Agilent Signal Analyzer Guide
Agilent Signal Analyzer Guide
Special Permission Amoco Corporation and Agilent Technologies, Inc are granted
full editing, duplication and distribution privileges.
AGILENT 35670A
Supplemental Operator’s Guide
SWEPT
SINE
LogMag
6 CORRELATN
decades
ANALYSIS
HISTOGRAM
/ TIME
CHANNELS
1 2 4
REF CHANS
1 1,3
10 TIME
nVrms CAPTURE
0Hz 51.2kHz
1
AGILENT 35670A
Supplemental Operator’s Guide
2
AGILENT 35670A
Supplemental Operator’s Guide
For AGILENT 35670A analyzers with four channels, these same procedures can be
repeated for Channels 3 and 4.
3
AGILENT 35670A
Supplemental Operator’s Guide
4
AGILENT 35670A
Supplemental Operator’s Guide
5
AGILENT 35670A
Supplemental Operator’s Guide
Any frequency below 51.2 kHz may of course be entered, however the analyzer will
select the lowest span that contains the desired frequency. For example, if 1200 Hz
is entered, the analyzer is switched to a span of 1600 Hz.
6
AGILENT 35670A
Supplemental Operator’s Guide
LogMag
6
decades
1
uVrms
0Hz 800Hz
Special marker features such as peak tracking can speed analysis. To always track
the peak amplitude and frequency, follow these steps:
• Press the [Marker] hardkey.
• Press [PEAK TRK ON OFF] softkey (F6) until ON is highlighted.
The frequency and amplitude of the peak bin in every trace will be determined and
displayed.
7
AGILENT 35670A
Supplemental Operator’s Guide
The default setting for input range is CH1 AUTO RANGE. For steady-state signals
this works well because the front end will find the input range that just exceeds the
signal level, and therefore maximize the dynamic range of the measurement. When a
front end is set up to autorange and the signal either overloads or becomes too faint
for the current range setting, the analyzer automatically searches for a new range and
then re-starts the measurement. To select CH1 AUTO RANGE:
• Select the [CH1 AUTO RANGE] softkey (F6)
If a signal is not steady, but has periods of high and low levels, then CH1 AUTO
RANGE may not set the front end at the most useful range. Instead, try CH1 AUTO
UP ONLY. This causes the input range to ratchet up when an overload is detected,
however when the signal is quiet, the front end range is not reduced. This prevents
an overload if another high level signal is expected. CH1 AUTO UP ONLY works
well with transient signals such as those expected during impact testing. To set up
this condition:
• Press the [Input] hardkey.
• Select the [CH1 AUTO UP ONLY] softkey (F5).
8
AGILENT 35670A
Supplemental Operator’s Guide
If engineering units are turned on ([XDCR UNIT ON OFF]) then the range can be
specified in EU. For example, if an accelerometer is connected to channel 1 with
XDCR UNIT turned on and XDCR UNIT LABEL specified as [g], then the input
range can be specified as 4 “g” as per the following:
• Press the [Input] hardkey.
• Select the [CHANNEL 1 RANGE] softkey (F3)
• Enter 4, then press the [EU] softkey (F8).
9
AGILENT 35670A
Supplemental Operator’s Guide
LogMag
6
decades
1
uVrms
0Hz AVG: 5 800Hz
•
10
AGILENT 35670A
Supplemental Operator’s Guide
11
AGILENT 35670A
Supplemental Operator’s Guide
12
AGILENT 35670A
Supplemental Operator’s Guide
Event Triggering
The AGILENT 35670A has the ability to trigger the acquisition of a time waveform
based on the slope and level of the measured signal. To set up triggering properly,
the polarity of the measured pulse must be known in advance: will the event start
with a negative slope or positive slope? Will the event be mostly positive or mostly
negative? Use trial measurements to determine polarity if unknown.
Proper triggering is also affected by the input range. Fortunately the AGILENT
35670A has the ability to set the trigger point to be a percentage of the input range,
so even with CH1 AUTO UP ONLY, successful triggering should be possible
automatically.
To specify trigger parameters for a “positive” event on channel 1 where the trigger
point should have a positive slope, follow these steps:
• Press the [Trigger] hardkey.
• Press the [CHANNEL 1 2 3 4] softkey (F3) until 1 is highlighted.
• Press the [TRIGGER SETUP] softkey (F6)
• Press the [CHANNEL LEVEL] softkey (F1)
• Enter 5 then press the [PERCENT (%)] softkey (F4)
• Press the [SLOPE POS NEG] softkey (F5) until POS is highlighted.
Use pre-triggering to acquire some data before the trigger point. To set up a 0.01s
pre-trigger condition, follow these steps:
• Press the [Trigger] hardkey.
• Press the [TRIGGER SETUP] softkey (F6)
• Press the [ALL CHANNELS] softkey (F8)
• Press the [CHANNEL * DELAY] softkey (F9)
• Enter -.01 then press the [S] softkey (F1)
13
AGILENT 35670A
Supplemental Operator’s Guide
Real
500
mV
/div
-2.5
V
-9.7656ms 489.7461ms
Peak tracking is very useful for determining maximum amplitude of positive peaks
in the time domain:
• Press the [Marker] hardkey.
• Press [PEAK TRK ON OFF] softkey (F6) until ON is highlighted.
Care must be taken with negative peaks, however, because PEAK TRK ON only
looks for the maximum value, and large negative peaks will be ignored.
14
AGILENT 35670A
Supplemental Operator’s Guide
To set up the analyzer for a dual channel spectrum measurement, proceed as follows:
15
AGILENT 35670A
Supplemental Operator’s Guide
LogMag
6
decades
1
uVrms
0Hz 800Hz
B: CH2 Pwr Spec X:60 Hz Y:80.4887 mVrms
1
Vrms
LogMag
4
decades
100
uVrms
0Hz 800Hz
16
AGILENT 35670A
Supplemental Operator’s Guide
Use the marker knob to read the amplitude at any measured frequency for Channel 1
in Trace A.
Coupled Markers
Markers on multiple traces can be coupled so that one marker will always track the
frequency specified in another trace. This makes comparisons of level differences
across channels simple. To set up coupled markers so that Trace B will follow the
marker frequency specified in Trace A, do the following:
• Press the [Active Trace] hardkey.
• Press the [A] softkey (F1).
• Press the [Marker] hardkey.
• Toggle the [COUPLED ON OFF] softkey (F1) until ON is highlighted.
Move the knob to move the marker in Trace A. Notice that the marker in Trace B
moves to the same frequency.
17
AGILENT 35670A
Supplemental Operator’s Guide
For signals that are not steady, use CH* AUTO UP ONLY. CH1 AUTO UP ONLY
can be used for all channels to set up modal impact testing. To set up this condition:
• Press the [Input] hardkey.
• Select the [CH* AUTO UP ONLY] softkey (F5).
18
AGILENT 35670A
Supplemental Operator’s Guide
LogMag
5
decades
10
uVrms
0Hz AVG: 10 800Hz
B: CH2 Pwr Spec X:60 Hz Y:95.0385 mVrms
100
mVrms
LogMag
4
decades
10
uVrms
0Hz AVG: 10 800Hz
19
AGILENT 35670A
Supplemental Operator’s Guide
T = N (1024
. F)
For 400 lines of resolution (default) and 1600 Hz span, the time record length should
be 400/(1.024*1600) ≈ 0.25 seconds. The table showing the relationship between
lines, span, and time record length can be found in the section on Single Channel
Time Waveforms.
20
AGILENT 35670A
Supplemental Operator’s Guide
21
AGILENT 35670A
Supplemental Operator’s Guide
Then set up trigger parameters (for a positive event, positive slope) as follows:
• Press the [TRIGGER SETUP] softkey (F6)
• Press the [CHANNEL LEVEL] softkey (F1)
• Enter 5 then press the [PERCENT (%)] softkey (F4)
• Press the [SLOPE POS NEG] softkey (F5) until POS is highlighted.
Trigger Delay
Both channels should acquire data using the same pre-trigger delay. To set up a
0.01s pre-trigger condition, follow these steps:
• Press the [Trigger] hardkey.
• Press the [TRIGGER SETUP] softkey (F6)
• Press the [ALL CHANNELS] softkey (F8)
• Press the [CHANNEL * DELAY] softkey (F9)
• Enter -.01 then press the [S] softkey (F1)
Real
100
mV
/div
-500
mV
-9.7656ms 489.7461ms
B: CH2 Time X:472.168 ms Y:-241.753 mV
500
mV
Real
100
mV
/div
-500
mV
-9.7656ms 489.7461ms
22
AGILENT 35670A
Supplemental Operator’s Guide
Coupled markers can provide a quick comparison of relative levels during a transient
event at a specific time. To set up peak tracking on Channel 1 and force the marker
on Channel 2 to the time of Channel 1 peak, do the follow:
• Press the [Active Trace] hardkey.
• Press the [A] softkey (F1).
• Press the [Marker] hardkey.
• Toggle the [PEAK TRK ON OFF] softkey (F6) until ON is highlighted.
• Toggle the [COUPLED ON OFF] softkey (F1) until ON is highlighted.
Peak tracking looks for the maximum positive level in a trace. Therefore do not use
peak tracking to determine the maximum absolute amplitude of negative pulses in a
time trace.
23
AGILENT 35670A
Supplemental Operator’s Guide
The default QUAD display shows Channel 1 power spectrum in Trace A, Channel 2
power spectrum in Trace B, Channel 1 time in Trace C, and Channel 2 time in Trace
D.
Individual markers can be used on each trace, or peak tracking and coupled markers
can be used as described for dual traces.
If averaging is applied to the power spectrum measurements, then the QUAD display
will show the averaged power spectra in Trace A and B, and the previously acquired
time traces in Trace C and D.
Scale Trace: A Ref Lvl: 1
Per Div: 5 Ref Pos: Top
Date: 10-02-96 Time: 08:54:00 PM
24
AGILENT 35670A
Supplemental Operator’s Guide
25
AGILENT 35670A
Supplemental Operator’s Guide
Then tap the structure with the hammer once. A yellow overload light by the input
BNC connectors will indicate an overload, and the range will increase one step.
Continue tapping on the structure, one tap at a time, allowing time for the input range
to increase, until an overload is no longer indicated. This may take many hammer
strikes. When an overload on either channel is no longer indicated, then the ranges
are set correctly.
26
AGILENT 35670A
Supplemental Operator’s Guide
27
AGILENT 35670A
Supplemental Operator’s Guide
28
AGILENT 35670A
Supplemental Operator’s Guide
window time constant of 62.5 ms. To set up these Force/Response windows, follow
these steps:
• Press the [Window] hardkey.
• Press the [FORCE EXPO] softkey (F4).
• Press the [FORCE EXPO SETUP] softkey (F6).
• Press the [FORCE WIDTH] softkey (F6).
• Enter 36 and then press the [mS] softkey (F2).
• Press the [EXPO DECAY] softkey (F7).
• Enter 62.5 and then press the [mS] softkey (F2).
29
AGILENT 35670A
Supplemental Operator’s Guide
significantly. Ten averages may take longer, but may be necessary when testing
difficult, non-linear structures. To set up five averages, follow these steps:
• Press the [Avg] hardkey.
• Toggle the [AVERAGE ON OFF] softkey (F1) until ON is highlighted.
• Press the [NUMBER AVERAGES] softkey (F2).
• Enter 5 and press the [ENTER] softkey (F1).
To avoid averaging overloaded data, turn on overload reject as follows:
• Press the [Avg] hardkey.
• Toggle the [OVLD REJ ON OFF] softkey (F1) until ON is highlighted.
LogMag
4
decades
10
e-3
0Hz AVG: 5 3.2kHz
B: Coherence2/1 X:2.96 kHz Y:963.738 e-3
1
Real
100
e-3
/div
30
AGILENT 35670A
Supplemental Operator’s Guide
Delay = 01
. (T )
ForceWidth = T 7
ExponentialDecay = T 4
When the frequency span is halved, the Time Record length doubles, as does the
Delay time, Force Width, and Exponential Decay. All three parameters should
change with the inverse of the change in frequency span.
31
AGILENT 35670A
Supplemental Operator’s Guide
On the other hand, power spectra obtained by measuring the single-channel response
to a non-instrumented hammer strike may also provide information regarding modal
frequencies and shapes, but success is too dependent on the consistency of the
hammer strike, and therefore this technique is not recommended.
LogMag
5
decades
10
e-3
0Hz AVG: 5 3.2kHz
B: Coherence2/1 X:656 Hz Y:939.573 e-3
1
Real
100
e-3
/div
32
AGILENT 35670A
Supplemental Operator’s Guide
To set up and measure a list of 10 orders for a machinery run-up, follow these steps:
• Press the green [Preset] hardkey
• Press [DO PRESET] softkey (F1).
• Press the [Pause/Cont] hardkey if desired.
• Press the [ORDER ANALYSIS] softkey (F3).
• Press [CHANNELS 1 2 4 ] softkey (F7) until 1 is highlighted.
• Press the [Input] hardkey
• Press the [CH1 AUTO UP ONLY] software (F5).
33
AGILENT 35670A
Supplemental Operator’s Guide
34
AGILENT 35670A
Supplemental Operator’s Guide
35
AGILENT 35670A
Supplemental Operator’s Guide
For this section, the comparison data is assumed to be stored in the disk file
“COMPARE1.DAT” and “COMPARE2.DAT.” To recall these spectra into D1 and
D2, respectively, follow these steps:
• Insert the 3.5” disk containing the files “COMPARE1.DAT” and
“COMPARE2.DAT.”
• Press the [Save/Recall] hardkey.
• Press the [RECALL DATA] softkey (F4).
• Press the [RCL TRACE AND SCALE] softkey (F1).
• Press the [FROM FILE INTO D1] softkey (F1).
• Press the [CLEAR ENTRY] softkey (F9).
• Enter COMPARE1.DAT and press the [ENTER] softkey (F1).
Data from the file “COMPARE1.DAT” now appears in data register D1. Next, recall
“COMPARE2.DAT” into dataregister D2 as follows:
• Press the [Save/Recall] hardkey.
• Press the [RECALL DATA] softkey (F4).
• Press the [RCL TRACE AND SCALE] softkey (F1).
• Press the [FROM FILE INTO D2] softkey (F2).
• Press the [CLEAR ENTRY] softkey (F9).
• Enter COMPARE2.DAT and press the [ENTER] softkey (F1).
36
AGILENT 35670A
Supplemental Operator’s Guide
The data registers now contain the comparison data. To make real-time
measurements, start the measurement as follows:
• Press the yellow [Start] hardkey.
Note: Usually one, two, or three decades offers the best scale for comparison. If the
scale is not identical, but the top reference is different, then the traces can be made
identical. Assuming the maximum peak of both traces is about 8 mV, then the top
reference can be set to 10 mV by the following:
• Press the [TOP REFERENCE] softkey (F2).
• Enter 0.01 and press the [ENTER] softkey (F1).
37
AGILENT 35670A
Supplemental Operator’s Guide
Scaling for traces A and B are now identical. To accomplish the same trace scaling
for C and D, follow these steps:
• Press the [Active Trace] hardkey.
• Press the [C D] softkey (F1).
• Press the [Scale] hardkey.
• Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.
• Press the [Y PER DIV (DECADES)] softkey (F6).
• Enter 2 and press the [ENTER] softkey (F1).
As with Traces A and B, if the top reference of C and D are not identical, set the top
reference level as shown above.
38
AGILENT 35670A
Supplemental Operator’s Guide
39
AGILENT 35670A
Supplemental Operator’s Guide
The default number of waterfall steps is 15, which means only 15 spectra are kept in
memory and as new data is taken, the oldest spectra is discarded. The maximum
number of waterfall steps which can be kept in memory depends on the RAM option
installed in the AGILENT 35670A and the lines of resolution in each spectra. With
standard memory and one channel active, 130 spectra can be stored with 400 lines of
resolution. With maximum optional RAM, over 1000 spectra may be kept in
memory.
The z-axis range describes how many spectra will be displayed in the waterfall
simultaneously. The default number is 15.
On the other hand if the time record length is equal to or shorter than the specified
time interval, the measurement will be armed (and triggered, when FREE RUN
TRIGGER is selected) exactly at the time interval increment.
To acquire a waterfall with 15 spectra at 1.5 second intervals, start the measurement:
• Press the yellow [Start] hardkey.
40
AGILENT 35670A
Supplemental Operator’s Guide
41
AGILENT 35670A
Supplemental Operator’s Guide
42
AGILENT 35670A
Supplemental Operator’s Guide
When Start is pressed, spectra will be acquired as soon as the tachometer reaches
1000 RPM, and every 50 RPM thereafter until 15 spectra have been measured.
43
AGILENT 35670A
Supplemental Operator’s Guide
A: D2 Pwr Spec
1
Vrms
LogMag
4
decades
100 u
0Hz 800Hz
B: CH1 Pwr Spec Slice:122 Hz Trace:4.311 kRPM
1
Vrms
LogMag
100
uVrms
4.883
kRPM
3.383
kRPM
0Hz 800Hz
44
AGILENT 35670A
Supplemental Operator’s Guide
The best technique is to use two identical accelerometers. If the transducers are not
the same model, then it is best to determine their relative polarity.
Accelerometer Polarity
To determine the polarity of two accelerometers, follow these steps:
• Press the green [Preset] hardkey
• Press [DO PRESET] softkey (F1).
• Press the [Pause/Cont] hardkey if desired.
• Press the [Freq] hardkey
• Enter 3200 and then press the [Hz] softkey (F2).
• Press the [Input] hardkey.
• Press the [ALL CHANNELS] softkey (F2).
• Select the [CH* AUTO UP ONLY] softkey (F5).
45
AGILENT 35670A
Supplemental Operator’s Guide
Mount the accelerometers on top of each other or very close together on a light test
structure, then:
• Press the yellow [Start] hardkey.
• Tap the structure in the area around the two accelerometers to trigger a
measurement.
• If the measurement overloaded, then tap again after the range has
increased. Tap again until a trace is not overloaded.
• Press the [Marker] hardkey.
• Toggle the [COUPLED ON OFF] softkey (F2) until ON is highlighted.
• Examine the two time traces with the coupled markers to determine if the
accelerometers are in phase or out of phase.
Note whether the accelerometers are in phase or out of phase. The function to
determine differential amplitude depends on this relationship.
46
AGILENT 35670A
Supplemental Operator’s Guide
47
AGILENT 35670A
Supplemental Operator’s Guide
Set up display
• Press the [Disp Format] hardkey.
• Press the [QUAD] softkey (F6).
After a instrument preset the QUAD display format normally places PWR SPEC
CHANNEL 1 in Trcae A and PWR SPEC CHANNEL 2 in Trace B. If this is not the
case, then follow these steps:
• Press the [Active Trace] hardkey.
• Press the [A] softkey (F1).
• Press the [Meas Data] hardkey.
• Press the [CHANNEL 1 2 3 4] softkey (F1) until 1 is highlighted.
• Press the [PWR SPEC CHANNEL 1] softkey (F3)
• Press the [Active Trace] hardkey again.
• Press the [B] softkey (F2).
• Press the [CHANNEL 1 2 3 4] softkey (F1) until 2 is highlighted.
• Press the [PWR SPEC CHANNEL 2] softkey (F3).
48
AGILENT 35670A
Supplemental Operator’s Guide
Start measurement
• Press the yellow [Start] hardkey.
The AGILENT 35670A will now measure motion of Channel 1 and 2 and show the
results in Traces A and B. The difference in motion between Channel 1 and Channel
2 is shown in Trace C. The phase angle between channels is shown in Trace D.
Disp Frmat
49
AGILENT 35670A
Supplemental Operator’s Guide
In either case the stimulus and the response measured with two channels of the
AGILENT 35670A. The stimulus is usually measured on Channel 1, and the
response is measured on Channel 2. The resulting frequency response is averaged to
reduce noise and variance, and the result is displayed in a variety of formats that will
be described in this section, including:
• Magnitude and phase
• Real and imaginary components (
• Nyquist Diagrams
The quality of the frequency response measurement is determined by how much of
the response is due to the stimulus instead of being due to noise or other undesirable,
unmeasured forces. Quality is described by coherence, which ranges from zero to
one, where zero means none of the response is due to the stimulus. A coherence of
one means that the response was entirely due to the stimulus
50
AGILENT 35670A
Supplemental Operator’s Guide
• Attach one end of stinger to shaker and one end to force transducer
• Connect amplifier output to shaker
• Connect amplifier input to AGILENT 35670A source
Connect the transducers
• Connect force transducer to Channel 1
• Connect accelerometer to Channel 2
Preset the AGILENT 35670A
• Turn on the AGILENT 35670A, or
• Press the green [Preset] hardkey
• Press the yellow [Pause/Cont] hardkey if desired.
51
AGILENT 35670A
Supplemental Operator’s Guide
Set up the AGILENT 35670A source (if amplifier has level control)
• Check the following connections:
AGILENT 35670A source connected to amplifier input
amplifier plugged in
amplifier output connected to shaker
force transducer connected to Channel 1
• Reduce amplifier level control the lowest setting
• Turn on amplifier
• Press the [Source] hardkey.
• Press the [RANDOM NOISE] softkey (F5)
• Press the [LEVEL] softkey (F2)
• Enter 0.5, then press the [Vpk] softkey (F1).
• Toggle the [SOURCE ON OFF] softkey (F1) until ON is highlighted.
52
AGILENT 35670A
Supplemental Operator’s Guide
• Turn up the level control on the shaker amplifier until the excitation is
noticeable, but does not buzz or rattle. If buzzing or rattling is heard,
then either the level is too high or the structure is loose and must be
quieted.
• Press the [MORE CHOICES] softkey (F9)
• Press the [BURST RANDOM] softkey (F2)
LogMag
3
decades
1
e-3
0Hz AVG: 10 400Hz
B: Freq Resp2/1 X:0 Hz Y:0 deg
180
deg
Phase
36
deg
/div
-180
deg
0Hz AVG: 10 400Hz
53
AGILENT 35670A
Supplemental Operator’s Guide
Imag
Nyquist
50
e-3
-1 +1
/div
-300
e-3
-366.456e-3 Real -j
AVG: 10 266.456e-3
54
AGILENT 35670A
Supplemental Operator’s Guide
55
AGILENT 35670A
Supplemental Operator’s Guide
56
AGILENT 35670A
Supplemental Operator’s Guide
57
AGILENT 35670A
Supplemental Operator’s Guide
To measure peak amplitudes, peak hold averaging is selected with enough averages
specified so that peaks can be accumulated during the entire run-up and run-down,
after which the measurement is paused and the results examined.
58
AGILENT 35670A
Supplemental Operator’s Guide
LogMag
4
decades
10
uVrms
0Hz AVG: 25 400Hz
59
AGILENT 35670A
Supplemental Operator’s Guide
If the trigger is TTL, with little noise and jitter, then trigger setup is accomplished as
follows:
• Press the [Trigger] hardkey.
• Select the [EXTERNAL TRIGGER] softkey (F2).
• Press the [TRIGGER SETUP] softkey (F6)
• Toggle the [EXT LEVEL USER TTL] softkey (F1) until TTL is
highlighted.
Other trigger parameters such as USER EXT LEVEL and EXT RANGE do not need
to be specified when using a TTL trigger.
Select the Histogram/Time measurement mode. This mode turns off the anti-alias
filter so that the actual analog signal can be examined.
• Press the [HISTOGRAM / TIME] softkey (F6).
• Press the [Meas Data] hardkey.
• Press the [UNFILTERD TIME CH 1] softkey (F6).
• Press the [Scale] hardkey.
• Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.
60
AGILENT 35670A
Supplemental Operator’s Guide
If the trigger appears noisy or erratic, try activating the 100 kHz external trigger
noise filter as follows:
• Toggle the [EXT FILTER ON OFF] softkey (F6) until ON is
highlighted.
61
AGILENT 35670A
Supplemental Operator’s Guide
62
AGILENT 35670A
Supplemental Operator’s Guide
63
AGILENT 35670A
Supplemental Operator’s Guide
64
AGILENT 35670A
Supplemental Operator’s Guide
65
AGILENT 35670A
Supplemental Operator’s Guide
66
AGILENT 35670A
Supplemental Operator’s Guide
Enter Trace A Title (if desired). Repeat for traces B, C, and D if displayed.
• Press the [Active Trace] hardkey.
• Press the [A] softkey (F1).
• Press the [Disp Format] hardkey.
• Press the [MORE] softkey (F9).
• Press the [TRACE TITLE] softkey (F7).
• Press the [CLEAR ENTRY] softkey (F9).
• Enter a description of the Trace, for example the accelerometer location,
such as “MACH 1 - Y” and then press the [ENTER] softkey (F1). Note
that trace titles can be up to 13 characters long.
67
AGILENT 35670A
Supplemental Operator’s Guide
68
AGILENT 35670A
Supplemental Operator’s Guide
69
AGILENT 35670A
Supplemental Operator’s Guide
Insert a 3.5” Disk into the AGILENT 35670A and ensure DEFAULT DISK is
INTERNAL DISK:
• Press the [Save/Recall] hardkey.
• Press the [DEFAULT DISK] softkey (F9).
• Press the [INTERNAL DISK] softkey (F3).
70
AGILENT 35670A
Supplemental Operator’s Guide
This does not turn off the analyzer, but it does stop data acquisition and reset most
measurement parameters. If the AGILENT 35670A is locked up due to a software
bug and no response can be generated by the Preset key, then a hard shutdown may
be required.
Two situations call for careful planning and precautions: the first is when
frequencies are low and data acquisition times are very long, such as very low
frequency seismic testing. An interruption may waste hours of data acquisition. The
second situation worthy of precaution is when the test involves a one time event, like
the initial run-up of large machine during a plant re-start, or like a crash test, where
the test article is destroyed. In either case the data is valuable and care should be
taken to save it to disk before the data is lost by another measurement.
Manual arming along with a Channel 1 trigger and pre-trigger delay is the preferred
technique of acquiring transient crash tests. The pre-trigger delay ensures that the
entire transient event is captured even if the trigger is a little late. Manual arming
prevents re-triggering and loss of data if the transducer encounters a post-test bounce
or hit. Of course manual arming REQUIRES that the measurement be manually
armed before the test, otherwise the measurement will not be armed and data will be
missed.
71
AGILENT 35670A
Supplemental Operator’s Guide
72
AGILENT 35670A
Supplemental Operator’s Guide
This accomplishes the equivalent of multiplying “g” by 386.4 to get “inches per
second squared” (ips^2), and dividing by (jω)^2 to integrate twice in the frequency
domain to get displacement.
73
AGILENT 35670A
Supplemental Operator’s Guide
Trace B will now convert “g/lb.” into “inch/lb.” This same technique could also be
used to convert “g” into velocity terms, or to convert between metric and English
units.
74
AGILENT 35670A
Supplemental Operator’s Guide
Section 30 : Using the Agilent Technologies Demostration Unit with the AGILENT
35670A
Several of the sections in this User’s Guide are best explained with live data using
transducers and machines. Because it is usually impractical to bring machinery into
a laboratory for training purposes, the Agilent Technologies Demonstration Unit
with the Dynamic Signals Test Disc #1 (The Red CD) provides a practical
alternative.
This section describes the AGILENT Demonstration Unit and how to use it. The
most important strating point is to MAKE SURE THE DEMONSTRATION UNIT
IS GROUNDED TO THE AGILENT 35670A! This can be done via software by
grounding Channel 1 via the Input, FRONT END CH1 SETUP. Alternatively a
BNC can be connected from the AGILENT 35670A Source output to the Demo
Unit’s notch filter.
75
AGILENT 35670A
Supplemental Operator’s Guide
The following descriptions link the Section in this User’s Guide to the appropriate
CD track. Also described herein is the recommended CD output channel connection
to the AGILENT 35670A. Finally any setup clues will be described to help utilize
the signals on the CD.
76
AGILENT 35670A
Supplemental Operator’s Guide
77
AGILENT 35670A
Supplemental Operator’s Guide
78