Lab Report 6
Lab Report 6
GROUP CEEH220 3F
EXPERIMENT 6
SEMESTER 3
Remarks:
Checked by: ---------------------------
Date:
TABLE OF CONTENTS
1.0 Abstract/Summary…………………………………………………………………….
2.0 Introduction……………………………………………………………………….....
3.0 Aims…………....………..…...…………………………………………………………
4.0 Theory…………………………………………………………………….………….
5.0 Apparatus………………………………………………..………………..…………..
6.0 Methodology/Procedure………………………………………………………….
7.0 Results/Data……………………...…………………………………………………...
8.0 Discussion…………………………………………………………………...……..
9.0 Conclusion…………………………………………………………………...…….….
10.0 Recommendations…………………………………………………………………….
11.0 Reference/Appendix……………………………………………………………....
1.0 ABSTRACT/SUMMARY
In this experiment, we will use UV-Vis and IR Spectroscopy to classify organic compounds. The
goal of this experiment is to put the skills learned in the first two chapters of the Instrumental
Chemistry course into practise. Using UV-Vis and infrared spectroscopy data,As a result, various
types of bonds can be formed with other carbon atoms as well as with other elements such as
hydrogen, oxygen, and nitrogen. UV-Vis and IR spectroscopy were used in this experiment to
determine and identify 6 compounds (A, C, D, I, J and K). UV-Vis Spectroscopy, also known as
Spectrophotometry, is a quantitative method for calculating how much light a specific molecule
absorbs, whereas infrared spectroscopy (IR spectroscopy) is the study of how a molecule responds
to infrared light.The unidentified compounds A, B, C, D, E, F, G, H, I, J, and K were successfully
identified as Ethanol, Acethyldehyde, Ethanoic Acid, Ethyl Ethanoate, Carbamide, Acetoxy
benzoic Acid, Benzene, Alkyl Benzene, Benzoic Acid, Ethyl Bromide, and Diethyl Ether during
this experiment. The experiment was a success because all of the compounds had already been
identified.
2.0 INTRODUCTION
The method allows for the measurement and recording of a sample's light
transmittance as the light travels through it. One of the most well-liked analytical methods is
ultraviolet-visible (UV-Vis) spectroscopy since it can identify almost all molecules and is
very adaptable. With UVVis spectroscopy, a sample is exposed to UV-Vis light, and the
amount of light transmitted through the sample is then measured [3]. The absorbance can be
determined using the transmittance (T), using the formula A = -log (T). It is possible to
obtain an absorbance spectrum that displays a compound's absorption at various wavelengths.
The chemical structure of the molecule determines how much absorption there is at any given
wavelength. By comparing the absorbance spectra of two compounds, UV-Vis can be used to
verify the identity of a chemical compound or to identify functional groups in a qualitative
way.
n – σ* Transitions
Saturated compounds having atoms with lone pairs (non-bonding electrons) can undergo n - *
transitions. These transitions need less energy than n - * transitions. They may be started by light
with a wavelength of 150 - 250 nm. There are few organic functional groups with - * peaks in the
UV region.
n - π* and π - π*Transitions
The greatest absorption spectroscopy of organic compounds that has occurred is the transition of n
or p electrons to the p* excited state. This is because the absorption peaks for these transitions are
located in an experimentally useful area of the spectrum (200 - 700 nm). Aside from that, these
transitions require an unsaturated group in the molecule to supply the p electrons.
Beer-Lambert Law is the fundamental concept behind absorbance in this spectroscopy (Equation1).
The symbol of A is the absorbance (it is unitless and commonly perceived as arbitrary units) for a
single wavelength, is the molar absorptivity of the substance or molecule in solution (M-1cm-1), b
is the path length of the cuvette or sample container (1 cm long), and c is the concentration of the
solution (M).
A = εbc
However, UV-Vis spectroscopy can only detect chromophores or the functional group of a
molecule. Fourier transform infrared spectroscopy (FT-IR) can detect a compound's unique
functional group. The unique chromophores may be identified in the 400 to 600 cm-1 range. This
FT-IR is an infrared spectroscopy method that may be used to detect organic and inorganic
substances, as well as solids, liquids, and gases.
5.0 APPARATUS
● UV-Vis Spectrometer
● FT-IR Spectrometer
● 6 unknown compounds labelled as A, C, D, I, J and K.
● Dropper
6.0 METHODOLOGY / PROCEDURE
UV-Vis
4. The type of instrument (high performance UV/VIS/NIR) is selected and Next is clicked,
followed by the wavelength for the instruments.
2. The wavelength is chosen based on the manual in the lab and ranges from 300 nm to 500 nm.
4. Accessed sample info, the quantity of sample was inserted, ENTER was pressed.
9. The sample container is filled with both a blank sample and an ethanol-containing sample.
10. Repeat steps 8 and 9 by replacing one of the ethanol samples with samples A, C, D, I, J and K.
11. The sample of solid I is diluted first before being put in the sample holder.
FTIR spectroscopy
3. Clicked at the background and ‘Ready for scan sample’ notice appeared.
4. Clicked at instrument, scan. The box appeared where the information needed to be keyed in.
5. Scan was clicked, and the range to run the sample is entered.
6. Apply was clicked and the sample was scanned when it was ready.
12. The top plate was cleaned with acetone after the first sample.
Result UV-VIS
1. Compound A
6. Compound K
APPENDIX 1
O-H
1. Compound A
2. Compound C
4. Compound I
6. Compound K
In these experiments two graphs are used to identify the functional group of 6 unknown
organic compounds which are A, C, D, I, J and K. The maximum value of characteristic absorption
(cm-1) from UV-Vis spectrum graphs as well as IR spectrum graphs will be compared to
Appendices 1 and 2 to determine the functional group for each compound.
For compound A, the UV-Vis spectrum graph shows that the main characteristic adsorption
are found to be 2942.79 cm-1 and 2832.05 cm-1 where it is lead to an alkyl C-H stretch since both
value fell in the range of 2950 - 2850 cm-1 region(from appendix 1). Additionally, the peak at
3321.09 cm-1 indicates that compound A is an alcohol O-H stretch since the range fell between
3550 - 3200 cm-1 (from appendixes 1). As the result perceived IR functional groups, compound A
was predicted to be ethanol. This is due to alcohol exhibiting distinct infrared absorptions
associated with both the O-H and C-O stretching vibrations. However, by using an IR spectrum
graph, compound A is found to be an alcohol group but it is a methanol. This is because the band is
2887.15 cm-1 where it falls in range for alkyl C-H stretch and 3309.28 cm-1 where it is an alcohol
O-H stretch. Therefore by using an IR spectrum graph, compound A still falls in an alcohol group
but the compound is methanol. However, compound A is expected to be methanol since FTIR
result is more accurate compared to UV-Vis result because there might be slightly human error
while running UV-Vis experiment such as not cleaning the cuvette with distilled water before put
into spectrophotometer compared to FTIR which the data obtained from software. Therefore,
compound A is alcohol which is methanol.
For compound C, the UV-Vis spectrum graph shows that the absorption peak is 1757.01
-1
cm where it is a carboxylic acid C=O stretch since it fell in the range between 1780 - 1710
cm-1(from appendix 1). Carboxylic acid O-H stretch also could be found on the compound C since
the adsorption peaks are found to be 2629.96 cm-1 and 3038.96 cm-1 where it falls in the range
3000 - 2500 cm-1(from appendix 1). Compound C was figured as carboxylic acid as in the
following result above. These compounds C can be claimed as carboxylic acid even more because
carboxylic acid can be produced through oxidation reaction of aldehyde with presence of CrO3 and
H3SO4. Therefore, compound C is found to be ethanoic acid by using UV-Vis graph. However, by
using an IR spectrum graph, the adsorption peak of compound C is 1704.55 cm-1 where it is an
Aldehyde C=O stretch. Therefore, by using IR spectrum graph compound C is found to be ethanal.
Since both graphs gives different functional group for compound C, we should consider compound
C as carbonyl group which is ethanal since FTIR result is more accurate compared to UV-Vis result
because there might be slightly human error while running UV-Vis experiment such as not cleaning
the cuvette with distilled water before put into spectrophotometer compared to FTIR which the data
obtained from software. Therefore, compound C is ethanal.
compound I we recognize it as Benzoic Acid, (𝐶6𝐻5𝐶𝑂𝑂𝐻 )by using UV-VIS spectrum and also
IR test. Firstly, the compound’s infrared spectrum indicates the main absorption are found at
−1
2976.76, 2871.70 and 1770.48 𝑐𝑚 for sample I. These lead to the stretches of O-H stretch,
aromatic C-H stretch and C=O stretch which are already shown on the table above. Based on the
result of UV-Vis, 1770.48 mm shows that the function group is carboxylic acid C=O stretch. Then,
−1
for the absorption peak of O-H stretch was 2976.76 𝑐𝑚 since the value fell in between the range
−1
of 3000 to 2500 𝑐𝑚 frequency range. Furthermore, since the range of aromatic C-H stretch is on
−1
~3300 and based on the result the close value of the range is on 2871.70 𝑐𝑚 . Moving on, based
−1
on result of IR which the value of 1633.36 𝑐𝑚 which indicates the value of carboxylic Acid
−1
C=O Stretch and as well as for aromatic C-H stretch the peak is at 3308.32 𝑐𝑚 ,based on the
appendix 1 this clearly shows the value is not in the range because of the error while running the
test.
As for compound J, by using UV-Vis spectrum graph the C-H stretch appears on peaks of
2927.50 cm-1, 2945 cm-1 and 2990.31 cm-1 where all 3 peaks fall in range between 2990.31 cm-1 -
2889.58 cm-1(from appendix 1) that represent the C-H stretch. Furthermore, the 558.78 cm-1 band
indicates the presence of a halogen compound which is C-Br since the absorption range for C-Br
stretch is 670 cm-1 - 515 cm-1(from appendix 1). Therefore, the functional group for compound J by
using UV-Vis spectrum graph is an alkyl halide group which is ethyl bromide. By using IR
spectrum graph, the functional group of compound J is still the same as the UV-Vis spectrum graph
which is ethyl bromide since the peak is 2869.76 cm-1 and 2978.72 cm-1 which is an alkyl C-H
stretch range. The absorption peak also falls in 845.17 cm-1 where it is in the range for C-Br stretch.
Therefore compound J is ethyl bromide since both UV-Vis spectrum graph and IR spectrum graph
of compound J give the same functional group for compound J.
Based on the UV-Vis spectrum graph of compound K, there is presence of C-H stretch for
this compound where the peak is at 2868.34 cm-1 and 2977.22 cm-1. Compound K also has the
presence of C-O-C stretch where the peak is ±1100.00 cm-1. As a result, by using the UV-Vis
spectrum graph, compound K is diethyl ether. The reaction information is used to strengthen this
claim where compound A which is ethanol reacts with compound J which is bromoethane to
undergo oxidation reaction with the presence of NaH and THF to synthesise an ester group which
is diethyl ether. Besides, by using an IR spectrum graph, there is a presence of alkyl C-H stretch
where the peaks fall on 2870.43 cm-1 and 2979.20 cm-1. There is also a presence of C-O-C stretch
in compound K by using IR spectrum graph since the peak is fall on ±1142.37 cm-1. Therefore, it is
confirmed that compound K is diethyl ether since both graphs show the same functional group.
9.0 CONCLUSION
Several recommendations may be taken from this experiment to enhance the results.
We have six undiscovered chemical substances in this experiment. As all the compounds are
colourless, it is important to identify the sample bottles to avoid associating incorrect
chemical information with the acquired results.
The cuvettes should be handled with care when performing the UV-Vis analysis. The
cuvette should be held on the line sides to ensure that there are no fingerprints on the clear
and smooth sides, which would most likely impact the resultant peak. The outside of the
cuvettes must also be lightly wiped with a tissue to eliminate any remains of spillage and to
avoid dirt on the clear and smooth surface. Furthermore, the result must be recorded
immediately after the reaction has occurred since some chemical compounds might corrode
the surface of the cuvette and affect the results.
It is important to wipe the sample plate with acetone before doing the IR analysis.
Since the plate is highly sensitive, it should be cleansed carefully with tissue to prevent
immediate damage. Dropping the samples onto the plate must be done quickly because if
done too slowly, the volatile sample would evaporate and the result will be inaccurate.
11.0 REFERENCE / APPENDIX
Alkyl C-H Stretch 2950 - 2850 (m or s) Alkane C-H bonds are fairly ubiquitous
Alkynyl C=C
Stretch 2260 - 2100 (v)
Amine N-H Stretch 3500 - 3300 (m) Primary amines produce two N-H stretch
absorptions, secondary amides only one,
and tertiary none.
Alkyl C-H Stretch 2950 - 2850 (m or s) Alkane C-H bonds are fairly ubiquitous
Alkynyl C=C
Stretch 2260 - 2100 (v)
Amine N-H Stretch 3500 - 3300 (m) Primary amines produce two N-H stretch
absorptions, secondary amides only one,
and tertiary none.
Nitrile C=N Stretch 2260 - 2220 (m)
Figure 11.2: using dropper to have the solutions inserted into the top plate.
Figure 11.3: handling the top plate.
Figure 11.6: inserting the unknown solution into the UV-VIS spectrophotometer.