ISO 02400-2012 Non-Destructive Testing-Ultrasonic Testing-Specification For Calibration Block No 1
ISO 02400-2012 Non-Destructive Testing-Ultrasonic Testing-Specification For Calibration Block No 1
ISO 02400-2012 Non-Destructive Testing-Ultrasonic Testing-Specification For Calibration Block No 1
STANDARD 2400
Second edition
2012-09-15
Reference number
ISO 2400:2012(E)
© ISO 2012
ISO 2400:2012(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International
Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 2400 was prepared by the European Committee for Standardization (CEN) Technical Committee
CEN/TC 138, Non-destructive testing, in collaboration with Technical Committee ISO/TC 135, Non-destructive
testing, Subcommittee SC 3, Ultrasonic testing, in accordance with the Agreement on technical cooperation
between ISO and CEN (Vienna Agreement).
This second edition cancels and replaces the first edition (ISO 2400:1972), which has been technically revised.
1 Scope
This International Standard specifies requirements for the dimensions, material and manufacture of a steel
block for calibrating ultrasonic test equipment used in manual testing.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced document
(including any amendments) applies.
EN 10025-2, Hot rolled products of structural steels — Part 2: Technical delivery conditions for non-alloy
structural steels
4 Manufacture
4.1 Steel
Blocks shall be manufactured from steel grade S355J0, specified in EN 10025-2, or from steel of an
equivalent grade.
4.2 Dimensions
The dimensions of calibration block No. 1 shall be as shown in Figure 1.
NOTE This International Standard does not include any requirement for a plastic insert.
All external surfaces shall be machined to an Ra value not greater than 0,8 µm.
Prior to final machining, the block shall be proven to be free from internal discontinuities. For this purpose, an
ultrasonic examination shall be carried out after the heat treatment, with a longitudinal wave probe of at least
10 MHz nominal centre frequency and having a transducer size of 10 mm to 15 mm. The block shall be checked
on the basis of all four long faces to cover the complete volume. With the probe positioned on the largest face
of the block, the equipment gain shall be set to achieve a grain scatter noise of 10 % of the screen height. No
echo shall have an amplitude greater than that of the grain scatter noise.
5 Velocities
The velocities of longitudinal and transverse waves shall be determined as specified in Annex A. The velocities
shall be determined within a maximum permissible error of ±0,2 %, i.e. with an uncertainty of ±6 m/s for
transverse waves and ±12 m/s for longitudinal waves.
The determined longitudinal wave velocity, vl, shall be 5 920 m/s ± 30 m/s and the transverse wave velocity, v t,
shall be 3 255 m/s ± 15 m/s.
6 Marking
The block shall be permanently marked, in the area shown in Figure 1, with:
7 Declaration of conformity
A declaration shall be issued by the manufacturer for each block, containing:
a) a statement that the block complies with this International Standard, i.e. ISO 2400;
b) the mean value of the determined longitudinal wave velocities, vl; see Clause 5;
c) the mean value of the determined transverse wave velocities, v t; see Clause 5.
8.1 General
One of the modifications in 8.2 or 8.3 shall be made to the block.
9 Existing blocks
Existing blocks fulfil the requirements of this International Standard provided that they meet the requirements
on velocities (see Clause 5) and dimensions (see 4.2) with the exception that the small side-drilled hole may
have a diameter of 1,5 mm.
Dimensions in millimetres
Tolerances, unless otherwise stated ±0,10 mm Height of characters for angle indents 5 mm
Angle identification and index mark lengths ±0,4 mm Surface finish, all over Ra ≤ 0,8 µm
Dimensions in millimetres
Dimensions in millimetres
Annex A
(normative)
Determination of velocities
A.1 General
This method first measures the physical dimensions of the block to an uncertainty of 0,01 mm. The areas to
be used are checked to ensure that there are no variations in thickness greater than 0,01 mm over the area of
the probe face. An instrument in conjunction with a straight-beam probe is then used to measure the time of
flight (to an accuracy in the measurement of time difference of 0,2 %). Velocities are then calculated (path over
time)1). The time of flight is measured in different directions, i.e. at two well separated positions through the
25 mm thickness (one at the quadrant and one close to the large hole) and at one position through the 100 mm
thickness. The measurements shall be carried out within the temperature range 17 °C to 23 °C.
Because transverse waves are polarized, make two measurements in each location of the probe with the plane
of polarization in the second measurement perpendicular to the first measurement and parallel to one side of
the block. Thus, for each calibration block, there are at least six values for transverse wave velocity.
ICS 25.160.40
Price based on 7 pages