Principles of Chromatography
Principles of Chromatography
Principles of Chromatography
Principles of Chromatography
Method Development
Narasimha S. Lakka and Chandrasekar Kuppan
Abstract
This chapter aims to explain the key parameters of analytical method develop-
ment using the chromatography techniques which are used for the identification,
separation, purification, and quantitative estimation of complex mixtures of
organic compounds. Mainly, the versatile techniques of ultra−/high-performance
liquid chromatography (UPLC/HPLC) are in use for the analysis of assay and
organic impurities/related substances/degradation products of a drug substance
or drug product or intermediate or raw material of pharmaceuticals. A suitable
analytical method is developed only after evaluating the major and critical separa-
tion parameters of chromatography (examples for UPLC/HPLC are selection of
diluent, wavelength, detector, stationary phase, column temperature, flow rate,
solvent system, elution mode, and injection volume, etc.). The analytical method
development is a process of proving the developed analytical method is suitable
for its intended use for the quantitative estimation of the targeted analyte present
in pharmaceutical drugs. And it mostly plays a vital role in the development and
manufacture of pharmaceuticals drugs.
1. Introduction
1
Biochemical Analysis Tools - Methods for Bio-Molecules Studies
and purification of pharmaceutical drugs [3]. In this chapter, the authors discuss
the principles for chromatography method development using ultra/high-perfor-
mance liquid chromatography (UPLC/HPLC) techniques for the analysis of assay
and organic impurities/related substances/degradation products of pharmaceuticals
(any drug product/drug substance/intermediate/raw material of pharmaceuticals).
These techniques are developed substantially as a result of the work of Archer John
Porter Martin and Richard Laurence Millington Synge during the 1940s and 1950s, for
which they won the 1952 Nobel Prize in Chemistry [4]. Commonly used character-
izing technique in pharma industry is liquid chromatography (e.g., HPLC, UPLC,
and LC–MS). Each one varies in the stationary phase and operational conditions.
HPLC and UPLC can be used as a quantitative technique if coupled with a mass
detector (MS) to elucidate the structure of the molecule and quantification.
In pharma industry specific, stability-indicating HPLC/UPLC methods have to
be developed to estimate the assay and to quantitatively determine the impurities of
new drug substances and drug products [5]. Assay is a quantitative test of a sub-
stance to determine the amount of an individual components present in it. Impurity
is an unknown component of drug substance that is not the chemical entity. Assay
and impurity tests are major and critical quality attributes of the pharmaceutical
dosage forms which help to check and ensure the quality, safety, and efficacy of
drug substances and drug products. This chapter will discuss the various param-
eters that have to be chosen to run the chromatography in order to have a better
separation and maximum purity. The process of changing the conditions in order
to design a best method run for a particular drug mixture or compound is called the
analytical method development.
• Selection of detector
○○ pH of buffer
○○ Mobile-phase composition
2
Principles of Chromatography Method Development
DOI: http://dx.doi.org/10.5772/intechopen.89501
• Selection of diluent
• Methods of extraction
• Samples to be used
• Finalization of wavelengths
• Stability of solution
• System suitability
• Quantification methods
iii. Stability profile of the drug substance with respect to storage conditions
(sensitivity of the drug towards light, heat, moisture etc.) is useful as it helps
in adopting the suitable/adequate precautions while handling drug and its
formulated products.
3
Biochemical Analysis Tools - Methods for Bio-Molecules Studies
a lot in developing the method for separation of all possible impurities and
degradation products of targeted analyte. It should be borne in mind that
impurity profile may vary depending on the manufacturing process (which
uses different methods, precursors, and conditions), which makes it clear
that not all manufacturing processes yield the same impurity profile.
v. Metabolic pathway is a chemical reaction which occurs within a cell when the
drug molecule reacts with an enzyme and forms a metabolite [6]. Metabolic
pathway gives the information on oxidation, reduction, and hydrolysis
products which gives critical inputs on the possible degradation products.
vii. Checking the polarity of the drug molecule using the functional groups as
elucidated from structural analysis techniques. By comparing the structures
of impurities and degradation products with the structure of drug molecule,
it will help in understanding the polarity based on the nature of functional
groups. This makes the scientists’ job easy in choosing the right solvents with
either lesser or higher in polarity than the compound of interest.
viii. Estimation of maximum daily dose (MDD). Calculate the reporting, identi-
fication, and qualification thresholds of drug substance and drug product
based on the maximum daily dose as per ICH Q3A guideline [7, 8]. MDD info
can also be obtained from physical desk reference (PDR), innovator product
information leaflet (PIL), and the website of RX-list (www.rxlist.com).
Chromatography can be operated by two ways, normal mode and reverse phase
modes. The choice of the mode is very important, which is dependent on the type of
sample which has to be separated. In general, the usage of reversed-phase chro-
matography (in which the mobile phase is polar and stationary phase is nonpolar
in nature) is the preferred mode for most of the molecules, except in the case of
isomer (enantiomers) separation where the normal-phase chromatography (in
which the mobile phase is nonpolar and stationary phase is polar in nature) is used.
Revered-phase chromatography separates the components with a good resolution
based on their hydrophobicity. A compound with a greater polarity elutes earlier,
and those with the least polarity elute later.
4
Principles of Chromatography Method Development
DOI: http://dx.doi.org/10.5772/intechopen.89501
If the compounds do not absorb and if they do not have chromophores, other
detectors like refractive index detector (RID) and evaporative light scattering detec-
tor (ELSD)/corona-charged aerosol detector (CAD) can be used for the quantitative
determination of assay and impurities [11]. If the compounds of interest contain a
part, which is non-chromophoric, which may likely be cleaved and produce a non-
chromophoric impurity, then both UV and other detectors like RI/ELSD/CAD can
be coupled in order not to miss any impurity.
Alternatively, non-chromophoric compounds can also be analyzed by UV
after converting it into a derivative which will be active. But the usage of deriva-
tives has to be carefully assessed keeping in view the functional group involved
in the derivatization reaction [12, 13]. In case the molecule of interest is having
fluorescence properties, a fluorescence detector (FLD) can be used for com-
pounds for which structural information is available [14]. But when FLD is to
be used for estimation of unknowns, it needs to be carefully assessed whether
fluorescence properties are available in all possible impurities and
degradation products.
The choice of the right column (stationary phase) is the basis of the whole
technology. Most chromatographic separations are achieved due to a wide variety
of columns available in the market and due to their flexibility in changing and
controlling the parameters. A widely used choice of column material is silica
either as neat or modified depending on the nature of the solute mixture in
normal-phase chromatography, wherein the eluent (mobile phase) is nonpolar
an organic solvent. The silanol groups on the surface of the silica give it a polar
character.
Though silica remains the most common support for liquid chromatography
(LC) columns, other commonly used materials are cross-linked organic poly-
mers, zirconia, etc. The silica support for columns was gradually modified for
the betterment through the years by three different manufacturing technolo-
gies commonly described as “evolution through three generations.” The initial
process started with type A silica where the raw material used is from inorganic
sols. A slightly modified type A silica by performing a chemical treatment to
remove the metal impurities is termed as a second-generation material which is
called as base-deactivated silica. Third generation silica (type B) is an altogether
new process which uses organic sols instead of inorganic sols. These materials
are similar in properties to the second-generation silica because both have a
minimum level of metal impurities. Silica-based liquid chromatography columns
with a different percent of cross-linking and functionalization of silanol groups
with substituted aliphatic and aromatic moieties were designed for varying
polarities of the separating medium. An increasing order of functionalized
silica is represented below with alkyl groups at the nonpolar end, phenyl and
amino functionalized in the moderate polar region, and cyano and silica groups
at the polar end.
5
Biochemical Analysis Tools - Methods for Bio-Molecules Studies
• Short (30–50 mm)—can result in short run times and low back pressure
End capping: End capping reduces peak tailing of polar compounds that
interact excessively with the otherwise exposed, mostly acidic silanols. Non-end
capped packing provides a different selectivity than do end-capped packing,
especially for polar compounds. A pictorial representation of difference in end
capping is shown below.
6
Principles of Chromatography Method Development
DOI: http://dx.doi.org/10.5772/intechopen.89501
Buffer and its efficiency play an important role in deciding the peak symme-
tries (shapes) and peak separation. Various types of organic/inorganic buffers are
employed for achieving the required separation. The most commonly used buffers are:
• Triethylamine/diethylamine buffers
• Buffers with various ion-pair reagents like tetrabutyl ammonium hydrogen sul-
fate, butane sulfonic acid, hexane sulfonic acid, heptane sulfonic acids, etc.
The choice of buffer is to reduce the tailing factor for each peak separated which
occurs due to varying ionic strength. The retention time of analyte(s) is delayed and
got separated well when more concentrated buffer is used [15]. Better separation
happens when the molarity of buffer used is in the range of 0.05 to 0.20 M. The
concentration of buffer is chosen by carefully choosing the composition of organic
mobile phase.
Depending on the need of the chosen mixture of separation, the strength of the
buffer can be increased or decreased if necessary to achieve the required separa-
tion, and it can be varied between 10 and 20%, and the effect of variation has to be
studied in detail before using. But it should be ensured that increased or decreased
buffer strength should not result in precipitation or turbidity either in mobile phase
during operation or during storage in refrigerator. Before using the chosen buffer
of specific strength to run a column, test experiments have to be done in optimizing
the separation to avoid peak tailing, better separation, and reproducibility.
7
Biochemical Analysis Tools - Methods for Bio-Molecules Studies
2.5.2 pH of buffer
It is well reported in literature that to achieve better efficiency, binary and ter-
tiary solvent mixtures are used along with other components like buffer and acids
or bases. The ratio of the organic versus (vs.) aqueous or polar vs. nonpolar solvents
is varied accordingly to get better separation. This is due to the fact that a fairly large
amount of selectivity can be achieved by choosing the qualitative and quantitative
composition of aqueous and organic portions. Experiments shall be conducted with
mobile phases having buffers of different pH and different organic phases to check
for the best separations between the impurities. Most chromatographic separations
can be achieved by choosing the optimum mobile phase composition [18].
8
Principles of Chromatography Method Development
DOI: http://dx.doi.org/10.5772/intechopen.89501
reduces or increases the polarity of component and helps in increasing the elution
time difference. Careful choice of an appropriate ion-pair reagent is required in
such cases to get the necessary selectivity. A dedicated LC column is used when an
ion pair reagent (0.0005 M to 0.02 M) is intended to employ for specific analysis,
but an appropriate cleaning procedure has to be established to enhance the lifetime
of the column material. Alkyl ammonium salts (tertiary or quaternary) and alkyl
sulfonate salts are the most useful in the separation of acidic and basic compounds,
respectively. Sodium perchlorate can also be used for acidic components.
9
Biochemical Analysis Tools - Methods for Bio-Molecules Studies
phase is pumped. To avoid the pumping problems due to the low viscous solvents
like acetonitrile in mobile phase, at least 10% aqueous portion could be added to
the organic phase.
While optimizing the gradient program, it is important to monitor the follow-
ing. Pressure graph is needed to be monitored so as to ensure that the overall system
pressure will not cross 400 bar or 6000 psi at any point during the run. Flow rate
has to be physically cross-checked by collecting the output from the detector during
the run at different time intervals, especially when the gradient is running with
higher organic-phase composition so as to ensure that there were no pumping prob-
lems during the run when mobile phases of different compositions are pumped. It is
also important to optimize the program for initialization after each run and before
going for the next injection. The program for initialization shall be optimized such
that there shall be no carry-over to the next run and the system stabilizes with initial
composition before the next injection.
One standard program which can be used for optimizing is discussed below.
For starting a method development, a solvent gradient system is always preferred.
Initially, start with a gradient of 50:50 buffer and mobile phase, and change the
program linearly up to 5:95, and retain the ratio for at least 30 minutes. Then try
with a gradient of 95:5 and the program linearly changed up to 5: 95, and retain for
at least 30 minutes. The typical gradient program is as follows:
• Selection of diluent based on extraction efficiency and peak shapes: Select the
diluent for finished dosage forms, in which the analyte should be extracted at
least 95% for assay and 90% for organic impurities. Calculate the % extraction
against pure standard compound in the concentration of linear range, (prefer-
ably <1 AU) by diluting the test preparation.
• The peak shapes of all compounds should be good in the selected diluent:
Select an initial flow rate of 1.0 mL/min or 1.5 mL/min and select column
temperature as ambient (25–30°C).
10
Principles of Chromatography Method Development
DOI: http://dx.doi.org/10.5772/intechopen.89501
procedures, heating can be adapted if the substance is stable and should not pre-
cipitate upon cooling to room temperature [33–34].
• Use the reaction mass/mother liquor/what if study samples for the above study
if all the impurities samples are not available in the beginning for method
development.
The test concentration and injection volume are generally chosen based upon
the response of API peak at the selected detector wavelength [35]. However, the
test concentration shall be finalized after it is proven that drug (API) is completely
extractable at the selected test concentration. After finalizing the test concentra-
tion and diluent, prepare a test solution, and keep the filtered solution in closed
condition on a bench top, and check whether the solution has any precipitation or
turbidity after 24 hours. Generally, the test solution must be clear and should not
show any turbidity or precipitation.
It’s a method of subjecting the drug substance or drug product to stress with
varied strengths of stressing agents to obtain the degradation. The stressed samples
were analyzed using an LC system equipped with a PDA detector and monitored for
the separation of degradation products formed under the stressed conditions and
the peak purity of the analyte peak. The method is considered as stability-indicating
for the estimation of the drug if it meets the peak purity requirement [36, 37].
11
Biochemical Analysis Tools - Methods for Bio-Molecules Studies
The major forced degradation studies which are to be carried out are as follows:
a. Thermolytic degradation
This stress testing method studies the degradation that is caused by exposure
to temperature high enough to induce bond breakage. Solid-state reactions often
proceed in an autocatalytic pathway involving an induction period (lag), fol-
lowed by a period of rapidly increasing degradation and then slowing down of the
degradation rate as the compound is consumed. Thus, solid-state reaction kinetics
will often follow an S-shaped curve when degradation vs. time is plotted. Thus,
before conducting thermolytic degradation, determine the melting point of the
compounds of interest. Then, choose a temperature of 70°C for all the drugs for
which melting point is <100°C, or choose a temperature which is 40°C below the
melting point.
For the compounds for which melting point is >150°C, stress the samples at
105°C. Keep the samples directly exposed in the oven for 1 week or until about
2–20% degradation is achieved, whichever is earlier. Stress the drug substance,
placebo, and drug product separately. In the case of the multicomponent drug
products, stress testing of placebo with other actives excluding the one at a time
shall be performed additionally.
b. Hydrolytic degradation
12
Principles of Chromatography Method Development
DOI: http://dx.doi.org/10.5772/intechopen.89501
c. Humidity stress
Stress the samples to 90% humidity for 1 week. Stress the drug substance,
placebo, and drug product separately. In the case of the multicomponent drug
products, stress testing of placebo with other actives excluding the one at a time
shall be performed additionally.
d.Oxidative degradation
e. Photolytic degradation
13
Biochemical Analysis Tools - Methods for Bio-Molecules Studies
Peak purity can be evaluated for the main peak and the major degradants which
have the peak heights less than 1 AU. Identify the degradation products by co-
injection, in case of known impurities and have comparable spectra.
If any known impurity is observed to be increased in stress, it can be examined
properly. If process impurity is found to be increased in stress study, it needs to be
assessed whether there is any secondary pathway of formation of this impurity via
some other degradant route.
After conducting these studies, verify the chromatograms, and observe any
peaks merging with respect to main peak and any critical pairs. If any situations
were arrived, adjust the mobile-phase compositions, column parameters, etc. and
conclude the method parameters.
After method finalization, check the method using different detectors (RI/
ELSD/CE/LC–MS), and compare the data with other detectors like UV, fluores-
cence, etc. The UV inactive components can be found with these experiments.
Identify the mass of major degradant which may be formed greater than 1.0% in
stress studies, and try to establish the structures.
Mass balance is a process of adding together the assay value and levels of deg-
radation products to see how closely these add up to 100% of the initial value. It
is important to have methods that detect all major degradation products. This is
generally done by performing the assay of forced degraded samples and assesses the
mass balance. Mass balance has to be achieved at least up to 95% level. If it is less
than the required criteria, investigation has to be done and justified. The following
are some of the reasons for not achieving the mass balance.:
• Parent compound may be lost from the sample matrix, due to insolubility, vola-
tility, or adsorption losses
14
Principles of Chromatography Method Development
DOI: http://dx.doi.org/10.5772/intechopen.89501
System suitability tests verify and ensure whether the system’s performance is
acceptable at the time of analysis in accordance with the criteria set forth in the
procedure or not. System suitability parameters are chosen based on the criticality
of separation. In general, resolution factor for the two adjacent peaks or closely
eluting peaks is selected as a system suitability requirement. If the separation of
impurities from each other and from API peak is found to be satisfactory, there
is no need to keep a resolution factor as a system suitability parameter. In such a
case, only a diluted standard reproducibility can be adopted as a system suitability
requirement. Before finalizing the system suitability parameters, the separation
needs to be studied during the robustness study to understand its behavior during
the various deliberate changes in method.
System suitability checking must be performed on two different make of HPLC
systems whenever the separation of any impurities is critical. For in-process-related
impurity issues, the quantification limit (QL) concentration is to be injected, and
signal to noise ratio (S/N) must be kept as a system suitability parameter.
The relative response factor is used to correct the difference in the detector
response of impurities with respect to the main analyte peak. It is mainly used to
control the impurities or degradation products in a drug substance or drug product.
RRF is established for all the known impurities using any of the slope methods. The
standard solutions of API and all impurity can be prepared in at least five different
15
Biochemical Analysis Tools - Methods for Bio-Molecules Studies
concentrations in the range of 0.1–1.0% (e.g., 0.1, 0.3, 0.5, 0.7, and 1.0%) and
analyzed using the liquid chromatography. RRF is calculated by using the slope of
the respective impurity and slope of the main drug (API) [38, 39].
The following methods can be used for the quantitative determination of assay
and organic impurities [40, 41]:
a. External standard method: This method is used for the assay and impurity
estimation in a given sample, where the impurities are estimated using the
respective impurity standard and without the API standard peak. It’s possible
to estimate the concentration from calibration curve.
b. Area normalization: If the RRF value of known impurity is close to the API
(analyte), i.e., 0.9–1.1, the area normalization method is chosen for quan-
tification. The recovery needs to be established without using the response
factors.
c. Diluted standard method: If the RRF values of impurities are different from the
analyte, the diluted standard method can be chosen.
3. Conclusion
16
Principles of Chromatography Method Development
DOI: http://dx.doi.org/10.5772/intechopen.89501
Author details
© 2019 The Author(s). Licensee IntechOpen. This chapter is distributed under the terms
of the Creative Commons Attribution License (http://creativecommons.org/licenses/
by/3.0), which permits unrestricted use, distribution, and reproduction in any medium,
provided the original work is properly cited.
17
Biochemical Analysis Tools - Methods for Bio-Molecules Studies
References
18
Principles of Chromatography Method Development
DOI: http://dx.doi.org/10.5772/intechopen.89501
ASIA PACIFIC VOLUME 8. 2005. [22] Shibue M, Mant CT, Hodges RS.
Available from: http://alfresco. Effect of anionic ion-pairing reagent
ubm-us.net/alfresco_images/ hydrophobicity on selectivity of
pharma/2014/08/22/806d27ce- peptide separations by reversed-phase
71f0-4239-b9a6-f0541666d136/ liquid chromatography. Journal of
article-254604.pdf [Accessed: Chromatography A. 2005;1080(1):68-
10 July 2019] 75. DOI: 10.1016/j.chroma.2005.03.035
19
Biochemical Analysis Tools - Methods for Bio-Molecules Studies
[29] Gradient HPLC Solvent Delivery [37] Jain D, Basniwal PK. Forced
System: WPP10. Waters. Available from: degradation and impurity profiling:
https://gimitec.com/file/wpp10.pdf Recent trends in analytical perspectives.
[Accessed: 10 July 2019] Journal of Pharmaceutical and
Biomedical Analysis. 2013;86:11-35
[30] Gradient Design and Development,
Breaking the Bad Gradient Cycle: [38] Kalyana Chakravarthy VV,
Agilent Technologies. Available from: Kishore Babu G, Lakshmana Dasu R,
https://www.agilent.com/cs/library/ Prathyusha P, Aparna Kiran G. The role
slidepresentation/public/Gradient%20 of relative response factor in related
Design%20and%20Development.pdf substances method development by high
[Accessed: 10 July 2019] performance liquid chromatography
(HPLC). Rasayan Journal of Chemistry.
[31] Good Habits for Successful Gradient 2011;4(4):919-943
Separations: Getting the Most From Your
Method. Agilent Technologies. Available [39] Bhattacharyya L, Pappa H,
from: https://www.agilent.com/cs/ Russo KA, Sheinin E, Roger L. Williams:
library/slidepresentation/Public/ U.S. pharmacopeia. The use of
Successful%20Gradient%20Separations. relative response factors to determine
pdf [Accessed: 10 July 2019] impurities. Pharmacopeial Forum.
2005;31(3). Available from:
[32] Wrezel PW, Chion I, Pakula R. Key https://www.researchgate.net/
factors in sample diluent selection for publication/295423011_The_use_
HPLC assays of active pharmaceutical of_Relative_Response_Factors_to_
ingredients. LCGC North America. determine_impurities
2005;23(7):682-686
[40] Altria KD. Essential peak area
[33] Slack GC, Snow NH. 8 HPLC normalisation for quantitative impurity
sample preparation. Separation Science content determination by capillary
and Technology. 2007;8:237-268. DOI: electrophoresis. Chromatographia.
10.1016/S0149-6395(07)80014-6 1993;35(3-4):177-182
20