Supplementary information
Nature | www.nature.com/nature
Supplementary Information, Quantum-enhanced nonlinear
microscopy
Caxtere A. Casacio, Lars S. Madsen, Alex Terrasson, Muhammad Waleed, Kai Barnscheidt, Boris
Hage, Michael A. Taylor, and Warwick P. Bowen.
This Supplementary Information provides details about the theory, modelling, and experimental
techniques used in the paper “Quantum-enhanced nonlinear microscopy”.
1 Microscope design
The microscope has an inverted configuration. It is home-built to allow both conventional bright-
field imaging and quantum-enhanced stimulated Raman imaging (see Fig. 1a in the main text).
Near-infrared wavelengths are chosen to minimise laser absorption and photodamage in biological
specimens [1, 2]. We employ custom high numerical aperture water immersion microscope objec-
tives to both focus and collect the light. These objectives were made by Leica Microsystems. They
were based on Leica HC-PL-APO63x/1.20W objectives, but with custom anti-reflection coating
at 1064nm. The choice of high numerical aperture water-immersion objectives ensured tight fo-
cussing of the laser fields, and therefore high intensities and spatial resolution. The anti-reflection
coatings allowed ultrahigh Stokes transmission of >92%, minimising any degradation in quantum
correlations due to microscope losses. Compared to typical high quality objectives with ∼ 65%
transmission efficiency, the high transmission of our objectives also increases the number of col-
lected Raman scattered photons by 42%, with a commensurate increase in signal strength. To
ensure that the high transmission is maintained in both dry and aqueous media we under-fill the
objectives (see Section 6.1).
The objectives have a relatively low ∼32% transmission efficiency for pump light. However, the
pump intensity is constrained by the photodamage it causes to the sample rather than the laser
source (see Fig. 2 of the main text), and is not sent to the detector (see Fig. 1a of the main text)
so this attenuation of the pump has no consequence for the achievable signal-to-noise.
The objective focus is controlled using a Mad City Labs Nano-F200S focusing module. Samples are
positioned with nanometer scale precision using a Mad City Labs Nano-Bio100 stage, which was
used to perform raster scanning in imaging experiments. Samples are simultaneously illuminated
with incoherent light from a green LED and imaged using a camera (Allied Vision Manta).
The Stokes laser is provided by a low noise solid-state pulsed laser with 6 ps pulse length, an
80 MHz repetition rate, and a wavelength of λStokes = 1064 nm (Coherent, Plecter Duo). Part of
the laser output is frequency-doubled to 532 nm and used to pump both the home-built squeezer
(described below) and a commercial optical parametric oscillator (OPO; APE, Levante Emerald).
1
The OPO generates a 500 to 700 mW pump laser that is temporally-synchronised and frequency-
tuneable over a wavelength range from 750 to 900 nm. This provides the capacity to probe Raman
shifts over the range 1713–3935 cm−1 .
The pulse duration of the laser source was chosen to be comparable to the decay time of molecular
vibrations since this maximises the Raman interaction, with molecular decay times typically in the
range of a few picoseconds [3]. The choice of 80 MHz repetition rate is designed to approximately
minimise photodamage processes [1]. Lower repetition rates lead to higher peak intensity for a
given average intensity, thereby increasing the relative contribution of nonlinear damage processes
like ablation, while higher repetition rate increases the relative contribution of linear damage for
instance due to heating.
Quantum correlations are generated between photons in the Stokes laser by passing the beam
through a home-built squeezed light source, described below.
Many factors can prevent precision optics experiments from operating at the shot-noise floor; most
particularly technical laser noise and photodetector electronic noise. Both of these noise sources
are especially problematic at low frequencies. To evade low frequency noise, we implement the
technique developed in Ref. [3] wherein the pump laser is strongly amplitude modulated. This
modulates the stimulated Raman process, shifting the detected signal to sidebands around the
modulation frequency, and away from the low frequency laser and electronic noise. In our case we
use a resonant electro-optic modulator that provides a high depth of modulation at 20.0083 MHz
(approximated as 20 MHz elsewhere in the paper). A custom-built photodetector allows the
detection of the sidebands with noise floor dominated by optical shot-noise (or the reduced squeezed
noise) and good clearance from both technical laser noise and photodetector electronic noise.
Using this apparatus and approach, we find that the detected Stokes light is shot-noise limited
(see Section 8.4). It is then possible to estimate the relative intensity noise ∆I/I of our stimulated
Raman microscope without squeezing. We find ∆I/I = (2h̄c/ηP λStokes τ )1/2 ∼ 10−8 , competitive
with the best state-of-the-art stimulated Raman microscopes [3, 4], where P =3 mW is the Stokes
power reaching the detector, c the speed of light, τ = 1 s is the integration time, and η = 72% the
measured quantum efficiency of the photodiode. Using squeezed light further reduces this noise.
To detect and characterise the stimulated Raman scattering, the Stokes laser is measured on a
resonant feedback photodetector (see Section 8.1) and processed with a spectrum analyser. The
average Stokes power reaching the detector is kept constant at 3 mW for all experiments, ensuring
both shot-noise limited (or squeezed when the optical parametric amplifier is on) performance and
large clearance from the electronic noise of the detector. As discussed above, the stimulated Raman
process causes the 20 MHz modulation of the pump field to be imprinted on the Stokes field, with
2
the size of the imprinted modulation proportional to the strength of Raman scattering. As such,
measurement of the 20 MHz modulation of the Stokes field provides a direct measurement of the
Raman gain at the frequency shift between the pump and Stokes fields. The Raman spectrum
(Fig. 1b of the main text) was measured in polystyrene beads by monitoring this 20 MHz signal
while adjusting the OPO to sweep the pump laser wavelength between 800 and 816 nm.
5 Sample preparation
The Raman signal is characterized using polystyrene microspheres. Polystyrene has a relatively
simple spectrum with strong Raman bands with Raman shifts in the range of 600 to 3100 cm−1 ,
accessible using our pump and Stokes lasers. The uniform composition and size of the beads
allows this Raman spectrum, and the photodamage induced by the measurement, to be accurately
and reproducibly characterised. It also allows accurate benchmarking of the performance of the
microscope.
Polystyrene bead samples were prepared from dry powder of monodisperse spherical polystyrene
beads with 3 µm diameter. These were scattered onto a glass coverslip. Vacuum grease was spread
around the edge of the coverslip as a sealant, and a second coverslip pressed on top.
For cellular imaging, dehydrated yeast cells (Saccharomyces cerevisiae) were sourced from a local
supermarket, and hydrated in distilled water mixed with sugar (approximately 0.1% by weight)
and NaCl salt (approximately 0.1%) to provide sustenance and prevent osmotic shock. The cells
were kept in this media for at least an hour to provide time to recover from dehydration. Cells
were then transferred into sample chambers by pipetting 15 µL of the cell solution onto a glass
coverslip, adding a sealant of vacuum grease around the outer edge of the coverslip, and then
pressing a second coverslip on top. The grease forms a seal around the edge while the cell solution
fills the approximately 70 µm gap between the coverslips. The sample was then left to rest for an
hour before experiments to provide time for the cells to adhere to the surface, which is necessary
for raster imaging. The health of the cell culture was verified separately by adding extra sugar
and observing the production of carbon dioxide.
6 Quantifying photodamage
The damage threshold was characterised for samples of 3 µm polystyrene beads. To find the
damage threshold, stimulated Raman signals were recorded in beads as the pump power was
gradually increased. The Raman signal generally increases with pump power below the damage
threshold, but decays after the onset of damage (with an example shown in Fig. 2a of the main
text). In all cases the damage was confirmed to have occurred by bright-field imaging. This
protocol was repeated for 110 particles with the histogram of recorded damage thresholds shown
in Fig. 2b of the main text.
We verify that photodamage has occurred using bright-field imaging which, for the example mea-
surement in Fig. 2a of the main text, shows substantial deformation of the polystyrene bead due
3
Figure S1: Bright-field images of typical photodamage. Examples of the different photo-
damage modalities that were observed for 3 µm polystyrene beads. a, Undamaged bead. b, Slight
deformation or ablation. c, Deformation of target bead and adjacent bead. d, Ablation of two
beads that were separately exposed. e, Deformation. f, Significant deformation. g, Complete
destruction. h, Ablation. i, Complete disintegration. In each image, the cross-hairs indicate the
target bead(s).
The distribution of threshold pump intensities at which photodamage occurs shown in Fig. 2b of
the main text fits well to a Gaussian with mean power at the sample of 54.6 mW and standard
deviation of 7.4 mW. The mean power corresponds to a peak pulse intensity of around 150 W/µm2
consistent with known thresholds for laser ablation [5]. The probability of photodamage is highly
sensitive to the optical intensity, as shown by the cumulative probability distribution in Fig. 2c of
the main text. For instance, when using a pump power of 30 mW none of the polystyrene beads
exhibited photodamage, while 26% were damaged at 50 mW.
To estimate the peak pump intensity Ipeak at the sample from the average pump power at the
sample P we assume that the pump pulses have a Gaussian temporal profile with full-width-half-
4
maximum tFWHM = 6 ps, along with a Gaussian transverse profile with width at the focus w0 .
Given the second of assumptions, it is straightforward to show that the peak pump intensity is
given by
Ppeak
Ipeak = , (1)
(πw02 /2)
where Ppeak is the peak pump power at the sample. The first assumption allows us to relate the
peak power to the average power as
r
ln 2 τ
Ppeak = 2 P, (2)
π tFWHM
where here τ = 1/(pulse repetition rate in MHz) = 12.5 ns is the repetition time of the pump
pulse train.
To produce squeezed light we utilise a single pass optical parametric amplifier (OPA) that consists
of a second-order nonlinear crystal which is pumped at 532 nm and seeded by the Raman Stokes
field at 1064 nm (see Fig. S2). The OPA pump is generated by frequency doubling the 1064 nm
light from the Coherent Plecter Duo laser, also used to generate the Stokes field. The nonlinear
crystal is a 7 mm long Periodically Polled Potassium Titanyl Phosphate (PPKTP) crystal (Raicol
Crystals, s/n:7-112714933). The crystal is temperature stabilised to 32◦ C to phase match the
nonlinear process. The relative phase between the seed and OPA pump is locked using a standard
Pound-Drever-Hall type locking system, with a 104.7 kHz modulation applied to the green field,
the green light transmitted through the nonlinear crystal detected and demodulated at 104.7 kHz,
and the resulting error signal applied back to the phase of the 1064 nm Stokes field. Choosing
the sign of the lock correctly allow the squeezing light source to be stably locked to deamplify the
Stokes field, leading to amplitude squeezing. The output state is a bright amplitude squeezed state
of light which exhibits pair-wise correlations between photons [2].
5
8 Detecting squeezing and performing quantum-enhanced
measurements
To detect the squeezed light, and later to implement quantum-enhanced Raman microscopy, we
choose to use a direct detection scheme rather than the usual homodyne detection. Generally
homodyne detection is preferred since the interference with the local oscillator both acts to boost
the shot-noise of the signal beam above the electronic noise of the detector, and because it provides
a straightforward and robust way of determining the shot-noise level, and therefore the level of
quantum enhancement1 .
The primary reason to choose direct detection here is that homodyne detection generally requires a
local power at least an order of magnitude higher than the signal power. Unlike most experiments
with squeezed light, in our experiments the signal power must be relatively high to reach param-
eter regimes relevant to state-of-the-art Raman microscopy and for which photodamage starts to
become apparent. The local oscillator powers required for homodyne detection would then intro-
duce significant challenges in detector design. A second reason to choose direct detection is that
it avoids the need to mode-match the signal field to another field after it has passed through the
microscope, and therefore relaxes requirements on the beam quality of the transmitted signal field.
1
By blocking the signal beam, so that the homodyne detects the noise level of the vacuum entering its signal
port.
amplification
mixer
servo
controller
Figure S2: Schematic of squeezed light source. φs : phase shift on 1064 nm Stokes beam. φ532
phase shift on 532 nm squeezer pump beam. λ1064 and λ532 wavelengths of Stokes and squeezer
pump. LPF: low pass filter. HF: high frequency. LF: low frequency.
6
+5V BF862 a) 4.7µH 50Ω b)
+5V 5kΩ
50Ω
1nF 50Ω
1kΩ BF862 715Ω
10pF
68p 50Ω
-V 250Ω IG17X250S4i 1µH -
D -
+
+
1kΩ LMH6624 V
OPA690 sig
-5V
c) 20.5kΩ
100p
50Ω
-
+
OP27 Vmon
Figure S3: Detector circuit diagram. Circuit diagram for the custom-built detector used to
detect squeezed light and perform quantum-enhanced Raman microscopy.
To minimise electronic noise and suppress the signal from the laser repetition rate while still
providing a high measurement bandwidth, we custom-build a photodetector designed for resonant
detection at frequencies around the modulation sideband frequency [6]. Resonant detection is
achieved using resonant feedback in a transimpedance amplifier design, as shown in the circuit
diagram in Fig. S3. This has several advantages over more commonly used designs, as discussed
below.
To avoid saturation due to the large photocurrents, non-resonant detection of pulsed light generally
requires a balanced photodiode configuration which subtracts the photocurrents produced by two
photodiodes. This cancels both the mean photocurrent and the transient photocurrent at the
laser repetition rate, that would otherwise be present. Using resonant feedback eliminates the
need for this subtraction by suppressing the response of the photodiode both at DC and above
the resonance frequency. This makes it well suited for high power applications that use a single
photodiode. Furthermore, compared with typical non-resonant designs based on ∼ 50 Ω resistive
loads, the ∼ 3 kΩ resonant feedback impedance converts the photocurrent generated by the detector
into a larger voltage. This reduces requirements on the analog filters and buffer amplifiers needed
to maintain signal integrity.
Alternative resonant designs use the capacitance of the photodiode paired with a serial inductance
to achieve both low noise and high sensitivity on resonance. However, since high intensity ps-pulses
create free charges on the diode faster than they can be transported, the capacitance increases as
7
Figure S4: Raw photocurrent power spectral densities. Raw power spectral density of
photocurrent from custom built resonant detector with no light on the detector (orange), and with
increasing Stokes power levels (blue). Grey: spectrum analyser noise floor. RBW: 1 MHz, VBW:
300 Hz.
the voltage across the photodiode drops. The increased capacitance after the pulse leads to a
significant shift of the resonance frequency. As this shift depends on the intensity of the pulse,
such a design would require adjustments of either the signal frequency or the bias voltage to
maintain the resonance condition at different intensity levels.
Since detection efficiency has a direct influence on the achievable quantum enhancement, a InGaAs-
PIN photodiode was chosen for the detector. This choice of material generally offers a higher
efficiency at 1064 nm than alternative IR-enhanced Si-PIN photodiodes. The higher capacitance
per area and lower maximum reverse voltage apply stronger restrictions to the electronic circuit
compared to Si photodiodes. However, these photodiode parameters are less critical for resonant
feedback detector than other resonant designs since the resonance frequency is determined predom-
inantly by the feedback. State-of-the-art stimulated Raman microscopes require photodetectors
that have high power handling to maximise the useable Stokes laser power and therefore signal-to-
noise, high bandwidth to maximise the achievable imaging frame rate, and low electronic noise [4].
As a compromise between these requirements, we choose a photodiode of 250 µm diameter, with
larger diameter allowing better power handling at the cost of slower response and higher electronic
noise.
The raw power spectra observed for different Stokes power levels in the absence of squeezing are
compared to the electronic noise floor of the detector in Fig. S4. When detecting 3 mW of Stokes
light we find that laser noise dominates electronic noise over a wide frequency band, from 9 to
8
3
0
0 1 2 3 4 5 6
P at detector (mW)
Figure S5: Stokes laser is shot-noise limited. Power scaling of Stokes laser noise variance
without squeezing, detected with our custom-built resonant feedback stimulated Raman photode-
tector. The variance is measured at the stimulated Raman modulation sideband frequency of
20.0083 MHz using a spectrum analyser with 1 kHz resolution bandwidth, matching the resolution
bandwidth used to take data in the main text. Red line: linear fit to the expected scaling for
shot-noise, excluding measurement points at high powers where the photodetector saturates.
40 MHz. The peaks in the electronic noise and Stokes noise variances at around 21 MHz are due
to the resonant design of the photodetector. The sharp peak in the Stokes variance at 80 MHz
corresponds to the repetition rate of the laser.
Throughout the paper, the noise reduction due to quantum correlations was assessed by analysing
the noise variance of the photocurrent from the resonant feedback photodetector using a spectrum
analyser. The noise was calibrated against shot-noise using the Stokes laser with squeezer switched
off as a reference, since this was known to be shot-noise limited (see below). To ensure a fair
comparison, the same Stokes power of 3 mW is used at the photodetector with both squeezer on and
off. All reported noise floor reductions are raw: that is, the electronic noise of the photodetector
and spectrum analyser is not removed. As such, the measurements represent the actual useful
improvement in noise floor that is achieved, rather than the ideal improvement that would be
possible if there was no electronic noise.
The most direct method to determine the quantum enhancement of the stimulated Raman micro-
scope would be to compare identical measurements with and without squeezing of the Stokes field.
However, the Raman scattering signal depends on the spatiotemporal modeshape of the Stokes
field, since this affects the Stokes intensity at the sample, the spatiotemporal overlap of the Stokes
field with the Raman pump, and how well matched the Stokes pulse duration is to the coherence
time of the molecular vibration. Consequently, if the comparison involved the Raman signal (for
instance, a direct comparison of signal-to-noise ratios), it would be crucial to ensure that the
spatiotemporal modeshape of the Stokes field remain unchanged when the squeezing is switched
on – any modification of the modeshape would alter the Raman signal size in a purely classical
9
way, which should not be misinterpreted as a quantum enhancement (or dehancement) due to the
presence of correlated photons. Maintaining identical spatiotemporal profiles with and without
squeezing proved not to be possible in practise due to the non-perfect spatiotemporal overlap be-
tween the 532 nm OPA pump laser and the Stokes field within the squeezing crystal. The effect of
this non-perfect overlap is explored in some detail in Section 9.2 below. The important concept,
here, is that only the component of the Stokes field that is overlapped with the OPA pump is
deamplified, with the non-overlapped component left to transmit through the crystal unaffected.
When the squeezer is switched on, the deamplification reduces the overlapped Stokes intensity by
more than 50% (see Fig. S9b) but leaves the non-overlapped component unchanged. This changes
the spatiotemporal modeshape of the Stokes field.
We found experimentally that the change in Stokes field spatiotemporal modeshape due to deam-
plification caused some modifications to the Raman signal strength. Generally, the signal strength
was degraded by these changes. This is most likely due to a reduction in peak Stokes intensity
– the Stokes field experiences maximum deamplification at the spatiotemporal peak of the OPA
pump pulse and less amplification in the wings of the pulse, this broaden it in both space and
time. We note that, even with the degradation in signal strength due to deamplification, due to
the quantum-enhancement of the noise floor it was possible to achieve higher signal-to-noise using
squeezing than without it. For instance, Fig. S6 shows an example measurement of the Raman
signal and noise floor for the CH aromatic stretch band (Raman shift of 3055 cm−1 ) in polystyrene,
with and without squeezing. We see that, while the Stokes spatiotemporal modeshape modifica-
tion due to deamplification reduces the signal strength by 10%, the improved noise floor due to
squeezing still allows a 4% improvement in the signal-to-noise ratio. Moreover, the degradation in
signal strength is not fundamental – it could be compensated for using different Stokes focussing
or applying pulse shaping techniques without compromising the quantum correlations that provide
a quantum-enhanced noise floor.
10
25
20
SRS signal [dB]
15
10
Figure S6: Stimulated Raman signal-to-noise with and without squeezing. Example
power spectral density of the detected photocurrent showing the stimulated Raman signal and
noise floor for both quantum-enhanced (blue) and shot-noise-limited (red) measurements. The
quantum-enhanced measurement has a signal-to-noise-ratio of 260, while the signal-to-noise of the
shot-noise-limited measurement is 250. The measurements were performed sequentially on the CH
aromatic stretch band (Raman shift of 3055 cm−1 ) in polystyrene. 30 mW of pump power and
3 mW of Stokes power were used. Quantum correlations suppressed the noise floor of the quantum-
enhanced measurement by 14% (-0.64 dB). When using squeezed light, classical spatiotemporal
modeshape differences degrade the Raman signal strength by 10%. Even taking this classical
degradation into account, the signal-to-noise is enhanced by 4% when using squeezed light. RBW:
1 kHz. VBW: 10 Hz.
11
into the microscope.
To verify that the Stokes laser is shot-noise limited when the squeezer switched off, as is critical
when using it for noise calibration, we measure the variance of its noise as a function of power.
This is shown in the power spectra in Fig. S4. Over the full range of frequencies where the light
noise dominates the electronic noise, the light noise was found to scale linearly with Stokes power
for powers beneath detector saturation. This linear scaling is consistent with expectations for a
shot-noise limited laser. If the laser were instead limited by technical noise, such as laser relaxation
oscillation or spontaneous emission noise, quadratic scaling would be expected.
Most important for the experiments reported here, is that prior to squeezing the Stokes light
is shot-noise limited around the stimulated Raman modulation frequency. Measurements at that
frequency are shown as a function of power in Fig. S5. Linear scaling is observed for optical powers
below 4 mW, with detector saturation occurring at higher powers. As can be seen, that data is
inconsistent with the quadratic scaling expected for technical noise. As such it may be concluded
that with the squeezer switched off the Stokes laser is shot-noise limited at the 3 mW power levels
used in our experiments.
Fig. S7 shows a typical observed raw power spectrum of the squeezed light generation in our
OPA, compared to both the electronic noise floor of the detector and the shot-noise of the Stokes
laser observed with the same 3 mW of power incident on the detector. It can be seen that the
noise on the squeezed field is suppressed compared to the shot-noise limit over a wide frequency
range. Fig, 3a, in the main text normalises the squeezing to the shot-noise level to show the quan-
tum noise reduction more clearly as a function of frequency. In this figure, and for all reported
quantum-enhancement values, electronic noise is not corrected for; that is, the quantum enhance-
ment reported is the real enhancement of the measurement, not the enhancement that would be
achieved if the detector was replaced with an ideal detector with no electronic noise. As can be
seen in that figure, suppression of the noise beneath the shot-noise level is achieved for frequencies
between 5 MHz and 52 MHz, though it is degraded at the extremes due to the electronic noise of
the detector. As such, the quantum enhancement we observe could be applied in scenarios with
sub-microsecond pixel dwell times, allowing video rate stimulated Raman imaging.
12
Noise Variance [a.u.]
Figure S7: Raw photocurrent power spectral densities using squeezed and shot-noise
limited light. Raw power spectral density of photocurrent from custom built resonant detector
with no light on the detector (grey), 3 mW of shot-noise limited light on the detector (light blue),
and 3 mW of amplitude squeezed light on the detector (dark blue). RBW: 1 MHz. VBW: 10 Hz.
Averaged thirty five times.
13
9.2 Modelling the squeezing process
As described above and in the main text, to produce squeezed light we utilise a single pass optical
parametric amplifier. The simplicity of a single-pass configuration makes it technically desirable
from alignment, construction and control perspectives. It is possible here, since the use of pi-
cosecond pulses increases the peak intensity of the light, and therefore the nonlinear interaction,
making resonant enhancement in a cavity redundant. However, the lack of an optical cavity also
introduces the possibility of multi-spatial mode effects which could potentially deleteriously effect
the level of squeezing generated. To investigate this, and confirm that the squeezing generation is
well understood, we develop consider a simple model of the squeezing process, taking account of
non-perfect spatial overlap between the OPA pump field and the Stokes field.
The concept of the model is shown in Fig. S8. Fig. S8a shows a rough schematic of the squeezer
with OPA pump and Stokes partially misaligned. Only the spatially overlapped component of
the Stoke laser will experience amplification due to the pump. The non-overlapped component
will pass through the nonlinear crystal unaffected. Fig. S8b shows a basic model of this, with
a fraction ηoverlap of the Stokes field perfectly overlapped with the OPA pump, and a fraction
1 − ηoverlap bypassing the pump entirely. Both component of the field then propagate towards the
photodetector, and reach it with the same transmission efficiency ηdet (including the inefficiency
of the photodetector itself). They are then detected together on the photodetector.
a)
b) ηdet
ηoverlap
Figure S8: Model of squeezing process. a, Schematic showing that the pump (shaded in green)
and seed (shaded in red) fields incident on the nonlinear crystal that performs optical parametric
amplification are not perfectly overlapped. The red field is detected directly on a photodiode. b,
The non-perfect overlap can be modelled by treating a fraction of the seed light (given by ηoverlap )
as perfectly overlapped, and a fraction as bypassing the nonlinear crystal entirely. Since the direct
detection detects both components of the seed field, and these components are orthogonal, the
photocurrents they create can be added together to determine the total detected photocurrent.
Detection inefficiency is modelled by including beam splitters into each of the two components,
both with transmission efficiency given by ηdet .
14
The amplitude X̂1 and phase Ŷ1 quadratures of the component of the detected light that does not
experience squeezing are given by
√ p √ p
X̂1 = ηdet 1 − ηoverlap X̂in − ηoverlap X̂v1 + 1 − ηdet X̂v2 (4)
√ p √ p
Ŷ1 = ηdet 1 − ηoverlap Ŷin − ηoverlap Ŷv1 + 1 − ηdet Ŷv2 , (5)
where X̂in and Ŷin are the amplitude and phase quadratures of the input Stokes light, and the sub-
scripts v1 and v2 label the vacuum noise entering due to poor overlap and transmission inefficiency,
respectively.
Similarly, the amplitude X̂2 and phase Ŷ2 quadratures of the component of the detected light that
is perfectly overlapped to experience squeezing are given by
p √ p p
X̂2 = Gηdet ηoverlap X̂in + 1 − ηoverlap X̂v1 + 1 − ηdet X̂v3 (6)
r
ηdet √ p p
Ŷ2 = ηoverlap Ŷin + 1 − ηoverlap Ŷv1 + 1 − ηdet Ŷv3 , (7)
G
where the subscript v3 labels the vacuum noise entering due to transmission inefficiency for this
field, and G is the level of squeezing of the variance of the perfectly overlapped squeezed light,
prior to detection inefficiencies. Here, G < 1 for amplitude squeezing.
Since the perfectly overlapped and non-overlapped fields are spatially orthogonal, when they are
detected together on the photodiode they do not interfere, so that the resulting photocurrent is
simply the sum of the photocurrents that would be measured if each was detected independently:
where the a’s are annihilation operators for the fields, with X̂ = a + a† and Ŷ = i(a† − a), the
coherent amplitude is defined as α = hai and is defined here to be real without loss of generality,
and the noise operators are defined as δ X̂ = X̂ − hX̂i so that hδ X̂i = 0. Here, to get to Eq. (9)
we have used the usual linearisation approximation, assuming that {α1 , α2 } 1 so that noise
operator product terms can be neglected.
The squeezer acts to both deamplify the coherent amplitude of the Stokes field and decrease its
noise variance. The deamplification is given by
hii α2 + α2
D= = 2 1 2 2 . (10)
hiiG=1 α1 + α2,G=1
15
squeezed component in the absence of the non-overlapped field. We have also used the fact that
when the squeezer is switched off (G = 1) the overlapped field is shot-noise limited so that V2 = 1.
To determine the effect of mode mismatch and detection inefficiency on the observed amplification
and squeezing, we therefore need to determine α1 , α2 and V2 . From Eqs. (4) and (6), the detected
coherent amplitudes of the non-overlapped and perfectly overlapped fields when detected can be
shown to be
q
α1 = ηdet (1 − ηoverlap )αin (12)
p
α2 = Gηdet ηoverlap αin , (13)
where we have appropriately taken the input field and vacuum noise inputs to be shot-noise limited
2 2 2
(hδ X̂in i = hδ X̂v1 i = hδ X̂v3 i = 1) and uncorrelated with each other.
and
1 − ηoverlap + ηoverlap G2
V = ηdet + 1 − ηdet . (16)
1 − ηoverlap + ηoverlap G
As expected, we see that the deamplification is independent of detection efficiency - the detection
inefficiency affects the coherent amplitude in exactly the same way independent of whether the
light is squeezed or not. As G → 0 the deamplification asymptotes to 1 − ηoverlap . This makes
sense, since in that limit the spatially overlapped fraction of the light is perfectly deamplified,
while the non-overlapped component remains unaffected.
The predicted noise variance V exhibits behaviour unexpected from a normal single-mode squeezer.
Instead of the variance decreasing monotonically as G → 0 from above, it decreases below the shot-
noise limit and then rises again to exactly the shot-noise limit (V = 1) at G = 0. This can be
understood since, as discussed above, in this limit the coherent amplitude of the squeezed com-
ponent goes to zero. Since the magnitude of the detected photocurrent scales with the coherent
amplitude, the squeezed-light signal in the photocurrent is fully suppressed, only leaving the signal
from the shot-noise limited field that has no overlap. We note that this deleterious effect could be
avoided in future quantum-enhanced nonlinear microscopes by using a homodyne configuration,
and carefully mode matching the local oscillator field to only the squeezed component of the Stokes
field. Improvements in the observed squeezing could be obtained not only by controlling the trans-
verse spatial modeshape of the local oscillator, but also by engineering its temporal modeshape.
The pulsing of the parametric amplifier pump field leads to a time dependent parametric gain, with
the highest deamplification occurring at the peak of the pulse. As a consequence, the squeezing
of the Stokes field can be expected to be strongest near the peak of the pulse. An appropriately
temporally tailored local oscillator could extract out this highly squeezed component of the Stokes
field, lowering the measurement noise floor.
16
It is straightforward to calculate the gain at which the squeezing is maximised as well as the
maximum squeezing. Doing this, we find
p
1 − ηoverlap p
Gmin V = 1 − 1 − ηoverlap (17)
ηoverlap
p
2 1 − ηoverlap p
Vmin = 2Gmin V = 1 − 1 − ηoverlap . (18)
ηoverlap
Assuming no pump depletion, which is reasonable in our case, the seed light (the input Stokes
field) is exponentially amplified as it propagates through the OPA crystal to produce the gain G
on the final output amplitude quadrature (and 1/G on the phase quadrature). The gain can then
be expressed as G = e−κL where κ is the gain per unit length and L is the length of the crystal.
Recognising √ that the nonlinear interaction is proportional to the amplitude of the OPA pump
√ field,
−g P
so that κ ∝ P where P is the OPA pump power, this can be re-expressed as G = e , where
g is a gain coefficient that depends on the length, phase matching and nonlinear coefficient of the
crystal, as well as the optical focussing within the crystal. With this relationship it is possible to
compare the predictions of the model to experiment.
To compare model and experiment we measure both the deamplification and squeezing as the pump
power is varied from 0 to 70 mW. The results are shown in Fig. S9. We fit the deamplification
to Eq. (15) extracting the overlap efficiency ηo = 0.85 ± 0.02 and gain coefficient g = 8.8 ±
0.4 W−1/2 . Fixing these parameters and fitting the squeezed variance to Eq. (16) we determine the
detection efficiency in this case (without passing through the microscope) to be ηd = 0.65 ± 0.01,
consistent with expectations given the measured 72% efficiency of the detector and inefficiencies
in transmission from the nonlinear crystal to the detector. As can be seen from Fig. S9, the data
agrees well with the theory2 , validating this simple model. The agreement also provides further
validations that, prior to the squeezing process, the Stokes field is shot-noise limited and that our
method of determining the quantum enhancement is accurate.
As can be seen in Fig. S9 the variance of the detecter squeezing is optimised and relatively insen-
sitive to OPA pump power in the power range between 35 and 65 mW. We therefore operate our
experiments in this regime.
Optical inefficiencies reduce quantum correlations between photons and therefore degrade the level
of quantum enhancement that is possible using squeezed light. The path between the nonlinear
crystal generating squeezing and eventual photodetection, included a range of optical elements
2
We note that the variance data disagrees with more basic theory that does not account for the effect of
deamplification of the coherent amplitude of the overlapped field on the detected variance.
17
(a)
(b)
Figure S9: Validating the model and characterizing the performance of the squeezer.
(a) Measured reduction in the intensity of the seed field after the OPA as a function of OPA pump
power. Line: least squares fit to data with the overlap efficiency ηoverlap and squeezer gain g as
fitting parameters. The fit yielded values of ηoverlap = 0.85 ± 0.02 and g = 8.8 ± 0.4 W−1/2 . (b)
Observed squeezed variance at 20 MHz as a function of OPA pump power. Line: least squares fit to
data with only the detection efficiency ηdet as a free parameter, ηoverlap = 0.85, and g = 8.8 W−1/2 .
The fit yielded ηdet = 0.65 ± 0.01. snu: shot-noise units.
18
which introduce loss, as shown schematically in Fig. 1a of the main text. The microscope itself
was a significant source of loss, in the objectives, in dichroic mirrors, and due to reflections from the
microscope coverslips. The total efficiency of transmission through the microscope was measured
to be ηmicroscope = 0.75. The mirrors, lenses and dichroic mirrors between the nonlinear crystal and
microscope and between the microscope and photodetector were measured to have transmission
of ηpropagation = 0.82. The photodetector efficiency was determined by measuring photocurrent i
produced by PStokes = 3 mW of Stokes light and comparing the flux of electrons this corresponds
to to the flux of photons arriving at the detector. The measured photocurrent of i = 1.86 mA
resulted in a detector quantum efficiency of
h̄ω i
ηdetector = × = 0.72, (19)
PStokes e
where e is the charge on an electron, h̄ is the reduced Plancks constant, and ω = 2πc/λ is the
angular optical frequency, with c the speed of light and λ = 1064 nm the Stokes wavelength.
Combined, this gives a total efficiency of transmission of the squeezed light from the nonlinear
crystal to the measured photocurrent of
η = ηmicroscope ηpropagation ηdetector = 0.44. (20)
It is interesting to ask whether the measured squeezing after passing the squeezed field through
the microscope is consistent with this measured total efficiency. The theoretical squeezed variance
after
after transmission with an efficiency of η is Vsqz = ηVsqz +1−η, where Vsqz is the squeezed variance
directly after the OPA crystal. Vsqz can be estimated from the measured squeezed variance prior to
prior prior
the microscope Vsqz using then relation Vsqz = ηdet Vsqz + 1 − ηdet , with the efficiency ηdet = 0.65
prior
determined in Section 9.2. Under ideal operating conditions, Vsqz = 0.72 (or -1.4 dB), as shown
in Fig. S9. However, the operating conditions of the squeezed light source vary from experiment
to experiment due to variations in the overlap between the OPA pump and the input Stokes field,
and variation in the performance of the phase lock. This results in degraded performance in some
prior
experiments (for instance, see Fig. 3a in the main text, for which Vsqz = 1−0.22 = 0.78). Overall,
prior prior
we find that from experiment to experiment Vsqz varies in the range Vsqz = 0.76 ± 0.04. From
this, we find the range of variances for the amplitude squeezing directly after the OPA crystal of
Vsqz = 0.63 ± 0.06. After transmission through the microscope apparatus and photodetection with
after
a combined efficiency of η, the predicted detected squeezing is Vsqz = 0.84 ± 0.03. This is broadly
consistent with our measurements after the microscope, with squeezed variance in Fig. 3b of the
main text equal to 0.87, and that in the quantum image of polystyrene beads in Fig. 4a of the
main text equal to 0.81. In the case of live cell imaging, significant efforts were made to improve
the alignment of the system, including improvements in the overlap between pump and seed in
the optical parametric oscillator generating squeezing, of the alignment of pump and Stokes in the
Raman microscope, and of the microscope itself. Fresnel reflective losses were also lower at the
water-glass interfaces between sample and coverslips, than at the air-glass interfaces when imaging
dry polystyrene samples. This resulted in an improved squeezed variance of 0.74 (Fig. 4b).
The lower limit to detectable concentration is reached where SNR=1. Since the SNR increases
with averaging time, we use normalized units, equivalent to the sensitivity achievable for the
19
30
25
15
10
Polystyrene is a polymer formed from the monomer styrene, and does not have a well-defined
molecular weight or number of aromatic rings. However every aromatic ring corresponds to one
styrene monomer. Using the density of polystyrene, 1060 kg/m3 , and the molecular weight of
styrene, 104.15 g/mol, we therefore find the polystyrene concentration to be C = 10.2 M. Note
that the unit M corresponds to mol/L.
Optimisation of the microscope for cell imaging also allowed significant improvements in the signal-
to-noise of our polystyrene measurements. The highest damage-free shot-noise-limited signal-to-
noise that we determine for the CH aromatic stretch bonds of polystyrene is 1510. This was
recorded with 1 kHz resolution bandwidth which corresponds to an effective measurement time of
τ = 1 ms. Extrapolating to SNR=1 with 1 Hz resolution bandwidth, the minimum concentration
sensitivity for styrene is estimated as Cmin = 8.3 mM Hz−1/2 . The quantum enhanced SNR was
20
1730 (see Fig. S10), also recorded with 1 kHz resolution bandwidth. Similarly extrapolating this
to SNR=1 provides an estimated minimum concentration sensitivity of Cmin = 7.8 mM Hz−1/2 ,
enhanced by around 7% due to quantum correlations.
The sensitivity we achieve is broadly consistent with high performance stimulated Raman micro-
scopes reported in the literature. Refs. [8] and [9] both perform stimulated Raman microscopy on
the same CH aromatic stretch bond in polystyrene that we probe, allowing a quantitative compar-
ison. A fair comparison requires that differences in pump and Stokes power are taken into account.
In our experiments, the minimum detectable concentration scales linearly with pump power and as
the square root of Stokes power, which were 30 mW and 3 mW respectively. As such our minimum
resolvable concentration can be normalized to Cmin = 400 mM Hz−1/2 mW−3/2 .
Ref. [8] reports a shot-noise-limited stimulated Raman microscope using a two-color scheme to
improve signal-to-noise by subtracting common mode noise on a balanced detector. A maximum
signal-to-noise of 106 is reported, using 28 mW of Stokes power, 14 mW of pump power and a
total integration time of 0.1 ms. They detect the pump, rather than the Stokes, so that the power
dependences of the signal-to-noise are reversed compared to our experiments. Two alternative
definitions of signal-to-noise are common in the literature, the ratio of signal power to noise power
used in our work, and the ratio of signal amplitude to noise standard deviation. Ref. [8] does
not specify which definition they use. Assuming the same definition as we use, we infer from the
numbers above a minimum resolvable concentration of around 1040 mM Hz−1/2 mW−3/2 . With the
alternative definition, the minimum resolvable concentration is around 101 mM Hz−1/2 mW−3/2 .
Either way, the sensitivity is similar to our best reported sensitivity, with one definition of signal-
to-noise yielding a factor of 2.6 higher sensitivity, and the other a factor of 4 lower.
Ref. [9] makes measurements using both simultaneous Raman gain and stimulated Raman loss
detection modes, using a 0.25 ms integration time. Of these, the most direct comparison to our
experiments comes from the simultaneous Raman gain measurements, since this is the detection
mode we use. For these measurements, they use 10 mW of power for both pump and Stokes.
They record an amplitude signal-to-noise of 23.1 which corresponds to a power signal-to-noise of
23.12 = 534. Using these numbers we infer a concentration sensitivity of 220 mM Hz−1/2 mW−3/2 ,
around a factor of 1.8 lower than our best reported sensitivity. For their stimulated Raman loss
measurements, they use 8 mW of Stokes power and 6 mW of pump power, and report an amplitude
signal-to-noise of 27.0 (power signal-to-noise of 27.02 = 730). From these numbers we infer an
improved concentration sensitivity of 140 mM Hz−1/2 mW−3/2 . This is a factor of 2.9 lower than
our experiments, though not directly comparable since the measurement uses stimulated Raman
loss rather than gain.
The above comparisons show that the concentration sensitivity from our experiments is similar to
those reported for the same bond in polystyrene in other stimulated Raman microscopes. Apart
from the modest 7% improvement provided by quantum correlations in our case, the variations
between experiments most likely arise primarily from the difference in pulse durations – both
Refs. [8] and [9] use few-hundred femtosecond pulses compared to our picosecond pulses. Differences
in collection efficiency and optical focussing into the microscope can also be expected to contribute
to the variation, though to a lesser degree.
We would note that the concentration sensitivity that can be achieved depends strongly on the
particular molecular vibration that is probed. For example, a state-of-the-art stimulated Ra-
21
man microscope was reported to allow measurement of CH bonds in dimethyl sulfoxide (DMSO)
with a concentration limit of 28 mM with 2 µs time constant, and with pump and Stokes pow-
ers of 72 mW and 20 mW [10, 11]. This corresponds to a minimum concentration sensitivity
of 12.8 mM Hz−1/2 mW−3/2 , a factor of approximately thirty times lower than our system.
Other bonds with especially high Raman cross-sections allow even better concentration sensi-
tivity. For instance, sensitivities in the tens of micromolar range have be achieved for the alkyne
bond (C≡C) [11, 12, 13, 14].
12 Imaging
The quantum-enhanced images are formed by recording the stimulated Raman signal on the spec-
trum analyser while raster scanning the sample in 100 nm steps at a rate of 20 steps per second,
returning a 100×100 pixel image in 8.3 minutes. The stage position is recorded throughout the
raster scan and is used to assign positions to the measured Raman values. After recording the
stimulated Raman signal at each position, the shot-noise is also measured at each of the same
locations by blocking the pump field, keeping the Stokes power at the detector at 3 mW, and
recording shot-noise across a similar raster scan. The Raman signal is normalized by dividing by
the shot-noise level at the same locations, which corrects for any change in signal strength due to
attenuation of the detected Stokes field by the sample.
As described in the main text, our images of yeast cells show three key features, a relatively uniform
signal across the entire cell that likely originates from cytosol, five bright roughly-ellipsoidal shapes,
and a faint sub-200 nanometre-thick semicircular feature. The Raman shift of 3055 cm−1 used to
generate the images was chosen to selectively target lipids. The five roughly-ellipsoidal shapes
seen in Fig. 4b of the main text are likely to be lipid droplets, since multiple lipid droplets of this
shape and size are commonly observed in yeast cells (see e.g. Refs. [15, 16, 17]). The semicircular
feature is consistent in shape with a section of the cell membrane or wall. While the cell membrane
consists of lipids and the cell wall has a significant fraction of lipid [18], their tens-of-nanometre
thickness results in far weaker Raman signals than large structures such as lipid droplets, consistent
with what we observe. The weakness of the Raman signal could explain why we only observe a
semicircular feature rather than a full circle enclosing the cell contents, with the Raman signal
sinking beneath the optical noise (even though it is reduced using quantum correlations) across
the majority of the structure. Measurements that only partially resolve the cell membrane are
not uncommon in Raman imaging. For examples, see Ref. [19] where the Raman signal varies
strongly at different parts of the membrane, and Ref. [20] where it is only partially resolved. It
is also conceivable that this feature arises from some form of aberration, distortion, scattering or
transmission effect.
The straight horizontal line of high Raman signal visible in Fig. 4b is an artefact that we attribute
to optical trapping of a high-Raman-contrast particulate by the Raman pump laser. The particu-
late was then held stationary for some time within the optical excitation area as the sample stage
22
and cell were scanned, before becoming untrapped prior to a single line scan being completed. Our
scanning system was arranged to scan horizontally and then step vertically, consistent with this
hypothesis.
It is interesting to compare the cell images generated by our quantum-enhanced microscope (shown
in Fig. 4b of the main text) to previously reported yeast cell images using Raman microscopy. Here,
we compare to images previously reported using both coherent Raman microscopy [21, 16, 22] and
spontaneous Raman microscopy [15]. The shortest reported pixel dwell time in these experiments
is 50 ms [21], the same as was used in our experiments. In this case fourteen Raman lines were
probed simultaneously, in contrast to our single line, and the peak signal-to-noise was around 5,
compared to our peak signal-to-noise of 100. Ref. [16] does not report a pixel dwell time, but also
has a signal-to-noise of around 5. The other two experiments have pixel dwell times in the range
of 0.1-0.5 seconds, significantly longer than ours. We were unable to extract the signal-to-noise in
these cases.
In addition to quantifying the optical damage on polystyrene beads, we were able to observe
and illustrate light-induced damage on the yeast cells. Photodamage appeared for intensities of
470 W/µm2 . We document in Figure S11g a case of photodamage occurring during a scan. The
image was scanned in horizontal scans, stepping vertically between scans and starting an the
bottom of the image. The signal amplitude increased considerably after the appearance of visible
damage. The Raman signal increased by up to 40 dB and debris with high Raman signatures were
dragged outside of the cell by the horizontal motion of the beam. We also present in Figure S11
widefield images of damaged cells. The damage was often localized and associated with shrinking.
This suggests that the light may have caused membrane destruction followed by diffusion of the
organelles into the extracellular medium.
Figure S12 compares a quantum enhanced cell image (Fig. S12a), to the equivalent shot-noise
limited image obtained by artificially raising the noise floor to the shot noise floor (Fig. S12b), and
to an experimental shot noise limited image (Fig. S12c). The removal of quantum correlated light
23
a b c
d e f
2um
Figure S11: Bright-field images and Raman signal image of damaged yeast cells. Photo-
damage induced by the pump field on yeast cells with 30 mW of pump power at the sample and an
effective microscope objective numerical aperture of NAeffective ∼ 1.2, corresponding to an intensity
of around 470 W/µm2 . a to f: widefield images of damaged cells. g: Photodamage induced during
simulated Raman imaging. Left: widefield images before (top) and after (bottom) damage. Right:
Raman image showing damage occurring mid-scan.
24
in Fig. S12b&c results in an evident reduction in contrast. The experimental shot noise limited
image was taken 30 minutes after the quantum-enhanced image. The sample was found to drift
axially over these timescales. We attribute the degradation in resolution observable in Fig. S12c
to this drift, rather than the removal of quantum correlations. Fig. S12d plots the Raman signal
along the dashed cross sections on Figs. S12a&b, where the semi-circular feature that may be
the membrane is well resolved both for the quantum enhanced and shot-noise limited equivalent
images. Fig. S12e is the same plot in a region of the membrane-like feature where only the quantum
enhanced image resolves the feature with a signal-to-noise ratio higher than 1.
One way to illustrate the advantage provided by quantum correlations is to analyse the images
for features that are not resolvable without these correlations. As can be seen by comparing
the images in Figure S12a,b & c, the semicircular feature located vertically above main body of
cell (consistent with a segment of the cell membrane or wall) is significantly better resolved using
quantum correlations than without them. To quantify this improvement we estimate the resolvable
fraction of this feature for Figures S12 a and b3 . This is done by taking radial cross-sections of the
images through the feature, and interpolating the Raman signal along these cross-sections. The
cross-sections show a small bump at the feature (see Fig. S12d &e), which we define as resolvable
if its maximum is more than a factor of two higher than the noise floor (equalling a signal-to-noise
of 1). This yields an angular range for which the signal from the feature is resolvable for each
of the quantum-enhanced and shot-noise limited images. From this analysis, we found that the
quantum-enhanced image allows a 40% greater length of the feature to be resolved than the shot-
noise limited image. Assuming the feature to be roughly circular as expected for the cell membrane
or wall, this corresponds to 23% of the total feature, compared to 16% for the shot-noise limited
image (shown angularly by the double-headed angles in Figs. S12a & b).
It is interesting to ask what avenues might be available, other than quantum correlations, to evade
photodamage in biological imaging. An obvious approach is to introduce contrast enhancing labels
such as in Ref. [23]. However, this comes at the risk of the contrast label compromising the natural
behaviour of the specimen. An alternative would be to increase the pixel dwell time, allowing
longer acquisitions and therefore higher signal-to-noise for a fixed illumination intensity. In this
case, the signal-to-noise scales linearly with pixel dwell time (assuming that the pulse duration,
peak intensity and repetition rate are fixed), so that a 35% enhancement in signal-to-noise, such as
is demonstrated here, could be obtained using a 35% longer measurement. This assumes, however,
that the photodamage is due to the intensity of the illumination light. This is not always the case.
For instance, the buildup of reactive oxygen species due to optical illumination is one key cause
of photodamage. This build-up typically depends on the total irradiation energy rather than the
intensity, and so cannot be avoided by performing a longer but lower-intensity measurement.
25
a b c
Membrane Membrane
SNR=1
SNR=1
Noise floor
Noise floor
d
e
SNR=1
SNR=1
Noise floor
Noise floor
Membrane Membrane
Figure S12: Comparison of quantum-enhanced and shot-noise limited images. a,b & c
respectively show the quantum-enhanced image, shot-noise limited image produced by artificially
raising the background of the quantum enhanced image by 1.3 dB, and the shot-noise limited
image. An interpolation of the Raman signal is performed radially from the center of the cell to
quantify the resolvable fraction of the membrane-like feature. d Raman signals along the dashed
lines in a & b, corresponding to the cross-section with maximum signal-to-noise from the feature.
Here we see SRS signals are approximately 7 and 6 dB higher than the noise floor, respectively,
for the quantum-enhanced and shot-noise limited images. e Raman signals along the solid lines in
a & b, corresponding to cross-section with degraded feature signal-to-noise. Here the SRS signals
are approximately 3.1 and 2.5 dB higher than the noise floors, respectively, for the quantum-
enhanced and shot-noise limited images. In this case, only the quantum-enhanced measurements
has a signal-to-noise above one. In d&e: insets are magnified parts of the Raman images, showing
the membrane-like feature; grey shading shows the location of the membrane-like feature; upper
dashed horizontal lines show a signal-to-noise of one; and lower dashed horizontal lines show the
noise floor normalised to the shot-noise level.
26
this is especially true, since it is in principle possible to perform parallel multispectral imaging [23]
to extract very reach chemical information about the specimens under study, but this requires
multiple measurements at different Raman shifts and often requires probing of molecular bonds
with far weaker Raman cross sections than C-H bonds. As such, compromises that involve increased
measurement duration are not typically preferable.
The absolute quantum advantage we report is defined relative to classical experiments conducted
with the same spatiotemporal modeshapes and apparatus. It is also interesting to consider what
avenues might be available to improve signal-to-noise without quantum correlations, were these
conditions to be relaxed.
An obvious path to improvements, both for shot-noise limited and quantum-enhanced Raman mi-
croscopy would be to improve the efficiency of the microscope, as well as of the optical components
after it (and in the case of quantum-light, also between the squeezed light source and the micro-
scope). In our case, the total efficiency for squeezed light was around 40%, and somewhat higher
for the transmission fo the probe light through the microscope. As such, improvements in signal-
to-noise from enhanced efficiency are possible, but constrained (without quantum-correlations,
changing the pulse focussing/width, or increasing the light intensity at the sample) to a maximum
of around a factor of two. Improvements using quantum-correlation do not suffer this constraint,
in principle allowing arbitrarily high signal-to-noise. For example, a factor of thirty enhance-
ment has been demonstrated using continuous wave technology developed for gravitational wave
detection [24].
The signal-to-noise could also potentially be improved through careful tuning of the pulse widths
of the pump and Stokes fields. As discussed already in the main text and in Section 2 of the
Supplementary Information, these widths should be roughly matched to the decay time of the
Raman transition that is being probed. In biology, these widths tend to be in the picosecond
range, matching the pulse widths we have chosen for SRS system. Indeed, we based our system
design on previous literature on precision stimulated Raman microscopes (see e.g. Ref. [25]), taking
care to choose appropriate pulse width and repetition rate to give a strong Raman signal, while
also aiming to minimise photodamage (i.e. ultimately to maximise the signal-to-noise ratio). We
believe that the signal-to-noise of our system is close to optimised, in this respect. Nevertheless,
the exact optimal pulse width and repetition rate will vary from specimen to specimen and between
Raman transitions within a given specimen.
References
[1] Yan Fu, Haifeng Wang, Riyi Shi, and Ji-Xin Cheng. Characterization of photodamage in
coherent anti-stokes Raman scattering microscopy. Optics Express, 14(9):3942–3951, 2006.
[2] Michael A Taylor and Warwick P Bowen. Quantum metrology and its application in biology.
Physics Reports, 615:1–59, 2016.
[3] Christian W Freudiger, Wei Min, Brian G Saar, Sijia Lu, Gary R Holtom, Chengwei He,
Jason C Tsai, Jing X Kang, and X Sunney Xie. Label-free biomedical imaging with high
sensitivity by stimulated Raman scattering microscopy. Science, 322(5909):1857–1861, 2008.
27
[4] Brian G Saar, Christian W Freudiger, Jay Reichman, C Michael Stanley, Gary R Holtom,
and X Sunney Xie. Video-rate molecular imaging in vivo with stimulated Raman scattering.
Science, 330(6009):1368–1370, 2010.
[5] Alexander A Oraevsky, Luiz B Da Silva, Alexander M Rubenchik, Michael D Feit, ME Glinsky,
Michael D Perry, Beth M Mammini, W Small, and Brent C Stuart. Plasma mediated ablation
of biological tissues with nanosecond-to-femtosecond laser pulses: relative role of linear and
nonlinear absorption. IEEE Journal of Selected Topics in Quantum Electronics, 2(4):801–809,
1996.
[6] Takahiro Serikawa and Akira Furusawa. 500 mhz resonant photodetector for high-quantum-
efficiency, low-noise homodyne measurement. Review of Scientific Instruments, 89(6):063120,
2018.
[7] Gil Triginer Garces, Helen M Chrzanowski, Shakib Daryanoosh, Valerian Thiel, Anna L
Marchant, Raj B Patel, Peter C Humphreys, Animesh Datta, and Ian A Walmsley. Quantum-
enhanced stimulated emission detection for label-free microscopy. Applied Physics Letters,
117(2):024002, 2020.
[8] Youngjin Choi, Sohee Lim, Joong Won Shim, Bonghwan Chon, Jong Min Lim, and Minhaeng
Cho. Shot-noise-limited two-color stimulated Raman scattering microscopy with a balanced
detection scheme. The Journal of Physical Chemistry B, 124(13):2591–2599, 2020.
[9] Rajeev Ranjan, Annalisa Darco, Maria Antonietta Ferrara, Maurizio Indolfi, Michele
Larobina, and Luigi Sirleto. Integration of stimulated Raman gain and stimulated Raman
losses detection modes in a single nonlinear microscope. Optics Express, 26(20):26317–26326,
2018.
[10] Ke Wang, Delong Zhang, Kriti Charan, Mikhail N Slipchenko, Ping Wang, Chris Xu, and
Ji-Xin Cheng. Time-lens based hyperspectral stimulated Raman scattering imaging and quan-
titative spectral analysis. Journal of Biophotonics, 6(10):815–820, 2013.
[11] Ji-Xin Cheng and X Sunney Xie. Vibrational spectroscopic imaging of living systems: An
emerging platform for biology and medicine. Science, 350(6264):aaa8870, 2015.
[12] Lu Wei, Fanghao Hu, Yihui Shen, Zhixing Chen, Yong Yu, Chih-Chun Lin, Meng C Wang,
and Wei Min. Live-cell imaging of alkyne-tagged small biomolecules by stimulated Raman
scattering. Nature Methods, 11(4):410–412, 2014.
[13] Senlian Hong, Tao Chen, Yuntao Zhu, Ang Li, Yanyi Huang, and Xing Chen. Live-cell
stimulated Raman scattering imaging of alkyne-tagged biomolecules. Angewandte Chemie,
126(23):5937–5941, 2014.
[14] Hyeon Jeong Lee, Wandi Zhang, Delong Zhang, Yang Yang, Bin Liu, Eric L Barker, Kim-
berly K Buhman, Lyudmila V Slipchenko, Mingji Dai, and Ji-Xin Cheng. Assessing cholesterol
storage in live cells and C. elegans by stimulated Raman scattering imaging of phenyl-diyne
cholesterol. Scientific Reports, 5(1):1–10, 2015.
[15] Kamila Kochan, Huadong Peng, Bayden R Wood, and Victoria S Haritos. Single cell assess-
ment of yeast metabolic engineering for enhanced lipid production using Raman and AFM-IR
imaging. Biotechnology for Biofuels, 11(1):106, 2018.
28
[16] Fredrik Svedberg, Christian Brackmann, Thomas Hellerer, and Annika M Enejder. Nonlinear
microscopy with fiber laser continuum excitation. Journal of biomedical optics, 15(2):026026,
2010.
[18] JB Rattray, Angelo Schibeci, and Denis K Kidby. Lipids of yeasts. Bacteriological Reviews,
39(3):197, 1975.
[19] Matthias Hofer, Naveen K Balla, and Sophie Brasselet. High-speed polarization-resolved
coherent Raman scattering imaging. Optica, 4(7):795–801, 2017.
[20] Gianluigi Zito, Giulia Rusciano, Giuseppe Pesce, Alden Dochshanov, and Antonio Sasso.
Surface-enhanced Raman imaging of cell membrane by a highly homogeneous and isotropic
silver nanostructure. Nanoscale, 7(18):8593–8606, 2015.
[21] Masanari Okuno, Hideaki Kano, Philippe Leproux, Vincent Couderc, James PR Day, Mischa
Bonn, and Hiro-o Hamaguchi. Quantitative CARS molecular fingerprinting of single living
cells with the use of the maximum entropy method. Angewandte Chemie, 122(38):6925–6929,
2010.
[22] Ryo Furuta, Naoyuki Kurake, Kenji Ishikawa, Keigo Takeda, Hiroshi Hashizume, Hiroki
Kondo, Takayuki Ohta, Masafumi Ito, Makoto Sekine, and Masaru Hori. Intracellular-
molecular changes in plasma-irradiated budding yeast cells studied using multiplex co-
herent anti-stokes Raman scattering microscopy. Physical Chemistry Chemical Physics,
19(21):13438–13442, 2017.
[23] Lu Wei, Zhixing Chen, Lixue Shi, Rong Long, Andrew V Anzalone, Luyuan Zhang, Fanghao
Hu, Rafael Yuste, Virginia W Cornish, and Wei Min. Super-multiplex vibrational imaging.
Nature, 544(7651):465–470, 2017.
[24] Henning Vahlbruch, Moritz Mehmet, Karsten Danzmann, and Roman Schnabel. Detection of
15 db squeezed states of light and their application for the absolute calibration of photoelectric
quantum efficiency. Physical Review Letters, 117(11):110801, 2016.
[25] Christian W Freudiger, Wenlong Yang, Gary R Holtom, Nasser Peyghambarian, X Sunney
Xie, and Khanh Q Kieu. Stimulated Raman scattering microscopy with a robust fibre laser
source. Nature Photonics, 8(2):153, 2014.
29