Scil Quick Guide
Scil Quick Guide
1.866.382.6937 | www.scilvet.com
Quick Guide
scil Vet abc Plus+
Table of Contents
Start Up Cycle 1
Reagent Pack Change 1
Status Icon And Log In 2
Archive Results 2
Quality Control Target Values 3
Quality Control Handling 5
Running A Quality Control 6
Analyzing A Patient Sample 7
Machine Shutdown Procedure 8
Minimizing Reagent Usage Concentrated Cleaning 8
Blockage Removal Procedure 9
Chamber Cleaning Procedure 10
Calibration 11
Flag Descriptions 12
Troubleshooting Status Messages 13
Icon Descriptions 14
Photometer Adjustment 15
▶Start Up Cycle
A start up cycle is performed every time the machine is powered on by turning
on the switch at the back of the machine. A temperature check cycle will initiate
on power up and once it passes, the Start up cycle will begin. Samples should
not be run until a start up has passed.
To initiate a start up cycle press the status button located at the top left of the screen
and then press the green start up button on the bottom right corner of the screen.
1
Quick Guide
scil Vet abc Plus+
▶Status Indicator
The status Indicator button located in the top right of the screen has 3 states.
ORANGE Indicates that an alarm has been raised. You are still able to analyze a
sample
RED Indicates that an alarm has been raised and the machine is locked. You
cannot analyze samples until the alarm is resolved
The 5 possible categories for alarms in order as shown on the status screen are;
A message will be displayed explaining the reason for the alarm. More information on these
alarms can be found on pages 14 of this Quick Guide. Ensure your startup has passed, and
your QC has passed. This will help prevent and resolve most status alarms.
▶User Log In
• Admin – Password is SCIL
• Vet – No password required
• Tech – Used only by tech support
2
Quick Guide
scil Vet abc Plus+
▶Archive Results
1. Insert a USB Stick into the front of the machine
2. Press the maintenance menu button
3. Press the Arch. Tab at the top of the screen
4. Press the purple button beside analyses
5. Press the incremental archive button to save the last 30 days of results
6. If you want to save the entire history of results stored in the machine use the complete
archive button
3. Make sure there are no other files on the USB Stick that start with the name “qcget_mx”. If
there is one, delete it
4. Open Internet Explorer. Do not use Google Chrome, Firefox etc
5. Go to www.scilvet.com and scroll to the bottom of the home page to find the control file
6. Right click on the control file link and select “save target as”
7. Navigate to the removable disk and save the file on the usb stick. Do not open the file!
8. To remove the USB stick, right click on the usb icon near the time and date and select
eject usb mass storage device
3
Quick Guide
scil Vet abc Plus+
▶Loading QC Targets
1. Insert the USB stick into the front of the scil Vet abc Plus+
2. Select the QC menu icon from the main screen
3. Press the import button
4. (a) If you only run green top normal controls, select only the green control
(b) If you run all three levels of controls, press the ALL QC’s option
5. Press the import button again
6. Confirm the entries by pressing the check mark
7. A lot number can only be used once, if a multiple LOT # needs to be used you can
change one digit of the LOT number to allow entry of ther targets EX. MX075N - MX075G
To delete the QC target values from the scil Vet abc Plus+
4
Quick Guide
scil Vet abc Plus+
Caution:
The Blood Controls:
- Must be stored in a refrigerator between 2 and 8 degrees celsius
- Can never be frozen at any time
- Cannot be left out for more than 10 minutes at room temperature
- Cannot be held in your hand while waiting for control results, may get too warm
- Do not use a blood rocker
- Only remove the control that you are going to use from the refrigerator
- Do not store controls in the refrigerator door
TIPS:
If a control has been left out of the refrigerator too long the HGB value will be either too
high or too low and usually accompanied by the ! flag.
If a control is dropped it may damage the cells and cause SCH, MIC flags as well as the
$ flag.
If dried flakes of blood have fallen inside the vial due to flaking around the rim and cap
there is a chance that the * flag or the $ flag may appear beside the WBC or RBC
measure value and could possibly create a low value due to a blockage. PLT counts are
affected as well.
Minotrol is stable until the printed expiration date if unopened. Once the control is opened
it is good for 16 days in a refrigerator stored between 2 and 8°C as long as proper handling
procedures have been followed.
5
Quick Guide
scil Vet abc Plus+
6
Quick Guide
scil Vet abc Plus+
Requirements:
• Use EDTA K2 lavender top tubes
• Ensure tubes are filled at least half way
• Mix sample thoroughly by inverting for 2 minutes
• Check for clots
• Analyze the sample as soon as possible
Procedure:
1. Log on as Admin or Vet
2. Press the species selection button to change the species
3. Press the internal keyboard button and enter the sample ID and any other fields you wish
to input. An auto ID is generated for the sample ID
4. Press the OK button
5. When the needle comes down place the sample tube under the needle and press the
sample bar to aspirate the sample
6. Once the analysis is complete, the measured results and histograms will appear on the
screen
7. If you notice any flags or alarms and are unsure of what they mean, please consult the
flags and limits section of the Quick Guide
Sample ID► Is the only field that must be filled in to run a sample. AUTOSID is default
Animal ID► The patient will be saved internally and can be identified by this. Also
used to identify patient in Practice Management Software
Date of Birth► If you use this field, you must enter all fields (Month, Day, Year)
7
Quick Guide
scil Vet abc Plus+
It is very important to follow the shut down procedure for your scil Vet abc Plus+
1. Press the exit button located at the bottom left of your screen
2. Press the logout button
3. Power the machine off by switching the black button located at the back of the machine
to the off position
Other Buttons:
Reset button will reboot the instrument without switching off the power
Logout button will end a session and return to the log in screen
Always cover the scil Vet abc Plus+ after it is shut down with the provided dust cover
1. Never turn the scil Vet abc+ on unless you need to use it
2. Once the analyzer has been turned on, leave it on until the end of the day
3. Never press the shut down button
4. If your blood control does not pass, call scil animal care tech support. Do not continue
running controls until tech support has advised you to
5. Perform a blockage removal procedure instead of running a concentrated cleaning
when results indicate that a cleaning is required. The blockage removal procedure does
not require any reagents (Pg. 9)
6. Maintain your scil Vet abc Plus+ by performing one concentrated cleaning and one
chamber cleaning procedure per reagent pack change to reduce crystallization in the
chambers
8
Quick Guide
scil Vet abc Plus+
▶Concentrated Cleaning
Once per reagent pack change
This procedure will use no reagents from your pack and can be used to resolve
blockages (Low WBC or Low RBC) , constant flags (* and $) and start up failures.
9. Put 3ml of distilled water in each of the chambers and start the chambers draining cycle
again
10. Slide the black cover back on making sure the cover sits flush against the frame and no
tubes are pinched. Run a concentrated cleaning once you have finished
10
Quick Guide
scil Vet abc Plus+
▶Calibration
The scil Vet abc Plus+ should be calibrated every six months
Before Calibrating
• Perform a blockage removal and run a start up
• Check your photometer reading and adjust if necessary
• Value should read 220 - 245
• Instructions are included at the back of this Quick Guide
• Run 1 daily quality control and make sure the values aren’t too far out of range. Slightly
out of range is acceptable
• If you are unsure of your results please call technical support before proceeding with the
calibration
DO NOT remove the Calibration Fluid from the refrigerator until step 6 is complete and
you understand the instructions.
Do not follow the instructions that come with the calibrator. Only follow these
instructions to complete the calibration.
When calibrating your machine the measured value of each parameter may not reflect
the target values indicated on the sheet provided with the calibrator vial.
Your goal is not to recreate the values on the target sheet provided with the calibrator
vial; the targets are used as reference for calibrating the instrument.
If you are in the middle of doing a calibration and need to call technical support make
sure you return your calibration fluid to the refrigerator.
▶Flag Descriptions
* - following either WBC, RBC HCT or PLT -- system analyzed the sample 3 times but all 3
counts differed and were outside the system’s precision limits. Result should be verified by
repeating the sample.
$ - between the test result and the units -- 3 counts were made and 2 were within the
system’s precision limits. The result can be accepted.
! - next to the HGB result -- the difference between HGB blank done for this analysis and the
HGB blank done for the previous analysis was outside the system’s precision limits. Instrument
provides a result based on the previous HGB blank. Result can be accepted. If flag persists
for more than 3 consecutive times call scil animal care
---D or < printed instead of a result -- linearity range for that parameter has been exceeded .
Repeat using a 1:2 dilution with saline solution. Multiply your results by 3.
MIC following PLT Flags -- microcytes present in the platelet measurement zone. Verify result
with manual count.
SCH following PLT Flags -- schistocytes or platelet aggregates present in the platelet
measuring zone. Review slide before reporting results.
SCL following PLT Flags -- small particles in the 2 - 3 fl zone. Reanalyze sample. If flag persists,
perform an automatic cleaning cycle and repeat again. If flag still persists, perform manual
platelet count.
AG1 and AG2 – Flags for big Leucocytes aggregates (> 1,5 % / 2,5 % of total Leucocytes).
Mix sample properly and repeat measurement.
12
Quick Guide
scil Vet abc Plus+
13
Actions Quick Guide
Quality Control
Start Up – Initiates a Start Up Cycle scil Button
Import Vet abc Plus+
– Used to import Quality
Control Target values
Shutdown – Shuts machine down (DO Control History – Shows the control history
NOT USE) stored in machine
Restart – Restarts machine QC Expired – Loaded QC data is out of
date
Logout – Disconnects current user and Inter QC time has elapsed – Instrument
displays the Log In screen expects a QC to be run
Edit – Allows user to begin making chang- Invalid QC – Some of the parameters are
es within the current screen out of range on the last ran QC
OK – Validates an action QC Passed – All parameters are within
range
Cancel – Cancels an action QC Icon – displayed when proper QC lot
# is entered in Sample ID
Reset – Resets the entered fields or by- Calibration Creation - Used to calibrate
passes the need to archive machine, opens menu
Delete – Deletes the selected item Reagent
Species – Allows user to change species Settings – (Date, time, printer, network,
species, alarms)
OK – Confirms an action Customer Service – (Prime, back flush,
concentrated clean, chambers draining)
Cancel – Cancels an action Service Cycles – (Needle, carriage,
valves etc.)
Emergency Stop – Stops a cycle in prog- Archive
ress
New User Button – Used to create a new Complete Archive – Archives all data
user in the settings menu found in machine
Keyboard – Opens virtual Keyboard Incremental Archive – Archives the last 30
days of information
Valid – Validates keyboard entry and Restore Button – Settings and analyzer set-
moves to next field tings can be restored to an older configu-
ration
Close – Closes keyboard after entries are
all complete
Menu
Sample ID – Enter data for running a Status – Displays all alarms, run a Start Up
sample from this screen
Quality Control – (Quality control data, Exit – Change user, restart or shut down
calibration) machine
Results Archive – Stored and searchable Help – Provides help with the current
results up to 1000 results screen
Maintenance – (Reagents, archive, stats,
logs, service)
14
Quick Guide
scil Vet abc Plus+
▶Photometer Adjustment
Note: The photometer should read between 220 and 245
Tools Required:
• 2.5 mm allen wrench screw driver (scil Vet abc Plus+ screwdriver)
• Precision flat head screwdriver
Procedure:
1. Log in as TECH (password: 01HB05)
2. Press the maintenance button
3. Press the adjustment icon
4. Press the “gains” tab at the top of the screen
11. Once the number is displayed you can begin to turn the VR97 brass screw. Watch the
screen to make sure the number is changing. When you reach 240 approximately, stop
turning the screw and wait for the cycle to finish
12. Place your cover back on the machine and close the door without putting the cover
screws back in. Run a Start Up
13. If the Start up passed you can secure the cover to the machine with the screws and you
are done!
16
scil animal care company
www.scilvet.com