Experiment 1 CHM510
Experiment 1 CHM510
INTRODUCTION :
Resolu on is the measured of how well separa on in the column. The formula of Resolu on
is :
PROCEDURES :
For this experiment, gas chromatography GC was used to determine the op mum flow
rate and temperature of methyl esters to achieve the best separa on and resolu on
condi ons. This experiment was also used to determine the various components of the
standard methyl ester mixture. A standard methyl ester is made up of three components:
methyl laurate, methyl myristate, and methyl palmitate. Split injec on with ra o of 40:1 was
used to set up the GC instrument. The flame ionisa on detector (FID) was used in this
experiment because it is suitable for methyl esters due to their vola lity.
Firstly, the effect carrier gas flow rate on isothermal GC separa on of methyl esters were
inves gated. 0.4 L of standard mixture was injected at 210 ℃ at a carrier gas flow rate of 30
cm sec-1 isothermally. A er the reading, the flow rate raised to 50 cm sec-1. Allowed the
system to equilibrate for a few minutes before injec ng the standard mixture again. The
procedure repeated at a flow rate of 70 cm sec-1. Then, the most suitable flow rate was
decided to be used in the second part of the experiment.
For the second part of the experiment, effect of column temperature on the isothermal
GC separa on of methyl esters were inves gated. Firstly, 0.4 µL standard mixture inject was
isothermally at 170℃, followed by 190℃ at the op mal carrier gas flow rate. A er the reading,
this factor was determined by evalua ng peaks, resolu on, and analysis me for each
temperature 170 ℃, 190 ℃ and 210 ℃.
Next, the third part of this experiment, separa on of methyl esters using column
temperature programming. Temperature programming refers to changing the temperature of
the column during chromatography analysis which provides a stable thermal environment for
the analysis. Standard mixture was injected at the op mal carrier gas flow rate and op mal
flow rate using a linear temperature ramp from 100 ℃ to 290 ℃.
Then, the resolu on was calculated by using the resolu on formula. If the resolu on was
not adequate, the temperature programming should be modified by increasing the
temperature to permits the higher resolu on of lighter compounds and sharp peaks for
heavier compounds, reducing the long run mes generated by heavier compounds. For the
last part of this experiment, iden fica on of components in methyl esters mixture. Each
methyl ester was injected individually by turn in the op mized temperature and flow rate. The
reading was compared to each other to iden fy the compound in the standard mixture.
Sample calcula on :
a. To calculate resolu on
At 30ms-1 ,210 C
( . . )
𝑅𝑠 = =26.227
( . . )
b. To calculate average Rs
At 30ms-1 ,210 C
. .
=27.008
DISCUSSION
Reten on me (TR ) is the amount of me that elapses between the introduc on of
sample and the highest signal of a specific compound at the detector. Resolu on refer to the
ability to dis nguish between two signals. In chromatography, this is the ability to dis nguish
between two peaks.
A er obtaining the results, we proceed to calculate the resolu on to determine the
op mal condi ons which include the temperature and flow rate of the mobile phase. We
begin analyzing the op mum flow rate, using 30 ms-1 , 50 ms-1 and 70 ms-1 with an isothermal
temperature 210°C. We observed that at 70 ms-1, 210°C has the lowest resolu on for both
peak 2 and peak 3, and peak 3 and peak 4, which is the best for our me analysis. Therefore ,
we choose 70 ms-1 as our op mum flow rate. For temperature, we use 70 ms-1 as constant
flow rate and vary the temperature between 170 , 190 and 210 . We observe that at 210 has
the lowest resolu on for both peak 2 and 3, and peak 3 and peak 4, therefore we choose 210
as our op mum temperature.
Hence at 70 ms-1, 210 C is the op mum condi on for our analysis to be done prac cally.
Comparison of our sample injec on with each standard methyl ester shows that at 70 ms-
1,210 C chromatogram has the closest results with the standard methyl ester’s reten on me.
However , there are poten al errors in this experiment such as slow injec on which could lead
to bigger dead me (Tm) and cause the analysis to be done slower . addi onally, it is important
to choose the appropriate injec on mode when analysing concentrated methyl esters, as we
have used split mode to ensure the analysis is accurate for only the ra o taken from the
sample.
CONCLUSION
In a nutshell, the op mum gas carrier flow rate and op mum column temperature were
obtained which are 70 ms-1 and 210 °C respec vely. 70ms-1 and 210 °C are also the most
suitable methyl esters to obtain an ideal resolu on separa on. Overall the experiment was
successful in achieving its objec ves.
REFERENCES
1. Gurleen Kaur, & Sahil Sharma. (2018, July 2). Gas Chromatography -A Brief
Review.
h ps://www.researchgate.net/publica on/344042922_Gas_Chromatography_-
A_Brief_Review
3. Venton, D. B. (2018). JoVE Science Educa on. Retrieved from Gas Chromatography
(GC) with Flame-Ioniza on Detec on: h ps://www.jove.com/v/10187/gas-
chromatography-gc- with-flame-ioniza on-detec on