This document provides specifications for the ML801B multi-channel measurement card including that it can connect up to 8 strain gauge transducers in various bridge configurations, has internal completion resistors from 120-1000 ohms, bridge excitation voltages from 0.5-5V, measuring ranges from ±8-±80 mV/V depending on voltage, and noise levels below ±4 μV/VSS with filtering. It also lists technical details such as non-linearity of 0.05%, measurement frequency range of 1000Hz, operating temperature range of -20-60°C, and allowed cable length between the transducer and connection board of up to 100 meters.
This document provides specifications for the ML801B multi-channel measurement card including that it can connect up to 8 strain gauge transducers in various bridge configurations, has internal completion resistors from 120-1000 ohms, bridge excitation voltages from 0.5-5V, measuring ranges from ±8-±80 mV/V depending on voltage, and noise levels below ±4 μV/VSS with filtering. It also lists technical details such as non-linearity of 0.05%, measurement frequency range of 1000Hz, operating temperature range of -20-60°C, and allowed cable length between the transducer and connection board of up to 100 meters.
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1
Multi-channel measurement card ML801B (continued)
ML801B + connection board AP815i
Accuracy class 0.11)2)3) Transducers that can be connected 8 SG full bridges in 6-wire configuration or 8 SG half bridges in 6-wire configuration or 8 SG half bridges in 5-wire configuration or 8 SG quarter bridges in 4-wire configuration or 2 SG rosettes Internal completion resistors Ω 120, 350, 700, (optionally 1000) Total transducer resistance with half and full bridges Ω 240 … 4000 (2x120 ... 2000 with half bridges; 4x240 ... 4000 with full bridges) Bridge excitation voltage (DC) V 5; 2.5; 1; 0.5 Measuring ranges mV/V ±8 (UB=5 V) ±16 (UB=2.5 V) ±40 (UB=1 V) ±80 (UB=0.5 V) Control signal (shunt) mV/V 1.0078 ±0.1 % (at 350 Ω) Noise at 350 Ω Filter characteristics Bessel/Butterworth Filter frequency Hz 1.25/5 40/80 200/500 UB=5 V μV/VSS <±0.1 <±0.6 <±4 UB=2.5 V μV/VSS <±0.2 <±1.2 <±8 UB=1 V μV/VSS <±0.5 <±3 <±20 UB=0.5 V μV/VSS <±1 <±6 <±40 Transducer connection Two D jacks, 25-pin (4 channels each) Allowed cable length between transducer and connection board m 2004) Width mm 20.3 (4 HP) Non-linearity % 0.05 Measurement frequency range Hz 1000 (-1 dB) Effect of 10 K change in ambient temperature on the zero point (relative to the full scale value) % 0.12) on sensitivity % 0.1 Operating temperature range °C -20 … +60 1) 0.2 with irradiation as per EN 61000‐4‐3:1996 + A1:1998 2) 0.2 with 5 V bridge excitation voltage 3) If zero balancing is not possible, the following accuracy classes apply: 0.2 with R transducer > 2 kΩ and 0.3 with Rtransducer > 3 kΩ. 4) 100 m max. distance between connection board and T-ID/TEDS module