Catalogue Cuvet Hellma

Download as pdf or txt
Download as pdf or txt
You are on page 1of 30

High Precision in Spectro-Optics

Product Catalog BestCellers


OPTICAL COMPONENTS FOR UV/VIS/NIR SPECTROSCOPY

CELLS
OPTICAL REFERENCE MATERIALS
TRAYCELL® MICRO VOLUME ANALYSIS
OPTICAL IMMERSION PROBES & MEASURING CELLS
MADE IN GERMANY

YOUR PROCESS. OUR SOLUTIONS.


As a leading supplier of high-precision, optical solutions that are 'Made in Germany' from glass,
quartz glass and synthetic crystals, Hellma has been a by-word for outstanding quality for more
than 95 years. A key supplier, the company is an integral part of its clients’ value chains. Reliability,
trust and continuity are inherent to Hellma’s work, and the company believes it has both a duty and
responsibility to ensure these principles are upheld. Clients worldwide put their confidence in
Hellma’s exceptional level of performance and problem-solving skills to meet and exceed regulatory
requirements and to make their products safer.

Hellma Analytics for


Optical components and
assemblies for use in
High-quality synthetic
crystals for use in the
Premium-quality preci-
sion optics for use in
OPTICAL ANALYSIS WITH
systems that serve fields of microlithography, laser technology as well THE HIGHEST PRECISION
analytical technology. optics, laser technology as all areas of photonics
and radiation detection. and the optical industry.

www.hellma-analytics.com www.hellma-materials.com www.hellma-optics.com

2 3
WHERE ONLY
THE BEST QUALITY
WILL DO

All-round expertise – from materials to Hellma Analytics products and


custom solutions solutions
From raw materials to component manufacturing This product catalog lists our com-
and from specific technological knowledge to prehensive range of products and
certifica­tion, Hellma offers a unique range of solutions – from cuvettes for spectro-
products and technologies for collecting exact and scopy and cytometry to micro volume
reliable results in the field of optical analysis. analysis technology, certified reference
Extensive engineering prowess, consulting skills and materials for spectroscopy, and
services complement our work. While our compre- optical immersion probes for use in
hensive portfolio means we have the right products at laboratories.
our fingertips to suit a wide variety of requirements,
we are, of course, also able to produce customized
solutions tailored precisely to your needs.
Please don’t hesitate to contact us – we’re always
happy to help.

HOW TO REACH US IMPORTANT:


Select your local distribution partner: Please check your order
www.hellma-analytics.com/contacts for completeness with regard
to the following points
Article number
Quantity needed
Transmission matched........yes/no
Polarimetric certification....yes/no
Antireflection or reflective
coatings, if required............yes/no

4 5
UNIQUE
TECHNOLOGICAL OPTICAL PATH LENGTH TOLERANCES

EXPERTISE FOR
Optical path length tolerance is a MATERIAL OPTICAL PATH LENGTH TOLERANCE
particularly important parameter
Quartz 0.01 mm to 0.05 mm ± 0.003 mm
for photometric applications be­-

EXACT MEASUREMENT
cause it influences the accuracy Quartz 0.1 mm to 0.2 mm ± 0.005 mm
of the measurements and preci- Quartz 0.5 mm to 20 mm ± 0.01 mm
sion of the results. These tight
Quartz 30 mm to 100 mm ± 0.02 mm

AND PRECISE
tolerances make Hellma Analytics
products ideally suited for coll- Special Optical Glass 0.1 mm to 10 mm ± 0.01 mm
ecting reliable and reproducible Special Optical Glass 20 mm to 100 mm ± 0.03 mm

RESULTS
results.
Optical Glass 10 mm to 20 mm ± 0.05 mm

Optical Glass 20 mm to 100 mm ± 0.1 mm


These optical path length tolerances apply to absorption cells.
For fluorescence cells, both for the direction of excitation and emission the tolerance is ± 0.05 mm.

MATERIAL CODES

Accurate measurements guaranteed An original Hellma Analytics MATERIAL MATERIAL CODE WAVELENGTH REFRACTIVE INDEX nd
product can be identified from the AT 588 nm
Founded on more than 95 years of experience in glass pro-
material code on each cuvette. Optical Glass ■ OG ■ 360 nm – 2500 nm 1,523
cessing, Hellma Analytics provides an impressive range of
services when­ever high-precision, innovative optics are Borosilicate Glass ■ BF ■ 330 nm – 2500 nm 1,473
needed for use in analysis. From proven, standard high- Special Optical Glass ■ OS ■ 320 nm – 2500 nm 1,523
precision products to complex, technologically advanced
Quartz Glass ■ UV ■ 260 nm – 2500 nm 1,458
custom-built designs, we offer our clients a comprehen­
sive range of services and solutions for collecting reliable Quartz Glass High Performance ■ QS ■ 200 nm – 2500 nm 1,458
and exact results. Quartz Glass Extended Range ■ QX ■ 200 nm – 3500 nm 1,458

Batch-produced OEM products


Besides our extensive collection of standard products, we TRANSMISSION OF EMPTY CELLS MADE OF DIFFERENT MATERIALS
also manufacture custom products according to client
specifications. Our state-of-the-art production facilities When comparing the transmis­ 100
Quartz Glass Extended Range
and in-depth specialist knowledge enable us to make the sion values of various cuvettes, 90 Quartz Glass
seemingly impossible – possible. We are always on hand to it is essential to ensure that the 80
High Performance

provide detailed, expert advice to help make your ideas a same measurement conditions 70
reality.

Transmission (%)
Quartz Glass
are in place. Please note that the 60
Please don’t hesitate to contact us. transmission curves shown were
Special
Optical Glass
50
plotted using measurements Optical
Glass
40
taken from empty cuvettes
(2 windows). 30

20
Borosilicate
Glass
10

Wavelength 150 200 250 300 400 2000 3000 4000 5000
Wavenumber (cm-1) 66667 50000 40000 33333 25000 5000 3333 2500 2000

I N F O R M A TI O N A B O U T
TH E M A TE R I A L S
www.hellma.com/technical-info

6 7
CELLS

CELLS

FOR RELIABLE MEASUREMENTS


COLLECTED WITH UTMOST
PRECISION

Cuvettes for absorbance and fluorescence ADVANTAGES


measurements
● Extremely high parallelism of the windows with a


Hellma Analytics produces a wide range of cuvettes for maximum tolerance of ± 0.01 mm
use in spectroscopy and cytometry with optical path ● Exceptional optical path length accuracy down to
lengths spanning 0.01 mm to 100 mm and above. Thanks 0.003 mm (3 µm) for high dimensional accuracy and
to their stability, maximum precision and reliability when reproducible measurements results
HIGH-PRECISION QUARTZ CUVETTE, TYPE 100-QS used for absorbance and fluorescence measurements,
Hellma cuvettes work exceptionally well in a wide ● Unique surface flatness of the optical windows of
Tried and tested over decades, used in countless applications 0.001 mm (1 µm)
range of areas in the lab. With a surface flatness of
Utmost accuracy in terms of optical path length and parallelism 1 μm, our quartz windows set a benchmark in cuvette ● Very high temperature stability and chemical
production. resistance due to thermal bonding of individual
Very high temperature resistance
What’s more, the function-optimized design with beveled components (effectively monolithic)
Very high chemical resistance
edges and corners protects against the risk of damage ● Guaranteed transmission of at least 82% from
Outstanding measurement reproducibility caused by splitting and assists users in their daily work. 200 nm to 3500 nm depending on the material
On request, we are able to produce custom models
designed for specific areas of application.

Measuring of cuvette transmission

If required, cuvettes can be spectrally calibrated into sets


with equal transmission values (measurement uncertainty C U V E TTE F I N D E R
±1%). These cuvettes are given a three-digit calibration If you cannot find the right product in
code containing coded data about the material and the the following selection, please use our
transmission at a wavelength typical for the cuvette online cuvette finder.
material.
www.hellma.com/cuvette-finder

Polarimetric checking of cuvettes


Developed using our specialist
expertise, Hellma cuvettes stand Cuvettes with an inside width greater than 5 mm can be
polari­metrically checked on request. They are marked
out for their excellent quality. with a “P” and delivered with a test certificate confirming
Thomas Brenn, that the rotation of the polarization plane does not exceed
Product Manager Cuvettes 0.01 degrees.

8 9
CELLS

UV/VIS/NIR SPECTROSCOPY
PERFECTION IN DETAIL DECISIVE STRENGTHS OF HELLMA ANALYTICS CELLS
Cuvettes are not all the same, even if they sometimes appear identical. The difference lies in the details and can be
crucial for measurement results. Take our cuvette windows, for example, which boast outstanding quality and a flat-
ness tolerance of better than 0.001 mm (1 μm). The parallelism of both window surfaces relative to one another is just
as important. Our high-precision production guarantees that the frontal deformation of the light wave passing through
a cuvette window is less than 4 lambda, which works out at approximately 0.001 mm (1 μm) if lambda = 546 nm. The
high level of flatness demonstrates that the cuvettes from Hellma Analytics are setting standards. Overall, an ideal
foundation for conducting reliable, reproducible and exact measurements.

Figure 1: Measurement of the flatness of a Hellma cuvette - the frontal deformation of the wave is extremely low.

TRUST THE ORIGINAL

Hellma Analytics produces cuvettes in


unrivaled high quality that are used in Figure 2: Measurement of the flatness of a competitor's cuvette - the frontal deformation of the wave is more than
absorbance measurements, fluorescence twice that of a Hellma cuvette.
measurements and special applications
such as cytometry, light scattering or
reflection measurements and guarantee
precise reproducible results.
Our comprehensive range of products
and support provides the right solution
for almost every requirement.

10 11
CELLS

ABSORPTION CELLS

MACRO CELLS MACRO CELLS

with PTFE lid or stopper with PTFE lid or stopper


TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. ARTICLE NO. REMARKS TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. ARTICLE NO. REMARKS
MATERIAL LENGTH H×W×D mm mm µl MATERIAL LENGTH H×W×D mm mm µl
mm mm mm mm
100-OS 1 45 x 12.5 x 3.5 9.5 1.5 350 100-1-20 glass lid
402.000-OG 10 40 x 23.6 x 15 18.5 2.5 6000 402-10-10
2 45 x 12.5 x 4.5 9.5 1.5 700 100-2-20 glass lid
50 40 x 23.6 x 55 18.5 2.5 30000 402-50-10
5 45 x 12.5 x 7.5 9.5 1.5 1750 100-5-20
10 45 x 12.5 x 12.5 9.5 1.5 3500 100-10-20 404.000-QX 1 47.5 x 23.6 x 7.5 18.5 2.5 700 404-1-46
20 45 x 12.5 x 22.5 9.5 1.5 7000 100-20-20 2 47.5 x 23.6 x 7.5 18.5 2.5 1400 404-2-46
40 45 x 12.5 x 42.5 9.5 1.5 14000 100-40-20 10 47.5 x 23.6 x 12.5 18.5 2.5 7000 404-10-46
50 45 x 12.5 x 52.5 9.5 1.5 17500 100-50-20
100 45 x 12.5 x 102.5 9.5 1.5 35000 100-100-20 glass lid 6030-OG 10 45 x 12.5 x 12.5 9.5 1.5 3500 6030-10-10
20 45 x 12.5 x 22.5 9.5 1.5 7000 6030-20-10
100-QS 1 45 x 12.5 x 3.5 9.5 1.5 350 100-1-40 glass lid 40 45 x 12.5 x 42.5 9.5 1.5 14000 6030-40-10
2 45 x 12.5 x 4.5 9.5 1.5 700 100-2-40 glass lid
50 45 x 12.5 x 52.5 9.5 1.5 17500 6030-50-10
5 45 x 12.5 x 7.5 9.5 1.5 1750 100-5-40
10 45 x 12.5 x 12.5 9.5 1.5 3500 100-10-40 6030-UV 10 (± 0.05) 45 x 12.5 x 12.5 9.5 1.5 3500 6030-UV-10-531
20 45 x 12.5 x 22.5 9.5 1.5 7000 100-20-40
40 45 x 12.5 x 42.5 9.5 1.5 14000 100-40-40
50 45 x 12.5 x 52.5 9.5 1.5 17500 100-50-40
100 45 x 12.5 x 102.5 9.5 1.5 35000 100-100-40 glass lid
SEMI-MICRO CELLS
100-QX 1 45 x 12.5 x 3.5 9.5 1.5 350 100-1-46 glass lid
2 45 x 12.5 x 4.5 9.5 1.5 700 100-2-46 glass lid
5 45 x 12.5 x 7.5 9.5 1.5 1750 100-5-46 with PTFE lid or stopper
10 45 x 12.5 x 12.5 9.5 1.5 3500 100-10-46
50 45 x 12.5 x 52.5 9.5 1.5 17500 100-50-46 TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. ARTICLE NO. REMARKS
100 45 x 12.5 x 102.5 9.5 1.5 35000 100-100-46 glass lid MATERIAL LENGTH H×W×D mm mm µl
mm mm
110-OS 1 52 x 12.5 x 3.5 9.5 1.5 350 110-1-20
2 52 x 12.5 x 4.5 9.5 1.5 700 110-2-20 6040-OG 10 45 x 12.5 x 12.5 4 3.2 1400 6040-10-10
5 46 x 12.5 x 7.5 9.5 1.5 1750 110-5-20
10 46 x 12.5 x 12.5 9.5 1.5 3500 110-10-20 6040-UV 10 (± 0.05) 45 x 12.5 x 12.5 4 3.2 1400 6040-UV-10-531
50 46 x 12.5 x 52.5 9.5 1.5 17500 110-50-20 with 2 stoppers
104-OS 10 45 x 12.5 x 12.5 4 3.2 1400 104-10-20
110-QS 1 52 x 12.5 x 3.5 9.5 1.5 350 110-1-40 50 45 x 12.5 x 52.5 4 3.2 7000 104-50-20
2 52 x 12.5 x 4.5 9.5 1.5 700 110-2-40 104-QS 5 45 x 12.5 x 7.5 4 3.2 700 104-5-40
5 46 x 12.5 x 7.5 9.5 1.5 1750 110-5-40 10 45 x 12.5 x 12.5 4 3.2 1400 104-10-40
10 46 x 12.5 x 12.5 9.5 1.5 3500 110-10-40 20 45 x 12.5 x 22.5 4 3.2 2800 104-20-40
20 46 x 12.5 x 22.5 9.5 1.5 7000 110-20-40 50 45 x 12.5 x 52.5 4 3.2 7000 104-50-40
40 46 x 12.5 x 42.5 9.5 1.5 14000 110-40-40 from 40 mm
50 46 x 12.5 x 52.5 9.5 1.5 17500 110-50-40 with 2 stoppers
104-QX 10 45 x 12.5 x 12.5 4 3.2 1400 104-10-46
100 46 x 12.5 x 102.5 9.5 1.5 35000 110-100-40
110-QX 1 52 x 12.5 x 3.5 9.5 1.5 350 110-1-46 WINDOW MATERIAL
5 46 x 12.5 x 7.5 9.5 1.5 1750 110-5-46 OG Optical Glass 360 nm – 2500 nm QS Quartz Glass High Performance 200 nm – 2500 nm
10 46 x 12.5 x 12.5 9.5 1.5 3500 110-10-46 OS Special Optical Glass 320 nm – 2500 nm QX Quartz Glass Extended Range 200 nm – 3500 nm
UV Quartz Glass 260 nm – 2500 nm
WINDOW MATERIAL
OG Optical Glass 360 nm – 2500 nm QS Quartz Glass High Performance 200 nm – 2500 nm
OS Special Optical Glass 320 nm – 2500 nm QX Quartz Glass Extended Range 200 nm – 3500 nm

100 100 100 110 110 110 402.000 404.000 6030 6030-UV 6040 6040-UV 104
10 mm 50 mm 100 mm 10 mm 50 mm 100 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm

12 13
CELLS

ABSORPTION CELLS

SEMI-MICRO CELLS ULTRA-MICRO CELLS

with PTFE lid or stopper with PE stopper or open with pipette tips
TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. ARTICLE NO. REMARKS TYPE/WINDOW OPTICAL PATH CENTER OUTSIDE DIM. APERTURE CHAMBER FILLING ARTICLE NO. REMARKS
MATERIAL LENGTH H×W×D mm mm µl MATERIAL LENGTH HEIGHT H×W×D HxW VOL. VOL.
mm mm mm mm mm mm µl µl
104B-QS 10 45 x 12.5 x 12.5 4 3.2 1400 104-B-10-40 black side walls and 105.200-QS 10 15 45 x 12.5 x 12.5 8 x 2 160 180 105-200-15-40
base 10 8.5 45 x 12.5 x 12.5 8 x 2 160 180 105-200-85-40
108-QS 10 45 x 12.5 x 12.5 4 9 1000 108-000-10-40 105.201-QS 10 15 45 x 12.5 x 12.5 5 x 2 100 120 105-201-15-40
10 8.5 45 x 12.5 x 12.5 5 x 2 100 120 105-201-85-40
108B-QS 10 45 x 12.5 x 12.5 4 9 1000 108B-10-40 black side walls and
base
105.202-QS 10 15 45 x 12.5 x 12.5 2.5 x 2 50 70 105-202-15-40
114-OS 10 46 x 12.5 x 12.5 4 3.2 1400 114-10-20 10 8.5 45 x 12.5 x 12.5 2.5 x 2 50 70 105-202-85-40

114-QS 10 46 x 12.5 x 12.5 4 3.2 1400 114-10-40 105.210-QS 10 15 40 x 12.5 x 12.5  0.8 5 10 1052101015-40
10 8.5 40 x 12.5 x 12.5  0.8 5 10 1052101085-40
114B-QS 10 46 x 12.5 x 12.5 4 3.2 1400 114B-10-40 black side walls and
base

TRAYCELL 2.0 FOR MICRO VOLUME ANALYSIS


MICRO CELLS More information: www.hellma.com/traycell

with PTFE lid or stopper


TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. ARTICLE NO. REMARKS DEMOUNTABLE CELLS WITH SMALL VOLUME
MATERIAL LENGTH H×W×D mm mm µl
mm mm
TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. THICKNESS INSIDE WIDTH VOL. ARTICLE NO. REMARKS
104.002-OS 10 45 x 12.5 x 12.5 2 3.2 700 104-002-10-20 MATERIAL LENGTH H×W×D mm mm µl
mm mm
104.002-QS 10 45 x 12.5 x 12.5 2 3.2 700 104-002-10-40
106-QS 0.01 ± 0.003 45 x 12.5 2.5 9 2.6 106-0.01-40 demountable rectangular cells
104.002B-OS 10 45 x 12.5 x 12.5 2 3.2 700 104002B-10-20 black side walls and 0.1 ± 0.005 45 x 12.5 2.6 9 26 106-0.10-40 Please order cell holder separately –
base see article no. 013-000-71
0.2 ± 0.005 45 x 12.5 2.7 9 52 106-0.20-40
0.5 ± 0.010 45 x 12.5 3 9 130 106-0.50-40
104.002B-QS 10 45 x 12.5 x 12.5 2 3.2 700 104002B-10-40 black side walls and
base
013.000 55 x 12.5 x 12.5 013-000-71 cell holder for cell type 106
105-QS 10 25 x 12.5 x 12.5 2 1.5 300 105-10-40

108.002-QS 10 45 x 12.5 x 12.5 2 9 500 108-002-10-40 WINDOW MATERIAL


QS Quartz Glass High Performance 200 nm – 2500 nm
108.002B-QS 10 45 x 12.5 x 12.5 2 9 500 108002B-10-40 black side walls and
base

115-QS 10 40 x 12.5 x 12.5 2 1.25 400 115-10-40

115B-QS 10 40 x 12.5 x 12.5 2 1.25 400 115B-10-40 black side walls and
base

WINDOW MATERIAL
OS Special Optical Glass 320 nm – 2500 nm QS Quartz Glass High Performance 200 nm – 2500 nm

104B 108 108B 114 114B 104.002 104.002B 105 108.002 108.002B 115 115B 105.200 105.201 105.202 105.210 105.830-UVS 106 013.000
10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm

14 15
CELLS

ABSORPTION CELLS

CELLS FOR MAGNETIC STIRRERS CYLINDRICAL CELLS

macro, semi-micro, with PTFE lid or stopper macro, with PTFE stopper
TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. ARTICLE NO. REMARKS TYPE/WINDOW OPTICAL PATH OUTSIDE- INSIDE- OUTSIDE VOL. ARTICLE NO. REMARKS
MATERIAL LENGTH H×W×D mm mm µl MATERIAL LENGTH DIAMETER DIAMETER DEPTH µl
mm mm mm mm mm mm
109.000-QS 10 45 x 12.5 x 12.5 9.5 5 3500 109-000-10-40 120-OS 50 22 19 52.5 14000 120-50-20 from 50 mm
100 22 19 102.5 28000 120-100-20 with 2 stoppers
109.004-QS 10 45 x 12.5 x 12.5 4 5 1500 109-004-10-40
120-QS 1 22 19 3.5 280 120-000-1-40
119.000-QS 10 49.5 x 12.5 x 12.5 9.5 5 3500 119-10-40 2 22 19 4.5 560 120-000-2-40
5 22 19 7.5 1400 120-5-40
119.004-QS 10 49.5 x 12.5 x 12.5 4 5 1500 119-004-10-40 10 22 19 12.5 2800 120-10-40
20 22 19 22.5 5600 120-20-40
332.300 6x3 332-300-VE10 10-pack
50 22 19 52.5 14000 120-50-40 from 50 mm
PTFE coated with 2 stoppers
magnetic stir bar 100 22 19 102.5 28000 120-100-40

120-QX 10 22 19 12.5 2800 120-10-46

SEALABLE CELLS 121.000-QS 0.1 22 13 20 160 121-0.10-40


0.2 22 13 20 170 121-0.20-40
0.5 22 13 20 210 121-0.50-40
2 ports
macro, semi-micro, for anaerobic applications 1 22 13 20 280 121-1-40 and stoppers
(with ISO thread GL 14 and screw cap with silicone rubber seal, accessories see page 33)
225-BF 10 22 18 12.5 2600 225-10-23 Cell with tube
Tube: Duran
TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. ARTICLE NO. REMARKS
MATERIAL LENGTH H×W×D mm mm µl
165-QS 10 22 10 12.5 800 165-10-40 Temperature controlled
mm mm cell
1 stopper and
117.100-QS 10 56 x 12.5 x 12.5 9.5 1.5 3500 117-100-10-40 Open screw cap
2 thermostat ports
117.200-QS 10 56 x 12.5 x 12.5 9.5 1.5 3500 117-200-10-40 Closed screw cap

117.104-QS 10 56 x 12.5 x 12.5 4 1.25 1400 117-104-10-40 Open screw cap


DEMOUNTABLE CELLS WITH SMALL VOLUME
117.204-QS 10 56 x 12.5 x 12.5 4 1.25 1400 117-204-10-40 Closed screw cap

TYPE OPTICAL PATH OUTSIDE- THICKNESS INSIDE- VOL. ARTICLE NO. REMARKS
LENGTH DIAMETER mm DIAMETER µl
CELLS WITH TUBES mm mm mm
124-QS 0.01 ± 0.003 22 2.51 15 2 124-0.01-40 demountable circular cell
0.1 ± 0.005 22 2.6 15 18 124-0.1-40 Please order cell holder separately!
macro, tube  8 mm, tube length 80 mm 0.2 ± 0.005 22 2.7 15 35 124-0.2-40 Article No.: 020-001-761
0.5 ± 0.005 22 3 15 85 124-0.5-40
TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. ARTICLE NO. REMARKS
MATERIAL LENGTH HxWxD mm mm µl 020.001 0.01 – 1 27 x 23.5 x 11.5 020-001-761 cell holder for cell type 124 and 201/202
mm mm
202-QS 1.25 22 202-40 circular window made of Quartz Glass High Performance
220-QS 10 40 x 12.5 x 12.5 9.5 1.5 3500 220-10-40
®
Quartz/DURAN tube
202-QX 1.25 22 202-46 circular window made of Quartz Glass Extended Range

WINDOW MATERIAL 201 1 ± 0.01 21 201-1-23 ring made of Duran for cell holder 020.001
QS Quartz Glass High Performance 200 nm – 2500 nm

WINDOW MATERIAL
OS Special Optical Glass 320 nm – 2500 nm QS Quartz Glass High Performance 200 nm – 2500 nm
BF Borosilicate Glass 330 nm – 2500 nm QX Quartz Glass Extended Range 200 nm – 3500 nm

109.000 109.004 119.000 119.004 117.100 117.104 220 225 120 120 120 121.000 165
10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 10 mm 1 mm 10 mm 50 mm 1 mm 10 mm 020.001 124-QS 201 Duran 202

16 17
CELLS

ABSORPTION CELLS

CELLS FOR FLOW-THROUGH MEASUREMENTS CELLS FOR FLOW-THROUGH MEASUREMENTS

macro, with in/outlet tubes compact, with 2 screw connectors M 6 x 1 and FEP tubes
(outside  1.9 mm, inside  1.1 mm, 500 mm long)
TYPE/WINDOW OPTICAL PATH CENTER HEIGHT OUTSIDE DIM. APERTURE VOL. ARTICLE NO. REMARKS
MATERIAL LENGTH mm H×W×D H×W µl
mm mm mm TYPE/WINDOW OPTICAL PATH CENTER HEIGHT OUTSIDE DIM. APERTURE VOL. ARTICLE NO. REMARKS
MATERIAL LENGTH mm H×W×D H×W µl
130-QS 10 45 x 12.5 x 12.5 33 x 9.5 3200 130-10-40 mm mm mm

137-QS 1 45 x 12.5 x 3.5 20 x 9 260 137-1-40 176.700-QS 5 15 35 x 12.5 x 12.5 11 x 3.5 195 1767005-15-40
2 45 x 12.5 x 4.5 20 x 9 520 137-2-40 5 8.5 35 x 12.5 x 12.5 11 x 3.5 195 1767005-85-40
5 45 x 12.5 x 7.5 20 x 9 1300 137-5-40 10 15 35 x 12.5 x 12.5 11 x 3.5 390 1767001510-40
10 45 x 12.5 x 12.5 20 x 9 2600 137-10-40 10 8.5 35 x 12.5 x 12.5 11 x 3.5 390 1767008510-40
50 15 35 x 12.5 x 52.5 11 x 3.5 1950 1767001550-40
170-QS 1 8.5 – 15 35 x 12.5 x 12.5 17.5 x 6.5 120 170-000-1-40
2 35 x 12.5 x 12.5 17.5 x 6.5 240 170-000-2-40

175.000-QS 10 15 45 x 12.5 x 12.5 11 x 6.5 750 175-15-10-40 micro, ultra-micro, with in/outlet tubes
10 8.5 38.5 x 12.5 x 12.5 11 x 6.5 750 175-85-10-40
178.010-QS 10 8.5 38.5 x 12.5 x 12.5 3 80 178-010-10-40

178.011-OS 178011-85-20
compact, with 2 screw connectors M 6 x 1 and FEP tubes 10 8.5 38.5 x 12.5 x 12.5 2 30

(outside  1.9 mm, inside  1.1 mm, 500 mm long)

TYPE/WINDOW OPTICAL PATH CENTER HEIGHT OUTSIDE DIM. APERTURE VOL. ARTICLE NO. REMARKS compact, with 2 screw connectors M 6 x 1 and FEP tubes
MATERIAL LENGTH mm H×W×D H×W µl (outside  1.9 mm, inside  1.1 mm, 500 mm long)
mm mm mm
170.700-QS 0.1 35 x 12.5 x 12.5 17.5 x 3.5 6.2 170700-0.1-40 up to 0.5 mm 178.710-QS 10 15 35 x 12.5 x 12.5 3 80 178-710-10-40
0.2 35 x 12.5 x 12.5 17.5 x 3.5 12.4 170700-0.2-40 with bypass for flow
10 8.5 35 x 12.5 x 12.5 3 80 1787108510-40
optimization
0.5 8.5 – 15 35 x 12.5 x 12.5 17.5 x 3.5 31 170700-0.5-40 50 15 35 x 12.5 x 52.5 3 370 1787101550-40
1 35 x 12.5 x 12.5 17.5 x 3.5 62 170-700-1-40
2 35 x 12.5 x 12.5 17.5 x 3.5 124 170-700-2-40 178.711-OS 10 8.5 35 x 12.5 x 12.5 2 30 1787118510-20

178.712-OS 10 8.5 35 x 12.5 x 12.5  1.5 18 178712-10-20

semi-micro, with in/outlet tubes 178.712-QS 10 8.5 35 x 12.5 x 12.5  1.5 18 1787128510-40

178.765-OS* 10 8.5 45 x 12.5/17 x 12.5  1.5 18 178-765-10-20 without tubes


TYPE/WINDOW OPTICAL PATH CENTER HEIGHT OUTSIDE DIM. APERTURE VOL. ARTICLE NO. REMARKS
MATERIAL LENGTH mm H×W×D H×W µl * Please order tubes separately – see page 33.
mm mm mm
176.000-QS 10 15 45 x 12.5 x 12.5 11 x 4 450 176-15-10-40 WINDOW MATERIAL
50 15 45 x 12.5 x 52.5 11 x 4 2250 176-50-40 OS Special Optical Glass 320 nm – 2500 nm QS Quartz Glass High Performance 200 nm – 2500 nm
50 8.5 38.5 x 12.5 x 52.5 11 x 4 2250 176-50-85-40

WINDOW MATERIAL
OS Special Optical Glass 320 nm–2500 nm QS Quartz Glass High Performance 200 nm – 2500 nm

130 137 170 175.000 170.700 176.000 176.700 176.700 178.010 178.011 178.710 178.710 178.711 178.712 178.765
10 mm 10 mm 1 mm 10 mm 1 mm 10 mm 10 mm 50 mm 10 mm 10 mm 10 mm 50 mm 10 mm 10 mm 10 mm

18 19
CELLS

ALL-QUARTZ FLOW-THROUGH CUVETTE

THREEFOLD ADVANTAGE
UNIQUELY CONVINCING

3 IN 1:
  ABSORBANCE
  FLUORESCENCE Ideally suited for tablet dissolution analysis Clear advantages due to the innovative all-
  2 OPTICAL PATH (TDA) and flow-through spectroscopic quartz design
LENGTHS analysis
● Noliquid leakage, monolithic quartz glass construction
ALL IN ONE SINGLE
The all-quartz flow-through cuvette is a high-precision cell prevents this by design
CUVETTE
for applications in spectroscopy. New technology enables ● Suitable for high and low temperatures
the positioning of precise internal threads into the quartz
glass. Tubes can now be connected very easily and securely ● Fully autoclavable
directly to the cuvette. The second path length is available ● Securetube connection is ensured due to the innovative
by simply turning the cuvette through 90° – all tubes quartz glass internal threads
remain screwed in place.
Time consuming changing of the cuvette is no longer Special features for TDA applications:
ALL-QUARTZ FLOW-THROUGH CUVETTE
necessary. Two different path lengths have a beneficial
Flexible in application effect to the costs and the application. Furthermore, it is Second path length can be set without the time consuming
possible to measure the fluorescence with each optical replacement of the cuvette and the disconnecting and
2 optical path lengths in one single cuvette
path length – another benefit. reconnection of the tubing
Excellent reproducibility of measurement results
Everything in sight: Red point mark for the quick detection
Fluorescence measurement possible for each of the path length position
path length
High temperature and chemical resistance because
of all-quartz construction

V I D E O TU TO R I AL
Innovative all-quartz cuvette with Basic handling and
2 optical path lengths. advantages of an
all-quartz flow-through
The second path length is available, cuvette
by simply turning the cuvette
through 90°
20 21
CELLS

FLUORESCENCE CELLS

MACRO CELLS MICRO CELLS

with PTFE lid or stopper, triangular cell with and without PTFE stopper
TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. NO. OF ARTICLE NO. REMARKS TYPE/WINDOW OPTICAL PATH CENTER OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. NO. OF ARTICLE NO. REMARKS
MATERIAL LENGTH H×W×D mm mm µl WINDOWS MATERIAL LENGTH HEIGHT H × W × D H×W×D mm µl WINDOWS
mm mm mm mm mm mm
101-OS 10 x 10 45 x 12.5 x 12.5 10 1.25 3500 4 101-10-20 101.015-QS 3x3 21 x 5.4 x 5.4 19.9 x 3 x 3 1.1 130 5 101-015-40

101-QS 10 x 10 45 x 12.5 x 12.5 10 1.25 3500 4 101-10-40 013.013 15 50.5 x 12.5 x 12.5 013-013-15-71 holder for cell
10 x 20 45 x 12.5 x 22.5 10 1.25 7000 4 101-20-40 8.5 44 x 12.5 x 12.5 013-013-85-71 type 101.015

111-OS 10 x 10 46 x 12.5 x 12.5 10 1.25 3500 4 111-10-20 111.057-QS 5x5 46 x 7.5 x 7.5 38.75 x 5 x 5 1.25 850 5 111-057-40

111-QS 10 x 10 46 x 12.5 x 12.5 10 1.25 3500 4 111-10-40 013.011 44 x 12.5 x 12.5 013-011-71 holder for cell
type 111.057

SEMI-MICRO CELLS
ULTRA-MICRO CELLS
with PTFE lid or stopper
with PE stopper
TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. ARTICLE NO. REMARKS
MATERIAL LENGTH H×W×D mm mm µl TYPE/WINDOW OPTICAL PATH CENTER OUTSIDE DIM. APERTURE CHAMBER FILLING NO. OF ARTICLE NO. REMARKS
mm mm MATERIAL LENGTH HEIGHT H × W × D H×W VOL. VOL. WINDOWS
mm mm mm mm µl µl
104F-QS 10 x 4 45 x 12.5 x 12.5 4 1.25 1400 104F-10-40
105.250-QS 10 x 2 15 45 x 12.5 x 12.5 5 x 2 100 120 3 105-250-15-40
108F-QS 10 x 4 45 x 12.5 x 12.5 4 9 1000 108-F-10-40 10 x 2 8.5 45 x 12.5 x 12.5 5 x 2 100 120 3 105-250-85-40

114F-QS 10 x 4 46 x 12.5 x 12.5 4 1.25 1400 114F-10-40 105.251-QS 3x3 15 45 x 12.5 x 12.5 5 x 3 45 70 3 105-251-15-40
3x3 8.5 45 x 12.5 x 12.5 5 x 3 45 70 3 105-251-85-40

105.252-QS 1.5 x 1.5 15 45 x 12.5 x 12.5 5 x 1.5 12 30 3 105-252-15-40


MICRO CELLS 1.5 x 1.5 8.5 45 x 12.5 x 12.5 5 x 1.5 12 30 3 105-252-85-40

WINDOW MATERIAL
with PTFE lid or stopper QS Quartz Glass High Performance 200 nm – 2500 nm
TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. ARTICLE NO. REMARKS
MATERIAL LENGTH H×W×D mm mm µl
mm mm
104.002F-QS 10 x 2 45 x 12.5 x 12.5 2 1.25 700 104002F-10-40

108.002F-QS 10 x 2 45 x 12.5 x 12.5 2 9 500 108002F-10-40

115F-QS 10 x 2 40 x 12.5 x 12.5 2 1.25 400 115-F-10-40

WINDOW MATERIAL
OS Special Optical Glass 320 nm – 2500 nm QS Quartz Glass High Performance 200 nm – 2500 nm

101 111 104F 108F 114F 104.002F 108.002F 115F 101.015 111.057 013.011 105.250 105.251 105.252
10 x 10 mm 10 x 10 mm 10 x 4 mm 10 x 4 mm 10 x 4 mm 10 x 2 mm 10 x 2 mm 10 x 2 mm 3 x 3 mm 5 x 5 mm 013.013 10 x 2 mm 3 x 3 mm 1.5 x 1.5 mm

22 23
CELLS

FLUORESCENCE CELLS

FLUORESCENCE CELLS FOR MAGNETIC STIRRERS CELLS FOR FLOW-THROUGH MEASUREMENTS

macro, semi-micro, with PTFE lid or stopper macro, with in/outlet tubes
TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. NO. OF ARTICLE NO. REMARKS TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. APERTURE VOL. NO. OF ARTICLE NO. REMARKS
MATERIAL LENGTH H×W×D mm mm µl WINDOWS MATERIAL LENGTH H×W×D H×W µl WINDOWS
mm mm mm mm mm
109.000F-QS 10 x 10 45 x 12.5 x 12.5 10 5 3500 4 109000F-10-40 131-QS 10 x 10 45 x 12.5 x 12.5 33 x 10 3300 4 131-10-40 base and lid 6 mm

119.000F-QS 10 x 10 49.5 x 12.5 x 12.5 10 5 3500 4 119F-10-40

109.004F-QS 10 x 4 45 x 12.5 x 12.5 4 5 1500 4 109004F-10-40 compact, with 2 screw connectors M 6 x 1 and FEP tubes
(outside  1.9 mm, inside  1.1 mm, 500 mm long)
119.004F-QS 10 x 4 49.5 x 12.5 x 12.5 4 5 1500 4 119004F-10-40
TYPE/WINDOW OPTICAL PATH CENTER OUTSIDE DIM. APERTURE VOL. NO. OF ARTICLE NO.
332.300 6x3 332-300-VE10 see page 33 MATERIAL LENGTH HEIGHT H×W×D H×W µl WINDOWS
mm mm mm mm
176.751-QS 3x3 8.5 35 x 12.5 x 12.5 11 x 3 100 3 176-751-85-40
SEALABLE CELLS
176.754-QS 10 x 2.5 15 35 x 12.5 x 12.5 11 x 2.5 275 4 176-754-10-15-40

macro, semi-micro, for anaerobic applications


TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. NO. OF ARTICLE NO. REMARKS ALL-QUARTZ FLOW-THROUGH CELLS WITH TWO OPTICAL PATH LENGTHS
MATERIAL LENGTH H×W×D mm mm µl WINDOWS
mm mm
with screw connectors M6 x 1, with FEP tubing 500 mm length
117.100F-QS 10 x 10 56 x 12.5 x 12.5 10 1.25 3500 4 117100F-10-40 Open screw cap
TYPE/WINDOW OPTICAL PATH CENTER OUTSIDE DIM. APERTURE VOL. NO. OF ARTICLE NO. REMARKS
117.200F-QS 10 x 10 56 x 12.5 x 12.5 10 1.25 3500 4 117200F-10-40 Closed screw cap MATERIAL LENGTH HEIGHT H×W×D H×W µl WINDOWS
mm mm mm mm
117.104F-QS 10 x 4 56 x 12.5 x 12.5 4 1.25 1400 4 117104F-10-40 Open screw cap
176.760-QS 5 and 10 15 35 x 12.5 x 12.5 11 x 6/11 x 5 550 4 176-760-15-40 for further
117.204F-QS 10 x 4 56 x 12.5 x 12.5 4 1.25 1400 4 117204F-10-40 Closed screw cap 8.5 176-760-85-40 information, see
page 20 to 21
With ISO thread GL 14 and screw cap with silicone rubber seal; accessories see page 33. 176.761-QS 2.5 and 5 15 35 x 12.5 x 12.5 11 x 4/11 x 2.5 140 4 176-761-15-40
8.5 176-761-85-40

CELLS WITH TUBES QUARTZ/DURAN® 176.762-QS 1.5 and 3 15 35 x 12.5 x 12.5 11 x 2.5 /11x1.5 50 4 176-762-15-40
8.5 176-762-85-40
176.765-QS 1 and 10 15 35 x 12.5 x 12.5 11 x 6/11 x 1 110 176-765-15-40
macro, tube  8 mm, tube length 80 mm 8.5
4
176-765-85-40
TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. NO. OF ARTICLE NO. 176.766-QS 2 and 10 15 35 x 12.5 x 12.5 11 x 6/11 x 2 220 4 176-766-15-40
MATERIAL LENGTH H×W×D mm mm µl WINDOWS 8.5 176-766-85-40
mm mm
221-QS 10 x 10 40 x 12.5 x 12.5 10 1.25 3500 4 221-10-40 WINDOW MATERIAL
QS Quartz Glass High Performance 200 nm – 2500 nm
221.001-QS* 10 x 10 Tol.+- 0.2 40 x 12.5 x 12.5 10 1.25 3500 4 221001-10-80
* for measurements at high and low temperatures

WINDOW MATERIAL
QS Quartz Glass High Performance 200 nm – 2500 nm

109.000F 119.000F 109.004F 119.004F 117.200F-QS 117.204F-QS 221 221.001 131 176.751 176.754 176.760-QS
10 x 10 mm 10 x 10 mm 10 x 4 mm 10 x 4 mm 10 x 10 mm 10 x 4 mm 10 x 10 mm 10 x 10 mm 10 x 10 mm 3 x 3 mm 10 x 2.5 mm 5 and 10 mm

24 25
CELLS

FLOW CYTOMETRY AND PARTICLE ANALYSIS

EXCEPTIONALLY FINE
MICRO-CHANNELS FOR F O R I N F O RM ATION
For further information on the

EXACT MEASUREMENT topic of cytometry and the fields


of application, see
www.hellma.com/cytometry

AND RELIABLE RESULTS

Quality and technology Areas of application

A high-precision quartz glass flow-through cuvette with ● Medical diagnostics (HIV, TB, malaria)
Example of use: a superfine channel is at the heart of every cytometer. This
● Analysis of the cycle of the cell
channel lends stability to the fluidic system, enabling the
Mobile analysis in the fight against
accurate optical analysis of individual cells or particles. ● Cell biology
HIV, tuberculosis and malaria
At Hellma Analytics, the production of cytometer cuvettes ● Cell and particle separation
Flow cytometry is used to analyze particles draws on more than 95 years of experience in producing
● Bead-based assays
and cells quickly and accurately. Hellma glass and quartz components. Thanks to our use of advan-
Analytics micro-channel cuvettes can be ced glass processing we are able to manufacture micro- ● DNA analysis
employed in mobile laboratories for the channel, cytometer cuvettes featuring polished channel
detection of life-threatening diseases such surfaces and made from fluorescence-free materials in Benefits
as HIV, tuberculosis and malaria. Set up custom sizes as small as 50 μm x 50 μm. The consistently
in secure vehicles, these analytical facilities high production quality guarantees maximum repro- ● Extremely tight tolerances for easy system integration
can provide fast and reliable diagnoses. ducibility with minimum tolerances. Our exceptional
● Extremely flat and flawless channels for excellent cell
Hellma’s technology is able to simultaneously production expertise combined with our state-of-the-art
production machinery enable us to manufacture cuvettes morphology analysis
analyze up to 16 parameters in more than
100,000 cells per minute, making a vital with various cone shapes as well as solutions tailored ● Completely fluorescence-free quartz glass for accurate
contribution towards rapidly helping people to our clients’ requirements. detection of fluorescence
in need. ● Perfectly rectangular channels allow laser beams to
pass through uninhibited
● Unrivaled production capacity and flexibility thanks to
Hellma’s specially designed production plant

CYTOMETER CUVETTE Fantastic OEM product!


Channels with polished inner surfaces
Made on a batch production line and tailored
Manufacture of cones of various shapes
to the measuring equipment
and sizes possible
Very tight tolerances
Outstanding surface precision
Manufacture of extremely narrow channels

26 27
CELLS

CELLS AND OPTICAL ELEMENTS


FOR SPECIAL APPLICATIONS

DYE-LASER CELL CELLS FOR REFLECTION MEASUREMENTS

macro, with PTFE stoppers cylindrical cells without lid


TYPE/WINDOW OUTSIDE DIM. INSIDE CROSS VOL. NO. OF ARTICLE NO. REMARKS TYPE/WINDOW OUTSIDE DIM. INSIDE DIM. VOL. ARTICLE NO. REMARKS
MATERIAL H×W×D SECTION µl WINDOWS MATERIAL H × DIAMETER H × DIAMETER µl
mm mm mm mm
111.070-QS 46 x 12.5 x 12.5 10 x 10 3500 4 111-070-40 on request with a polished base 692.091-OG 25 x 34 23 x 31.6 12000 692-091-12

692.103-BF 30 x 50 27.5 x 45 32000 692-103-23

CELL WITH TWO CHAMBERS 692.104-BF 40.5 x 60 39 x 55.6 73000 692-104-23

692.455-BF 52 x 65 50 x 60 110000 692-455-23 acc. to ISO 17223 with markings at 25 mm and 45 mm


TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE WIDTH BASE THICKN. VOL. ARTICLE NO. REMARKS
MATERIAL LENGTH H×W×D mm mm µl
mm mm WINDOW MATERIAL
OG Optical Glass 360 nm – 2500 nm BF Borosilicate Glass 330 nm – 2500 nm
238-QS 2 x 4.375 46 x 12.5 x 12.5 9.5 1.5 2 x 1000 238-000-40 with 2 stoppers

CELLS FOR LIGHT SCATTERING MEASUREMENTS

with PTFE stoppers


TYPE/WINDOW OUTSIDE DIM. INSIDE DIM. VOL. ARTICLE NO. REMARKS
MATERIAL H × DIAMETER H × DIAMETER µl
mm mm
540.110-QS 75 x 10 74 x 8 3200 540-110-80

540.111-QS 75 x 10 74 x 8 3200 540-111-80 polished outer cylinder

540.135-QS 75 x 20 74 x 18 14000 540-135-20-40

WINDOW MATERIAL
QS Quartz Glass High Performance 200 nm – 2500 nm

111.070 238 540.110 540.135 692.091 692.103 692.104 692.455


10 x 10 mm 2 x 4.375 mm

28 29
CELLS

CELLS AND OPTICAL ELEMENTS


FOR SPECIAL APPLICATIONS

CELL FOR TURBIDITY MEASUREMENTS OPTICAL PARTS

rectangular cell TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. THICKNESS INSIDE VOL. ARTICLE NO. REMARKS
MATERIAL LENGTH H×W×D mm WIDTH µl
TYPE/WINDOW OPTICAL PATH OUTSIDE DIM. INSIDE DIM. VOL. ARTICLE NO. REMARKS mm mm mm
MATERIAL LENGTH H×W×D H×W×D µl 665.000-QS 45 x 12.5 1.25 665-000-40 rectangular window made of Quartz Glass High
mm mm mm Performance
402.013-OG 25 x 25 70 x 30 x 30 67 x 25 x 25 35000 402-013-10 25 ml marking, 5 windows
665.000-QX 45 x 12.5 1.25 665-000-46 rectangular window made of Quartz Glass Extended
Range

LARGE CELLS TYPE/WINDOW OUTSIDE- THICKNESS INSIDE- VOL. ARTICLE NO. REMARKS
MATERIAL DIAMETER mm DIAMETER µl
with glass lids mm mm
202-QS  22 1.25 202-40 circular window made of Quartz Glass High
700.000-OG 10 ± 0.2 53 x 55 x 15 50 x 50 x 10 20000 700-000-10-10 Performance
with glass lid
20 ± 0.2 53 x 55 x 25 50 x 50 x 20 40000 700-000-20-10
202-QX  22 1.25 202-46 circular window made of Quartz Glass Extended
700.010-OG 20 ± 0.2 82 x 44.4 x 24.4 80 x 40 x 20 56000 700-010-20-10 without lid Range

700.015-OG 28 ± 0.2 35 x 35 x 32 33 x 31 x 28 22000 700-015-10 without lid


WINDOW MATERIAL
700.016-OG 18 ± 0.2 38 x 22 x 22 36 x 18 x 18 10000 700-016-10 without lid QS Quartz Glass High Performance 200 nm – 2500 nm QX Quartz Glass Extended Range 200 nm – 3500 nm

700.061-OG 50 ± 0.5 100 x 150 x 55 96.5 x 143 x 50 600000 700-061-10 without lid

704.000-OG 20 ± 0.2 22.5 x 25 x 25 20 x 20 x 20 6000 704-000-20-10 with glass lid

704.001-OG 30 ± 0.2 32.5 x 35 x 35 30 x 30 x 30 22500 704-001-30-10 with glass lid

704.002-OG 40 ± 0.2 42.5 x 45 x 45 40 x 40 x 40 56000 704-002-40-10 with glass lid

704.003-OG 50 ± 0.5 52.5 x 55 x 55 50 x 50 x 50 88000 704-003-50-10 with glass lid

740.000-OG 34.5 ± 0.2 100 x 50 x 39.5 97 x 44 x 34.5 100000 740-000-10 with markings at 100 ml
without lid

WINDOW MATERIAL
OG Optical Glass 360 nm – 2500 nm

402.013 700.000 700.010 700.015 700.016


10 mm 20 mm 30 mm 18 mm

704.000 704.001 704.002 704.003 740.000 665.000 202


20 mm 30 mm 40 mm 50 mm 34.5 mm

30 31
CELLS

QUARTZ MICROPLATES
AND ACCESSORIES FOR CELLS

QUARTZ MICROPLATES TUBINGS

made of quartz TYPE DESCRIPTION ARTICLE NO. REMARKS

TYPE/WINDOW DESCRIPTION OUTSIDE DIM. BASE WELLS ARTICLE NO.


MATERIAL H×W×D mm 040.111 FEP tubing set 500 mm long; 040-111-722 for compact and 3-in-1 cells; with one short and one long
outside  1.9 mm; inside  1.1 mm screw fitting
mm DIAMETER DEPTH VOLUME
mm mm µl
040.222 PTFE tubing set 500 mm long with Omnifit gripper 040-222-72 for compact and 3-in-1 cells; with one short and one long
730.009-QG Quartz Microplate** 14.5 x 127 x 85.5 2* 6.6 12.5 300 730-009-44 outside  1.6 mm; inside  1.0 mm Omnifit Gripper
with 96 wells
Base: Synthetic Quartz Glass

730.009B-QG Black Quartz Microplate 14.5 x 127 x 85.5 2* 6.6 12.5 300 730009-B-44 LIDS, STOPPERS AND OTHER ACCESSORIES
with 96 wells
Base: Synthetic Quartz Glass
010.010 PTFE lid 010.010, 10 mm 010-001-10-VE10-72 for cell models with 10 mm path length
10 pcs pack
QG is synthetic quartz glass with a transmission over 80% between 200 nm and 2500 nm for an empty cell.
* On request base with reduced thickness down to 0.5 mm.
** Available made of Borosilicate Glass on request. 010.050 PTFE lid 010.050, 50 mm 010-001-50-VE5-72 for cell models with 50 mm path length
5 pcs pack
011.001 PTFE stopper with fitting NS 5 011-001-VE5-72 for cell types:
5 pcs pack 110, 111, 114, 120 with 1 – 5 mm path length; and for cell
type 404 with 1 – 10 mm path length

011.002 PTFE stopper with fitting NS 7 011-002-VE5-72 for cell types:


5 pcs pack 110, 111, 114/114F, 115/115F, 119/119F,
120 with 10 – 100 mm path length; and for cell type 770

011.103 PE stopper, 10 mm 011-103-VE10-73 for cell types:


10 pcs pack 105.200, 105.201, 105.202, 105.203, 105.204, 105.250,
105.251, 105.252, 105.253, 105.254

011.550 Pipette tip for Ultra-Micro cells 011-550-VE10 for cell types:
10 pcs pack 105.210-QS

011.600 Open screw caps, with ISO GL 14 thread and 011-600-VE10-734 for cell types:
silicone seal (septum) 117.100; 117.100F, 117.104, 117.104F
730.009-QG 730.009B-QG
10 pcs pack

011.601 Closed screw caps, with ISO GL 14 thread and 011-601-VE10-734 for cell types:
silicone seal 117.200; 117.200F, 117.204, 117.204F
10 pcs pack
ALUMINUM SPACERS
011.650 Replacement silicone rubber seals (septum) 011-650-VE10-72 for cell types:
10 pcs pack 117.100; 117.100F, 117.104, 117.104F, 117.200; 117.200F,
TYPE DESCRIPTION ARTICLE NO. REMARKS 117.204, 117.204F

013.101 Aluminum spacer 013-101-71 to fit cells with 1 mm optical path length 011.651 Replacement silicone rubber seals (septum), 011-651-VE10-72 for cell types:
into 10 mm cell holder PTFE coated on one side 117.100; 117.100F, 117.104, 117.104F, 117.200; 117.200F,
38 x 12.5 x 9 mm 117.204, 117.204F
10 pcs pack
013.102 Aluminum spacer 013-102-71 to fit cells with 2 mm optical path length
38 x 12.5 x 8 mm into 10 mm cell holder 332.300 PTFE coated magnetic stir bars 332-300-VE10  ca. 3 – 4 mm,
10 pcs pack Length 6 – 7 mm
013.105 Aluminum spacer 013-105-71 to fit cells with 5 mm optical path length for cell types:
38 x 12.5 x 5 mm into 10 mm cell holder 109.000, 109.000F, 109.004, 109.004F, 119.000, 119.000F,
119.004, 119.004F

013.102
040.111 040.222 010.010 010.050 011.650 011.651 011.600 011.601

32 33
CELLS

CLEANING CUVETTES AND OPTICAL PARTS TIPS ON HANDLING CUVETTES

FOR ACCURATE, UNTAINTED RESULTS PLEASE PAY ATTENTION TO THE FOLLOWING

Regularly using Hellmanex® III to clean your cuvettes and 1. Our precision cuvettes are made of glass or quartz glass SPECIAL INSTRUCTIONS FOR CUVETTES
optical parts ensures accurate measurement and precise and have all the advantages and disadvantages (such as the
SEALED WITH STOPPERS
results. This highly effective alkaline cleaning concentrate inherent fragility) of these materials. Once the measurement
is ideal for use on glass and quartz glass cuvettes, sensitive process is complete, we generally recommend that you Cuvettes containing liquid and sealed with stop-
optical parts and laboratory equipment made of glass, quartz, immediately clean, dry and store the cuvettes in cases. pers may break if the internal pressure increases.
sapphire and porcelain. It effectively removes dirt and prevents
2. Do not store the cuvettes in the open in a corrosive atmo­- The most common reason for such an increase
loosened dirt particles from redepositing. After cleaning, the
s­phere, and do not leave the polished windows in contact in pressure is the expansion of the liquid in the
parts can be rinsed without leaving any residue, including
with liquids for an extended period of time. This could lead cuvette due to a rise in temperature. This may
UV/Vis active substances, being left on the optical surfaces.
to the formation of deposits or stains on the polished be caused by:
surfaces, rendering the cuvettes unusable. ● heat
from an external source, such as thermal
3. In order to avoid scratching the precision-polished windows, conduction via the cuvette holder
USE cuvettes should never come into contact with objects made ● a chemical reaction in the liquid
of hard materials, such as glass or metal.
Cleaning ● radiation absorption in the liquid
1. Place the cuvettes in a bath of Hellmanex III diluted to 2%
with water. Clean flow-through cuvettes by pumping the
cleaning solution through the cuvette.
320.003 325.000 IMPORTANT TIPS TAKING THE FOLLOWING PRECAUTIONS
2. The cleaning process can be sped up by gently heating the
● Careis required when inserting cuvettes into a metal
WILL HELP PREVENT CUVETTES FROM
solution.
cuvette holder. BREAKING:
TYPE DESCRIPTION ARTICLE NO.
3. Agitate the cleaning solution to boost cleaning performance. 1. Fill the cuvette just high enough for the light beam to pass
● When using a pipette to fill cuvettes with liquids,
4. After cleaning, thoroughly rinse the cuvettes using ultrapure 320.003 Hellmanex III
®
9-307-011-4-507 never touch the polished window with the pipette. through the liquid unimpeded. This allows the liquid to
Liquid cleaning concentrate, for glass, expand into the remaining air volume if the temperature in-
water. Replace the contents of the cleaning bath at least
quartz cells and optical components ● Never
use metal tweezers or pliers to carry or hold
three times during rinsing. creases.
1.4 kg PE bottle (1.0 l) cuvettes.
5. Blow the cuvettes dry using clean air and leave them to dry 325.000 CleanAssist plastic cell holder for 4 325.000 2. If you fill the cuvette to the rim, put the stopper on loosely so
out in a dust-free environment. Alternatively, rinse them cells with 10 mm optical path lenght for that any excess liquid can escape.
cleaning purposes
with a highly volatile solvent, such as alcohol. Then allow 3. Do not try to force the stopper into place, as this will inevi-
the solvent to evaporate. tably damage the cuvette.

4. Use stoppers with a capillary hole.

PLEASE NOTE
● Avoid extreme changes in temperature.
! Cleaning and Dilution
CARE MUST ALSO BE TAKEN AT
LOW TEMPERATURES.
Fragile! Although it is possible to cool an empty cuvette down to a few
The optimal dilution depends on several factors, such as the
Avoid exposure to ultrasound waves: Excessive degrees Kelvin without breaking it, when filled with water and

hardness of the water, the degree and type of contamination,
energy density and/or unfavorable frequencies cooled to just a few degrees below the freezing point, the same
the temperature etc. The use of demineralized water improves
may break the cuvettes. Cuvettes made of multiple cuvette may burst, even if it is not sealed. This is because
the cleaning characteristics.
materials (glass, metal, etc.) are especially at risk. water expands in all directions when cooled and if it freezes
Cavitation attacks polished surfaces, rendering them may cause the cuvette to burst.
unusable. CONCENTRATION TEMPERATURE (°C) TIME
(% BY VOL.) (MINUTES)
● Do not leave the cleaning solution in the cuvette
0.5 – 2 20 – 25 120 – 180
at high temperatures for so long that it evaporates.
This is because an increase in concentration and 0.5 – 2 30 – 35 30 – 40
the high pH value may damage the surface of the
0.5 – 2 50 – 60 (Quartz only) 10 – 15
glass.
0.5 – 2 70 – 80 (Quartz only) <5

34 35
CELLS

CERTIFIED REFERENCE MATERIALS

CERTIFIED UV / VIS REFERENCE


MATERIALS / CALIBRATION
STANDARDS

Continuously reliable measurements and Benefits


precise results through certified validation
● Outstanding quality and high levels of compliance
The regular testing of spectrophotometers ensures accu- (Ph. Eur., USP, DAB, GLP, DIN 9001, etc.)
rate results and plays an important role in quality assuran-
● Traceable to NIST standards
ce and production control. Hellma Analytics reference
materials for UV/Vis spectrophotometry meet internatio- ● With DAkkS certificate
nally recognized standards and provide ultimate process
● Possible to certify each filter separately
transparency. Besides enabling the spectral resolution and
EP 10 FILTER SET wavelength accuracy to be reliably tested, they can check ● No need to prepare solutions
for stray light and photometric accuracy, too.
● Cost-effective and environmentally friendly
Wavelength accuracy,
photometric accuracy,
photometric linearity, Areas of application
stray light and spectral resolution
in one set The safe and easy testing of
UV/Vis spectrophotometers for
Can be extended according to the following parameters:
requirements
● Wavelength accuracy
● Photometric accuracy
● Photometric linearity
● Stray light level
● Spectral resolution
V I D E O TU TO R I AL
Preparation and
carrying out of
measurements
With the certified UV/Vis reference materials with glass filters

from Hellma Analytics, we create the basis Handling instructions


for reliable measurement results for our for download
customers. www.hellma.com/CRM-en

Birgit Kehl, Compliance Manager


Calibration Laboratory

40 41
CERTIFIED REFERENCE MATERIALS

ACCURATE
MEASUREMENTS
GUARANTEED

UV/Vis certified reference materials


Quality assurance and quality control regulations, such
as ISO 9001, GLP, GMP and pharmacopoeias, require
companies to verify that any spectrophotometers in use are
performing at a consistently high level. Hellma Analytics'
certified reference materials can be used to test the
following parameters easily and efficiently: wavelength
accuracy, photometric accuracy, photometric linearity,
stray light and spectral resolution in UV/Vis spectrophoto-
meters.

Hellma Analytics calibration laboratory:


accredited to DIN EN ISO / IEC 17025 TRACEABLE
Our calibration laboratory is accredited by DAkkS to
DIN EN ISO/IEC 17025. By achieving this accreditation, we
have provided proof of our expertise in the calibration acti-
vities that we perform and are authorized to issue interna-
tionally recognized DAkkS calibration certificates. All cer- DIN EN ISO/IEC 17025
tified measurement results can be traced back to NIST
standard reference materials (SRMs) and to the PTB
(Physikalisch-Technische Bundesanstalt – Germany’s na-
tional metrology institute).

Quality management
Hellma Analytics’ reference materials
comply with the provisions stipulated
by quality management systems and
Spectrophotometers in pharmacopoeias, meeting the highest
instrumental analytics are quality requirements and ensuring that
measurement results can be compared
considered as subject to internationally.
examination.

42 43
CERTIFIED REFERENCE MATERIALS

UV/VIS CERTIFIED REFERENCE MATERIALS

GLASS FILTERS WITH DAkkS CERTIFICATE


TYPE MATERIAL WAVELENGTH nm ARTICLE NO. TYPE MATERIAL WAVELENGTH nm ARTICLE NO.

Glass Filters for testing the wavelength accuracy Sets for testing the photometric accuracy, photometric linearity and wavelength accuracy

666-F1 Holmium Glass Filter F1 279; 361; 453; 536; 638 666F1-339 666-S000 Complete Glass Filter Set: A: 440; 465; 546.1; 590; 635 666S000
F1, F2, F3, F4, F0 (Abs: 0.25; 0.5; 1.0) W: 279; 361; 453; 536; 638
666-F7W Didymium Glass Filter F7W 329; 472; 512; 681; 875 666F7W-323
666-S001 Glass Filter Set: A (F7): 270; 280; 297; 321; 342 666S001
Glass Filters for testing the photometric accuracy F3, F4, F7 (Abs: 0.5; 1.0; F7: ca. 0.5 – 1.0) A (F3, F4): 440; 465; 546.1;
590; 635
666-F390 Neutral Density Glass Filter F390; 0.04 Abs 440; 465; 546.1; 590; 635 666F390-25 W (F7): 329; 472; 512; 681; 875

666-F2 Neutral Density Glass Filter F2; 0.25 Abs 440; 465; 546.1; 590; 635 666F2-39 666-S002 Glass Filter Set: F2, F3, F4 (Abs: 0.25; 0.5; 1.0) A: 440; 465; 546.1; 590; 635 666S002

666-F201 Neutral Density Glass Filter F201; 0.3 Abs 440; 465; 546.1; 590; 635 666F201-39 666-S003 Glass Filter Set: A (F7): 270; 280; 297; 321; 342; 666S003
F1, F2, F3, F4, F7; (Abs: 0.25; 0.5; 1.0; F7: ca. 0.5 – 1.0) A (F2, F3, F4): 440; 465; 546.1;
666-F3 Neutral Density Glass Filter F3; 0.5 Abs 440; 465; 546.1; 590; 635 666F3-38 590; 635
W (F1): 279; 361; 453; 536; 638
666-F204 Neutral Density Glass Filter F204; 0.7 Abs 440; 465; 546.1; 590; 635 666F204-37 W (F7): 329; 472; 512; 681; 875

666-F4 Neutral Density Glass Filter F4; 1.0 Abs 440; 465; 546.1; 590; 635 666F4-37 666-S004 Glass Filter Set: F201, F202, F203, F0 (Abs: 0.3; 1.5; 2.0) A: 440; 465; 546.1; 590; 635 666S004

666-F202 Neutral Density Glass Filter F202; 1.5 Abs 440; 465; 546.1; 590; 635 666F202-36 666-S005 Glass Filter Set: F0, F1, F3, F4; (Abs: 0.5; 1.0) A: 440; 465; 546.1; 590; 635 666S005
W: 279; 361; 453; 536; 638
666-F203 Neutral Density Glass Filter F203; 2.0 Abs 440; 465; 546.1; 590; 635 666F203-36
666-S006 Glass Filter Set: F0, F2, F3, F4; (Abs: 0.25; 0.5; 1.0) A: 440; 465; 546.1; 590; 635 666S006
666-F301 Neutral Density Glass Filter F301; 2.5 Abs 440; 465; 546.1; 590; 635 666F301-361
666-S200 NEW Glass Filter Set: F2, F4, F203; (Abs: 0.25; 1.0; 2.0) A: 440; 465; 546.1; 590; 635 666S200
666-F303 Neutral Density Glass Filter F303; 3.0 Abs 440; 465; 546.1; 590; 635 666F303-361
666-S300 Glass Filter Set: F390, F301, F303 (Abs: 0.04; 2.5; 3.0) A: 440; 465; 546.1; 590; 635 666S300
666-F7A Didymium Glass Filter F7A; ca. 0.5 – 1.0 Abs 270; 280; 297; 321; 342 666F7A-323
A: Wavelength for absorbance   W: Wavelength for wavelength accuracy
Glass Filter for testing the photometric accuracy and wavelength accuracy

666-F7 Didymium Glass Filter F7 A: 270; 280; 297; 321; 342 666F7-323
W: 329; 472; 512; 681; 875

Empty filter mount

666-F0 Reference filter frame made of aluminum (without glass) 666F0-71

A: Wavelength for absorbance   W: Wavelength for wavelength accuracy

Complete Set 666-S000

666-F1 666-F2 666-F3 666-F4 666-F201 666-F202 666-F203 666-F7 666-F0 666-F1 666-F2 666-F3 666-F4

44 45
CERTIFIED REFERENCE MATERIALS

UV/VIS CERTIFIED REFERENCE MATERIALS

LIQUID FILTERS WITH DAkkS CERTIFICATE


TYPE CONTENT WAVELENGTH nm ARTICLE NO. TYPE CONTENT WAVELENGTH nm ARTICLE NO.

Potassium dichromate Liquid Filters for testing the photometric accuracy Liquid Filters for testing the wavelength accuracy according to EP and USP

667-UV20 20 mg/l potassium dichromate in HClO4 (0.1–0.3 Abs) 235; 257; 313; 350 667020 667-UV5 Holmium in perchloric acid acc. to EP10 241; 287; 361; 451; 485; 536; 640 667005

667-UV40 40 mg/l potassium dichromate in HClO4 (0.2–0.6 Abs) 235; 257; 313; 350 667040 667-UV5USP Holmium in perchloric acid acc. to USP857 241; 250; 278; 287; 333; 345; 361; 667005USP
385; 416; 452; 468; 485; 536; 640
667-UV60 60 mg/l potassium dichromate in HClO4 (0.3–0.9 Abs) 235; 257; 313; 350 667060
667-UV25 Didymium in perchloric acid 329; 469; 575; 740; 864 667025
667-UV80 80 mg/l potassium dichromate in HClO4 (0.4–1.2 Abs) 235; 257; 313; 350 667080
667-UV25EPUSP UV25, Didymium Liquid Filter for testing the wavelength accuracy 512; 732; 740; 794; 801; 864 667025EPUSP
667-UV0100 100 mg/l potassium dichromate in HClO4 (0.5–1.45 Abs) 235; 257; 313; 350 6670100 NEW above 640 nm (according to EP10 and USP857)

667-UV0120 120 mg/l potassium dichromate in HClO4 (0.6–1.7 Abs) 235; 257; 313; 350 6670120 667-UV35 Rare Earth (Cerium) Liquid Filter for testing the wavelength accuracy 201; 211; 222; 239; 252 667035
below 240 nm (according to EP10 and USP857)
667-UV0140 140 mg/l potassium dichromate in HClO4 (0.7–2.0 Abs) 235; 257; 313; 350 6670140
667-UV45 Holmium/Didymium in perchloric acid 241; 354; 444; 575; 641; 740; 864 667045
667-UV0160 160 mg/l potassium dichromate in HClO4 (0.8–2.3 Abs) 235; 257; 313; 350 6670160
667-UV45EPUSP UV45, HoDi Liquid Filter (Holmium und Didymium in perchloric acid) 241; 287; 361; 451; 482; 512; 537; 667045EPUSP
667-UV0180 180 mg/l potassium dichromate in HClO4 (0.9–2.6 Abs) 235; 257; 313; 350 6670180
NEW for testing the wavelength accuracy from 240 to 870 nm (according to 641; 732; 740; 794; 801; 864
667-UV0200 200 mg/l potassium dichromate in HClO4 (1.0–3.0 Abs) 235; 257; 313; 350 6670200 EP10 and USP857)

667-UV600 600 mg/l potassium dichromate in HClO4 (1.0 Abs) 430 667600 667-UV400 Liquid Filter Set: UV5, Holmium and UV14, perchloric acid 241; 287; 361; 451; 485; 536; 640 667400
(Reference Filter) – acc. to EP10
667-UV14 Perchloric acid (reference filter) 235; 257; 313; 350 667014
667-UV425 Liquid Filter Set: UV5, Holmium and UV25, Didymium for testing the UV5: 241; 250; 278; 287; 333; 345; 667425
667-UV301 Filter Set for UV-range: UV60, UV14 235; 257; 313; 350 667301 wavelength accuracy from 240 to 870 nm (according to USP 857) 361; 385; 416; 452; 468; 485; 536; 640
UV25: 732; 740; 794; 801; 864
667-UV304 Filter Set for Vis-range: UV600, UV14 430 667304
Liquid Filters for testing stray light
667-UV305 Filter Set for UV/Vis-range: UV60, UV600, UV14 235; 257; 313; 350; 430 667305
667-UV1 Potassium chloride in pure water, PL 10 mm, ≥ 2 Abs at 198 nm 200 cut-off 667001
Potassium dichromate Liquid Filter Set for testing the photometric linearity
667-UV1H* Potassium chloride in pure water, reference filter PL 5 mm 200 cut-off 667001H
667-UV307 Filter Set: UV20, UV40, UV60, UV80, UV0100, UV14 235; 257; 313; 350 667307
667-UV10 Sodium iodide in pure water, PL 10 mm, ≥ 3 Abs at 220 nm 259 cut-off 667010
Niacin Liquid Filter Sets for testing the photometric accuracy
667-UV10H* Sodium iodide in pure water, reference filter PL 5 mm 259 cut-off 667010H
667-UV506 6 mg/l Niacin in HCl (0.25 Abs) 213; 261 667506
667-UV23 NEW Potassium Iodide in pure water, PL 10 mm, ≥ 3 Abs at 250 nm 259 cut-off 667023
667-UV512 12 mg/l Niacin in HCl (0.5 Abs) 213; 261 667512
667-UV11 Sodium nitrite in pure water, PL 10 mm, ≥ 3 Abs at 340 nm 385 cut-off 667011
667-UV518 18 mg/l Niacin in HCl (0.75 Abs) 213; 261 667518 and 370 nm

667-UV524 24 mg/l Niacin in HCl (1.0 Abs) 213; 261 667524 667-UV11H* Sodium nitrite in pure water, reference filter PL 5 mm 385 cut-off 667011H

667-UV536 NEW 36 mg/l Niacin in HCl (1.5 Abs) 213; 261 667536 667-UV12 Pure water, reference filter PL 10 mm 198; 200; 300; 400 667012

667-UV548 NEW 48 mg/l Niacin in HCl (2.0 Abs) 213; 261 667548 667-UV19 Acetone (pure), PL 10 mm, ≥ 2 Abs at 300 nm 325 cut-off 667019

667-UV599 Hydrochloric Acid (HCl), (reference filter) 213; 261 667599 667-UV19H* Acetone (pure), reference filter PL 5 mm 325 cut-off 667019H

Liquid Filter Sets for testing stray light according to EP10 and USP857 Procedure B
Niacin Liquid Filter Sets for testing the photometric linearity
667-UV100 Filter Set contains: UV1 and UV12; ≥ 2 Abs at 198 nm 200 cut-off 667100
667-UV350 Filter Set: UV506, UV512, UV518, UV524, UV599 213; 261 667350
667-UV101 Filter Set contains: UV10 and UV12; ≥ 3 Abs at 220 nm 259 cut-off 667101
667-UVS501 NEW Filter Set: UV506, UV512, UV524, UV599 213; 261 667S501
667-UV107 NEW Filter Set contains: U23 and UV12; ≥ 3 Abs at 250 nm 259 cut-off 667107

667-UV102 Filter Set contains: UV11 and UV12; ≥ 3 Abs at 340 and 370 nm 385 cut-off 667102

667-UV103 Filter Set contains: UV1, UV10, UV11 and UV12 cut-off: 200; 259; 385 667103

667-UV104 Filter Set contains: UV10, UV11 and UV12 cut-off: 259; 385 667104

46 47
CERTIFIED REFERENCE MATERIALS

UV/VIS CERTIFIED REFERENCE MATERIALS NEW BASIC FILTER SETS FOR EP 10 COMPLIANCE

LIQUID FILTERS WITH DAkkS CERTIFICATE EP BASIC SET UV/VIS


TYPE CONTENT WAVELENGTH nm ARTICLE NO.
(for wavelength range: 240–640 nm and absorbance range: 0.25–1.0)

Liquid Filter Sets for testing stray light according to USP857 Procedure A PARAMETER ARTICEL NO. CONTENT WAVELENGTH nm

667-UV100H Filter Set contains: UV1 and UV1H, PL 10 and 5 mm 200 cut-off; SR: 190–210 667100H Photometric accuracy 667S501 Filter set S501 includes: 213, 261
UV range UV506, Niacin filter 6 mg/l (0.25 Abs)
667-UV101H Filter Set contains: UV10 and UV10H, PL 10 and 5 mm 259 cut-off; SR: 210–270 667101H UV512, Niacin filter 12 mg/l (0.5 Abs)
UV524, Niacin filter 24 mg/l (1.0 Abs)
667-UV102H Filter Set contains: UV11 and UV11H, PL 10 and 5 mm 385 cut-off; SR: 300–400 667102H UV599, Hhydrochloric acid (reference filter)
667-UV119H Filter Set contains: UV19 and U19H, PL 10 and 5 mm 325 cut-off; SR: 250–330 667119H Photometric accuracy 666S002 Glass Filter Set S002 includes: 440, 465, 546.1, 590, 635
Vis range F2 (0.25 Abs), F3 (0.5 Abs) and F4 (1.0 Abs)
667-UV105H Filter Set contains: UV1/UV1H; UV10/UV10H; UV19/UV19H; cut-off: 200; 259; 325; 385 667105H
PL 10 mm and 5 mm Wavelength accuracy 667400 Filter set, UV5 Holmium in Perchloric acid and UV14 Perchloric acid 241, 287, 361, 451, 485, 536, 640
(reference filter)
667-UV106H Filter Set contains: UV1/UV1H; UV10/UV10H; UV19/UV19H; Cut-off: 200; 259; 325; 667106H
SD 10 mm and 5 mm Stray Light 667102 Filter set, UV11 Sodium nitrite in pure water and UV12 pure water 385 cut-off
(reference filter), ≥ 3 Abs at 340 and 370 nm
Liquid Filters for testing the resolution
Resolution 667200 Filter set, UV6 Toluene in Hexane and UV9 Hexane (reference filter) Scan from 265 to 270 slit width:
667-UV6* Toluene in Hexane Scan: 265–270 667006
0.5; 1.0; 1.5; 2.0; 3.0
667-UV9* Hexane, reference filter Scan: 265–270 667009
Total 667EP1001 EP 10 Basic set UV/Vis
667-UV200* Filter Set contains: UV6 and UV9 Scan: 265–270 667200
Slit widths: 0.5; 1.0; 1.5; 2.0; 3.0

PL: Path length   SR: Spectral range


* with Hellma Analytics calibration certificate

EP BASIC SET UV

NEW BASIC FILTER SET FOR USP ‹857› CONFORMITY (for wavelength range: 240–400 nm and absorbance range: 0.25–1.0)

PARAMETER ARTICEL NO. CONTENT WAVELENGTH nm


Update Dec. 2019 Photometric accuracy 667S501 Filter set S501 includes: 213, 261
UV range UV506, Niacin filter 6 mg/l (0.25 Abs)
UV512, Niacin filter 12 mg/l (0.5 Abs)
UV524, Niacin filter 24 mg/l (1.0 Abs)
USP BASIC SET UV/VIS UV599, Hydrochloric acid (reference filter)

(for wavelength range: 240–640 nm and absorbance range: 0.2–2.0) Wavelength accuracy 667400 Filter set, UV5 Holmium in Perchloric acid and UV14 Perchloric acid 241, 287, 361, 451, 485, 536, 640
(reference filter)
PARAMETER ARTICEL NO. CONTENT WAVELENGTH nm
Stray Light 667107 Filter set, UV23 Potassium iodide in pure water and UV12 pure water 259 cut-off
Photometric accuracy 667020 UV20: 20 mg/l Potassium dichromate in HClO4 (0.1–0.3 Abs) 235, 257, 313, 350 (reference filter), ≥ 3 Abs at 250 nm
UV range 667060 UV60: 60 mg/l Potassium dichromate in HClO4 (0.3–0.9 Abs)
6670140 UV0140: 140 mg/l Potassium dichromate in HClO4 (0.7–2.0 Abs) Resolution 667200 Filter set, UV6 Toluene in Hexane and UV9 Hexane (reference filter) Scan from 265 to 270
667014 UV14: Perchloric acid (reference filter) slit width: 0.5; 1.0; 1.5; 2.0; 3.0

Photometric accuracy 666S200 Glass Filter Set S200 contains: 440, 465, 546.1, 590, 635 Total 667EP1002 EP Basic Set UV
VIS range F2 (0.25 Abs), F4 (1.0 Abs), F203 (2.0 Abs)

Wavelength accuracy 667005USP UV5USP: Holmium in perchloric acid 241; 250; 278; 287; 333; 345; 361;
385; 416; 452; 468; 485; 536; 640

Stray Light 667019 UV19: Acetone (pure) measured against air, ≥ 2 Abs at 300 nm Spectral range: 250–330

Resolution 667200 Filter Set, UV6 Toluene in Hexane and UV9 Hexane (reference filter) Scan from 265 to 270,
Slit width: 0.5; 1.0; 1.5; 2.0; 3.0

Total 667857V3 USP 857 Basic Set

48 49
CERTIFIED REFERENCE MATERIALS

REFERENCE PLATES FOR QUALIFYING MICROPLATE READERS WITH 30 YEAR MANUFACTURER’S WARRANTY ON ALL HELLMA ANALYTICS
DAkkS CERTIFICATE REFERENCE MATERIALS

With reference plates from Hellma Analytics you can check the photometric and wavelength accuracy of microplate readers. We’re confident of our quality and you can be confident before being sent back with a new DAkkS calibration cer-
They have the same dimensions as a microplate with 96 wells and a 6.6 mm diameter per window (height 13.0 x width 127 x of reliable results! tificate and calibration mark. Damaged filters and filters
length 85.5 mm). All Hellma Analytics reference materials come with that deviate significantly from nominal values are usually
a 30-year warranty, provided that they are regularly replaced in consultation with the customer.
recertified – every two years – at the Hellma Analytics
TYPE USAGE MATERIAL WAVELENGTH ARTICLE NO.
Nominal value of absorption (Abs.) nm
calibration laboratory. Certified reference materials sent
for recertification are carefully cleaned and recertified
666-R013 To check photometric accuracy Neutral Density Glass Filters A: 405; 450; 490; 650 666R013
NG 11 (0.25), NG 5 (0.5), NG 4 (1.0),
NG 3 (1.5), (2.5) All of our reference materials
666-R113 To check photometric accuracy Neutral Density Glass Filters A: 405; 450; 490; 650 666R113 come with a 30-year warranty,
and wavelength accuracy NG 5 (0.5), NG 4 (1.0), NG 3 (1.5), (2.0) W: 279; 361; 453; 536; 638 provided that they are regularly
Holmium Glass Filter recertified (at least every
A: Wavelength for absorbance   W: Wavelength for wavelength accuracy
two years) at the Hellma
Analytics calibration
laboratory.

30 YEAR
WARRANTY

666-R113

50 51
CERTIFIED REFERENCE MATERIALS

THE DAkkS ACCREDITED CALIBRATION LABORATORY


FROM HELLMA ANALYTICS

Proven reliability, completely documented


The Hellma Analytics calibration laboratory is the only

Sample
calibration laboratory in Germany accredited for the
certification of UV/Vis reference materials.
After careful manufacture, the reference materials are
certified according to the DAkkS regulations, using a high
performance UV/Vis spectrometer.

Advantages

● Most important information at a glance: Users are


able to test and calibrate their spectrometers by using
the values documented and certified on the calibration
certificate.
● Legally secure: DAkkS calibration certificates are
certificates from the “Deutsche Kalibrierdienst”
(German Calibration Authority) and may only be issued
by accredited partners.

Optical measuring instruments


must be regularly calibrated
or validated according to the
standard I M P O R TA N T I N F O R M A TI O N
Only if the DAkkS calibration
certificate has been issued and
the calibration mark has been
affixed, do the reference mate-
rials actually become certified
reference materials.

52 53
CERTIFIED REFERENCE MATERIALS

RECERTIFICATION

Continuously assured quality: Important parameters for recertification Fast and reliable – recertification service Returning your reference materials for
Recertification intervals for reference materials recertification
The length of intervals between the recertification of refer­ In our DAkkS-accredited calibration laboratory, your refer-
As is the case for all measuring devices, the reference ence materials depends on how frequently materials are ence materials are cleaned and are recertified in accord- Efficient processing of the reference materials you send us
materials used to verify spectrophotometers must also used, the wear associated with this, accuracy requirements, ance with your requirements using a high-performance ensures that you will be able to use your filters again with-
be checked and recertified at regular intervals (see for and the requirements of a company’s internal auditing. In spectrophotometer. If necessary, filters are repaired, or are in just a few days.
example ISO 9001: “Control of Monitoring and Measuring general, a recertification interval of 12 months is recom- exchanged following a consultation. We need your support to make this possible. Please in-
Equipment”). This allows you to ensure that you consi- mended for checking and recertifying glass filters during You will receive your filters with a new DAkkS calibration clude all information needed to process the reference
stently fulfill your in-house quality requirements and gua- the first two years of use, with an interval of 24 months certificate or Hellma Analytics calibration certificate. materials:
rantees high levels of accuracy and reliability in your thereafter. We recommend verifying and recertifying liquid Filters are usually recertified within five working days of
measurements. filters within a maximum of 12 months. Intervals should their arrival at the calibration laboratory. ● Article no.*
be specified individually in accordance with your QA ● Serial no.*
system. Recertification of reference materials from other manu­- ● Wavelength(s) to be measured*
facturers. ● Slit width(s) to be measured*
We also recertify reference materials for UV/Vis spectros- ● Documentation of measurement data prior to cleaning**
copy from other suppliers. If you require a quotation first, Yes / No
please send your inquiry via email to your local Hellma ● Quotation no. (if you have already received a quotation
partner. from us)
● Billing address
● Delivery address (if different from billing address)
● Special requests, e.g. additional wavelengths etc

GLASS FILTERS RECERTIFICATION EVERY 24 MONTHS Please enclose a copy of your order or send this via email
P L E A S E N O TE : to your local Hellma partner.
If you send your reference materials with only a delivery
PURCHASE 1st YEAR 2nd YEAR 4th YEAR 6th YEAR 8th YEAR Liquid filters can only be sent when the external note, it is essential that you indicate your order number.
temperature is above 4°C, as the liquid can freeze, Please include this on the delivery note, otherwise we will
which will destroy the reference materials. be unable to process your order.

*This information is not necessary if you enclose a copy of the current calibration certificate.
** Documentation of measurement data prior to cleaning
If you require documentation of measurement data prior to cleaning, please note this on your
order. Depending on your quality management requirements, you have two options:
1. Documentation of measurement data prior to cleaning with DAkkS certificate.
2. Documentation of measurement data prior to cleaning with simple measurement report.

LIQUID FILTERS RECERTIFICATION INTERVAL OF 12 MONTHS IS RECOMMENDED

PURCHASE 1st YEAR 2nd YEAR 3rd YEAR 4th YEAR 5th YEAR 6th YEAR 7th YEAR 8th YEAR

54 55
OPTICAL IMMERSION PROBES AND FIBER OPTIC CELLS NEW
!
a n t ly e x p a n d e dr
ific an
gn

EXTREMELY ROBUST AND

ge
si
CONFIGURATOR
HIGHLY PRECISE – IN THE
LABORATORY AND IN THE

on
us
r c

ti
e
or le
ien e
te d p ro d u ct s

PROCESS

EXCALIBUR STANDARD IMMERSION PROBE Online, atline, inline – continously reliable Benefits at a glance
Suitable for a variety of process and laboratory
measurements with high reproducibility
● World’s largest range of optical immersion probes
applications UV/Vis/NIR/IR/Raman
Fiber optic immersion probes and cells for flow-through ●

High optical path length accuracy (± 0.01 mm) measurements support engineers, researchers and ana- ● Fiber optic cells and immersion probes available

lysts to make their processes more efficient. They collect ● Various sealing technologies to choose from
ensures accurate results
accurate and reliable results when used to take online ● Extremely robust products
Chemical-resistant measuring head made from measurements during production processes. ● Interfaces to connect to instruments from
Quartz Glass Extended Range different manufacturers
In the lab, they are especially useful in cleanrooms, for
automated methods and scaling up processes.

Areas of application

● Chemical and petrochemical engineering


● Pharmaceuticals and life sciences Here you will find further information
● Food and beverage industry as well as a product configurator for
immersion probes and fiber optic cells.
● And many more
configurator.hellma.com

The benefits of optical immersion probes


and fiber optic cells include their ease of
use, flexibility and ability to boost efficiency.
Our comprehensive portfolio enables us to
meet our clients’ requirements as closely
as possible.
Dr. Oliver Mandal,
Product Manager Fiber Optic Systems

56 57
OPTICAL IMMERSION PROBES AND FIBER OPTIC CELLS

OPTICAL IMMERSION PROBES – TRANSMISSION


FOR THE UV/VIS/NIR RANGE

EXCALIBUR LAB EXCALIBUR ALL-QUARTZ

All-Round Probe All-Quartz Probe


This classic transmission probe features a broad range These Hellma Analytics all-quartz probes are outstanding
of possible applications. Whether for use in the lab, for due to their unique design which makes additional sealing
online monitoring in process environments or even for material superfluous. This makes them the ideal tool for
TDA measurements – it is always the right choice. measuring aggressive samples even at the lowest tempera-
tures – measuring beyond the limits.

UV NIR
Optical Path Length Optical Path Length Path Length mm ARTICLE NO.
1 mm, 2 mm, 5 mm, 10 mm, 20 mm 1 mm, 2 mm, 5 mm, 10 mm, 20 mm
mm (tolerance ± 0.01) mm (tolerance ± 0.01) 1 LQID01UE20AN LQID01NE20AN
probe head: 15 mm Probe head: 15 mm 2 LQID02UE20AN LQID02NE20AN
Outer Diameter
Outer Diameter probe shaft: 20 mm Quartz barrel: 18 mm 5 LQID05UE20AN LQID05NE20AN
protective sleeve: 20 mm Optical Material Quartz 10 LQID10UE20AN LQID10NE20AN
Optical Material Quartz Probe Body Material Quartz 20 LQID20UE20AN LQID20NE20AN

Probe Body Material Stainless steel (316 L) Sealing Technology Directly fused
Viton UV/Vis ((210 nm – 1100 nm, low solarization)
Sealing Technology Spectral Range WITH VACUUM JACK FOR LOW TEMPERATURE APPLICATIONS
Kalrez 4079
®
NIR (400 nm – 2300 nm)
UV NIR UV NIR
UV/Vis ((210 nm – 1100 nm, low solarization) 1.8 m external fiber optic cables with
Spectral Range Path Length mm ARTICLE NO. Fiber Optic Connection Path Length mm ARTICLE NO.
NIR (400 nm – 2300 nm) SMA connectors
1 LQID01UE20AY LQID01NE20AY
1.8 m external fiber optic cables with 1 LEIX01UE20AN LEIX01NE20AN 5 °C to 150 °C
Fiber Optic Connection Temperature Range 2 LQID02UE20AY LQID02NE20AY
SMA connectors 2 (-180 °C to 150 °C with vacuum jack)
LEIX02UE20AN LEIX02NE20AN 5 LQID05UE20AY LQID05NE20AY
Temperature Range 5 °C to 150 °C 5 Pressure Range -1 bar to 6 bar 10
LEIX05UE20AN LEIX05NE20AN LQID10UE20AY LQID10NE20AY
Pressure Range -1 bar to 6 bar 10 Immersion Depth 210 mm (10 mm optical path length) 20 LQID20UE20AY LQID20NE20AY
LEIX10UE20AN LEIX10NE20AN
Immersion Depth 100 mm (10 mm optical path length) 20 LEIX20UE20AN LEIX20NE20AN

EXCALIBUR ALL-QUARTZ NS19/35

All-Quartz Probe, tapered version with ground glass joint NS 19/35


Immersion Probe Configurator
Hellma offers the world’s largest product selection Optical Path Length
1 mm, 2 mm, 5 mm,10 mm, 20 mm
mm (tolerance ± 0.01)
and the highest level of expertise to support your
online measurements from UV to IR. Probe head: 15.5 mm
Outer Diameter
Taper: NS 19/35

Find the right immersion probe/measuring cell quic- Optical Material Quartz
kly and easily or configure your individual Probe Body Material Quartz
solution at: Sealing Technology Directly fused
UV NIR
UV/Vis (210 nm – 1100 nm, low solarization)
configurator.hellma.com Spectral Range
NIR (400 nm – 2300 nm) Path Length mm ARTICLE NO.
1.8 m external fiber optic cables with 1 LQID01UE20DN LQID01NE20DN
Fiber Optic Connection
SMA connectors 2 LQID02UE20DN LQID02NE20DN
Temperature Range* 5 °C to 150 °C 5 LQID05UE20DN LQID05NE20DN
Pressure Range -1 bar to 6 bar 10 LQID10UE20DN LQID10NE20DN
Immersion Depth 130 mm (10 mm optical path length) 20 LQID20UE20DN LQID20NE20DN
* With vacuum connection for use at low temperatures possible, from -180 °C to 150 °C

58 59
CELLS IMMERSION PROBES AND FIBER OPTIC CELLS
OPTICAL

OPTICAL IMMERSION PROBES – TRANSFLECTION OPTIC CELLS – ATR


FOR THE UV/VIS/NIR RANGE FOR M-IR RANGE

These transflection immersion probes have been specifically designed for laboratories and small volume analyses. They are KATANA TUNNEL CELL
available with fixed path lengths and very small outer diameters e.g. 3.2 mm/4 mm/6 mm. The 6 mm version offers increased
flexibility due to interchangeable path length tips. ATR tunnel cells provide a high degree of analytical accuracy.
They are ideal for the chemical analysis of liquids in the IR range.

FALCATA LAB
KATANA TNL 120 (Micro) KATANA TNL 130
with 3.2 mm and 4 mm diameter Reflections 10 11
These micro immersion probes have been specifically developed for
measurements in small volumes. Due to their slim form, less sample Aperture 8 mm 16 mm
material is required for a measurement to be taken. Filling Volume 40 µl 2 ml

Optical Path Length FALCATA STANDARD IMMERSION PROBE 3.2 MM DIAMETER Optical Material ZnSe, (AMTIR-1, ZNS, others on request)
5 mm, 10 mm
mm (tolerance ± 0.02) Spectral Range NIR, IR to: 670 cm-1 (ZnSe); 880 cm-1 (AMTIR-1); 980 cm-1 (ZnS)
UV NIR
Outer Diameter 3.2 mm/4 mm Body Material 316 L 316 L
Optical Path Length mm ARTICLE NO.
Optical Material Quartz Sealing Technology Kalrez® 6375 Kalrez® 6375
5 LFCI05UE20AN LFCI05NE20AN
Probe Body Material Stainless steel 1.4435 (316 L) ARTICLE NO. CKHZZSEIA000WNNAN CKDZZSEIA000WNNAN
10 LFCI10UE20AN LFCI10NE20AN
Sealing Technology Epoxy glue
Also available as heatable version.
UV/Vis (210 nm – 1100 nm, low solarization)
Spectral Range
NIR (400 nm – 2300 nm)
1.8 m external fiber optic cables with FALCATA STANDARD IMMERSION PROBE 4 MM DIAMETER
Fiber Optic Connection
SMA connectors UV NIR
Temperature Range 5 °C to 100 °C Optical Path Length mm ARTICLE NO.
Pressure Range -1 bar to 6 bar 5 LFDI05UE20AN LFDI05NE20AN
Immersion Depth 75 mm/130 mm (10 mm path length) 10 LFDI10UE20AN LFDI10NE20AN

FALCATA LAB OPTICAL PROBE – RAMAN


with 6 mm diameter
Increased flexibility due to interchangeable path length tips. UV NIR
Optical Path Length mm ARTICLE NO. TIDUS LAB
1/2/5/10/20 LFEITSUE20AN LFEITSNE20AN
The RFP-400 Raman series was developed for laboratory use.
1 LFEI01UE20AN LFEI01NE20AN
A big advantage of this series is the high flexibility due to
2 LFEI02UE20AN LFEI02NE20AN exchangeable optics and filters.
5 LFEI05UE20AN LFEI05NE20AN
10 LFEI10UE20AN LFEI10NE20AN TIDUS RFP 442
Optical Path Length 1 mm, 2 mm, 5 mm, 10 mm, 20 mm
mm (tolerance ± 0.02) through interchangeable tips 20 LFEI20UE20AN LFEI20NE20AN Outer Diameter 12,7 mm
Outer Diameter 6 mm Optical Material Sapphire
Optical Material Quartz Probe Body Material 316 L
ACCESSORIES INTERCHANGEABLE PATH LENGTH TIPS
Probe Body Material Stainless Steel 1.4435 (316 L) Path Length Tips Sealing Technology PTFE coated nickel alloy
Sealing Technology Epoxy glue Optical Path Length mm ARTICLE NO. Spectral Range 150 – 4000 cm-1 (Raman shift)
UV/Vis (210 nm – 1100 nm, low solarization) 1/2/5/10/20 Set IPLTL1SQSI Excitation wavelength 532 nm (633, 785 nm possible)
Spectral Range
NIR (400 nm – 2300 nm)
1 IPLTL1AQSI Fiber Optic Connection FC (Excitation); F-SMA (Detection)
1.8 m external fiber optic cables with
Fiber Optic Connection Temperature Range 5 to 200 °C
SMA connectors 2 IPLTL1BQSI
Temperature Range 5 °C to 100 °C 5 IPLTL1CQSI Pressure Range 0 to 100 bar
Pressure Range -1 bar to 6 bar 10 IPLTL1DQSI Immersion Depth 200 mm
Immersion Depth 175 mm (10 mm optical path length) 20 IPLTL1EQSI ARTICLE-NO. RA2BSSHDKALM200N

60 61
OPTICAL IMMERSION PROBES AND FIBER OPTIC CELLS
CELLS

ACCESSORIES ACCESSORIES

CONNECTION TO THE SPECTROPHOTOMETER FIBER OPTIC INTERFACE FIBER OPTIC CABLES


This accessory is used when the available spec- Fiber optic cables can be supplied with either SMA connectors
Direct Connection via SMA Connector trophotometer does not have an SMA connector. or special collimating lenses to suit the application that they
Immersion probes from Hellma Analytics can be connected Probes with collimators as a connection option can are being used for.
to spectrophotometers in two different ways. If the spectro- be directly connected to the interface. The adapter
photometer has SMA sockets, an immersion probe with SMA has the dimensions of a standard cuvette and fits Core Diameter 600 µm
connectors can be directly connected. For this purpose, fiber into a standard cuvette holder. Numerical Aperture 0.22
optic cables with SMA Collimator are suitable.
Beam Diameter (lens) 3.7 mm
Max. Temperature 150 °C
fiber optic cable

spectrophotometer Effective Aperture 4 mm Diameter


fiber optic cable connector
Outside Dimension 60 mm x 12.5 mm x 12.5 mm
or SMA sockets immersion probe
Center Height 8.5/15/20 mm
light path 190 nm to 2300 nm
Wavelength Range
depends on cables used
Notes Other center heights on request

Center Height ARTICLE NO. A large selection of fiber optic


8.5 mm 662-85-UVNIR-46 cables can be found on:
source light beam detector 15 mm 662-15-UVNIR-46 configur ator.hellma.com
20 mm 662-20-UVNIR-46
"Accessories".

Connection via Fiber Optic Interface EXTERNAL CELL HOLDER


For spectrophotometers equipped only with a cell holder, an
immersion probe can be connected to the instrument using the The external cell holder is useful when the spectrophotometer does
not have an internal cell holder or when measurements with cells
fiber optic interface 662.000 and special fiber optic cables with
are to be made at some distance from the spectrophotometer, e.g. in
2 collimating lenses.
a fume hood. To connect this cell holder properly to your system you
will require 2 x 1 m fiber optic cables in the corresponding spectral
range. You should select the option “SMA-Collimator”.

Material Aluminum, Black Anodized

spectrophotometer Dimensions 123 mm x 40 mm x 45 mm


fiber optic cable
Temperature of
immersion probe Max. 120 °C (Quartz Cells Only)
solution in cell
light path These must be ordered separately
light beam Fiber Optic Cables
(see page 65)
Suitable for cells with path length 1 mm to
Notes
20 mm
ARTICLE NO. 664-15-71
detector
source
fiber optic interface

62 63

You might also like