SYMS II-User Manual EN 10-2012
SYMS II-User Manual EN 10-2012
MANUAL SEALER
Table of contents
1 Foreword ............................................................................................4
2 Introduction .........................................................................................4
2.1 Applications....................................................................................4
2.2 Protection and safety ........................................................................5
2.2.1 Electrical safety standards .............................................................5
2.3 Installation security requirements .........................................................5
2.3.1 Installation ...............................................................................5
2.3.2 Cautionary note ..........................................................................5
2.3.3 Recommendations .......................................................................5
2.4 Transport and storage ........................................................................6
3 SYMS II® ..............................................................................................7
3.1 Overview .......................................................................................7
3.2 Welding unit ...................................................................................8
3.3 Protective cover ..............................................................................9
3.4 Control touch screen ....................................................................... 10
3.5 Supports ...................................................................................... 11
3.5.1 CBS™ Straw and HSV Kit Support .................................................... 11
3.5.2 CBS™ Tube Support .................................................................... 11
4 First commissioning .............................................................................. 12
4.1 Sealer installation .......................................................................... 12
4.1.1 Electrical connection ................................................................. 12
4.1.2 Support positioning .................................................................... 12
4.2 Sealer activation ............................................................................ 13
4.3 Access to settings ........................................................................... 15
4.3.2 Audio settings .......................................................................... 16
4.3.3 Language settings ..................................................................... 16
4.3.4 Screen settings ......................................................................... 17
5 Operating procedure ............................................................................. 18
5.1 Manual mode ................................................................................ 18
5.1.1 CBS™ High Security straws............................................................ 18
5.1.2 CBS™ High Security tube .............................................................. 19
5.1.3 HSV High Security Vitrification Kit .................................................. 19
5.2 Automated mode ............................................................................ 20
5.2.1 CBS™ High Security straws............................................................ 20
5.2.2 CBS™ High Security tube .............................................................. 21
5.2.3 HSV High Security Vitrification Kit .................................................. 21
5.3 Run-up time and transfer time from a mode to another .............................. 22
5.4 Welding time ................................................................................ 22
6 Maintenance ...................................................................................... 23
6.1 Servicing...................................................................................... 23
6.2 Accessories and spare parts ............................................................... 23
6.3 Replacing Teflon strips ..................................................................... 24
6.4 Alarm management ......................................................................... 28
6.4.1 Changing Teflon strips ................................................................ 28
6.4.2 Opening the cover ..................................................................... 29
6.4.3 System error............................................................................ 29
Version 6/2012 2
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
Version 6/2012 3
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
1 Foreword
The operating instructions are designed to provide an accurate description of our SYMS II®
sealer and provide you with a full description.
Included with the operating instructions, you will find a description of the various
components of the sealer, in particular the parts with their references, which are required
when you order spare parts (see 6.2 Accessories and Spare Parts).
The SYMS II® sealer cannot be used to seal any products other than those specified on the
list below:
• Sterile products:
o CBS™ High Security 0.3 mL and 0.5 mL straws
o Co-X High Security 0.3 mL and 0.5 mL straws
o CBS™ High Security 2 mL straws
o CBS™ High Security tubes
o HSV High Security vitrification Kits
• Non-sterile products:
o CBS™ High Security 0.3 mL and 0.5 mL straws
o Co-X High Security 0.3 mL and 0.5 mL straws
o CBS™ High Security 2 mL straws
o CBS™ High Security tubes
To maximize seal integrity of Cryo Bio System products, only use Cryo Bio System products
and accessories when using the system. Filling nozzles, tips and other items have been
specifically designed for use with the CBS™ High Security 0.3 and 0.5 mL straws, CBS™ High
Security tube, HSV High Security Vitrification Kit and CBS™ High Security 2 mL straw.
Caution: Only products at room temperature can be sealed with the SYMS II sealer.
2 Introduction
2.1 Applications
The SYMS II® sealer is a bench top unit that can be used under laminar flow cabinet and is
using a thermal procedure to seal CBS™ High Security 0.3 mL, 0.5 mL and 2 mL straws,
CBS™ High Security tubes as well as HSV High Security vitrification Kits for the following
applications:
Version 6/2012 4
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
2.3.1 Installation
Safe use of this unit requires its connection to a grounded, compliant electrical source.
The SYMS II® sealer is supplied with low tension current through an external power supply.
The voltage and frequency should comply with the instructions in the rating plate.
The power cable should be in good working order and without any damage. If the cable is
damaged in any way, power off the sealer until a replacement cord is available.
If the unit is used in a manner that is not compliant with the specifications of these
operating instructions, the warranty provided with the unit may be compromised.
Any change and/or electrical repair should be made solely by a qualified individual under
the supervision of a customer service technician of Cryo Bio System.
2.3.3 Recommendations
The full seal of all CBS™ High Security products (specified on page 4) is made by the
precise combination of three welding parameters. The parameters include temperature,
pressure and time. These parameters are different for each of the different types of CBS™
High Security products. Failure to use the pre-determined parameters could compromise
the integrity of the seal.
Keep the inside and outside of the CBS™ High Security products sealing area clean and dry.
The sealing support must be adapted to the welding product (3.5: “Supports”).
Version 6/2012 5
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
Do not operate the sealer without a CBS™ High Security product loaded into the machine
(CBS™ High Security straw, CBS™ High Security tube or HSV High Security Vitrification Kit).
Store and transport the unit and its accessories in the original packaging to prevent any
damage.
Version 6/2012 6
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
3 SYMS II®
3.1 Overview
Touch screen
Protective
cover ON/OFF
switch
Presence Base
detectors
Side view
Version 6/2012 7
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
Front view
Top view
Version 6/2012 8
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
When the SYMS II® sealer is operating, the removal of the protective cover will
immediately trigger an audible signal. The display will show an alarm message and ask you
to replace the cover or power down the sealing unit.
Version 6/2012 9
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
1. Types of welding
o Automatic
o Manual
The selected type remains highlighted.
3. Types of container
o Sterile products
o Non-sterile products
Version 6/2012 10
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
3.5 Supports
Two types of supports are delivered with the SYMS II® sealer to provide optimal welding of
all the products suggested by Cryo Bio System:
• CBS™ Straw and HSV Kit Support for CBS™ High Security 0.3, 0.5 and 2 mL straws
and the HSV High Security Vitrification Kit, and
Version 6/2012 11
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
4 First commissioning
Reminder:
The SYMS II® sealer is configured only for the welding of CBS™ High Security sterile and
non-sterile products (specified on page 4).
Place the CBS™ straw and HSV kit support or the CBS™ tube support corresponding to the
type of CBS™ High Security product you are intending to seal.
Version 6/2012 12
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
The following display allows for the changing of the Teflon strip:
125
weldings
Do it now ?
YES NO
Version 6/2012 13
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
Version 6/2012 14
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
1. Standard settings:
o Language
o Audio
o Screen
o Exit
“SETUP” menu
Version 6/2012 15
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
Version 6/2012 16
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
In case of display errors, the screen settings can be reconfigured by selecting the “SETUP”
submenu of the “MAIN SCREEN” then “SCREEN”.
Version 6/2012 17
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
5 Operating Procedure
1. Select “MANUAL” in the main menu, then select “sterile” or “non-sterile”, and
choose the CBS™ High Security product to be sealed
2. Wait until the preheating is completed
3. When the preheating temperature is reached, the unit will make an audible signal
4. Place the straw on the CBS™ straw and HSV Kit support and advance it away from
the user until the straw stops against the presence detector,
5. Press the “WELD” button keeping the straw in position,
6. Wait until the seal is completed, it is indicated by a blue bar,
7. Upon completion of the seal, the jaws will re-open and the “READY” screen is
displayed again,
8. Remove the straw from its support,
9. Repeat to seal the other end of the straw (steps 4 to 8).
CLEANING
HSV
Straw
TEFLON
CHANGE WELD
BACK BACK
High
SETUP BACK Security
TUBE
Version 6/2012 18
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
CLEANING
High
Security
TEFLON TUBE
CHANGE WELD
BACK BACK
SETUP BACK
Follow the same procedure for the CBS™ High Security straw, selecting “TUBE CBS High
Security” in step 1 and using the support adapted to CBS™ High Security tube in step 4.
Follow the same procedure for the CBS™ High Security straw and select “HSV Kit” in step 1.
Version 6/2012 19
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
The automated welding starts when the presence detector detects one of the containers
supported by the SYMS II® sealer.
1. Select “AUTOMATIC” in the main menu, then select “sterile” or “non sterile”, and
choose the CBS™ High Security product to be sealed
2. Wait until the preheating is completed
3. When the preheating temperature is reached, the unit will make an audible signal
4. Place the straw on the CBS™ straw and HSV Kit support and advance it away from
the user until the straw stops against the presence detector
5. When the end of the straw is detected, it is sealed automatically
6. Wait until the seal is completed, it is indicated by a blue bar
7. Upon completion of the seal, the jaws will re-open and “READY” screen is displayed
again,
8. Remove the straw from its support,
9. Repeat to seal the other end of the straw (steps 4 to 8).
Version 6/2012 20
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
SETUP BACK
Follow the same procedure for the CBS™ High Security straw, selecting “TUBE CBS High
Security” in step 1 and using the support adapted to CBS™ High Security tube in step 4.
Follow the same procedure for the CBS™ High Security straw and select “HSV Kit” in step 1.
Version 6/2012 21
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
Data Time
Run-up time for HSV Kit 1 min 25 s
Run-up time for CBS™ straws 2 min 20 s
Run-up time for CBS™ tube 2 min 30 s
HSV Kit/CBS™ straws transfer time 1 min
HSV Kit/CBS tube transfer time 1 min 10 s
CBS™ straws/HSV Kit transfer time 30 s
CBS™ tube/HSV Kit transfer time 40 s
CBS™ tube/CBS™ straws transfer time 10 s with support change
CBS™ straws/tube transfer time 10 s with support change
Hold time at temperature before standby 30 min
Tube or straws cooling time up to 39°C 24 min
The following data are provided for your information only and were measured during the
use of the SYMS II® unit at an ambient temperature of 25°C.
Version 6/2012 22
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
6 Maintenance
6.1 Servicing
There is no specific maintenance required for the main mechanical and electrical parts of
the sealer.
However, it is strongly recommended to regularly clean the sealing plates, especially after
heat sealing of the CBS™ High Security straws, HSV High Security Vitrification Kit, and CBS™
High Security tube. In the event of stain or area of burns on the Teflon strips, the Teflon
should be replaced.
Name Reference
CBS™ straw and HSV Kit Support 022285
CBS™ Tube Support 022286
Protective cover 022288
Teflon strip / 2 022847
Reminder:
Only a professional who is authorized by Cryo Bio System or a customer service technician
of Cryo Bio System shall perform potential repairs.
Version 6/2012 23
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
The “TEFLON CHANGE” function keeps the sealing plates split in order to replace the
Teflon strips.
The Teflon strips must be replaced every 1,000 welds or in case of stains or area of burns
on the Teflon strips.
Version 6/2012 24
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
Version 6/2012 25
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
4. Remove the worn Teflon strip or the strip showing burn marks
5. Apply a new Teflon strip, then close the sealing plate clip
Version 6/2012 26
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
Version 6/2012 27
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
The Teflon strips can be replaced only if the temperature of the dies is lower than 40°C.
Version 6/2012 28
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
As the protective cover is removed, an “ALERT” screen is displayed on the touch screen
and the sealer will generate an audible signal. As the cover is closed, the “MAIN MENU”
appears.
In case of system error, an “ALERT” screen is displayed on the touch screen and the sealer
will generate 3 audible signals every 2 seconds. The sealer is automatically locked.
You should restart the sealer.
If the problem remains, please contact a customer service technician of Cryo Bio System.
Version 6/2012 29
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
7.1 Precautions
All the maintenance operations for the SYMS II® sealer must be performed with the unit
turned off. In case of injury, Cryo Bio System disclaims responsibility.
Cleaning can be performed only if the temperature of the sealing plates is lower than
40°C.
7.2 Cleaning
The “CLEANING” function spreads the sealing jaws in order to clean the unit.
If the sealing plate temperature is higher than 40°C, an “ALERT” screen appears and
indicates the exact temperature. It is essential to wait until the welding unit cools down.
When the temperature is lower than 40°C, the jaws separate.
The SYMS II® sealer can be decontaminated, provided that the unit is not immersed.
Version 6/2012 30
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
8 Technical Specifications
• Dimensions
Length: 363 mm
Width: 276 mm
Height: 152 mm
• Weight: 6 kg
• Indoor use
• Usage environment
The unit cannot be used at an altitude higher than 2,000 m, or in an explosive or
corrosive environment.
• Sealing Plates
Voltage: 12 – 15 V
Version 6/2012 31
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
We, the undersigned, Cryo Bio System, ZI n°1 Est, 61300 L’Aigle, certify that:
• 2004-108/CE and
• 2006-95/CE.
Mr. BEAU
Version 6/2012 32
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
9 Important Information
9.1 Orders
For any spare parts order, thank you for referring to 6.2 : “Accessories and Spare Parts” in
these operating instructions.
Cryo Bio System guarantees its equipment against all malfunctions caused by faulty
material, manufacturing defect, or design error for one year from the date which is
provided on the certificate of guarantee.
Free parts and service will be included during this period.
Based on the above warranty, Cryo Bio System will repair or replace the parts that become
unusable during the warranty period for reasons outlined above.
The implementation of this warranty - that is, the repairs and replacements of the whole
unit or parts of the delivered unit, is not likely to extend the warranty time.
Any dispute caused by the interpretation and/or execution of these warranty conditions,
shall be exclusively subjected to the French law. The Courts of Alençon shall be the only
competent courts including in the event of a third party complaint and several defendants.
Version 6/2012 33
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
The Cryo Bio System Company is not responsible for damages caused by external
influences as well as inappropriate and non-compliant treatments with the
recommendations stipulated in these operating instructions.
Refer to the application fields (see 2.1 Applications) and electrical specifications (see 8.3
Power supply) of these instructions.
9.4 Contact
Version 6/2012 34
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications
This document cannot be disclosed to third parties without written authorization by Cryo Bio System.
SYMS II® Operating instructions
Notes:
Version 6/2012 35
Copyright Cryo Bio System 2010 – Non contractual pictures and specifications