Experiment 7
Experiment 7
FLOW SENSOR
1
Procedure
1) Connect, through leads, pin No. 16 of the PUMP DIRIVER to the bush of SET
POINT 1, pin No. 17 to adjacent bush and pin No. 18 to its adjacent bush
5) Regulate the valve MANUAL VALVE to read on the FLOW METER 6 1t/min: if the
flow is slightly lower, regulate MANUAL VALVE to reach the value closely higher.
6) Start the pump regulating the voltage on SET POINT 1 at 10V and at the same time
start the chronometer.
7) Read the value shown in the FLOW METER and write it down.
8) Wait until the water level reaches 13cm, stop the chronometer and stop the pump.
10) Slightly turn counter-clockwise the knob NEEDLE VALVE and or the knob DRAIN
VALVE to bring the water level back to 10cm and close the valves.
11) Repeat the procedure from step 4: the two values read on the chronometer must be
equal.
13) Calculate the flow of the pump and verify that the value is equal to the value read on
the flow meter.
14) Connect pins No. 5 and 6 of the flow sensor FLOW to its adjacent pins of the relevant
interface
2
15) Insert one terminal of the digital voltmeter, set in dc, in pin No. 11 and the other one
in the earth bush.
16) Regulate the trimmer OFFSET to read the value 0V on the multimeter.
18) Slightly turn counter-clockwise the knob NEEDLE VALVE and/or the valve DRAIN
VALVE.
19) Start the pump regulating the voltage on set point 1 at 10V
20) Regulate the trimmer GAIN to read on the multimeter the value 9V, in case the flow
read and calculated is 6 lt/min.
21) Regulate the valve MANUAL VALVE for all the value of flow listed in Table 7.1
and write down the voltage values read on the voltmeter.
3
Table 7.1
Flow
1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6
(liters/m)
Voltage
(V)
fig. 7.2