Morehouse

Calibration
Software
Instruction Manual
Data acquisition software using polynomial equations

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 1 of 29 Rev: 0
Table of Contents
1. Software Summary

1.1. Hardware Devices

1.2. Hardware Communication

1.3. Software Appearance

1.4. Software Releases

1.5. Data Storage

2. Installation and Setup

2.1. Download

2.2. Install Morehouse Software Calibration

2.3 Windows Defender SmartScreen

2.4. Necessary Components

2.4.1. Microsoft Excel

2.5. Indicator Drivers

3. Operating the Software

3.1. Launching the Software

3.2. Interacting with the Software

3.2.1. Start/Setup Tab

3.2.2. Calibration Tab

3.2.3. Capturing Data

3.2.4. Report Tab

3.2.5. Help Tab

3.2.6. Indicator Screen

3.2.7. Shell Screen

3.2.8. Instrument Screen

3.3. Detailed Operation

3.3.1. Importing Certificates

3.3.2. Configuring Active Indicator Reading Panes

3.3.3. Understanding the Data Grid

3.3.4. Exporting Captured Data

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 2 of 29 Rev: 0
4. Quick Start Guide

4.1. Checklist

4.2. Detailed Guide

5. FAQ - Frequently Asked Questions

5.1. Removing Force Units

5.2. Force Summation

5.3. Missing Indicator

5.4. Missing HADI Indicator

5.5. Resetting COM Ports

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 3 of 29 Rev: 0
1. Software Summary
The Morehouse Calibration Software (the software) is provided by Morehouse Instrument Company (Morehouse)
and is designed to be used with indicators and calibration coefficients. The software is capable of supporting an
unlimited number of devices, simultaneously. Each device connected via the software will have its mV/V value read
from it, converted into a reading using the user supplied coefficients, and displayed on the screen. The software is
also capable of reading from multiple devices simultaneously and adding their calculated values together.

1.1. Hardware Devices


The software is designed to initiate communication with hardware devices over a Virtual Communication Port
(VCP). Most devices will require a third party driver to be installed on the computer to turn the USB connection
into a VCP. These drivers are included in the ZIP archive to install for the specified models. Some devices may also
require a custom RS232 cable as well as a customer supplied RS232 to USB adapter. The adapter may have its own
driver the customer may need to install.

Currently, the only known limitation on the number of simultaneously connected devices is determined by the
computer they’re connected to. Most computers are limited to a maximum of 255 Serial Ports. Theoretically, the
software would be able to simultaneously read from all 255 ports. Using more ports than necessary will cause the
hardware discovery process to take longer, so it is recommended to limit the number of connected devices to
about 10-15. Leaving more connected should not cause an issue with the program, but it will significantly increase
the chance of a user (technician) making a mistake.

1.2. Hardware Communication


The software is designed to communicate with various devices utilizing a Serial Port object. This behavior can also
be reproduced using an RS232 terminal or even via PowerShell. As each device is different, a collection of baud
rates and other configurations are tried to establish a connection with a device. Unfortunately, some serial port
devices are sensitive to unknown commands being sent to them. This may cause unexpected behavior, including
displaying an error message on the screen. To prevent this issue from happening, it is recommended to disconnect
these devices from the computer first.

The software is also designed to gracefully ignore serial devices that don’t respond well to attempted
communications. This may include a Bluetooth device, the computer manufacturer’s enabled serial device, or other
devices. These devices may be left connected to the computer, but if an issue arises, it is recommended to
disconnect them to ensure they aren’t the reason for the issue.

For devices that a successful communication has been established, the software can communicate with the device
using a handful of commands. Devices that require remote operation, like the HBM DMP40, the software is able to
enter/exit remote operation mode by itself. For some devices, this generally locks out the front panel, preventing
human interaction without restarting the device. The software is designed to enter remote mode when attempting
to communicate with the device, but will automatically exit remote mode when communication has been finished.
The software will also exit the device from remote mode if the technician requests that live readings are turned off.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 4 of 29 Rev: 0
The software is also capable of requesting the device’s serial number (used to verify the manufacturer and model),
tare the meter, and obtain the current reading from the device. The serial number is only requested during the
discovery phase to look for a response. Once a response has been received, the software will request the serial
number again to verify. The software will only tare the meter upon the technician’s request, either via individual
device taring or by requesting a tare of all devices.

It has been reported that some devices do not always tare when prompted. This
may be due to the current reading of the device being higher than the
manufacturer's programmed threshold. This may also indicate possible damage to
the equipment. Please contact Morehouse for further assistance.

When the software attempts to retrieve a reading from the device, it will typically request the mV/V reading of the
device. Not every device has a mV/V reading, like the Morehouse G501F, so the software will opt to read what is
currently displayed on the device. This means that if the calibration was done in LBF, the G501F will need to be set
to the LBF units to get accurate readings. If a different unit is required than what the device was calibrated for (KGF
instead of LBF), use the software’s built in unit conversion.

The software also has a built-in terminal for communicating with a device. This allows messages to be sent and
received from devices from within the program without requiring a different application.

1.3. Software Appearance

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 5 of 29 Rev: 0
The software is designed using the Metro theme based on the Windows 8 operating system. This theme utilizes a
green style to closely match the colors of Morehouse. The software’s user interface (UI) is tailored towards smaller
resolutions. Currently, the software supports resolutions of 720p (100% UI scaling) and greater and 1080p (150%
UI scaling) and greater. Due to the target resolutions, the UI may not appear optimally on better resolutions. To
alleviate this, it’s recommended to manually set the size to provide a pleasant appearance.

The software is also equipped with a dark mode. The ability to change the software between light and dark mode
gives the technician the flexibility to determine the correct contrast to best observe the readings. The current light/
dark mode is set in the main window and is applied to all other windows of the software. Dark mode can be
toggled directly within the software. Dark mode will also be enabled when the software starts if the computer’s
color personalization is set to Dark or the default app mode is set to Dark.

Internationalization has been implemented within the software. This feature is limited to numbers that are
displayed throughout the software. The numbers will be displayed with the thousands and decimal separator
according to the computer’s locale.

1.4. Software Releases


To better manage the software’s releases, all releases are now hosted on GitHub. Morehouse’s GitHub
organization is located at https://github.com/mhforce but only the releases are available to the public. This
provides Morehouse more flexibility with uploading new versions and updating information for previous versions.

For customers that are unable to access https://github.com from work and can’t get an exception for Morehouse’s
GitHub organization, please contact Morehouse for a copy of the latest release.

1.5. Data Storage


The software is designed to store all data in a JSON file in the software’s program data directory
(%ProgramData%/Morehouse Calibration Software/data.json). This is a plain text file format and can be
edited by any text editor program. Manual modifications to this file may cause unexpected behavior. It is
recommended to only modify this file if instructed by Morehouse.

If a version before 1.2.0 of the software was used, the program used a SQL database file. Version 1.2.0 is able to
migrate the data from the SQL database to the JSON file. If the software is unable to migrate the data, an error will
be presented and the program will exit. In the event the migration fails, the database file can either be deleted and
start fresh, or Morehouse can be contacted for assistance.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 6 of 29 Rev: 0
2. Installation and Setup
2.1. Download
To download the latest version of the software, visit https://github.com/mhforce/MorehouseCalibrationSoftware-
Releases/releases/latest/download/Morehouse_Calibration_Software.zip. A ZIP archive containing the latest
versions of the installer, this instruction manual, and meter drivers will be downloaded. Unzip the archive and
follow the next steps to get the software installed.

2.2. Install Morehouse Software Calibration


The software is installed using an offline installer and is named Morehouse Calibration Software.msi.
Follow the instructions in the installer. The installer may request elevated permissions or ask to approve the
installation. Failure to complete this step will prevent the software from being installed. If admin privileges are not
available, choose to install the software for Just me instead of for Everyone and change the installation
directory to a location that the user has access to. Program Files and Program Files (x86) are generally
restricted to admin access only. A recommended location to install software just for the user is in
%LocalAppData%/Programs (e.g. C:\Users\CalTech\AppData\Local\Programs\Morehouse
Calibration Software).

2.3 Windows Defender SmartScreen


Previous versions of the software were unsigned and would trigger Windows
Defender SmartScreen. This is no longer the case. The software is now signed
with an Extended Validation (EV) code signing certificate.

2.4. Necessary Components


Previous versions of the software required Microsoft SQL Server Compact 4.0 SP1
to be installed. This is no longer required. The software now utilizes a JSON
file to store the data and will migrate an existing database to the JSON
file. .NET Framework 4.8 is still required, but is included on modern Windows
PCs. If .NET Framework 4.8 is not installed, please contact the company's IT
department.

2.4.1. Microsoft Excel


The software requires that a version of Microsoft Excel is installed to use the Excel interoperability
functionality. Due to the widespread use of Excel, no installer is provided with the software. If a copy of Excel is
needed, please contact the company’s IT department.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 7 of 29 Rev: 0
2.5. Indicator Drivers
To communicate with an indicator, the devices need to be detected by the computer as a Serial Port in the Device
Manager. If a device is not visible under the Serial Port section or if the device isn’t being detected by the
software, the corresponding driver may need to be installed. The drivers included with the installer are USB to VCP
drivers and should allow the device to appear under the Serial Port section. Also see Missing Indicator §5.3,
Missing HADI Indicator §5.4, and Resetting COM Ports §5.5 for more troubleshooting steps.

All the device drivers are located in the Drivers directory. Each device model supported by the software has its
own directory listed. Within each directory is the USB to VCP driver for the specific model. If a needed model isn’t
listed, the device likely doesn’t include a USB connection port and offers an RS232 port instead.

If the device utilizes an RS232 port, a custom RS232 cable and 3rd party RS232 to USB adapter will be required to
establish communications. The chosen RS232 adapter will offer its own driver to enable communication with the
device. For more information, please contact the manufacturer or check the User’s Manual of the device.

3. Operating the Software


3.1. Launching the Software
By default, the software is installed at C:\Program Files (x86)\Morehouse Calibration
Software\Morehouse Calibration Software. Double-click the Morehouse Calibration Software file to
run the program. Alternatively, a shortcut to the software is available on the desktop and start menu.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 8 of 29 Rev: 0
3.2. Interacting with the Software
3.2.1. Start/Setup Tab

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 9 of 29 Rev: 0
1. The software starts with the Main Screen visible and the Start/Setup tab selected. This is the landing
screen of the software. To return to this screen at any time, select Start/Setup at the top of the screen.
2. Selecting Calibration will switch to the Calibration Tab §3.2.2 and allow a technician to read from
any discovered indicators and capture data.
3. Selecting Report will switch to the Report Tab §3.2.4 and will allow a technician to review the captured
data from each indicator and compare the percent error. Captured data can also be modified here.
4. Selecting Help will switch to the Help Tab §3.2.5 and will present contact information, relevant links,
and software information to the technician.
5. Clicking on the moon/sun in the upper right will toggle between light and dark modes §1.3. If the moon is
visible, it will toggle to dark mode and change to a sun. If the sun is visible, it will toggle to light mode and
switch to a moon. This toggle is always available on the Main Screen, regardless of which tab is being
shown.
6. Before being able to search for connected devices (7), the checkbox corresponding to the manufacturer
and model (6) of the meter must first be checked. Alternatively, the Search For All option can be checked
to discover all supported, connected devices. The software will not experience any issues if models are
selected that are not currently connected to the computer.
7. Once the model(s) are selected, clicking on Search Indicators will cause the software to begin searching for
connected devices. While the software is searching, a loading screen will be displayed. Once finished, the
Indicator Screen §3.2.6 will be displayed.
8. Toggling Ignore Calibration Units prevents units from being shown throughout the software. Additionally,
unit conversions are disabled.
9. Selecting Enter Calibration Coefficients will allow a technician to create, delete, or modify coefficients for
a load cell. This will open the Instrument Screen.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 10 of 29 Rev: 0
3.2.2. Calibration Tab

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 11 of 29 Rev: 0
1. Main Reading - Displays the sum of the most recent reading of each device that is included in the
summation.
2. Reading Units - The units (13) shown by the main reading.
3. Peak Reading - Shows the highest value the main reading (1) reached since the values were last reset
(8), captured (9), or a meter was tared (7).
4. Valley Reading - Shows the lowest value the main reading (1) reached since the values were last reset
(8), captured (9), or a meter was tared (7).
5. Enable Readings - Enables/disables actively reading from the activated indicators (18).
6. Settings Toggle - Shows/hides the Settings pane.
7. Tare All - Will attempt to tare all actively read devices simultaneously and reset values (8).
8. Reset Values - Reset the peak (3) and valley (4) readings to the current reading.
9. Capture Data - Resets values (8) and captures data. More information in the Capturing Data §3.2.3
section.
10. Mass Conversion - Enables/disables mass conversion calculations. Changes force units to mass units.
Enables gravity controls (11).
11. Local Gravity - If converting to mass (10), the local gravity needs to be entered for an accurate
conversion.
12. Reading Interval - Determines how fast readings should be updated for each active device.
13. Engineering Units - Shows either force or mass units. Changes the calculated value to account for the unit
conversion, if applicable.
14. Reading Resolution - The resolution to calculate readings to as a power of 10.
15. Close Settings - Closes the Settings pane.
16. Inactive Indicators - Inactive, connected devices are shown here. Clicking on one will move it to Active
Indicators (18).
17. Close Indicators - Closes the Indicators pane.
18. Active Indicators - Shows connected devices that will be read from when live readings (5) is enabled.
19. Indicators Toggle - Shows/hides the Indicators pane.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 12 of 29 Rev: 0
3.2.3. Capturing Data

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 13 of 29 Rev: 0
1. Contribution Percentage - Displays the percentage this instrument contributes to the summation.
2. Instrument Reading - Instrument’s calculated value or raw reading.
3. Displayed Value - Click to toggle.
◦ Load is the calculated value.
◦ mV/V is the raw reading from the instrument.
4. Add to Summation - Toggle to include this instrument in the total.
5. Indicator - Image of the indicator with its serial number underneath.
6. Instrument - Load cell image with the instrument’s serial number underneath.
7. Loading Mode - Click to toggle. Switches between the compression and tension coefficients.
8. Tare - Requests the indicator to tare and reset the peak and valley values.
9. Remove indicator - Click to remove. Moves the indicator to the Indicator pane and stops reading from the
indicator.
10. Reading Range - Calculates the difference between the largest and smallest reading. Visible with 2 or more
active indicators.
11. Capture Data - Captures the current raw reading of each active instrument as well as the coefficients and if
the instrument is included in the summation. The data is recorded into a data cell (13) and is recalculated
whenever any Setting is modified.
12. Target Cells - Enter the target value for each data point here. Used to calculate % error on the Report
Tab §3.2.4 or when data is exported (16).
13. Data Cells - Select the cell to capture data to it. When capturing data (11), the selected cell’s information
will be updated.
14. Captured Data - Contains indicator/instrument serial numbers and the raw reading with the calculated
value.
15. Clear Data - Clears the Target Cells and Data Cells.
16. Excel Export - Exports the captured data to Excel [§3.3.4].

See Configuring Active Indicator Reading Panes §3.3.2 for more information.

See Understanding the Data Grid §3.3.3 for more information about Target Cells (12),
Data Cells (13), and Captured Data (14).

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 14 of 29 Rev: 0
3.2.4. Report Tab

1. Load Cell Serial Number - Review captured data from the selected load cell.
2. Resolution - Performs the same as (14) in Calibration Tab §3.2.2.
3. Units - Performs the same as (13) in Calibration Tab §3.2.2.
4. Mass Conversion - Performs the same as (10) in Calibration Tab §3.2.2.
5. Local Gravity - Performs the same as (11) in Calibration Tab §3.2.2.
6. Coefficient Calculation - Shows the calculation used to turn a raw reading into the calibrated engineering
unit. Updates when selecting a cell in the table (11). Copy button on the left will copy an Excel friendly
version of the equation to the clipboard.
7. Unit & Resolution Calculation - Shows the calculation for resolution and unit conversion (including mass
if applicable). Updates when selecting a data or % error cell in the table. Copy button on the left will copy
and Excel friendly version of the equation to the clipboard.
8. Captured mV/V - Allows adjusting the raw reading if necessary. Updates when selecting a cell in the table
(11). Forces all calculations to update when changing the value.
9. Final Value - The final calculated value with the units displayed (if applicable). Useful for viewing how
modifying the raw reading affects the calculated value.
10. Coefficient List - Shows all coefficients for the selected load cell (Compression & Tension).
11. Data Grid - Shows the target and calculated values as well as the % error, using the reading resolution to
improve accuracy. Can show NaN (not a number) or infinity. % error cell is yellow if the error less than 1%.
An error over 1% turns red.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 15 of 29 Rev: 0
3.2.5. Help Tab

1. Email Software - Creates a new email draft and populates the To and Subject fields to submit feedback to
the software team.
2. Email Sales - Creates a new email draft and populates the To and Subject fields to contact the Morehouse
sales team.
3. Website Links - Opens the Morehouse website. The Software link will open directly to the software page
on the website.
4. Non-Digital Communications - Morehouse contact information for users that prefer non-digital
correspondence.
5. Version Number - Displays the current version of the program. Additionally, the program will attempt to
check for updates when the software is opened.
◦ Version x.x.x Available - An update is available that will be downloaded when clicked. The update
will still need to be installed manually.
◦ Up to Date - The latest version is currently being used.
◦ Update Check Failed - The software was unable to locate the latest version.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 16 of 29 Rev: 0
3.2.6. Indicator Screen

1. Indicator Pane - Shows connected/all indicators. Can be toggled (7).


2. Load Cell - The load cell’s coefficients to use. Warns if a load cell isn’t selected.
3. Serial Number - The meter’s serial number.
4. Serial Port Info - Various configurations to establish a connection with the device.
5. Communication Commands - The commands the software uses to communicate with the device.
6. Indicator Shell - Opens a communication terminal to the device to send/receive messages.
7. Connected Only - Toggles between hiding/showing disconnected meters. Disconnected meter models are
unknown so (4) and (5) are empty.
8. Proceed Button - Closes this window. If a connected device doesn’t have a selected load cell (2), a
prompt will appear before closing.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 17 of 29 Rev: 0
3.2.7. Shell Screen

1. Serial Number - Indicator’s serial number for easy reference.


2. Command History - Shows all responses from the device.
3. Command Input - Text entered here will be sent to the device once the Enter key has been pressed.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 18 of 29 Rev: 0
3.2.8. Instrument Screen

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 19 of 29 Rev: 0
1. Instruments - Collection of all instruments currently saved in the database by serial number. Selecting one
will update the other fields.
2. Add Instrument - Creates a new instrument. A new instrument can only be created if the previously
created instrument has been assigned a name.
3. Delete Instrument - Deletes the currently selected instrument.
4. Instrument Name - Allows the currently selected instrument’s name to be updated. An error will appear if
the name is not unique or is blank/contains only spaces.
5. Serial Number - Allows the currently selected instrument’s serial number to be updated. An error will
appear if the serial number is not unique, is blank, or contains only spaces.
6. Compression Units - Compression force units the instrument was calibrated for. For non-force instruments,
leave the units as LBF and be sure to enable Ignore Units (8) on the Start/Setup Tab §3.2.1.
7. Compression Coefficients - Coefficients for the compression mode found on the certificate. For non-force
instruments, this would be one mode, like clockwise, vacuum, etc.
8. Tension Units - Tension force units the instrument was calibrated for. For non-force instruments, leave the
units as LBF.
9. Tension Coefficients - Coefficients for the tension mode found on the certificate. For non-force
instruments, this would be the other mode (if applicable), like counter-clockwise, pressure, etc.
10. Import Certificates - Drag & Drop Morehouse calibration certificates to automatically import their
information. See Importing Certificates §3.3.1 for more information.
11. Proceed Button - Closes the window. If there are any errors on an instrument, the user will be prompted
before exiting.

3.3. Detailed Operation


3.3.1. Importing Certificates
Only digital calibration certificates generated by Morehouse for Single Run or
E74 calibrations are supported at this time. Other certificate formats must be
entered manually.

To import a calibration certificate, drag and drop the certificate file(s) (from File Explorer) onto the Instrument
Screen §3.2.8. If importing a single file, the calibration report number from the certificate is matched against
the currently selected instrument. The comparison made is <serialNumber><DateCode> and
<serialNumber>. In the event of a match, the current instrument’s data is updated with the information
contained in the certificate. If the comparison doesn’t match, other existing instruments are checked for a match. If
one is found, the user is asked to update the existing one or the currently selected one. When there is no match to
any existing instrument, the user will be asked to either update the currently selected instrument or create a new
one.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 20 of 29 Rev: 0
When importing multiple certificates at once, there are no prompts. A status bar is displayed to show the progress
of the import. During the import process, each certificate report number is compared against all existing
instruments in the software using the above-mentioned comparison method. If an existing instrument matches, its
information is updated with the information in the certificate. If no existing instrument matches, a new instrument
is created in the software with the information from the certificate.

When bulk importing, the process will continue silently until complete. Upon completion, the user will be notified
of any certificates that failed to import and the file’s name. If importing a single certificate, and it isn’t recognized,
the user is notified it doesn’t match a supported format.

3.3.1.1. Importing a Single Certificate Example


If the report number of the certificate is S-12345B2023, the software will look to see if an instrument with the
name or serial number of S-12345B2023 exists. If it doesn’t, it checks to see if an instrument with the name or
serial number of S-12345 exists. The software extracts the instrument’s serial number as S-12345 and the date
code as B2023 (2/20/2023).

• The currently selected instrument is S-11111 and there were no matches


◦ The user is asked to create S-12345B2023 or update S-11111
• The currently selected instrument is S-11111 and S-12345B2023 exists
◦ The user is asked to update S-12345B2023 or update S-11111
• The currently selected instrument is S-12345
◦ The instrument S-12345 is updated silently

3.3.1.2. Importing Multiple Certificates Example


During the import process, no prompts are provided. Like in Importing a Single Certificate Example
§3.3.1.1, a match with the report number or just the serial number is looked for. If there is a match to an
existing instrument, it is updated silently. Otherwise, a new instrument is created.

• Importing S-12345B2023 and S-12345 exists


◦ S-12345 is updated silently
• Importing S-12345B2023 and there were no matches
◦ S-12345B2023 is created silently

3.3.2. Configuring Active Indicator Reading Panes


Active indicators have several settings that can be changed independently of other indicators. Clicking on Load to
show the mV/V (raw) reading is beneficial for troubleshooting if the software is not operating as expected.
Otherwise, the Load mode is generally used.

Toggling between loading modes (Compression/Tension) switches between the two sets of calibration
coefficients.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 21 of 29 Rev: 0
Including an indicator in the summation of readings will add its force value to other indicators that are included in
the summation. The sum is shown on the Main Display (1) §3.2.2. Some use cases may require standard(s)
and unit(s) under test (UUT). To support this operating mode, the UUT(s) can be removed from the summation as
their readings are still captured for data export. Any combination of standard(s) and UUT(s) are supported.

Loading mode and summation inclusion settings must be configured before


capturing any data. If they were misconfigured during data capture, the data
must be recaptured.

3.3.3. Understanding the Data Grid


If a technician is targeting a predetermined set of values (for example 10% points), the target values (12) §3.2.3
can be entered before capturing any data. Alternatively, if target values are unknown ahead of time or not needed,
they can be left blank. Regardless, a data cell (13) §3.2.3 must be selected to record data. Once the data has
been captured, the cell below the current one is selected, adding a new row to the table if necessary. Focus is
returned to the Capture Data button for quick, repeated data capture.

When the Capture Data button is pressed while a target value (12) §3.2.3 is selected, the technician will be
asked if they want to capture data in the adjacent data cell. If the technician chooses no, nothing is done. A
technician can also overwrite existing data by selecting a data cell that already contains data and press Capture
Data again. Previous data will be lost.

When capturing data, only the raw reading, loading mode (compression/tension) coefficients, and summation
inclusion state are recorded. This allows existing data to be recalculated if the units or resolution of the software
are changed. It also allows a technician to modify captured data.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 22 of 29 Rev: 0
3.3.4. Exporting Captured Data

The software is capable of exporting the captured data either to a new or existing Excel workbook. After selecting
to either create a new workbook or open an existing one, the software will work through the workbook over a few
different phases. Additionally, several Excel named ranges will be used by the software. If a named range already
exists in the workbook, the contents of that range are updated (beneficial for template workbooks). If the range
doesn’t exist, it is created. The following ranges are used by the software:

NOTE: The <serial number> is sanitized for use as a named range. This replaces
every character that isn't a letter, number, or period with an underscore (_).

NOTE: The <serial number> is sanitized for use as a worksheet name. This
replaces the characters : \ / * [ ] with an underscore (_).

• RAW_<serial number> - Captured data from the instrument.


• CALC_<serial number> - Calculated values after polynomial, unit conversion, mass conversion, and
resolution.
• NEG_<serial number> - Negative (compression) coefficients.
• POS_<serial number> - Positive (tension) coefficients.
• LocalGravity - The local gravity in m/s^2. Only used if doing mass conversion.
• StandardGravity - The average gravity value in m/s^2. Only used if doing mass conversion.
• Resolution - The precision to round the values to and/or the number of decimals to show (if significant).
• TargetValues - The values the totals should add up to.
• <unit name> - First letter capitalized. Values are per pound.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 23 of 29 Rev: 0
Named ranges can be managed via the Name Manager in Excel. The Name Manager can be found under the
Formulas tab. Existing named ranges can have its location changed or be deleted. While a named range tutorial is
beyond the scope of this manual, the INDEX function can be used to access the contents of a named range. For
example, =INDEX(RAW_S_12345,2,1) will return the value in the second row of the first column of the
RAW_S_12345 named range. For an example, check the formulas in the calculated tables of an instrument or the
overview worksheet.

3.3.4.1. Phase 1: Creating the Overview Worksheet


If the spreadsheet MCS Calculated exists, its contents are cleared, otherwise the software will create that
spreadsheet. Unit conversion rates and the software’s configured resolution are written. If mass conversion was
used, the configured local gravity and average gravity values are written as well.

The named ranges LocalGravity and StandardGravity are used to store the local gravity and average gravity
values. The named range Resolution is used to store the software’s configured resolution. The named ranges
<unit name> are used to store the unit conversion rates. If these named ranges already exist, their contents are
updated. Otherwise, they are created in their respective locations on this sheet.

Additionally, the software will create the overview run tables. These tables are named Run<run number>Table
and are used to store the summation of the chosen captured data for a given run. The named range
TargetValues is used to store the target values. If the named range doesn’t already exist elsewhere in the
workbook, it is applied to the left most column of the first run table. The tables will include the aforementioned
Target column, as well as a total column and a column for each instrument.

3.3.4.2. Phase 2: Creating the Instrument Worksheets


NOTE: The <serial number> is sanitized for use as a named range and table names.
This replaces every character that isn't a letter, number, or period with an
underscore (_).

NOTE: The <serial number> is sanitized for use as a worksheet name. This
replaces the characters : \ / * [ ] with an underscore (_).

For each instrument’s serial number, the software will create a new worksheet named Data <serial number>,
following the above note. If the worksheet already exists, its contents are cleared first. The coefficients for the
instrument (both positive and negative coefficients) are written to the worksheet. The named ranges
NEG_<serial number> and POS_<serial number> are used to store the negative (compression) and positive
(tension) coefficients respectively. If these named ranges already exist, their contents are updated. Otherwise, they
are created in their respective locations on this sheet.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 24 of 29 Rev: 0
Additionally, the software will create the captured data table and the calculated data table. These tables are named
Captured<serial number>Table and Calculated<serial number>Table respectively. The named ranges
RAW_<serial number> and CALC_<serial number> are used to store the captured and calculated data
respectively. If these named ranges already exist, their contents are updated. Otherwise, they are created in their
respective locations on this sheet.

To provide transparency, the software will create Excel formulas to calculate


the values the same way it does. If there is no captured data for a specified
point, the calculated value will be blank.

3.3.4.3. Phase 3: Filling the Run Overview Tables


Back on the MCS Calculated spreadsheet, the run data tables are filled in with references to the captured and
calculated data named ranges for each instrument. If a reading was omitted from summation, the value will be
turned gray to help distinguish from the included values. The total column is also filled in with the sum of the
included values.

4. Quick Start Guide


4.1. Checklist
1. Verify the instrument’s coefficients are set and correct.
2. Verify indicators are connected to the computer and not in use in another program.
3. Search for indicators to read from.
4. On each indicator, select the connected instrument from the drop-down.
5. Navigate to the Calibration tab and activate indicators to read from.
6. Configure indicator settings.
◦ Tension/Compression
◦ Summation
7. Configure reading settings.
◦ Reading Interval
◦ Resolution
◦ Units
◦ Mass/Gravity
8. If necessary, pre-enter target values in the first column of the data grid.
9. Enable live readings.
10. Choose the cell in the data grid where desired to capture the first reading.
11. Press the capture data button to capture a reading.
12. Review captured data, if desired.
13. Repeat steps 10 - 12 until enough data has been collected.
14. Export to Excel.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 25 of 29 Rev: 0
4.2. Detailed Guide
Opening the Instrument Screen §3.2.8 allows a technician to view all configured instruments. Selecting any
of them will allow both sets of coefficients to be displayed. If necessary, modify the coefficients to match the
current ones for the instrument. Once all instruments are properly configured, exit the Instrument Screen
§3.2.8. If necessary, Ignore Calibration Units can be enabled here to hide the units throughout the software.
See Removing Force Units §5.1 for more information.

To discover indicators, ensure all devices are connected to the computer, their drivers have already been installed,
and none of the devices are currently in use by another program or the system. If necessary, verify the indicator is
plugged in and turned on. Choose the indicator models on the Start/Setup Tab §3.2.1 and press Search
Indicators. Once indicators have been found, verify that each indicator on the Indicator Screen §3.2.6 is
paired with the correct instrument and close the screen. If 1 or more connected indicators isn’t found, refer to
Missing Indicator §5.4 for more information.

Switch to the Calibration Tab §3.2.2 to obtain readings from the newly discovered indicators. Select
whichever indicators to use from the inactive indicators’ pane. If all inactive indicators are activated, the pane will
automatically close, otherwise, the red X will close the pane manually. If the pane needs to be reopened, press the
green plus at the bottom right of the active indicators’ pane. Each active indicator can be configured
independently for tension/compression and to include the indicator’s value into the summation shown on the
Main Display. Additionally, the reading can be changed from the calculated value (Load) to show the raw reading
(mV/V).

Reading settings can also be modified. The system’s resolution can be configured in powers of 10, rounding
accordingly. The resolution can be changed at any time, even during live readings or after data has been captured.
Engineering units, mass, and gravity can also be set now, during live readings, or after data has been captured.
When enabling mass readings, the engineering units options will change from force units to mass units. All of
these settings will cause the readings to be recalculated. Additionally, the Reading Interval at which a new
reading is displayed for the software can be configured. A longer interval could allow the readings to be more
stable.

The Settings Pane can be closed with the red X or by pressing the gear icon next to the Live Reading toggle. To
open the settings again, simply press the gear icon next to the live reading toggle. If target values are known prior
to starting the calibration, they should be entered into the first column of the Data Logging data grid. Just select
the first data cell and type the target value in. Pressing Enter after the value has been entered, the next cell down
will be selected, making it easy to enter all the values quickly. If necessary, a new row will be automatically added
once the last row is reached.

Before Capturing Data §3.2.3, the Live Readings toggle must be enabled. This will begin communication
with each of the active indicators, causing their displays to update within their corresponding panels. This update
always occurs after the reading interval has elapsed. If necessary at this time, an indicator can be tared using the
button on its panel or all indicators can be tared simultaneously via the Tare All button. Taring can also be
performed at any time the software is actively reading from 1 or more indicators.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 26 of 29 Rev: 0
Choose a cell in the Data Logging data grid. The chosen cell will have data entered into it once the Capture Data
button is pressed. After the data has been entered into the chosen cell, the next cell down will automatically be
selected, creating a new row if that cell was in the last row. Repeat this step as many times as necessary until all
desired data has been captured. If additional runs are required, select a cell in the next run’s column and repeat
this entire step as necessary.

While Capturing Data §3.2.3, the Report Tab §3.2.4 can be chosen at any point to review collected data.
The tab provides the calculated % error for the reading from the target value. For convenience, coefficients are
shown for the selected instrument. Readings and settings can also be modified on this screen as necessary.

If no more data is to be collected, live readings can be disabled without losing data. Optionally, an active indicator
can be removed by pressing the red X in the top right of an active indicators’ panel. The data can also be exported
to an Excel spreadsheet §3.3.4.

5. FAQ - Frequently Asked Questions


5.1. Removing Force Units
Question: Our company does more than just force and mass calibration/measurements. Is there a way to remove
the unit label to help remove confusion for our technicians?

Answer: Yes. Enabling the Ignore Calibration Units (8) checkbox on the Start/Setup Tab §3.2.1, all
unit labels are removed from the program. Due to a technical limitation though, the Instrument Screen
§3.2.8 still requires a unit to be selected.

Please note that when this feature is enabled, the unit of Force/Mass dropdown is disabled and
simply displays “As Calibrated”. Additionally, the Mass toggle and Gravity text box are also
disabled. Finally, any points captured while this feature is enabled will not record any unit of
measurement data.

5.2. Force Summation


Question: Can I use multiple indicators and load cells to calculate force?

Answer: Yes. Force summation is supported.

1. Ensure all load cell coefficients and indicators are set up. See Instrument Screen §3.2.8 and
Indicator Screen §3.2.6 for instructions.
2. Select the indicators to be used in the calculation. See Capturing Data §3.2.3 for instructions.
◦ Each selected indicator/instrument will have its own region for control and display purposes
3. An indicator can be ignored temporarily by unchecking the Add to Force Sum checkbox for the indicator.
4. Set the Force/Mass toggle.
5. Capture data and export as before.
6. The exported data will contain all devices.

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 27 of 29 Rev: 0
5.3. Missing Indicator
Question: Why isn’t my indicator showing up?

Answer:

1. If using a Morehouse 4215 indicator, ensure that the “Add Line Feeds” option is enabled via the front panel
controls.
2. The indicator may not be properly connected. Please ensure the connection to the computer is good.
3. The indicator’s driver may not have been installed. Please refer to Indicator Drivers §2.5 in this
document for further instructions.
4. The indicator may be in use by another program. Please stop any other programs that may be using the
indicator or restart the PC.
5. As a last attempt, please see Resetting COM Ports §5.5.

5.4. Missing HADI Indicator


Question: Why isn’t my HADI indicator being detected?

Answer:

1. Complete the steps in the Missing Indicator §5.3 first


2. Open the computer’s Device Manager via Control Panel or right-clicking on the start menu and selecting
Device Manager
3. Find the Universal Serial Bus controllers device category and expand to view all devices
4. Locate the USB Serial Converter device and right click on it and select Properties.
5. Navigate to the Advanced tab and enable Load VCP

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 28 of 29 Rev: 0
5.5. Resetting COM Ports
Question: I am having unexplained indicator issues, how do I reset my COM ports?

Answer: Follow the instructions below, but be aware of the warning.

WARNING: These steps will also remove any other COM devices. Use with caution!

1. Open the computer’s Device Manager via Control Panel or right-clicking on the start menu and selecting
Device Manager
2. Find the Ports (COM & LPT) device category and expand to view all devices
◦ USB Serial Port (HADI), USB COM Port (4215), or USB DSC Port (DSC) may be listed
3. Select each listed device from the Ports (COM & LPT) category and right click and select Uninstall Device
4. Disconnect the indicator’s USB cable from the PC and then reconnect it. If it is not automatically detected,
click on the Scan for hardware changes button at the top of the device manager

Document #: PM-4114 Copyright © 2023 v1.2.6

Page 29 of 29 Rev: 0