1. INTRODUCTION
The crycooler cools the heat shields inside the magnet, thereby reducing the helium boil-off rate.
2. INSTALLATION REQUIREMENTS
Note:
The hospital should be warned of the possible risks of connecting the compressor to a cold water
distribution system. A failure of this system will result in a shutdown of the compressor which will cause
increased helium boil-off.
The helium gas lines are made of soft cupper, which can be easy bend.
Minimum bend radius : 457 mm without a tube bender
: 57 mm when using a tube bender
The two helium gas lines (supply & return) are the same.
With the compressor of the crycooler the following items for connecting the water supply and return line
are delivered:
- 12 metre nylon water hose, 1/2 inch outer diameter.
- 2 compression fittings for 1/2 inch nylon water hose.
- 2 hose couplings, 1/2 inch outer diameter.
3. SPECIFICATIONS
*) The lowest inlet pressure can be 300 kPa when short connections to the water supply are used.
4. PREVIOUS SYSTEMS
4.1. INTRODUCTION
Systems delivered before November 1996 have other helium gas lines.
These systems have: two He gas lines for supply, instead of one
two He gas lines for return, instead of one
RF feedthrough couplings, instead of RF feedthrough clamps
If a system has to be moved, it is possible the He gas lines get damaged or can not be used anymore.
He gas line inside 10 m: 4522 150 12251
He gas line outside 15 m: 4522 150 12261
It is possible to replace the He gas lines by the new type of He gas lines.
He gas line 20 m: 4522 150 15771
RF feedthrough clamp: 4522 150 15781
In this case it is also possible to use the 10m or 15m He gas lines instead of the 20m He gas lines.