ARCADIS
SP

Software
System

Software Clone Installation Version VC10A

ARCADIS Varic
© Siemens, 2007

ARCADIS Orbic
ARCADIS Avantic
10048590
08080108
08080017

© Siemens, 2007
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.

Print No.: SPR2-310.816.03.04.02 English


Replaces: SPR2-310.816.03.03.02 Doc. Gen. Date: 09.09
CS PS SP
n.a.

2007
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer

Copyright
“© Siemens, 2007“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Mindit Magnetic Resonance Ltd. - China, Siemens Shang-
hai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA and/or
Siemens Healthcare Diagnostics Inc. - USA.

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.

Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.

ARCADIS SPR2-310.816.03.04.02 Page 2 of 48 © Siemens, 2007


09.09 CS PS SP
Table of Contents 3
0Table of Contents

1 _______ General information______________________________________________ 5

General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Software Clone installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Passwords. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Service, shutdown, hibernation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Handover to the customer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

2 _______ Software installation _____________________________________________ 6

Installation prerequisites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
System condition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Required Software CDs / DVDs / Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Installation time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Software Updates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Local Printer (if applicable) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
USB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
WLAN option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Functional Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Patient Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Password list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
HIPAA Administrator Logon (only if HIPAA is configured) . . . . . . . . . . . . . . . . . . . . . . . . 9
Entering and saving data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Service key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
BIOS password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Computer name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Administrator logon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
HIPAA logon (only if HIPAA is configured) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Backup procedure and SW-installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Systems with PC type M450 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Systems with PC type M460 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Downloading the main system software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
ARCADIS Avantic, downloading the generator software . . . . . . . . . . . . . . . . . . . . . . . . 16
Software installation VC10A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Installation prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Installation preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
PC type M450: image system installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
PC type M460: image system installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
PC type M450 and M460: Installation of Windows XP and syngo application . . . . . 25
Local service configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Restore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Restore main system settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Image text configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

© Siemens, 2007 SPR2-310.816.03.04.02 Page 3 of 48 ARCADIS


09.09 CS PS SP
4 Table of Contents

Country-specific keyboard layout and user interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . 33


Quick guide for system operation (only if available) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Test images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Patient Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Software Updates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Creating a Recovery DVD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Reconnecting the network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Closing the covers and protective conductor resistance test. . . . . . . . . . . . . . . . . . . 40

3 _______ Local printer installation (option) __________________________________ 41

General information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Sony UPD 970 and UPD 990 local printer installation . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Printer setup in local service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Log off as “meduser” and log on as “administrator”. . . . . . . . . . . . . . . . . . . . . . . . . . 43
Sony UPD 970/990 printer driver installation and settings . . . . . . . . . . . . . . . . . . . . 43
Printer setup in SynGo application software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

4 _______ Changes to previous version _____________________________________ 47

ARCADIS SPR2-310.816.03.04.02 Page 4 of 48 © Siemens, 2007


09.09 CS PS SP
General information 5
1-
General information
1General information

Software Clone installation 0

CAUTION The Clone software DVD is available as a spare part.


Only install the Clone software if the recovery DVD is defec-
tive or lost.
¹ Due to licensing, do not leave a Clone SW DVD at the
customer site.

Passwords 0

Required passwords are not included in this document. You will find them in the CS Knowl-
edge Base on the intranet or you can obtain them from the relevant USC or RSC.

Service, shutdown, hibernation 0

CAUTION Switching the system off via hibernation after service or


before handover to the customer is not sufficient.
Various error messages can appear after the next system
boot and configuration changes are not saved.
¹ To ensure that the configuration changes are saved, a
proper shutdown must be performed. Shut down the
system via the upper monitor menu bar
<Options>-<End Software Session>-<Shut Down> and
then press the "off" key on the monitor trolley.

Handover to the customer 0

NOTE Before the system is handed over to the customer, the sys-
tem must be shut down via the menu bar <Options>-<End
Software Session>-<Shut Down>.
Switching the system off via hibernation (on/off button on the
monitor trolley) is not sufficient.

Safety information 0

When performing the work steps and checks, the general safety information for medical
products must be observed.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 5 of 48 ARCADIS


09.09 CS PS SP
6 Software installation
2-
Installation prerequisites
2Software installation

System condition 0

• The recovery DVD is defective or lost.


• ARCADIS GEN2 systems are delivered from the factory with software version VC10A.
¹ Systems with software version VB13C cannot be upgraded.
• After the complete software installation, create a new recovery DVD.
See the software installation instructions in “Software Installation Recovery VC10A”,
print no. SPR2-310.816.04...

Required Software CDs / DVDs / Documents 0

• Image System Clone Software VC10A DVD must be ordered as a spare part.
¹ Do not leave the clone software DVD at the customer site.
• The backup/restore CD of the system. See system binder.
¹ If the backup/restore CD of the system is also defective or lost, then the imaging
PC must be configured manually.
See document SP00-000.843.30...
• 2 blank DVD-R media to create a new recovery DVD after the complete software instal-
lation.
• COBR tool as described in the document “Software Installation Recovery VC10A”, print
no. SPR2-310.816.04...
• Quick Reference Guide CD for Varic / Orbic / Avantic (if available)
• All software updates released for the system in the meantime. Check the system binder
for software updates and check the Healthcare Customer Service Intranet for more
details.
• System binder
• Software installation instructions “Software Installation Recovery VC10A”, print no.
SPR2-310.816.04...

Installation time 0

• The complete software installation procedure takes about 2 hours. The time needed to
save patient images is not included in the calculation.

Software Version 0

• The system must already be at software level VC10A or, if the PC is defective, an
up-to-date backup of the VC10A settings must be available.
¹ If the backup/restore CD of the system is also defective or lost, then the imaging
PC must be configured manually.
See document SP00-000.843.30...

ARCADIS SPR2-310.816.03.04.02 Page 6 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 7

Software Updates 0

• After the VC10A image system clone software has been installed, check whether addi-
tional software updates were released in the meantime and need to be reinstalled.
• Use the update instructions included with the updates to install the software updates.

Monitors 0

• Make sure that both monitors are connected and switched on.

Local Printer (if applicable) 0

• USB Printer (Sony): Make sure that the local printer is switched off and the USB con-
nection is not connected to the imaging system PC.

USB 0

• Make sure that no USB memory stick or USB hard drive is connected to the imaging
system PC.

NOTE Be aware that the dongle is not a memory stick.

Network 0

• The system must be physically disconnected from the network during software installa-
tion.

WLAN option 0

• If the WLAN option is installed on the system, ensure that the WLAN client is also dis-
connected from the imaging PC.
• Open the rear cover of the monitor trolley.
• Disconnect the RJ45 network cable (coming from the network switch, or hub) from the
RJ45 network plug of the imaging PC network adapter.
¹ No network cable is connected to the imaging PC any more.

Functional Test 0

• Check the functional status of the system.


¹ No error message should appear on the system.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 7 of 48 ARCADIS


09.09 CS PS SP
8 Software installation

Patient Images 0

Patient images that are not saved will be lost during a software installation.
• If required, save the patient images.
- Send the patient images to the archive
- or burn them to CD/DVD
- or save them to USB hard drive

Password list 0

• The knowledge base password list must be available.

ARCADIS SPR2-310.816.03.04.02 Page 8 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 9

HIPAA Administrator Logon (only if HIPAA is configured) 2.1

NOTE If a login window appears after the system startup, then


HIPAA is configured.

1. If you are uncertain of the administrator password, switch the system on and login into
the service application.
2. Click <Configuration>-<Next>-<Users>.
3. Enter the password (e. g. default password, see SP password list) in the “Password”
and “Confirm” fields below “Account for Administrator”
¹ With HIPPA configured, the administrator password is necessary for the first
HIPAA login after restoring “SW_Settings 02”.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 9 of 48 ARCADIS


09.09 CS PS SP
10 Software installation

Entering and saving data 2.2

Service key 0

• Record the syngo service key ......................................................................................


¹ It can be found in the system binder or can be obtained from the responsible
USC.

BIOS password 0

• Record the BIOS password...........................................................................................


¹ This is listed in the knowledge base password list.

Computer name 0

• Record the computer name..........................................................................................


¹ It can be found in the service software under Configuration/Local host/TCP IP
LAN/Computer name.

Administrator logon 0

• Record the administrator name ...................................................................................


¹ This is listed in the knowledge base password list.
• Record the administrator password .............................................................................
¹ This is listed in the knowledge base password list.

HIPAA logon (only if HIPAA is configured) 0

• Record the HIPAA user name.......................................................................................


¹ A user name can be obtained from the HIPAA administrator.
• Record the HIPAA user password.................................................................................
¹ A user password can be obtained from the HIPAA administrator.

ARCADIS SPR2-310.816.03.04.02 Page 10 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 11

Backup procedure and SW-installation procedure 0

Fig. 1: Overview Backup procedure and SW installation

© Siemens, 2007 SPR2-310.816.03.04.02 Page 11 of 48 ARCADIS


09.09 CS PS SP
12 Software installation

Backup 2.3

NOTE All network settings, including the static route, are a part of
“SW-Settings02”.

NOTE The monitor trolley and the basic unit must be connected
(plug X10 is connected to socket X10).

NOTE Main system backup for systems with SW version VC10A and
SP1 service package installed:
The programming of the country-specific configuration for
Australia and Switzerland is not stored with the backup of the
main system settings and cannot be restored by restoring
the main system settings.
If necessary, program and check the country-specific config-
uration for Australia or Switzerland as described in the Con-
figuration Instructions SP00-000.843.30...

Prerequisites 0

• The system must be switched on.


• The main system must be connected to the monitor trolley (monitor cable plug X10).

Systems with PC type M450 0

1. Open the DVD drive tray and insert a blank CD-R into the DVD drive.
2. Close the DVD drive tray.
3. Open the local service.
4. Click <Backup & Restore>
5. Select “Command” and “Backup”.
6. Select “(R) CD-R” in the “Drives” list field as the backup drive.
7. Perform a backup of the following packages:
¹ SW-Settings02
¹ ASPIASettings
¹ ExamSet
¹ Security-Settings (Option HIPAA)
¹ MainSystem
¹ 3D-Settings (Option Orbic 3D)
8. Remove the backup CD-R from the DVD drive tray.

ARCADIS SPR2-310.816.03.04.02 Page 12 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 13

9. Label the backup CD-R with the following data:


¹ Product name (ARCADIS Varic, Orbic, or Avantic)
¹ Product serial number
¹ “Backup CD”
¹ “Software version (Enter the current software version and, if installed, the current
bugfix version or service package version)”
¹ Date
10. Click <Home>.

Systems with PC type M460 0

1. Open the DVD drive tray and insert a blank CD-R into the DVD drive.
2. Close the DVD drive tray.
3. Open the local service.
4. Click <Backup & Restore>
5. Select “Command” and “Backup”.
6. Select “(R) CD-R” in the “Drives” list field as the backup drive.
¹ After a few seconds, a message “Please insert Media in Drive R:\ and press ’OK’
to proceed the Backup & Restore” is displayed.

NOTE This message pops up repeatedly approx. every 10 seconds,


whether a CD-R medium was inserted or not. If a CD-R
medium was inserted in the DVD drive, ignore the repeated
message.

7. Click ’OK’ to close the message.


8. Perform a backup of the following packages:
¹ SW-Settings02
¹ ASPIASettings
¹ ExamSet
¹ Security-Settings (Option HIPAA)
¹ MainSystem
¹ 3D-Settings (Option Orbic 3D)

NOTE After each backup of a selected package, the tray of the


CD/DVD drive opens automatically. Close the tray and start
the backup of the next package until all required packages
are stored on the CD-R medium.

9. Remove the backup CD-R from the DVD drive.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 13 of 48 ARCADIS


09.09 CS PS SP
14 Software installation

10. Label the backup CD-R with the following data:


¹ Product name (ARCADIS Varic, Orbic, or Avantic)
¹ Product serial number
¹ “Backup CD”
¹ “Software version (Enter the current software version and, if installed, the current
bugfix version or service package version)”
¹ Date
11. Click <Home>.

ARCADIS SPR2-310.816.03.04.02 Page 14 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 15

Downloading the main system software 2.4

NOTE A download of the main system software is required only if a


newer version is released or if the main system D1 board has
been replaced.

1. In the Service menu, click “Main System” and <Next>.


2. Under the <Download> menu, click <C-arm>.
3. Insert the main system software CD into the DVD drive.
¹ Wait until the CD is recognized by the DVD drive.
4. Under drive, select “R” (DVD drive).
5. Under “Directories & files” select “:
- for ARCADIS Varic/Orbic --> “SMC_Plus_V...”
- for ARCADIS Avantic --> “ARC C08 V...”
6. Click <Compare>.
¹ The file version is displayed under <New version>, and the recommended action
(e. g. update necessary) is displayed under <Status/Action>.
7. Click <Download>.
¹ A progress bar indicates the progress of the download.
¹ After the download, the host reboots automatically.
8. When the download is completed, click <Finish> and remove the software CD from the
DVD drive.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 15 of 48 ARCADIS


09.09 CS PS SP
16 Software installation

ARCADIS Avantic, downloading the generator software 2.5

NOTE The generator software only needs to be downloaded if a


newer version has been released or if the generator D21
board has been replaced.

1. In the Service menu, click “Main System” and <Next>.


2. Under the <Download> menu, click <C-arm>.
3. Insert the generator software CD into the DVD drive.
¹ Wait until the CD is recognized by the DVD drive.
4. Under drive, select “R” (DVD drive).
5. Under “Directories & files” select the binary file “gen_Vxxxx.bin” stored on the CD.
6. Click <Compare>.
¹ In line “GEN”, the file version is displayed under <New version>, and the recom-
mended action (e. g. update necessary) is displayed under <Status/Action>.
7. Click <Download>.
¹ A progress bar indicates the progress of the download.
¹ After the download, the generator and host reboots automatically.
8. When the download is completed, click <Finish> and remove the software CD from the
DVD drive.

NOTE After the successful download of the generator software and


the first reboot of the main unit, the main unit must be
switched off and then on again. Otherwise, error messages
can occur.

9. Disconnect the X10 connector from the main unit.


¹ The main unit is off.
10. Connect the X10 connector again and wait until the main unit is ready for use.

ARCADIS SPR2-310.816.03.04.02 Page 16 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 17

Software installation VC10A 2.6

NOTE ARCADIS systems with software version VB13C cannot be


upgraded to software version VC10A!

Installation prerequisites 0

• Ensure that the USB local printer (SONY) is disconnected from the imaging system PC.
• Ensure that the any network connection is disconnected from the monitor trolley.
• WLAN option: Ensure that the WLAN client is disconnected from the imaging PC net-
work adapter. If the WLAN option is installed, open the rear cover of the monitor trolley
and disconnect the RJ45 cable connector from the imaging PC network adapter (net-
work cable between imaging PC network adapter and internally installed network
switch (hub).
• Make sure that no USB memory stick or USB hard drive is connected to the imaging
system PC.
• Ensure that both monitors are connected and switched on.
• Ensure that a backup of the current system configuration is on hand.
• Ensure that all patients are archived.

Installation preparations 0

1. Disconnect the main system from the monitor trolley.


2. Switch the system on if it is switched off.
3. Shut down the system via the menu bar <Options>-<End Software Session>-<Shut-
down> .
4. Immediately after this, switch off the system by pressing the power off button on the
monitor trolley.
¹ Wait until the shutdown is finished.

PC type M450: image system installation 0

NOTE Systems with PC type M460: Skip this section.


Continue with section “PC M460: image system installation”.

PC type M450: changing the boot sequence and USB BIOS settings of the BIOS

NOTE To enable bootup from the image system DVD, some BIOS
settings have to be changed.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 17 of 48 ARCADIS


09.09 CS PS SP
18 Software installation

1. Switch on the system using the monitor trolley power on button.


2. As soon as text is displayed on the left monitor, press the “F2” key on the keyboard.
¹ The BIOS login window is displayed.

Fig. 2: Bios_PW
3. Enter the BIOS password and confirm it by pressing the “Enter” key.
4. ARCADIS Varic or ARCADIS Orbic: Insert the “C06/C07 Image system VC10A” soft-
ware DVD into the DVD drive.
5. ARCADIS Avantic: Insert the “C08 Image system VC10A” software DVD into the DVD
drive.
6. Press the “F9” key on the keyboard.
¹ “Load default configuration now?” is displayed.
7. Confirm the “Yes” button by pressing the “Enter” key .
8. Select <Advanced> with the cursor keys and press the “Enter” key.
9. Select <USB BIOS Supported Devices> with the cursor keys and press the “Enter” key.
10. Using the cursor keys, change the setting from “Auto” to “Keyboard/Mouse” and press
the “Enter” key.
¹ “Keyboard/Mouse” is displayed on the screen.
11. Press the “ESC” key to leave the menu.
12. Press the “ESC” key to select “Exit”.
13. Confirm the question “Save Changes & Exit” by pressing the “Enter” key.
¹ The message appears: “Save configuration changes and exit now?”

Fig. 3: Setup confirmation


14. Confirm the question “Save configuration changes and exit now?” by selecting the [Yes]
and pressing the “Enter” key.
¹ The system will be restarted and boots from the inserted “Image System” DVD.
15. Continue with the section <PC type M450: BIOS installation>.

ARCADIS SPR2-310.816.03.04.02 Page 18 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 19

PC type M450: BIOS installation

NOTE The BIOS installation is part of the Image System VC10A


installation and starts automatically after the previous reboot
of the system.
The Image System VC10A DVD must lie in the DVD drive tray.

1. The system boots from the previously inserted “Image System” DVD.
¹ The system installs the BIOS settings. This takes approx. 5 to 10 minutes.
¹ Some scripts are executed.
¹ During the BIOS installation, the system reboots again.
2. Wait until the message “Varic, Orbic, Avantic (V, O, A)” is displayed.
¹ The BIOS installation is finished.
¹ The “Image System VC10A” DVD remains in the DVD drive.
3. Continue with section <PC type M450: image system installation>.

PC type M450: image system installation

NOTE The image system installation is part of the Image System


VC10A installation and starts automatically after the previ-
ously performed BIOS installation.
The Image System VC10A DVD remains in the DVD drive.

1. The message “Varic, Orbic, Avantic (V, O, A)” is still displayed.


2. Select the software installation routine for your system by pressing the correct letter on
the keyboard:
“V” for ARCADIS Varic software installation.
“O” for ARCADIS Orbic software installation.
“A” for ARCADIS Avantic software installation.
¹ As soon as the letter key on the keyboard is pressed, the image installation starts.
¹ Some scripts are running.
¹ After a few seconds, a license agreement window is displayed.

NOTE If the message “A:\GHOSTERR.TXT” appears, the letter for


the system was incorrect and the installation must started
from the beginning.

3. Confirm the <OK> button by pressing the “Enter” key


¹ The image installation starts.
¹ The “Ghost” installation takes about 30 minutes.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 19 of 48 ARCADIS


09.09 CS PS SP
20 Software installation

NOTE The Ghost progress bar and time indication are incorrect.
After a progress bar display of about 40% the installation is
completed.

Fig. 4: Ghost_License
Fig. 5: ghost_progress

4. After successful installation of the image files, the system reboots automatically.
¹ After the reboot, the Windows XP Setup window is displayed on the left-hand
monitor.
5. Skip the next sections under <PC type M460: image system installation> and continue
with the section <Installation of Windows XP and syngo application>.

PC type M460: image system installation 0

NOTE Systems with PC type M450: Skip this section.

PC type M460: Changing the RAID settings for the BIOS installation

NOTE To enable bootup from the image system DVD, some BIOS
settings have to be changed.

NOTE To enable the image system installation, the section


“Changing the RAID settings for the BIOS installation”
must be performed prior to the section
“Changing the boot sequence for the BIOS installation”.
Otherwise the installation routine cannot create a partition
on one of the hard disk drives that is temporarily necessary.
An error message appears and the installation is unsuccess-
ful.

ARCADIS SPR2-310.816.03.04.02 Page 20 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 21

1. Switch on the system using the monitor trolley power on button.


2. As soon as text is displayed on the left monitor, press the “F2” key on the keyboard.
¹ The BIOS login window is displayed.

Fig. 6: Bios_PW
3. Enter the BIOS password and confirm it by pressing the “Enter” key.
4. Select the menu <Advanced>.
5. Select the menu line <Advanced System Configuration> and press the “Enter” key.
6. Select the line <SATA RAID Enabled> and press the “Enter” key.
7. Select <Disabled> and press the “Enter” key.
¹ <SATA RAID Enabled> is now disabled.
8. Select the line <SATA AHCI Enabled> and press the “Enter” key.
9. Select <Disabled> and press the “Enter” key.
¹ <SATA AHCI Enabled> is now disabled.
10. Press the “ESC” key till the BIOS main menu is displayed.
11. To store the BIOS settings, press the “F10” key.
¹ The message window <Setup configuration> is displayed.

Fig. 7: Setup confirmation


12. Confirm the question “Save configuration changes and exit now?” by selecting [Yes]
and pressing the “Enter” key.
¹ The BIOS window disappears and the system reboots.
13. As soon as text is displayed on the left monitor, press the “F2” key again.
¹ The BIOS login window is displayed again.

Fig. 8: Bios_PW
14. Continue with the next section <PC type M460: changing the boot sequence for the
BIOS installation>.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 21 of 48 ARCADIS


09.09 CS PS SP
22 Software installation

PC type M460: Changing the boot sequence for the image system installation

NOTE The work steps described in section “PC type M460: chang-
ing the RAID settings for the BIOS installation” must be per-
formed prior to the work steps described in this section.

1. ARCADIS Varic or ARCADIS Orbic: Open the DVD drive tray and insert the “Image
System Clone Install M460 C06/C07” CD into the DVD drive and close the drive tray.
2. ARCADIS Avantic: Open the DVD drive tray and insert the “Image System Clone Install
M460 C08” DVD into the DVD drive and close the drive tray.
3. Enter the BIOS password and confirm it by pressing the “Enter” key.
4. In the upper menu line, select <Boot>, using the cursor keys, and press the “Enter” key.
5. Using the up and down cursor keys, select the line <SATA CD: Optiarc DVD-RW-(S3)>.
¹ The characters in the line <SATA CD: Optiarc DVD-RW-(S3)> are highlighted.

NOTE Depending on which BIOS version is installed, BIOS settings


or the DVD drive, the name <SATA CD: Optiarc
DVD-RW-(S3)> displayed in the BIOS may differ.
Example for another displayed name:
<PCI BEV: P3 Optiarc DVD RW AD 7200>

NOTE As long as the WINDOWS operating system is not running,


the US keyboard layout is active. Use the layout diagram
illustrated here: (Fig. 9 / p. 22) to activate the “+” key func-
tion.

Fig. 9: Keyboard US layout


6. Using the cursor keys, select the line <SATA CD: Optiarc DVD-RW-(S3)>.

ARCADIS SPR2-310.816.03.04.02 Page 22 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 23

7. Move the line <SATA CD: Optiarc DVD-RW-(S3)> to the top position by entering the “+”
sign (highlighted line moves upward) or “-” sign (highlighted line moves downward).
¹ The <SATA CD: Optiarc DVD-RW-(S3)> is now in the top position under <Boot
Sequence>.
Selecting the “+” sign at the ARCADIS keyboard:
-> Press the “shift” key (2/Fig. 9 / p. 22) and the “=” key(1/Fig. 9 / p. 22) simultaneously.
Selecting the “-” sign at the ARCADIS keyboard:
- Press the “-” key (first key left of the “=” key (1/Fig. 9 / p. 22)).
¹ The line <SATA CD: Optiarc DVD-RW-(S3)> is now in the top position.

NOTE Do not change the boot sequence of the other displayed hard
disk drives!

8. Press the “F10” key to save the BIOS change


¹ The “Setup Confirmation” window is displayed.

Fig. 10: Setup confirmation


9. Confirm the text line <Save configuration changes and exit now?> by selecting [Yes]
and pressing the “Enter” key.
¹ The system will restart, and boot from the inserted “Image System Clone Install”
DVD.

PC type M460: BIOS installation (first part)

NOTE The BIOS installation is part of the Image System VC10A


installation and starts automatically after the previous reboot
of the system.
The Image System VC10A DVD is still in the DVD drive tray.

1. After the system restart, the system boots from the inserted “Image System Clone
Install M460 ...” DVD.
¹ The system installs the first part of the BIOS settings. This takes approximately 3
minutes.
¹ Some scripts are executed.
¹ During the BIOS installation, the system reboots several times.
2. Wait until the message “Varic, Orbic, Avantic (V, O, A)” or “Varic, Orbic (V, O)” is dis-
played.
¹ The BIOS installation is finished.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 23 of 48 ARCADIS


09.09 CS PS SP
24 Software installation

3. Continue with the next section <PC type M460: image system installation>.

PC type M460: image system installation

NOTE The image system installation is part of the Image System


VC10A installation and starts automatically after the previ-
ously performed BIOS installation.
The Image System VC10A DVD is still in the DVD drive tray.

1. The message “Varic, Orbic, Avantic (V, O, A)” or “Varic, Orbic (V, O)” is still displayed.
2. Select the system by pressing the correct letter on the keyboard:
“V” for ARCADIS Varic software installation,
“O” for ARCADIS Orbic software installation,
“A” for ARCADIS Avantic software installation,
¹ As soon as the letter key on the keyboard is pressed, the image installation starts.
¹ After a few seconds, a license agreement window is displayed.

NOTE If the message “A:\GHOSTERR.TXT” appears, the letter for


the system was incorrect and the installation must started
from the beginning.

3. Confirm the <OK> button by pressing the “Enter” key


¹ The “Ghost” image installation takes about 30 minutes.

NOTE The Ghost progress bar and time indication are incorrect.
After a progress bar display of about 40% the installation is
completed.

Fig. 11: Ghost_License


Fig. 12: ghost_progress

¹ The image installation starts.


4. After successful installation of the image files, the system reboots automatically.
¹ After the reboot, the second part of the BIOS installation starts.

ARCADIS SPR2-310.816.03.04.02 Page 24 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 25

5. After successful installation of the image files, the system reboots automatically.
¹ After the reboot, the Windows XP Setup window is displayed on the left-hand
monitor.
6. Continue with the next section <Installation of Windows XP and syngo application>.

PC type M460: BIOS installation (second part)


• After the installation of the Ghost image, the second part of the BIOS installation starts
automatically.
¹ The second part of the BIOS installation takes approximately 7 minutes.
¹ Some scripts are executed.
¹ During the BIOS installation, the system reboots several times.
• Wait until the following three text lines are displayed on the left monitor:
¹ Installation Done
¹ Please remove CD!
¹ D:\>
• Open the DVD drive tray and remove the “Image System Clone Install M460...” DVD.
• Close the DVD tray.
• Press the keys “CTRL” + “ALT” + “DEL” to restart the system, or switch the system off
and then on again (ON/OFF keys on the monitor trolley).
¹ The system reboots.
¹ The WINDOWS XP and syngo installation starts automatically.

PC type M450 and M460: Installation of Windows XP and syngo application 0

The system was restarted manually. The WINDOWS XP and syngo installation starts auto-
matically. The installation takes several minutes.

1. Wait until the window “Computer Name and Administrator Password” is displayed.

Fig. 13: compute_name

© Siemens, 2007 SPR2-310.816.03.04.02 Page 25 of 48 ARCADIS


09.09 CS PS SP
26 Software installation

2. When a window appears with “Computer name” and “Administrator” enter the previ-
ously noted computer name in the “Computer name” field.

NOTE DO NOT enter any password into the “Administrator pass-


word” fields!

3. In the window click the <Next> button.


¹ Windows XP Professional installation takes about 10 minutes.
¹ After Windows XP Professional is successfully installed, the system reboots auto-
matically.
¹ Some installation scripts will be started automatically.

NOTE If the window “Welcome to found new Hardware Wizard”


appears, select “No, not this time”--> <Next>, --> “Install the
software automatically (Recommended)” and <Next>-<Fin-
ish>. The message “The system cannot find the file speci-
fied” disappears after a while, and the installation continues.

¹ After a second reboot, the login dialog window is displayed.

Fig. 14: meduser

Fig. 15: ARCADIS Keyboard, US layout

ARCADIS SPR2-310.816.03.04.02 Page 26 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 27

NOTE Regardless of the keyboard, the keyboard layout is English.


The @ sign can be selected by pressing the <Shift> + <2>
keys simultaneously.

4. In the login dialog, change the “User name” to “meduser” and enter the “Password”
¹ User name: meduser
¹ Password: @med@User... (“...” = the last 3 digits of the previously noted com-
puter name)
5. Click <OK>.
¹ Autoconfiguration starts.
¹ Wait for the Explorer window “Please enter license” (see next section).

Local service configuration 0

Starting the Local Service Configuration

NOTE If no backup is on hand, configure the system according to


the configuration guide.

1. Remove the installation DVD from the drive.


2. Select “Click here to start the Local Service Configuration”.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 27 of 48 ARCADIS


09.09 CS PS SP
28 Software installation

3. Enter the service key in the service window that appears (under password).
4. Select <Set as default>.
5. Confirm by clicking <OK>.
¹ The <Home> menu window of the local service is displayed.

Restore 0

1. Open the DVD drive tray and insert the backup CD into the DVD drive.
2. In the Service menu, click <Backup & Restore>.
3. Under “Command” select “Restore” and select “(-R-) CD-R” under “Drives”.

SW-Settings 02

1. Under “Archive”, select the correct (newest) backup file of “SW-Settings 02”
2. Highlight all the files in “Groups” (press the “Shift” key and select the first and last entry
of groups).
3. Click <Go>.
4. Wait until “======= End Restore: Day...., Date....., Time...., Duration....” is displayed in
the status window. The next settings group can be selected.
5. Do not restart the ASPIA PC. First restore the following settings.

ASPIA Settings

1. Under “Archive”, select the correct backup file of the “ASPIA Settings”
2. Highlight all the files in "Groups”.
3. Click <Go>.
4. Wait until “======= End Restore: Day...., Date....., Time...., Duration....” is displayed in
the status window. The next settings group can be selected.
5. Do not restart the ASPIA PC. First restore the following settings.

3D Settings (only for ARCADIS Orbic)

1. Option for Orbic: Under “Archive” select the correct backup file of 3D Reconstruction
settings.
2. Highlight all the files in "Groups”.
3. Click <Go>.
4. Wait until “======= End Restore: Day...., Date....., Time...., Duration....” is displayed in
the status window. The next settings group can be selected.
5. Do not restart the ASPIA PC. First restore the following settings.

ARCADIS SPR2-310.816.03.04.02 Page 28 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 29

Exam Set

NOTE ARCADIS GEN2 systems with software version VC10A and


installed imaging PC type M460:
At the factory-installed new default exam sets on the imaging
PC were installed.
ARCADIS Gen2 systems with software version VC10A and
installed imaging PC M450:
The CD with the default exam sets “AutoLUT - Autodose -
Examset V2” is part of the software package delivered with
the Update Instructions SP018/09/R (Service pack SP1).
Upon customer request, the new “AutoLUT - Autodose -
Examset V2” can also be installed on these systems.
ARCADIS Gen2 systems (M450 and M460):
Based on the reasons given above, the default exam sets
must be restored too.

1. Under “Archive” select the correct backup file of “Exam Sets”.


2. Highlight “Filter_Files”, “User Defined_ExamSets” and “Default_Exam_Set
(only_for_headquarters).
3. Close the message box “Default Exam Sets should be restored only on request of the
Headquarters! Continue?” by clicking on the OK button.
4. Click <Go>
5. Only if an “Exam sets” window is displayed: Select all exam sets in the window ”Exam-
Sets in backup package” and click <Import> to load the exam sets into the window
“Installed ExamSets”.
6. Click <Go>
7. Wait until “======= End Restore: Day...., Date....., Time...., Duration....” is displayed in
the status window. The next settings group can be selected.
8. Do not restart the ASPIA PC. First restore the following settings.

Security-Settings (only if HIPAA is configured)

NOTE If HIPAA is configured, an HIPAA password is required after


the restore of the security settings, on every system restart.
For login, the “HIPAA” user name and password is needed.

1. Under “Archive”, select the correct backup file of “Security Settings”.


2. Click <Go>.
¹ Systems with the HIPAA option activated: After the next restart, the user names
and passwords stored in the security settings are valid again.
¹ Next time, log in with the user name/password obtained from the HIPAA adminis-
trator.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 29 of 48 ARCADIS


09.09 CS PS SP
30 Software installation

3. Wait until “======= End Restore: Day...., Date....., Time...., Duration....” is displayed in
the status window.

NOTE The main system is still not connected to the monitor trolley.
At this time, the restore of the main system parameters is not
possible.
The main system parameters are stored on the main system
D1 board. A restore is necessary only if D1 is defective and
was replaced.

4. Remove the backup CD from the DVD drive.


5. Restart the system via the windows menu bar, click <Start>-<Shut Down>, select
<Restart> an click <OK>.

Database creation and activating the restored values

1. After the restart of the system, the patient database creation starts and takes about 15
minutes.
¹ The syngo counter screen appears.
¹ When this is completed, a window appears with the following text:
"Warning: database creation in progress, this will last a long time. Wait until
the patient registration dialog is open! Close the window then! Press any
key to continue...”
¹ Wait until the patient registration dialog or the patient browser is displayed.
2. Close the patient registration dialog or the patient browser.
3. Click the window “Warning: Database...” and press the “Enter” key.
¹ The window “Warning: Database...” disappears.

Final steps of software installation

1. Connect the main system.


2. Continue with “Restore main system settings”.

Restore main system settings 0

NOTE Now the main system is connected to the monitor trolley.


The main system settings are stored on the main system D1
board.
A restore is only needed if the D1 board is was replaced.

1. Open the DVD drive tray and insert the backup CD into the DVD drive again.
2. Open the local service again.
3. In the Service menu, click <Backup & Restore>.
4. Under “Command” select “Restore” and select “(-R-) CD-R” under “Drives”.

ARCADIS SPR2-310.816.03.04.02 Page 30 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 31

5. Under “Archive” select the correct backup file “Main system”.


6. Click <Go>.
7. Wait until “======= End Restore: Day...., Date....., Time...., Duration....” is displayed in
the status window. The next settings group can be selected.
8. Wait until the main system restarts automatically.
9. Go back to the main menu of the local service.
10. Do not restart the system.
11. Open the DVD drive tray and remove the backup CD from the drive.
12. Continue with the next section “Image text configuration”.

NOTE Systems with SW version VC10A and installed SP1 service


package:
The programming of the country-specific configuration for
Australia and Switzerland is not stored with the backup of the
main system settings and cannot be restored by restoring
the main system settings.
If necessary, program and check the country-specific config-
uration for Australia or Switzerland as described in the Con-
figuration Instructions SP00-000.843.30...

© Siemens, 2007 SPR2-310.816.03.04.02 Page 31 of 48 ARCADIS


09.09 CS PS SP
32 Software installation

Image text configuration 2.7

1. In the syngo menu bar, select <Options>-<Configuration>.


2. Double-click “Image text editor”.
3. Under “View Name” (Basisformat), select “XA Image”.
- Deselect software version
¹ Software version = Q
4. Click <Apply>.
5. Under “View name”, select “ARCADISAcqView”.
6. Click “Customized text” (“Ausgewählte Texte”).
¹ “Image Date” = Q
¹ “Image Time” = Q
¹ “Image Number” = Q
7. Click <Apply>.
8. Click <OK> and close the Configuration panel.

ARCADIS SPR2-310.816.03.04.02 Page 32 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 33

Country-specific keyboard layout and user interface 2.8

1. In the "Configuration" panel double-click “Regional and Language...”.


2. Select “Regional Options”.
3. Under “Standards and formats”, select the appropriate language from the drop-down
menu:
- English (United States)
- German (Germany)
- French (France)
- Spanish (Spain)
4. Under “Location”, select your corresponding location.
5. Click <Apply>

NOTE “English (United States)” is the default setting. As a result


“Apply” is not available.

6. Select the “Languages” tab card and click <Details...>.


7. Under “Installed services” click “Add...”
¹ The window “Add input language” appears.
8. Under “Installed services”, the following “input language” can be selected:
- English (United States)
- German (Germany)
9. Under “Keyboard layout / IME”, select the appropriate keyboard language from the
drop-down menu:
- US
- United Kingdom
- German
- Swedish
- Portuguese
- Italian
- French
- Spanish

NOTE If a keyboard language (e. g. “US”) is displayed under


“Installed services”, it disappears from the “Keyboard lay-
out/IME” drop-down menu.

NOTE The correct keyboard language can also be found in the


delivery note from the factory.

10. Click <OK> to close the “Add input language” window.


11. Click <Apply> and <OK> to close the window “Text Services and Input Languages”.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 33 of 48 ARCADIS


09.09 CS PS SP
34 Software installation

12. Click <Apply> and <OK> to close the window “Regional and Language Options”.
¹ If the language have been changed, the following message appears:
The following is displayed in the case that values have been changed:
“The regional settings have been changed.
Therefore a restart of the application is necessary.
Do you want to restart the application now?”
¹ In case of language change click on <OK>. This will apply the settings and the
system restarts.
¹ Otherwise shut down the system via <Options>-<End Software Session>-<Shut-
down>.

ARCADIS SPR2-310.816.03.04.02 Page 34 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 35

Quick guide for system operation (only if available) 2.9

1. In the Service menu, click <Configuration>.


2. Under “Help / Documentation”, click “Help Update”
3. Place the “Quick Reference Guide” CD-R into the DVD drive.
4. Select the path and file(s) under “Directories and file”.
5. Click <Save> and remove the CD-R from the DVD drive.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 35 of 48 ARCADIS


09.09 CS PS SP
36 Software installation

Test images 2.10

• Test images are automatically installed during software installation.


¹ The “Service Patient” and corresponding images are shown only when the “Ser-
vice menu” is opened and the login has been performed.

ARCADIS SPR2-310.816.03.04.02 Page 36 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 37

Patient Images 2.11

• If necessary, restore the patient images.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 37 of 48 ARCADIS


09.09 CS PS SP
38 Software installation

Software Updates 2.12

• The image system Clone CD software package does not include any software update
released later for this system.
• Check the system binder for previously installed software updates for this system.
• Check the Healthcare Customer Service Home page for further released software
updates that apply to your system.
• If needed and before creating a new recovery DVD, install the corresponding software
updates, using the documentation delivered with the software updates.

ARCADIS SPR2-310.816.03.04.02 Page 38 of 48 © Siemens, 2007


09.09 CS PS SP
Software installation 39

Creating a Recovery DVD 2.13

• After a complete software installation, create a new recovery DVD.


• Use the document “Software Installation Recovery VC10A” to create the new DVD.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 39 of 48 ARCADIS


09.09 CS PS SP
40 Software installation

Final work steps 2.14

Reconnecting the network 0

1. Reconnect the network connectors to the monitor trolley.


2. If previously disconnected, reconnect the RJ45 network connector of the PC to the net-
work switch (hub).
3. If a local printer is present, DO NOT connect the printer to the imaging PC. First per-
form the work steps described in chapter “Local printer installation (option)”.

Closing the covers and protective conductor resistance test 0

1. If covers were removed, reconnect all disconnected protective conductor wires and
reinstall the covers.
2. If covers were removed, perform a protective conductor resistance test.
¹ The protective conductor resistance must not exceed 0.2 ohms.

ARCADIS SPR2-310.816.03.04.02 Page 40 of 48 © Siemens, 2007


09.09 CS PS SP
Local printer installation (option) 41
3-
General information
3Local printer installation (option)

NOTE Perform the steps in this chapter only if the system is


equipped with a local printer.

NOTE Local printer, PostScript printer


These printer types are not approved for diagnostic pur-
poses.
Notify the responsible personnel about this limitation.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 41 of 48 ARCADIS


09.09 CS PS SP
42 Local printer installation (option)

Sony UPD 970 and UPD 990 local printer installation 3.1

Prerequisites 0

NOTE Do not connect the USB printer until you are instructed to in
the course of the software installation process.

1. For the necessary test exposures, make sure the printer has paper or film.
2. Unplug the printer's USB connection to the PC.
3. Switch on the system and wait until it has booted.

Printer setup in local service 0

1. In the "Service" menu, click <Configuration>.


2. In the “List of system options” window the “Paper Printer” check box H must be
checked.
3. Click <Next>.

Paper printer

1. Under the “External devices” menu, click (or <Next> to go to) “Paper Printer”.
2. In “Available paper printer media formats”, mark “A4” and “8x10”. All other formats must
be deselected.
3. Click <Save>.

Printer settings

1. Under the “External Devices” menu, click (or <Next> to go to ) “Printer settings”.
¹ Make sure that the default value “black” is selected and click <Finish>.

PS LUT files

1. Under the “External devices”, menu click (or <Next> to go to) “PS LUT files”.
2. Ensure that the “Use LUT for Postscript Printers” check box is unchecked.
¹ The “PS LUT directory” field is not relevant.
3. Ensure that the “Use floating Zoom factors” check box is checked.
4. Click <Save> and <Finish>.

NOTE Do not shut down the system.

ARCADIS SPR2-310.816.03.04.02 Page 42 of 48 © Siemens, 2007


09.09 CS PS SP
Local printer installation (option) 43

Log off as “meduser” and log on as “administrator”. 0

1. In the "Service" menu, click <Home> -<Utilities>.


¹ If a window appears “A system reboot is necessary...” click <Cancel>.
2. Under “Source”, select “System”.
3. Select “Shut down Application” and click <Go>.
¹ The message box appears: “Application shutdown was selected. Are you sure?”
4. Click <OK>.
¹ The system shuts down to the Windows XP level.
¹ The message “Shutdown of Application finished...” appears.
5. In the Windows menu bar, click <Start>-<Shut Down...>.
6. Select “Log off meduser”.
7. Press and hold the shift key and click <OK>.
¹ A “Log on” window appears.
- Under “User name”, enter “administrator”.
- Under “Password”, enter the administrator password (see SP password list).

NOTE If the administrator password is not valid, proceed as fol-


lows: In the “Service” menu click <Configuration> - <Next>
then under “Local Host” click <Users>. In the “Account for
Administrator” fields, enter the password from the SP pass-
word list, confirm it, and click <Save>.

8. Click <OK>.
¹ The system is now at the Windows XP level with administrator rights.

Sony UPD 970/990 printer driver installation and settings 0

Printer driver installation


1. In the Windows task bar, click <Start> with the right-hand mouse button, then select
and click <Explore>.
¹ Windows Explorer opens.
2. Select and open the path “C:\ASPIA\driver\UPD9X0”.
¹ The contents of the subdirectory “C:\ASPIA\driver\UPD9X0” are displayed.
3. Double-click the file “printer_Sony_UP-970AD_990AD.reg” and confirm the subse-
quent message windows by clicking <Yes> and <OK>.
4. Connect the USB cable to the printer and switch the printer on.
¹ The hardware installation procedure starts automatically.
¹ Follow the procedure:

© Siemens, 2007 SPR2-310.816.03.04.02 Page 43 of 48 ARCADIS


09.09 CS PS SP
44 Local printer installation (option)

- Select “No, not this time”.


- Click <Next>.
- Select “Install from a specific location (Advanced)”.
- Click <Next>.
- Select “Include this location in the search”.
- Select “Browse”
- Select the path “My Computer\C:\ASPIA\driver\UPD9X0”.
- Click <OK>.
- Click <Next>.
¹ The driver installation starts automatically.
- Click <Finish>.
Printer driver settings
1. In the Windows task bar, click <Start>-<Settings>-<Printers and Faxes>.
2. Click the driver name “Sony UP-970AD/990AD” with the right mouse button and select
the <Properties> menu.
3. Select <Advanced>.
4. Click <Printing Defaults>-<Density Adjust> and set the following values:
- Gamma = “Tone1”
- Advanced = “0”
- Dark = 0
- Light = “- 15” (minus 15)
- Sharpness = “4”
5. Click <Apply> and <OK>.
6. In the window “Sony UP-970AD/990AD Properties”, click “General” and “Print Test
Page”.
¹ After a few seconds a message window is displayed.
¹ Wait until a Windows test page is printed out.
¹ Close the displayed message window by clicking <OK>.
¹ Close the “Sony UP-970AD/990AD Properties” window.
7. In the Windows task bar, click <Start>-<Shut Down...>-<Restart> and confirm with
<OK> to restart the system.
¹ After the restart, you are logged onto the system as “meduser”.

Printer setup in SynGo application software 0

Set the film size under “Filming”


1. On the right-hand monitor, select the FTC (filming” tab).
¹ The “Filming” tab is active on the right-hand monitor.
2. Select the “Camera” menu tab card on the “Filming” tab.

ARCADIS SPR2-310.816.03.04.02 Page 44 of 48 © Siemens, 2007


09.09 CS PS SP
Local printer installation (option) 45

3. In the “Camera” list field, select and click “Sony UP-970AD/990AD”.


¹ The “Sony UP-970AD/990AD” list field must be highlighted.
4. Open the patient browser and send an image to the FTC.
5. In the “Default Film size” list field, select and click “8x10”.
¹ The “8x10” list field must be highlighted.
Set the default printer for “Structured reporting”
1. On the “Examination” tab, click <Options>-<Configuration>.
¹ The “Configuration” panel is displayed.
2. Double click the “Structured reporting” icon ("Strukturierte Befundung”).
3. In the “Default Printer” list field, select “SonyUP-970AD/990AD” - even if
“SonyUP-970AD/990AD” is already displayed.
¹ The “SonyUP-970AD/990AD” list field must be highlighted.
4. In the “Default paper size” list field, select and click “8x10”.
¹ The “8x10” list field must be highlighted.
5. Click <Apply> and <OK>.
Configure “Filming Layout”
1. Double click the <Filming Layout> icon.

NOTE If a message box appears, confirm with OK and double-click


the <Filming Layout> icon again.

2. On the “Filming” tab, enter the following parameters:


- New film job by patient = H
- New film sheet by (study or series) = Q
- New row of images by (patient, study or series) = Q
- Number of Copies = “1”
- Exposure every X document = “1”
- Film size = “8x10” --> click in the “8x10” window.
¹ The “8x10” list field must be highlighted.

NOTE If the film size is not displayed, restart the system via
<Options>-<End Software Session>-<Shutdown> and repeat
the section “Printer setup in syngo application software”.

- Segment lines = “Yes”


- Page number on print out = H
3. Click “Apply”.
4. On the “Series” tab, enter the following parameters:

© Siemens, 2007 SPR2-310.816.03.04.02 Page 45 of 48 ARCADIS


09.09 CS PS SP
46 Local printer installation (option)

- Layout divisions = “1x1”


- Orientation = “Portrait” (“Hochformat”)
- Image order = “Horizontal”
- Aspect Ratio = “Keep visible part”
- Copy Series = Q
5. Click <Apply>.
6. Shut down the system via <Options>-<End Software Session>-<Shutdown>.

ARCADIS SPR2-310.816.03.04.02 Page 46 of 48 © Siemens, 2007


09.09 CS PS SP
Changes to previous version 47
4Changes to previous version 4-

Chapter
Software Installation: Software Clone installation for PC type M450 and M460
revised.

© Siemens, 2007 SPR2-310.816.03.04.02 Page 47 of 48 ARCADIS


09.09 CS PS SP
48 Changes to previous version

ARCADIS SPR2-310.816.03.04.02 Page 48 of 48 © Siemens, 2007