
Footswitch
Connector
AC Main Switch
Fuses (x2)
Footswitch
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 3-9
Console Body – Rear Panel
Storage Compartment
Cord Wrap
3-10 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Console Body – Rear Panel
Component Description
Volume Control Adjusts the active vibration audible tone. You cannot adjust
the alarm tone.
VOLUME
AC Main Switch Controls AC power input to the system. See Section 3: About
the Power Switches.
Cord Wrap Use the four cord wraps to hang the cord on the back of the
unit, keeping the cord safely out of the way.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 3-11
Console Body – Rear Panel
Component Description
Storage Stores the footswitch and manual (includes one interior shelf)
Compartment
3-12 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
About the Power Switches
AC Main Switch
Located on the rear panel, this switch controls AC power input to the
system. It does not activate the system; instead, it provides AC power to
the System Power Switch (on the front panel) so that you can activate the
system with the System Power Switch.
Once the AC Main Switch is on ( | ), you may leave it on, even when you
unplug the console from the wall receptacle.
Note: To isolate the CUSA Excel/CUSA Excel+ from the AC power supply,
you must unplug the console from the power source.
Warning
System
TO AVOID RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE
Power
CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE EARTH.
Warning
The power cord on this product contains lead, a chemical known to the State of
California, USA, to cause cancer, and birth defects or other reproductive harm.
Wash hands after handling. The power cord complies with the DIRECTIVE
2011/65/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of
8 June 2011 on the restriction of the use of certain hazardous substances in
electrical and electronic equipment (recast).
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 3-13
Notes
3-14 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 4
Control Panel Display and
4
Functions
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 4-1
About the Control Panel Display
4-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
About the Control Panel Display
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 4-3
About the Control Panel Display
Alarm Indicators
The CUSA Excel/CUSA Excel+ activates an alarm to indicate a technical
problem with the system. All alarms on the CUSA Excel/CUSA Excel+ are
technical, low-priority alarms, for example, mechanical or equipment-
related. There are no physiological alarms on the CUSA Excel System.
When the CUSA Excel/CUSA Excel+ System activates an alarm, it:
• Illuminates the general alarm symbol in cyan on the control panel. This
indicates that an alarm exists on the system
• Illuminates one or more alarm indicators in cyan on the control panel.
This indicates the type of alarm(s) that exist
• Sounds the alarm tone four times
Table 4-1 describes each of the alarms on the CUSA Excel/CUSA Excel+
System. The general alarm symbol and alarm indicator(s) remain
illuminated until you resolve the corresponding problem(s). For
information on troubleshooting alarms, see Troubleshooting the System,
page 14-1.
4-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
About the Control Panel Display
See Figure 4-1 and Figure 4-2 for the location of the alarm indicators on
the control panel.
There is no method to silence alarms from the control panel, therefore, to
manually turn off an alarm, you must turn off the console. If the system
shuts down unexpectedly (for example a power failure) with an alarm
illuminated on the control panel, the alarm remains illuminated when you
power the system on again.The only method of turning off the alarm
indicator on the control panel is to resolve the corresponding alarm.
Recommended: When you set up the CUSA Excel/CUSA Excel+ for
surgery, make sure that the control panel is always clearly visible to the
surgeon in the event of an alarm. Remove any obstructions that may block
the surgeon’s view of the control panel.
See Figure 4-1 and Figure 4-2 for the location of the Lap Mode, Prime,
Test, and Status buttons on the control panel.
Setpoint Indicators
Green setpoint indicators show user adjusted setpoints for Aspiration,
Irrigation, and Amplitude:
• Not Activated: One setpoint indicator illuminates to show the setpoint
value
• Activated: All setpoint indicators up to and including the setpoint value
light up.
For Tissue Select®, “Std” (standard) is the first setpoint value; “++++” is the
highest setpoint value and indicates increased selectivity.
See Figure 4-1 and Figure 4-2 for the location of the setpoint indicators on
the control panel.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 4-5
About the Control Panel Display
Status Indicators:
Run (Green)
Standby (Orange)
Wait/Cooling Water (Orange)
4-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
About the Control Panel Display
Status Indicators:
Run (Green)
Standby (Orange)
Wait/Cooling Water
(Orange)
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 4-7
About the Control Panel Display
Status Indicators
To change from Standby to Run, or from Run to Standby, press the Status
button.
Run: A green status indicator illuminates to show when the system is in
Run status. The CUSA Excel/CUSA Excel+ must remain in Run mode for
a minimum of two minutes (continuous) before it is ready for use; once the
two minutes elapses, the system is ready for use.
Standby: An orange status indicator illuminates when the system is in
Standby.
Wait/Cooling Water: The Wait status indicator flashes and the Cooling
Water status indicator illuminates in the following circumstances:
• Immediately after turning on ( | ) the System Power Switch, the system
automatically fills the handpiece with cooling water.
• Immediately after turning off ( ) the System Power Switch, the
system automatically drains handpiece cooling water.
While the Wait/Cooling Water status indicators remain lit, you cannot use the
control panel or footswitch.
See Figure 4-3 and Figure 4-4 for the location of the status indicators on
the control panel.
+ 40 ms 10 ms
++ 30 ms 10 ms
+++ 20 ms 10 ms
++++ 10 ms 10 ms
4-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
About the Control Panel Display
Push a button once for a single change in setpoint value. Note: Holding
down the button will only produce a single change in setpoint value.
See Figure 4-3 and Figure 4-4 for the location of the setpoint adjustment
buttons on the control panel.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 4-9
Signs and Symbols on the Excel8 and Excel9 Control Panel
Table 4-2 Signs and Symbols on the Excel8 and Excel9 Control Panel
Symbol Description
4-10 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Signs and Symbols on the Excel8 and Excel9 Control Panel
Table 4-2 Signs and Symbols on the Excel8 and Excel9 Control Panel
Symbol Description
Suction adjust
Irrigation adjust
Amplitude adjust
Std Standard
Standby
Run
Prime
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 4-11
Signs and Symbols on the Excel and Excel 2 Control Panel
Table 4-3 Signs and Symbols on the Excel and Excel 2 Control Panel
Symbol Description
4-12 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Understanding the Control Panel at System Startup
When you turn on ( | ) the System Power Switch (on the front panel), the
CUSA Excel/CUSA Excel+ System illuminates all indicators on the control
panel, including the alarm indicators, for approximately four seconds; then
it turns off all indicators.
The CUSA Excel/CUSA Excel+ System then checks for the footswitch and
the handpiece connections. If the system detects the footswitch and
handpiece, the Wait (flashing) and the Cooling Water (solid) status
indicators turn on. After one minute, the Wait and Cooling Water status
indicators turn off. The Test and Prime indicators then turn on. If the
system cannot detect either the handpiece or footswitch connections, it
activates the appropriate alarm.
Footswitch Alarm
When turned on ( | ), if the system does not find the footswitch connection,
it illuminates the general alarm symbol and the Footswitch alarm indicator,
and it sounds the alarm tone four times. Then it turns off the Wait status
indicator. If this occurs, make sure the footswitch is connected to the
console rear panel. If not, connect it. The Footswitch alarm will turn off and
the Wait status indicator will turn on.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 4-13
Understanding the Control Panel at System Startup
Handpiece Alarm
When turned on ( | ), if the system does not find the handpiece connection,
it illuminates the general alarm symbol and the Handpiece alarm indicator,
and it sounds the alarm tone four times. Then it turns off the Wait status
indicator. If this occurs, make sure the handpiece is connected to the
console. If not, connect it. The Handpiece alarm will turn off and the Wait
status indicator will turn on.
Note: When turned off ( ), the system automatically drains the cooling
water from the handpiece and the cooling water system; the Wait status
indicator and Cooling Water status indicator illuminate. Do not disconnect
the handpiece while the system is draining cooling water. If you do, the
system illuminates the Handpiece alarm and turns off the Wait status
indicator. If this happens, reconnect the handpiece to the console. The
system turns the Wait status indicator on and the alarm off, and resumes
draining the cooling water. When the system completes draining the
cooling water, the entire control panel turns off (goes blank). Then you can
disconnect the handpiece.
System startup will not proceed until you clear a handpiece or footswitch
alarm. If you connect the handpiece and footswitch and the corresponding
alarm indicator remains on, call Integra for assistance.
4-14 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Understanding the Control Panel at System Startup
Warning
Ignoring alarms on the CUSA Excel/CUSA Excel+ System while continuing to
use the system may result in injury to the patient and/or surgical personnel, or
equipment damage.
Test Test
Use the Test button (orange box on a black button) at the top of the
Amplitude display to test the handpiece function.
To test the handpiece, press the Test button. The system automatically
activates the handpiece at 100% vibration for 4 seconds. During the test,
green setpoint indicators in the amplitude display light up to 100% to show
that the handpiece is working correctly.
If the setpoint indicators do not light up to 100%, a problem may exist. To
troubleshoot:
• Push the Test button again
• Verify proper tip torquing
• If the problem persists, perform more complete troubleshooting (see
Section 14: Troubleshooting the System)
Vibration Alarm
If the handpiece is not working properly, the system illuminates the general
alarm symbol and the Vibration alarm (left side of the control panel), and it
sounds the alarm tone four times. Possible causes for this alarm are:
• Damaged tip
• Loose tip (not properly attached to handpiece)
Prime Prime
To start Prime, press the Prime button (blue box on a black button at the
top of the Irrigation display). You need to repeat Prime until you see
irrigation fluid flow at the handpiece tip. If it doesn’t, verify that all
connections are tight, then press the Prime button again.
The Prime button automatically increases the rate of the irrigation pump to
25 to 30 milliliters per minute (ml/min). Prime allows the system to pump
irrigation fluid through the manifold tubing, through the flue, and onto the
tip as quickly as possible, reducing setup time. Watch for irrigation fluid
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 4-15
Understanding the Control Panel at System Startup
dripping from the handpiece tip. Priming takes about one minute. As the
system primes, one setpoint indicator in the Irrigation display illuminates
every 6 seconds to show its progress.
To stop Prime at any time, press the Prime button again.
You can repeat the Prime process at any time by pressing the Prime
button once to start Prime, and pressing it again to stop Prime.
After priming, the system automatically proceeds to Standby status.
4-16 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 5
Setting Up the System
5
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 5-1
Quick Reference – Setup
Handpiece Assembly
1. Attach the tip to the handpiece.
5. Attach the sterile manifold tubing (if not attached before sterilization).
Console Assembly
1. Plug the power cord into a wall receptacle.
2. Confirm that the AC Main Switch located on the rear panel is on.
9. Connect the suction manifold tubing to the console and the suction
canister.
10. Connect the irrigation manifold tubing to the console and the IV
administration set. Open the IV clamp.
11. Wait until the Wait (flashing) and Cooling Water status indicators go off.
12. Press the Test button to check the setup of the handpiece. If the cyan
vibration alarm indicator does not illuminate, the handpiece is
functioning properly.
13. Press the Prime button to prime the handpiece with sterile irrigation
fluid.
14. Adjust settings for amplitude, suction, and irrigation.
15. Press the Status button to select the Run mode when the surgeon is
ready to use the system.The CUSA Excel/CUSA Excel+ System must
remain in Run mode for a minimum of two minutes (continuous) before
it is ready for use; once the two minutes elapses, the system is ready
for use.
5-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Handpiece Assembly Options
You also have the option of attaching manifold tubing to the handpiece
cable before or after handpiece sterilization. If you chose to attach the
manifold tubing after handpiece sterilization, you must attach it in the
sterile field, see Section 10: Completing Handpiece Setup in the Sterile
Field.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 5-3
Preparing the System for Startup
a. Press the button (just above the reservoir) on the Arm Housing to
release the latch inside the housing.
b. Slide the reservoir toward the rear of the console, then remove it
from its slot.
2. Gently open the black rubber lid on the reservoir top.
3. Fill the cooling water reservoir to the line with 1000 ml distilled water
(sterile water is distilled).
Warning
Do not fill the Cooling Water reservoir with tap water or saline solution. Use
distilled water only.
off.
Min.
on, 5
SN Min.
loading,10
intermittent
VA with
Hz 900
110-120/220-240V~ operation
50/60
Continuous
Date
Mfg.
BF
Type
USA
Classification: 80301-3299S.A.
CO Center,
Boulder, Services
Inc,
Valleylab European
France
Suture
Auto Elancourt,
78990in USA
Made
5-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Preparing the System for Startup
Notice
To prevent fluid flowing into the vacuum line, only use a suction canister that has
a non-return valve.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 5-5
Preparing the System for Startup
Warning
When you connect the
handpiece to the console,
the handpiece becomes a
functional surgical device.
Caution
Sharp edge at the handpiece
connection point.
5-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Turning On the System
Warning
To avoid injury to surgical personnel, keep fingers away from the suction pinch
valve.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 5-7
Connecting the Suction Tubing
The following figure shows the suction system as it looks after successful
setup.
Figure 5-1 The assembled
suction system
5-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Connecting the Irrigation Tubing
Warning
To avoid injury to surgical personnel:
• When closing the irrigation pump latch, keep fingers away from the
area between the V-shaped tubing retainers.
• If the pump latch is open, keep fingers away from the pump rollers.
Caution
Make sure that the irrigation tubing centers between the V-shaped tubing
retainers before you close the pump latch. Otherwise, the pump latch will pinch
the tubing, preventing the flow of irrigation fluid.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 5-9
Connecting the Irrigation Tubing
Direction of
Irrigation Flow
5-10 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Testing the Handpiece
Warning
Do not use a damaged handpiece with the CUSA Excel/CUSA Excel+ System.
This may result in injury to the patient or surgical personnel.
Warning
When the handpiece is connected to a CUSA Excel/CUSA Excel+ System that
is powered on, but the handpiece is not in use, keep the handpiece away from
the patient. Place the handpiece on a sterile, flat, dry, nonconductive, and highly
visible surface.
Warning
When you connect the handpiece to the console, the handpiece becomes a
functional surgical device.
Caution
When you test the handpiece, do not allow the handpiece tip to contact anyone
or anything during tip activation. Contact may result in patient injury, user injury,
or handpiece tip damage.
When the handpiece cooling water system fills, the system automatically
proceeds to the next step and illuminates the Test and Prime setpoint
indicators.
1. Press the Test button
(orange box on a black
button at the top of the
amplitude display
column) on the control
panel.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 5-11
Testing the Alarm Tone
3. Verify that the system sounds the alarm tone four times.
If the alarm tone doesn’t sound, contact your Integra representative.
5-12 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
When the Surgeon Is Ready
Warning
Ignoring alarms on the CUSA Excel/CUSA Excel+ System while continuing to
use the system may result in injury to the patient and/or surgical personnel, or
equipment damage.
Warning
The handpiece and handpiece accessories must be sterile before surgical use.
Caution
During surgery, under maximum loading conditions, the
CUSA Excel/CUSA Excel+ console is suitable for ultrasonics activation times of
10 minutes on, 5 minutes off.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 5-13
When the Surgeon Is Ready
The CUSA Excel/CUSA Excel+ System must remain in Run mode for
a minimum of two minutes (continuous) before it is ready for use; once
the two minutes elapse, the system is ready for use. Vibration will
activate when the surgeon presses the Vibration pedal on the
footswitch.
5. Position the footswitch where it is easily accessible to the surgeon.
Caution
Before surgery, apply the brakes locks to all wheels on the console to stop the
wheels from rolling.
Notice
During surgery, do not allow the handpiece tip to touch metal objects such as
staples, clips, instruments, etc. Handpiece tip damage will result.
Warning
Touching of the tip of the handpiece by the operator, while the handpiece is
powered on, can result in personal injury.
Warning
When the handpiece is powered on, contact of the tip with a hard surface (e.g. a
metal instrument, tray, staples, clips, instruments, etc) may damage the tip of the
handpiece and require replacement before use.
Warning
CUSA Excel tips utilize silicone flues. Compressing the flue against the side of
the vibrating surface along the length of the tip can cause excessive heating and
potential hazard to adjacent tissue, such as burns.
Warning
Excessive loading of CUSA Excel tips at the surgical site can induce heating
due to vibration and acoustic power transmissions. Thermal management of the
surgical site with the aid of the appropriate irrigation and aspiration settings is
essential.
5-14 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
When the Surgeon Is Ready
Warning
Avoid excessive lateral loading of CUSA Excel tips.
Warning
Avoid contacting bone with the CUSA Excel tips (excluding SaberTip™).
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 5-15
Notes
5-16 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 6
Using the Console Controls
6
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 6-1
Adjusting Setpoint Values
Aspiration
Aspiration
The aspiration display includes 10 settings, ranging from 10 to 100% of
available suction (up to 660 mm mercury at sea level). The default setting
is 100%. Green up/down arrows mark the aspiration adjustment buttons.
Irrigation Irrigation
The irrigation display includes 10 settings. The irrigation rate ranges from
1 ml/min to 10 ml/min. The default setting is 3 ml/min. Blue up/down
arrows mark the irrigation adjustment buttons.
Amplitude Amplitude
The amplitude display includes 10 settings, from 10 to 100% of the total
amplitude (distance traveled by the tip as it vibrates) available for the
handpiece.The default setting is 10%. Most surgical procedures require a
higher setting. Orange up/down arrows mark the amplitude adjustment
buttons.
The setpoint indicators in the amplitude display show real-time feedback
from the system about the tip performance. The number of setpoint
indicators lit shows the tip performance at any given time.
6-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Changing Functions
Changing Functions
When starting the CUSA Excel/CUSA Excel+ System, switch to Run
status by pressing the green Status button.
Fast Flush
Fast Flush increases irrigation rate from the setpoint value to greater than
25 ml/min to allow additional irrigation at the surgical site. When the
surgeon activates Fast Flush, the system illuminates all setpoint indicators
in the Irrigation display to show the higher rate. When the surgeon
releases the pedal, the system returns the setpoint indicator display to the
setpoint value.
To activate Fast Flush, press the Fast Flush (left) pedal on the footswitch.
To stop Fast Flush, release the pedal.
Quiet Function
When the surgeon has not pressed either footswitch pedal for at least four
minutes, the system changes to a Quiet Function. In the Quiet Function,
the irrigation and suction pumps run at reduced speed and the machine is
quieter.
As soon as the surgeon presses either pedal, or the operator presses a
control panel button, the system returns to normal operation.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 6-3
Using the Tissue Select Feature
The Tissue Select feature increases both the inherent selectivity and the
tactile feedback of the vibrating tip by reducing the reserve power the
ultrasonic generator provides to the handpiece.
One of the variables of power consumption is time: reduce time, reduce
power consumed. To reduce power consumption, the ultrasonic generator
reduces the time that it provides electrical power. Specifically, rather than
providing power continuously, it provides power at measured on-off
intervals:
6-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Using the Tissue Select Feature
Standard Operation
Power is continuous. The ultrasonic generator provides ample reserve
power; more than is necessary to drive stroke amplitude under heavy load:
• The tip fragments “soft” tissue easily.
• The tip fragments “strong” tissue, but with more difficulty (slower,
requires more surgeon pressure).
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 6-5
Using Other Features
6-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Using Other Features
Alarm Tone
This tone (on/off) sounds when an alarm activates. You cannot adjust the
volume of the alarm tones. The alarm tone is less than 80dB.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 6-7
Using Other Features
Important
Do not try to raise the pole 1. Grasp the pole and rotate it to position B.
from Position A. Do not try to
rotate the pole when it is in the 2. With one hand, hold the pole; with the other hand, press the button to
upper position. release the locking pin.
3. Raise the pole. As the pole clears its lower locking position, release the
button.
4. Continue to move the pole upward until it locks into position.
8. After the pole has locked into its lower position, you can rotate it to
position A.
6-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 7
Handpiece Components
7
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 7-1
Components of Assembled Handpieces
Connector
Cable
O-Rings
Handpiece
Body
Flue
Connecting
Body Tip
Handpiece
The handpiece consists of a connector, a cable, a handpiece body, and a
connecting body (see Figure 7-1). The connector attaches the cable to the
CUSA Excel/CUSA Excel+ console. The cable contains:
• Cooling water tubes (deliver water from the console to cool the
handpiece)
• Electric wires (deliver electric power from console to drive the
handpiece)
The handpiece body contains these working items:
• An electric coil, which creates a magnetic field that excites the
transducer
• A transducer, which converts electric energy to mechanical motion
• Cooling water tubes; the water removes the heat generated by the
transducer in the energy conversion
The connecting body serves as the connecting point for the tip and
transfers the vibrations from the transducer to the tip.
Remember:
• The 23 kHz handpieces are gray with green connectors
7-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Components of Assembled Handpieces
O-Rings
The o-rings fit onto the connecting body in two places:
• A large o-ring fits into a groove on the handpiece housing neck
• A smaller o-ring fits into a groove near the tip end of the connecting
body
The o-rings provide stability for the nosecone and prevent fluid leaks into
the connecting body.
Use different colored o-rings for different frequency handpieces:
23 kHz 36 kHz
Nosecone
The nosecone attaches to the handpiece and covers the connecting body.
The standard nosecone is nonsterile and reusable. It serves two
purposes:
• Holds the flue in place
Standard Nosecone
• Anchors the suction tube
Tip
The tip, a hollow titanium tube, touches patient tissue. When active, the tip
vibrates at an ultrasonic frequency, causing it to fragment tissue.
Threads
The tip has two patented pre-aspiration holes, one on either side, which
help to keep the tip clear and provide better visibility at the surgical site.
Threads on one end allow you to attach the tip to the connecting body on
the handpiece.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 7-3
Components of Assembled Handpieces
For more information on various tip diameters and lengths, see Appendix
A, Technical Specifications, in this guide.
Flue
The flue (see Figure 7-1), a translucent silicone tube tapered at one end,
provides a sleeve over the tip. Irrigation fluid flows through an irrigation
connection tube at one end of the flue and down the tip to the surgical site.
Each tip size requires a tip-specific flue; therefore, you will find the tip-
specific flue packaged with the appropriate tip.
Manifold Tubing
The manifold tubing (silicone) consists of two tubes:
• Suction tubing – one end (with a light green fitting) connects to a
suction canister on the console and passes through the suction pinch
valve; the other end connects to the suction port on the handpiece
nosecone.
• Irrigation tubing – one end connects to a standard IV set at the console
and passes through the irrigation pump; the other end connects to the
flue.
The manifold tubing set also includes clips (not shown) that attach the
manifold tubing to the handpiece cable.
Manifold tubing is sterile, single patient use only. You can resterilize
unused manifold tubing once. You have two options for attaching the
tubing to the handpiece:
• Attach the manifold tubing before sterilization
• Attach the manifold tubing in the sterile field
Whether you attach the manifold tubing before or after sterilization,
discard the tubing after one use.
Notice
Attach manifold tubing only if the handpiece is to be sterilized with manifold
tubing at 132°C (270°F). If handpieces are to be sterilized at 134°C (273°F),
assembly of manifold tubing should be done in the sterile field. Refer to
Sterilization Parameters, page 9-1.
Warning
Single Use devices are for single patient use only. Do not reprocess or re-use.
Devices (s) is (are) intended to be used for one procedure only. If reprocessed
or re-used may result in the infection of patient (or patient specimen) through
cross-contamination, as well as would incur the risk of modifying the properties
and performance of the device, and of increasing the likelihood of complications
and/or undesirable effects. Once used, devices must be disposed of in
accordance with hospital policies.
7-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Additional Handpiece Components
Color coding in the tip torquing base slots, on the ends of the torque
wrench, and on each handpiece connector makes it easy to determine
where to put the handpiece in the base and which end of the torque
wrench to use when attaching a tip to a handpiece.
23 kHz green
36 kHz blue
Notice
Do not sterilize the tip torquing base or the torque wrench with steam. Steam
destroys the lubrication in the torquing mechanism, resulting in product damage.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 7-5
Additional Handpiece Components
36 kHz
Sterilize the torque bases with steam. To hold the torque bases in position
in the sterilizer case, install the restraining devices in the sterilizer case.
These restraining devices comprise of pegs that snap into the tray and
silicone straps that hold the bases to the pegs, and they are packaged
with the torque bases.
Warning
Do not use the sterile wrench for more than one surgical procedure.
7-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Additional Handpiece Components
Sterilizable torque bases and sterile wrenches bear the same color coding
as other CUSA Excel/CUSA Excel+ System components:
23 kHz green
36 kHz blue
Sterilizer Cases
The sterilizer cases hold the handpiece and accessories during
sterilization.
There are separate sterilizer cases for each handpiece:23 kHz (gray case)
and 36 kHz (white case). The dimensions for both cases are:
• length – 54.6 cm (21.5 in.)
• width – 25.4 cm (10 in.)
• depth – 12.7 cm (5 in.)
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 7-7
Additional Handpiece Components
l
ce
® Ex
SA
CU
7-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 8
Assembling the Handpiece
8
in a Nonsterile Area
In this section:
Notice
Use the information in this section only if the handpiece is to be sterilized with
items, accessories and manifold tubing at 132°C (270°F). If handpieces are to
be sterilized at 134°C (273°F), assembly should be done in the sterile field.
Refer to Sterilization Parameters, page 9-1.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 8-1
Items Needed for Handpiece Assembly
Standard
nosecone
8-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Items Needed for Handpiece Assembly
O-rings
Flue
Tip cleaner
(shipped
inside the
flue)
You also need a tip torquing base and torque wrench to assemble a
handpiece.
Manifold tubing is sterile, single use only. Discard the manifold tubing after
one use. You have two options for attaching the tubing to the handpiece:
• Attach the manifold tubing before sterilization
• Attach the manifold tubing in the sterile field
Notice
Attach manifold tubing before sterilization only if the handpiece is to be sterilized
with manifold tubing at 132°C (270°F). If handpieces are to be sterilized at
134°C (273°F), assembly of manifold tubing should be done in the sterile field.
Refer to Sterilization Parameters, page 9-1.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 8-3
Attaching the Tip to the Connecting Body
Warning
Single Use devices are for single patient use only. Do not reprocess or re-use.
Devices (s) is (are) intended to be used for one procedure only. If reprocessed
or re-used this may result in the infection of patient (or patient specimen)
through cross-contamination, as well as would incur the risk of modifying the
properties and performance of the device, and of increasing the likelihood of
complications and/or undesirable effects. Once used, devices must be disposed
of in accordance with hospital policies.
8-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Attaching the Tip to the Connecting Body
Caution
To avoid product damage, NEVER hold the handpiece in your hand while
using the torque wrench to tighten or loosen the tip.
Warning
Turning the torque wrench
further clockwise will damage
the handpiece.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 8-5
Putting On the O-Rings
23 kHz 36 kHz
Caution
Do not assemble the nosecone to the handpiece until both the handpiece and
nosecone are sterile.
8-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Packaging the Handpiece for Sterilization
Notice
Attach manifold tubing before sterilization only if the handpiece is to be sterilized
with manifold tubing at 132°C (270°F). If handpieces are to be sterilized at
134°C (273°F), assembly of manifold tubing should be done in the sterile field.
Refer to Sterilization Parameters, page 9-1.
If you decide to clip the manifold tubing onto the handpiece cable before
sterilization, follow this procedure:
1. Remove the manifold tubing with clips from the packaging. Uncoil the
tubing.
2. Snap the handpiece clip
in place on the handpiece
housing.
Items Needed
You will need the sterilizer case provided by Integra.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 8-7
Packaging the Handpiece for Sterilization
l
ce
® Ex
SA
CU
1. Put the handpiece into the case first. Align the handpiece with the
outline on the case bottom. Secure the handpiece with the silicone
strap.
2. Place the nosecone and flue with the tip cleaner in the handpiece
compartment.
3. Close the protective cover over the handpiece.
4. Coil the cable and manifold tubing (if attached) around the inside of the
case.
5. Put the handpiece connector in the center compartment and secure it
in place.
6. Put on the lid. Close and latch the case.
8-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Packaging the Handpiece for Sterilization
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 8-9
Notes
8-10 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 9
Sterilizing Handpieces and
9
Accessories
In this section:
Sterilization Parameters
Sterilize CUSA® Excel/CUSA® Excel+ handpieces and accessories with
steam. Sterilizing the handpiece, along with manifold tubing, increases the
required sterilization times. For information on preparing the handpiece for
sterilization, see Packaging the Handpiece for Sterilization, page 11-15.
The following tables provide the minimum parameters:
Packaging
Wrapped: Sterilizer case double wrapped in hospital CSR material.
Flash: Sterilizer case unwrapped.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 9-1
Sterilization Parameters
Notice
Accessories include the nosecone, manifold tubing, tip pack contents (tip, flue,
tip cleaner, o-rings) and sterilizable tip torquing base.
9-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Sterilizing the Handpiece with Steam
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 9-3
Notes
9-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 10
Completing Handpiece Setup
10
In this section:
Notice
Retaining a spare handpiece in the sterile field is highly recommended.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 10-1
Quick Reference – Setup
Handpiece Assembly
1. Attach a standard nosecone onto the handpiece.
3. Attach the sterile manifold tubing (if not attached before sterilization).
Items Needed
You need the following items to set up the handpiece for surgery:
• Standard nosecone
• Flue
• Manifold tubing
10-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Completing the Handpiece Assembly
Attach a Nosecone
1. Attach a standard
nosecone. Holding the
handpiece, insert the tip
into the nosecone.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 10-3
Completing the Handpiece Assembly
10-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Completing the Handpiece Assembly
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 10-5
Completing the Handpiece Assembly
10-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 11
Assembling or Changing
11
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 11-1
Items Needed to Change Tip in Sterile Field
You need the following items to change tips in the sterile field:
• Sterilized handpiece(s)
• Sterilized standard nosecone
• Sterile tip/flue pack(s) or sterilized extended life tips and flues
• Sterilized tip torquing base
• Sterile torque wrench (23 kHz or 36 kHz)
• Sterile manifold tubing
The following figures illustrate these items.
Figure 11-1 The sterilized
handpiece
11-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Items Needed to Change Tip in Sterile Field
The contents of sterile tip/flue packs are sterile, single use only. You can
resterilize unused tips and flues once.
The contents of extended life tip/flue packs are nonsterile. You can use
them six times. If you use extended life tips and intend to attach the tip in
the sterile field, ensure tips and flues are sterile before they are introduced
to the sterile field.
You have two options for attaching tips to the handpiece:
• Attach the tip in the sterile field
• Attach the tip before sterilizing the handpiece (If you use this option, do
not change tips in the sterile field because the area under the tip
threads will not be sterile)
Figure 11-3 Sterile tip/flue pack
contents. The tip cleaner is inside
the flue.
Tip
O-rings
Flue
Tip
cleaner
Steam sterilize the tip torquing base before each surgical procedure in
which you change tips in the sterile field.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 11-3
Items Needed to Change Tip in Sterile Field
36 kHz
(Blue)
You can use the sterile torque wrench to change as many as five tips in
one surgical procedure. Do not use the sterile torque wrench in more than
one surgical procedure. Do not sterilize the sterile torque wrench; dispose
of the used sterile torque wrench in accordance with your Healthcare
Facility protocols.
Figure 11-5 Sterile torque
wrench (23 kHz or 36 kHz)
Manifold tubing is sterile, single patient use only. You can resterilize
unused manifold tubing once. You have two options for attaching the
tubing to the handpiece:
• Attach the manifold tubing in the sterile field
• Attach the manifold tubing before sterilizing the handpiece
Figure 11-6 Sterile manifold
tubing
11-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Assembling the Tip
Notice
Attach manifold tubing only if the handpiece is to be sterilized with manifold
tubing at 132°C (270°F). If handpieces are to be sterilized at 134°C (273°F),
assembly of manifold tubing should be done in the sterile field. Refer to
Sterilization Parameters, page 9-1.
Warning
Single Use devices are for single patient use only. Do not reprocess or re-use.
Devices (s) is (are) intended to be used for one procedure only. If reprocessed
or re-used this may result in the infection of patient (or patient specimen)
through cross-contamination, as well as would incur the risk of modifying the
properties and performance of the device, and of increasing the likelihood of
complications and/or undesirable effects. Once used, devices must be disposed
of in accordance with hospital policies.
Warning
The handpiece and handpiece accessories must be sterile before surgical use.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 11-5
Assembling the Tip
Caution
To avoid product damage, NEVER hold the handpiece in your hand while
using the torque wrench to tighten or loosen the tip.
Warning
Turning the torque wrench
further clockwise will damage
the handpiece.
11-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Assembling the Tip
23 kHz 36 kHz
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 11-7
Assembling the Tip
5. Remove the tip cleaner from the flue and set it aside. You may need it
during the surgical procedure to remove tissue blockage from the tip,
the suction connection to the handpiece, or both.
6. Slide the flue that
corresponds to the
selected tip over the tip
and onto the nosecone.
Markings on the flue indicate the tip the flue should cover:
S Standard Tip
Mi MicroTipTM, ShearTipTM,
SaberTipTM
P PrecisionTipTM
Ma MacroTipTM
Make sure to push the flue base completely over the nosecone base,
and that the end of the flue lines up with the pre-aspiration holes.
11-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Assembling the Tip
4. (Viewed from rear) Connect the manifold irrigation tubing to the Luer
fitting on the handpiece flue.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 11-9
Disassembling Handpiece to Change Tips in the Sterile Field
11-10 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Disassembling Handpiece to Change Tips in the Sterile Field
DO NOT DISCARD A
STANDARD
NOSECONE.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 11-11
Disassembling Handpiece to Change Tips in the Sterile Field
2. Locate the sterile torque base in which the handpiece support color
matches the color on the handpiece connector (23 kHz – green;
36 kHz – blue).
3. Put the handpiece in the
torque base so that the
metal connecting body
fits snugly in the metal
slot at the end of the
support.
Caution
To avoid product damage, NEVER hold the handpiece in your hand while
using the torque wrench to tighten or loosen the tip.
11-12 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Disassembling Handpiece After the Surgical Procedure
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 11-13
Resterilizing the Handpiece
23 kHz 36 kHz
11-14 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Packaging the Handpiece for Sterilization
Cleaning efficacy has been validated using enzymatic cleaning agents (for
example, Klenzyme® and Enzol®) according to the manufacturer's
instructions.
After disassembling the tip from the handpiece, clean, then sterilize, the
sterilizable torque base. Refer to the cleaning instructions below.
Notice
Use the information in this section only if the handpiece is to be sterilized with
items, accessories and manifold tubing at 132°C (270°F). If handpieces are to
be sterilized at 134°C (273°F), sterilize the handpiece separate from the
accessories and manifold tubing. Refer to Sterilization Parameters, page 9-1.
1. Put the disassembled handpiece into the case first. Align the
handpiece with the outline on the case bottom. Secure the handpiece
with the silicone strap.
2. Close the protective cover over the handpiece.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 11-15
Packaging the Base for Sterilization
4. Coil the cable and manifold tubing (if attached) around the inside of the
case.
5. Put the handpiece connector in the center compartment and secure it
in place.
Figure 11-7 The handpiece and
connector secured into the
sterilizer tray
Do not close the sterilizer case until you clean and load the sterilizable
torque base.
Notice
Package the torque base with handpiece only if the handpiece is to be sterilized
with the torque base at 132°C (270°F). If the handpiece is to be sterilized at
134°C (273°F), sterilize the torque base separate from the handpiece. Refer to
Sterilization Parameters, page 9-1.
1. Load the sterilizable torque base into the CUSA® Excel/CUSA® Excel+
sterilizer tray and secure it with the silicone strap.
11-16 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Sterilize the Handpiece and Base
23kHz 36kHz
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 11-17
Notes
11-18 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 12
Shutting Down, Disconnecting
12
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 12-1
Shutting Down the System
3. While the system displays the flashing Wait status indicator and the
Cooling Water status indicator, remove the green-banded suction
tubing from the pinch valve.
When the flashing Wait status indicator goes off, the control panel goes
blank and the system automatically powers down. When this occurs, the
pinch valve automatically closes. If you haven’t already removed the
suction tubing before the pinch valve closes, use the valve override button
on the front of the pinch valve to open the valve and remove the tubing.
Manually opening the valve requires a strong push on the button.
Disconnect the tubing and the handpiece, as described in the next
procedure. Power down the system from the rear of the unit and unplug
the system from the wall receptacle.
Notice
Do not disconnect the handpiece until the control panel goes completely blank.
Otherwise, product damage may result.
12-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Disconnecting Suction Tubing, Irrigation Tubing, and the Handpiece
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 12-3
Disconnecting Suction Tubing, Irrigation Tubing, and the Handpiece
Warning
To avoid injury to surgical
personnel:
• When closing the
irrigation pump latch,
keep fingers away from
the area between the V-
shaped tubing retainers.
• If the pump latch is open,
keep fingers away from
the pump rollers.
Caution
Sharp edge at the handpiece
connection point.
Important
A standard nosecone is not a Disconnect the aspiration line from the nosecone.
disposable component. Do not
discard a standard Remove the flue from the tip.
nosecone. Remove the tip from the handpiece.
Unclip the manifold tubing from the handpiece and handpiece
cable.
Discard all of these items in the patient’s biohazard waste container.
Do not discard a reusable tip. Reusable tips can be reused six times.
12-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Cleaning the Console
a. Press the button (just above the reservoir) on the Arm Housing to
release the fittings inside the housing.
b. Slide the reservoir toward the rear of the console, then remove it
from its slot.
c. Gently open the black rubber lid on the reservoir top and discard
the distilled water.
d. Gently close the black rubber lid.
e. Slide the empty reservoir back into the slot at the base of the
adjustable arm until it snaps into place.
DO NOT DISCARD THE COOLING WATER RESERVOIR.
Warning
Electric Shock Hazard – Always unplug the CUSA Excel/CUSA Excel+ System
before cleaning it.
Notice
Do not rub, press, or touch any panels with solvents; caustic, corrosive, or
abrasive cleaning or disinfectant compounds, or other materials that could
scratch the panels. Do not use a betadine-based solution because it will cause
discoloration. Do not allow fluids to enter the chassis.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 12-5
Notes
12-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 13
Disassembling and Cleaning
13
Handpieces
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 13-1
Disassembling the Handpiece
tip cleaner
o-rings
tip
Note: Do not discard a reusable tip. Reusable tips can be reused six
times.
When cleaning handpieces, wear protective clothing, gloves, and safety
glasses in accordance with your hospital’s cleaning policy.
1. Using a soft cloth
moistened with a
germicidal solution, wipe
the handpiece to remove
gross contaminants.
a. Release the
nosecone by twisting
it counterclockwise.
b. Pull the nosecone
away from the
handpiece.
13-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Disassembling the Handpiece
Caution
To avoid product damage, NEVER hold the handpiece in your hand while
using the torque wrench to tighten or loosen the tip.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 13-3
Cleaning the Handpiece
When you have disassembled the handpiece, clean the handpiece, the tip
torquing base, and the torque wrench as described in the procedures
below.
Notice
Product damage will result if you do not follow these notices when cleaning the
handpiece:
• Do not immerse the handpiece cable electrical connector in liquid.
• Do not use ultrasonic washers.
• Do not use chlorinated substances such as bleach solution.
• Do not clean the handpiece with abrasives such as Ajax®, Comet®, or steel
wool.
13-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Cleaning the Handpiece
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 13-5
Cleaning the Handpiece
13-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Permitted Number of Reprocessing Cycles for Handpieces when Cleaned with an Automatic Washer
Handpiece Cycles
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 13-7
Notes
13-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 14
14 Troubleshooting the System
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 14-1
Responding to Alarms
Responding to Alarms
Footswitch Footswitch cable not properly Connect the footswitch to the console (rear panel).
connected to console Turn the connector until you feel it lock into place.
If the alarm persists after you connect the footswitch
to the console, call Integra for assistance.
Handpiece Handpiece cable not properly Connect the handpiece to the console. Twist the
connected to console handpiece connector clockwise until it locks into
place, then verify that the yellow dots align.
If the alarm persists after you connect the handpiece
to the console, call Integra for assistance.
Cooling Water Not enough distilled water in the Add distilled water to the cooling water reservoir up
cooling water reservoir to the fill line.
Cooling water reservoir not properly Re-seat the cooling water reservoir in the console.
seated
A pinch or kink in the handpiece cable Remove the pinch or kink in the cable.
Handpiece cable not properly Connect the handpiece to the console. Twist the
connected to console handpiece connector clockwise until it locks into
place, then verify that the yellow dots align.
Cooling Water–On Possible air bubble in the handpiece The CUSA Excel/CUSA Excel+ System corrects this
Intermittently cooling system condition. However, if the Cooling Water alarm
remains lit, refer to the condition immediately before
this one: Cooling Water – continuously on.
Note: If the Cooling Water alarm persists (it
continues to illuminate intermittently), call Integra for
assistance.
A pinch or kink in the handpiece cable Remove the pinch or kink in the cable.
14-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Responding to Alarms
Vibration–In Test Tip loose due to incorrect assembly Testing before sterilization:
mode only
1. Use the tip torquing base and wrench to tighten
the tip.
2. Test the tip by testing the handpiece. See Testing
the Handpiece, page 5-11.
3. Sterilize the handpiece.
Tip contacting another object Remove the contact with any other object.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 14-3
Responding to Alarms
Vibration–In Test Blocked tip 1. Verify that the manifold suction tube connects
mode only properly to the nosecone by disconnecting, then
reconnecting, the tubing and its clip.
2. Use the tip cleaner to dislodge the blockage in
the tip. Continue to push the tip cleaner into the
tip until the cleaner becomes visible in the
tubing.
3. When you have cleared the blockage, verify
again that the manifold suction tube connects
properly to the nosecone.
4. If the tip remains blocked, disconnect the
manifold suction tube connector from the
nosecone. Use the tip cleaner to dislodge
blockage from the tubing, then reconnect the
tubing to the nosecone.
5. Immerse the handpiece tip in sterile irrigation
fluid briefly. This may clear the suction tubing.
14-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
General Troubleshooting
General Troubleshooting
No power at the console AC Main Switch (on the rear Turn on the AC Main Switch (on the
panel) off rear panel).
System Power Switch (front of Turn on the System Power Switch (on
console) off the front of the console).
Power cord not plugged in Plug the power cord into the wall
receptacle.
Power interruption to the console Power cord becomes unplugged 1. Turn off the System Power Switch
(on the front of the console).
2. Plug the power cord into the wall
receptacle.
3. When power is restored, turn on
the System Power Switch.
4. When the Wait status indicator
(flashing) turns off, press the
Prime button to make sure the
irrigation system is ready to use.
5. Adjust the Aspiration, Irrigation,
Amplitude, and Tissue Select®
settings.
6. Push the Status button to change
to Run mode.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 14-5
General Troubleshooting
Power interruption in the operating Facility power loss 1. Turn off the System Power Switch
room (on the front of the console).
2. When power is restored, turn on
the System Power Switch.
3. When the Wait status indicator
(flashing) turns off, press the
Prime button to make sure the
irrigation system is ready to use.
4. Adjust the Aspiration, Irrigation,
Amplitude, and Tissue Select
settings.
5. Push the Status button to change
to Run mode.
No irrigation flow from the IV administration set roller clamp Open the roller clamp.
handpiece flue; the handpiece tip closed
gets hot
Irrigation tubing not routed properly Route the irrigation tubing properly
through irrigation pump through the irrigation pump.
Not enough time set at Prime to allow Press the Prime button, and wait until
fluid to pass through the irrigation irrigation fluid drips from the tip.
tubing
Irrigation tubing pinched or kinked Remove the pinch or kink from the
tubing.
14-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
General Troubleshooting
Excessive misting at the Flue not lined up with tip Adjust the flue position to line up with
handpiece tip pre-aspiration holes the pre-aspiration holes in the tip. See
Attach the Tip and O-rings, page 11-5.
Low or no suction due to suction Verify that the manifold suction tube
tubing improperly connected to connects properly to the nosecone by
handpiece disconnecting, then reconnecting, the
tubing and its clip.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 14-7
General Troubleshooting
Excessive misting at the Blocked tip 1. Verify that the manifold suction
handpiece tube connects properly to the
tip–continued nosecone by disconnecting, then
reconnecting, the tubing and its
clip.
2. Use the tip cleaner to dislodge the
blockage in the tip. Continue to
push the tip cleaner into the tip
until the cleaner becomes visible
in the tubing.
3. When you have cleared the
blockage, verify again that the
manifold suction tube connects
properly to the nosecone.
4. If the suction remains low,
disconnect the manifold suction
tube connector from the
nosecone. Use the tip cleaner to
dislodge blockage from the
tubing, then reconnect the tubing
to the nosecone.
5. Immerse the handpiece tip in
sterile irrigation fluid briefly. This
may clear the blockage.
14-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
General Troubleshooting
Excessive misting at the Suction connections at the suction 1. Remove any unused suction
handpiece canister not seated properly canister “elbow” connectors.
tip–continued
2. Reseat all connections.
Open port(s) on the suction canister Close any open ports on the suction
canister.
Suction pinch valve not working Check the Lap mode indicator.
properly
• If Lap mode is off, in Run mode the
pinch valve should be open.
• If Lap mode is on, in Run mode the
pinch valve should be open only
when you activate vibration.
If these conditions do not exist, call
Integra for assistance.
The Run mode status indicator System in Standby mode Press the Status button to switch the
does not illuminate when you system to Run mode.
activate vibration.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 14-9
General Troubleshooting
Little or no vibration or Excessive tip/tissue pressure (stalling 1. Verify that console settings are at
fragmentation (Amplitude setpoint tip) desired levels.
indicators may illuminate well
2. Remove the tip from contact with
below setpoint)
tissue, then reactivate vibration.
Note: When first activating
vibration, make sure that there is
no tip/tissue pressure.
14-10 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
General Troubleshooting
Little or no vibration or Tip loose due to improper assembly 1. Press the Status button to place
fragmentation (Amplitude setpoint the system in Standby mode.
indicators may illuminate well
2. Press the Test button to test the
below setpoint)–continued
handpiece and tip.
3. If the Vibration alarm activates, do
one of the following:
If tip assembled in a nonsterile area:
• Remove the handpiece from the
sterile field.
• Use the tip torquing base and
wrench to tighten the tip.
• Test the tip by testing the
handpiece.
• Sterilize the handpiece.
If sterile tip assembled to handpiece
in the sterile field:
• Use the sterilizable torque base
and a disposable sterile wrench
to tighten the tip. (See Section
11: Assembling or Changing
Tips in the Sterile Field)Test the
tip by testing the handpiece.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 14-11
General Troubleshooting
Little or no vibration or Damaged or cracked tip 1. Press the Status button to place
fragmentation (Amplitude setpoint the system in Standby mode.
indicators may illuminate well
2. Press the Test button to test the
below setpoint)–continued
handpiece and tip.
3. If the Vibration alarm activates, do
one of the following:
If tip assembled in a nonsterile area:
• Remove the handpiece from the
sterile field.
• Use the tip torquing base and
wrench to replace the tip.
• Test the tip by testing the
handpiece.
• Sterilize the handpiece.
Suction tube won’t assemble Nosecone not assembled to Align the dot on the nosecone with
correctly to handpiece nosecone handpiece correctly the dot on the handpiece.
14-12 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
General Troubleshooting
Low suction at the handpiece tip Low suction setting on the control Increase the setting.
panel
System inadvertently in Lap mode Turn off the Lap mode by pressing the
Lap mode button.
Suction tubing improperly connected Verify that the manifold suction tube
to handpiece connects properly to the nosecone by
disconnecting, then reconnecting, the
tubing and its clip.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 14-13
General Troubleshooting
Low suction at the handpiece A blockage in the suction tubing 1. Examine the suction tubing along
tip–continued between the handpiece and the its entire length for a blockage.
suction canister
2. Squeeze the blocked area to
loosen it.
3. Immerse the handpiece tip in
sterile irrigation fluid briefly. This
may clear the blockage.
Open port(s) on the suction canister Close any open ports on the suction
canister.
Problem at the console suction port Remove the contamination guard and
put a finger in the suction system port.
If you feel no suction, call Integra for
assistance.
14-14 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
General Troubleshooting
The handpiece gets hot Surgeon holding handpiece at tip and Hold the handpiece at the nosecone.
flue
The nosecone is loose Large o-ring not installed correctly, or Install the large o-ring in the proper
missing position on the handpiece.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 14-15
Notes
14-16 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
SECTION 15
15 Maintaining the System
In this section:
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 15-1
Quick Reference
Quick Reference
The following chart lists routine maintenance tasks, when you perform
each task, and the equipment on which you perform them.
Table 15-1
Routine maintenance tasks for
CUSA® Excel/CUSA® Excel+ When Equipment Task
System
Daily or when used CUSA Excel/CUSA Disinfect and wipe dry.
Excel+ console
Daily or when used Cooling Water Discard water from the reservoir.
Reservoir
15-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Handpiece Maintenance
Handpiece Maintenance
Maintaining the handpiece involves lubricating the handpiece connector o-
rings or replacing the handpiece connector o-rings.
Handpiece Cycles
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 15-3
Sterilizer Case Maintenance
Insertion Tool
Notice
Do not clean the sterilizer case with abrasives. Product damage will result.
15-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Storage of the System and Accessories
Console
Drain all liquids. Make sure the system is clean. Store the system in a low
traffic area that is free of dirt, blood, water, and other contaminants. Store
the system at an ambient temperature between -34°C and 65°C.
Handpiece
Drain all liquids. Store the handpiece in the sterilizer case according to
your facility’s policy.
Store the handpiece at an ambient temperature between -34°C and 65°C.
Footswitch
You do not need to disconnect the footswitch from the console except for
maintenance or service. When you are not using the
CUSA Excel/CUSA Excel+ System, secure the footswitch cord on cord
wraps provided on the back of the console and store the footswitch in the
storage compartment in the rear of the console.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 15-5
Return Equipment for Service
Returning Handpieces
Clean, disinfect, and sterilize a handpiece before you package it for
shipping.
Package a handpiece to protect the handpiece and handpiece cable
connector from damage.
Package each handpiece in a separate packaging container.
15-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Integra Service Centers
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide 15-7
Notes
15-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
APPENDIX A
A Technical Specifications
Console Dimensions
Footswitch
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide A-1
Console Subsystems
Console Subsystems
Ultrasonic
Fluidic System
A-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Electrical Requirements
Electrical Requirements
Frequencies 50 Hz ± 2 Hz
60 Hz ± 2 Hz
Power Cords
Warning
Explosion Hazard – Do not use the CUSA Excel/CUSA Excel+ System in the
presence of flammable anesthetics or any potentially explosive or flammable
atmosphere.
Warning
TO AVOID RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE
CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE EARTH.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide A-3
Environment
Warning
The console must be earthed, and therefore it MUST only be fitted with a 3-pin
plug, or a 2-pin plug that has an integral earth grounding connection. Mains plug
type and construction MUST comply with Legal requirements within country of
installation. Only Integra Service personnel or Integra authorized
representatives or agents can change the mains plug on the console.
Duty Cycle
Under maximum loading conditions, the CUSA Excel/CUSA Excel+ console is
suitable for ultrasonics activation times of 10 minutes on, 5 minutes off.
Environment
The system requires a minimum of one hour exposure at its operating temperature
range before you use it.
See section EMC Compatibility on page A-5 for further information on
environmental conditions.
Electromagnetic Interference
The CUSA Excel/CUSA Excel+ System console minimizes electromagnetic
interference to other equipment used in the operating room. The system complies
with the requirements of IEC 60601-1-2:2007.
Note that other devices in the operating room may generate electromagnetic
interference. Use caution in locating equipment within the room to reduce the
electromagnetic interference.
A-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Voluntary Standards
Voluntary Standards
The CUSA Excel/CUSA Excel+ System meets the following standards:
• IEC 60601-1:2005 Medical electrical equipment: General requirements for
basic safety and essential performance
• cUL/CSA Test requirements: (CSA 22.2, No. 601.1.8)
• IEC 60601-1-6:2005 Medical electrical equipment: General requirements for
basic safety and essential performance. Collateral Standard: Usability
• IEC 60601-1-2:2005 Medical electrical equipment: General requirements for
basic safety and essential performance. Collateral standard: Electromagnetic
compatibility - Requirements and tests
• CISPR11 Electromagnetic compatibility for industrial-process measurement
and control equipment: Emissions requirements
• ISTA Pre-Shipment Test Procedures
EMC Compatibility
Notice
The CUSA Excel/CUSA Excel+ System should not be used adjacent to or
stacked with equipment other than the equipment specified in the
CUSA Excel/CUSA Excel+ System User Guide. If adjacent or stacked use is
necessary, the CUSA Excel/CUSA Excel+ System should be observed to verify
normal operation in the configuration in which it will be used.
The use of accessories, other than the accessories specified in the
CUSA Excel/CUSA Excel+ System User Guide may result in increased
emissions or decreased immunity of the CUSA Excel/CUSA Excel+ System.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide A-5
Voluntary Standards
The CUSA Excel/CUSA Excel+ System is intended for use in the electromagnetic environment specified below. The
customer or the user of the CUSA Excel/CUSA Excel+ System should assure that it is used in such an environment.
RF emissions Class A The CUSA Excel/CUSA Excel+ System is suitable for use in
all establishments other than domestic and those directly
CISPR 11 connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonics emissions Class A
IEC 61000-3-2
IEC 61000-3-3
A-6 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Voluntary Standards
The CUSA Excel/CUSA Excel+ System is intended for use in the electromagnetic environment specified below. The
customer or the user of the CUSA Excel/CUSA Excel+ System should assure that it is used in such an environment.
Electrical fast ± 2 kV for power supply ± 2 kV for power Mains power quality should be that of a
transient/burst lines supply lines typical commercial or hospital
environment.
IEC 61000-4-4 ± 1 kV for input/output ± 1 kV for input/output
lines lines
Surge ± 1 kV line(s) to line(s) ± 1 kV line(s) to line(s) Mains power quality should be that of a
typical commercial or hospital
IEC 61000-4-5 ± 2 kV line(s) to earth ± 2 kV line(s) to earth environment.
Voltage dips, short <5 % UT <5 % UT Mains power quality should be that of a
interruptions and (>95 % dip in UT) (>95 % dip in UT) typical commercial or hospital
voltage variations on for 0,5 cycle for 0,5 cycle environment. If the user of the CUSA
power supply input Excel/CUSA Excel+ System requires
lines. <40 % UT <40 % UT continued operation during power mains
(>60 % dip in UT) (>60 % dip in UT) interruptions, it is recommended that the
IEC 61000-4-11 for 5 cycles for 5 cycles CUSA Excel/CUSA Excel+ System be
powered from an uninterruptible power
<70 % UT <70 % UT supply or a battery.
(>30 % dip in UT) (>30 % dip in UT)
for 25 cycles for 25 cycles
<5 % UT <5 % UT
(>95 % dip in UT) (>95 % dip in UT)
for 5 s for 5 s
Power frequency (50/60 3 A/m 3 A/m Power frequency magnetic fields should
Hz) magnetic field be at levels characteristic of a typical
location in a typical commercial or hospital
IEC 61000-4-8 environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide A-7
Voluntary Standards
The CUSA Excel/CUSA Excel+ System is intended for use in the electromagnetic environment specified below. The
customer or the user of the CUSA Excel/CUSA Excel+ System should assure that it is used in such an environment.
Conducted RF 3 Vrms
3 Vrms
IEC 61000-4-6 150 kHz to 80 MHz
IEC 61000-4-3
3 V/m Where P is the maximum output power rating of the
80 MHz to 2,5 GHz
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
A-8 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Voluntary Standards
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the CUSA Excel/CUSA Excel+ System
is used exceeds the applicable RF compliance level above, the CUSA Excel/CUSA Excel+ System should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orientating or relocating the CUSA Excel/CUSA Excel+ System.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide A-9
Voluntary Standards
The CUSA Excel/CUSA Excel+ System is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the CUSA Excel/CUSA Excel+ System can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the CUSA Excel/CUSA Excel+ System as recommended below, according to the
maximum output power of the communications equipment.
W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
100 12 m 12 m 24 m
For transmitters rated at maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
A-10 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Handpieces
Handpieces
Nominal Frequencies
Dimensions
23 kHz Straight
Length 22.2 cm (8.74 in)
23 kHz Angled
Length 22.2 cm (8.74 in)
36 kHz Straight
Length 13.9 cm (5.47 in)
Tip Specifications
Table A-1 on page A-12 shows specifications for tips that attach to the 23 kHz
handpiece.
Table A-2 on page A-13 shows specifications for tips that attach to the 36 kHz
handpiece.
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide A-11
Tip Specifications
Standard Tip 79.8 [3.140] 1.98 [0.078] 2.54 [0.100] Straight Handpiece: 3.54
287 to 355 [0.0113 to 0.0140]
Angled Handpiece:
112 to 163 [0.0044 to 0.0064]
MicroTip™ 92.2 [3.630] 1.57 [0.062] 1.93 [0.076] Straight Handpiece: 3.99
279 to 355 [0.0110 to 0.0140]
Angled Handpiece:
127 to 183 [0.005 to 0.0072]
Curved Extended 183.8 [7.238] 1.98 [0.078] 2.54 [0.100] Straight Handpiece: 16.15
Standard Tip 287 to 355 [0.0113 to 0.0140]
Not compatible with angled
handpiece
MacroTip™ 79.5 [3.130] 2.64 [0.104] 3.18 [0.125] Straight Handpiece: 3.84
254 to 307 [0.0100 to 0.0121]
Angled Handpiece:
102 to 142 [0.0040 to 0.0056]
Laparoscopic Tip 301.4 [11.865] 1.98 [0.078] 2.54 [0.100] Straight Handpiece: 32.13
178 to 229 [0.007 to 0.009]
Not compatible with angled
handpiece
Extended Life 79.8 [3.140] 1.98 [0.078] 2.54 [0.100] Straight Handpiece: 3.54
Standard Tipa 287 to 355 [0.0113 to 0.0140]
Angled Handpiece:
112 to 163 [0.0044 to 0.0064]
Extended Life 79.5 [3.130] 2.64 [0.104] 3.18 [0.125] Straight Handpiece: 3.84
MacroTipb 254 to 307 [0.0100 to 0.0121]
Angled Handpiece:
102 to 142 [0.0040 to 0.0056]
A-12 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
Tip Specifications
PrecisionTip™ 56.4 [2.221] 1.14 [0.045] 1.45 [0.057] 191 to 210 [0.0075 to 0.0083] 1.37
Straight Extended 125.1 [4.924] 1.14 [0.045] 1.45 [0.057] 191 to 210 [0.0075 to 0.0083] 7.86
PrecisionTip
Curved Extended 125.1 [4.924] 1.14 [0.045] 1.45 [0.057] 191 to 210 [0.0075 to 0.0083] 7.86
PrecisionTip
MicroTip™ 52.8 [2.080] 1.57 [0.062] 1.93 [0.076] 175 to 193 [0.0069 to 0.0076] 1.30
Straight Extended 121.5 [4.783] 1.57 [0.062] 1.93 [0.076] 175 to 193 [0.0069 to 0.0076] 7.50
MicroTip
Curved Extended 121.5 [4.783] 1.57 [0.062] 1.93 [0.076] 175 to 193 [0.0069 to 0.0076] 7.50
MicroTip
Extended 192.7 [7.588] 1.57 [0.062] 1.93 [0.076] 137 to 155 [0.0054 to 0.0061] 7.97
MicroTip Plus
Standard Tip 45.7 [1.800] 1.98 [0.078] 2.54 [0.100] 137 to 155 [0.0054 to 0.0061] 1.31
Straight Extended 114.4 [4.504] 1.98 [0.078] 2.54 [0.100] 137 to 155 [0.0054 to 0.0061] 7.48
Standard Tip
Curved Extended 114.4 [4.504] 1.98 [0.078] 2.54 [0.100] 137 to 155 [0.0054 to 0.0061] 7.48
Standard Tip
CUSA® 114.9 [4.525] 1.14 [0.45] 2.23 [0.088] 117 to 135 [0.0046 to 0.0053] 9.55
SaberTip™
CUSA® 117.9 [4.643] 1.57 [0.062] 2.33 [0.0919] 165 to 203 [0.0065 to 0.0080] 7.77
ShearTip™
Extended Life 125.1 [4.924] 1.14 [0.45] 1.45 [0.057] 175 to 210 [0.0069 to 0.0083] 7.86
Curved Extended
PrecisionTipa
Extended Life 121.5 [4.783] 1.57 [0.062] 1.93 [0.076] 165 to 193 [0.0065 to 0.0076] 7.50
Curved Extended
MicroTipb
Extended Life 114.4 [4.504] 1.98 [0.078] 2.54 [0.100] 137 to 155 [0.0054 to 0.0061] 7.48
Curved Extended
Standard Tipc
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide A-13
Notes
A-14 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
APPENDIX B
B Sterilization Validation
Challenge
A minimum of 1 x 106 bacillus stearothermophilus on a spore strip/disk carrier and
a braided nylon suture carrier. (D121°C = 1.5 and 1.6 minutes).
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide B-1
Validation of Steam Sterilization Parameters
Acceptance Criteria
In order for the handpiece to be considered validated, none of the biological
indicators demonstrated growth of the indicator organism. A 6-log reduction at the
half cycle exposure time indicates that a full cycle (twice the half cycle exposure
time) will produce sufficient lethality to effect at least a 12-log reduction and
provide a 10-6 sterility assurance level (SAL).
B-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
APPENDIX C
C Warranty
Integra Warranty for CUSA® Excel and CUSA® Excel+ Ultrasonic Surgical
Aspirator Systems
1. Warranty.
INTEGRA LIFESCIENCES CORPORATION and its wholly owned subsidiaries
("Integra") warrant to Integra authorized distributors and the original purchaser
only that each new Integra CUSA Excel and CUSA Excel+ Ultrasonic Surgical
Aspirator (hereinafter the Equipment or the product) is free from manufacturing
defects in material and workmanship under normal use and service for a period of
one (1) year (except as otherwise expressly provided as to accessory items) from
the date of invoice by Integra (or its authorized distributor) to the original
purchaser, but in no event beyond the expiration date stated on any product
labeling (hereinafter the Warranty Period). For purposes of products sold by
Integra through an authorized distributor of Integra, "original purchaser" shall
include the purchaser of Integra products to whom the distributor first sells the
product. The original purchaser is hereinafter referred to as Customer.
1.1. Coverage. During the Warranty Period, Integra shall provide free-of-
charge service and maintenance consistent with the provisions of Section 3 of
this Warranty, so that the Equipment conforms to the specifications defined in
the CUSA Excel and CUSA Excel+ Ultrasonic Surgical Aspirator Operator’s
Manuals, as such Operator’s Manuals may be modified by Integra from time to
time (the “Specifications”).
1.2. Exclusions. The Warranty shall not apply in any manner to service or
maintenance of the Equipment, or to replacement of its parts, with respect to:
(i) use of Equipment with any tips, flues, and manifold tubing sets and
accessories other than those manufactured by Integra LifeSciences;
(ii) defects arising out of materials or parts provided, modified or designed
by anyone other than an authorized Integra service agent (the Integra
“Service Agent”);
(iii) defects emanating from improper or negligent installation, storage or
use of the Equipment or any component thereof, including but not limited to
operating the Equipment not in accordance with instructions provided in the
Operator’s Manual;
(iv) defects arising from improper or negligent cleaning or sterilization
methods or improper maintenance of the Equipment;
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide C-1
(v) defects resulting from repairs or service of the Equipment provided
other than by Integra or its authorized representatives;
(vi) defects arising from accidental damage to the Equipment, acts of God,
electrical power damage, equipment malfunction, unusual stress,
unreasonable operating procedures or abnormal or extreme operating
conditions; and
(vii) normal wear and tear.
C-2 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide
4. Quality Control.
4.1. Customer shall maintain reasonable standards of quality control,
operations, procedures, safety testing and inspection of Equipment to ensure
that unnecessary service or maintenance is not required hereunder.
4.2. Customer shall provide a technical counterpart to Integra's Service Agent
for assistance in Integra's telephonic diagnosis of the malfunction with the
Equipment. Customer shall reasonably accept Integra's determination whether
a repair or service is an in-warranty repair or an out-of-warranty repair.
5. Limitation of Liability.
5.1. INTEGRA's sole responsibility under the warranties described in Section 1
shall be repair or replacement, at INTEGRA's sole discretion at INTEGRA's
expense, subject to the terms of this warranty and applicable agreements. THE
WARRANTIES DESCRIBED IN SECTION 1 HEREOF ARE EXCLUSIVE AND
ARE GIVEN AND ACCEPTED IN LIEU OF ALL OTHER WARRANTIES OF
INTEGRA OR ITS SERVICE AGENTS WITH RESPECT TO THE QUALITY,
PERFORMANCE AND OPERATION OF THE EQUIPMENT, WRITTEN OR
ORAL, EXPRESS OR IMPLIED, AND WHETHER OR NOT ATTRIBUTABLE
TO SERVICE PERFORMED PURSUANT TO THE WARRANTY. ALL OTHER
REPRESENTATIONS OR WARRANTIES OF INTEGRA OR ITS
REPRESENTATIVES, EXPRESS OR IMPLIED, WITH RESPECT TO THE
EQUIPMENT OR THE SERVICES, DIAGNOSIS, ADVICE, ASSISTANCE OR
PARTS TO BE TENDERED PURSUANT TO THE WARRANTY, INCLUDING,
WITHOUT LIMITATION, THOSE OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE, ARE HEREBY EXPRESSLY DISCLAIMED.
IN NO EVENT SHALL INTEGRA, ITS ASSIGNEES OR SERVICE AGENTS
BE LIABLE FOR LOSS OF USE, REVENUE OR PROFIT OR ANY OTHER
INDIRECT, INCIDENTAL, EXEMPLARY, CONSEQUENTIAL, SPECIAL OR
OTHER DAMAGES, WHETHER ARISING IN CONTRACT OR IN TORT, BY
VIRTUE OF THE WARRANTY OR ANY PERFORMANCE OR BREACH BY
INTEGRA, ITS ASSIGNEES OR SERVICE AGENTS HEREUNDER OR
PURSUANT HERETO.
5.2. Customer agrees that, notwithstanding the technical assistance provided
pursuant to the Warranty by Integra or its representatives, Customer shall be
fully and solely responsible for all treatments performed or attempted with the
Equipment. INTEGRA MAKES NO REPRESENTATION OR WARRANTY AS
TO THE EFFICACY OF THE EQUIPMENT OR OF THE TECHNICAL
ASSISTANCE TO BE RENDERED BY INTEGRA, ITS ASSIGNEES OR
SERVICE AGENTS, FOR PURPOSES OF THE PARTICULAR TREATMENT
THAT CUSTOMER INTENDS TO PERFORM FOR THIRD PARTIES.
Moreover, Integra disclaims any liability with respect to the efficacy of the
Equipment or of said technical assistance or with respect to any claims by third
parties related to any treatment performed by Customer.
5.3. THIS INTEGRA LIMITED WARRANTY IS EXCLUSIVE AND IN LIEU OF
ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, AND OF ALL OTHER
OBLIGATIONS OR LIABILITIES ON INTEGRA'S PART OR THE PART OF ITS
DISTRIBUTORS, AND INTEGRA NEITHER ASSUMES NOR AUTHORIZES
ANY REPRESENTATIVE OR OTHER PERSON TO ASSUME FOR IT ANY
OTHER LIABILITY IN CONNECTION WITH INTEGRA'S PRODUCTS.
INTEGRA DISCLAIMS ALL OTHER WARRANTIES, EXPRESS OR IMPLIED
INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR OF
FITNESS FOR A PARTICULAR PURPOSE OR APPLICATION OR
WARRANTY OF QUALITY AS WELL AS ANY EXPRESS OR IMPLIED
WARRANTY TO PATIENTS. No warranty or guarantee may be created by any
act or statement nor may this Standard Warranty be modified in any way,
CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide C-3
except as a result of a writing signed by an officer of INTEGRA. These
limitations on the creation or modification of this warranty may not be waived or
modified orally or by any conduct.
5.4 IN NO EVENT SHALL INTEGRA AUTHORIZED DISTRIBUTORS BE
LIABLE TOWARDS CUSTOMER FOR ANY INCIDENTAL, INDIRECT,
CONSEQUENTIAL OR PUNITIVE DAMAGES IN CONNECTION WITH THE
ACQUISITION OR USE OF ANY INTEGRA PRODUCT. Further, this warranty
shall not apply to, and INTEGRA authorized distributors shall not be
responsible towards Customer for, any loss, arising in connection with the
purchase or use of any INTEGRA product that has been repaired by anyone
other than an authorized INTEGRA service representative or altered in any
way so as to affect its stability or reliability, or which has been subject to
misuse, negligence or accident, or which has been used otherwise than in
accordance with the instructions furnished by INTEGRA. THIS INTEGRA
DISTRIBUTOR LIMITED WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL
OTHER WARRANTIES TOWARDS CUSTOMER, EXPRESS OR IMPLIED,
AND OF ALL OTHER OBLIGATIONS OR LIABILITIES TOWARDS
CUSTOMER ON INTEGRA AUTHORIZED DISTRIBUTOR'S PART.
INTEGRA AUTHORISED DISTRIBUTORS DISCLAIM ALL OTHER
WARRANTIES TOWARDS CUSTOMER, EXPRESS OR IMPLIED
INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR OF
FITNESS FOR A PARTICULAR PURPOSE OR APPLICATION OR
WARRANTY OF QUALITY AS WELL AS ANY EXPRESS OR IMPLIED
WARRANTY TO PATIENTS OR OTHER THIRD PARTIES.
5.5 Force Majeure. Notwithstanding anything to the contrary herein
contained, if the performance of the Warranty by Integra, Integra authorized
distributors, or Customer or any obligation of Integra, Integra authorized
distributors, or Customer hereunder is prevented, restricted or interfered with
by reason of fire, explosion, act of God, labor disputes or accidents affecting
performance under the Warranty, or war, mobilization, civil commotions,
blockade or embargo, or any future law, regulation, ordinance or requirement
of any government or regulatory agency or any other act, whatsoever similar to
those above enumerated, or any other circumstance being beyond the
reasonable control of Integra, Integra authorized distributors, or Customer,
then and in that event Integra, Integra authorized distributors, or Customer, as
the case may be, shall promptly notify the other party hereto of the difficulties
resulting therefrom, and any of the foregoing events shall excuse any
performance required under the Warranty.
C-4 CUSA® Excel/CUSA® Excel+ Ultrasonic Surgical Aspirator System User’s Guide