Service Manual - Plus - Core
Service Manual - Plus - Core
Important
• Qualified service personnel must perform these procedures.
• See the Carestream Service Portal for possible updates before you begin.
• When doing the procedures in this document, you must use safe work practices and wear the correct personal
protective equipment (for example, safety eyewear) according to your company’s standard operating procedures.
• When servicing this equipment, be sure to read and follow the Manufacturer’s Safety Data Sheets for any chemical,
cleaner, solvent, paint, lubricant, adhesive, or replacement battery to understand proper use and disposal
precautions.
Confidential
Publication No. 6K8894
2017-08-01
Supersedes:
2016-12-15
All rights reserved. No part of this manual may be reproduced or copied in any form by
any means—graphic, electronic, or mechanical, including photocopying, typing, or
information retrieval systems—without written permission.
Contents
1 Safety
2 Replacements
Remove the Bumper...................................................................................................................... 2-1
Install the Bumper ......................................................................................................................... 2-2
Remove the Access Cover Seal Assembly and/or Backup Battery ............................................... 2-3
Install the Access Cover Seal Assembly and/or Backup Battery .................................................. 2-4
Remove the Battery Latch Assembly ............................................................................................ 2-6
Install the Battery latch Assembly ................................................................................................ 2-7
Remove the Battery Contacts ....................................................................................................... 2-8
Install the Battery Contact .......................................................................................................... 2-10
3 Diagnostics
LED Display Matrix......................................................................................................................... 3-1
Reset Button .................................................................................................................................. 3-5
DRX Diagnostic Tool...................................................................................................................... 3-6
Publication History
6K8894 | 2017-08-01 i
1 Safety
Safety
This information defines the safety and information icons used in this publication.
Radiation
This icon is used for conditions that could cause personal injury from radiation.
Caution
This icon is used for conditions that could cause injury to a person, or damage to the equipment
or software data.
ESD
Possible damage from electrostatic discharge
This icon is used for conditions that could cause damage to the equipment.
Electrostatic discharge (ESD) is a primary source of:
• Equipment failure
• Equipment repairs
Electrical components in this equipment can be damaged by a low level electrical discharge
imperceptible to a human being.
Preventive Measures
• Check for an ESD warning label before doing any procedure with ESD-sensitive components.
All sensitive components have graphic labels that frequently include instructions. Use all label
instructions.
• Wear a grounding strap when you touch ESD-sensitive components. Check that the clip
remains fastened to a ground that has a clean surface with no paint.
• Repair components in an ESD-protection area or use a portable grounding mat.
• When moving ESD-sensitive components from area to area, insert and transport the
components in the special material made for the transport of these components.
Important
This icon is used for important information.
Note
This icon is used for additional information.
See the Safety and Regulatory Information for additional safety information.
1. Place the detector on a flat surface with the tube side (1) facing up.
2. Locate the 4 corner bumpers (2)
3. Peel back the trim label (3).
Save the trim label for re-use, unless damaged.
4. Use fine-tip needle nose pliers to remove the 2 pins (4) holding the bumper (5).
Pull the pins straight up to remove.
5. Remove the bumper.
2 4
1 5
H224_6512_C1
2 4 6
3
1 5
H224_6512_C2
ESD
Possible damage from electrostatic discharge
Remember to wear a grounded ESD wrist strap.
3
2
H224_6510_C1
ESD
Possible damage from electrostatic discharge
Remember to wear a grounded ESD wrist strap.
3
2
H224_6510_C1
1. For Backup Battery replacement only. Connect the backup battery wire (5) to the
circuit board.
2. For Backup Battery replacement only. Install the backup battery (6).
Important
Install the Kapton tape to the detector if not currently installed.
3. Remove the release paper from the Kapton tape.
4. Place the Kapton tape across the three bosses to prevent the wires from moving up
on top of the bosses (see photo).
5. Remove the release paper from the foam pad on the bottom side of the seal
assembly.
6. Install:
a. New seal assembly (4)—One corner of the seal assembly is notched for proper
orientation.
b. Access cover (3).
c. 12 screws (2)—use a T5 TORX driver.
Torque the screws to 40 in oz.
d. Label (1).
3 2
H224_6513_C1
1. Install:
a. Battery latch assembly (3)
b. 2 Screws (2)—use a T6 TORX driver.
Tighten the screws to 40 oz-in.
2. Install the battery into the detector.
3 2
H224_6513_C1
ESD
Possible damage from electrostatic discharge
Remember to wear a grounded ESD wrist strap.
2. Remove:
a. 4 screws (1)—using a T6 TORX driver.
b. Plate (2)
2 1
H224_6505_C1
Important
The battery contact is not symmetrical. Observe the orientation of the battery contact before
removing.
3. Remove:
a. 2 screws (1)—use a T6 TORX driver or 1.5 mm ALLEN wrench
b. Battery contact assembly (2)—carefully observe the orientation of the battery
contact before removing
1
H224_6506_C1
4. Clean all the black RTV from around the housing for the battery contact.
Do not use cleaners or solvents.
ESD
Possible damage from electrostatic discharge
Remember to wear a grounded ESD wrist strap.
1. Carefully clean the surfaces (2) on the detector that mate with the battery contact.
Do not use cleaners or solvents.
2. Apply a continuous bead of RTV (3) approximately 1.00 mm (0.03 in.) in diameter to
the front edge (3) of the housing and to the left and right edge (1) of the battery
contacts.
Make sure that the RTV does not get on the plated contacts of the circuit board.
1
2 3
H224_6507_C1
Note
You must install the new battery contact in the same orientation as the old one was installed.
3. Install:
a. Battery Contact (2)
b. 2 screws (1)—use a T6 TORX driver and tighten to 40 oz-in.
4. Apply black RTV to the cracks on the left, right, and top edges of the new battery
contact.
5. Apply a continuous bead of RTV (3) approximately 1 mm (0.03 in.) in diameter in a
closed loop that flows along the front edge of the battery contacts, then crosses over
the caulked top cracks, and follows the overlap area in the housing for the battery
retainer plate.
1 2 3
H224_6508_C1
6. Install:
a. Plate (2)
b. 4 Screws (1)
7. Use foam-tipped swabs to clean up any squeezed-out RTV.
Do not use cleaners or solvents.
With practice, it is possible to do this operation with no squeeze-out and no clean up.
1
2
H224_6509 C1
Postrequisites:
None.
The drawing above locates the Reset button (1), and the LEDs that indicate Error (2),
Communication (3), and Power (4).
The yellow communications LED is an indicator of AP mode for detectors using firmware version 80 or
higher.
AP Mode Client
Mode
Key
Green
Magenta
Yellow
Blue
Orange
AP Client
mode mode
Boot up
Authenticated on wire-
Normal boot up and
less network or Ethernet
connecting sequence
on tether
Firmware update
Reset Button
The Reset button is located on the side of the detector. It is intended for a quick restart
of the detector software, similar to the action of removing and reinserting a battery.
Note
When the reset button is pressed the LEDs will brighten as an immediate feedback response.
Pressing and holding the Reset button for more than 5 seconds disrupts the detector and causes
an immediate hard shutdown. A hard shutdown should only be used when directed by a Service
provider or as a part of troubleshooting a serious condition.
Pressing and releasing the reset button 3 times in 3 seconds is only used in the most serious
conditions. It is intended for Service use only.
Pressing and releasing the reset button 4 times in 3 seconds is only used for detector AP mode.
Press and release 3 times in 3 seconds Resets the detector to private default settings.
(For service use only) • IP Address: 10.0.1.150
• SSID: CSH_DRX_FACTORY
• Passphrase: sundance2009
Press and release 4 times in 3 seconds Places the detector in AP mode and sets/restores the
detector's public default settings:
• IP Address: 192.168.1.150
• SSID: Carestream-serial # of detector
• Passphrase: sundance2009
The country code and channel are not changed
Important
• The diagnostic tool can only be used with the DRX Plus and DRX Core detectors. It cannot be
used on previous detectors such as DRX-1, DRX-1C, DRX 2530C or TDR 3543.
• The detector must be powered on and connected to the console in either wireless or tether
mode.
1. From the DR console or Image Suite Application, exit to desktop to shut down the
Directview or Image Suite software.
2. For Image Suite only, it is necessary to ensure that the Detector Integration
Package (DIP) is not running on the system.
Important
When you are done using the tool, it is necessary to restart the DIP services.
a. Click the Windows Start button.
b. Select Task Manager.
c. Click the Services tab.
d. Select Detector Integration Manager from the list of running services.
e. Right-click and select Stop from the drop-down menu.
f. Close the Task Manager dialog by clicking the X.
4. To connect the application to the detector type the following into the screen.
a. Serial number
b. IP Address—If the IP Address is unknown, leave the field blank.
5. Click Connect.
The application will search for the IP Address if it was not entered in the previous
step.
6. Navigate through the following tabs on the screen to capture the desired
information.
Tab Description
View battery information for the detector. You can click Main
Battery Battery or Backup Battery to view information specific to that
component.
Clear orphan images from the detector. Note that this just deletes
Orphan Images
the images – they will not be saved anywhere.
Tab Description
Note
For Image Suite software, the DRX Remote Command tool can also be launched separately as
an application from the bin folder (same location as DRX Diagnostic Tool).
8. Type the detector IP Address.
9. Select the desired toggle button for information regarding the private network and
Enterprise Wireless performance.
Note
Set Network Mode and the buttons below it are only for Enterprise Wireless configurations.
10. Click Disconnect when you are done using the tool.
Important
For Image Suite, after you are done using the tool, it is necessary to restart the DIP services.
Navigate to the same location in Task Manager, right-click on Detector Integration Manager,
and Start the service.