HV Test Kit
HV Test Kit
HV Test Kit
�ET
1S0/IEC 17025 : 2017 - NABL Accredited Laboratory
At 5th Floor, 'Voltech Eco Tower',
No.2/429, Mount Poonamallee Road, Ayyappanthangal,
Chennai - 600056, Tamil Nadu.
Phone +91-44-43978021
CALIBRATION SERVICES Fax +91-44-42867746
CERTIFICATE NUMBER
CC2904
Email [email protected]
Web www.voltechgroup.com
GST No: 33AACCV0108M1ZK
CALIBRATION CERTIFICATE
SRF No: 75 Certificate No :MET/23/L/308
Mis.Voltech Engineers Pvt Ltd,
Customer Name And Address "Voltech Eco Tower"No.2/429,Mount Poonamalle
Road,Ayyapanthangal,Chennai-600 056
Customer Reference DC No.: 1303
ULR Number CC-290423000000308F
Details of the lnstrument
Description High Voltage Test Kit
Make Sudharsan
Model 30/40KV
Serial number 18/96
Identification Number VE-4749
Range Vide Respective Calibration Table
Resolut!cf! Vide Respectiv:.:: Calibration Table
Accuracy Vide Respective Calibration Table
Equipment Received on 15.03.2023
Condition of the equipment on receipt Good
Date of calibration 22.03.2023
Date of next calibration suggested 21.03.2024
Date of Certificate issued 23.03.2022
Calibration Environments
Tempernture 24.3°C
Relative humidity 55%RH
Master Instrument Details Master I
Description High Voltage Probe With Multimeter
Make Agilent
Type / Model 34136A
Serial number 20140105/29930160
Report Number TSC/22-23/20097-7
Calibration valid till 09.08.2023
Traceable to Transcal Bangalore
Method of Calibration
I. The instrument was calibrated as per W0rk Instruction no. MET!CC/WI-E12
--, ----�
2. The Standards used are traceable to National / I
Calibrated By
'�)
V Mohan K.Saravanakumar
Team Member Quality Manager
Page OJ of 02
tncT tlt$numcilrs cnuonnflofl cctTRC
ffi
n u|nof uoll€(fi cllclxc€ffi pw. ltD.
ISO/IEC 17025 : 2017 - NABL Accredited Laboratory
At 5th Floor, 'Voltech Eco Tower',
No.21429, Mount Poonamallee Road, Ayyappanthangal,
Chennai - 600056, Tamil Nadu.
CALIBRATION SERVICES :
Phone +91-44-43978021
Fax : +91-44-42867746 CERTIFICATE NUMBER
cc2904
Email : [email protected]
Web : www.voltechgroup.com
GST No : 33AACCV0108M1ZK
AC High Voltase@s0k
IIUC Mean Deviation Expanded Coverage
Range Standard Reading Deviation
Value Allowed (+) Uncertainty Factor
KV KV KV KV o/
KV /o K
I 1.0199 I -0.0199 0.300 0.8327 2.00
t0 10.208 l0 -0.2080 0.500 l. I 553 2.00
t5 r5.1 13 l5 -0.1 1 3 ' 0.s00 1.1562 2.00
20 20.299 20 -0.299 r.00 2.0835 2.00
25 25.289 25 -0.289 2.80 5.6306 2.00
Remarks
1. The above UUC was calibrated at Lab"
Calibrated By
o4
K.Saravanakumar