Logiq e
Logiq e
Logiq e
LOGIQ® e
This is huge. General Imaging Suite.
Product Description
The LOGIQ e is a high performance multipurpose color compact ultrasound system, which includes many new
imaging technologies to help improve your diagnostic confidence. B-Steer Plus combines two GE technologies, B-
Steer and CrossXBeam™, so you can see both anatomy and the needle in one image with amazing clarity. Virtual
apex on phased array probes provides a wider field of view. The LOGIQ e is designed for cardiac, abdominal,
obstetrics, gynecology, vascular, musculoskeletal, small parts, pediatric, neonatal, nerve block (assisting needle
guidance in the administration of peripheral nerve block), urology and intraoperative applications.
TruScanTM Architecture
GE’s exclusive, software-intensive ultrasound
imaging platform gives you computational power,
image-manipulation capability, workflow flexibility
and product upgradeability.
Page 1 of 12
GE Healthcare
System Overview • ACE™(Adaptive Color Enhancement)
• TruAccess, Raw Data Processing
• Patient Information Database
Applications
• Image Archive 55.8GB on Hard Drive
General Specification • Abdominal
• Full M&A Calculation Package with
• Cardiology
Real Time Auto Doppler Calculations
• Obstetrical
Dimensions and Weight • Vascular Calcs
• Gynecological
• Height: 61 mm (2.40 in) console only • Cardiac Calcs
• Musculoskeletal
100 mm (3.94 in) with handle • OB Calcs and Tables
• Vascular
• Width: 340 mm (13.38 in) • Fetal Trending
• Urological
• Depth: 287 mm (11.30 in) console only • Multi Gestational Calcs
• Small Parts
337 mm (13.27 in) with handle • Hip Dysplasia Calcs
• Pediatric and Neonatal
• Weight with Battery: approx. 4.6 kg • Gynecological Calcs
• Intraoperative
(10.1 lb.) • Urological Calcs
• Nerve block
Electrical Power • Renal Calcs
Scanning Methods
• Voltage: 100- 240 V AC • Report Package
• Phased Array Sector
• Frequency: 50/60 Hz
• Electronic Convex
• Power: Max. 130 VA with Peripherals
• Electronic Linear Software Options
Console Design
Transducer Types
• Laptop Style • Easy 3D
• Convex Array
• Integrated HDD (80GB) • DICOM 3.0 Connectivity
• Microconvex Array
• Lithium-Ion Battery Pack (standard) • LOGIQ View
• Linear Array
• Wired, wireless LAN Support • B Steer+
• Phase Array
- For DICOM communication • CrossXBeam
Operating Modes
- For Network Storage • CMM and AMM
• B-Mode
(Image store to PC without DICOM • Touch Mode
• M-Mode
system)
• Anatomical M-mode Hardware Options
• USB ECG (AHA / IEC) (Optional) Support
• Color Flow Mode (CFM) • Battery Pack
• CWD (Optional) Support
• Power Doppler Imaging (PDI) • CWD
• 1 probe port with micro-connector
• Continuous Wave Doppler (optional) • USB ECG (AHA / IEC)
• Rear handle
• Pulse Wave Doppler (PWD)
• Harmonic Imaging, 2 method Filtered,
User Interface Phase inversion
Media & Peripherals
• 3 pedal Foot Switch (IPX8)
• CrossXBeam ™ • 1 pedal Foot Switch (IPX1)
Operator Keyboard • Virtual convex, Virtual apex • Isolation Cart
• Alphanumeric Keyboard • Color M mode(CMM) • Docking Cart
• Ergonomic Hard Key Operations • Range Focus Adjustable platform height
• Integrated Recording Keys for Remote • LOGIQ View Digital / Analog video output
Control of Peripheral Devices and Standard Features -3 Probe port adapter option
DICOM Devices • High Resolution 15 inch Color LCD -External speaker option
• 6 TGC Pods, with Re-mapping • Over 1000 frames or over 60sec CINE -17" Extra LCD option
functionality at any depth Memory (64MB) depend on FOV, - 15" Extra touch screen option
• Backlight keys Scanning Lines etc. -Panasonic™ DVR option support
Display Screen • 80GB Hard Drive -Total height with LCD:
• 15 inch High-Resolution Color LCD • External DVD R/W storage 1385mm±3mm
- Display size: 1024x768 • Loops storage-from ‘on the fly’ Depth: 617mm±2mm
• Interactive Dynamic Software Menu scanning Width: 470mm±1.5mm
• Open Angle Adjustable and from memory Weight : 59kg±1.0kg
- 0 to 160° • Automatic Optimization - Total height with Touch screen:
• Integrated Speakers - Auto Tissue Optimization: ATO 1740mm ±3mm
• Brightness Adjustment - Auto CFM Optimization: ACO Depth : 625mm±2mm
• Audio Volume Adjustment - Auto Spectrum Optimization: ASO Width: 473mm±1.5mm
- Auto TGC Weight: 70kg±1.0kg
- Auto Clarity Suite (Auto Focus and • External USB DVD-RW (standard)
Auto Frequency)
Page 2 of 12
• USB thermal B&W printer, Sony™ • Date: 2 types selectable Depth/Width, Depth, 2
UPD-897 (option) MM/DD/YY, DD/MM/YY • M Scale Markers: 2 types
• USB thermal color printer, Sony UPD- • Time: 2 types selectable Time/Depth, Time
24 hours, 12 hours • Image Management Menu: Menu,
23 MD (option)
• Operator Identification Delete, and Image Manager
• USB DeskJet™ color printer, HP470/HP
• Patient Name: First, Last & Middle • Image Palette
Pro K550 (option) • Caps Lock: On/Off
• Patient Identification: 64 characters
• NETGEAR™ USB Wireless Adapter • Gestational Age from • System Messages Display
WG111 V3 supporting the 802.11a/b/g LMP/EDD/GA/BBT • Trackball Functionality Status: Scroll,
formats, where available (option) • Power Output Readout M&A (Measurement and Analysis),
• Memory Stick (option) - MI: Mechanical Index Position, Size, Scan Area Width and Tilt
• USB2.0 Hub (option) - TIS: Thermal Index Soft Tissue • Battery status
• USB HDD - TIC: Thermal Index Cranial (Bone) • Biopsy Guide Line and Zone
• Measurement transfer Kit • Heart Rate
- TIB: Thermal Index Bone
• Barcode scanner(1D, 2D), HPP4600g • Primary Parameter Menu (depend on
• System Status (real-time or frozen)
current mode)
• Barcode scanner(1D), HHP3800g • Probe Orientation Marker: Coincides
with a probe orientation marking on - B Mode
• 120W Universal Power Supply (UPS)
the probe. Frequency
Display Modes
• Image Preview Gray Map
• Simultaneous Capability
• Gray/Color Bar Dynamic Range
- B/PW
• Cine Gauge Image Rotate
- B/CFM or PDI
• Measurement Summary Window Focus Position
- B/M or AMM
• Measurement Results Window: pre- Colorize
- Dual B (B/B) settable display location Edge Enhance
- Dual B + CFM or PDI • Probe Type Focus Number
- Real-time Triplex Mode • Application Name
Virtual Convex / Virtual Apex
• Selectable Alternating Modes • Imaging Parameters by Mode (current
- Color Flow Mode
- B/M mode)
Frequency
- B/PW - B/M-Mode
Frequency Spatial Filter
- B/CW
Gain Angle Steer
- B + CFM (PDI)/M (optional)
Edge Enhance/Frame Averaging Packet Size
- B + CFM (PDI)/PW
Gray Map PRF
- B + CFM (PDI)/CW Image Depth Color Map
- 3D – Mode (option) Dynamic Range Threshold
• Multi Image Split Screen Frame Rate
Color Invert
- Live and/or frozen % of Power Output
Wall Filter
- B + B/CFM or PDI - Color Flow Mode
Color Flow Frequency - M Mode
- Independent Cine playback
Color Gain Gray Map
• Zoom: Read/Pan and from archive
Line Density/Frame Average Dynamic Range
• Colorized Image
% of Power Output Sweep Speed
- Colorized B PRF Display Format
- Colorized M Wall Filter Colorize
- Colorized PW Spatial Filter/Packet Size
Edge Enhance
- Colorized CW - PW-Mode
Full Timeline
• Time line Display Doppler Frequency
Doppler Gain Power Output
- Independent Dual B/PW/CW
% of Power Output Anatomical M
Display
PRF Rejection
- Display Formats:
Wall Filter - PW Mode
Top/ Bottom or Side/ Side selectable Sample Volume Width Frequency
Format Size: Vert1/3 B; Vert1/2 B; Dynamic Range Baseline
Vert2/3 B; Horiz 1/2 B; Horiz1/4 B; TL - CW-Mode Quick Angle
Only Doppler Frequency
Angle Steer
format, switchable after freeze Doppler Gain
% of Power Output Sweep Speed
- Update mode: timed based on
PRF PRF
sweep
Wall Filter SV Length
• Quad Screen Display access from split
Dynamic Range Angle Correct
Screen
• Focal Zone Markers Spectral Invert
Display Annotation • Body Pattern: 219 types
• Institution/Hospital Name Wall Filter
• B Scale Markers: 2 types
Page 3 of 12
- Cine Mode Colorize • B/M-Mode
Loop Speed Spectral Average - Gain
Cycle select Power Output - CrossXBeam
Start Frame Cycles, to Change - PIH
End Frame - TGC
Frame by Frame - Image Reverse
Run/Stop System Parameters - Depth
Select All - Scan Area
Cine Mode select - Auto Optimize (ATO)
First System Setup - Dynamic Range
Last • Diagnostic Categories: 6 types, pre- - Focus Number
settable
Secondary Parameters Menu - Focus Position
OB, GYN, Cardiac, Urology, Smallparts,
- B Mode Pediatric - Line Density
Frame Average • User Programmable Preset Capability - Frequency
Biopsy • Factory Default Preset Data - Frame Average
Line Density • Languages setup: - Edge Enhance
Focus Width English, Norwegian, French, German, - Focus Width
B Softener Spanish, Italian, Portuguese, Russian, - M/D Cursor
Suppression Greek, Finnish, Swedish, Dutch, - Sweep Speed for M-Mode
Danish , Japanese • PW-Mode
Power Output
• Languages for Manuals:
Bsteer+ - Gain
English, Norwegian, Polish, French,
Range Focus German, Spanish, Italian, Portuguese, - Sample Volume Gate Position,
- CF Mode Russian, Greek, Finnish, Swedish, Length
Baseline Dutch, Danish , Japanese - PRF
Line Density • Operation Error Beep - Doppler Frequency
Transparency Map • Body Surface Area: 2 types - Dynamic Range
Focus Position Oriental, Occidental - Auto Optimize (ASO)
Flash Suppression • OB Report Format 5 types - Audio Volume
Tokyo Univ., Osaka Univ., USA, • CW-Mode
Power Output
Europe,ASUM
Frame Average - Gain
• EFBW: 8 types
- PW Mode Tokyo Univ., Osaka Univ., USA and - Velocity
Rejection Europe (Shephard, Merz, - Doppler Frequency
Dynamic Range Hadlock/Shephard, Williams, Brenner) - Dynamic Range
Display Format • CUA/AUA for Hadlock - Auto Optimize (ASO)
Full Timeline • Body Pattern Copy to Active Side: - Audio Volume
Trace Direction On/Off • Color Flow Mode
Auto Calculations • Colorized B/M: 5 types for each. - Gain
PWD/CWD: 6 types for each
Modify Calcs - ROI Position, Size
• Programmable Annotation Library:
PW/CF Ratio - PRF
44 annotations
Triplex (Duplex) on/off • Customized Common Home Position - Color Line Density
Colorize • Menu Selection at New Patient: 2 - Color Frequency
Gray Map types - Packet Size
Cycles, to Average Patient Entry, Schedule - Threshold
Trace Method • Sort Criteria for Schedule List: 2 types - Frame Average
Trace Sensitivity Date&Time, Name - Focus Position
Time Resolution • Patient Name Format: 2 types - Auto Optimize (ACO)
Full Name, Last&First • 3D Acquisition (option)
Spectral Average
• Pre-settable Doppler Audio Volume
Power Output - Scan Distance
• Measurement Clear Operation: 2 types
-CW-Mode - Scan Plane
Meas.-only, with-Comment
Display Format • Display Unit Age: 3 types - Acquisition Mode
Full Timeline "Year,month" "Week.day", "No display" Post-Processing
Trace Direction • System Boot Up: 25 seconds • TruAccess: GE-exclusive, raw-data
Trace Method • Full maintenance Reboot: 120 sec digital processing
Trace Sensitivity • Probe Change: 3-5 sec • Read Zoom up to 8x
Auto Calculations Pre-Processing • B/M-Mode
Modify Calcs • Acoustic Power Output - Gain
Time Resolution • Read Zoom up to 8x - Image Reverse
Page 4 of 12
- Auto Optimize (ATO) - Texture • Scrolling Timeline Memory
- Image Rotation - Gray Surface Image Archive/Connectivity
- Gray Map - Scalpel • Clipboard: displays thumbnail images
- Colorize - Auto Movie of the acquired data for the current
- Rejection - Undo exam
- B Softener - Reset • Previewing Clipboard Images: An
- Sweep Speed for M-Mode enlarged preview of the image
- Auto TGC • Recalling Images from the Clipboard
• PW-Mode Imaging Processing and • Image Browser: Archived images from
- Gain past patient exams appear as well as
- Baseline
Presentation images stored for the current exam
- Angle Correct - Previewing an Image
- Quick Angle TrueScan : software Intensive - Grouping a Set of Images
- Doppler Invert Ultrasound Imaging Platform - Analyzing Images
- Display Format • Digital Beamformer • Image Management
- Sweep Speed • Beamformer Operating Frequency - Select All/Unselect All
- Full Timeline Range: 1.7– 18 MHz - Permanent Store
- Rejection • 1024 Digital Processing Channel - Delete all the Temporary Images
- Colorize Technology - Delete Selected Image
- Compression (Dynamic Range) • Displayed Imaging Depth: Minimum - Analyze
- Auto Calcs Depth of Field: 2 cm (Zoom and probe • Ethernet Network Connection
- Trace Direction dependent); Maximum Depth of Field: • Configurable 3 Print (Recording) Keys
- Modify Calcs 30 cm (probe dependent) (P1-P3) to Multiple Output
- Number of Average Cycles • Transmission Focus Devices/Workflows
- Trace Method - 1 – 8 Focus Points Selectable (probe • Archiving Format:
- Trace Sensitivity and application dependent) - DICOM with ultrasound raw data
- Auto Optimize (ASO) - Focal Zone Position - DICOM
• CW-Mode • Continuous Dynamic Receive • Capture Area: pre-settable for each
- Gain Focus/Aperture print key
- Baseline • Multi-Frequency/Wideband - Video Area
- Angle Correct Technology - Application Window
- Quick Angle • 256 Shades of Gray (VGA) - Whole Screen
- Doppler Invert • 174 dB System Internal Dynamic • Archiving Image Frames: / pre-
- Display Format Range settable for each print key
- Sweep Speed • Adjustable Field of View (FOV) - Single: stores single frame only
- Full Timeline • Image Reverse: Right/Left - Multiple: stores cineloop
- Rejection • Image Rotation: 2 steps - Secondary Capture: screen shot
- Colorize Rotation: 0°, 180° • Image Compression/Picture Quality:
- Compression (Dynamic Range) CINE Memory/Image Memory pre-settable for each print key
- Auto Calcs • Over 1000 frames or over 60sec CINE - Quality: 1% to 100%
- Trace Direction Memory (64MB) depend on FOV, • Dataflow: a set of pre-configured
- Modify Calcs Scanning Lines etc. services
- Number of Average Cycles • CINE Gauge and CINE Image Number - When you select a dataflow, the
- Trace Method Display Ultrasound system automatically
- Trace Sensitivity • CINE Review: works according to the services
- Auto Optimize (ASO) Frame-by-frame, Loop associated with the dataflow
• Color Flow Mode • CINE Review Speed: 9 types • Configurable Examination List
- Gain 48%;31%;25%;22%;17%;14%;13%;11 Window, Patient Information Window,
- Baseline %; and Search/Create Patient Window
- Color Invert • Selectable CINE Sequence for CINE - Free text addresses, birth date,
- Color Map Review extended patient dialog in Pts Info
- Threshold • Start and End Frame Selections for window
- Frame Average (in loop images) Loop Playback - Extended search dialog, auto
- Flash Suppression • Separation Maker to Indicate Time search for patient in Search/Create
• Easy 3D (option) Discontinuity Pts window
- Threshold (Opacification) • Measurements, Calculations and - Pre-defined text directly in Exam
- Render Annotations on CINE Playback List window
Page 5 of 12
- Automatic generation of patient ID • Focus Depth: steps dependent on • B/M Acoustic Output: 0 – 100 %, 2 %
- Request acknowledge of End Exam probes step
action • Rejection: 6 steps • Rejection: 6 steps
- Go directly screen from search • Frequency: 3-5steps, probe dependent • Dynamic Range: 30 – 120 dB, 3 dB
- Detect unfinished examination Color Flow Mode step
• Tools • Base Line • Edge Enhance: 6 steps
- Verify DICOM directory on • Invert: On/Off • Gray Scale Map: 40 types
removable media • CF/PDI Focus Depth: 11 steps default • M Gain: 0 – 98 dB, 2 dB step
- Format removable media pre-settable PW/CW-Mode
(rewritable DVD) • CF/PDI Acoustic Output: 0 – 100%, • Maximum and Minimum Velocity
• Views: shows you an overview of the 10% step Scales
Ultrasound system’s connectivity • Packet Size: 6-18, probe dependent. PW
Max: 870cm/s, 19800Hz
architecture • Line Density: 5~8 steps (probe
Min Velocity Scale 15cm/s, 700Hz
- The currently selected dataflow dependent) CW
- All configured data flows • Frame Average: 7 steps Max: 1460cm/s, 40000Hz
- The network structure tree • PRF: 0.3K-11.4K Hz (Probe dependent) Min Velocity Scale 40cm/s, 2100Hz
- The configured buttons data flows • Spatial Filter: 6 steps
• AVI and JPEG Export • Gain: 0 – 40 dB, 0.5 dB step • Gray Scale Map:7 types
DICOM Support (option) • Wall Filter: 2~ 6 steps, Application and • Dynamic Range: 24 - 60, 4 dB step,
-Verify probe dependent depend on Application.
• Base Line: 5- 95 %
-Print • Angle/Width (deg, mm): probe
• SV Gate: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 12,
-Store dependent
14,16mm, depend on Application.
-Modality Worklist • CF/PDI Vertical Size (mm): default pre-
• Angle Correct: +/- 90°, 1° step
-Multiframe settable
• Spectral Color: 6 types
-Storage Commitment • CF/PDI Center Depth (mm): default
• PW Sweep Speed: 8 steps
-Modality Performed Procedure Step pre-settable • Invert: On/Off
(MPPS) • CF/PDI Frequency: 2-4steps, depend • M/PW Display Format: V-1/3B, V-1/2B,
-Media Exchange on probe. V-2/3B, H-1/2B, H-1/4B, TLOnly
-Off network/mobile storage queue • CF/PDI Focal Number: 1 • PW Acoustic Output: 0 - 100 %, 10 %
• Color Map: 13 types at most, depend step
on probe and application • Spectral Averaging: 5 steps pre-
settable
Scanning Parameters • Color Threshold: 10 – 100 %, 5 % step
• Time Resolution: 4 steps
PDI-Mode
• PW/CF Ratio: 1, 2, 4
B-Mode • PDI Map: 13 types
• Rejection: 15 steps
• CF/PDI Acoustic Output: 0 – 100%,
• B/M Acoustic Output: 0 – 100%, 2% • Gain: 0 - 32 dB, 1 dB step
10% step
step • Wall Filter: 5 - 1500 Hz, 22 steps,
• Packet Size: 6-18,depend on probe. depend on probe/application
• Image Reverse: On/Off
• PW Angle Steer: 0, +/- 10, 15, 20°
• B Colorize: 5 types
• Spatial Filter: 6 steps • PRF: 700 – 19800 Hz with PW,
• Thermal Index: TIC, TIS, TIB
• Frame Average: 7 steps (Probe 2100Hz-40000 Hz with CW
• Softener: 4 steps • Sample Volume Depth: 29 steps
dependent)
• Focus Number: 8 steps default pre-settable
• PRF: 0.3K-11.4K Hz (Depth dependent)
• Line Density: 5~8 steps (Probe • Audio Volume
• Power Threshold: 10 – 100 %, 5 % step
dependent) • PW Frequency 2-4steps, depend on probe
• CF/PDI Vertical Size: default pre-
• Frame Average: 6 steps
settable
• Edge Enhance: 6 steps LOGIQ view
• CF/PDI Center Depth: default pre-
• Angle (deg): probe dependent, 10° – Available on all probes
settable
133°
• CF/PDI Focal Number: 1
• Gray Scale Map: 40 types Virtual Convex
• Gain: 0 – 40 dB, 0.5 dB step
• Gain: 0 – 98 dB, 2 dB step Available on the all Linear probes
• Wall Filter: 6 steps, depend on probe
• Dynamic Range: 30 – 150 dB, 3 dB
• CF/PDI Frequency: 2-4 steps depend
step, depend on probe
on probe
• Harmonic start: on/off
M-Mode
• Virtual Convex: on/off Measurements /
• Sweep Speed: 8 steps
• Depth: 2 – 30 cm, 1 cm step
• M Color: 5 types Calculations
0.5cm step for Linear probe when
• M/PW Display Format: V-1/3B, V-1/2B,
less than 5cm depth.
V-2/3B, H-1/2B, H-1/4B, TL Only
Page 6 of 12
General - Adb.Aorta: Prox Ao, Mid Aorta, FL/HC
Measurements/Calculations Distal Ao, Prox Ao Trans, Mid Ao CI (Cephalic Index)
Mode Measurement Trans, Distal Ao Trans, AAA Long, AAA OB Graphs
• B-Mode Trans • Fetal Growth Curve Graphs
- Distance - Gallbladder: CBD, Gallbladder Wall, - Normal growth curve, positive and
- Circumference/Area (Ellipse/Trace) Thickness negative standard deviations or
• M-Mode • Max PG applicable percentiles, and
- Tissue Depth (Distance) • Mean PG ultrasound age of the fetus
- Time Interval • Testicle Length, width, and Height - One measurement per graph
- Depth Difference with Time Interval Obstetrics - Single or Quad views
and Slope Measurements/Calculations • Fetal Growth Bar Graph
• Doppler Mode • Open trance - Ultrasound age and gestational
- Velocity • Abdominal Circumference (AC) age
- TAMAX, TAMIN, and TAMEAN • Amniotic Fluid Index (AFI) [Moore] - Plots all measurements on one
(Manual/Auto Trace) • Antero-PosteroTrunk Diameter and graph
- Two Velocities with Slope and Time Transverse Trunk Diameter (APTD- Gynecology
Interval TTD) Measurements/Calculations
- Time Interval • Antero-PosteroTrunk Diameter by • Ovary Length, Width, and Height
Transverse Trunk Diameter (AxT) • Uterus Length, Width, and Height
Generic Measurement • Biparietal Diameter (BPD) • Ovarian Follicle Measurements
• B-Mode • Crown Rump Length (CRL) - 1 distance
- % Stenosis • Cardio-Thoracic Area Ratio (CTAR) - 2 distances
- Volume • Estimated Fetal Weight (EFW) - 3 distances
- Angle • Femur Length (FL) • Endometrium thickness (Endo)
- A/B Ratio • Foot Length (Ft) Cardiac
• M-Mode • Gestational Sac (GS) Measurements/Calculations
- % Stenosis • Head Circumference (HC) B-Mode Measurements
- A/B Ratio • Humerus Length (HL) • Aorta
- Heart Rate • Length of Vertebra (LV) - Aortic Arch Diameter (Ao Arch
• Doppler Mode • Occipitofrontal Diameter (OFD) Diam)
- PI (Pulsatility Index) • Transverse Abdominal Diameter (TAD) - Ascending Aortic Diameter (Ao Asc)
- RI (Resistive Index) • Transverse Cerebellar Diameter (TCD) - Descending Aortic Diameter (Ao
- S/D Ratio • Thorax Transverse Diameter (ThD) Desc Diam)
- D/S Ratio • Tibia Length (Tibia) - Aorta Isthmus (Ao Isthmus)
- A/B Ratio • Ulna Length (Ulna) - Aorta *** (Ao st junct)
- Max PG (Pressure Gradient) • Multi-Gestational Calculations • Aortic Valve
- Mean PG (Pressure Gradient) - Up to 4 fetuses - Aortic Valve Cusp Separation (AV
- SV (Stroke Volume) - Comparison of multiple fetus data Cusp)
- Heart Rate on a graph and a worksheet - Aortic Valve Area Planimetry (AVA Pl
Abdomen and Small Parts OB Worksheet ani met ry)
Measurements/Calculations • Patient Information - *** ( Trans AVA)
• Aorta Diameter Fetus Number • Left Atrium
• Renal Length CUA/AUA Selection - Left Atrium Diameter (LA Diam)
• Doppler Abdomen and Renal Artery Fetus Position - LA Length (LA Major)
Exam Calcs Placenta - LA W idth (LA Minor)
- Acceleration • Measurement Information - Left Atrium Diameter to AoRoot
- Acceleration Time (AT) AFI Diameter Ratio (LA/Ao Ratio)
- Peak Systole (PS), End Diastole (ED), AC - Left Atrium Area (LAA(d), LAA(s))
or Mid Diastole (MD) HC - Left Atrium Volume, Method of Disk
- Pulsatility Index (PI) BPD (LAEDV A2C, LAESV A2C ) (LAEDV
- S/D or D/S Ratio FL A4C, LAESV A4C)
- Resistive Index (RI) • Calculation Information • Left Ventricle
- TAMAX EFW - Left Ventricle Mass (LVPWd, LVPWs)
• Thyroid Length, Width, and Height EFW GP (growth percentile) - Left Ventricle Volume,
• Renal Volume FL/BPD Teichholz/Cubic (LVIDd, LVI Ds)
• Emergency medicine FL/AC - Left Ventricle Internal Diameter
HC/AC (LVIDd, LVI Ds)
Page 7 of 12
- Left Ventricle Length (LVLd, LVLs) - Pulmonic Diameter (Pulmonic Diam) - Aortic Valve Ejection Time (LVET)
- Left Ventricle Outflow Tract • Right Atrium • Left Atrium
Diameter (LVOT Diam) - Right Atrium Diameter, Length - Left Atrium Diameter to AoRoot
- Left Ventricle Posterior Wall (RAD Ma) Diameter Ratio (LA/Ao Ratio)
Thickness (LVPWd, LVPWs) - Right Atrium Diameter, Width - Left Atrium Diameter (LA Diam)
- Left Ventricle Length (LV Major) (RAD Mi) • Left Ventricle
- Left Ventricle Width (LV Minor) - Right Atrium Area (RAA) - Left Ventricle Volume,
- Left Ventricle Outflow Tract Area - Right Atrium Volume, Single Plane, Teichholz/Cubic (LVIDd, LVI Ds)
(LVOT) Method of Disk (RAAd) - Left Ventricle Internal Diameter
- Left Ventricle Area, Two - Right Atrium Volume, Systolic, (LVIDd, LVI Ds)
Chamber/Four Chamber/Short Axis Single Plane, Method of Disk (RAAs) - Left Ventricle Posterior Wall
(LVA (d), LVA (s)) • Right Ventricle Thickness (LVPWd, LVPWs)
- Left Ventricle Endocardial Area, - Right Ventricle Outflow Tract Area - Left Ventricle Ejection Time (LVET)
Width (LVA (d), LVA(s)) - Left Pulmonary Artery Area - Left Ventricle Pre-Ejection Period
- Left Ventricle Epicardial Area, (LPA Area) (LVPEP)
Length (LVAepi (d), LVAepi (s)) - Right Pulmonary Artery Area - Interventricular Septum (IVS)
- Left Ventricle Mass Index (LVPWd, (RPA Area) - Left Ventricle Internal Diameter (LVI
LVPWs) - Right Ventricle Internal Diameter D)
- Ejection Fraction, Teichholz/Cube (RVIDd, RVIDs) - Left Ventricle Posterior Wall
(LVIDd, LVIDs) - Right Ventricle Diameter, Length Thickness (LVPW)
- Left Ventricle Posterior Wall (RVD Ma) • Mitral Valve
- Fractional Shortening (LVPWd, - Right Ventricle Diameter, Width - E-Point-to-Septum Separation
LVPWs) (RVD Mi) (EPSS)
- Left Ventricle Stroke Index, - Right Ventricle Wall Thickness - Mitral Valve Leaflet Separation
Teichholz/Cube (LVIDd, LVIDs, and (RVAWd, RVAWs) (D-E Excursion)
Body Surface Area) - Right Ventricle Outflow Tract - Mitral Valve Anterior Leaflet
- Left Ventricle Fractional Shortening Diameter (RVOT Diam) Excursion (D-E Excursion)
(LVIDd, LVIDs) - Left Pulmonary Artery (LPA) • Pulmonic Valve
- Left Ventricle Stroke Volume, - Main Pulmonary Artery (MPA) - QRS complex to end of envelope
Teichholz/Cubic (LVIDd, LVIDs) - Right Pulmonary Artery (RPA) (Q-to-PV close)
- Left Ventricle Stroke Index, Single • System • Right Ventricle
Plane, Two Chamber, Method of - Interventricular Septum Thickness - Right Ventricle Internal Diameter
Disk (LVI Dd, LVIDs) (IVSd, IVSs) (RVIDd, RVIDs)
- Left Ventricle Stroke Index, Single - Inferior Vena Cava - Right Ventricle Wall Thickness
Plane, Four Chamber, Method of - Pulmonary Artery Diameter (MPA) (RVAWd, RVAWs)
Disk (LVI Dd, LVIDs) - Systemic Vein Diameter (Systemic - Right Ventricle Ejection Time (RVET)
- Left Ventricle Stroke Index, Bi-Plane, Diam) - Right Ventricle Pre-Ejection Period
Bullet, Method of Disk (LVAd, LVAs) - Patent Ductus Arterosis Diameter (RVPEP)
- Interventricular Septum (IVS) (PDA Diam) • System
- Left Ventricle Internal Diameter - Pericard Effusion (PEs) - Interventricular Septum Thickness
(LVI D) - Patent Foramen Ovale Diameter (IVSd, IVSs)
- Left Ventricle Posterior Wall (PFO Diam) - Pericard Effusion (PE(d))
Thickness (LVPW) - Ventricular Septal Defect Diameter - Interventricular Septum (IVS)
• Mitral Valve (VSD Diam) Fractional Shortening (IVSd, IVSs)
- Mitral Valve Annulus Diameter (MV - Interventricular Septum (IVS) • Tricuspid Valve
Ann Diam) Fractional Shortening (IVSd, IVSs) - QRS complex to end of envelope
- E-Point-to-Septum Separation • Tricuspid Valve (Q-to-TV close)
(EPSS) - Tricuspid Valve Area (TV Panimetry Doppler Mode Measurements
- Mitral Valve Area by Pressure Half Æ TVA Panimetry) • Aortic Valve
Time - Tricuspid Valve Annulus Diameter - Aortic Insufficiency Mean Pressure
- Mitral Valve Area Planimetry (MVA (TV Annulus Diam) Gradient (AR Trace)
Planimetry) M-Mode Measurements
• Pulmonic Valve • Aorta - Aortic Insufficiency End Diastole
- Pulmonic Valve Area (PV • Aortic Valve Pressure Gradient (AR Trace)
Planimetry) - Aortic Valve Diameter (AV Diam) - Aortic Insufficiency Mean Velocity
- Pulmonic Valve Annulus Diameter - Aortic Valve Cusp Separation (AV (AR Trace)
(PV Annulus Diam) Cusp)
Page 8 of 12
- Aortic Insufficiency Mean Square - Mitral Valve Regurgitant Flow • Pulmonic Valve
Root Velocity (AR Trace) Acceleration (MR Trace) - Pulmonic Insufficiency Peak
- Aortic Insufficiency Velocity Time - Mitral Valve Regurgitant Mean Pressure Gradient (PR Vmax)
Integral (AR Trace) Velocity (MR Trace) - Pulmonic Insufficiency End-
- Aortic Valve Mean Velocity (AV - Mitral Regurgitant Mean Square Diastolic Pressure Gradient
Trace) Root Velocity (MR Trace) (PRTrace)
- Aortic Valve Mean Square Root - Mitral Regurgitant Mean Pressure - Pulmonic Valve Peak Pressure
Velocity (AV Trace) Gradient (MR Trace) Gradient (PV Vmax)
- Aortic Valve Velocity Time Integral - Mitral Regurgitant Velocity Time - Pulmonic End-Diastolic Pressure
(AV Trace) Integral (MR Trace) Gradient (PR Trace)
- Aortic Valve Mean Pressure - Mitral Valve Mean Velocity (MR - Pulmonic Insufficiency Peak
Gradient (AV Trace) Trace) Velocity (PR Vmax)
- Aortic Insufficiency End-Diastolic - Mitral Valve Mean Square Root - Pulmonic Insufficiency End-
Velocity (AR Trace) Diastolic Velocity (Prend Vmax)
- Aortic Valve Peak Velocity (AV Velocity (MR Trace) - Pulmonic Valve Peak Velocity
Vmax) - Mitral Valve Velocity Time Integral (PV Vmax)
- Aortic Valve Peak Velocity at Point E (MR Trace) - Pulmonic End-Diastolic Velocity
(AV Vmax) - Mitral Valve Mean Pressure (PV Trace)
- Aorta Proximal Coarctation (Coarc Gradient (MR Trace) - Pulmonary Artery Diastolic Pressure
Pre-Duct) - Mitral Regurgitant Peak Pressure (PV Trace)
- Aorta Distal Coarctation (Coarc Gradient (MR Vmax) - Pulmonic Insufficiency Mean
Post-Duct) - Mitral Valve Peak Pressure Gradient Pressure Gradient (PR Trace)
- Aortic Valve Insufficiency Pressure (MR Vmax) - Pulmonic Valve Mean Pressure
Half Time (AR PHT) - Mitral Regurgitant Peak Velocity Gradient (PV Trace)
- Aortic Valve Flow Acceleration (AV (MR Vmax) - Pulmonic Insufficiency Mean
Trace) - Mitral Valve Peak Velocity (MR Velocity (PR Trace)
- Aortic Valve Pressure Half Time (AV Vmax) - Pulmonic Insufficiency Mean
Trace) - Mitral Valve Velocity Peak A Square Root Velocity(PR Trace)
- Aortic Valve Acceleration Time (AV (MV A Velocity) - Pulmonic Insufficiency Velocity
Acc Time) - Mitral Valve Velocity Peak E Time Integral (PR Trace)
- Aortic Valve Deceleration TIme (AV (MV E Velocity) - Pulmonic Valve Mean Velocity
Trace) - Mitral Valve Area according to PHT (PV Trace)
- Aortic Valve Ejection Time (AVET) (MV PHT) - Pulmonic Valve Mean Square Root
- Aortic Valve Acceleration to - Mitral Valve Flow Deceleration Velocity (PV Trace)
Ejection Time Ratio (AV Acc Time, (MV Trace) - Pulmonic Valve Velocity Time
AVET) - Mitral Valve Flow Acceleration Integral (PV Trace)
• Left Ventricle (MV Trace) - Pulmonic Insufficiency Pressure
- Left Ventricle Outflow Tract Peak - Mitral Valve E-Peak to A-Peak Ratio Half Time (PR PHT)
Pressure Gradient (VLOT VmaxÆ (A-C and D-E) (MV E/ARatio) - Pulmonic Valve Flow Acceleration
LVOT Vmax) - Mitral Valve Acceleration Time (PV Acc Time)
- Left Ventricle Outflow Tract Peak (MV Acc Time) - Pulmonic Valve Acceleration Time
Velocity (LVOT Vmax) - Mitral Valve Deceleration Time (PV Acc Time)
- Left Ventricle Outflow Tract Mean (MV Dec Time) - Pulmonic Valve Ejection Time (PVET)
Pressure Gradient (LVOT Trace) - Mitral Valve Ejection Time - QRS complex to end of envelope
- Left Ventricle Outflow Tract Mean (MV Trace) (Q-to-PV close)
Velocity (LVOT Trace) - Mitral Valve A-Wave Duration - Pulmonic Valve Acceleration to
- Left Ventricle Outflow Tract Mean (MV A Dur) Ejection TIme Ratio (PV Acc Time,
Square Root Velocity (LVOT Trace) - Mitral Valve Time to Peak PVET)
- Left Ventricle Outflow Tract Velocity (MV Trace) - Pulmonic Valve Pre-Ejection to
Time Integral (LVOT Trace) - Mitral Valve Acceleration Ejection Time Ratio ( PVET)
- Left Ventricle Ejection Time (LVET) Time/Deceleration Time Ratio • Right Ventricle
- Cardiac Output by Aortic Flow (AVA (MVAcc/Dec Time) - Right Ventricle Outflow Tract Peak
Pl ani met ry, AV Trace) - Stroke Volume Index by Mitral Flow Pressure Gradient (RVOT Vmax)
- Stroke Volume Index by Aortic Flow (MVA Planimetry, MVTrace) - Right Ventricle Systolic Pressure
(AVA Planimetry, AV Trace) - Mitral Valve Area from Continuity (RVOT Vmax)
• Mitral Valve Equation (MVAPlanimetry, LVOT - Right Ventricle Outflow Tract Peak
Vmax) Velocity (RVOT Vmax)
Page 9 of 12
- Right Ventricle Diastolic Pressure - Tricuspid Regurgitant Velocity Time - Cardiac Output by Aortic Flow (AVA
(RVOT Trace) Integral (TR Trace) Planimetry, AV Trace, HR)
- Right Ventricle Outflow Tract - Tricuspid Valve Mean Velocity - Aortic Valve Area by Continuity
Velocity Time Integral (RVOTTrace) (TV Trace) Equation VTI (Ao Root Diam, LVOT
- Tricuspid Valve Mean Square Root Vmax, AV Trace)
- Stroke Volume by Pulmonic Flow ( Velocity (TV Trace) • Left Ventricle
RVOTTrace) - Tricuspid Valve Velocity Time - Cardiac Output, Teichholz/Cubic
- Right Ventricle Stroke Volume Integral (TV Trace) (LVIDd, LVI Ds, HR)
Index by Pulmonic Flow (RVOT - Tricuspid Valve Time to Peak - Cardiac Output Two Chamber,
Planimetry, RVOT Trace) (TV Acc/Dec Time) Single Plane, Area-Length/ Method
- Tricuspid Valve Ejection Time of Disk (Simpson) (LVAd, LVAs, HR)
• System (TV Acc/Dec Time) - Cardiac Output Four Chamber,
- Pulmonary Artery Peak Velocity - Tricuspid Valve A-Wave Duration Single Plane, Area-Length/ Method
(PV Vmax) (TV A Dur) of Disk (Simpson) (LVAd, LVAs, HR)
- Pulmonary Vein Velocity Peak A - QRS complex to end of envelope - Ejection Fraction Two Chamber,
(reverse) (P Vein A) (Q-to-TV close) Single Plane, Area-Length/ Method
- Pulmonary Vein Peak Velocity - Tricuspid Valve Pressure Half Time of Disk(Simpson) (LVAd, LVAs)
(P Vein D, P Vein S) (TV PHT) - Ejection Fraction Four Chamber,
- Systemic Vein Peak Velocity - Stroke Volume by Tricuspid Flow Single Plane, Area-Length/ Method
(PDA Diastolic, PDA Systolic) ( TV Trace) of Disk (Simpson) (LVAd, LVAs)
- Ventricular Septal Defect Peak - Tricuspid Valve E-Peak to A-Peak - Left Ventricle Stroke Volume, Single
Velocity (VSD Vmax) Ratio (TV E/A Velocity) Plane, Two Chamber/Four
- Pulmonary Artery Velocity Time Color Flow Mode Measurements Chamber, Area-Length (LVAd, LVAs)
Integral (PV Trace) • Aortic Valve - Left Ventricle Stroke Volume, Single
- Pulmonary Vein A-Wave Duration - Proximal Isovelocity Surface Area: Plane, Two Chamber/Four
(P Vein A Dur) Regurgitant Orifice Area (AR Radius) Chamber, Method of Disk(Simpson)
- IsoVolumetric Relaxation Time - Proximal Isovelocity Surface Area: (LVIDd, LVIDs, LVAd, LVAs)
(IVRT) Radius of Aliased Point (AR Radius) - Left Ventricle Volume, Two
- IsoVolumetric Contraction Time - Proximal Isovelocity Surface Area: Chamber/Four Chamber, Area-
(IVCT) Regurgitant Flow (AR Trace) Length (LVAd, LVAs)
- Pulmonary Vein S/D Ratio - Proximal Isovelocity Surface Area: - Ejection Fraction, Bi-Plane, Method
(P Vein D, P Vein S) Regurgitant Volume Flow (AR Trace) of Disk (LVAd, LVAs, 2CH, 4CH)
- Ventricular Septal Defect Peak - Proximal Isovelocity Surface Area: - Left Ventricle Stroke Volume, Bi-
Pressure Gradient (VSD Vmax) Aliased Velocity (AR Vmax) Plane, Method of Disk (LVAd, LVAs,
- Pulmonic-to-Systemic Flow Ratio • Mitral Valve 2CH, 4CH)
(Qp/Qs) - Proximal Isovelocity Surface Area: - Left Ventricle Volume, Bi-Plane,
• Tricuspid Valve Regurgitant Orifice Area (MR Method of Disk (LVAd, LVAs, 2CH,
- Tricuspid Regurgitant Peak Radius) 4CH)
Pressure Gradient (TR Vmax) - Proximal Isovelocity Surface Area: - Left Ventricle Stroke Index, Single
- Tricuspid Valve Peak Pressure Radius of Aliased Point (MR Radius) Plane, Two Chamber/Four
Gradient (TV Vmax) - Proximal Isovelocity Surface Area: Chamber, Area-Length (LVSd, LVSs,
- Tricuspid Regurgitant Peak Velocity Regurgitant Flow (MR Trace) and BSA)
(TR Vmax) - Proximal Isovelocity Surface Area: - Left Ventricle Volume, Single Plane,
- Tricuspid Valve Peak Velocity Regurgitant Volume Flow (MR Two Chamber/Four Chamber,
(TV Vmax) Trace) Method of Disk (LVAd, LVAs)
- Tricuspid Valve Velocity Peak A - Proximal Isovelocity Surface Area: - Left Ventricle Volume, Apical View,
(TV A Velocity) Aliased Velocity (MR Vmax) Long Axis, Method of Disk (LVAd,
- Tricuspid Valve Velocity Peak E Combination Mode Measurements LVAs)
(TV E Velocity) • Aortic Valve - Stroke Volume by Aortic Flow (AVA
- Tricuspid Regurgitant Mean - Aortic Valve Area (Ao Root Diam, Planimetry, AV Trace)
Pressure Gradient (TR Trace) LVOT Vmax, AV Vmax) • Mitral Valve
- Tricuspid Valve Mean Pressure - Aortic Valve Area by Continuity - Stroke Volume by Mitral Flow (MVA
Gradient (TV Trace) Equation by Peak Velocity (Ao Root Planimetry, MV Trace)
- Tricuspid Regurgitant Mean Diam, LVOT Vmax, AV Vmax) - Cardiac Output by Mitral Flow (MVA
Velocity (TR Trace) - Stroke Volume by Aortic Flow (AVA Planimetry, MV Trace, HR)
- Tricuspid Regurgitant Mean Square Pl ani met ry, AV Trace) • Pulmonic Valve
Root Velocity (TR Trace)
Page 10 of 12
- Stroke Volume by Pulmonic Flow • Generic Worksheet - Applications: Vascular, Small Parts,
(PV Planimetry, PV Trace) • Intravessel Ratio - Probe Frequency Range : 4-12 MHz
- Cardiac Output by Pulmonic Flow Pediatrics - Number of Element: 128
(PV Planimetry, PV Trace, HR) Measurements/Calculations - FOV(max) : 39 mm
• Tricuspid Valve • Alpha HIP - Lens Foot Print : 42 x 8 mm
- Cardiac Output by Tricuspid Flow • d: D Ratio - B-mode Imaging Frequency : 6.0,
(TV Planimetry, TV Trace, HR) • HIP (BA) 7.0, 8.0, 10.0,11.0 MHz
Urology Measurements/Calculations - Harmonic Imaging Frequency : 6.0,
• Bladder (0.7) Length, Height and Width 8.0, 10.0, 11.0, 12.0 Mhz
Cardiac Worksheet • Prostate Length, Height and Width - CFM Imaging Frequency : 4.0, 4.4,
Vascular • Renal Length, Height and Width 5.0 MHz
Measurements/Calculations • STVOL - Doppler Frequency : 4.0, ,4.4, 5.0
Exam Categolies MHz
• Generic - Steered Angle :+/-20°
• Carotid Artery - Biopsy Guide Available : Multi Angle
• Lower Extremity Artery
Probes
• Lower Extremity Vein • 12L-RS Wide Band Linear Probe
• Renal Artery • 4C-RS Wide Band Convex Probe - Applications: Vascular, Small Parts,
• Upper Extremity Artery - Applications: Abdomen, OB Gyn, Neonatal, Pediatrics
• Upper Extremity Vein Urology - Probe Frequency Range : 5-13 MHz
B-Mode Measurements - Probe Frequency Range : - Number of Element: 192
• % Stenosis 2.0~5.5MHz - FOV(max) : 39mm
- Diameter - Number of Element: 128 - Lens Foot Print : 42 x 7 mm
- Area - Convex Radius : 60 mmR - B-mode Imaging Frequency : 7.0,
• Volume - FOV : 55° 8.0, 10.0, 12.0 MHz
- One distance - Lens Foot Print : 65 x 16 mm - Harmonic Imaging Frequency : 8.0,
- Two distances - B-mode Imaging Frequency : 2.0, 10.0, 12.0, 13.0 MHz
- Three distances 3.0, 4.0, 5.0 MHz - CFM Imaging Frequency : 5.0, 6.7
• A/B Ratio - Harmonic Imaging Frequency : MHz
- Diameter 4.0, 5.0, 5.2, 5.5 Mhz - Doppler Frequency : 5.0, 6.7 MHz
- Area - CFM Imaging Frequency : 2.5, 3.3 - Steered Angle :+/-20°
M-Mode Measurements Mhz - Biopsy Guide :
• % Stenosis - Doppler Frequency : 2.5, 3.3 MHz - Multi Angle Biopsy attachement
- Diameter - Biopsy Guide Available :
• A/B Ratio Reusable Bracket, Disposable
- Diameter Sleeve
- Time • E8C-RS Wide Band Microconvex Probe
- Velocity • 3S-RS Wide Band Phase Probe - Applications: OB, Gyn, Urology,
Doppler Mode Measurements - Applications: Cardiac, Abdomen, Endocavity
Auto Vascular Calculation Gyn - Probe Frequency Range : 4.0 – 10.0
• Acceleration - Probe Frequency Range : 1.7- 4 MHz
• Acceleration Time (AT) MHz - Number of Element: 128
• End Diastole (ED), Mid Diastole (MD) or - Number of Element: 64 - Convex Radius : 11 mmR
Peak Systole (PS) - FOV : 90° - FOV : 133°
• ED/PS or PS/ED Ratio - Lens Foot Print : 21 x 15 mm - Lens Foot Print : 26 x 10 mm
• Heart Rate - B-mode Imaging Frequency : 2.0, - B-mode Imaging Frequency : 6.0,
• Pulsatility Index (PI) 2.5, 3.0MHz 8.0, 10.0 MHz
• Resistive Index (RI) - Harmonic Imaging Frequency: - Harmonic Imaging Frequency : 8.0,
• TAMAX 2.8, 3.0, 3.2, 3.6, 3.8, 4.0 MHz 10.0 MHz
• Edit Trace - CFM Imaging Frequency: 1.7, 2.0, - CFM Imaging Frequency : 4.0, 5.0
• TAMEAN 2.2, 2.6 MHz MHz
• Volume Flow - Doppler Frequency : 1.7, 2.0, 2.2, - Doppler Frequency : 4.0, 5.0 MHz
• PV 2.6MHz - Biopsy Guide Available : Fixed Angle,
Vascular Worksheet - Biopsy Guide Available : Multi Disposable, or Reusable.
• Vessel Worksheet Angle
• Vessel Summary • 8C-RS Wide Band Microconvex Probe
• Examiner’s Comments • 8L-RS Wide Band Linear Probe - Applications: Pediatrics
Page 11 of 12
- Probe Frequency Range : 4.0 – 10.0 - Lens Foot Print : 44 x 9mm display (MI, TIS, TIB, TIC)
MHz - B-mode Imaging Frequency : 6.0,
Not all features or specifications
- Number of Element: 128 8.0, 10.0 MHz described in this document may
- Convex Radius : 11 mmR - Harmonic Imaging Frequency : 8.0, available in all probes and/or modes.
- FOV : 133° 10.0, 12.0 MHz
- Lens Foot Print : 26 x 10 mm - CFM Imaging Frequency : 4.0, 4.4, General Electric Company reserves the
right to make changes in
- B-mode Imaging Frequency : 6.0, 5.0 MHz
specifications and features shown
8.0, 10.0 MHz - Doppler Frequency : 4.0, 4.4, 5.0 herein, or discontinue the product at
- Harmonic Imaging Frequency : 8.0, MHz any time without notice or obligation.
9.0, 10.0 MHz - Steered Angle :+/-10°, +/-15° Contact GE Representative for the
- CFM Imaging Frequency : 4.0, 5.0 - Biopsy Guide Not Available most current information.
MHz © 2009 General Electric Company - All
- Doppler Frequency : 4.0, 5.0 MHz rights reserved.
- Biopsy Guide Available : Biopsy not Inputs and Outputs GE, GE Monogram, TruScan and
support CrossXBeam are trademarks of
• Outputs General Electric Company.
- VGA Sony is a registered trademark
• i12L-RS Wide Band Linear Probe of Sony Corporation
- Earphone Port
- Applications: Vascular, Small Parts, Panasonic is a registered trademark of
• Connectors
Intra-operative Panasonic
- USB (Footswitch, DVD-RW, video NETGEAR is a registered trademark of
- Probe Frequency Range: 4 -10 MHz
printer) NETGEAR Inc.
- Number of Element: 96 DeskJet is a registered trademark of
- DC Power input
- FOV(max) : 25mm Hewlett-Packard
- Ethernet port
- Lens Foot Print : 29 x 10 mm
- Docking Connector
- B-mode Imaging Frequency : 6.0,
8.0, 10.0 MHz
- Harmonic Imaging Frequency : 8.0,
10.0 MHz Safety Conformance
- CFM Imaging Frequency : 4.0, 5.0
MHz LOGIQ e is:
- Doppler Frequency : 4.0, 5.0 MHz • Listed to UL 60601-1 by a
- Steered Angle :+/-20° Nationally Recognized Test Lab
- Biopsy Guide Not Available • Certified to CAN/CSA-C 22.2
No.601.1 by an SCC accredited Test
• 9L-RS Wide Band Linear Probe Lab
- Applications: Vascular, Small Parts • CE Marked to Council Directive
- Probe Frequency Range:3.33 -10 93/42/EEC on Medical Devices
MHz • Conforms to the following
standards for safety:
- Number of Element: 192
- IEC 60601-1 Electrical medical
- FOV(max) : 44mm equipment
- Lens Foot Print : 44 x 7.0mm - IEC 60601-1-1 Electrical medical
- B-mode Imaging Frequency : 5.0, equipment
7.0, 9.0 MHz - IEC 60601-1-2 Electromagnetic
- Harmonic Imaging Frequency : 8.0, compatibility
10.0 MHz - IEC 60601-1-4 Programmable
medical systems
- CFM Imaging Frequency : 3.33, 4.0,
- IEC 60601-1-6 2004 Medical
5.0MHz Electrical Equipment – Part 6:
- Doppler Frequency : 3.33, 4.0, General Requirements for
5.0MHz safety – Usability
- Steered Angle :+/-20° - IEC 61157 Declaration of
- Biopsy Guide Available : Multi Angle acoustic output
- IEC60601-2-37: Particular
requirements for the safety of
• I/T739L-RS Wide Band Linear Probe
ultrasonic medical diagnostic and
- Applications: Vascular, Small Parts,
monitoring equipment
Intra-operative - ISO 10993 Biological evaluation
- Probe Frequency Range:4 -12 MHz of
- Number of Element: 96 medical devices
- FOV(max) : 39mm - NEMA UD3 Acoustic output
Page 12 of 12