HS OperationGuide
HS OperationGuide
Jul. 2010
LabSolutions
TurboMatrixHS Control Software
Operation Guide
Read the instruction manual thoroughly before you use the product.
Keep this instruction manual for future reference.
This page is intentionally left blank.
Introduction Read this Instruction Manual
thoroughly before using the
product.
Thank you for purchasing Shimadzu analytical instrument
workstation “LabSolutions TurboMatrixHS Control Software”
(hereafter referred to as “the software”).
Indication Meaning
Emphasizes additional information that is provided to ensure
the proper use of this product.
^ Reference Indicates the location of related reference information.
[] Indicates the names of buttons, menu options, setting
options, windows/sub-windows, and icons that are displayed
in a window.
Example: Click [OK].
1. Period: Please contact your Shimadzu representative for information about the
period of this warranty.
3. Limitation of (1) In no event will Shimadzu be liable for any lost revenue, profit or
data, or for special, indirect, consequential, incidental or punitive
Liability: damages, however caused regardless of the theory of liability,
arising out of or related to the use of or inability to use the product,
even if Shimadzu has been advised of the possibility of such
damage.
(2) In no event will Shimadzu’s liability to you, whether in contract, tort
(including negligence), or otherwise, exceed the amount you paid
for the product.
4. Exceptions: Failures caused by the following are excluded from the warranty, even
if they occur during the warranty period.
1) Improper product handling
* Warranty periods for products with special specifications and systems are provided separately.
* The license cannot be reissued if you lose the USB dongle provided with the product.
1 Installation
1.1 Confirmation of parts ...........................................................................................1
1.3 Connection...........................................................................................................1
2 Operations
2.1 Method Development...........................................................................................7
2.1.1 Edit Method File ................................................................................................. 7
2.1.2 Parameters......................................................................................................... 9
TurboMatrixHS GC 1
Start N.O. #5 START IN
#6 1
1
2 Connect the ready signal cable.
Before connecting this cable, power switch to both the HS main unit and the GC must be turned off.
TurboMatrixHS GC 1
Ready In #9 READY OUT
Ground #12
1
1
TurboMatrixHS Control Software 1
1 Installation
5 Click the (Instruments) icon, and double-click the icon of the instrument to use.
If the (System Configuration) icon is not displayed, click on the title of the assistant
bar.
9 Click [Communication].
Upon establishment of connection, the information on the ROM version, largest vial number, serial
number, and presence/absence of optional units is sent to the PC and displayed.
If the (Data Acquisition) icon is not displayed, click on the title of the assistant bar.
2
2
2
2
2
TurboMatrixHS Control Software 7
2 Operations
3 Set the analysis conditions for the HS on the [TurboMatrixHS] tab in [Instrument
Parameter View].
^ Reference
If the [TurboMatrixHS] tab is not displayed in [Instrument Parameter View], refer to "1.4 Sofeware
System Configuration" P.2.
There are items that cannot be displayed depending on options. The following figure shows the state
in which all the items are displayed for explanation. So, it is different from the actual screen.
2.1.2 Parameters
Setting Range
Parameters Description Default
(Unit)
Mode Select an operating mode here. Constant Consta
Progressive nt
MHE
Injection Mode Select either of time or volume as a measure for determining Time Time
injection amount. Volume
2
Time: set Injection Time.
Volume: set Injection Volume and Transfer Flow.
Injection time calculated using the formula below is set to the
instrument.
(Injection Time) = (Injection Volume) / (Transfer Flow)
Injection Time Specify the injection amount by time. 0.00 ~ 9.99 min 0.08
Injection Volume Specify the injection amount by volume. 0.00 ~ 9.99 mL 2.40
Transfer Flow Specify the transfer flow rate. 0.01 ~ 50.00 30
mL/min
Temperature Zone Specify a method for setting the temperatures for heated O/N/T O/N/T
zones. Auto
O/N/T: Allows you to specify each temperature
independently from others.
Auto: Only the oven temperature is specified. The
temperatures for other units are automatically set to 5oC
above the oven temperature.
Oven Temp. Specify the oven temperature. 35 ~ 210 oC 60
Thermostatting Time Specify the heating time in the thermostat oven. 0.0 ~ 999.9 min 12
Pressurization Time Specify the time to pressurize sample vials. 0.0 ~ 300.0 min 3
Withdrawal Time Specify the time to withdraw sample vials. 0.0 ~ 99.9 min 0.2
GC Cycle Time Specify the cycle time from sample injection to the state 0.1 ~ 999.9 min 10
ready for analyzing next sample.
HS Carrier Gas Pressure Specify carrier gas pressure. 0.0 ~ 400.0 kPa 100
0.0 ~ 58.0 psi 14.5
High Pressure Turns ON/OFF the high-pressure sampling function. ON, OFF OFF
Inject Pressure Turns ON/OFF the cryofocus function. 0.0 ~ 414.0 kPa 100
0.0 ~ 60.0 psi 14.5
Vial Venting Specify the injection pressure for high-pressure sampling. ON, OFF ON
Sample Shaker Turns ON/OFF the vial venting function. ON, OFF OFF
Water Trap Turns ON/OFF the sample shaker function. ON, OFF OFF
Cryofocus Turns ON/OFF the water trap function. ON, OFF OFF
Precryofocus Time Specify the time for performing precryofocus. 0.0 ~ 300.0 min 0
Postcryofocus Time Specify the time for performing postcryofocus. 0.0 ~ 300.0 min 0
Cryofocus Temp. Specify the temperature for cryofocus. -180 ~ -10 oC -40
Back Flush Turns ON/OFF the back flush function. ON, OFF OFF
Back Flush On Time Specify the time to start back flushing. 0.0 ~ 300.0 min 0
Setting Range
Parameters Description Default
(Unit)
Back Flush Off Time Specify the time to stop back flushing. 0.0 ~ 300.0 min 0
Back Flush Pressure Specify the pressure for back flushing. 0.0 ~ 2.6 kPa 0
0.0 ~ 0.4 psi 0
^ Reference
• The time event function cannot be set.
• For GC Cycle Time (Thermostatting Time when in the MHE mode), set a time longer than the total time of
GC Analysis Time plus Cooling Time. Cooling time should be about 10 minutes when the initial
temperature is 50oC or higher and about 15 minutes when less than 50oC.
In cases where your system has no gradient program for column oven or incorporates a cooling program,
there is no need to worry about cooling time.
• Usually, set Temperature Zone as shown below:
(Oven Temp.) < (Needle Temp.) ≤ (Transfer Temp.) ≤ (GC Injection Port Temperature)
• The unit of pressure follows the <Configuration> - <GC> - <General> settings.
However, since TurboMatrixHS is only the unit of kPa or psi, in the case of the other unit, it is set to kPa.
1 Click the
Analysis] program.
(Realtime Batch) icon on the [Main] assistant bar in the [Realtime
If the (Realtime Batch) icon is not displayed, click on the title of the assistant bar.
2 Edit a batch table by referring to "4. Realtime Batch" in the Operators Guide.
You can create a batch table with simple operations by using the wizard.
Set the Vial # in ascending sequence.
Be sure that the vial number settings in the batch table match the vial numbers of the HS main unit.
Constant/Progressive
Set the vial numbers in ascending sequence.
Examples) Line Number Vial #
1 1
2 2
3 3
4 4
5 5
. .
. .
MHE
Since up to 9 repeat elutions may be conducted, set the same vial number to the number of times
equivalent to the number of repetitions.
Examples) Line Number Vial #
1 1
2 1
3 1
4 2
5 2
6
.
.
2
.
.
2
3 repeat
elutions
• When different HS parameters are set, an error is generated at the time of batch processing.
• When setting vial numbers, DO NOT skip numbers in the sequence. When any number is skipped, an
error is generated at the time of batch processing.
• Be sure to set a vial to the position in the HS main unit corresponding to the vial number specified in the
batch table. If no vial is present at the position, batch processing must be stopped.
For HSS analysis, partial execution and [Start Priority Batch] function are not available. Note that after batch
analysis is stopped, it can restart only from the first row.
1 Select [Add to Queue with Selected Batch File] on the [Batch] menu during realtime
batch.
2 Select the batch file to continuously perform realtime batch on in the [Open Batch File]
sub-window, and click [Open].