CLM 3100 Service Manual
CLM 3100 Service Manual
CLM 3100 Service Manual
1. INTRODUCTION ................................................................................... 4
1.1. COMPONENTS LIST ........................................................................................................................ 4
1.2. REPAIR PROCEDURE ...................................................................................................................... 6
1.3. CAUTIONS ................................................................................................................................... 7
1.4. DISPLAY LAYOUT FOR ROM INFORMATION ................................................................................. 7
1.5. OPTICAL SYSTEM .......................................................................................................................... 7
1.6. MEASUREMENT PRINCIPLE ......................................................................................................... 9
2
Auto Lensmeter CLM-3100P
4.3 LCD TEST ........................................................................................................................................ 50
4.4. LED TEST ....................................................................................................................................... 51
4.5. UV MODULE TEST ....................................................................................................................... 52
4.6. PD TEST..................................................................................................................................... 53
4.7. THERMAL PRINTER TEST ..................................................................................................... 54
3
1. Introduction
4
Auto Lensmeter CLM-3100P
Foot Swi t ch
Connect or
Interface
Connect or
Power
Swi t ch
Power
Conncentor
5
1.2. Repair Procedure
Power ON
Normal
State
Checking
Abnormal
State
Cleaning the 4
Pinhole
Normal
State
Checking
Abnormal
State
Setup Again
AUTO-LENSMETER
Exit
6
Auto Lensmeter CLM-3100P
1.3. Cautions
VER. 1.01.14
Copyright (C) 2002-2004
Huvitz. Co., Ltd.
CLM-3100Premium
ROM information will be shown on top of the screen if you turn on the power switch.
Optical system resolves the features of the lens such as diopter and prism. It is
composed of emitting light part, transmit part and receiving light part: transmit part
composed of collimator lens and 4 pinhole and receiving light part composed of 2-
dimentional CMOS sensor. The followings are the components of each part:
Transmit Part
Collimator Lens: makes parallel light.
4 Pinhole : disperses parallel light into four signals. Prevent the other light
7
except the light source
MLA :micro lens array. Focus the every four spots.
Emitting light part and receiving light part also built optical system for UV
measurement. One is composed of UV lamp and the other receives UV light. The
wavelength of the UV light is 370nm.
AUTO-LENSMETER
8
Auto Lensmeter CLM-3100P
1.6. Measurement Principle
Basic Principle
When there is no lens (0D state), parallel light that has transmitted the collimator lens
passes the 4 pinhole and makes image on CCD. In case of convex lens (lens with plus
diopter), parallel light that has transmitted the collimator lens converges through the
measuring lens, passes the 4 pinhole and then makes image on CCD. On the other
hand, in case of concave lens (lens with minus diopter), parallel light that has
transmitted the collimator lens diverges through the measuring lens and then passes
the 4 pinhole and at last makes image on CCD.
Initial State
Convex Lens
Concave Lens
9
2. Checking and Setup Method
When you turn on the power switch, you should hear the beep sound and see the
ROM information. At this point, press the three buttons in the right side one after
another from the rightmost one. Then, after the beep sound, your lensmeter,
CLM3100P will start in setup and check mode. While this mode, ‘SETUP’ mark will
flicker in the left-upper area of the centric circles. The booting steps with ‘SETUP’
mode are as follows:
A. Turn on the power switch. Then you can hear the beep sound and see the ROM
information.
B. Press the three buttons in the right side one after another from the rightmost
one.
C. Check that ‘SETUP’ mark flickers in the left-upper area of the centric circle.
SETUP
MARKING OK
AUTO-LENSMETER
Setting and Controlling can be processed only in SETUP mode; 0D, prism, diopter,
and right/left switching position of PD. When setting 0D or diopter, you should
convert the display step (STEP) into 0.01. In case of prism, convert the prism display
into P-B format. Display step (STEP) and prism display format can be changed in
‘Setting the system variables’ screen. To enter the mentioned ‘Setting the system
variables’ screen, press the ‘MODE’ button for a while (2~3 seconds) in the basic
display screen.
10
Auto Lensmeter CLM-3100P
11
2.2. How to Set Origin
Press the ‘CLEAR’ button for a while (2~3 seconds). And then press the ‘MEASR’
button. Next, determine the lens to measure (0D) by using left two arrow buttons in
the following screen. Go ahead and calibrate by using two arrow buttons in the right
side. When setting is completed, press the rightmost ‘SAVE & QUIT’ button to save
and exit. The steps of setting origin are as follows:
Buttons for Selecting Diopter Buttons for Calibrate Diopter SAVE & QUIT button
If you want to return to the previous one without saving in the SAVE screen, press
12
Auto Lensmeter CLM-3100P
the leftmost button to render the order of sixth row ‘-1th (Dip : -30)’. And then
press the ‘SAVE & QUIT’ button.
13
2.3. How to Set Prism
When prism power does not fall in the range ±0.01 referred to the given value in
the standard model-eye specification although you place the 2 lens from the
standard lens set, you should set it up again.
14
Auto Lensmeter CLM-3100P
2.4. How to Set 12 standard lenses
If the measured value is not accurate, you should set the diopter variables again by
using the standard lens-set: for all the 12 lenses in the standard lens set. Always
keep the standard lens clean out of dust or stain.
S –D.00
C +0.00
A 180
PI 0.00
D 0.00
MARKING OK
A. Set the diopter display step into 0.01.
B. Place the standard lens and move it so that ‘Marking OK’ will appear and the
prism X and Y become 0.
C. Press the ‘CLEAR’ button for a while.
D. Press the ‘MEASR’ button.
E. Using the two arrow buttons in the left side, set the diopter to the value of the
lens to measure.
F. Go ahead and calibrate by using two arrow buttons in the right side by the
given value in the standard model-eye specification.
G. To save and exit, press the rightmost button.
H. Repeat the procedures from A to F for 12 standard lenses.
I. Finally, don’t forget to press the ‘PRGA’ button at the second page for a while.
J. After setting diopter, return the diopter display step to 0.25.
The information displayed in the ‘Setting Diopter’ screen is the same as that of
‘Setting 0D’ screen. After finishing setting up 12 lenes, press the ‘PRGA’ button for 2
seconds.
15
2.5. How to Set the Right/left Switching Position of PD
You should specify the encoder position that the PD sensor(=nose) exactly lies on
the center of the lens cap.
A. Fix the PD sensor on the origin and turn on the power switch.
B. Press the ‘CLEAR’ button for a while.
C. Press the ‘DISP’ button three times.
D. Find the location of PD sensor on which PD sensor is the center of the lens cap.
And remember the encoder value at the moment.
E. By using the right and left arrow, you should make the encoder value of the
first row become the very thing at the switching point, which is the value of the
third row.
F. Press the ‘SAVE’ button
G. Finally, press the ‘QUIT’ button to exit.
The information displayed in ‘setting the right/left switching position of PD’ screen is
as follows:
- Set PD Enc : 182 : Position of the Encoder which has been or will be
saved.
- RIGHT : Current position information of the PD sensor (RIGHT or LEFT)
- Now PD Enc : 85 : Position of the current Encoder of the PD
sensor
AUTO-LENSMETER
16
Auto Lensmeter CLM-3100P
2.6. Diopter Setting Variables List
If diopter controlling is performed normally, the length of each row (L) decreases
row by row.
17
2.7. The calibration of the cylindrical axis
With the rectangular cylindrical lens, we can calibrate the cylindrical axis of our
machine. The procedures are as followings:
18
Auto Lensmeter CLM-3100P
2.8. –5D setup for the progressive measurement
For measuring the accurate for the high powered progressive lens, it is needed to
calibrate with –5D lens, which you can get easily the normal glass store. The
procedures are as followings :
19
3. Repair Standard
2
P
5
I 3
c 6
t 1
u
r
e
4
AUTO-LENSMETER
20
Auto Lensmeter CLM-3100P
3 Cnnt bracket - Next, separate 'Knob pin 2' unscrewing two M2.6X8L.
4 RS 232c
5 Foot switch - Remove six screws at the back cover, and then separate
10
'Body rear cover'. And then separate the 'Foot switch'
12
- Be careful that you never scratch on the surface of
13
'Cover' while assembling or disassembling.
14
15
- Assembly is the reverse procedure of disassembly
16
.
17
18
21
3.2. Removing Body front cover (Upper)
2
P
I
c
t
1 3
4
u
5
r
e
6
AUTO-LENSMETER
22
Auto Lensmeter CLM-3100P
3 PCB cover
4 Button PCB - Next, remove 4 spring screws at the TFT-LCD 2' and
6 Brightness knob
10
12 knob 6' and the 'Pcb cover' and 'Button Pcb 4'.
13
15 assembly or disassembly.
16
19 in spare parts.
23
3.3 Removing Printer assembly
P
I
c
t
u 2
r 1
e 6
5 4
AUTO-LENSMETER
24
Auto Lensmeter CLM-3100P
3 Printer cover
6 Printer header
8 take out the 'Printer paper 4' and the 'Printer holder 5'.
12
15
25
3.4 Removing Pd assembly
P
I
c 4
3
t 2
5
u 1 6
r
7
e
12 9
10
11
AUTO-LENSMETER
26
Auto Lensmeter CLM-3100P
3 PD shaft2
4 PD main connect - Unscrew the 'PD shaft2,3' from the 'PD shaft(1).(2)' and
6 Left PD connect
7 PD guide cover - Pull to take out the 'PD shaft1' keeping the orientation
8 Ink case suppt 20deg clockwise rotated. And then, separate the 'PD shaft2',
10 Ink sponge
- Remove the dish-head M2X6L screws and take out the
11 Ink cover(Lower)
'PD guide cover'.
12 Ink cover(Upper)
13
- Separate the 'Ink sub assembly' and the body and take
14
out the 'Ink case support 8' with unscrewing the M3X8L
15
screws.And then separate parts No 12, 11, 9 sequencially.
16
27
3.4.1 Removing Pd sub assembly
P 3 12
15
11 13
I 14
c 10
t
16
u
r 21
19 21
e
4 9
7 1
17
20
18
6
5
2
8
AUTO-LENSMETER
28
Auto Lensmeter CLM-3100P
3 PD lock and pull out the 'Nose shaft 5' and separate parts
5 Nose shaft
6 PD nose spring - Separate the 'PD frame' through removing the M3X6L
10 PD shaft(1)
- Separate the 'Right/left PD sensor', 'Main connector 14'
11 PD shaft(2)
and the PD case through removing M2X5L screws.
12 PD main connect
17 Encoder shaft
- Separate the 'PD etching 20', the 'Encoder etching' and
18 Encoder housing
the 'Encoder housing sub assembly' through removing
19 Enooder etching
M2X4L screws.
20 PD etching
PD shaft
- Be careful not to bend etching parts.
21
29
3.5 Removing Body front cover(Lower)
P
I
c
2
t 5
3
u
r
e
1
4
AUTO-LENSMETER
30
Auto Lensmeter CLM-3100P
3 Filter cover(Upper)
5 Membrane switch and the 'UV cover' from the 'Body front cover'.
10
- Remove the 'Membrane switch sticker'. Clear the gluing
14
- Assembly is the reverse procedure of disassembly.
15
16
17
18
19
20
21
31
3.6 Removing UV assembly
P
3
I
c 4
t
u 2
r
6
e
5
1
AUTO-LENSMETER
32
Auto Lensmeter CLM-3100P
3 UV led
4 UV teflon(Upper) And then separate the 'UV LED 3', 'UV teflon 4',
6 UV teflon(Lower)
33
3.7 Removing Movement assembly
12 11 10
P 6 1
9
I
7 2
c
8
t
4
u
r 3
e
13 5
14
15
21
16
17
18
20
19
AUTO-LENSMETER
34
Auto Lensmeter CLM-3100P
2 Foot stopper(Upper) 'Shaft support 3', 'Control position foot 4' and
5 Pen control suppt - At 'Control position foot 4' and 'Pen control support 5',
8 Position spring
10 Washer
back of 'Position support'. Take out
13 E-ring(E2)
- Next, we have to disassemble the 'foot rubber
14 Position foot spring
sub ass'y' from the 'Control position foot'.
15 Position foot rubber sub ass'y
Remove the 'E-ring(2)', 'Spring 14', 'Spring 15' and
16 Pen guide shaft
'Foot rubber sub ass'y' in order.
17 Pen guide spring
18 E-ring(E2.5)
- Next, we have to disassemble the
19 Pen control housing
'Pen control support 5' and the 'Pen control housing 20'.
20 Pen spring
21 Pen
Take out the 'Pen guide shaft 16' and remove the
35
3.8 Removing Gear assembly
P 8
7 5
4
I
3
c
t
6
u
2
r
1
e
AUTO-LENSMETER
36
Auto Lensmeter CLM-3100P
3 Shaft nut
4 Band washer - Remove the 'Shaft holder' by wrench for M3 Set screw.
5 Gear washer
6 Gear washer - Separate the 'Gear nut 3', 'Band washer 4',
7 Gear shaft 'Teflon washer 5', 'Teflon washer 6', 'Gear shaft 7'
10
- Next, separate the 'Gear shaft 7', 'Gear 8' through
37
3.9 Removing Main board assembly
P
I
c
t
u 1
r
e
2
AUTO-LENSMETER
38
Auto Lensmeter CLM-3100P
5
- Assembly is the reverse procedure of disassembly.
39
3.10 Removing Back bone assembly
P
I 1
c
t
u
r 3
2
AUTO-LENSMETER
40
Auto Lensmeter CLM-3100P
3 Back bone space - Separate the 'Back bone stopper 2' and 'Back bone'
41
3.10.1 Removing Lens housing bone and Led sub assembly
P
I 5
4
c
t 8
7
u
6
r
e
3
AUTO-LENSMETER
42
Auto Lensmeter CLM-3100P
4 LED housing - Remove the 'Lens housing bone retainer 3', the
6 Pin hole(Upper)
7 Teflon scatter - Separate the 'LED housing' and 'Lens housing bone'
9
- Remove the 'Pin hole(Upper) 6', 'Teflon 7' and
43
3.10.2 Removing CCD Camera assembly
P
I 7
c
t
u 6
3
r 5
e 2
1
AUTO-LENSMETER
44
Auto Lensmeter CLM-3100P
4 CCD cover
5 CCD holder(Upper) - Separate the 'CCD holder 5' and 'Back bone', and
6 CCD holder remove the 'CCD holder' and 'CCD camera PCB ass'y'
10
'Filter housing' and after combining the
12
- Assembly is the reverse procedure of disassembly.
45
3.11 Removing SMPS Assembly and Power cover
P
I
c
1
t
5
u 2
r
e 3
6
AUTO-LENSMETER
46
Auto Lensmeter CLM-3100P
3 Power socket
6 Foot rubber
10
- Separate the 'Frame support(UV) 4' and the 'Frame'
11
through removing four M3X5L screws.
12
47
4. Electrical System
text text
10P
DS1
4P
ROM
text
J3 J4
2OP
THERMAL LEFT PHOTO
PRINTER
4P
J5
RIGHT PHOTO
3P
J7 6P
UV SEN IN JP2
3P JP3
CPU J10
J8 ENCODER
UV LED OUT
3P
J9
5P FOOT S/W
2P
J11
DC POWER
J10
MEM S/W
1
4
P
5P
J13 J12
LCD CCFL EEP LCD DATA
ROM
text 4P text
3P 2P 6P 2P 2OP
AUTO-LENSMETER
48
Auto Lensmeter CLM-3100P
4.2 Inspecting SMPS
49
4.3 LCD TEST
YES
Change LCD
50
Auto Lensmeter CLM-3100P
4.4. LED TEST
Check1. Change MAIN BOARD PCB ASS’Y and test again. NO Change CCD module
51
4.5. UV MODULE TEST
52
Auto Lensmeter CLM-3100P
4.6. PD TEST
Change PD MODULE
PD value is so strange…
Check1. Check that PD LEVER is in initial position before you turn the
power switch.
YES
YES
Check1. First, turn off the POWER switch and place PD LEVER in initial
position. And then turn on the power switch.
53
4.7. THERMAL PRINTER TEST
54
Auto Lensmeter CLM-3100P
5.1 Introduction
Downloading program for CLM3100P with RS232 cable (Lensmeter RS232) provides
downloading function of the firmware software from PC to Lensmeter, CLM3100P
with serial cable.
Check the following items before installing and running this software:
See the operation manual for the Lensmeter to learn how to connect PC and
Lensmeter with serial cable.
55
5.2 Installation and Uninstallation
A. Installation
You can take the following steps to install this software. There is no preliminary
condition and if you run the providing setup program, all the components required to
communicate to the Lensmeter will be installed by default.
① Open the Windows explorer and move to the directory that contains setup
program. You will see the screen as shown in [Figure 1].
[Figure 1] Windows explorer view of the directory containing the setup package.
56
Auto Lensmeter CLM-3100P
③ The next dialogue box will inquire the installation directory. You don‟t have
to change the default path but you can do so if there are some reason such
as shortage of the disk space.
④ Choose program group. Select the default value if there is no special reason.
57
⑤ Installation will be processed. All the programs needed will be copied to the
predefined path and the system files such as dll and ocx to the Windows
system path. After completing installation, you see the message as shown in
[Figure 5]. There is no need to restart your system for this software: just
execute this program immediately.
58
Auto Lensmeter CLM-3100P
B. Uninstallation
Uninstallation follows the standard Windows uninstallation method. First, press the
“Close” and “Exit” button to end the downloading program normally before
uninstallation. And follow the below steps.
① Open Windows control panel and double click the “Add/Remove Programs”
icon.
59
② Select the “Downloading Program for CLM3000” in the “Install/Uninstall” tab
as shown in [Figure 7]. And press the “Add/Remove…” button at the bottom
of the screen.
③ Press the “Yes” button in the application removing dialogue box as shown in
[Figure 8] to remove the application
60
Auto Lensmeter CLM-3100P
④ Now all the tasks are completed. Press the “OK” button in the application
removal dialogue confirmation box [Figure 9].
61
5.3 Usage
Now that installation is completed, let’s execute the downloading program. Press the
“Start” menu and select the “Programs”. After that, execute the “Downloading
Program for CLM3000” of the MiraeOptics program group as shown in [Figure 10].
After doing so, you should see the “Download” page shown in [Figure 11].
[Figure 10] Execute the “Downloading Program for CLM3000” in the “Start” menu
AUTO-LENSMETER
62
Auto Lensmeter CLM-3100P
To download the firmware image files to the Lensmeter, take the following steps.
You should keep the order as follows. Otherwise, downloading might not be
completed successfully. Notice that there are two firmware image files to be
downloaded. One is for flash memory and the other is for main program.
① First thing that you need to do is to check the COM port number at PC side of
the serial cable connected to the Lensmeter. Then move to the top of the
screen and set the correct comport number at “1. ComPort” item.
② Click on the “Connect” button in “2.Initialize ComPort” item to initialize the
COM port. Next, switch off the power of the Lensmeter in advance if power
is on. And then pressing the „leftmost button‟ and the „fourth button‟ of the
Lensmeter at the same time, turn on the power of it to enter the PROM mode.
You should hear the beep sound and see the welcome message in the PROM
log textbox as shown in [Figure 12]. (PROM log textbox is located at the left
side of the “Download” screen, [Figure 11].) If you cannot see the welcome
message, try one more time to enter the PROM mode.
63
③ Move to the next item, “3. Choose File & Download” and press the “…”
button in “(1) Flash File” to choose the “almflash.bin” image file for flash
memory.
④ Press “Start” button in (1) Flash File to start downloading. At this point, you
should see the message starting “Now, Downloading… [FILESIZE: 40578”.
The message shows the size of the file in parenthesis that has been sent to
the Lensmeter by that time. After completing downloading, you can see the
messages like [Figure 14].
64
Auto Lensmeter CLM-3100P
⑤ Next, press the “…” button in “(2) Main File” of “3. Choose File & Download”
item to select the image file for main program, “alm_main_bin”.
If you encounter problems when downloading, see the following chapter, “5.4.
Troubleshooting”
65
[Figure 16] Completion message for downloading main image file
AUTO-LENSMETER
66
Auto Lensmeter CLM-3100P
5.4 Troubleshooting
Here are some hints for the troubles that might occur during downloading process.
Before you contact your agency or Mirae Optics, check the following list carefully.
67
firmware image files should be done again.
⑧ There are strange characters in the PROM log textbox.
It might be the case that you run PROM mode before you enter the
CONNECT mode. If so, switch off the power of the Lensmeter and do the
download procedure again from the first step.
⑨ I have received the welcome message in the PROM log textbox but there is
no response after start downloading. Of course, I checked the COM port
number.
Check your cable connection. Test the cable connection by using the
communication software like HyperTerminal provided basically in
Windows 95/98™. First, select COM port and change the number of bits
per second into “115200” in the property menu, stream control into “No”.
Then do the test. If there is no response, it may be the fault cable.
⑩ The size of the downloaded file in the PROM log textbox does not coincide
with the real one.
Stop the process and restart the program to download the image files
again. In case it occurs again, test your Lensmeter after downloading. If
there are some problems, consult your agency immediately.
⑪ The vertical scroll bar doesn‟t work.
Normally, the vertical scroll bar doesn‟t work until you press the “Close”
button to close the session.
If you cannot solve the problems with above tips, contact your agency after checking
the version of your hardware/software and determining the abnormal case.
AUTO-LENSMETER
68
Auto Lensmeter CLM-3100P
5.5 Appendix
Firmware image files are not the accessories of this software. Please ask your agency
for image files.
69