EFDA - Guidelines For IVD Registration Requirements
EFDA - Guidelines For IVD Registration Requirements
EFDA - Guidelines For IVD Registration Requirements
December 2020
Addis Ababa, Ethiopia
Page 1 of 56
Table of Contents
Table of Contents...........................................................................................................................................................2
ACRONYMS ................................................................................................................................................................ 6
ACKNOWLEDGEMENT .............................................................................................................................................7
1. INTRODUCTION .................................................................................................................................................8
2. SCOPE ...................................................................................................................................................................9
3. PURPOSE.............................................................................................................................................................. 9
4. DEFINITIONS ......................................................................................................................................................9
5. GENERAL GUIDANCE AND PRINCIPLES .................................................................................................... 12
6. PRODUCT DOSSIER CLARITY AND COMPLETENESS.............................................................................. 14
7. APPLICABILITY OF SUPPORTING EVIDENCE TO THE PRODUCT UNDER REVIEW ......................... 14
CHAPTER ONE .......................................................................................................................................................... 17
ADMINISTRATIVE ................................................................................................................................................... 17
1.1. Cover Letter ............................................................................................................................................... 17
1.2. Agency Agreement..................................................................................................................................... 17
1.3. Evidence for registration application fee .................................................................................................... 17
1.4. Quality Management System, Full Quality System ................................................................................... 17
1.5. Good Manufacturing Practice (GMP) Certificates ..................................................................................... 17
1.6. Free Sale Certificate/Certificate of Marketing Authorization .................................................................... 18
1.7. Declarations of Conformity ........................................................................................................................ 18
CHAPTER TWO ......................................................................................................................................................... 19
SUBMISSION CONTEXT ......................................................................................................................................... 19
2.1. General Summary of Submission ............................................................................................................... 19
2.2. Device Description ..................................................................................................................................... 20
2.2.1. Comprehensive Device Description and Principle of Operation ........................................................... 20
2.2.2. Description of Device Packaging ........................................................................................................... 21
2.2.3. History of Development ......................................................................................................................... 22
2.2.4. Reference and Comparison to Similar and/or Previous Generations of the Device ............................... 22
2.2.5. Substantial Equivalence ......................................................................................................................... 22
2.3. Indications for Use and/or Intended Use and Contraindications ................................................................ 23
2.3.1. Intended Use; Intended Purpose; Intended User; Indications for Use ................................................... 23
2.3.2. Intended Environment/Setting for use ................................................................................................... 23
2.3.3. Pediatric Use .......................................................................................................................................... 24
2.3.4. Contraindications for Use ...................................................................................................................... 24
2.4. Global Market History ............................................................................................................................... 24
2.4.1. Global Market History ........................................................................................................................... 24
2.4.2. Global Incident Reports and Recalls ...................................................................................................... 25
Page 2 of 56
2.4.3. Sales, Incident and Recall Rates ............................................................................................................ 25
2.4.4. Evaluation/Inspection Reports ............................................................................................................... 26
CHAPTER THREE ..................................................................................................................................................... 27
ANALYTICAL PERFORMANCE AND OTHER EVIDENCES .............................................................................. 27
3.1. Risk Management ...................................................................................................................................... 27
3.2. Essential Principles (EP) Checklist ............................................................................................................ 28
3.3. Standards .................................................................................................................................................... 28
3.4. Analytical Performance .............................................................................................................................. 29
3.4.1. Stability of Specimen(s)......................................................................................................................... 29
3.4.2. Validation of Specimens ........................................................................................................................ 29
3.4.3. Metrological traceability of calibrator and control material values ....................................................... 30
3.4.4. Accuracy of Measurement ..................................................................................................................... 30
3.4.5. Analytical Sensitivity ............................................................................................................................. 31
3.4.6. Analytic Specificity ............................................................................................................................... 32
3.4.7. High dose hook effect ............................................................................................................................ 33
3.4.8. Measuring range of the assay ................................................................................................................. 33
3.4.9. Validation of assay cut off ..................................................................................................................... 34
3.4.10. Validation of the assay procedure ..................................................................................................... 34
3.5. Other Studies .............................................................................................................................................. 34
3.5.1. Electrical Systems: Safety, Mechanical and Environmental Protection, and Electromagnetic
Compatibility ....................................................................................................................................................... 34
3.5.2. Software/Firmware ................................................................................................................................ 35
3.5.2.1. Software/Firmware Description ........................................................................................................ 35
3.5.2.2. Hazard Analysis ................................................................................................................................ 35
3.5.2.3. Software Requirement Specification ................................................................................................. 36
3.5.2.4. Architecture Design Chart ................................................................................................................. 36
3.5.2.5. Software Design Specification .......................................................................................................... 36
3.5.2.6. Traceability Analysis ......................................................................................................................... 36
3.5.2.7. Software Life Cycle Process Description .......................................................................................... 36
3.5.2.8. Software Verification and Validation ................................................................................................ 36
3.5.2.9. Revision Level History and Unresolved Anomalies (Bugs or Defects) ............................................ 37
3.5.2.10. Cybersecurity ................................................................................................................................ 37
3.5.3. Cleaning and Disinfection Validation .................................................................................................... 37
3.5.4. Usability/Human factors ........................................................................................................................ 38
3.5.5. Stability of the IVD................................................................................................................................ 39
3.5.5.1. Claimed Shelf-life ............................................................................................................................. 40
3.5.5.2. In Use Stability .................................................................................................................................. 41
Page 3 of 56
3.5.5.3. Shipping Stability .............................................................................................................................. 41
CHAPTER FOUR ....................................................................................................................................................... 42
CLINICAL EVIDENCE .............................................................................................................................................. 42
4. Clinical Evidence ....................................................................................................................................... 42
4.1. Overall Summary of Clinical Evidence ................................................................................................. 42
4.2. Expected values/reference ranges .......................................................................................................... 42
4.3. IVD medical device specific clinical studies ......................................................................................... 42
4.4. Qualification of Usability ...................................................................................................................... 43
CHAPTER FIVE ......................................................................................................................................................... 44
LABELLING AND PROMOTIONAL MATERIAL .................................................................................................. 44
5. Labelling Requirements ............................................................................................................................. 44
5.1. Product/package labels .......................................................................................................................... 44
5.2. Package insert/ Instructions for use ....................................................................................................... 44
5.3. Technical/operators manual ................................................................................................................... 47
5.4. Other labelling and promotional materials ............................................................................................. 47
5.5. e-labeling ............................................................................................................................................... 47
CHAPTER 6A ............................................................................................................................................................. 48
QUALITY MANAGEMENT SYSTEM PROCEDURES .......................................................................................... 48
6A.1. Administrative ............................................................................................................................................... 48
6A.1.1. Product Descriptive Information ............................................................................................................. 48
6A.1.2. General Manufacturing Information ....................................................................................................... 48
6A.2. Quality management system procedures ........................................................................................................ 48
6A.3. Management responsibilities procedures ....................................................................................................... 48
6A.4. Resource management procedures ................................................................................................................. 49
6A.5. Product realization procedures ....................................................................................................................... 49
6A.5.1. Design and Development Procedures ..................................................................................................... 49
6A.5.2. Purchasing Procedures ............................................................................................................................ 49
6A.5.3. Production and service controls procedures ............................................................................................ 49
6A.5.4. Control of monitoring and measuring devices procedures ...................................................................... 50
6A.6. QMS measurement, analysis and improvement procedures .......................................................................... 50
CHAPTER 6B ............................................................................................................................................................. 51
QUALITY MANAGEMENT SYSTEM DEVICE SPECIFIC INFORMATION ....................................................... 51
6B.1. Quality management system information ...................................................................................................... 51
6B.2. Management responsibilities information ...................................................................................................... 51
6B.3. Resource management information ................................................................................................................ 51
6B.4. Product realization information ...................................................................................................................... 51
6B.4.1. Design and development information ..................................................................................................... 52
Page 4 of 56
6B.4.2. Purchasing information ........................................................................................................................... 52
6B.4.3. Production and service controls information........................................................................................... 52
6B.4.4. Control of monitoring and measuring devices information .................................................................... 52
6B.5. QMS measurement, analysis and improvement information ......................................................................... 53
ANNEX ONE .............................................................................................................................................................. 54
CHECKLIST FOR ESSENTIAL PRINCIPLES OF BASIC SAFETY AND ESSENTIAL PERFORMANCE OF
IVD MEDICAL DEVICES ......................................................................................................................................... 54
Page 5 of 56
ACRONYMS
EU European Union
NB Notified Body
Page 6 of 56
ACKNOWLEDGEMENT
Ethiopian Food and Drug Authority is very grateful to those who have participated directly or
indirectly in preparing this Guideline.
The Authority would like to express its special thanks and appreciation to the following experts
who have dedicated their precious time, energy and resources to contribute to develop this
Guideline from its drafting until its approval.
Kidanemariam Gebremichael (Senior Pharmacist), JSI Technical Advisor at EFDA
Keneni Benti (MSc in Biomedical Engineering), Medical Device Registration Dossier
Assessor at EFDA
Daniel Takele (BSc in Biomedical Engineering), Medical Device Registration Dossier
Assessor at EFDA
Solomon Shiferaw (Senior Pharmacist), Mckinsey Technical Advisor at EFDA
Page 7 of 56
1. INTRODUCTION
Ethiopian Food and Drugs Authority is the national regulatory authority responsible to
regulate food, medicines, medical devices, cosmetics and other health products mandated
by Ethiopian Food and Medicine Administration Proclamation No: 1112/19. The regulated
products including medical devices follow various regulatory processes from their
premarket assessment to post-market control in order to protect the public health by
ensuring the safety, performance and quality of the products. One of the EFDA’s premarket
assessments is to evaluate the product dossiers of the medical devices. This guideline is
therefore intended to provide a comprehensive and well-organized structure for premarket
In Vitro Diagnostic Medical device submissions that assists the manufacturers to submit
uniform registration dossier when registering their products.
This document provides harmonized, modular, format for use when filing In Vitro
Diagnostic medical device submissions to EFDA for market authorization.
The outline of documentation is to support a smooth documentation process. It remains the
applicant’s responsibility to ensure all regulatory requirements are met, and that clear and
transparent evidence of conformity to these requirements are provided.
Manufacturers of all classes of medical device are expected to demonstrate conformity of the
device to the Essential Principles of Safety and Performance (EPSP) through collection and
examination of evidence of conformity in technical documentation that shows how each
medical device was developed, designed and manufactured together with the descriptions
and explanations necessary to understand the manufacturer’s determination with respect to
such conformity. This technical documentation must be updated as necessary to reflect the
current status, specification and configuration of the device.
The evidence of conformity of medical device to the EPSP must be compiled and submitted
as per the requirements set in this guideline for the purpose of conformity assessment and
submission of application for medical device registration. The compiled documents need to
be kept in the premise for audit or inspection purposes.
The EFDA will monitor the use of these structures and work to continually improve the
documents at appropriate intervals based on sufficient use and experience.
Page 8 of 56
2. SCOPE
This Guideline is applicable for registration of In-vitro diagnostic Medical devices. This
guideline does not apply to medical devices listed under the “Guideline for Premarket
Notification of Low Risk Medical Devices”. The Medical devices other than In Vitro
diagnostic medical devices are also not in the scope of this guideline as these devices have
separate guideline stipulating requirements for their registration.
3. PURPOSE
4. DEFINITIONS
Page 9 of 56
changing the content of the data, and/or (b) communicate with each other, and/or (c) work
together as intended.
Cyber-security: Means a state where information and systems are protected from
unauthorized activities, such as access, use, disclosure, disruption, modification, or destruction
to a degree that the related risks to confidentiality, integrity, and availability are maintained at
an acceptable level throughout the life cycle.
Reprocessing- Means a process carried out on a used device in order to allow its safe reuse
including cleaning, disinfection, sterilization and related procedures, as well as testing and
restoring the technical and functional safety of the used device;
Conformity assessment- Means the process demonstrating whether the requirements of this
Regulation relating to a device have been fulfilled;
CE marking of conformity or ‘CE marking’- Means a marking by which a manufacturer
indicates that a device is in conformity with the applicable requirements set out in the
European Union Medical Devices Regulation and other applicable Union harmonization
legislation providing for its affixing;
Clinical evaluation- Means a systematic and planned process to continuously generate,
collect, analyze and assess the clinical data pertaining to a device in order to verify the safety
and performance, including clinical benefits, of the device when used as intended by the
manufacturer;
Clinical investigation- Means any systematic investigation involving one or more human
subjects, undertaken to assess the safety or performance of a device;
Clinical data- Means information concerning safety or performance that is generated from the
use of a device and is sourced from the following:
clinical investigation(s) of the device concerned,
clinical investigation(s) or other studies reported in scientific literature, of a device for
which equivalence to the device in question can be demonstrated,
reports published in peer reviewed scientific literature on other clinical experience of
either the device in question or a device for which equivalence to the device in question
can be demonstrated,
clinically relevant information coming from post-market surveillance, in particular the
post-market clinical follow-up;
Page 10 of 56
Clinical evidence- Means clinical data and clinical evaluation results pertaining to a device of a
sufficient amount and quality to allow a qualified assessment of whether the device is safe and
achieves the intended clinical benefit(s), when used as intended by the manufacturer;
Informed consent- Means a subject's free and voluntary expression of his or her willingness to
participate in a particular clinical investigation, after having been informed of all aspects of the
clinical investigation that are relevant to the subject's decision to participate or, in the case of
minors and of incapacitated subjects, an authorization or agreement from their legally designated
representative to include them in the clinical investigation;
Adverse event- Means any untoward medical occurrence, unintended disease or injury or any
untoward clinical signs, including an abnormal laboratory finding, in subjects, users or other
persons, in the context of a clinical investigation, whether or not related to the investigational
device;
Submission – Means a regulatory submission can be any type of information related to a
medical device regulatory process. This includes but is not limited to a request for
approval/authorization to market a device, any communications relating to the original
submission, and any request for modification to an existing approval.
Medical Devices other than In Vitro Diagnostic Devices: Means all medical devices including
Active implantable devices, Anesthetic and respiratory devices, Dental devices, Electro
mechanical medical devices, Hospital hardware, Non-active implantable devices, Ophthalmic
and optical devices, Reusable devices, Single use devices, Assistive products for persons with
disability, Diagnostic and therapeutic radiation devices, Complementary therapy devices, and
Biological-derived devices that do not take human samples and perform in vitro analysis for
screening, diagnosis or monitoring of diseases.
In vitro diagnostic devices: Means reagents, instruments, and systems intended for use in
diagnosis of disease or other conditions, including a determination of the state of health, in order
to cure, mitigate, treat, or prevent disease or its sequelae. Such products are intended for use in
the collection, preparation, and examination of specimens taken from the human body.
Authority- Means Ethiopian Food and Drug Authority
Page 11 of 56
5. GENERAL GUIDANCE AND PRINCIPLES
The content of this Guideline should be read in conjunction with relevant information described
in other existing International Medical Devices Regulators Forum (IMDRF) reference documents
and WHO IVD prequalification product dossier requirements.
The quality and performance of the intended product to be registered should not be inferior to the
available options.
Alternate approaches to the principles and practices described in this Guideline may be
acceptable provided they are supported by adequate scientific justification. It is important to note
that the Authority may request information or material, or define conditions not specifically
described in this guidance, in order to adequately assess the safety, performance, and quality of a
medical device prior to and after approval.
General format and guidance for preparation of dossiers
Well-organized and compiled documents will facilitate the evaluation process and decrease the
delay in the screening time. In contrast, poorly compiled documents may lead to unnecessary
loss of time, both for the applicant and the Authority. Therefore, documents should have
unambiguous contents: title, nature, and purpose should be clearly stated. They should be laid
out in an orderly fashion and be easy to check.
Formats of all the files to be submitted:
1. Paper size is A4; top, bottom, header, and footer margins are 12.5 mm; left and right
margins are 25mm
2. Single-spaced paragraphs
3. Times New Roman font, 12-point; letter space 0%.
4. The weight of the font should be legible when copied.
5. The attached data and documents should appear in the English language.
6. Any abbreviations should be clearly defined.
Applications submitted for registration will be screened chronologically by the date of
submission to the Authority, and the applicant will be notified of the evaluation results.
Some products are given assessment priority and they are listed in the “Medical Device Fast-
track Directive”.
Page 12 of 56
A request to add supplemental materials must be submitted within six months of being notified
of missing elements and/or clarification. If the supplemental submission is not implemented
within that time period, urge to be supplemented within 15 days shall follow.
If the supplemental document is not submitted within the urge period, or the contents of the
replenishment is inappropriate, the speculation shall be clarified and the document shall be
returned and/or rejected. However, if the applicant calls for an extension, the submission period
shall be determined based on the speculation.
The agent or the manufacturer should appoint a technical person who is able to understand this
and related guidelines of the Authority, who is familiar with the registration process of the
products, and who can freely communicate with the assessors should clarification be needed
(product-related or administrative) on queries raised by the Authority.
The Authority will not accept applications for registration by different applicants or local agents
for the same product manufactured by the same manufacturer and/or subsidiaries of one
manufacturer.
If the manufacturer of a medical device has one or more subsidiaries, the applicant is responsible
for submitting the technical documents of each specific product under registration from each of
the subsidiaries and/or from the site where the file is kept.
If the medical device use accessory and/or consumable (such as reagents, controls, etc.)
which is manufactured by a company‘s subsidiary, the free sale certificate should indicate the
same and/or a separate free sale certificate should be a part of the document.
If the medical device use accessory and/or consumable (such as reagents, controls, etc.) which is
manufactured by an independent manufacturer, information regarding the manufacturer and the
technical documents should be submitted. The authority will review on a case-by-case basis.
A medical device that an applicant has registered with the USFDA, Health Canada,
European Union, Ministry of Health, Labour and Welfare (Japan), Therapeutic Goods
Administration (Australia), or WHO Prequalification Programme is considered to be registered
with Stringent Regulatory Authority approved devices registration procedures.
All the required documents should be attached at its respective attachment spaces in PDF
format.
Page 13 of 56
The application shall be made online through the Authority’s regulated products registration
platform (www.eris.efda.gov.et ) after obtaining the username and password from the concerned
authorized person of the Registration department of the Authority.
Page 14 of 56
The analytical and clinical performance characteristics described in chapters 3 and 4 of this
guideline may not necessarily apply to all types of products submitted for registration
assessment.
For each performance study submitted in a product dossier, the following shall be provided:
Study Description: A description of the study that includes information to facilitate record
traceability: study identifier, product identifier (for example, lot numbers), IFU version
used, the date of initiation and the date of completion. All data shall be clearly labelled, and
clearly linked to the study report.
Study summary: A summary of the study findings including a conclusion that clarifies
how the study objectives have been met
Full study protocol and report: The study protocol and full report, which incorporates at
a minimum, the following information:
• study objectives, study design, the methodology used and data collected the site(s)
where the study was performed (for example, manufacturers R&D laboratory,
hospital laboratory, health care clinic)
• operator(s) of the assay
• the reference standard/method, if applicable
• specimen acceptance/selection criteria, specimen characterisation
• specimen type(s) (e.g. serum, plasma, finger stick whole blood, venous whole
blood) and numbers of each type
• actual test result summaries with their acceptance criteria and not just pass/fail
statements
• results that are reported in sufficient detail to allow the detection of potential
differences in performance between the conditions being investigated (e.g.
depending on the product this might require the use of either a semi-quantitative
scoring system or a calibrated, graduated colour chart to record line intensity)
• the numbers of invalid tests observed
• photographs of all test results, wherever possible
• details of statistical methods, estimations and calculations applied
• the study conclusion
Page 15 of 56
• when performed by a party other than the manufacturer, details of this third party
and the relationship to the manufacturer as well as copy of the contract between the
manufacturer and the third party identifying roles and responsibilities of each party
Page 16 of 56
CHAPTER ONE
ADMINISTRATIVE
1.1. Cover Letter
The cover letter should state applicant (Local agent), Manufacturer and License holder and/or
their authorized representative name and full address.
It should also state the type of registration (application), the common name of the device,
device trade name or proprietary name (both of the base device and a new name if one is given
to the new version/model of the device) and include the purpose of the application.
The cover letter should be stamped and signed by a person authorized by the applicant.
Page 17 of 56
1.6. Free Sale Certificate/Certificate of Marketing Authorization
Certificate issued by the National Regulatory Authority where the medical device is marketable,
attesting that the device is marketable, without any restriction at their jurisdiction shall be
submitted. This certificate shall indicate the name and full address of the manufacturer, the
name(s) of the device(s) (with model if applicable) and explains whether the products are freely
sold in the country of origin; if not, the reasons thereof should be clearly stated with appropriate
justification. If the manufacturer of the medical device has subsidiaries, a free sale certificate
should indicate the name and address of the subsidiaries with the name of the device they
manufacture, and/or a separate free sale certificate should be submitted for each subsidiary.
The certificate should be original, and valid.
In this section, the manufacturer shall provide (if applicable):
List the National Regulatory Authorities that have provided current regulatory approval
for the supply of this product in their country/region of authority
Provide details of the type of regulatory approval obtained from each National
Regulatory Authority
Provide current evidence of the regulatory approval, such as certificates provided by the
National Regulatory Authority.
o The evidence should clearly show that the product under assessment falls within
the scope of the submitted regulatory approval.
o Copies can be certified by a notary public. The manufacturer may be asked to
present the original copy at any time.
Information relating to export-only regulatory approvals should be clearly identifiable as
export-only approvals.
Page 18 of 56
c) Generic name, trade name and models (if applicable) of the device(s).
d) An appropriate conformity assessment procedure
The declaration must include the following information:
Date from which the Declaration of Conformity is valid;
Name and address of the device manufacturer; and,
The name, position, and signature of the responsible person who has been authorized to
complete the Declaration of Conformity on behalf of the manufacturer.
CHAPTER TWO
SUBMISSION CONTEXT
a) Statement of the device type (e.g. Tacrolimus test system, blood specimen collection device,
calibrator) and name (e.g. trade name, proprietary name), its general purpose, and a high-level
summary of key supporting evidence (i.e. studies that are unique to the risks of this device type)
b) Summary of submission, including
i. The type of submission (e.g. new, amendment, change of existing application, renewal);
ii. Any high-level background information or unusual details that the manufacturer wishes to
highlight in relation to the device, its history or relation to other approved devices or previous
submissions (provides context to submission).
Page 19 of 56
2.2. Device Description
2.2.1. Comprehensive Device Description and Principle of Operation
Page 20 of 56
f) Description of the accessories, other IVD or non-IVD medical devices and other products,
which are intended to be used in combination with the IVD medical device.
g) If approved by the regulator, provide the approval number and identification for each of the
accessories, other IVD or non-IVD medical devices and other products, which are intended to be
used in combination with the IVD medical device.
h) If applicable, indication of biological material or derivate used in the medical device,
including: origin (human, animal, recombinant or fermentation products or any other biological
material) and source (e.g. blood, bone, heart, any other tissue or cells). Where a significant risk is
identified, a brief summary of evaluations performed to minimize biological risks, in particular,
with regard to viruses and other transmissible agents.
i) If the device contains an active pharmaceutical ingredient (API) or drug, an indication of the
substance, should be provided. This should include its identity and source, and the intended
reason for its presence and its primary mode of action.
j) Description of the collection and/or transport container(s) provided with the IVD medical
device or a description of specifications or recommended collection and/or transport container(s).
k) If applicable, a listing of assays that are compatible with the instrument. l) If applicable, a
listing of compatible instruments.
m) A list of any software to be used with the IVD medical device and a description of its role in
the delivery of the intended purpose
NOTE: The sponsor/applicant should explicitly address any existing regional regulatory
guidance related to the comprehensive device description and principles of operations provided
in this section regarding the subject device.
a) Information regarding the packaging of the devices, including, when applicable, primary
packaging, secondary and any other packaging associated;
b) Specific packaging of accessories marketed together with the IVD medical devices shall also
be described;
c) If the user needs to package the IVD medical device or its accessories before they perform
sterilization, information about the correct packaging (e.g. material, composition, dimension)
should be provided.
Page 21 of 56
2.2.3. History of Development
For any device versions/prototypes referenced in the evidence presented in the submission, a
table describing the version/name, with 4 columns (Device Name and/or Version; Description of
changes from previous row; motivation for the change; list of verification/validation activities,
including clinical studies, conducted using this version).
For any design verification or validation activities presented in this submission (including
clinical studies) performed on any earlier versions of the subject device, include a justification
for why the changes do not impact the validity of the data collected under those activities in
supporting the safety and performance of the final IVD medical device design.
2.2.4. Reference and Comparison to Similar and/or Previous Generations of the Device
a) A list of the similar devices (available on local and international market) and/or previous
generation of the devices (if existent) relevant to the submission. This should include any
similar/previous generation devices that were previously reviewed and refused by the subject
regulator.
b) Description of why they were selected.
c) A key specification comparison, preferably in a table, between the references (similar and/or
previous generation) considered and the device.
Page 22 of 56
2.3. Indications for Use and/or Intended Use and Contraindications
2.3.1. Intended Use; Intended Purpose; Intended User; Indications for Use
a) Intended Use: The statement of intended use should specify the therapeutic or diagnostic
function provided by the device and may describe the medical procedure in which the device is
to be used.
b) Intended Purpose: What is expected with the use of this medical device? Which results are
expected?
c) Intended user and skills/knowledge/training that the user should have to operate or use the
device.
d) Identify if the device is intended for single or multiple use
e) Indications for Use:
i. Disease or medical condition that the device will diagnose, parameters to be monitored
and other considerations related to indication for use.
ii. If applicable, information about patient selection criteria.
iii. If applicable, information about intended patient population (e.g. adults, pediatrics or
newborn) or a statement that no subpopulations exist for the disease or condition for which the
device is intended.
NOTES:
i. The statements of intended use and purpose and the intended user and indications for use
must be as presented in the labeling.
ii. If more than one device is included, the information should be provided for each device.
Page 23 of 56
a) The setting where the device is intended to be used (e.g. domestic use, hospitals,
medical/clinical laboratories, ambulances, medical/dental offices). Multiple options can be
indicated.
b) If applicable, environmental conditions that can affect the device’s safety and/or
performance (e.g. temperature, humidity, power, pressure, movement).
a) Description of any pediatric subpopulations that suffer from the disease or condition that
the device is intended diagnose.
b) The number of affected pediatric patients, as a whole and within each pediatric
subpopulation.
OR
c) Statement that no pediatric subpopulation exists for the disease or condition for which the
device is intended.
If applicable, specify the disease or medical conditions that would make use of the device
inadvisable due to unfavorable risk/benefit profile.
NOTE: The statement if contraindications for the device must be as presented in the labeling.
a) Up to date indication of the markets (all countries or jurisdictions) where the device is
approved for marketing, including any marketing under compassionate use regulations.
b) Should include history of the marketing of the device by any other entity in as much
detail as possible, acknowledging that detailed information may not be available in all cases.
c) If the subject device is different in any way (e.g. design, labeling, specifications) from
those approved or marketed in other jurisdiction, the differences should be described.
Page 24 of 56
d) The month and year of market approval in each country or jurisdiction where the device
is marketed. If the device has been marketed for greater than 10 years, a statement of greater than
10 years can be made.
e) For each of the markets listed in (a) above, and statement of the commercial names used
in those markets OR a clear statement that the commercial names are the same in all
jurisdictions.
f) State the date of data capture for the market history data.
g) If the subject device has been the subject of any previous compassionate use and/or clinical
trials this should be identified and, if applicable, relevant reference numbers provided.
a) List adverse events/incidents associated with the device and a statement of the period
associated with this data.
b) If the number of adverse events is voluminous, provide a summary by event type that state
the number of reported events for each event type.
c) List of the medical device recalls and/or advisory notice, and a discussion of the handling
and solution given by the manufacturer in each case.
d) A description of any analysis and/or corrective actions undertaken in response to items
listed above.
a) A summary of the number of units sold in each country/region and a statement of the
period associated with this data.
b) Provide the rates calculated for each country/region, for example:
i. Incident rate = # adverse events/incidents divided by # units sold x 100
ii. Recall rate = # recalls divided by # units sold x 100
Rates may be presented in other appropriate units such as per patient year of use or per use. In
this case, methods for determining these rates should be presented and any assumptions
supported.
Page 25 of 56
c) Critical analyses of the rates calculated (e.g. why are they acceptable? How do they break
down in terms of incidents? Is there some outlier data that has driven the rates up? Are there any
trends associated with any sub-groups of the devices that are subject of the submission (e.g. size,
version)?).
NOTES
i. Sales in this context should be reported as the number of units sold.
ii. The summary of sales should be broken down by components when appropriate.
Copies of Evaluation/Inspection Reports from other parties (e.g. Notified Body inspection
reports).
Page 26 of 56
CHAPTER THREE
ANALYTICAL PERFORMANCE AND OTHER EVIDENCES
A risk analysis shall be undertaken to identify and address all known or foreseeable hazards
related to the product, taking into account such aspects as the user(s) of the device, and the
technology involved.
The information provided in this section shall contain:
A summary report of the risks identified during the risk analysis process, including, but
not limited to:
o Risk to the patient arising from false positive or false negative results
o Indirect risks that may result from product-associated hazards, such as instability,
which could lead to erroneous results
o The risk of delays in availability of results
o User-related hazards, such as reagents containing infectious agents
o Production-related risks
o Risks arising from the use of the product by users with minimal skills, training or
experience.
A description of how these risks have been controlled to an acceptable level. This may be
demonstrated by provision of documented evidence using risk assessment tools such as
failure modes and effects analysis (FMEA), failure reporting and corrective action system
(FRACAS), fault tree analysis (FTA) or other methods.
Measures to inform users of any residual risks.
A conclusion with evidence that the remaining risks are acceptable when compared to the
benefits. This conclusion shall be dated and signed by senior management.
Evidence that the risk analysis is part of the manufacturer's risk management plan
(inclusion of documented evidence in this regard).
Where a standard has been followed, identify the standard
Page 27 of 56
3.2. Essential Principles (EP) Checklist
a) An EP checklist (see Annex-I) established for the IVD medical devices, information about
method(s) used to demonstrate conformity with each EP that applies, references for the method
adopted and identification of the controlled document with evidence of conformity with each
method used.
b) For the controlled documents indicated which are required for inclusion in the submission: a
cross-reference of the location of such evidence within the submission.
c) If any EP indicated in the checklist does not apply to the device: a documented rationale of the
non-application of each EP that does not apply.
NOTE:
Methods used to demonstrate conformity may include one or more of the following:
a) Conformity with recognized or other standards;
b) Conformity with a commonly accepted industry test method(s);
c) Conformity with an in-house test method(s);
d) The evaluation of pre-clinical and clinical evidence;
e) Comparison to a similar device already available on the market.
3.3. Standards
a) List the standards that have been complied with in full or in part in the design and manufacture
of the device.
b) At a minimum should include the standard organization, standard number, standard title,
year/version, and if full or partial compliance.
c) If partial compliance, a list the sections of standard that
i. Are not applicable to the device, and/or
ii. Have been adapted, and/or
iii. Were deviated from for other reasons – discussion to accompany
Page 28 of 56
3.4. Analytical Performance
3.4.1. Stability of Specimen(s)
The manufacturer shall provide studies to support the stability, storage and where appropriate,
transport, of all specimen type(s) identified in the labelling, including any and all recommended
additives (e.g. anticoagulants). This information shall include:
Storage conditions (lower and upper limits of claimed temperature range, duration at each
temperature, variation in humidity, freeze/thaw cycles)
Transport conditions, where applicable
How the storage conditions, as well as the maximum allowable time between specimen
collection and its processing, or addition to the IVD, takes into consideration the settings
of intended use
Details of the specimen collection media, collection devices and transfer devices, whether
these contain anticoagulants and whether they can be sealed, if applicable.
The manufacturer shall provide studies to support the validity of specimen type(s) used in the
analytical and clinical studies as representative of all of the sample type(s) identified in the
labeling, including any and all recommended additives (e.g. anticoagulants), as well as contrived
specimens used in certain analytical studies. This should include:
a) A list of the specimen type(s) used, including any additives (e.g. anticoagulants), in each of
the analytical performance studies. If the same specimens are used for all analytical studies, this
can be stated and the specimen type identified.
b) For any or all of the analytical and clinical studies, if a particular specimen type(s) including
additives (e.g. anticoagulants), has been chosen as representative of other specimen types
identified in the labeling, this should be described and supported.
c) If the preparation of the specimen has not followed the protocol described in the current
labeling, this should be identified and validated.
d) A justification of the selection of the studies performed.
e) Provide summary of the evidence that falls within this category
Page 29 of 56
3.4.3. Metrological traceability of calibrator and control material values
The manufacturer shall provide evidences that support the metrological traceability of values
assigned to calibrators and trueness control materials. This should include:
a) A description of all calibrators and trueness control materials associated with the system.
b) A justification of the selection of the studies performed.
c) Provide summary of the evidence that falls within this category, including for example,
methods and acceptance criteria for the metrological traceability to reference materials and/or
reference measurement procedures and a description of value assignment and validation.
NOTE: Precision control materials used during analytical studies to establish the reproducibility
of a measurement procedure do not require the assessment of metrological traceability to a
reference material or a reference method.
3.4.4.1. Trueness
The manufacturer should provide a summary of information and evidence relating to the trueness
of the measurement procedure. Trueness measures apply to both quantitative and qualitative
assays only when a reference standard or method is available. This should include:
a) A rationale for the reference standard or method(s) used
b) A summary of the evidence that falls within this category
Page 30 of 56
The manufacturer shall provide evidence supporting claims for the precision of the product: i.e.
repeatability (e.g. within-run variability) and reproducibility (e.g. between-run, -lot, -day, -
operator, -site etc. variability, as appropriate). The studies provided in this section shall include:
Estimation of precision for each analyte for which detection is claimed. A justification
shall be given if this is not provided.
Testing in a panel of specimens that reflects the main specimen types intended for use
with the IVD.
For products that will be used at point-of-care, testing is likely to be undertaken by users who
represent a diversity of skills, training and experience. This section should include studies where
operator-to-operator variability has been investigated using representative of likely end-users of
the product (e.g., non-laboratory trained personnel: healthcare workers and trained lay providers)
using only those materials provided with the IVD (e.g. instructions for use, labels and other
instructional materials) who have undertaken the testing unassisted, following only those
instructions provided with the product. Personnel shall be selected who reflect the diversity of
intended users and operational settings so as to challenge the usability of the product.
The manufacturer shall provide evidence that demonstrates the analytical sensitivity of the
product. Analytical sensitivity shall be determined for each claimed variant, type and/or subtype,
where a suitable biological reference material exists. Depending on the intended use of the
product, this may include studies that establish limit of detection (LoD): the lowest concentration
of analyte (measurand) in a specimen that can be reliably detected.
The studies that establish analytical sensitivity shall include:
A description of specimen type and preparation, including the matrix used, the amount of
analyte in each specimen and how this was established. Analytical sensitivity shall be
demonstrated in a clinical sample matrix and shall use the entire assay system from
sample preparation to interpretation
The number of replicates tested at each concentration
A description of the calculation used to determine assay sensitivity
Page 31 of 56
3.4.6. Analytic Specificity
The manufacturer shall provide validated data that describes interference and cross reactivity
studies to determine the analytical specificity of the product. Analytical specificity is defined as
the ability of a measurement procedure to detect or measure only the analyte (measurand) to be
detected, in the presence of other substances/agents in the specimen.
Substances with the potential to cause interference or cross-reactivity will vary depending on the
assay type and design and may arise from either exogenous or endogenous sources. Typically,
interference and cross-reactivity studies involve adding the substance under evaluation to the
specimen and determining any bias of the test parameter relative to the control specimen to
which no such substance has been added.
Analytical specificity shall be evaluated in relation to the potential not only to cause false
positive results (using specimens that do not contain the analyte) but also to cause false negative
results (using specimens with the analyte at or added to a low level of reactivity on the product).
Substances for which the potential for interference or cross-reactivity can be reasonably expected
should be identified as part of a risk assessment for the product, taking into consideration the
populations and settings in which the product will be used.
Common interferants and cross-reacting substances/agents may include, as appropriate:
Substances used for patient treatment (e.g. therapeutic drugs, anticoagulants, etc.)
Substances ingested by the patient (e.g. over the counter medications, alcohol, vitamins,
foods, etc.)
substances added during specimen preparation (e.g. preservatives, stabilizers)
substances encountered in specific specimen types (e.g. haemoglobin, lipids, bilirubin,
proteins)
analytes of similar structure (e.g. precursors, metabolites) or medical conditions unrelated
to the test condition, including specimens negative for the assay but positive for a
condition that may mimic the test condition (e.g. for a hepatitis A assay: specimens
negative for hepatitis A virus, but positive for hepatitis B virus)
specimens from unrelated infections that cause false negative results
Page 32 of 56
substances used in, or related to, the design of the product: e.g. biotin and/or avidin;
interference from human antibodies to components of the vector used for expression of
product reagents
The manufacturer shall provide studies that evaluate the effects on the product of potentially
interfering and cross-reacting substances/agents. These studies shall include:
The substance/agent type, numbers of each corresponding specimen, and concentration
tested
Specimen type
Measurand (e.g. Analyte concentration)
Test results that are reported with respect to each condition (and analyte, as appropriate)
and not reported as an aggregate of the total number of specimens tested in the study
Evidence that any observed interference or cross-reactivity is reported as a limitation of
performance in the product IFU.
A study design that includes appropriate interferents and cross-reacting
substances/agents.
The manufacturer shall provide evidence that supports the absence of high-dose hook, or prozone
effects (if applicable). Specimens used to investigate high dose hook effect shall be chosen that
have a high analyte concentration, as determined using a method other than the product intended
to be prequalified. This second method shall be of a design not subject to prozoning.
The manufacturer shall provide information on studies that define the measuring range of the
assay (linear and non-linear measuring systems), including the lower and upper limits of
quantification (LLoQ and ULoQ), as appropriate, and describes information on how this has
been established. The extent of correlation of quantitation with a suitable reference test shall also
be determined.
In this section, the manufacturer shall provide the studies and information that allow an
understanding of the approach and its validity.
Page 33 of 56
3.4.9. Validation of assay cut off
The manufacturer shall provide an explanation, with supporting evidence describing how the
assay cut-off (or the algorithm/method for determining a cut-off for different assay runs) has
been established (if applicable). Depending on the intended use of the product, this may require
an explanation of the statistic approach (e.g. use of Receiver Operator Characteristics [ROC]
curve). This section shall include study description, study summary as well as full study and
report. For products that include the use of test reader, the way in which the reader has been
designed to differentiate reactive specimens from those that are non-reactive shall be
demonstrated.
The manufacturer shall provide a demonstration of how the assay procedure was validated, with
regard to important reaction conditions (e.g. reaction times, reaction temperature, reagent
volume, reading time,) and validation of controls (if applicable).
For example, for products where a reading interval is specified (i.e. time when result can first be
read; time beyond which result should not be read), validation of critical time points shall be
provided.
This section shall include study description, study summary as well as full study and report.
These studies may be conducted as part of investigations into the robustness of the product.
The manufacturer shall provide evidence supporting electrical safety, mechanical and
environmental protection, and electromagnetic compatibility.
This should include:
a) A justification of the selection of the studies performed.
b) A summary of the evidence that falls within this category
Page 34 of 56
c) A discussion and a conclusion to support why the evidence presented is sufficient to support
the application.
3.5.2. Software/Firmware
The manufacturer shall provide studies and supporting information on the software design,
development process and evidence of the validation of the software, as used in the finished IVD
medical device. It should also address all of the different hardware configurations and, where
applicable, operating systems identified in the labeling.
The Hazard Analysis should take into account all device hazards associated with the IVD
medical device’s intended use, including both hardware and software hazards.
NOTE:
i. This document can be in the form of an extract of the software-related items from
comprehensive risk management documentation, described in ISO 14971.
ii. Hazard analysis, should address all foreseeable hazards, including those resulting from
intentional or inadvertent misuse of the IVD medical device.
Page 35 of 56
3.5.2.3. Software Requirement Specification
The manufacturer shall provide the Software Requirements Specification (SRS) that documents
the requirements for the software. This typically includes functional, performance, interface,
design, developmental, and other requirements for the software. In effect, this document
describes what the Software Device is supposed to do. For example, hardware requirements,
programming language requirement, interface requirements, performance and functional
requirements,
The manufacturer shall provide the detailed depiction of functional units and software modules.
This may include state diagrams as well as flow charts.
The manufacturer shall provide the Software Design Specification (SDS) that describes the
implementation of the requirements for the Software Device. The SDS describes how the
requirements in the SRS are implemented.
The manufacturer shall provide a Traceability Analysis that links together your product design
requirements, design specifications, and testing requirements. It also provides a means of tying
together identified hazards with the implementation and testing of the mitigations.
The manufacturer shall provide a summary describing the software development life cycle and
the processes that are in place to manage the various life cycle activities.
Page 36 of 56
a) Include an overview of all verification, validation and testing performed both in-house and in
a simulated or actual user environment prior to final release.
b) Discussion to support why the evidence presented is sufficient to support the application.
3.5.2.10. Cybersecurity
The manufacturer should provide evidence to support the cybersecurity. These evidences
include, but not limited to:
a) Cybersecurity vulnerabilities and risks analysis
b) Cybersecurity controls measures
c) Traceability matrix linking cybersecurity controls to the cybersecurity vulnerabilities and risks
The manufacturer shall provide data that contains information on the validation of cleaning and
disinfection instructions for reusable devices, including evidence to support maintenance of
performance when subject to this procedure over a number of cycles that is representative of the
IVD medical device’s expected useful life. Information to be included in this section includes:
a) If applicable, a discussion of how the number of cycles that is representative of the IVD
medical device’s expected useful life has been determined.
b) A justification of the selection of the studies performed.
c) A summary of the evidence that falls within this category
NOTES:
This applies most typically in near patient testing involving whole blood.
Page 37 of 56
3.5.4. Usability/Human factors
The manufacturer shall provide studies that specifically assess the robustness of the product, a
term that denotes a product’s resilience to variations in either its environment or its usage.
Robustness (flex) studies consider the labelling and/or design of the product with respect to the
potential impact of human behaviour, abilities, limitations and other characteristics on the ability
of the product of fulfil its intended use. The impact of reagent variations is also considered.
Robustness studies should challenge the product under conditions of stress that allow an
understanding of any potential product deficiencies, including where and how a product might
fail. The manufacturer shall consider multiple skill levels of users, as well as potential instrument
and reagent problems.
Depending on the intended use of the product, the influence of the following factors should be
included in this section:
Operator error/ human factors, including use of incorrect specimen type, Incorrect
application of the specimen to the device (e.g., incorrect placement, incorrect volume),
incorrect handling of reagents including those in self-contained unitized test devices,
incorrect placement of device (e.g., non-level surface), incorrect placement of reagents,
including strips, or other components that contain reagent, use of incorrect reagents (for
example, reagents that are not specific for the particular device or lot or generic reagents),
incorrect order of reagent application, use of incorrect amount of reagent, incorrect
timing of procedures (e.g., specimen application, running the test, or reading results),
incorrect reading of test results, incorrect reading due to color blindness, etc..
Specimen integrity and handling including errors in specimen collection, clotted
specimens, error in specimen handling, incorrect specimen transport and/or storage,
presence of bubbles in the specimen etc.
Reagent integrity (Reagent viability) including use of improperly stored reagents, use of
outdated reagents, use of improperly mixed reagents, use of contaminated reagents, etc..
Hardware, software, and electronics integrity including power failure, power fluctuation,
incorrect voltage, repeated plugging and unplugging of the device, hardware failure,
software failure, electronic failure, physical trauma to unit, etc..
Page 38 of 56
Stability of calibration and internal controls including factors that affect calibrator and
calibration stability, factors that may interfere with calibration.
Environmental factors including impact of key environmental factors (temperature,
relative humidity, barometric pressure changes, altitude (if applicable), sunlight, surface
angle, device movement, etc.) on reagents, specimens, and test results, impact of key
environmental factors (including changes in parameters such as pH or temperature) etc.
In addition to the description, summary and detail and report, the studies in this section shall
include:
A summary of the evidence that falls within this category
Testing in specimens that represent all relevant test results and/or interpretations (e.g.
reactive and non-reactive; enzyme deficiency at key decision points, etc)
Details of the test environment and relation to the intended use environment
A discussion of what tests were considered for the device and why they were or were not
performed
A discussion to demonstrate why the evidence presented is sufficient to support the
application
Depending on the product (e.g. products likely to be used in point-of-care settings by users with
limited training, skills and/or experience) this section shall include complete studies that
demonstrate:
Label comprehension - Questionnaire-based testing of subjects representative of end
users undertaken to assess the ability of intended users to correctly comprehend key
messages from packaging and labelling.
Interpretation of results – Testing to assess the ability of subjects to correctly interpret
contrived test results.
If a clinical study has been conducted that includes usability/human factor endpoints (e.g. for
self-testing), reference to the studies and endpoints shall be made in this section but full results
do not need to be repeated here and shall be included in Chapter 4 – clinical evidence.
Page 39 of 56
This section shall describe in detail the claimed shelf life (including transport stability) as well as
in-use stability of a product.
Claims for stability shall be based on the second-last successful data point from the least stable
lot. For example: for testing conducted at 3, 6, 9, 12 and 15 months, if stability was observed at
15 months, then the maximum stability claim shall be 12 months.
Each of the studies referred to in this section for product stability must be presented with the
study description, protocol, full study report and conclusion. In addition to specifying acceptance
criteria, the study protocol must specify appropriate testing intervals and ensure that testing
extends beyond the projected claim of shelf life.
This section shall contain details and evidence supporting the claimed shelf-life of the IVD
medical device components (e.g. reagents, calibrators/reference materials, control material, any
other components susceptible to degradation).
Information provided in this section should include:
a) A description of recommended environmental conditions for storage of the IVD medical
device (e.g. temperature, pressure, humidity, light conditions).
b) A statement of the claimed shelf-life indicated as a period of time or any other means of
appropriate quantification.
c) An indication of the packaging used in any studies conducted in support of the shelf- life. If
the packaging used in the studies differs from the final device packaging, a discussion of why the
evidence can be considered valid in support of the claimed shelf life.
d) A description of the simulated transport conditions that the IVD was exposed to before the
start of shelf-life studies.
e) A justification of the selection of the studies performed.
f) A summary of the evidence that falls within this category
g) A discussion and a conclusion to support why the evidence presented is sufficient to support
the claimed shelf-life.
OR
h) A rationale that, for an indefinite period, the storage conditions could not affect IVD medical
device safety or performance.
Page 40 of 56
3.5.5.2. In Use Stability
This section shall contain details and evidence supporting the stability during actual routine use
of the IVD medical device (real or simulated), including all applicable components (e.g.
reagents, reaction cartridges). This may include open vial stability and/or, for automated
instruments, onboard stability. Information provided in this section should include:
a) A description of recommended environmental conditions for use of the IVD medical device
(e.g. temperature, pressure, humidity, light conditions).
b) A justification of the selection of the studies performed.
c) A summary of the evidence that falls within this category
d) A discussion and a conclusion to support why the evidence presented is sufficient to support
the application.
OR
e) A rationale that, for an indefinite period, the storage conditions could not affect IVD medical
device safety or performance.
This section shall contain details and evidence supporting the tolerance of IVD medical device,
or if provided separately, the components (e.g. reagents, calibrators/reference materials) to the
specified or expected shipping conditions. Information provided in this section should include:
a) An indication of environmental conditions for correct shipment of the IVD medical device
(temperature, pressure, humidity, light conditions, mechanical protection etc.).
b) A justification of the selection of the studies performed.
c) A summary of the evidence that falls within this category
d) A discussion and a conclusion to support why the evidence presented is sufficient to support
the application.
OR
e) A rationale that, for an indefinite period, the storage conditions could not affect IVD medical
device safety or performance.
Page 41 of 56
CHAPTER FOUR
CLINICAL EVIDENCE
4. Clinical Evidence
Clinical evaluation is the assessment and analysis of data generated from the clinical intended
use of the product in order to verify the clinical safety and performance of the device. Clinical
evidence is the combined information from the clinical data and its evaluation. A manufacturer
shall have clinical evidence to support any clinical claims. This shall include claims for clinical
or diagnostic sensitivity and specificity.
If applicable, the manufacturer shall provide in this section information on what values to expect
in healthy individuals versus those in individuals affected by a corresponding infection, disease
and/or condition.
All claims for the clinical performance of the product shall be supported by well-designed
performance evaluations. These may include evaluations that have been carried out or
coordinated by the manufacturer, as well as evaluations carried out by bodies wholly
independent of the manufacture. In addition to the detail description, full reports and
information, clinical evaluation studies provided in this section shall include:
Any anomalous results, or results that are not within predetermined specifications, shall
be clearly explained or justified. All invalid results shall be recorded and evaluated in
comparison to the reference result. Invalid results shall not be excluded from estimates of
sensitivity or specificity.
Estimates of diagnostic/clinical sensitivity and specificity shall be reported with 95%
confidence intervals
Page 42 of 56
Where an IVD is intended to detect multiple analytes without differentiating which
analyte is detected, specimens chosen for the testing panel shall comprise those that are
reactive only for each individual analyte.
Results shall be reported with respect to each study site and not be reported as an
aggregate of the total number of specimens tested to establish these characteristics
Details of the product lots/batches used for the evaluation, including lot number, date of
expiry, and the storage conditions of the product before and during the study
Details of the geographical region, clinical status, age and sex, as appropriate, of the
subjects from which specimens have been drawn for the clinical evaluation
Full details of the method used to select specimens for testing that would allow it to be
understood that selection biases have been either minimized or eliminated. This shall
include any acceptance/exclusion criteria as well as details of any specimens that were
excluded from selection using these criteria
Full details of the methods used to define the clinical status of the subjects and to
characterize the specimens
Evidence that the outcomes of the performance studies have been reviewed by the
manufacturer’s management and accepted for implementation.
All abbreviations used in reports and on data records shall be explained and clearly
defined
If the study has been published in peer-reviewed scientific literature, provide publication
details for the study
Testimonials from hospitals, laboratory staff, product users, patients, or testimonials of
any other kind are not considered to be evidence of performance. Testimonials shall not
be included in the dossier as they will not be considered during review
In this section, the manufacturer shall provide information that demonstrates the performance of
the product when used by observed, untrained self-testing users.
Page 43 of 56
CHAPTER FIVE
LABELLING AND PROMOTIONAL MATERIAL
5. Labelling Requirements
In this section, the manufacturer shall provide all packaging labels used in the product. This
includes primary and secondary labelling of all devices, accessories and components (but
exclusive of labels for shipping). Labels shall include at least the following information:
The product name and product identification number (product code/catalogue number)
The name and contact details of the manufacturer, or an authorized representative of the
manufacturer, on the outer package labels
The name of the reagent/ingredient
The expiry date (or a statement as to where and how this will be displayed)
An indication of any special storage and/or handling conditions that apply
The warnings and precautions
The lot/batch and/or serial number (or a statement as to where and how this will be
displayed)
The information regarding product conditions such as product sterility
The names of all included reagents in each box on the outer package label, where
possible
If a component is too small to contain all the above information, it shall at a minimum
contain the name, lot number expiration date, volume, and storage conditions.
If the product requires associated instrumentation, the requirements listed above also
apply to the instrument.
Page 44 of 56
In this section, the manufacturer shall provide the current product instructions for use (IFU). The
information provided in a product IFU shall be clear, correct, suitable for intended users and
consistent with that provided in the product dossier.
The product IFU shall at a minimum include the following information:
Product identification (name of the product and variants and corresponding product
codes)
A clearly stated intended use, including:
o what is detected by the assay (that is, the analytical use of the assay e.g. the
marker or nucleic acid sequence being detected)
o the clinical indication for the test (e.g. if it is for a specific disorder, or a condition
or risk factor of interest that the test is intended to detect, define or differentiate)
o the function of the product (screening, monitoring, diagnostic or aid to diagnosis,
staging or aid to staging of disease)
o the intended user (laboratory professional and/or at point-of-care)
o the intended testing population (e.g. neonates, antenatal women)
o the type of specimen(s) required (e.g. serum, plasma, whole blood, sputum, urine)
o whether the assay is automated
o what the instrument is intended for
o whether the test is qualitative or quantitative
o an indication that the product is for in vitro use
A general description of the principle of the assay method or instrument principles of
operation
A description of all components of the assay (e.g. reagents, assay controls and
calibrators) and a description of the reactive ingredients of relevant components (e.g.
antibodies, antigens, nucleic acid primers etc.)
A description of the specimen collection and transport materials provided with the
product or recommended for use
For instruments of automated assays: a description of the appropriate assay
characteristics or dedicated assays
For automated assays: a description of the appropriate instrumentation characteristics or
dedicated instrumentation
Page 45 of 56
If applicable, a description of any software to be used with the product
If applicable, a description or complete list of the various configurations/variants of
product that will be made available
If applicable, a description of the accessories, and other products that are intended to be
used in combination with the product but are not provided with the product
Storage conditions, including storage conditions and stability of both the unopened and
opened product, and working solutions. When applicable, these instructions shall include
such information as conditions of temperature, light, humidity, and other pertinent factors
Specimen exclusion criteria (e.g. specimens with visual evidence of hyperlipidaemia or
haemolysis, excessive specimen age, excessive number of freeze/thaw cycles).
If the test kit includes sterile accessories, an indication of that condition and any
necessary instructions in the event of damage to sterile packaging
If the test kit includes accessories that have been specified by the manufacturer as
intended for single-use only, an indication of that status
Clear instructions on how to perform the assay, including instructions on specimen
collection, handling, preparation and storage of reagents, the use of assay calibrators and
controls as well as the reading and interpretation of test results
Recommendations for quality control procedures
Clear instructions on the correct usage of any equipment or software that is required for
the performance of the assay
Any warning and precautions to be considered related to the use of the assay including
but not limited to interpreting the results, the disposal of the assay and/or its accessories
(e.g. lancets), to any consumables used with it (e.g. reagents) that may be carcinogenic,
mutagenic or toxic, or to any potentially infectious substances of human or animal origin
Any residual risks
Precautions and measures to be taken in the event of performance changes or product
malfunction
Limitations of the assay, including information on interfering substances that may affect
the performance of the assay
Performance characteristics including diagnostic sensitivity and specificity,
seroconversion sensitivity, accuracy, dynamic range, lower limit of detection, and
Page 46 of 56
reproducibility, as appropriate, and any other performance aspects that are relevant to the
product
Any requirements for special training or particular qualifications of the assay user
Any requirements for routine maintenance. Include details of frequency of maintenance
and who should perform this maintenance (for example: the user, a representative of the
manufacturer, or a third party)
Where relevant, a bibliography
Document control details, such as a document version number and release date
Definition of terms and abbreviations (if applicable)
The name and contact details of the manufacturer or an authorized representative of the
manufacturer, in order for the user to obtain assistance.
If the product requires associated instrumentation or the requested product for registration itself
is instrument, the manufacturer shall provide a copy of the instrument’s manual and/or associated
operator manuals.
Provide copies of any other instructional materials that are provided to the user such as job aids,
information resources on a website, CD-ROM etc.
5.5. e-labeling
For eligible medical devices and stand-alone software, the manufacturer shall identify which
form of e-labeling is being used in case of e-labeling (e.g. electronic storage system or built-in
system, website). The manufacturer shall also provide details of risk management in relation to
e-labeling. If this is part of the overall risk management, refer to it here.
A description of the procedure and operations on providing IFU's when requested.
The manufacturer also need to submit a written information for user Information on webpage
where IFU and further information can be found in relevant languages. Description on how the
requirements detailed for the website have been met should also be provided.
Page 47 of 56
CHAPTER 6A
QUALITY MANAGEMENT SYSTEM PROCEDURES
6A.1. Administrative
Administrative information needed to evaluate the premarket submission related to the QMS.
Abbreviated description of the device, operating principles and overall manufacturing methods
a) Address and contact information for all sites where the device or its components are
manufactured.
b) Where applicable, addresses for all critical subcontractors, such as outsourced production,
critical component or raw material production (e.g. animal tissue, drugs), and sterilization will
need to be provided.
High level quality management system procedures for establishing and maintaining the quality
management system such as the quality manual, quality policy, quality objectives, and control of
documents and records ISO 13485 Elements– SOPs to satisfy clause 4.
Procedures that document the management commitment to the establishment and maintenance of
the QMS by addressing quality policy, planning, responsibilities/authority/communication and
management review.
ISO 13485 Elements – SOPs implementing clause 5.
Page 48 of 56
6A.4. Resource management procedures
Procedures that document the adequate provision of resources to implement and maintain the
QMS including human resources, infrastructure and work environment. ISO 13485 Elements –
SOPs implementing clause 6.
High level product realization procedures such as those addressing planning and customer related
processes
ISO 13485 Elements – SOPs implementing sub clause 7.1 and 7.2.
Procedures that document the systematic and controlled development of the device design from
initiation of the project to transfer to production. ISO 13485 Elements – SOPs for implementing
sub clause 7.3.
This section should include the following information:
a) Design Control Procedure(s) b) Design & Development Planning Procedure(s) c) Design Input
Procedure(s) d) Design Output - Procedure(s) e) Design Review Procedure(s) f) Design
Verification Procedure(s) g) Design Validation Procedure(s) h) Risk Analysis Procedure(s) i)
Design Transfer Procedure(s) j) Design Changes Procedure(s) k) Design History File
Procedure(s).
Procedures that document that purchased products/services conform to established quality and/or
product specifications.
ISO 13485 Elements – SOPs to implement sub clause 7.4.
Procedures that document the production and service activities are carried out under controlled
conditions. These SOPS address issues such as cleanliness of product and contamination control;
Page 49 of 56
installation and servicing activities; process validation; identification and traceability; etc. ISO
13485 Elements – SOPs implementing sub clause 7.5.
Procedure that document that monitoring and measuring equipment used in the QMS is
controlled and continuously performing per the established requirements.
ISO 13485 Element- SOPs for implementing sub clause 7.6.
Procedures that document how monitoring, measurement, analysis and improvement to ensure
the conformity of the product and QMS, and to maintain the effectiveness of the QMS. ISO
13485 Element – SOPS for implementing clause 8.
Page 50 of 56
CHAPTER 6B
QUALITY MANAGEMENT SYSTEM DEVICE SPECIFIC
INFORMATION
Documentation and records specific to the subject device that results from the high level quality
management system procedures for establishing and maintaining the quality management system
such as the quality manual, quality policy, quality objectives, and control of documents, noted in
Chapter 6A.
ISO 13485 Elements – documentation specific to the subject device for the implementation of
clause 4.
Documentation and records specific to the subject device that result from the implementation the
management responsibilities procedures noted in Chapter 6A. ISO 13485 Elements –
documentation specific to the subject device for the implementation of clause 5.
Documentation and records specific to the subject device that result from the implementation the
resource management procedures noted in Chapter 6A.
ISO 13485 Elements – documentation specific to the subject device for the implementation of
clause 6.
Documentation and records specific to the subject device that results from the implementation of
the high level product realization procedures noted in Chapter 6A. ISO 13485 Elements –
documentation specific to the subject device for the implementation of sub clause 7.1 and 7.2.
Page 51 of 56
6B.4.1. Design and development information
Documentation and records specific to the subject device that results from the implementation of
the design and development procedures noted in Chapter 6A. NOTE: The source of this
information is the Design and Development Records (e.g. DHF - Design History File). ISO
13485 Elements – documentation specific to the subject device for the implementation of sub
clause 7.3.
Documentation and records specific to the subject device that results from the implementation of
purchasing procedures noted in Chapter 6A ISO 13485 Elements – documentation specific to the
subject device for the implementation of sub clause 7.4.
List of suppliers of goods or services that affect product conformity with requirements (critical
suppliers) and a description of how purchasing requirements are fulfilled for these suppliers.
Documentation and records specific to the subject device that results from the implementation of
the control of monitoring and measuring device procedures noted in Chapter 6A.
Page 52 of 56
ISO 13485 Elements – documentation specific to the subject device for the implementation of
sub clause 7.6.
Documentation and records specific to the subject device that results from the implementation of
the QMS measurement, analysis and improvement procedures noted in Chapter 6A. ISO 13485
Elements – documentation specific to the subject device for the implementation of clause 8.
Page 53 of 56
ANNEX ONE
Page 54 of 56
Ethiopian Food and Drug Authority, In Vitro Diagnostic Medical Device Registration
Requirements-
Checklist for Essential Principles of Safety and Performance Requirements
General Requirements
5.1.1 Medical devices and IVD medical devices should achieve the performance intended
by their manufacturer and should be designed and manufactured in such a way that,
during intended conditions of use, they are suitable for their intended purpose. They
should be safe and perform as intended, should have risks that are acceptable when
weighed against the benefits to the patient, and should not compromise the clinical
condition or the safety of patients, or the safety and health of users or, where applicable,
other persons.
5.1.2 Manufacturers should establish, implement, document and maintain a risk
management system to ensure the ongoing quality, safety and performance of the
medical device and IVD medical device. Risk management should be understood as a
continuous iterative process throughout the entire lifecycle of a medical device and IVD
medical device, requiring regular systematic updating. In carrying out risk management
manufacturers should:
a) establish and document a risk management plan covering each medical device and
IVD medical device;
b) identify and analyze the known and foreseeable hazards associated with each medical
device and IVD medical device;
c) estimate and evaluate the risks associated with, and occurring during, the intended use
and during reasonably foreseeable misuse;
d) eliminate or control the risks referred to in point (c) in accordance with the
requirements of points 5.1.3 and 5.1.4 below;
e) evaluate the impact of information from the production and postproduction phases, on
the overall risk, benefit-risk determination and risk acceptability. This evaluation should
Page 55 of 56
include the impact of the presence of previously unrecognized hazards or hazardous
situations, the acceptability of the estimated risk(s) arising from a hazardous situation,
and changes to the generally acknowledged state of the art.
f) based on the evaluation of the impact of the information referred to in point (e), if
necessary amend control measures in line with the requirements of points 5.1.3 and 5.1.4
below.
5.1.3 Risk control measures adopted by manufacturers for the design and manufacture of
the medical device and IVD medical device should conform to safety principles, taking
account of the generally acknowledged state of the art. When risk reduction is required,
manufacturers should control risks so that the residual risk associated with each hazard as
well as the overall residual risk is judged acceptable. In selecting the most appropriate
solutions, manufacturers should, in the following order of priority:
a) eliminate or appropriately reduce risks through safe design and manufacture;
b) where appropriate, take adequate protection measures, including alarms if necessary,
in relation to risks that cannot be eliminated; and
c) provide information for safety (warnings/precautions/contra-indications) and, where
appropriate, training to users.
5.1.4 The manufacturer should inform users of any relevant residual risks.
5.1.5 In eliminating or reducing risks related to use, the manufacturer should: a)
appropriately reduce the risks related to the features of the medical device and IVD
medical device and the environment in which the medical device and IVD medical
device are intended to be used (e.g. ergonomic/usability features, tolerance to dust and
humidity) and b) give consideration to the technical knowledge, experience, education,
training and use environment and, where applicable, the medical and physical conditions
of intended users.
5.1.6 ------------------------------- etc. --------------------------------------------
Page 56 of 56