Product: XVI
Date: May 2014
FCO: 200 01 507 086
FCO: 200 01 507 086 VID: 1.0 Elekta Limited, Linac House, Fleming Way, Manor Page 2 of 7
Royal, Crawley, West Sussex. RH10 9RR.
Tel: +44 (0)1293 544422 Fax: +44 (0)1293 654321
Copyright © 2013 Elekta AB. All rights reserved www.elekta.com gPOL0007-1 VID:2.0
Figure 1: Table Move Assistant dialog box showing blank Relative Actual values
The Relative Actual column or the Absolute Actual of the Table Move
Assistant dialog box incorrectly appears blank.
When this occurs, XVI does not read the Absolute Actual position of
the treatment table. This causes XVI to calculate the treatment table
moves from the zero position of the treatment table
(+0.0, +0.0, +0.0).
If the Relative Set values are less than or equal to 2 cm, you can do
remote automatic table movement (RATM) from the function keypad
(FKP).
When you press the Table ASU (automatic setup) buttons, the
treatment table moves to a position that was incorrectly calculated
from XVI.
FCO: 200 01 507 086 VID: 1.0 Elekta Limited, Linac House, Fleming Way, Manor Page 3 of 7
Royal, Crawley, West Sussex. RH10 9RR.
Tel: +44 (0)1293 544422 Fax: +44 (0)1293 654321
Copyright © 2013 Elekta AB. All rights reserved www.elekta.com gPOL0007-1 VID:2.0
Scenario 2:
The Absolute Actual column of the Table Move Assistant dialog box
contains values which appear correct. However, this data is
inaccurate.
If the Relative Set values are less than or equal to 2 cm, you can do
remote automatic table movement (RATM) from the function keypad
(FKP).
When this occurs XVI does not receive Absolute Actual positions of
the treatment table. This causes XVI to incorrectly calculate the
target treatment table position.
When you press the Table ASU (automatic setup) buttons, the
treatment table moves to the incorrect position.
FCO: 200 01 507 086 VID: 1.0 Elekta Limited, Linac House, Fleming Way, Manor Page 4 of 7
Royal, Crawley, West Sussex. RH10 9RR.
Tel: +44 (0)1293 544422 Fax: +44 (0)1293 654321
Copyright © 2013 Elekta AB. All rights reserved www.elekta.com gPOL0007-1 VID:2.0
Clinical impact: Collision Risk
WARNING
Make sure that the gantry and treatment table movements are in view
at all times when you do ASU. If you ignore this warning, it can cause
fatal injury and clinical mistreatment.
If you do not monitor the patient and treatment table movements, and you
keep the Table ASU buttons pressed, the treatment table continues to
move. It only stops when a touchguard activates. Therefore, this error can
cause a collision between the patient and an external beam limiting device
(BLD) if:
The external BLD does not operate with the touchguard attached.
The external BLD extends out more than the touchguard.
If this error occurs with an external BLD attached to the digital accelerator,
you can cause fatal injury to the patient.
Clinical mistreatment
Scenario 1:
When the error occurs, it is not possible to get the Table Move Assistant
values in tolerance. Therefore, it is not possible to deliver treatment to a
patient in the incorrect position.
Scenario 2:
There is no potential for clinical mistreatment if the Relative Set is more than
0.1 cm. See Figure 3.
FCO: 200 01 507 086 VID: 1.0 Elekta Limited, Linac House, Fleming Way, Manor Page 5 of 7
Royal, Crawley, West Sussex. RH10 9RR.
Tel: +44 (0)1293 544422 Fax: +44 (0)1293 654321
Copyright © 2013 Elekta AB. All rights reserved www.elekta.com gPOL0007-1 VID:2.0
Solution: Monitor the Relative Actual and Absolute Actual columns of the Table Move
Assistant dialog box, if these values are blank do not press the Table ASU
buttons on the FKP.
Or
If you see a movement larger than the Relative Actual and the Absolute
Actual does not update, follow this procedure to correct the error.
A later software release of XVI will be supplied free of charge which gives a
solution to this problem.
Technical Reference: CLM 01690462, CLM01795287
Contact: If you have any queries about this Notice, contact your local Elekta office.
Safety reference
The following warnings and cautions are associated with this notice:
WARNING
Make sure that the gantry and treatment table movements are in view at all times
when you do ASU. If you ignore this warning, it can cause fatal injury and clinical
mistreatment.
FCO: 200 01 507 086 VID: 1.0 Elekta Limited, Linac House, Fleming Way, Manor Page 6 of 7
Royal, Crawley, West Sussex. RH10 9RR.
Tel: +44 (0)1293 544422 Fax: +44 (0)1293 654321
Copyright © 2013 Elekta AB. All rights reserved www.elekta.com gPOL0007-1 VID:2.0
Please complete the details below and sign the appropriate acknowledgement section:
Existing installations; Acknowledgement by the customer
New installations: New installation confirmation by the installing Elekta or Representative employee
Please return this report to your local Elekta Office or Representative, as soon as possible and within 30 days at the latest.
*The information in this FCO has been provided to address a safety issue and therefore the customer is expected
to acknowledge and accept the recommendations given, and ensure they are implemented. By refusing to
implement the recommendations, the customer assumes full responsibility and liability for all matters (including
costs, losses, claims, and expenses) resulting, whether directly or indirectly from not implementing such
recommendations. Further the customer will hold Elekta harmless from all matters (including costs, losses, claims
and expenses) resulting, whether directly or indirectly from not implementing such recommendations.
Failure to sign and return the acknowledgement may affect any follow-up actions necessary for us to take, and
may require Elekta to report to the Regulatory Authorities in your country.
Classification: Important Field Safety Notice FCO Ref: 200 01 507 086
FCO
Possible risk of a collision when you use XVI and an external beam limiting device
description:
Scope: XVI R4.0 to R4.2.1, R4.5.0, R4.5.1 and R5.0.0 to R5.0.1
Hospital:
*
Acknowledgement by customer :
This notification to be signed by the customer
I acknowledge that I have read and understood this Notice and accept implementation of any given
recommendations:
Name: Title:
Signature: Date:
Signature: Date:
FCO: 200 01 507 086 VID: 1.0 Elekta Limited, Linac House, Fleming Way, Manor Page 7 of 7