This document contains 12 problems related to instrumentation and measurement of various systems. Specifically: 1) It determines the overall sensitivity and temperature change corresponding to a recorder pen movement for a temperature measuring system. 2) It plots data from a tachogenerator measuring angular speed and determines the order of the system. 3) It calculates the time constant and static sensitivity of a mercury-in-glass thermometer described by a differential equation. 4) It finds the time at which the ratio of output to input signals is 0.8 for a first order instrument with a given time constant, subjected to a step input.

Uploaded by

danielruthers2
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
54 views

Tutorial 2

This document contains 12 problems related to instrumentation and measurement of various systems. Specifically: 1) It determines the overall sensitivity and temperature change corresponding to a recorder pen movement for a temperature measuring system. 2) It plots data from a tachogenerator measuring angular speed and determines the order of the system. 3) It calculates the time constant and static sensitivity of a mercury-in-glass thermometer described by a differential equation. 4) It finds the time at which the ratio of output to input signals is 0.8 for a first order instrument with a given time constant, subjected to a step input.

Uploaded by

danielruthers2
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

INSTRUMENTATION & MEASUREMENT

TUTORIAL NO: 2

(ZERO AND FIRST ORDER SYSTEMS)


1. The individual sensitivities of different elements comprising a temperature measuring system
are:
Transducer 0.3 /0C
Wheatstone bridge 0.01 V/
Amplifier gain 80 V/V
Pen recorder 0.12 cm/V
Determine the overall sensitivity and the temperature change corresponding to a recorder pen
movement of 3 cm.
2. The following data pertains to measurement of angular speed by a dc tachogenerator.
Angular Velocity (rev/min) 0 500 1000 1500 2000 2500 3000
Output Voltage (Volts) 0 8.9 15.1 23.5 29.7 38.8 47.5
Plot the results and determine the order of the system and make an estimate of sensitivity of
the instrument.
3. Determine the time constant and static sensitivity of mercury-in-glass thermometer whose
dynamic performance is prescribed by the following differential equation:
dq o
4.5 + 3 q0 = 1.5  10-3 qi
dt
where q0 is the height of the mercury column in m and qi is the input temperature, in 0C.
4. A first order instrument has a time constant of 0.2 s. For a step input to the instrument, find
the time at which the ratio of the output to the input signals is 0.8.
5. Assume that a particular temperature probe approximates first-order behavior in a particular
application, that it has a time constant of 6 s, and that it is suddenly subjected to a
temperature step of 750C – 3000C.
(a) What temperature will be indicated 10 s after the process has been initiated?
(b) Find the indicated temperature after 10 s if the same probe is subjected to a step 3000C –
750C.
6. A thermometer is initially at a temperature of 200C and is suddenly plunged into a liquid
bath, which is maintained at 1500C. The thermometer indicated 950C after time interval of 3
s. Estimate the time constant for the thermometer. Also calculate the indicated temperature
after five time constants and comment upon the results.
7. A measuring element with a time constant of 0.4 s and a static sensitivity of 0.05 mV/ 0C is
used to measure the temperature of a medium, which changes from 20 to 600C. Taking the
output as zero at 200C, find the output voltage at the end of 5 s, if the temperature changes
from 20 to 600C, occurs at a constant rate in 5 s.

INSTRUMENTATION & MEASUREMENT 1 TUTORIAL NO. 2/Khem Gyanwali


8. In a response test of a thermometer, which was suddenly put in water bath at 1000C, the
following data was obtained:
Time, t (s) 0 1 3 6 8 11 15 18
Temperature, (0C) 30 50 65 80 90 95 98 99
(a) Find the time constant of the thermometer.
(b) Find the steady state error if the thermometer is used to measure temperature of liquid
cooling at a rate of 10C every 6 s.
9. A first order instrument has a time constant of 2 s. For sinusoidal input to the instrument, find
the frequency at which the amplitude of the output signal is 0.8 times the amplitude of the
input signal.
10. A first order instrument must measure signals with frequency content upto 100 Hz with an
amplitude inaccuracy of 5 %. What is the maximum allowable time constant? What will be
the phase shift at 50 and 100 Hz?
11. The input signal of qi = sin 0.2t + 0.3 sin 2t is given to two different first order instruments
with time constants 1 = 2 s and 2 = 0.02 s respectively. Find out the output response of the
above instruments. Comment upon the results.
12. A thermocouple with a time constant of 0.3 s and a static sensitivity of 0.04 mV/ 0C is used to
measure the temperature of a medium, which varies as shown in Figure P2.12. If the initial
temperatures of the thermocouple reference and measuring junctions was 250C, find the
output of the thermocouple at t = 0.6 s and 30 s.

1000C

25 0C time
0 0.6 30.0
Figure P2.12
ANSWERS
1. 0.0288 cm/0C, 104.1670C 8. 4.5 s, 0.750C
2. Zero order, 0.0153 V/RPM 9. 0.597 Hz
3. 1.5 s, 0.5  10-3 m/0C 10. 0.523 ms, – 9.330C, – 180C
4. 0.32 s 11. 0.93 sin (2t – 21.80) + 0.073 sin (20t –
5. 2570C, 1170C 760); sin (2t – 0.230) + 0.3 sin (20t –
6. 3.487 s, 149.120C 2.30C)
7. 1.84 mV 12. 2.6 mV, 0.03 mV

INSTRUMENTATION & MEASUREMENT 1 TUTORIAL NO. 2/Khem Gyanwali