Helium Leak Detector
Helium Leak Detector
User Manual
Notices
Manual Part Number Warranty Safety Notices
G8610-90002 The material contained in this
Edition B.00, July 2023 document is provided “as is,” and is CAUTION
subject to being changed, without
notice, in future editions. Further, to the A CAUTION notice denotes a hazard. It
Copyright maximum extent permitted by calls attention to an operating
applicable law, Agilent disclaims all procedure, practice, or the like that, if
© Agilent Technologies, Inc. 2023 warranties, either express or implied, not correctly performed or adhered to,
No part of this manual may be with regard to this manual and any could result in damage to the product
reproduced in any form or by any information contained herein, including or loss of important data. Do not
means (including electronic storage but not limited to the implied proceed beyond a CAUTION notice
and retrieval or translation into a warranties of merchantability and until the indicated conditions are fully
foreign language) without prior fitness for a particular purpose. Agilent understood and met.
agreement and written consent from shall not be liable for errors or for
Agilent Technologies, Inc. as governed incidental or consequential damages in
by United States and international connection with the furnishing, use, or WARNING
copyright laws. performance of this document or of
any information contained herein. A WARNING notice denotes a hazard.
Should Agilent and the user have a It calls attention to an operating
Agilent Technologies Italia S.p.A. separate written agreement with procedure, practice, or the like that, if
Vacuum Products Division warranty terms covering the material not correctly performed or adhered to,
Via F.lli Varian, 54 in this document that conflict with could result in personal injury or death.
10040 Leinì (TO) these terms, the warranty terms in the Do not proceed beyond a WARNING
ITALY separate agreement shall control. notice until the indicated conditions
are fully understood and met.
Technology Licenses
www.agilent.com
The hardware and/or software
described in this document are
Instrument Manufacturing furnished under a license and may be
Agilent Technologies LDA Malaysia used or copied only in accordance with
Sdn Bhd Bayan Lepas the terms of such license.
Free Industrial Zone 11900 Bayan Lepas,
Penang Malaysia Restricted Rights Legend
U.S. Government Restricted Rights.
Software and technical data rights
granted to the federal government
Printed in Italy include only those rights customarily
provided to end user customers.
Agilent provides this customary
commercial license in Software and
technical data pursuant to FAR 12.211
(Technical Data) and 12.212
(Computer Software) and, for the
Department of Defense, DFARS
252.227-7015 (Technical Data -
Commercial Items) and DFARS
227.7202-3 (Rights in Commercial
Computer Software or Computer
Software Documentation).
Contents
2 Technical Information 45
Description of the Leak Detector 47
Leak Detector Configurations 49
Technical Specifications 50
Front Panel Display and Operations 52
Home Screen 53
Front Panel Display - Home Screen Display 55
Rear Panel 57
Features 61
Operating the Leak Detector 62
Screens 62
Basic Functions 62
Home Screen 63
Graphical Display Screen 69
Menus Screen 70
Calibration Screen 71
Calibration Set-Up 71
Test Configuration Screen 77
Test Configuration 77
Reject and Audio Set Points 77
Test Configuration Advanced Features 80
Transfer to Test 83
Stabilization Wait Timer 84
High Pressure Test Set-Up Procedure 84
Home Screen - Activate Gross Test 85
Application Setup 86
Sniff Method 87
Spray Method 89
High Sensitivity 91
Auto Sequencer 92
Auto Sequencer Set-Up 95
Rough Time Setup 95
Test Time Setup 96
Set Point 1 Setup 96
Set Point 2 Setup 96
Using Set Point 1 and 2 together (Optional) 96
Pressure Set point (High & Low Limit) - Optional 96
Split Flow 97
PPM (parts per million) Sniff Method 100
Settings Screen 110
Settings Menu 110
Main Settings 111
Configure Date and Time 113
Screen Settings 115
Analog Output 117
Remote Control 119
Password and Restrictions 120
Wireless 121
Maintenance Screen 122
Maintenance Menu 122
Software Update 123
Internal Calibrated Leak Setup 124
Manual Tuning 127
Validity
This manual lists the instructions for the users of the Helium Leak Detector (HLD),
with particular reference to the notions relating to safety, operation and first level
maintenance, limited to maintenance operations for which the user is responsible.
The maintenance operations, illustrated in the specific sections, with specific
provisions relating to the higher level of maintenance (personnel specifically trained
for maintenance operations) must not be carried out by the user.
For a correct installation and start/stop, please refer to "Technical Information"
section.
NOTE 1 This manual contains useful information so that all personnel using the
Helium Leak Detector (HLD) can operate it safely and guarantee perfect
efficiency, for its entire life span.
2 Keep this manual, together with all the related publications, in an accessible
place known to all operators/maintenance personnel.
CAUTION Caution messages are displayed before procedures which, if not observed, could
cause damage to the equipment.
NOTE Notes are intended to call attention to important information and provide more
detail regarding specific steps.
Warning Symbols
The following is a list of symbols that appear in conjunction with warnings on the
HLD. The hazard they describe is also shown.
A triangular symbol indicates a warning. The meanings of the symbols that may
appear alongside warnings in the documentation are as follows:
The following symbol may be used on warning labels attached to the instrument.
When you see this symbol, refer to the relevant operation or service manual for the
correct procedure referred to by that warning label.
Generic Hazard
CE certification
Safety
This section contains the information, prescribed by the Low Voltage Directive
2014/35/EU, which is essential for the compliance and observance of the safety
regulations both generally and in relation to the specific use of the product.
Failure to comply with these instructions and the other instructions contained in this
manual may render the safety conditions envisaged in the design phase inefficient and
cause accidents to those operating the product.
Agilent Technologies declines all responsibility for damage to the product or for the
physical safety of the operator or third parties deriving from the non-observance of the
safety rules indicated in the technical documentation.
Solvents
WARNING The mechanical components of leak detectors are typically cleaned with
alcohol, methanol, or other solvents.
When heated, sprayed, or exposed to high-temperature equipment, these
solvents become flammable and explosive, causing serious injury or death.
Do not use these solvents near a high-temperature source. Ventilate the
working area with a blower and work in a large, well-ventilated room.
Alcohol, methanol, or other solvents are irritants, narcotics, depressants
and/or carcinogens. Their inhalation and/or ingestion may produce serious
side effects. Prolonged or continued contact with the skin results in
absorption through the skin and moderate toxicity. Always ensure that
cleaning operations are carried out in large, well-ventilated rooms, and wear
eye shields, gloves, and protective clothing.
To clean the leak detector enclosure and the Touch Panel display use only a
soft cloth slightly dampened with water or a mild soap.
DO NOT USE EXCESS WATER OR CLEANING SOLVENTS OF ANY KIND.
Avoid splashing any cleaning solvents into the unit through the ventilation
openings or Front Panel buttons. Wipe the surface with a dry lint-free cloth.
CAUTION Do not clean any aluminum parts with Alconox®. Alconox is not compatible with
aluminum and will cause damage.
O-ring Care
When removing, checking or replacing O-rings, keep in mind the following:
NOTE Agilent recommends replacing all O-rings during routine maintenance or during any
maintenance procedure requiring that O-rings be removed.
NOTE Due to the effective cleaning nature of VacuSolv solvent and its residue-free
properties, Agilent’ Component and Spectrometer Cleaning Kit (Part Number
670029096), used in accordance with the kit instructions, is recommended for
cleaning spectrometer components. The kit can also be used for fine cleaning of
other parts in the leak detector’s vacuum system such as valves and fittings. No
rinsing steps or high-temperature drying are required following cleaning with
VacuSolv. Although appropriate precautions are advised, VacuSolv is compatible
with most materials and does not contain toxic chemicals or CFCs
(chlorofluorocarbons).
Other acceptable solvents are isopropyl alcohol (IPA) or Dow Corning® OS-20.
CAUTION Remove O-rings carefully with your fingers. Do not use metal tools for this
task; this prevents scratching of any sealing surfaces.
• Wipe all O-rings clean with a lint-free cloth before installation to ensure that no
foreign matter is present to impair the seal.
• Do not use grease or any other substance on O-rings that will come in contact
with the spectrometer.
• Do not use alcohol, methanol or other solvents on O-rings. Doing so causes
deterioration and reduces their ability to hold a vacuum.
• Agilent does not recommend the use of vacuum grease. If applicable, apply a
small amount of Apiezon® L grease and wipe the O-rings shiny dry.
CAUTION Metal Seals must be replaced during routine maintenance or during any
maintenance procedure required. All fasteners must be installed and torqued per
assembly procedure specifications. Remove Metal Seals carefully with your
fingers. Do not use metal tools for this task, this prevents scratching of any
sealing surfaces. Wear non-powdered, ESD safe Nitride or equivalent gloves prior
to removing or replacing Metal Seals.
• All Metal Seals come in clean room condition packages. No cleaning is
required. If necessary, Metal Seals can be cleaned using alcohol or methanol.
Wipe all Metal Seals clean with a lint free cloth before installation to ensure
that no foreign matter is present to impair the seal.
• Do not use grease or any other substance on Metal Seals that will come in
contact with the spectrometer.
Equipment, General
Environment of intended use:
Indoors use in an industrial and laboratory installation only.
Altitude up to 2000m
POLLUTION DEGREE 2, Material group III
Operating Temperature: +12 °C to +40 °C, Relative Humidity (RH) up to 90%
max., non-condensing.
Storage ambient conditions: RH from 0% to 95%, non-condensing, temperature:
-20°C to +60°C.
WARNING The leak detector is not designed for use with hazardous gases.
Verify that the system to be tested has been purged of all hazardous gases
prior to using the leak detector. When testing a system that contained
hazardous gases, connect the exhaust of the leak detector to a scrubbed or
toxic containment exhaust. Exposure to hazardous gases could result in
serious injury or death.
WARNING
Equipment tightness is guaranteed for normal operating conditions when the
equipment leaves the factory. It is the user’s responsibility to maintain the
level of tightness, particularly when pumping dangerous products.
CAUTION The performance and operating safety of this equipment can only be guaranteed
if it is operated according to normal conditions of use.
Allow at least 2 to 3 feet of clearance adjacent to the slots at the front, sides and
back of the equipment enclosure.
Allow sufficient room for the operator to safely maneuver to perform testing.
CAUTION Do not apply excessive pressure to the Touch Panel display as this may cause
smears.
In general, the lower the temperature, the longer it takes the Touch Panel display
to turn on. Performance of the Touch Panel display may deteriorate in low
temperature.
Radio
The Wireless Remote Control and Base Board PCB are built around Laird's, model
RM024, RF transceiver module operating in 2.4 GHz ISM band. The same type of
the RF module is built in an Agilent Technologies Helium Mass Spectrometer Leak
Detector Wireless Base Station PCB and Wireless Remote unit (see Agilent
Technologies Helium Mass Spectrometer Leak Detector Wireless Remote
Operational Manual (6999099945A).
Agency Identification Numbers for Laird, Model: RM024 AC4424-100:
US/FCC ID: KQL-RM024
CAN/IC: RSS-210
EUR/ETSI EN: CE
Taiwan: CCAI13LP3400T1, CCAI13LP3330T3
are also applied to the Agilent's Wireless Remote and Agilent Technologies Helium
Mass Spectrometer Leak Detector Wireless Base Station, when the units are used
the approved antennas.
WARNING The Wireless Remote complies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions: (1) This device may not cause
harmful interference, and (2) This device must accept any interference
received, including interference that may cause undesired operation.
To satisfy RF exposure requirements for mobile type transmitting devices,
maintain a separation distance of 20 cm or more between the antenna of the
Wireless Remote and persons during operation.
Maintain a separation distance of 20 cm or more between the operator and
the RF antenna of the leak detector with a wireless base station.
WARNING Depending on the unit configuration ordered, the mains power supply
requirements for Agilent Technologies Helium Mass Spectrometer leak
detectors are:
• 100 VAC, 50 Hz, 5 A
• 115 VAC, 60 Hz, 5 A to 9 A (G8611A)
or
• 230 VAC, 50/60 Hz, 3 A to 4 A
Before powering the unit for the first time, verify that the unit is configured to
operate for the local mains supply voltage. Need nominal current for all
models.
CAUTION Use surge protection to improve the immunity of the leak detector against
unidirectional transients caused by the following phenomena:
• Switching phenomena in the power network (e.g., switching of capacitor
banks, inductive loads, electrical motors, etc.)
• Faults in the power network
• Indirect lightning strokes.
This equipment has been tested and found to comply with the limits for Class A
CAUTION digital device, pursuant to Part 15 of FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when it is operated in
a commercial environment.
This equipment generates, uses, and can radiate RF energy, which if not installed
and used in accordance with the instructional manual, can cause harmful
interference to radio communications.
When this equipment is operated in a commercial environment, operation is
subject to the following two conditions:
• This equipment must not cause harmful interference.
• This equipment must accept any interference received, including interference
(RF and ESD) that may cause undesired operation.
The equipment may need to be reset after RF and/or ESD events by cycling the
Power Switch/Circuit Breaker on the back panel of the unit.
Operation of this equipment in a residential area is also likely to cause harmful
radio communications interference, in which case, the users will be required to
correct the interference at their expense.
Spectrometer
Store the Ion Source/Preamplifier sub-assembly in a cool, dry area in a
WARNING tightly sealed, ESD protected container. Wear lint-free gloves when handling
the spectrometer. Wash hands thoroughly after handling the spectrometer
filaments and especially before smoking or eating.
CAUTION The spectrometer operates at a very high vacuum produced by the high vacuum
pump. Service of the spectrometer require that this vacuum be vented to the
atmosphere.
CAUTION Do not use grease or any other substance on O-rings or metal seals that will
come in contact with the spectrometer.
CAUTION If the spectrometer magnet comes in contact with a magnetic surface, the
magnet can become degraded causing the spectrometer to lose sensitivity.
Pumps
To avoid injury, use proper lifting techniques when moving pumps. Two
WARNING people may be required to move pumps safely.
To avoid injury, wait until the turbo pump is completely stopped before
WARNING disconnecting it from the vacuum system.
Check the oil level often. Do not allow oil-based mechanical pumps to run
when the oil level is below the LOW mark.
Damage to pumps can occur if operated with no oil, if the oil level is too high
or if the oil has become discolored.
Protective equipment
The protective equipment of the operators who are executing the maintenance of
the Leak Detector must always be adequate for the type of operation being
executed. Furthermore, it must comply with the safety requirements of the
legislation in force in the country in which the machine is used.
In general, the operator must wear accident-prevention shoes while handling the
Helium Leak Detector and during installation.
• Wear safety shoes with steel toe cap according to directive EN 347.
Product Description
This equipment is designed for use by trained personnel. The user should read this
instruction manual and any other additional information supplied by Agilent before
operating the equipment. Agilent will not be held responsible for any events that
occur due to non-compliance with these instructions, improper use by untrained
persons, non-authorized interference with the equipment, or any action contrary to
that provided for by specific national standards.
The following paragraphs contain all the information necessary to guarantee the
safety of the operator when using the equipment. Detailed information is supplied
in “Technical Information”.
If the leak detector is not used immediately, it can be stored as received without
special precautions. A dry, relatively dust-free area is preferred. The required
environmental conditions for storage are:
0% to 95% relative humidity, non-condensing
-20 °C (-4 °F) to + 60 °C (+ 140 °F) ambient temperature.
CAUTION Never leave the AA batteries in the Wireless Remote unit if it is will not be in use
for a long period of time.
Installation
Preparation for Installation
The following items are included in the shipment:
Agilent Technologies leak detector configured and completely assembled as
ordered
All necessary power cords, as ordered
Any optional equipment ordered.
Unpacking
WARNING Unpacking the Leak Detector: Always work in a team of at least two people.
Wear personal protective equipment (PPE), including gloves, boots, and a
weight belt to avoid injury.
ON
2 Examine the leak detector for damage (whether a mobile unit with cart and
wheels, portable, or benchtop), and remove the remaining shipping straps and
top plastic shower cap cover (see following figure).
1 Use a flat tip screwdriver to remove the side walls from the leak detector cart.
Remove the panels and set aside.
2 For the:
IDP-15 model, also remove the rear panel of the cart, opposite of the handle
to access the rear of the pump, opposite the pump cooling fan. If any,
remove and file paperwork attached to the pump.
Use the provided 4 mm T-handle Allen wrench to turn the locks clockwise
to loosen; the IDP-15 ship locks only tighten/loosen, and are not removable
(see following figure).
For the TS620 pump, use a 10 mm driver to remove the shipping bolts (one
on each side of the pump base) (see following figure).
3. Adding Oil
Oil Rotary Vane Pumps must be filled with oil before operation. These include the
DS40-M, DS302, and DS602 pump model leak detectors (see following figure).
a) Use a funnel when adding oil, fill to the center of the sight glass marking.
b) For the:
DS40-M, open the top unit of the leak detector side away from screen
(see following figure).
Mobile units that require oil (DS302 and DS602) units, open the bottom
cart and fill to middle of sight glass marking (see following figure).
Powering Up
1 Ensure that the roughing pump is switched on and if applicable (IDP-15), set to
the correct voltage setting.
2 Switch the leak detector on, cover the leak detector inlet with a blank-off cap
(provided and already in place), and wait for the screen to read System Ready
before operating.
Installation requirements
Installation requirements are described in the following table.
Item Requirement
NOTE: When installing a bench system with a mechanical pump, the power
cord for that pump, if connected to the back of the leak detector, cannot exceed
three meters. Additionally, connect the mechanical pump to the leak detector
using the shortest possible vacuum connection to maximize vacuum
performance.
Mains Supply Voltage Fluctuations Must not exceed ±10% of the nominal voltage.
Figure 11 Bench Mount Leak Detector with External DS302/DS602 Wet Rotary Vane Pump,
Model G8612A\GS8612
The following figure is a drawing of a Bench Mount Leak Detector with External
TS620 Dry Scroll Pump, Model G8612B with individual components shown.
NOTE Cut the ¾” black tubing into two equal parts as shown.
Figure 12 Bench Mount Leak Detector with External TS620 Dry Scroll Pump, Model G8612
The following figure is a drawing of a Bench Mount Leak Detector with External IDP-
15 Dry Scroll Pump, Model G8612C with individual components shown.
Figure 13 Bench Mount Leak Detector with External IDP-15 Dry Scroll Pump, Model G8612C
Helium
Welding grade helium in a standard cylinder with pressure regulating valve and
hose is required for testing products and leak checking.
Startup Procedure
To start the leak detector:
1 Plug the leak detector power cord into an appropriate mains supply receptacle.
2 Move the power switch on the rear power control panel to the l position. If you
have a bench system with a separate backing/roughing pump, turn the pump
on. The Test button is disabled on start up until System Ready appears on the
Home screen. Once the leak detector boots up a startup wizard appears on the
touch panel display allowing you to change the time and language, and the
ability to scan a QR code for product registration.
3 Select choice of language, time/type (AM, PM & 24) and system setup.
The System Ready indication appears on the Home screen. If the leak detector
is being started after a long period of being off (several hours), it can take up to
five and a half minutes to stabilize and provide reliable, quantitative leak rate
readings.
4 The leak detector is capable of performing an automated calibration routine
using either an internal calibrated leak (factory installed), or an external
calibrated leak placed into the system test port. Refer Section “Calibration Set-
Up” to properly set-up the leak detector for auto-calibration.
NOTE If the system was recently calibrated before shut down, it can be used for
qualitative testing. Read the standard leak to verify accuracy. For more accurate
quantitative leak values, perform a calibration 20 minutes after start-up.
To accurately read in the most sensitive decades (E-10 and E-11), warm up the leak
detector for at least one hour prior to performing a full calibration. When leak
checking in the E-9 atm-cc/sec range or higher, you can use the leak detector prior
to one hour and verify system performance by assuring that the system can
accurately read the internal calibrated leak.
The system is shipped from Agilent with the Minimum Range function turned ON
and the bottom range set at the E-10 range (atm-cc/sec). To accurately test in the
E-11 range, careful control of materials and the environment is required, consult a
Agilent sales engineer for support.
NOTE The firmware initially starts at a large time interval, but recalculates the required
time as the pumping continues. The new time is reflected on the timer.
NOTE Test Configuration Menu > Applications Setup > High Sensitivity automates and
walks you through the E-10 External Calibration procedure.
NOTE When calibration is complete, the leak detector reverts to Fine Test mode.
Shutdown Procedure
The system can be in any mode when shut down. When the system is shut down,
all valves in the system close. If the system is not in Vent mode, the test port
remains under vacuum and may be difficult to open:
To do a shutdown:
CAUTION A soft power shut down (CPU and Electronics and Turbo Pump) is highly
recommended before shutting off the main power switch (circuit breaker).
NOTE Do not turn off the main power switch until the power down status bar is complete
and final power off message is accepted (OK).
To do a hard shut down of the leak detector for 24 VDC power supply and primary
pump:
Move the power switch on the rear panel to the
° position.
Agilent Helium Leak Detector User Manual 37
Instructions for Use
Maintenance
WARNING Disconnect power from the leak detector before performing any
maintenance procedure that requires physically disconnecting any part of
the system.
The leak detector has two battery backed clocks which use a primary (non-
rechargeable) lithium coin battery from Panasonic, P/N BR2477A/HB (3.0V,
1Ah) that is permanently soldered onto a Motherboard PCB and Panasonic's
P/N BR2032 (3.0V, 500 mAh) located on the CPU PCB.. Never replace this
battery with a rechargeable type. Use only the primary type of lithium
battery. Only a qualified technician should replace any battery. The battery
has a life of 7 to 10 years.
The leak detector uses fuses and circuit breakers that are not operator
replaceable. Only a qualified technician should replace any fuse.
Information about the fuses and circuit breakers is given in Table 4.
Cleanliness is vital when servicing the leak detector or any vacuum equipment.
There are some techniques more important in leak detector servicing than in
general vacuum work:
NOTE Agilent does not recommend the use of vacuum grease. Vacuum grease absorbs
helium tracer gas and releases this slowly resulting in helium background during
leak test operations. If it must be used, use it sparingly and avoid silicone types.
Apiezon® L grease is recommended (Agilent Part No. 695400004).
CAUTION Remove O-rings carefully with your fingers. Do not use metal tools for this task.
This prevents scratching of any sealing surfaces.
Wipe all O-rings clean with a lint-free cloth before installation to ensure that no
foreign matter is present to impair the seal.
Do not use grease or any other substances on O-rings or metal seals that will
come in contact with the spectrometer.
Do not use alcohol, methanol or other solvents on O-rings as this causes
deterioration and reduces their ability to hold a vacuum.
NOTE If applicable, apply a small amount of Apiezon®L grease and wipe the
O-rings shiny dry.
NOTE Due to the effective cleaning nature of VacuSolv solvent and its residue-free
properties, Agilent’ Component and Spectrometer Cleaning Kit (Part Number
670029096), used in accordance with the kit instructions, is recommended for
cleaning spectrometer components. The kit can also be used for fine cleaning of
other parts in the leak detector’s vacuum system such as valves and fittings. No
rinsing steps or high-temperature drying are required following cleaning with
VacuSolv. Although appropriate precautions are advised, VacuSolv is compatible
with most materials and does not contain toxic chemicals or CFCs
(chlorofluorocarbons). Other acceptable solvents are isopropyl alcohol (IPA)
or Dow Corning® OS-20.
Calibration check X
Spectrometer cleaning The spectrometer is designed to operate for years without maintenance, however actual
performance depends on the leak detector operational environment. Spectrometer
cleaning restores optimum performance if the unit shows significant sensitivity loss or
high background signal.
Sensitivity loss appears as a high gain value after a calibration or failure to calibrate with a
gain too high error message.
Filament replacement Filaments are designed to operate for several years with no maintenance. When a
filament burns out, the system automatically switches to the other filament and notifies
you to recalibrate. The calibration notice also serves to inform you of a filament failure.
Agilent recommends that the burned out filament be replaced at a convenient time so
that a spare is always available during leak test operations to prevent unplanned system
down time.
Tuning adjustments If calibration is not successful or if the system is calibrated to an external leak, tuning can
maximize the helium signal.
Mechanical pump fluid change Persistent, high helium background signal. Fluid contamination (dirty brown color
indicates burnt or contaminated fluid, milky white consistency indicates high water
vapor content in fluid).
NOTE Always perform a Full Internal Calibration after any spectrometer maintenance:
• Spectrometer header replacement
• Filament replacement
• Ion source assembly replacement or cleaning.
NOTE Only a qualified technician should replace circuit breakers and fuses. Use only
circuit breakers and fuses of the same type and rating.
Daily Maintenance
Sensitivity Check
To perform a sensitivity check:
1 Turn on the Std Leak in the Control Panel screen.
2 Compare the value on the Home screen to the value of the known calibrated
leak on the Internal Calibrated Leak screen.
3 If the values do not match, press Calibrate on the Control Panel screen to
perform an automated calibration and then repeat step 1.
If specification is still not met, Manual Tuning may be required.
See Section “Manual Tuning.
NOTE The nitrogen supply line should have a pressure relief valve installed so that the
pressure does not exceed two psi.
1 Press down on top of fitting and remove filter (see the following figure).
2 Install 6mm tubing for Dry Nitrogen venting.
Disposal
Service
Agilent offers:
Rebuilt spectrometers on an exchange basis.
NIST-traceable calibrated leak testing and verification services.
Preventive maintenance services.
Overhaul services.
System recertification.
Support agreements.
On-site support.
2 Technical Information
Description of the Leak Detector 47
Leak Detector Configurations 49
Technical Specifications 50
Front Panel Display and Operations 52
Home Screen 53
Front Panel Display - Home Screen Display 55
Rear Panel 57
Features 61
Operating the Leak Detector 62
Screens 62
Basic Functions 62
Home Screen 63
Graphical Display Screen 69
Menus Screen 70
Calibration Screen 71
Calibration Set-Up 71
Test Configuration Screen 77
Test Configuration 77
Reject and Audio Set Points 77
Test Configuration Advanced Features 80
Transfer to Test 83
Stabilization Wait Timer 84
High Pressure Test Set-Up Procedure 84
Home Screen - Activate Gross Test 85
Application Setup 86
Sniff Method 87
Spray Method 89
High Sensitivity 91
Auto Sequencer 92
Auto Sequencer Set-Up 95
Rough Time Setup 95
Test Time Setup 96
Set Point 1 Setup 96
Set Point 2 Setup 96
Using Set Point 1 and 2 together (Optional) 96
Pressure Set point (High & Low Limit) - Optional 96
Split Flow 97
PPM (parts per million) Sniff Method 100
The leak detector is a wide-range Helium Mass Spectrometer Leak Detector, which
is comprised of a turbo molecular high vacuum pump, internal mechanical rough
pump (model G8610A), internal scroll pump (model G8610B), external mechanical
rough pump (models G8611A, G8611B, G8611C, G8612A, G8612B, G8612C and
G8612D), spectrometer tube, valve block, leak detector electronics, an operator
interface (TFT-Touch Panel display and optional features, including:
Wireless remote control that includes:
Wireless Base PCB (P/N G8600-68001) installed in a leak detector
Wireless Remote unit (P/N G8600-60002).
Optional I/O PCB (P/N VSFLDIO).
Models G8611Ax, G8611Cx, G8611Bx and G8612D include an optional four-wheel cart.
All models are available in 100/115 VAC and 230 VAC configurations (see the
following table).
The wireless remote control has the following capabilities:
RF communication between Agilent Technologies Helium Mass Spectrometer
leak detector (Base Station or Host) and the Wireless Remote unit from the
distance up to 100 m (300 ft).
Enter the leak detector unit into Test and Hold modes from the Wireless
Remote unit keypad.
Read the leak detector unit's leak rate in a linear and logarithmic bar graph
format with a numeric range indicator and System status in alphanumeric
format on the Wireless Remote unit Touch Panel display.
Configure up to ten different Agilent Technologies Helium Mass Spectrometer
leak detector units for wireless control by toggling through a MAC address list
using one Wireless Remote unit.
The optional I/O PCB compromises:
Optically isolated parallel I/O (DB25 female connector with 12 output and 8
input lines) to use with an external PLC and should be powered by the user
(+5-24 VDC) via a cable with length less than 10’ and provided by the user
(see the "I/O Connector" section).
The usage of a longer cable for the parallel I/O output connection with an
CAUTION external PLC is fully responsibility of the user. Some EMC prevention measure
may be required by the user.
USB type A connector and one located on the back of the leak detector.
Model Number Leak Detector Type Agilent P/N Mains Voltage 50/60 Hz (Amp)
G8611D
Mobile G8611-64006 230 VAC 4.4/2.8
(DS602)
G8612D
(DS602) Benchtop G8612-64006 230 VAC 4.4/2.8
Outline drawings for each configuration are shown in Section “Leak Detector
Configurations”.
Figure 17 Portable Leak Detector with Internal DS40M Wet Rotary Vane Pump, Model G8610A,
and Portable Leak Detector with Internal IDP-3 (G8610B) Dimensions
The following figure shows the outline drawing and physical dimensions of a
mobile leak detector with Models G8611x.
764mm 572mm
Figure 18 Mobile Leak Detector with External Wet Rotary Vane Pump or Dry Scroll Pump on a Cart,
Model G8611A, G8611B and G8611C Dimensions
Technical Specifications
Table 6 Leak Detector Series Specifications
Turbo Molecular Pump Navigator™ Turbo molecular pump specifically designed for leak detection
Helium Response Time 0.5 second, within 66% of peak reading within the given time. Per AVS 2.1.
Test Port NW-25; A 11/8" (28 mm) ID Compression Port is available as an accessory.
Installation Indoor use only. Altitude up to 2000 m (6,400’), Material Group III, POLLUTION DEGREE 2
according to EN61010-1 Standard, Class 1
Operating Temperature System: +12° C to +45° C (wet and dry pumps); Fore pump limited on the low ends and
turbo pump limited on the high ends.
Air Pressure 563 Torr (75 kPa) to 795 Torr (106 kPa)
Touch Panel display 8.4 inch color TFT LCD module G084SN03 V3.
The screen format is intended to support the SVGA (800(H) x 600(V)) screen and 16.2M
(RGB 8-bits) or 262k colors (RGB 6-bits). Viewing area 175.1 mm x 132.2 mm
Regulatory CAN/CSA-C22.2 No. 61010-1-12 - Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use - Part 1: General Requirements
UL Std. No. 61010-1, 3rd Edition - Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use - Part 1: General Requirements
EMC Emissions and Immunity -ETSI EN 301 489-1 v2.2.0 (2017-03) Electromagnetic
Compatibility and Radio Spectrum Matters (ERM); Electromagnetic Comparability (EMC)
standard for radio equipment and services; Part1: Common technical requirements:
• ETSI EN 301489 v2.2.1 (2012-09) Electromagnetic Compatibility and Radio Spectrum
Matters (ERM); Electromagnetic Comparability (EMC) standard for radio equipment;
Part 17: Broadband Data Transmission Systems
• EN 61326-1: 2013 EMC requirements for Electrical equipment for measurement, control
and laboratory use - General Use
For particular Safety and EMC Standards see Declaration of Conformity.
Maximum spectrometer 1E-04 Torr - Filament turns off when pressure exceeds this point
pressure
All operator control and monitoring are achieved via the resident software accessed
using the Touch Panel Display (Table 7).
Touch Panel Display
Represents:
Leak Rate - in a Bar Graph format with a numeric range indicator
Leak Rate, System Status and Parameters - in alphanumeric format
Series of screens with the soft key buttons for operator interface
Leak Rate, System Status and Parameters in alphanumeric format:
Leak Rate is presented in a scientific notation: XXE-YY with appropriate units
Where: X .X mantissa is a leak rate value E-Y = 10-YY exponent is a leak rate
range X and Y are any numbers from 0 to 9
System Status and/or Mode are represented on the screen in alpha format on
the screen and an italic font in this manual
Parameters are represented on the screen in alphanumeric format a bold font
in this manual.
Leak Rate Bar Graph
The large bar graph displays the leak rate in one of three forms:
A linear bar graph
A logarithmic bar graph
Home Screen
VENT The VENT button advances the leak detector to the vented state.
The red vent symbol in the middle of the switch remains red while
the leak detector is in Vent.
TEST When the system is in the Vent or Hold states, pressing the TEST
button automatically advances the leak detector through roughing,
then through the various test states, depending on the achievable
test port pressures and leak rates.
Press the TEST button while in TEST to isolate the test port from
the leak detector vacuum system.
HOLD With the leak detector in Test/Fine Test mode, press TEST again
and the leak detector goes into HOLD mode, isolating the leak
detector from the unit under test.
The leak detector operator interface functions through a series of screens, which
are navigated as in the following figure:
Screen Purpose
Maintenance
• General - Configures:
Internal Calibrated Leak Setup & Expiration
Software Update Utility
• Manual Tuning - Configures:
Toggles between Filament 1 & 2
Sets Emission, Ion and Gain settings
• Valve Control - Configures:
Access to control and monitor valves
Monitor turbo pump parameters
• Pump - Configures:
Fore Pump description type, operating hours and Last Service Date
Enable Exhaust Pump
• Gauge Calibration - Configures:
Calibrates the Test Port TC's Vacuum and ATM reference points
Displays Test port and spectrometer pressures
• System Defaults - Configures:
Ability to reset Turbo and Filament hours for maintenance purposes
Ability to reset back to Factory Defaults
System Information:
• Displays part number and serial number
• Display Software Version
• Displays the status of the spectrometer and turbo pump parameters
• Current Setup
• Displays the status of active items
Rear Panel
The rear panel of the leak detector with the wireless remote control option is shown
in the following figure.
The rear panel of the leak detector (Model G8610A and G8610B) is shown in the
following figure.
I/O Connector is a 25-pin female D-sub connector to interface with an external PLC
via a parallel cable.
Inputs:
Four level inputs: optically isolated, (+5 to+24) VDC.
Four momentary inputs: optically isolated, (+5 to+24) VDC, requires 200 ms
pulse width.
All inputs are opto-isolator LEDs with a series 3300 Ohm resistor.
Outputs: Eight optically isolated emitter followers with a 10 Ohm series
resistor, 14 mA max drive current, +24VDC max.
Power: An external power supply is supplied by the user.
Features
The features for the leak detector are provided in the following table:
Table 8 Features
Feature Description
Calibration Fully automated tuning and calibration using the internal or external calibrated
leak. The fast calibrate feature allows calibration in less than one minute.
Zero Control Selectable mode background elimination with a unique, patented Autozero<0
feature.
Leak Indication A bar graph (linear or log) indicator displays leak rates automatically; alphanumeric
indication on Touch Panel display. A leak can also trigger an audible alarm
frequency that varies in proportion to the size of the leak, dedicated volume control
and programmable audio threshold. Analog voltage is proportional to the leak rate.
Pressure Indication Test port and spectrometer pressure with alphanumeric indication on the Touch
Panel display.
Spectrometer Sensitivity optimized design, header mounted ion source with dual Yitria-coated
iridium filament, a high sensitivity, wide dynamic range preamplifier, and vacuum
gauge.
Automatic Valve Sequencing Microprocessor-controlled valve operation provides automatic system cycling.
Safety Interlocks A valve system design that completely protects the high vacuum system under
test against pressure surge and power interruption.
Optional
• The ability to read the leak detector’s leak rate in a linear and logarithmic bar
graph format with a numeric range indicator and System status in
alphanumeric format on the Wireless Remote unit's Touch Panel display.
• Configuration up to 10 different leak detector unit’s for wireless control by
toggling through a MAC address list using one Wireless Remote unit
Discrete I/O Control This provides 12 outputs and 8 inputs lines, optically isolated, to control the leak
detector using an external PLC via a 25-pin DB25 female connector and a user-
provided parallel cable.
Screens
The leak detector display is used for the initial set-up and configuration of the leak
detector. Once the leak detector is set up and configured for a specific application,
basic operation is controlled primarily using the TEST and VENT buttons on the leak
detector and the Home screen Zero button (Section “Home Screen”).
Basic Functions
The following table explains the icons that appear on various screens and their
meaning.
Icon Meaning
Indicates a feature is locked and cannot be activated due to a higher level selection
in the setup.
Indicates the status of the internal calibrated leak: If flashing, the leak calibration
will expire within 10 days. Steady and blue indicates that the leak has expired.
Steady and red indicates a temperature sensor failure or a disconnected cable to
the calibrated leak temperature sensor.
Indicates that the split flow leak rate being entered is greater than 1000 times the
indicated leak rate; the maximum permitted.
Home Screen
The Home screen displays a summary of the following status and configuration settings:
Digital Leak Rate Display
Test Port Pressure
Gross Test button (not shown)
Leak Detector Status
Condition Indication
Reject Status Indicato
System Ready Indicator
Menus button
The following figure shows the screen controls.
As of February 2023
(LD01.06) the Calibration
button can be enabled on
the Home Screen from
the Calibration Menu
Hold Places the system in Hold (while in Test, press Test again to
starts Hold); keeps the test port under vacuum and does not
vent. Use Hold to keep the system under vacuum, while taking
the leak detector out of test. This is useful if you want to let the
leak detector clean up the helium background, while keeping the
system under vacuum.
Since the test port pressure gauge is located on the test port
side of the isolation valve, a subsequent rise in test pressure
can be observed while in this mode.
Digital Leak Rate Correlates directly with the log and linear bar graph leak rate
display. The units for the measurements are selected in the
Settings > Main Settings menu (see Section “Settings Menu”).
Test Port Pressure Appears on the Home screen and represents the test port
pressure as measured by a thermocouple device mounted on
the valve block just below the test port. The pressure transducer
and associated electronics are intended to protect the leak
detector from damage due to over-pressure conditions. Use an
external pressure transducer if the test process requires a more
accurate monitoring of the test object pressure.
Leak Detector Status Appears below the test port pressure reading and indicates the
current leak detector operating state. The normal operating
states are detailed in the Table 11.
Leak Detector Appears below the Leak Detector Status and indicates the
Condition current leak detector condition during abnormal conditions. See
the Table 11 for a list of leak detector conditions.
Setpoints Status Appears below the Leak Detector Condition on the left-hand
Box center of the Home screen when any of the four independent
set points are enabled and activated or de-activated.
Zero Zeroes out the helium leak indicated on the display. The Zero
button darkens while zeroing is taking place. The leak rate must
be less than 4.0E-7, while in Test, to allow a cross over to Fine
Test.
The Zero button is only active while the leak detector is in the
Test and Fine Test states of operation.
The leak detector automatically subtracts the recorded
background signal to display only the real leak rate.
NOTE Pressing Zero while introducing helium tracer gas to the test object or while the
helium background is varying significantly could result in the suppression of a real
leak. Only use the Zero button after the helium tracer gas source has been removed.
The Zero capability allows the testing of sensitivity levels that are below the system
background. This feature also reduces test cycle times by reducing the amount of
time required for background levels to naturally clean up or by allowing testing to
begin at relatively high test pressures. It is recommended that only two decades of
signal be zeroed out.
Zero Enable Turns on or off the zeroed out background. This feature
excludes the background leak from the leak rate signal.
Helium background represents helium that is in the system
but does not come from the part under test. If the
background is zeroed out, a more accurate measurement
of the leak rate of the part is determined. If Zero Enable is
turned ON, the background is excluded; if it is turned OFF,
the system displays the full helium signal, including total
background
Display Description
Calibrating Indicates that the leak detector is currently performing a calibration routine.
Indicates that the leak detector is in Test mode. When in Test mode, the Test Port valve opens
Test and the leak detector is ready for leak testing. Test leak rate range is 1E-4 to 1E-8 (MDL 5E-9).
Fine Test Indicates that the leak detector is in Fine Test mode. When in Fine Test mode, the test port
valve opens and the leak detector is ready for fine leak testing. Fine Test leak rate range is 9E-7
to 1E-11 (MDL 5E-12).
Hold Indicates that the leak detector is in Hold mode. When in Hold mode, all but the foreline valves
are closed, isolating the test port and any test objects or fixturing from the leak detector
vacuum system. Use Hold when you want to keep the test port under vacuum without venting
the test volume; this helps the system clean up.
Roughing Indicates that the leak detector is currently performing a Roughing level evacuation of the test
port and any attached test objects or fixturing attached to the test port.
Std Leak Indicates that the optional internal calibrated leak standard valve is open and that the leak
detector is reading the leak value.
Vented Indicates that the test port is vented to atmosphere. The test port and all objects attached to
the test port are isolated from the leak detector internal vacuum system.
Display Description
Backing Indicates that the leak detector is momentarily backing the foreline of the high vacuum pump
during an extended rough period. Backing occurs if the roughing time is greater than 60
minutes or if the turbo pump is not ready.
Both Filaments/ No Indicates that no filament emission is detected. This can result from a PCB or cable problem,
Emission or burned out filaments.
Cal OK Indicates that the calibration routine was executed successfully.
Calibration Prep Indicates that the leak detector is preparing for a calibration routine.
Filament 1 No Emission Indicates that filament 1 of the ion source has burnt out. Filament 2 automatically lights if it is
still good, and a flashing C appears in the leak rate exponent, indicating that a calibration
routine is required.
Filament 2 No Emission Indicates that filament 2 of the ion source has burnt out. Filament 1 automatically lights if it is
still good, and a flashing C appears in the leak rate exponent, indicating that a calibration
routine is required.
Filament Wait Indicates that the system is lighting the filament.
Display Description
Gain Too High Indicates that the gain value required to calibrate the leak detector during the calibration
routine is greater than the maximum allowable value. This is generally the result of system
sensitivity being too low, due to normal system aging.
Gain Too Low Indicates that the gain value required to calibrate the leak detector during the calibration
routine is less than the minimum allowable value. This is generally the result of a system leak.
High Background Indicates a high Helium background detected during calibration self-test. This typically,
indicates a vacuum system leak.
No Ion Peak Indicates a tuning/calibration failure due to the system not detecting an ion peak during the
auto-tuning routine.
Offset Fault Indicates an electronics failure identified during calibration self-test.
Offset Wait Indicates that the system is setting the Offset value.
Reject Indicates a test object failure during an Auto Sequencer Test cycle. See “Auto Sequencer”.
Spec-Tube Flashes red when the spectrometer pressure > 1.0E-4 Torr.
Stabilization Wait Indicates that the system is waiting for the electronics to stabilize before completing the
startup routine; up to two and a half minutes.
Std Leak Prep Indicates that the optional internal calibrated leak is being roughed before it is introduced
directly to the system for calibration verification.
Switching Filaments Indicates that the system is momentarily in the process of switching from a burnt out filament
to the next available filament.
System Pressure Wait Indicates that the system (spectrometer) pressure is too high to light the ion source filament.
System Ready Indicates that the system is ready for testing. This message appears only during the initial
startup and is cleared after the first test cycle.
Zeroing Indicates that the leak detector is in the process of Zeroing out a background signal. The Zero
routine is initiated automatically during the startup and calibration routines, or manually when
the operator presses the Zero button on the Home screen.
Zeroing Wait Indicates that the system is setting the initial zero values during the startup routine.
Audio Speaker The audio signal increases from a low to high frequency as
Function the leak rate increases within each decade. The pitch cycles
low to high as it passes through each decade. The audio
response corresponds directly to the size of the leak.
Speaker Volume To turn the sound on/off, press the speaker button. The
volume level is restored to the level it was at prior to being
turned off.
Timer Toggles the graphing time between 120 and 720 seconds.
The timer automatically defaults to 120 seconds on leak
detector start up. Press the toggle button while graphing to
clear the current graph. The data rate for the timers are:
120 seconds - two measurements/second
720 seconds - one measurement/three seconds.
As of February 2023 (Rev. LD01.06) extended the time
base from 720 seconds (12 min) to 3600 seconds
(60 min) with a 0.3 second resolution.
NOTE Pressing Zero while introducing helium tracer gas to the test object, or while the
helium background is varying significantly, could result in the suppression of a real
leak. Only use the Zero button after the helium tracer gas source has been removed
The Zero capability allows the testing of sensitivity levels that are below the system
background. This feature also reduces test cycle times by reducing the amount of
time required for background levels to naturally clean up or by allowing testing to
begin at relatively high test pressures. It is recommended to zero out only two
decades of signal.
Menus Screen
Press Menus on the Home screen. The Menus screen appears (see the following
figure). Use the arrow to navigate to the previous screen.
The Menus screen consists of buttons to access the lower level screens used for
application-specific configuration.
Press a
button to
access specific
configurations
Calibration Screen
Calibration Set-Up
Press Calibration Menu screen.
The Calibration screen appears (see the following figure). Use this screen to:
View the value and expiration date of the Internal Calibrated Leak
Enter the External Calibrated leak value
Select whether to use an Internal or external calibrated leak
Set the calibration mode: Full or Fast Calibration
The calibration routine can only be executed while in Test/FineTest.
NOTE Factory default – Show Home Cal & Show Cal Pop (not selected/not enabled).
Calibration Status Messages – Cal Ok & Cal Failed pop up messages can be
enabled on the Home Screen (customer selectable) from the Calibration Menu.
NOTE The internal leak is a factory supplied helium leak internal to the leak detector.
In order to perform an external calibration, you must supply a separate calibrated
leak and mount it on top of the leak detector.
NOTE The following figure appears within the first twenty minutes of operation if calibrate
is selected. This message advises you that a twenty minute warm up is strongly
recommended after a cold start up. This twenty minute warm up ensures optimum
instrument calibration and peak performance.
• Select No - No calibration is performed
• Select Yes - Calibration begins.
• No action - Keeps the message displayed and the leak detector automatically
calibrates after the 20 minute timer times out.
Read Internal Leak Opens the Internal leak valve so the leak can be seen by the
system.
The Read Internal Leak button is only functional while the
leak detector is in Test/Fine test and activates the internal
calibrated leak, allowing verification of the system
calibration. When the system is in Std Leak mode, Std Leak
appears on the Home screen.
The Std Leak mode isolates the test port from the system
and exposes the leak to the spectrometer. This allows you
to compare the displayed leak rate value to the calibrated
leak value on the Calibration Menu screen to verify proper
system operation.
Because the test port is isolated from the fore pump in this
mode, a slight increase in test port pressure may be
indicated on the test port pressure display.
Full Calibration Performs a thorough tuning process and a system gain
(Internal Only) adjustment (calibration). The full tuning process involves
scanning the ion source chamber voltage, and optimizing
each value for a maxi- mum helium signal. Once the tuning
process is complete, system gain is adjusted to bring the
leak detector into calibration. A full calibration takes less
than three minutes.
Fast Calibration Compares the leak rate signal with the calibrated leak value
and adjusts the gain to bring the system into calibration. If
the gain value is above the upper gain limit for the working
range as defined in the following table, then the fast
calibration process is terminated and a flashing c symbol
appears on the Home screen. A fast calibration takes less
than 30 seconds.
* If the system cannot read the internal calibrated leak temperature, the system
displays the leak value and the temperature shown on the calibrated leak label. If
this happens, the calibrated leak symbol on the main screen turns red.
NOTE The firmware initially starts at a large time interval, but recalculates the required
time as the pumping continues. The new time is reflected on the timer.
The new Applications Setup Menu (Test Confutations Menu) helps guides you
through the process - Select High Sensitivity.
Test Configuration
The Test Configuration Menu allows you to:
Program and activate process setpoints (leak rate pressure and audio)
Enable or disable leak detector advanced features
Select Leak Detection specific applications with auto parameters setup with
step-by -step instructions for enhanced operation.
Toggles these
options on/off.
Activates the
reject
the programmed
set point if above
value
Reject Set Points Activate and non-activate set points appear on the
Home screen inside the Set Point Status window.
The status of the set points are also available through
the RS-232 line (see Appendix A “Communications
Protocol”). There are four possible set points. Three set
points can be programmed to a leak rate value and one
can be programmed to test port pressure.
Set points may be turned ON or OFF (enabled or
disabled) for each set point. If a Reject Set Point is ON,
then the rear panel output for that set point goes active
when its leak rate value is exceeded and an activated
red box indication appears on the Home screen.
Each set point is set to activate above or below the
programmed value. To activate the reject set point:
Above the programmed value, select the >.
Below the programmed value, select the <.
NOTE For Plugged Probe Alarm - High Pressure Test (Sniffing Mode) see Section “Sniff
Method”.
Off:
The audio sounds for all leak rates above zero.
Set points may be turned ON or OFF (enabled or disabled). When the Audio Set
Point is OFF, audio sounds for all leak rates above zero. When the audio set
point is ON, audio sounds only for leak rates above the audio set point.
High Pressure Test Select the High Pressure Test mode. When High
Pressure Test is selected, Range Stop defaults to 10-7
atm cc/sec. High Pressure Test sensitivity mode range
is 10-4 to 10-8 atm cc/sec.
When the leak detector is in Vent mode:
Turn the High Pressure Test feature on to start the
test cycle. This automatically locks the leak
detector into Contra-Flow™ Leak mode, and auto-
ranges through all decades down to the most
sensitive Contra-Flow range.
When the leak detector is in Test mode:
Turn the High Pressure Test feature on to perform
the same function, with the exception of
performing the test cycle start sequence.
Fine Test Only Select the Fine Test Only switch to enable the feature.
When selected and Test is initiated on the Home
Screen the leak detector transfers from Test to Fine
Test and remains in Fine Test regardless of leak rate. If
the test port pressure is greater than 1.4E-1 Torr the
leak detector transfers back to the Roughing Mode.
NOTE For High Pressure Test/Sniffing mode, de-activation (Inactive AZ<0) is highly
recommended. Manual Zeroing produces a more desirable performance, thus
eliminating false leaks due to fluctuations in test port pressure while in test.
Gross Test Allows the operator to locate leaks while the system is
in roughing mode. The test indicates the presence of a
leak at high pressures only and does not display a leak
rate value.
Gross Test is activated in the Test Configurations > Advanced Features menu.
Once activated, the GROSS TEST button appears on the Home screen. Press
GROSS TEST to enter the screen.
There is an audio tone for Gross Leak testing, which increases as the leak
increases.
Gross Test only works when the test port pressure is between 10 and 150 Torr
and remains constant (± 2% between previous and current pressure readings).
The Gross Leak test stops if the system goes into Backing, Test or Fine TEST
mode.
Activating the Auto Sequencer button terminates the Gross Leak test.
The HOLD button is inactive while in Gross Leak test mode, as the system is in
roughing mode.
Transfer to Test
The value input and hidden keypad is to the right of the Transfer to Test feature.
Use it to set the pressure at which the system goes into Test.
The Transfer to Test setting can be adjusted to lower values for specific
applications. The limit on the maximum test pressure prevents the operator from
inadvertently setting the leak detector to transfer into test at too high a pressure,
which would result in excessive spectrometer contamination or ion source filament
failure.
Press the
value to access
its keypad
Maximum Test Determines the test pressure cross over point at which
Pressure the system transfers from Roughing into Test and then
into the Contra-Flow Test mode. In the Contra-Flow
Test mode, the gas sample is pumped through the leak
detector and swept across the high vacuum pump
foreline. The Contra-Flow mode allows for higher sensi-
tivity testing at exceptionally high test pressures.
The maximum test port pressure is 10 Torr.
NOTE The transfer pressure from the Test mode to the Fine Test mode is set at 100
milliTorr (1.0E-1 Torr). You may wish to lower that value under certain
circumstances. The mid-stage transfer pressure limit is adjustable from 1 to 100
milliTorr via an RS-232 command (see ML-XFER and INIT-ML-XFER in Section A
“Communications Protocol”). The pressure limit is not adjustable using the display.
Press to
activate
Gross Test
Gross Test Allows the operator to locate leaks while the system is in
roughing mode. The test indicates the presence of a leak at
high pressures only and does not display a leak rate value.
1 Activate Gross Test. The GROSS TEST button appears on
the Home screen when the test port pressure is between
10 and 150 Torr.
2 Press GROSS TEST to enter the screen.
There is an audio tone for Gross Leak testing, which
increases as the leak increases.
Gross Test only works when the test port pressure is
between 10 and 150 Torr and remains constant (± 2%
between previous and current pressure readings).
The Gross Leak test stops if the system goes into
Backing, Test or Fine TEST mode.
Activating the Auto Sequencer button terminates the
Gross Leak test.
The HOLD button is inactive while in Gross Leak test
mode, as the system is in roughing mode.
Application Setup
Sniff Method
Utilizing the Agilent Power Probe connected to the test port of the Helium Leak
Detector, the sniff method pinpoints the location and magnitude of leaks from
volumes that have been filled with helium. Settings available in this mode include:
Establishing the minimum displayed range - avoids the distraction of
inconsequential leaks.
Entering set point values for audible and visual alarms.
Adjusting the sampling pressure for best response and clean up and
establishing an alarm should the sniffer probe become plugged.
To access:
1 Select Test Configurations > Applications Setup.
2 Select Sniff Method and press Set-Up and the following figure appears:
At this point:
The leak detector locks into High Pressure Test (Test only)
Auto Zero < 0 is disabled
Minimum Range defaults to 10-7 - This value can be adjusted from 10-8 to 10-4
Buttons and Fields
Set Point 1 Desired Alarmed Leak Rate value (> <) - This value can be
adjusted from 10-8 to 10-4. Set Point values in Sniffing
applications are usually programmed for greater than the
acceptable leak rate limits.
Audio Set Point Desired Audio Alarmed Leak Rate Value (> <). This
value can be adjusted from 10-8 to 10-4. The Audio Set
Point triggers with its' own distinct tone if different
from Set Point 1.
NOTE The Audio Setpoint is very useful in the Sniffing Application. The programmed Audio
Set Point value can be programmed 50% < Set Point 1, which alarms the operator
that they are approaching a leak (leak location).
Plugged Probe Alarm A distinct high frequency tone different from the Audio
& Set Point.
1 This alarm is useful in alerting the operator that the
desired Power Probe/Sniffing Probe adjustment
(pressure) has changed. A High Limit can be
selected in alerting that too much flow (high test
port pressure) is present. A Low Limit trigger is
typically the norm indicating that the Power
Probe/Sniffing is blocked or the tubing is kinked or
pinched off.
Spray Method
The spray method is for leak testing vacuum systems of all sizes. When properly
connected to the vacuum system under test, the Helium Leak Detector will locate
and indicate the relative size of leaks. Settings available in this mode include:
Entering set point values for audible and visual alarms.
Connecting the wireless remote - allows testing of large systems by one operator.
To access:
1 Select Test Configurations > Applications Setup.
2 Select Spray Method and press Set-Up and the following figure appears:
NOTE The Wireless Remote is a factory and field installed customer option.
NOTE The external calibration routine for the Spray Method is typically performed on large
chamber applications for proper characterization of the leak rate measurement.
High Sensitivity
Often, calibration to a known, external helium leak standard is required for internal
quality requirements or additional accuracy. High Sensitivity Test guides you
through this procedure. Settings available and features in this mode include:
Establishing the minimum displayed range - avoids the distraction of
inconsequential leaks.
Step-by-step guidance on the external calibration procedure.
Employs additional software routines to improve stability at low leak rates.
To access:
1 Select Test Configurations > Applications Setup.
2 Select High Sensitivity and press Set-Up and the following figure appears:
To perform a calibration process (internal and external) for a precision 10-10 leak
rate accuracy and stability:
NOTE Minimum Range can be programmed - Default 10-10. Auto Zero < 0 is disabled.
Auto Sequencer
The Agilent Auto Sequencer facilitates programming a repetitive test cycle for leak
testing parts that have been sealed with helium inside. This can improve
productivity and reduce testing errors. Settings available in this mode include:
Program timers for the roughing and test modes.
Entering set point values for audible and visual alarms.
Establishing a lower level verification leak to improve reliability of the test.
Setting a pressure set point that saves time by rejecting a part if the pump
down characteristics fall outside of the norm.
The auto sequencer is used to automatically perform testing on a part, to maximize
testing throughput and to reduce the chances of accidentally accepting a bad part.
The auto sequencer is used in conjunction with the reject and audio set point
settings, which provide the criteria for pass or fail conditions. Within the auto
sequencer feature, there are two variables that must be set, rough time and test
time. The ROUGH TIME setting indicates to the system the maximum time it is
allowed to be in roughing before the test fails. Ideally, the system goes into the
roughing phase and then proceeds to TEST. The TEST TIME setting tells the
system when to perform the test against the set point values. For example, you
may set 20 seconds for the TEST TIME. At the end of 20 seconds, the system
determines if the part has passed or not and then makes an indication by either
going to VENT (Pass) or going to HOLD (Fail) and displaying the reject set point
failure on the front panel
Press Auto Sequencer Set-Up on the Menus screen.
The Auto Sequencer Set-Up screen appears (see the following figure). The leak
detector can be configured to automatically run through a complete test cycle
(Start/Rough/Test/Vent).
During the test cycle, the system ensures that during the Rough Time the user-
entered pressure transfer set points are met.
A Pass condition is indicated by the system reverting to the Vent mode upon
completion of the test cycle. A Fail condition is indicated by the system
reverting to the Hold mode, and all operating buttons being disabled except
Vent. A failed condition also results in a red activated set point box in the Set
Point window appearing on the Home screen, along with a reject set point
activation upon completion of the test cycle. The status of the reject set points
are available through the RS-232 line (see Appendix A “Communications
Protocol”). On completion of an automated test cycle, the leak detector
captures the leak rate signal measured just prior to the end of the automated
test cycle. This leak rate appears on the Home screen until the next test cycle is
initiated.
Rough Determines the amount of time that the leak detector is allowed to
Time evacuate the test object to the Maximum Test Pressure. In Auto
Sequencer mode, the leak detector transfers into test as soon as
the specified transfer pressure is met. If the specified test transfer
pressure is not achieved within the pre-selected rough time, the
system aborts the cycle and goes into Hold mode.
The Rough Time value is determined empirically and reflects the
typical time required to rough the test object to the specified test
pressure. Failure to achieve the test pressure within the pre-
selected rough time indicates a gross leak in the test object or test
fixturing, or that the time allowed was insufficient.
The rough timer records the amount of time the system spent in
the roughing phase prior to it entering the TEST mode. The rough
timer is intended to help you set up a rough time value since it can
be difficult to predict accurately how long the system is in the
roughing phase when you cycle. Put the system into TEST mode,
the rough timer records and displays the rough time so that you
can enter a rough time value close to what was recorded. For
example, if the rough timer records a value of 12 seconds, you can
then enter a value in the rough time that is slightly longer than
what was recorded, 15 seconds for example. The rough timer only
records a time when the auto sequencer is ON.
Test Time Determines the amount of time that the leak detector remains in
test before it compares the measured leak rate with the selected
reject set point. Allow sufficient time for the leak rate reading to
stabilize before the system makes a pass/fail decision. This is a
function of the part configuration and the leak rate specification. In
Auto Sequencer mode, the leak detector remains in test for the
entire duration of the preselected test time.
To configure the auto sequencer mode:
1 Press the blue check to turn on the Auto Sequencer.
The pass/fail criterion is based on the reject set points that are enabled.
2 Press the value button associated with the Rough time.
The data entry pad appears.
3 Enter the time and press OK.
4 Press the value button associated with the Test time.
The data entry pad appears.
5 Enter the time and press OK.
Press TEST and let the system record the actual time spent in roughing. After the test
is complete, use the recorded rough timer value to set an appropriate rough time.
To access:
1 Select Test Configurations > Applications Setup.
2 Select Auto Sequencer and press Set-Up and the following figure appears:
Test Time This setting tells the system when to perform the test against the
set point values. For example, you may set 20 seconds for the
Test Time. At the end of 20 seconds, the system determines if
the part has passed or not and then makes an indication by
either going to VENT (Pass) or going to HOLD (Fail) and
displaying the reject set point failure on the Home screen.
The final leak rate value appears on the Home screen regardless
of a pass or failed criteria. The next test cycle (pressing Test)
clears the leak rate value (setup below).
NOTE The Test, Vent and Zero functions are available in this application setup. The Test
and Vent buttons aid in setting up the Rough and Test Timers. The Zero function is
extremely helpful in zeroing out unwanted background associated with the test
fixture. The zeroing set up helps in reducing the test time.
NOTE A Typical Auto Sequencer application uses either Set Point 1 or Setpoint 2 for the
desired leak rate reject set point.
NOTE If the leak rate falls below the reference leak value (Set Point 2) at the end of the test cycle,
the leak detector reports Rejected/Hold. This type of a reject is a system level failure. This
could be related to a fixturing issue, operator error or system performance.
Split Flow
Activating the split flow option properly characterizes the system to an external
calibrated leak, which is influenced by additional vacuum pumping sources. These
separate vacuum pumping sources reduce the amount of helium measured by the
leak detector. The split flow option is very useful for external calibrations were
external gain factors exceed the limit of 50 due to a high helium background (Range
Stop E-8). The split flow factor is also useful for calibrating to a standard leak while
in the sniffing mode (High Pressure Test).
Leak detectors are frequently connected to vacuum systems in parallel with other
vacuum pumps. This can cause errors in readings when testing to a specific value
is important. Split Flow properly compensates for flow differences caused by
additional pumps. Settings available in this mode include:
Setting the value of the installed reference leak.
Step-by-step guidance in setting the split flow value.
Entering set point values for audible and visual alarms.
Enabling an Extended Split Flow routine when testing with low concentrations
of helium and/or to quantify large leak values.
Ground rules and assumptions:
The software limits split flow to no more than 1000 times (3 decades). If the
split flow is turned ON and the leak rate entered is more than 1000 times higher
than the system leak rate, the system will not allow the split flow to turn ON
and <> appears.
At the main screen, the system blinks the split flow symbol when split flow
is ON.
When split flow is turned ON, it displays the corrected leak rate but it knows the
real leak rate value.
When ZERO is pressed with when split flow ON, the system zeros out
background up to a limit equal to the maximum split flow value.
To activate split flow:
1 Place the system into TEST mode and let the leak rate stabilize. Calibrate the
system.
2 Press the Split Flow blue check button. The leak detector prompts you to enter
the leak rate for the customer's calibrated standard leak (what the leak rate
should be).
The leak detector displays the corrected leak rate, essentially calculating a split
flow gain value. The leak detector automatically calculates the split flow factor,
turns on and calculates the range stop, based on the split flow factor.
When Split Flow is turned OFF, the system reverts back to the uncorrected leak
rate. This process must be repeated when reactivating split flow.
To setup:
1 Connect the leak detector and secondary rough pump to the test chamber as
seen in the previous figure.
2 Install the External Calibrated Leak on the test chamber.
3 Press Test on the leak detector.
4 Open valve on secondary rough pump to evacuate test chamber.
5 Select Test Configurations > Applications Setup and the following figure
appears.
6 Select Split Flow and press Setup and the following figure appears:
The software limits split flow to no more than 1000 times (3 decades). If the split
NOTE
flow is turned ON and the leak rate entered is more than 1000 times higher than the
system leak rate, the system will not allow the split flow to turn ON and <> appears.
At the main screen, the system blinks the split flow symbol when split flow
is ON .
When split flow is turned ON, it displays the corrected leak rate but it knows the real
leak rate value. When ZERO is pressed with when split flow ON, the system zeroes
out background up to a limit equal to the maximum split flow value.
NOTE Calibrate the leak detector prior to the PPM Sniff Application.
To run this:
1 Select Test Configurations > Applications Setup.
2 Select PPM Sniff Method, press Set-Up and the following figure appears:
3 Ensure that the leak detector must be in Test and setup for a sniffing
application as shown in the following figure. The test port pressure appears in
the top left hand corner of the setup screen (see previous figure).
4 If the leak detector is vented - Press the Test button in the bottom right hand
corner. Adjust the metering valve on the Sniffing Probe until a desired value
between 5.0E-1 and 1 Torr is obtained.
In this Setup Screen the leak detector can be configured to measure PPM
concentrations of Helium from the following three methods:
Calibrating to a known Helium PPM concentration (“Calibrating to a Known
Helium PPM Concentration”).
Monitoring the actual flow rate in real time from a recommended Mass Flow
Meter manufacturer (Alicat Scientific) via the RS232 port to obtain the
calculated Helium PPM concentration at the adjusted test port pressure
(“Measuring PPM of Helium with an Alicat Scientific Brand Mass Flow Meter”).
Inputting the flow rate (sccm) from any Mass Flow Meter or Mass Flow
Controller via the displayed flow rate reading/value at the adjusted test
port pressure to obtain the calculated Helium PPM concentration
(see “Measuring PPM of Helium with a non-Alicat Scientific brand Mass
Flow Meter”)
NOTE Parts per million (PPM) of helium may be measured using a helium leak detector
that is either used in conjunction with a mass flow meter, or is calibrated using a
known concentration (PPM) of helium. PPM of helium is described as the flow rate
of helium divided by the flow rate of total gas, multiplied by 100. If there are 5 moles
of helium, and 999,995 moles of other gas, then the concentration is 5 PPM helium.
This is a typical concentration of helium at sea level on earth.
NOTE This requires calibrating using a known PPM mixture, such as a 50 PPM helium
mixture. To sample the gas concentration, the operator may use either the standard
power probe sniffing gun (Agilent part number K9565306), or any small ID SST
or copper tubing attached to Swagelok fine metering valve (Swagelok
part number: SS-SS4).
To do this:
1 Select Test Configurations > Applications Setup.
2 Select PPM Sniff Method > Application Setup.
3 Ensure the system is in Test and the test port pressure adjusted between
5E-1 Torr 1 Torr.
4 Enter the Helium concentration in PPM in the associated field.
5 Press the Calibrate button (see Figure 46) and follow the step-by-step
instructions. The leak detector is now ready to measure the concentration
of Helium in PPM.
Measuring PPM of Helium with an Alicat Scientific Brand Mass Flow Meter
Scientific confidence added through accurate computation of both helium leak
rate, and gas flow rate.
Slower response time than without using mass flow meter at all.
Requires Alicat Scientific 1 SLM Whisper series mass flow meter.
• Alicat Scientific part number MW-1SLPM-D-DB15/5M, GAS: Air. Comes with
DC power supply & RS232 cable.
Requires a mid E-5 std. cc/sec He calibrated standard leak.
Application Setup
1 Assemble a fine metering valve, as described in Figure 49 and follow the
assembly notes.
2 Connect the fine metering valve in series with the leak detector (see the
following figure).
Item A. (2X) Description: NW25 to 1/4 NPT stainless steel male pipe
adapter Make: Agilent Technologies Inc. - Agilent part
number: KAMP250250S
Item B. (2X) Description: 1/4 NPT to 1/4" tubing adapter
Make: Swagelok - Swagelok part number: SS-4-TA-1-4
Item C. (1X) Description: 316 stainless steel very fine metering valve,
1/4" swage, 0.004 max
Make: Swagelok - Swagelok part number: SS-SS4
Assembly notes:
Apply Teflon pipe tape onto threads of item B to ensure proper seal with
minimal leakage.
When connecting item B. to item C., push in 1/4" pipe as far into nut as can be
allowed. Then snug nut to finger tight torque. Using a 9/16" wrench, turn nut 1 and
1/4 turns (360° + 90° = 450° total) for proper crush seal with the ferrule inside.
This fine metering valve is the same type of valve as used in the Agilent sniffing
power probe. If desired, that valve may be taken out of the power probe and
reassembled in a similar technique to the one described here.
NOTE The Calibration button on the PPM Sniff application is only to calibrate using a
known PPM mixture.
The leak detector is now ready to measure the concentration of Helium in PPM.
Figure 51 Alicat Scientific Mass Flow Meter Supported Method of PPM Testing
The hose may be connected to the inlet side of the mass flow meter. One
connection technique is described here, if you desire to purchase the Agilent sniffer
power probe. 1/16" ID hose may be used alone, by itself as a sniffing probe, though
it will not have a handle.
Figure 52 Alicat Scientific Mass Flow Meter Supported Method of PPM Testing
Agilent part number R0305, an adapter from KF25 to 3/16" ID hose may be
purchased as part of the power probe kit, Agilent part number K95565305, the
entire power probe sniffer assembly. The molded plastic casing of the power probe
may be re-purposed by removing the valve and copper tubing, and inserting plain
1/8" ID hose into the shell of the gun. Remove the screws holding the case together
using a 7/64" Allen key.
Connecting from 3/16" tubing to 1/16" tubing may be accomplished using
McMaster-Carr part number 2808K105. 1/16" tubing should be used as the long
line sniffer in this application, not the thicker 3/16" ID tubing. The thinner tubing
gives a faster response time for a helium signal. 3/16" ID tubing may be purchased
as McMaster-Carr part number 6516T53. This tubing should only be used as an
adapter section or as use if not using a mass flow meter. The mass flow meter
restricts flow speed, requiring a thinner tube to be used for a quick response time.
1/16" ID tube may be purchased as McMaster-Carr part number 6516T62.
Since a thin hose is required without a fine metering valve on the top of the system,
if the operator already has a power probe handle that is not in use, the gun portion
of the probe may be disassembled, striped of all valves and copper tubing, and
reassembled with a 1/16" hose inside. It is important, however, not to have another
metering valve in series above the mass flow meter. This is for optimizing response
time. Of course, this only provides a convenient handle. A naked hose may be used
for sniffing.
Measuring PPM of Helium with a non-Alicat Scientific brand Mass Flow Meter
Scientific confidence added through accurate computation of both helium leak
rate, and gas flow rate.
Does not require use of known PPM gas mixture (though use is suggested to
check calibration).
Risk of choosing a mass flow meter that is not in the proper range. The mass
flow meter selected should be in the range of SUGGESTED 0 to 1 SLM
(equivalently, 0 to 1000 SCCM).
NOTE The smallest mass flow meter that may be used while still functioning. It could be
as small as 300 SCCM, though this flow rate may not be capable of reading mass
flow when measuring flows of mobile unit leak detectors (units with larger roughing
pumps).
Alternatively, if using a mass flow meter that is not supported by the leak detector
software (anything other than Alicat Scientific brand mass flow meter):
1 Attach the mass flow meter in series above the Agilent power probe sniffing
gun (Agilent part number K9565306).
2 Attach the mass flow meter to the sniffing gun through use of a KF to
compression port adapter, 1/8" to NW25 (Agilent part number KCP250125),
if using the power probe as the metering valve (see the following figure).
3 Perform an internal calibration.
4 Perform an external calibration using the sniffing probe by:
a. Selecting Calibration > Calibrate With > External Leak.
b. Entering the value of the external calibrated leak. Keep the probe at the inlet
of the standard leak.
c. Pressing Calibrate.
2 Measure the flow rate. Write this number down for later.
3 Vent the system and remove the mass flow meter. If the test port pressure
remains constant, the flow rate should not change. Do not adjust the metering
valve on the back of the sniffing power probe without recalibrating.
4 Vent the leak detector and remove the mass flow meter, if desired for faster
helium signal response time.
5 Install the power probe back onto the leak detector and put the system into Test.
Theory
Parts per million (PPM) of helium may be measured using a helium leak detector
that is either used in conjunction with a mass flow meter, or is calibrated using a
known concentration (PPM) of helium. PPM of helium is described as the flow rate
of helium divided by the flow rate of total gas, multiplied by 100. If there are 5 moles
of helium, and 999,995 moles of other gas, then the concentration is 5 PPM helium.
This is a typical concentration of helium at sea level on earth.
A helium leak detector by itself does not measure PPM. A helium leak detector
measures a leak rate of helium in units of (atm cc)/sec which is a measure of mass
flow rate (cubic centimeter volumes filled with 1atmosphere of pressure that flows
into the leak detector inlet every second). The unit atm, atmospheres, is the same
type of unit as Torr, a measure of pressure. 1 atm = 760 Torr. Additionally, cc, cubic
centimeters, is the same type of unit as liters. 1 liter = 1000 cc.
The helium leak rate measurement taken from the leak detector by itself does not
provide enough information to describe the PPM of helium present. To get that
information, the mass flow rate of helium must be known and the mass flow rate of
total gas passing into the leak detector must be known. An external mass flow
meter may be used for this measurement. With a known mass flow rate of helium
and a known mass flow rate of total gas, a simple percentage of helium in the total
gas can be computed.
The additional 1E6 converts the fraction into a part per million.
Settings Screen
Settings Menu
The Settings Menu configures the leak detector for a desired appearance and
functionality. The Settings Menu consist of the following:
Main Settings - Date and Time setup, Leak Rate and Pressure Units selection
and Language selection.
Screen& Output Settings - Bar graph appearance, Leak Rate Ranging Selections
and Analog Output selections.
Remote Control - Wireless activation and channel selection.
Passwords and Restrictions - Restriction level and password selection plus leak
detector control lockouts.
Main Settings
To open this screen select Menu tab > Setting > Main Settings.
With the Settings Menu open, select the Main Settings Menu as shown below:
Vent Pop Up and Filament Status Pop Up Selections Released - LD01.07 2023
Screen Settings
The Screen Settings is a personalized menu that configures the leak detector's bar
graph, leak rate ranging and the analog output voltage type.
Range Configures Minimum Range (Range Stop) and Only Show Range
(Manual) Leak Rate options
Only Show Range - Configures the leak detector so that it only
displays the pre-selected decade entered in the manual range
variable field. For example, if the manual range feature was
enabled and the manual range variable was set to E-6, then the
leak rate decade displayed on the LED bar graph during testing
would be the E-6 atm cc/sec range only. This feature is useful
when the concern is only with monitoring the leak rate
measurement within a single decade.
For example, if the Only Show Range feature was enabled and
the Only Show Range variable was set to 10-06, then the leak rate
decade displayed on the LED bar graph during testing would be
the 10-06 atm cc/sec range only.
Minimum Range - Configures the leak detector so that it only
auto-ranges down to the pre-selected decade. To allow auto-
ranging to the maximum sensitivity of the system, turn
Maximum Range Off.
This feature is useful when a product specification is more than
one decade less stringent than the actual sensitivity of the leak
detector. If the leak detector is capable of displaying full scale <
E-11 atm cc/sec, but the test specification is only in the E-7 atm
cc/sec range, then setting the range stop to E-8 atm cc/sec
would reduce the test cycle time by eliminating the need for the
leak detector to reach its most sensitive scale during each test
cycle. Activating the Minimum Range extends the sensitivity
requirements for calibration, thus eliminating unnecessary
calibration errors.
NOTE Activating the range stop extends the sensitivity requirements for calibration, thus
eliminating unnecessary calibration errors (see the following table).
ON at
-10 10 5.00E-11
-9 20 5.00E-10
-8 50 5.00E-09
Analog Output
Remote Control
The remote control feature is a factory or field installed wireless option that
establishes communication between the leak detector and Agilent's wireless
hand-held remote control.
Use this menu to activate the leak detector's wireless baseboard option and to
assign the RF channel number that the remote hand-held device will link with.
Communications with the leak detector
Control of the leak detector.
NOTE VENT, TEST and ZERO buttons are still accessible in all password levels.
NOTE The password can also be read through the RS-232 port in the event of a lost
password. See Section A “Communications Protocol”.
Lock Vent Control Disables the VENT button. This option prevents
accidental system venting during test while under
remote or PLC control. The Vent Lock feature is not
active when the Auto Sequencer is on.
Lock All Control Disables the Test, Vent and Zero buttons. This option
prevents accidental system venting, testing, system
hold and Zeroing during test while under remote or PLC
control.
Wireless
Information relating to wireless communications
Maintenance Screen
Maintenance Menu
The Maintenance Menu allows you to view or change the settings/parameters of
the leak detector, including
General
• Internal Calibrated Leak
Manual Tuning
• Spectrometer tuning parameters
Valve Control
• Valve status
Pump
• Primary/Roughing Pump configuration
Gauge
• Test Port TC calibration
System Defaults
• Reset Turbo & Filament hours
This Maintenance menu (General) allows you to view and input the value and
parameters of the installed internal calibrated leak. A Software Update utility is also
now available.
Software Update
Contact Agilent Customer Service/Technical Support for the latest software
upgrade. The software update utility is necessary when added features and
software upgrades are needed. A USB jump drive is all that is needed. The USB port
is located in the back of the leak detector. See the following figure.
1 Select the Software Update option. You are prompted to insert the USB drive
into the back of the leak detector. See the following figure.
2 Enter the required values as shown on the calibrated leak label or calibration
certification.
3 Repeat step 1 and 2 for each desired item.
4 Press OK to exit the screen.
5 The value of the internal calibrated leak, as shown on the CALIBRATION Menu,
is compensated for temperature using the calibrated leak temperature sensor.
The calibrated leak is not compensated for temperature if any of the following
conditions exist:
The temperature sensor on the calibrated leak fails or there is a cable
connection problem.
The temperature sensor reading is out of the operating range. The
operating temperature range is from 10.0 to 45.0° C (50 to 113° F).
The internal calibrated leak is not a E-7 or E-8 range leak.
NOTE The actual value and information of the internal calibrated can be found on the
calibration certificate that shipped with the leak detector or shipped with the field
installed calibrated leak. The label on the internal calibrated contains also contains
all the necessary information as shown in the figure above.
Manual Tuning
Press Manual Tuning and the Manual Spectrometer Tuning screen appears
(see the following figure).
Use this screen to view and manually adjust the parameters of the ion source in the
spectrometer for a maximum helium signal.
NOTE For all tuning parameters, press Cancel to revert back to the original value or press
Done to initiate/store the desired.
Valve Status -
color code333
Figure 74 Valve Control Models G8610B, G8611B, G8611C, G8612B & G8612C
CAUTION Manipulation of the valves using the Manual Valve Control screen must only be
performed by persons who are extremely familiar with the leak detector as
damage to critical components (e.g., spectrometer, turbo) can occur.
Press the circle corresponding to that valve to change the state of a specific valve.
Vent O C C C O C O
Rough C O C C C C C
Contra-Flow Test O O C C C C O
Midstage Test O C O C C C O
Hold O C C C C C O
Sniff O O C C C C O
O: Open C: Closed
Maintenance: Pump
The Pump Menu screen displays the Fore pump type (Dry or Oil) - Factory
configured. Operating hours and Last Service are also displayed.
This menu also displays the installation of a diaphragm exhaust pump and its'
operating hours and Last Service Date.
NOTE When the Last Service date is updated/changed the Operating Hours reset to Zero.
Test Port Pressure and These values appear on this screen for reference.
Spectrometer Pressure
System Information
Press System Information on the Menus screen.
The System Information screen appears (see the following figure).
This screen gives details of the installed software version, spectrometer
parameters and the turbo frequency.
Ion Source Filament Filament 1 Off Filament 1 is selected and not lit.
See Section “Manual
Tuning”. Filament 1 On Filament 1 is selected and lit.
Gain Gain (Gain Value) Displays the gain that applied to the leak
rate. The actual helium signal is multiplied
by the gain value to read the correct,
calibrated leak rate.
As of February 2023 Firmware Rev LD01.06 120822 Filament status check for both
Filament 1 & 2 at startup and during normal operation.
Filament good and active
Filament good and not active
Filament burnt out (No Emission)
Audio Set Point Audio Set Point: Active Audio Set Point is active.
See Section
“Current Setup Menu”. Audio Set Point: Inactive Audio Set Point is inactive.
Leak Rate Range Auto Ranging System is set to auto range through
See Section the leak rate scale.
“Calibration Screen”
Manual Ranging System is set for manual range
control.
Range Stop Range Stop -[Exp] System is set to display the most
See Section sensitive leak rate range, 10-[Exp].
“Calibration Screen” The range stop line shows only if
range stop is on.
Reject Set Point Reject# 1,2,3,4 Active Reject set points displayed are active;
See Section this line is blank when all set points are
“Current Setup Menu”. disabled or inactive.
Current Setup
The current setup menu displays all activated parameters including selected
applications.
This feature provides a soft shut down for the CPU and turbo pump prior to turning
off the main power/circuit breaker switch.
NOTE This power off feature provides a two-minute braking procedure for the turbo pump.
This aides in turbo pump life and for safe transportation of the leak detector.
Communications Protocol
This section defines the technical specifications for the RS-232 protocol used with
the leak detector and describes the customer-configured inputs/output.
This section defines the technical specifications for the RS-232 protocol used with
the leak detector and describes the customer-configured inputs/outputs.
Protocol (RS-232)
S-232 ports operate at 9600 baud, 8 bits, no parity, and one stop bit.
All characters transmitted to the leak detector are echoed by the leak detector.
Commands, inquiries, and strings of both commands and inquiries must be
terminated by a carriage return <CR>; the carriage return is echoed as a space
character.
The maximum length of the input is 80 characters; if no carriage return has been
received before the 80th character, execution of the command string begins
anyway. Words that begin with:
? are inquiries for the controlling device to determine the current state or value
of a leak detector parameter.
PUT- are commands for the controlling device to set the current state or value
of a volatile leak detector parameter.
INIT- are commands for the controlling device to set non-volatile leak detector
parameters.
Other commands do not require a parameter and do not begin with any special
characters.
Successful inquiries respond with data, followed by a space, then: ok<CR><LF>, as
specified in the tables of this appendix. Unsuccessful inquiries respond with the
offending inquiry followed by a space, then: #?<CR><LF> (line feed, <LF>).
Inquiries, parameters, and commands can be concatenated. Each word or numeric
parameter is followed by one or more spaces. The string is terminated by a <CR>,
which causes execution to begin.
Successful strings respond with the specified data for the input inquiries in the
order the inquiries were issued, followed by: ok<CR><LF>.
Unsuccessful strings respond with the first offending command followed by a
space, then: #?<CR><LF>. All commands and inquiries following the failed word are
ignored; all parameters are discarded.
Tables 16-17-18-19 outline the available control and query commands.
The following table lists inquiries that are used to determine internal operating
parameters. All inquiries, except VER, begin with ? and end with <CR>.
Inquiry Response
?ALL Responds with four lines. Each begins with a <cr><lf>. The first line reports the ion chamber
value. The second line reports the emission value. The third line reports the value of the
offset variable. The fourth line reports the gain value.
?CPLOCKOUT Reports the status of the control panel lockout feature, ON or OFF.
?CURRDATE Reports the current system date and time in the following format: 3 05/25/2005
04:19:03.35 1 where 3 is the day of the week (0 = Sunday, 1 = Monday, etc), 05/25/2005
is mm/dd/yyyy (date), 04:19:03.35 is hh/mm/ss/ms (time) and 1 is PM
(0=AM, 1=PM, 2=24 hour).
?DISPLAY Reports what the display bar graph is set to, log or lin.
?EXPONENT Reports the current manual-mode exponent with a two-character (minus followed by one
digit) number. The exponent is within the least sensitive range (as reported by ?RANGE) and
the three lower (more negative exponent) ranges. See also INIT-EXPONENT.
?EXTLEAK Reports the currently stored value of the external standard leak.
?GAIN Reports the system gain value by returning a three-character number consisting of a two-
digit gain factor with a decimal point after the first digit.
?INTEXT Indicates which leak, Internal or External has been selected for use during calibration.
?LANGUAGE Reports the language that has been selected to display on the front panel (0=English,
1=French, 2=German, 3=Korean, 4=Spanish, 5=Japanese, 6=Chinese or Russian
depending on software version purchased).
Inquiry Response
?LEAK Reports a six-character number consisting of a two-digit leak rate mantissa with decimal
point after the first digit followed by E - then a two digit number, which is the leak rate range
exponent (e.g., 1.3E-08).
?LEAKTEMP Reports the internal calibrated leak temperature that the factory calibration was performed
at (degrees C).
?LP Reports the current leak rate along with the test port and system pressures.
?LPV Reports the current Leak Rate, Pressures and Valve state.
?ML-XFER Reports the current Test to Fine Test crossover pressure setpoint in Torr. The response is in
exponential format (in Torr): X.XE-XX
?PRESSURES Reports two lines. Each begins with a <cr><lf>. The first line consists of the words test port
TC followed by a number in mTorr. The second line consists of the words system TC
followed by a number in uTorr.
?QUICK-CAL Reports whether Fast or Full Calibration is chosen if a calibration were to be started.
?REJECT Reports the reject leak rate setpoint (e.g., 7.0E-05) (where x is a number from 1 to 4 or the
letter A).
?>xREJECT Reports the status of a reject set point (where x is a number from 1 to 4 or the letter A).
?xREJECT Reports the reject leak rate setpoint (e.g., 7.0E-05) (where x is a number from 1 to 4 or the
letter A).
?ROUGH Reports the currently stored Test Sequencer rough time used as part of setting up the Auto
Sequencer.
Inquiry Response
?SETUP Reports seven lines, each begins with a <cr><lf>. The first line reports the turbo pump speed
with the RPM value or OFF. The second line reports the selected manual leak rate range,
and the ranging method as auto or manual. The third line reports the least sensitive leak
rate range. The fourth line reports the reject set point leak rate. The fifth line reports the
internal calibrated leak rate value. The sixth line reports the DAC output method as Linear,
Log(2V), or Log(3V). The seventh line reports the active filament status as One or Two,
followed by Lit or Out.
?STDLEAK Reports a six-character number consisting of a two-digit calibration standard leak rate
mantissa with a decimal point after the first digit followed by E-, and then by a two-digit
exponent (e.g., 1.3E-07). This inquiry is used for reporting the calibration standard leak rate
entered using INIT-STDLEAK.
?SYSLEAK Reports the leak rate value that has been stored that relates to the split flow functionality.
?TEST Reports the currently stored Test Sequencer test time used as part of setting up the Auto
Sequencer.
?TEMPFACTOR Reports the internal calibrated leak temperature factor that is used as part of the
temperature correction logic (degrees C).
?TURBO Reports four lines, each begins with a <cr><lf>. The first line consists of the word turbo
followed by either READY or NOT READY. The second line consists of the word turbo
followed by either Fault or No Fault. The third line consists of the words turbo speed RPM
followed by the RPM value:
?VALVESTATE xxxxx. The fourth line consists of the words "Turbo Temp (Celsius):" followed by the bearing
temperature value: XX.
VER Reports the CPU software version in the form LC02.05 followed by a six digit hexadecimal
checksum.
?0ENABLE Responds with the status of the Zero Enable function, ON or OFF.
The commands listed in the following table are used to set non-volatile operating
parameters. The current value of the operating parameter is changed to the new
value. If the leak detector is in back panel control, the response is cant.
Command Parameter
INIT-CL-XFER Preceded by X.XE-X, sets the value of the contra-flow mode crossover pressure in Torr.
INIT-CPLOCKOUT Sets the status of the control panel lockout. Preceded by (0) or (1), 0 = OFF, 1 = ON.
INIT-DATE Sets the system date and time. Preceded by: dow, mm dd yyyy hh mm ss AmPm 24-12
mode, Where 2 is the day of the week (dow) (0 = Sunday, 1
INIT-DISPLAY = Monday, etc), mm/dd/yyyy (date), hh:mm:ss (time) AmPm (0=AM, 1=PM)24-12 mode (0
for 24 hr, 1 for 12 hr). Examples: 2 02 12 2005 04 16
INIT-EXPONENT Sets the range that the leak detector will operate in when Manual Range is ON. Preceded by
0 to -11.
INIT-EXTLEAK Sets the value of the external leak. Preceded by a two-digit leak rate mantissa with a
decimal point after the first digit followed by E-, and then by a two-digit leak range exponent:
the helium leak rate number in atm cc/sec of the calibration standard leak.
INIT-GAIN Sets the value of the gain used for adjusting the helium signal to match a calibration
standard leak. Preceded by a two-digit number with a decimal point after the first digit, in
the range 1.0 to 6.0.
Note: Changing Gain changes Zero by the same ratio.
Command Parameter
INIT-ION Sets the ion voltage of the ion source. Preceded by a three-digit number (volts) in the range
220 to 280.
INIT-LANGUAGE Selects the language to be shown on the display. Preceded by 0-6 where (0=English,
1=French, 2=German, 3=Korean, 4=Spanish, 5=Japanese, 6=Chinese).
INIT-LINEAR Sets the leak rate analog output voltage to Linear. Not preceded by a value.
INIT-LKEXPIRE Sets the date on which the internal calibrated leak expires. Preceded by the date.
Example: mm dd yyyy INIT-LKEXPIRE.
INIT-LEAKTEMP Sets the internal calibrated leak temperature (printed on the leak label) that the factory
calibration was performed at. Preceded by 12.0 to 50.0 (degrees C).
Example: 23.5 INIT-LEAKTEMP
INIT-1LOG Sets the leak rate analog output voltage to logarithmic at 1 volt per decade. Not preceded
by a value.
INIT-ML-XFER Sets the value of the Fine Test-flow mode crossover pressure in Torr. Preceded by X.XE-XX.
The range is 1.0E-03 to 1.0E-01 Torr. The default is 1.0E-01 (100 mTorr).
INIT-PASSWORD Sets the system password. Preceded by a 1 to 6 digit password and followed by a digit
indicating the number of digits in the password. Example: 123123 6 INIT-PASSWORD.
Password is read back using the ?SYSPWD command.
INIT-QUICK-CAL Sets which type of calibration will be run if the CALIBRATE command is initiated. Preceded
by 0 or 1, 0 for FULL cal, 1 for FAST cal.
INIT-RANGESTOP Sets the value of the most sensitive range exponent. Preceded by XX.
INIT-1REJECT Sets the leak rate value for Reject Set Point #1. Preceded by a two-digit leak rate mantissa
with a decimal point after the first digit followed by E-, and then by a two-digit leak range
exponent, the helium leak rate number is in atm cc/sec.
Example 3.4E-08 INIT-1REJECT
Command Parameter
INIT-4REJECT Sets the test port pressure value for Reject Set Point #4. Preceded by a two-digit leak rate
mantissa with a decimal point after the first digit followed by E-, and then by a two-digit
exponent: the test port TC pressure in millitorr.
Example 1.5E-02 INIT-4REJECT.
INIT-AREJECT Same as INIT-1REJECT, but for the Audio Reject Set Point.
HIGH-xREJECT Sets the reject set point to activate when the actual test value is higher than the set point.
Where x is a number from 1 to 4 or the letter A. For example, the reject set point is set for
7.0E-05 and the actual test value is 9.2E-05.
LOW-xREJECT Sets the reject set point to activate when the actual test value is lower than the set point.
Where x is a number from 1 to 4 or the letter A. For example, the reject set point is set for
7.0E-05 and the actual test value is 4.3E-07.
INIT-ROUGH Sets the Auto Sequencer rough time in seconds. Preceded by XX.
INIT-SYSLEAK Sets the split flow leak value, system must be in TEST and the split flow feature must be
OFF when this parameter is set. Preceded by a two-digit leak rate mantissa with a decimal
point after the first digit followed by E-, and then by a two-digit leak range exponent: the
helium leak rate number in atm cc/sec.
Example 3.4E-08 INIT-SYSLEAK.
INIT-STDLEAK Sets the value of the internal leak. Preceded by a two-digit leak rate mantissa with a decimal
point after the first digit followed by E-, and then by a two-digit leak range exponent: the
helium leak rate number in atm cc/sec.
Example: 3.2E-08 INIT-STDLEAK.
INIT-TCTPZERO Used to calibrate the low-pressure end of the test port thermocouple to a known pressure
higher than zero. Preceded by a number (XX) of mTorr representing the test port pressure at
the time of the command.
INIT-TEMPFACTOR Sets the temperature correction factor that is printed on the internal calibrated leak label.
Preceded by –10 to +10.
Example: -7 INIT-TEMPFACTOR
INIT-TEST Sets the Auto Sequencer test time in seconds. Preceded by XX.
LOCK-MS Enables the MS lock feature. If the system exceeds 150 mTorr, the system goes from Fine
test directly into Roughing mode (no Contra mode). If the leak rate exceeds the 7 range, the
over range symbol appears.
?MS-LOCK Queries the status of the MS lock feature; either on or off appears.
The commands listed in Table A-3 are used to cause an immediate change in the
spectrometer operating parameters. These Commands Do Not Change The Non-
Volatile Operating Parameters.
Command Parameter
PUT-EXPONENT Sets the exponent leak range for Manual mode. Preceded by a two-character number
Acceptable values are 0 through -10.
PUT-ION Sets the ion value of the ion source. Preceded by a three-digit number of volts, in the range
200 to 350.
PUT-RANGESTOP Sets the Range Stop exponent. Preceded by two or three characters -XX or -X.
The commands listed in the following table are used to cause certain leak detection
actions.
Command Parameter
CALIBRATE Initiates a Full or Fast calibration depending on system settings. The CPU software tunes,
then adjusts the gain so that the current helium signal causes the current leak rate
measurement to be the same as the most recently input using INIT-STDLEAK.
DECREMENT Subtracts 1 from the Manual Ranging leak rate exponent (making it more negative).
Success is indicated by the normal ok response. Does nothing in auto-ranging mode. Does
nothing upon reaching the most sensitive range.
DISABLE-xREJECT Turns the Reject Set Point OFF. Where x is a number from 1 to 4 or the letter A.
Command Parameter
ENABLE-xREJECT Turns the Reject Set Point ON. Where x is a number from 1 to 4 or the letter A.
EXTERNAL Sets the External Calibrated Leak (in test port) as the leak for use during calibration.
INCREMENT Adds 1 to the Manual Ranging leak rate exponent (making it less negative).
Success is indicated by the normal ok response. Does nothing in
auto-ranging mode. Does nothing upon reaching the least sensitive range.
INTERNAL Sets the Internal Calibrated Leak as the leak to be used during calibration.
KEEP Causes the leak detector to close most valves and enter the Hold state.
MANUAL Turns the Manual Ranging Mode ON. Success is indicated by the normal ok response.
ROUGH Causes the leak detector to begin a roughing on the test port only.
STDLEAK Turns Read Std Leak ON/OFF. Verifies calibration by exposing internal cal leak to the
system.
TPTCATM Sets the current test port thermocouple reading to represent atmospheric pressure when
the test port is exposed to atmosphere. Success is indicated by the normal ok response.
TPTCZERO Sets the current test port thermocouple reading to represent a ZERO pressure reading when
the test port pressure is less than 1.0E-04 Torr. Success is indicated by the normal ok
response.
TUNE Adjusts the ion source voltage to cause a maximum response to helium.
VENT Causes the leak detector to isolate and vent the test port to the atmosphere.
XYZZY Unlocks the system so that all RS-232 commands are accessible.
ZERO Sets the current leak rate measurement to be 0.0 atm cc/sec in the most sensitive range.
0 9600
1 19200
2 38400
3 57600
4 115200
Table 21
Command Action
Table 22
Command Action
Command Action
DISMISS_CAL OK Automatically selects OK for both “Cal OK” and “Cal Failed” pop ups
Command Action
Command Action
Table 26 Startup Calibration Messages – Cancel of Execute Calibration at startup (20-minute bypass)
Command Action
Serial Connector
Use the SERIAL connector on the Rear Panel to interface the leak detector to a PC
via an isolated RS-232 connection. Table A-5 and Figure A-1 provide details of the
connection. A typical connection from a 9-pin PC serial connector requires a
female-to-female, 9-pin, null modem cable, where pins 2 and 3 are crossed.
2 3 5
I/O Connector
The following table lists the optional discrete I/O PCB (P/N# R2111501) pin signals.
This I/O connector provides a means for a controller, usually a PLC, to control or to
determine the leak detector status. The I/O is accessible via a type DB25S (female
socket) connector.
Table 28 I/O Connector Pin Signals
1 CAL_OUT O Level Active HIGH when the leak Performs calibrating routine with
detector is in CALIBRATION the Internal Calibrated Leak;
mode. Zeroing routine is included.
2 ZERO_OU O Level Active HIGH when the leak Performs the Zeroing routine.
T detector is in ZEROING mode.
3 HOLD_OU O Level Active HIGH when the leak Performs a HOLD routine and
T detector is in HOLD mode. isolates the test port and any
test objects from the leak
detector vacuum system.
4 VENT_OU O Level Active HIGH when the leak Performs a VENT routine,
T detector is VENT mode. isolates the test port from the
leak detector vacuum system
and vents the test port to
atmosphere.
6 READY_ O Level Active HIGH when the leak Indicates that the leak
OUT detector is in READY mode. detector is completing a
STARTING routine and that
the filament and turbo are OK.
7 TEST_OU O Level Active HIGH when the leak Indicates that the leak
T detector is in Test and detector is ready for fine leak
Fine Test modes. testing.
8 REJECT1_ O Level Active HIGH when the leak rate is Indicates that the value of the
OUT over set point #1. leak rate exceeds the value set
in the 1REJECT variable.
The following figure shows the output circuit sketch. Level outputs are optically
isolated emitter followers with 10 Ohm series resistors, and a 14 mA max drive
current (24 VDC max).
The following figure shows the input circuit sketch. All inputs are optically isolated
3300 Ohm resistive loads in series with the LEDs of the opto-isolators and voltage
level from +5 VDC to +24 VDC. The required pulse inputs are 200 ms minimum
pulse width.
The only molecule smaller than Helium (mass 4) is Hydrogen (mass 2), which is not
inert. Helium is much lighter than the next heavier inert molecule, Neon (mass 20)
which is much more expensive. Helium is present at a concentration of only 5 ppm
in normal atmospheric conditions.
Our SIPD family of products is also sensitive to helium and is based on a patented
technology called Selective Ion Pump Detection (see the following figure). The
sensor technology incorporates an ion pump connected to a quartz capillary tube
and maintained under high vacuum. This membrane is heated with a coiled
platinum filament. Once heated, the membrane becomes permeable to helium. As
the partial pressure of helium in the ion pump increases, so does the current draw
of the ion pump. This current is proportional to the pressure and is therefore
representative of the helium at the test probe of the instrument.
A selective ion pump detector consists of the following:
An ion pump and controller
A permeable quartz capillary
A heater coil that surrounds the quartz capillary
Electronics to process the signal
Display for access to leak rate and other unit function.
Locating Leaks
To pinpoint the location of the leak(s) (but not measure the total leakage rate),
helium is administered to the suspected leak sites of the part using a spray probe
with an adjustable flow (see the following figure).
Measuring Leak
To determine the total quantity of leakage (but not the number or location of leaks),
the part is connected to the leak detector and shrouded by a helium environment.
This helium environment can be contained in many methods ranging from a simple
plastic bag to more complex bell jar arrangements (see the following figure).
Mount Test
Part
In this technique, the part is pressurized with helium or a mixture of helium and air,
and tested by one of the following methods.
Measuring Leaks
To determine the total quantity of leakage (but not the number or location of leaks),
the part is pressurized with helium (or a mixture of helium and air or nitrogen).
This can be done by bombing or backfilling small hermetically sealed parts. Larger
parts can be actively pressurized using a hose or tubing to deliver the helium. The
part is placed in a volume that is then evacuated by the leak detector. All the helium
escaping from the part is captured and quantified (see the following figure).
Mount Test
Part
Locating Leaks
To pinpoint the location of the leak(s) (but not measure the total leakage), the likely
potential leak sites of the part are scanned using a Sniffer Probe connected to the
inlet of the leak detector (see the following figure).
Mount Test
Part
Figure 93
Vacuum Systems
In general, vacuum systems are tested with a portable leak detector. Typically the
leak detector is connected by means of a tee connected in between the foreline of
the high vacuum pump and the inlet of its backing pump. A system should be
capable of maintaining a foreline pressure low enough to operate the leak detector
at this location. Helium is supplied to potential leak site using a spray probe or
bagging suspected areas. If a leak exists, helium enters the system and rapidly
diffuses through it. The leak detector should respond within several seconds or
less. Leak detector sensitivity is diminished in systems with large backing pumps. If
a system is using a cryopump as a high vacuum pump, it must be valved off before
helium is introduced as cryopumps have limited helium pumping capacity
(see the following figure).
Mount Test
Part
Pressurized Systems
Many different types of pressurized systems also need to be leak-free. These
systems can be charged with helium or some mixture of helium and another gas
such as nitrogen. If a diluted helium mixture is used, the helium signal is diminished
proportionally. For example, if a mixture of 10% helium and 90% nitrogen is used,
the signal reads 10% of the actual value of the leak, or a decade lower. This may be
acceptable in many cases as system leak checking is usually to locate rather than
quantify leaks. Once the system has been charged with an appropriate amount of
helium, leak checking can be performed by means of a sniffer probe, or by bagging
suspected leak sites so that leaking helium accumulates to a detectable level.
Maintenance of Systems
Many vacuum process tools in fabs of all types require occasional leak checking.
This may be part of a preventative maintenance schedule or in the event of an
unexpected failure.
Downtime in either case must be minimized. A rugged, dependable, fast starting
leak detector is essential to maximize up time of production tools. Agilent Agilent
Technologies Helium Mass Spectrometer portable leak detectors keep industry
moving. Examples of some of these applications are:
Pressurized systems
Power Plants
Underground tanks, cables, and pipes
High purity gas handling systems
Bioreactors and fermenters
Liquid gas manufacturing facilities
Fuel tanks and bladders.
R3391301 TC Gauge
X3760-64005 G8611A, G8612A, G8610A, G8611D AVF 60 Gold Mechanical Pump Oil
*Most accessories with NW25 fittings are available with 1 1/8 compression fittings.
Call technical support for more information.
Dear Customer,
Please follow these instructions whenever one of our products needs to be returned.
Complete the attached Request for Return form and send it to Agilent Technologies (see below), taking particular care to include the
completed Health and Safety declaration Section. No work can be started on your unit until we receive a completed copy of this
form.
After evaluating the information, Agilent Technologies will provide you with a Return Authorization (RA) number via email or fax, as
requested. Note: Depending on the type of return, a Purchase Order may be required at the time the Request for Return is submitted. We will
quote any necessary services (evaluation, repair, special cleaning, eg).
Product preparation
o Remove all accessories from the core product (e.g. inlet screens, vent valves).
o Prior to shipment and if applicable for your product, drain any oils or other liquids, purge or flush all gasses, and wipe off any
excess residue.
o If ordering an Advance Exchange product, please use the packaging from the Advance Exchange to return the defective
product.
o Seal the product in a plastic bag, and package product carefully to avoid damage in transit. You are responsible for loss or
damage in transit.
o Include a copy of the Health and Safety Declaration in the shipping documentation on the outside of the shipping box of your
returning product.
o Clearly label package with RA number. Using the shipping label provided will ensure the proper address and RA number are on
the package. Packages shipped to Agilent without a RA clearly written on the outside cannot be accepted and will be returned.
o Return only products for which the RA was issued.
Shipping
o Ship to the location specified on the printable label, which will be sent, along with the RA number, as soon as we have
received all of the required information. Customer is responsible for freight charges on returning product.
o Return shipments must comply with all applicable Shipping Regulations (IATA, DOT, ADR, etc.) and carrier requirements.
RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION:
Fax: 00 39 011 9979 330 Fax: 1 781 860 9252 Please visit our website for
Toll Free: 00 800 234 234 00 Toll Free: 800 882 7426, Option 3 individual office information
o Unless otherwise pre-negotiated, customer is responsible for the freight charges for the returning product. Return shipments
must comply with all applicable Shipping Regulations (IATA, DOT, etc.) and carrier requirements.
o Agilent Technologies is not responsible for returning customer provided packaging or containers.
o Customers receiving an Advance Exchange product agree to return the defective, rebuildable part to Agilent Technologies
within 15 business days. Failure to do so, or returning a non-rebuildable part (crashed), will result in an invoice for the non-
returned/non-rebuildable part.
o Returns for credit toward the purchase of new or refurbished Products are subject to prior Agilent approval and may incur a
restocking fee. Please reference the original purchase order number.
o Units returned for evaluation will be evaluated, and a quote for repair will be issued. If you choose to have the unit repaired,
the cost of the evaluation will be deducted from the final repair pricing. A Purchase Order for the final repair price should be
issued within 3 weeks of quotation date. Units without a Purchase Order for repair will be returned to the customer, and the
evaluation fee will be invoiced.
o Products returned that have not been drained from oil will be disposed.
o A Special Cleaning fee will apply to all exposed products
o If requesting a calibration service, units must be functionally capable of being calibrated.
Equipment
Product description Agilent PartNo Agilent Serial No Original Purchasing Reference
Failure description Type of process (for which the equipment was used)
Type of return
Non Billable Billable New PO # (hard copy must be submitted with this form): ______________________________________
Exchange Repair Upgrade Demo Calibration Evaluation Return for Credit
Goods preparation
If you have replied YES to one of the above questions. Has the product been purged? YES NO
If yes, which cleaning agent/method:
Has the product been drained from oil? YES NOT APPLICABLE
I confirm to place this declaration on the outside of the shipping box.
I declare that the above information is true and complete to the best of my knowledge and belief.
I understand and agree to the terms and conditions on page 2 of this document.
Name: Authorized Signature:
Position:
Date:
NOTE: If a product is received at Agilent which is contaminated with a toxic or hazardous material that was not disclosed, the customer will be held
responsible for all costs incurred to ensure the safe handling of the product, and is liable for any harm or injury to Agilent employees as well as to any third
party occurring as a result of exposure to toxic or hazardous materials present in the product.