MM Merged 1
MM Merged 1
Measurements
By
Dr. Ch. Sampath Kumar
Assistant Professor
Mechanical Engineering Department
Nit Warangal
What is Measurement?
Act of quantitative comparison between a predetermined standard and an unknown
magnitude.
Applications of Measurement
1. Monitoring of processes and operations
2. Control of processes and operations
3. Experimental engineering analysis
Fundamental Measuring Process
Measurand is used to designate the particular physical parameter being observed and quantified.
Measurand
By
Dr. Ch. Sampath Kumar
Assistant Professor
Mechanical Engineering
Department
Nit Warangal - 506004
Measurement Terminologies
Readability – closeness or ease with which the scale of an instrument may
be read. Fine and widely spaced graduation lines improve the readability.
Eg.- To make the micrometres more readable they are provided with
vernier scale or magnifying devices.
Measurement Terminologies
Resolution: It is basically the smallest input that an instrument can detect and
its output changes.
Measurement Terminologies
Hysteresis – difference in output signal for a given input signal depending on whether the input
signal is approached in increasing levels or decreasing levels.
Causes - dry friction, elastic elements…
Accuracy –Accuracy is about how close measurements are to the ‘true answer’.
The degree of closeness of the output to the known input. It is usually indicated as
percentage of full scale reading. The deviation is called as Error.
Precision – Precision is about how close measurements are to one another. Ability
to reproduce a certain output with a given accuracy.
Calibration – comparison of the output signal of a given instrument with a standard or
instrument of known accuracy or with known input source, to improve its accuracy.
The role of standards is to achieve uniform, consistent and repeatable measurements throughout
the world. It facilitates interchangeability of parts and identifies the method of manufacture.
By
Dr. Ch. Sampath Kumar
Assistant Professor
Mechanical Engineering Department
Nit Warangal - 506004
Generalized Measurement System
Three stages
1. Detector or Transducer
stage (Input stage) – Detects
and transforms the signal to
more usable form.
2. Intermediate stage –
enhances the signal by
amplification, filtering,
attenuation etc…
3. Final or Output stage –
indicates, records or
control the variable being
measured.
Generalized Measurement System
Types of inputs based on time dependence
1. 11.
Static Dynamic
1. Desired input
Spurious inputs
2. Interfering input (when they affect the measurement)
3. Modifying input
Types of inputs and their Correction
1. Desired input Spurious inputs
2. Interfering input (when they affect the measurement)
3. Modifying input
By
Dr. Ch. Sampath Kumar
Assistant Professor
Mechanical Engineering Department
Nit Warangal - 506004
Response of Measuring systems
Response: It may be defined as an evaluation of the system's ability to faithfully sense, transmit,
and present all the pertinent information included in the measurand and to exclude all else.
Response of Measuring systems Cont….
Static measurement is done when the input (physical quantity measured) is not
changing with time
Input
signal
Time
Dynamic Measurement
The fundamental concepts of dynamic response, however, can be understood by studying relatively simple
mathematical models
Dynamic Measurement
Dynamic Measurement
Dynamic Measurement
2 5
3
Dynamic Measurement
Dynamic Measurement
This involves the transformation of a force input function into a displacement function.
Dynamic Measurement
Dynamic Measurement
Dynamic Measurement
Dynamic Measurement
Under a static input a second order linear instrument tends to oscillate about its position of equilibrium.
The natural frequency of the instrument is the frequency of these oscillations.
Friction in the instrument opposes these oscillations with a strength proportional to the rate of change of
the output. The damping factor is a measure of this opposition to the oscillations.
Dynamic Measurement
By
Dr. Ch. Sampath Kumar
Assistant Professor
Mechanical Engineering Department
Nit Warangal - 506004
Types of elementary inputs
Step
Input
Time
Response to change of input
Suppose also that y(t) = 0 for t < 0. We want to find y(t) for t > 0
Response to change of input
Response to change of input
Response to change of input
Response to change of input
Response to change of input
Response to change of input
τ
Response to change of input
By
Dr. Ch. Sampath Kumar
Assistant Professor
Mechanical Engineering Department
Nit Warangal - 506004
Response to change of input
Unit Step Function
Second Order Instrument
Consider that f(t) = 0 at t = 0 and f(t) = F1 when t > 0, with the initial condition y =
y0 at t = 0. The solution for the above equation with the given condition is considered
for four cases.
(i) Undamped system (ζ =0)
(ii) Underdamped system (ζ < 1)
(iii) Critically damped system (ζ = 1)
(iv) Over damped system (ζ > 1)
Response to change of input
Case 0. For,
or
Response to change of input
(a)
Response to change of input
Response to change of input
Harmonic Function
Second Order Instrument
Spring balance
Response to change of input
Response to change of input
Harmonic Function
Second Order Instrument
Harmonic Function
First Order Instrument
Example 1: The approximate time constant of a thermometer is determined by immersing it
in a bath and noting the time it takes to reach 63% of the final reading. If the result is 28 s,
determine the delay when measuring the temperature of a bath that is periodically changing 2
times per minute.
(td = 6.7 s)
Response to change of input
Unit Step Function
Second Order Instrument
Example 1: A pressure transducer has a natural frequency of 30 rad/s, damping ratio of 0.1
and static sensitivity of 1.0 μV/Pa. A step pressure input of 8x105 N/m2 is applied. Determine
the output of the transducer.
Response to change of input
Harmonic Function
Second Order Instrument
Example 1: A second order instrument is subjected to a sinusoidal input.
Undamped natural frequency is 3 Hz and damping ratio is 0.5. Calculate the
amplitude ratio and phase angle for an input frequency of 2 Hz.
ME304 – Mechanical Measurements
By
Dr. Ch. Sampath Kumar
Assistant Professor
Mechanical Engineering Department
NIT Warangal
Error: Deviation from the true value
• Errors
•Systematic
•Random
Systematic errors: Errors that are not determined by chance but are introduced by an inaccuracy
inherent in the system. They are also called as fixed or bias error.
Random errors: Random errors are errors in measurement that lead to measurable values being
inconsistent when repeated measures of a constant attribute or quantity are taken.
Introduction
Analysis of data pertains to determine errors, precision, and general validity of experimental
measurement
Errors will creep into all experiments regardless of the care taken.
Our task is to determine just how uncertain a particular observation may be and to devise a
consistent way of specifying the uncertainty in analytical form
Systematic error, is predictable and typically constant or proportional to the true value. If the
cause of the systematic error can be identified, then it usually can be eliminated.
Random errors show up as different results for ostensibly the same repeated measurement.
They can be estimated by comparing multiple measurements, and reduced by averaging multiple
measurements.
Systematic error cannot be discovered this way because it always pushes the results in the same
direction.
If the cause of a systematic error can be identified, then it can usually be eliminated.
Error
Example for simple error calculation
Power P=EI
E = 100 ± 2 V
I = 10 ± 0.2 A
Pmax = (100+2)(10+0.2) = 1040.4 W
Pmin = (100-2)(10-0.2) = 960.4 W
Uncertainty of measurement is the doubt that exists about the result of any measurement.
Error is the difference between the measured value and the ‘true value’ of the thing being
measured.
Whenever possible we try to correct for any known errors: for example, by applying corrections
from calibration certificates. But any error whose value we do not know is a source of uncertainty.
Uncertainty
Uncertainty is expressed in general with two parameters. One is the width of the margin, or
interval. The other is a confidence level, and states how sure we are that the ‘true value’ is
within that margin.
For example:
We might say that the length of a certain stick measures 20 centimetres plus or minus 1
centimetre, at the 95 percent confidence level. This result could be written:
The statement says that we are 95 percent sure that the stick is between 19 centimetres and 21
centimetres long.
Uncertainty Analysis
•
Uncertainties for Product Functions
Suppose the result function takes the form of a product of the respective primary variables raised to exponents
and expressed as
(1)
Inserting this relation in the following equation 4 gives the final equation 5
(4)
(5)
Uncertainties for additive Functions
(1)
(3)
Uncertainty Analysis
Problem: The resistance of a certain copper wire is given as
R = R0[1+𝛼(T-20)]
where R0 = 6Ω ± 0.3% at 20oC, 𝛼 = 0.004oC-1 ± 1% and the temperature of the wire is T = 30 ±
1oC. Calculate the resistance of the wire and its uncertainty.
Soln: R = 6.24 ohms and WR = 0.0305 ohms or 0.49%
By
Dr. Ch. Sampath Kumar
Assistant Professor
Mechanical Engineering Department
NIT Warangal
Uncertainty Analysis
•
Uncertainties for Product Functions
Suppose the result function takes the form of a product of the respective primary variables raised to exponents
and expressed as
(1)
Inserting this relation in the following equation 4 gives the final equation 5
(4)
(5)
Uncertainties for additive Functions
(1)
(3)
Our primary concern is the application of probability and statistics to the analysis of experimental data.
When a set of readings of an instrument is taken, the individual readings will vary somewhat from each other,
and the experimenter may be concerned with the mean of all the readings
Arithmetic
mean,
Sometimes it is appropriate to use a geometric mean when studying phenomena which grow in proportion to their size
Probability is a mathematical quantity that is linked to the frequency with which a certain phenomenon occurs after a
large number of tries
Binomial distribution:
Poisson distribution: The limit of the binomial distribution as N → ∞ and p → 0 such that Np = a = const
The Gaussian or Normal Error Distribution: If the measurement is designated by x, the gaussian distribution gives
the probability that the measurement will lie between x and x+dx and
is written
The Gaussian or Normal Error Distribution
For any large number of measurement made to determine a particular value, the
random errors are expected to have a normal distribution, given that the
experimental system is controlled well.
The Gaussian or Normal Error Distribution
•
The Gaussian or Normal Error Distribution
•
The Gaussian or Normal Error Distribution
•
The Gaussian or Normal Error Distribution
Problem
: A certain steel bar is measured with a device which has a known
deviation of ±0.5mm w
hen a large number of measurement is taken. How many
measurements are necessary to establish the mean length with a 5%
level of significance such that xm= xm±0.2mm.
(Soln: 25)
R (kΩ): 1.22, 1.23, 1.26, 1.21, 1.22, 1.22, 1.22, 1.24, 1.19
What is the best estimate for the resistance and what is the error with 95%
The Gaussian or Normal Error Distribution
ME304 – Mechanical Measurements
By
Dr. Ch. Sampath Kumar
Assistant Professor
Mechanical Engineering Department
NIT Warangal
Problem: Resistance of a certain copper wire is measured repeatedly to obtain the following data.
R (kΩ): 1.22, 1.23, 1.26, 1.21, 1.22, 1.22, 1.22, 1.24, 1.19
What is the best estimate for the resistance and what is the error with 95% confidence?
(Soln: 1.22kΩ, 0.04kΩ)
Sol:
Regression analysis aims to determine the best global representation of experimental data using a chosen functional
form
Linear relation between Y and X Linear relation between log Y and log X Non Linear relation between Y and X
The non-linear relationship follows a polynomial relationship of the form
The polynomial form is treated as a linear regression noting that it may be written in the linear form
•
Regression Analysis
•
Polynomial Regression
If it were desired to obtain a least-squares fit according to the quadratic function
Regression Analysis
The goodness of the fit is given by a term called as the correlation coefficient defined by,
Problem: Obtain the equations to find the fit parameters for y as a linear function of x.
We seek an equation of the form y = ax + b
Problem: The following data is expected to follow a quadratic relationship. Determine the fit parameters and
the correlation coefficient.
By
Dr. Ch. Sampath Kumar
Assistant Professor
Mechanical Engineering Department
Nit Warangal - 506004
Pressure
Force per unit area exerted by the fluid on the walls of its container
• Pressure difference is used many a time as a means of measuring the flow rate of
a fluid
Methods of Pressure Measurement
• 1Pa = 1 N/m2
• 1 atm= 760 mm mercury column = 1.013×105
Pa
• 1 mm mercury column = 1 Torr
• 1 Torr = 1.316×10−3 atm= 133.3 Pa
• 1 bar = 105 Pa
Dead Weight Tester
1 – Hand pump
2 – Testing Pump
3 – Pressure Gauge to be calibrated
4 – Calibration Weight
5 – Weight Support
6 – Accurately Machines Piston
7 – Close fitting Cylinder
8 – Filling Connection
Dead Weight Tester
Dead weight testers can measure pressures of up to 10,000 bar, attaining accuracies of between
0.005% and 0.1% although most applications lie within 1 - 2500 bar.
•
Manometer
•
Manometer
Example 1: (a) A U-tube manometer employs a special oil of specific gravity 0.82 as the
manometer fluid. One limb of the manometer is exposed to an atmospheric pressure of 740mm
of Hg. The difference in column heights is measured as 20cm ± 1mm when exposed to an air
source at 25 deg C. Calculate the pressure differential in Pa and the uncertainty.
Soln: 1598.5 ± 8 Pa
(b) If the above manometer was carelessly mounted with an angle of 3o to the vertical, what is
the error in the indicated pressure due to this corresponding to the data given above?
Soln: -2.2 Pa
Manometer Variants
• Well type manometer
• Well type manometer with Inclined leg
Manometer Variants
•
Manometer Variants
Well type manometer with Inclined leg
In case the measured pressure difference is small one may use a well type manometer with
inclined tube shown in figure. This gives better sensitivity
h=Rsinθ
ME304 – Mechanical Measurements
3. Pressure Measurement
By
Dr. Ch. Sampath Kumar
Assistant Professor
Mechanical Engineering Department
Nit Warangal - 506004
Example 1: (a) A U-tube manometer employs a special oil of specific gravity 0.82 as the manometer
fluid. One limb of the manometer is exposed to an atmospheric pressure of 740mm of Hg. The
difference in column heights is measured as 20cm ± 1mm when exposed to an air source at 25 deg C.
Calculate the pressure differential in Pa and the uncertainty.
1598.5 ± 8 Pa
(b) If the above manometer was carelessly mounted with an angle of 3o to the vertical, what is the error in the
indicated pressure due to this corresponding to the data given above?
The mounting error has been given as θ = 3◦ with respect to the vertical.
It is clear that the manometer liquid height difference, with mounting error is given by
Note that there is thus a systematic error of 1596.3−1598.5 = −2.2 Pa because of mounting error.
This is about 25% of the error due to the error in the measurement of h.
Example 1: (a) A U-tube manometer employs a special oil of specific gravity 0.82 as the
manometer fluid. One limb of the manometer is exposed to an atmospheric pressure of 740mm
of Hg. The difference in column heights is measured as 20cm ± 1mm when exposed to an air
source at 25 deg C. Calculate the pressure differential in Pa and the uncertainty.
A)
Sol:
0.00009=0.74x0.0005(sinθ+0.1111)
θ =7.6 degrees
Accuracy Factors
To achieve the best accuracy from a dead weight tester the following factors should be
considered when calculating the pressure reading
Local Gravity
If the gravity at the location where the dead weight tester is being used is different to where it
was calibrated then the readings will need to be corrected since the force generated by the
weights will change.
Pressure Coefficient
The effective area of the dead weight tester will change with pressure due to the expansion and
contraction of the piston and cylinder at different pressures.
Tare Force
This represents a force error in the determination of the mass of the weights, the calculation of
the piston/cylinder and carrier weight or a known characteristic of the dead weight tester.
Horizontal Levelling
If the dead weight tester is not level with the horizontal plane and therefore the piston cylinder
assembly is not perfectly upright there will be a friction force due to a leaning towards one side.
Air Buoyancy
The air pressure and humidity will affect the density of air which contributes a relatively small
lifting force to the weights.
Thermal Expansion
As the temperature changes the cross sectional area of the piston and cylinder will change
affecting the effective area of the dead weight tester.
Surface Tension
For fluid filled dead weight testers a surface tension force is exerted on the piston. This surface
tension force depends on the type of fluid used and the circumference of the piston.
Height Difference
If the dead weight tester is used at a different height to the unit being tested, a force will be
generated due to the head of air or fluid in the connecting test hose .
Dynamic Response of Pressure Measurement System
•
Dynamic Response of Pressure Measurement System
•
Dynamic Response of Pressure Measurement System
•
Dynamic Response of Pressure Measurement System
•
Dynamic Response of Pressure Measurement System
•
Dynamic Response of Pressure Measurement System
•
Dynamic Response of Pressure Measurement System
•
Dynamic Response of Pressure Measurement System
•
Response of a U-tube manometer for step input
Dynamic Response of Pressure Measurement System
Example 1 : A U-tube manometer uses mercury as the manometer fluid having density of
13580kg/m3 and kinematic viscosity of 1.1 x 10-7m2/s. The total length of the liquid is 0.6m. The
tube diameter is 2mm. Determine the characteristic time and the damping ratio for this
installation. Redo the above with water as the manometer fluid. The density and kinematic
viscosity of water are 996kg/m3 and 10-6m2/s respectively.
[Soln: 0.247s, 0.109]
Hence the characteristic time using Equation 7.12 is τ = L g = 0.6 9.81 = 0.247 s
Hence the characteristic time using Equation 7.12 is τ = L g = 0.6 9.81 = 0.247 s
The manometer liquid properties are changed to ρm = 996 kg/m3;ν = 10−6 m/s2
The characteristic time remains unchanged since it is purely a function of length of manometer fluid column.
The damping ratio changes because of the change in the liquid viscous resistance.
With water as the manometer liquid we have RH2O = 128×10−6 ×0.6 π×0.0024 = 1527887.5 Pa s/kg
•
McLeod Gauge
•
ME304 – Mechanical Measurements
By
Dr. Ch. Sampath Kumar
Assistant Professor
Department of Mechanical Engineering
NIT Warangal 1
1
Introduction
• Temperature: A measure proportional to the average translational kinetic energy associated with the
disordered microscopic motion of atoms and molecules.
• The flow of heat is from a high temperature region toward a lower temperature region.
• The higher is the temperature of the object, the more kinetic energy the particles of an object have
The International Practical Temperature scale (ITS-90) assigns values of temperature to a few highly
reproducible states of matter, such as certain freezing points and triple point.
2
2
Temperature Scales
• These are based on specification of number of increments (100 in Celsius scale and 180 in
Fahrenheit scale) between the freezing point and boiling point of water at standard atmospheric
pressure.
• The absolute Celsius scale is called Kelvin scale and that of Fahrenheit is called as the Rankine
scale.
3
3
Temperature measurement
• Parameters like
• Pressure
• Volume
• Electrical resistance
• Expansion coefficients
changes with temperature. Hence these changes can be used to measure temperature.
4
4
Common temperature-sensing techniques.
6
6
The Ideal Gas Thermometer
• Basis of working – behavior of ideal gas at low pressure
V – volume occupied by the gas, m – mass, Rg – Gas constant for a particular gas given by Rg =
R/M, where R is universal gas constant (8.314 J/mol.K) and M is the molecular weight of the gas.
• The figure above shows a constant volume ideal gas thermometer. The pressure gauge reads
absolute pressure P and the metal ball has a fixed volume V
7
7
The Ideal Gas Thermometer
The metal ball is exposed to an unknown temperature (to be measured) and a known standard
temperature. The pressure in both cases are measured
Now,
T = Tref (p/pref)const. vol
8
8
The Ideal Gas Thermometer
9
9
The Ideal Gas Thermometer
The copper bulb of a constant volume gas thermometer has a radius of 0.05 meters. When the bulb is
filled to 1 atm pressure, there are ≈ 0.021 moles of gas in the bulb. Using the slope of the pressure vs
temperature plot, estimate the ideal gas constant (R). Calculate the percentage error with the
standard value for R of 8.314 J /mol.K .
10
10
Temperature Measurement by Mechanical effects
11
11
Temperature Measurement by Mechanical effects
Liquid-in-glass thermometer
• An inert gas, mainly argon or nitrogen is filled inside the thermometer above
mercury to trim down its volatilization.
12
12
Temperature Measurement by Mechanical effects
Liquid-in-glass thermometer
• The extent to which these thermometers are inserted into the medium under temperature
measurement principally decides the accuracy of results.
• Typically, three classes of immersion exist which are total, partial and complete immersion classified
according to the level of contact between the medium and the sensing element.
• An error can be produced when the thermometer is not immersed to the same extent as it was
when it was originally calibrated.
• An 'emergent stem correction' may be necessary when it is not possible to immerse the
thermometer sufficiently deeply.
13
13
Temperature Measurement by Mechanical effects
• The response time of a liquid-in-glass thermometer varies according to the kind of thermometer, its bulb
volume, thickness and overall weight.
• For getting quick response, the bulb of the thermometer should be designed in such a way that it results in small
and the bulb wall thin.
• Their sensitivity is based upon the reversible thermal expansion characteristics of the liquid in comparison to the
glass. The more the thermal expansion of the liquid, the highly sensitive the thermometer is.
• Organic liquids which are usually employed for construction of liquid-in-glass thermometers include toluene,
ethyl alcohol and pentane.
• Although their thermal expansion is high but they are non linear and their use is restricted for high
temperatures.
14
14
Temperature Measurement by Mechanical effects
Liquid-in-metal thermometer
15
15
Temperature Measurement by Mechanical effects
Bimetallic strips
The principle behind a bimetallic strip thermometer relies on the fact that different metals expand
at different rates as they warm up. By bonding two different metals together, you can make a simple
electric controller that can withstand fairly high temperatures.
16
16
Temperature Measurement by Mechanical effects
Bimetallic strips
Bulb thermometers are good for measuring temperature accurately, but they are harder to use
when the goal is to control the temperature. The bimetallic strip thermometer, because it is
made of metal, is good at controlling things.
Eskin and Fritze [3] have given calculation methods for bimetallic strips. The radius of curvature r may be
calculated as
17
17
Temperature Measurement by Mechanical effects
Bimetallic strips
Main Features
• These types of thermometers work best at higher temperatures, since their accuracy and
sensitivity tends to reduce at low temperatures.
• Bimetallic strip thermometers are manufactured in various designs. One of the most popular
design i.e. flat spiral is shown in the figure below. They can also be wound into a single helix or
multiple helix form.
18
18
Temperature Measurement by Mechanical effects
Bimetallic strips
Main Features
• Bimetallic thermometers can be customized to work as recording thermometers too by
affixing a pen to the pointer. The pen is located in such a way that it can make recordings on a
circling chart.
• Bimetallic strips often come in very long sizes. Hence, they are usually coiled into spirals which
make them compact and small in size. This also improves the sensitivity of bimetallic strips
towards little temperature variations.
• The bimetallic strip can be scaled up or down. On a large scale, it can provide literally tones of
force for mechanical control or other purposes. On a smaller scale, it can provide the force and
movement for micro machine integrated circuits (MMIs).
19
19
Temperature Measurement by Thermoelectric effects
▪ A thermoelectric device creates voltage when there is a different temperature on each side.
Conversely, when a voltage is applied to it, it creates a temperature difference.
▪ This effect can be used to generate electricity, measure temperature or change the temperature of
objects.
▪ Because the direction of heating and cooling is determined by the polarity of the applied voltage,
thermoelectric devices can be used as temperature controllers.
20
20
Temperature Measurement by Thermoelectric effects
The Seebeck effect is a phenomenon in which a temperature difference between junctions of two
dissimilar electrical conductors or semiconductors produces a voltage difference between the two
substances.
21
21
Temperature Measurement by Thermoelectric effects
▪ A thermocouple is a device consisting of two dissimilar conductors or semiconductors that contact each
other at one or more points.
▪ A thermocouple produces a voltage when the temperature of one of the contact points differs from the
temperature of another, in a process known as the thermoelectric effect (Seebeck Effect).
▪ The Seebeck effect creates an emf wherever there is a temperature gradient. This emf is used to develop an
open-circuit voltage. Under open-circuit conditions where there is no internal current flow, the gradient of
voltage ΔV is directly proportional to the temperature gradient ΔT.
▪ ΔV = S(T) ΔT, where S(T) is called as the Seebeck coefficient. It is a temperature dependent material
property.
22
22
Temperature Measurement by Thermoelectric effects
23
23
Temperature Measurement by Thermoelectric effects
Advantages of Thermocouples
Temperature range: Most practical temperature ranges, from cryogenics to jet-engine exhaust,
can be served using thermocouples. Depending on the metal wires used, a thermocouple is capable
of measuring temperature in the range –200°C to +2500°C.
Robust: Thermocouples are rugged devices that are immune to shock and vibration and are
suitable for use in hazardous environments.
Rapid response: Because they are small and have low thermal capacity, thermocouples respond
rapidly to temperature changes, especially if the sensing junction is exposed. They can respond to
rapidly changing temperatures within a few hundred milliseconds.
No self heating: Because thermocouples require no excitation power, they are not prone to self
heating and are intrinsically safe.
24
24
Temperature Measurement by Thermoelectric effects
Thermocouple circuit
Figure shows that the measured output voltage, VOUT, is the difference between the measuring (hot)
junction voltage and the reference (cold) junction voltage.
Since VH and VC are generated by a temperature difference between the two junctions
The scale factor, a, which relates the voltage difference to the temperature difference, is known as the
Seebeck coefficient.
25
25
Temperature Measurement by Thermoelectric effects
Disadvantages
26
26
Temperature Measurement by Thermoelectric effects
▪ When working with thermocouples, you must establish a reference point because thermocouples
are differential temperature-measurement devices.
▪ If you know both the temperature of the cold junction and the temperature of the hot junction
relative to the cold-junction temperature, you can determine the actual hot-junction temperature.
▪ When the two dissimilar materials are connected to a measuring device, there will be another
thermal emf generated at the junction of the materials and the connecting wires to the voltage-
measuring instrument.
27
27
Temperature Measurement by Thermoelectric effects
Rules for analysis of thermoelectric circuits
1. Law of intermediate materials
The net emf of the circuit, where any number of dissimilar materials are introduced, remains the
same as long as the new connections (junctions)are at same temperature.
Let E2 be the Seebeck emf with measuring junction at T2 and reference junction at T3.
Then the Seebeck emf E3 = E1+E2 will be the Seebeck emf with the measuring junction at T1 and
reference junction at T3.
28
28
Temperature Measurement by Thermoelectric effects
Commonly used thermocouple circuit
In this example, the open ends of each wire are electrically connected to wires or traces made of
copper. These connections introduce two additional junctions into the system.
As long as these two junctions are at the same temperature, the intermediate metal (copper) has
no effect on the output voltage.
This configuration allows the thermocouple to be used without a separate reference junction. VOUT
is still a function of the difference between hot- and cold-junction temperatures, related by the
Seebeck coefficient.
29
29
Temperature Measurement by Thermoelectric effects
Conventional reference cold junctions
In the early days of thermocouples, the ice-bath reference served as the standard in thermocouple
applications. Implementing an ice bath today is impractical in most situations.
Therefore, when the cold junction is not at 0°C, the temperature of this junction must be known in
order to determine the actual hot-junction temperature.
30
30
Temperature Measurement by Thermoelectric effects
Cold junction compensation
The output voltage of the thermocouple must also be compensated to account for the voltage created by
the nonzero cold-junction temperature.
To implement cold-junction compensation, the temperature of the cold junction must be determined.
32
32
33
33
Temperature Measurement by Thermoelectric effects
Thermopile
34
34
Temperature Measurement by Thermoelectric effects
Thermocouples in parallel
35
35
Temperature Measurement by Electrical effects
•
36
36
37
37
Temperature Measurement by Electrical effects
▪ Resistance thermometers, also called resistance temperature detectors (RTDs), are sensors used
to measure temperature by correlating the resistance of the RTD element with temperature.
▪ Care must be taken to ensure that the resistance wire is free of mechanical stresses and so
mounted that moisture cannot come in contact with the wire and influence the measurement
▪ Most RTD elements consist of a length of fine coiled wire wrapped around a ceramic or glass
core. The element is usually quite fragile, so it is often placed inside a sheathed probe to protect
it.
▪ They are slowly replacing the use of thermocouples in many industrial applications below 600 °C,
due to higher accuracy and repeatability.
38
38
Temperature Measurement by Electrical effects
Limitations:
1. Compared to thermistors, platinum RTDs are less sensitive to small temperature changes and
have a slower response time.
2. RTDs in industrial applications are rarely used above 660 °C due to contamination of resistance
element at higher temperature.
3. RTDs are also prone to self-heating and lead wire errors.
39 39
39
Temperature Measurement by Electrical effects
Choice of RTDs vs thermocouples
1. Temperature: If process temperatures are between −200 to 500 °C (−328.0 to 932.0 °F), an
industrial RTD is the preferred option. Thermocouples have a range of −180 to 2,320 °C and
so for temperatures above 500 °C (932 °F) they are the only contact temperature
measurement device.
1. Response time: If the process requires a very fast response to temperature changes—
fractions of a second as opposed to seconds then a thermocouple is the best choice.
1. Size : A standard RTD sheath is 3.175 to 6.35 mm (0.1250 to 0.2500 in) in diameter; sheath
diameters for thermocouples can be less than 1.6 mm (0.063 in).
For steady-state measurements a null condition will suffice, while transient measurements will
usually require the use of a deflection bridge.
The floating-potential arrangement in Fig. C is the same as the Siemen’s connection, but an extra lead is inserted.
This extra lead may be used to check the equality of lead resistance. 42
42
Temperature Measurement by Electrical effects
•
43
43
The thermistor is a very sensitive device, and consistent performance within 0.01◦C may be anticipated with
proper calibration.
44
44
Temperature Measurement by Electrical effects
Thermistor
Advantages
Limitations
45
45
Temperature Measurement by Radiation
▪ The emissive power of blackbody varies with wavelength as per the eqn.
Ebλ = C1λ-5/(eC2/ λT – 1)
Ebλ– monochromatic black body emissive power , W/m2
λ – wavelength (μm), T – temperature (K), C1=3.74x108 W. μm4/m2
C2 = 1.4387x104 μm.K
46
46
Temperature Measurement by Radiation
When thermal radiation strikes a surface then,
α+ρ+τ=1
α – absorprtivity
ρ – reflectivity
τ – transmissivity
Emissivity is defined as ε = (E/Eb), where E is the emissive power of the actual surface and Eb is the
emissive power of black body at the same temperature.
ε=α
47
47
Temperature Measurement by Radiation
Real surfaces exhibit highly variable emissivities over the wavelength spectrum.
Hence the real surface is frequently approximated as a gray body having an emissivity equal to the
average total emissivity of the real surface as defined by
ελ = (Eλ /Ebλ ),
48
48
Temperature Measurement by Radiation
The shift in these maximum points explains the change in color as the body is heated i.e., higher
temperature results in concentration of radiation in lower wavelength portion of the spectrum.
49
49
Temperature Measurement by Radiation
Optical Pyrometer
It is a device that from a distance determines the temperature of a surface from the spectrum of
the thermal radiation it emits.
50
50
Temperature Measurement by Radiation
Optical Pyrometer
It is a device that from a distance determines the temperature of a surface from the spectrum of
the thermal radiation it emits.
51 51
51
Temperature Measurement by Radiation
Optical Pyrometer
It is a device that from a distance determines the temperature of a surface from the spectrum of
the thermal radiation it emits.
E = εσT4
T = (E/εσ )(1/4)
52
52
Transient Temperature Measurement
53
53
Transient Temperature Measurement
•
54
54
1. A certain mercury-in-glass thermometer has been calibrated for a prescribed immersion
depth. The thermometer is immersed too much, such that the extra depth is equal to a
distance of 10◦ on the scale.
The true temperature reading may be calculated with
Where Tind = indicated temperature Tamb = ambient temperature of the exposed stem
D = extra immersion depth of the thermometer past the correct mark
Calculate the thermometer error for an indicated temperature of 210◦F and an ambient
temperature of 70◦F.
55
55