ECI User Manual Dynamic - EnG v1 - 7

This document is the user manual for the ECI Multi Functional Weighing Indicator (Dynamic). It contains information about installing and connecting the indicator, descriptions of the different screens and symbols that will appear, and instructions for using the device features through the menu system. The manual includes specifications, considerations for installation and use, diagrams of physical connections, and step-by-step guides for common tasks like changing parameters and selecting programs.

Uploaded by

dnjovic369
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
22 views354 pages

ECI User Manual Dynamic - EnG v1 - 7

This document is the user manual for the ECI Multi Functional Weighing Indicator (Dynamic). It contains information about installing and connecting the indicator, descriptions of the different screens and symbols that will appear, and instructions for using the device features through the menu system. The manual includes specifications, considerations for installation and use, diagrams of physical connections, and step-by-step guides for common tasks like changing parameters and selecting programs.

Uploaded by

dnjovic369
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 354

ECI Multi Functional Weighing Indicator (Dynamic)

User Manual
ECI
Multi Functional
Weighing Indicator
(Dynamic)

Manual V1.7
Esit Technical Support
Esit Support Line for Turkey

Istanbul
Phone: Direct Line +90 216 585 18 88 / +90 216 585 18 18 Extension: 141 /142
Fax : +90 216 585 18 89
Hendek
Phone: +90 264 614 16 41 / +90 264 614 16 42
Fax : +90 264 614 16 43
Izmir
Phone: +90 232 433 02 99
Fax : +90 232 433 12 42
Ankara
Phone: +90 312 397 01 10
Fax : +90 312 397 07 73
Adana
Phone +90 322 428 11 71
Fax : +90 322 428 11 72

Esit Istanbul Head Office


Address : Nisantepe Mahallesi Gelin Cicegi Sokak No:36 Cekmekoy 34794 Istanbul
Phone : 0216 585 18 18
Fax : 0216 585 18 19
E-mail : [email protected] - [email protected]

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 2


ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 3
1 Considerations
 Do not power-up the device before making sure that the device connections are made according to this
manual.
 Do not open the case or disconnect-insert the connections before disconnecting the device from the power
source.
 Note that the load cell and communication cables are not very close to the energy lines.
 Note that the cables are not crushed.
 Operate the device within the stated temperature range.
 The device has been developed for indoor use. Do not expose sunlight, rain and other external factors.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 4


Table of Contents
1 Considerations ............................................................................................................................................................. 4
2 Introduction ................................................................................................................................................................ 18
3 Physical Dimensions .................................................................................................................................................. 20
4 Specifications............................................................................................................................................................. 22
4.1 Technical Specifications ...................................................................................................................................... 22
4.2 Use Properties..................................................................................................................................................... 23
5 Installation and Connections ...................................................................................................................................... 25
5.1 Standard Modules ............................................................................................................................................... 27
5.1.1 Slot1: Terminal (Supply and Communication Module) ................................................................................... 28
5.1.2 Slot2: Weighing (Load Cell) Module ............................................................................................................. 33
6 Screenshots ............................................................................................................................................................... 37
6.1 Weighing Main Screen ........................................................................................................................................ 37
6.1.1 Screen Layout ............................................................................................................................................... 38
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 5
6.1.2 Symbols ........................................................................................................................................................ 38
6.1.3 Keys.............................................................................................................................................................. 39
6.2 Menu Screen ....................................................................................................................................................... 40
6.2.1 Screen Layout ............................................................................................................................................... 40
6.2.2 Symbols ........................................................................................................................................................ 41
6.2.3 Keys.............................................................................................................................................................. 42
6.3 Change Parameter Screen (Numeric or Alpha-Numeric Entry Type) ................................................................... 43
6.3.1 Screen Layout ............................................................................................................................................... 43
6.3.2 Symbols ........................................................................................................................................................ 44
6.3.3 Keys.............................................................................................................................................................. 44
6.4 Change Parameter Screen (Selective Type) ........................................................................................................ 45
6.4.1 Screen Layout ............................................................................................................................................... 45
6.4.2 Symbols ........................................................................................................................................................ 46
6.4.3 Keys.............................................................................................................................................................. 46

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 6


6.5 Ordered Menu Screen ......................................................................................................................................... 47
6.5.1 Screen Layout ............................................................................................................................................... 47
6.5.2 Symbols ........................................................................................................................................................ 48
6.5.3 Keys.............................................................................................................................................................. 49
6.6 Confirmation Screen ............................................................................................................................................ 50
6.6.1 Screen Layout ............................................................................................................................................... 50
6.6.2 Symbols ........................................................................................................................................................ 51
6.6.3 Keys.............................................................................................................................................................. 51
6.7 Message Screen ................................................................................................................................................. 52
6.7.1 Screen Layout ............................................................................................................................................... 52
6.7.2 Symbols ........................................................................................................................................................ 53
7 Using the Device........................................................................................................................................................ 55
7.1 Program Selection ............................................................................................................................................... 56
7.1.1 Start-up Screen ............................................................................................................................................. 57

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 7


7.1.2 Program Selection Screen ............................................................................................................................ 58
7.2 Menu List............................................................................................................................................................. 59
7.3 Using the Menu ................................................................................................................................................... 60
7.3.1 Terminal Menu .............................................................................................................................................. 60
7.3.2 Scale Menu ................................................................................................................................................... 87
7.3.3 Optional Communication Menu ................................................................................................................... 106
7.3.4 Optional In/Out Menu .................................................................................................................................. 121
7.3.5 Optional Input Output Menu ........................................................................................................................ 139
7.3.6 Optional Expansion Menu ........................................................................................................................... 155
7.4 Zeroing and Tare Operations ............................................................................................................................. 161
7.4.1 Zeroing ....................................................................................................................................................... 161
7.4.2 Tare ............................................................................................................................................................ 161
8 Calibration ............................................................................................................................................................... 163
8.1 Load Calibration ................................................................................................................................................ 163

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 8


8.2 mV/V Calibration ............................................................................................................................................... 168
8.3 Coefficient Calibration ....................................................................................................................................... 172
8.4 Zero Calibration ................................................................................................................................................. 175
9 Software Update ...................................................................................................................................................... 179
10 Annex 1: Optional Modules .................................................................................................................................... 184
10.1 Slot3 (Optional Communication Slot): .............................................................................................................. 184
10.1.1 E-MDL-COM (Serial Communication Module): ......................................................................................... 184
10.1.2 E-MDL-PRF (Profibus Module) .................................................................................................................. 188
10.2 Slot4 (Optional Input/Output Menu): ................................................................................................................ 189
10.2.1 E-MDL-RLY (Relay Module): ..................................................................................................................... 189
10.2.2 E-MDL-DAC (Analogue Output Module): ................................................................................................... 191
10.3 Slot5 (Optional Input/Output Menu) ................................................................................................................. 194
10.4 Slot6 (Optional Expansion Slot) ....................................................................................................................... 194
10.4.1 E-MDL-BCD (Binary Output Module) ......................................................................................................... 194

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 9


10.4.2 E-MDL-EXP (Relay Expansion Module) ................................................................................................... 197
11 Annex 2 Belt Scale Program .................................................................................................................................. 201
11.1 Specifications .................................................................................................................................................. 201
11.2 Main Program Screen ...................................................................................................................................... 202
11.2.1 Screen Layout ........................................................................................................................................... 203
11.2.2 Keys.......................................................................................................................................................... 204
11.2.3 Symbols .................................................................................................................................................... 205
11.3 Beltscale Menu ................................................................................................................................................ 206
11.4 Beltscale Menu Usage ..................................................................................................................................... 207
11.5 Alibi Memory ................................................................................................................................................... 217
11.6 Auto Zero ........................................................................................................................................................ 218
11.7 Calibration of BeltScale ................................................................................................................................... 219
11.7.1 Load Calibration ........................................................................................................................................ 219
11.7.2 Zero Calibration ........................................................................................................................................ 224

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 10


11.8 BeltScale Connections .................................................................................................................................... 227
11.8.1 Input Connections ..................................................................................................................................... 227
11.8.2 Output Connections .................................................................................................................................. 228
12 Annex 3 Dosage Belt Scale Program ..................................................................................................................... 230
12.1 Specifications .................................................................................................................................................. 230
12.2 Main Program Screen ...................................................................................................................................... 231
12.2.1 Screen Layout ........................................................................................................................................... 232
12.2.2 Keys.......................................................................................................................................................... 233
12.2.3 Symbols .................................................................................................................................................... 234
12.3 Dosage Belt Scale Menu ................................................................................................................................. 235
12.4 Dosage Belt Scale Menu Usage ...................................................................................................................... 236
12.5 Alibi Memory ................................................................................................................................................... 248
12.5 Auto Zero ........................................................................................................................................................ 249
12.6 Calibration of Dosage Belt Scale ..................................................................................................................... 250

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 11


12.6.1 Load Calibration ........................................................................................................................................ 250
12.6.2 Zero Calibration ........................................................................................................................................ 255
12.7 Dosage Belt Scale Connections ...................................................................................................................... 258
12.7.1 Input Connections ..................................................................................................................................... 258
12.7.2 Output Connections .................................................................................................................................. 260
13 Annex 4 Impact Scale Program .............................................................................................................................. 261
13.1 Impact Scale Program Specifications .............................................................................................................. 261
13.2 Home Screen View .......................................................................................................................................... 262
13.2.1 Screen Layout ........................................................................................................................................... 263
13.2.2 Keys.......................................................................................................................................................... 264
13.2.3 Symbols .................................................................................................................................................... 265
13.3 Impact Scale Menu .......................................................................................................................................... 266
13.4 Use of the Impact Scale Menu ......................................................................................................................... 267
13.5 Alibi Memory ................................................................................................................................................... 270

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 12


13.6 Auto Zero ........................................................................................................................................................ 270
13.7 Impact Scale Calibration .................................................................................................................................. 270
13.7.1 Load Calibration ........................................................................................................................................ 270
13.7.2 Zero Calibration ........................................................................................................................................ 270
13.8 Impact Scale Connections ............................................................................................................................... 270
14 Annex 5: CheckWeigher Program .......................................................................................................................... 272
14.1 The Operation Sequence................................................................................................................................. 272
14.1.1 Dynamic Mode .......................................................................................................................................... 272
14.1.2 Static Mode ............................................................................................................................................... 272
14.2 CheckWeigher Program Screen ...................................................................................................................... 273
14.2.1 Screen Layout (Dynamic Mode) ................................................................................................................ 275
14.2.2 Screen Layout (Static Mode) ..................................................................................................................... 275
14.2.3 Screen Layout (Information Screen) .......................................................................................................... 276
14.2.4 Symbols .................................................................................................................................................... 277

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 13


14.2.5 Keys.......................................................................................................................................................... 278
14.3 CheckWeigher Menu ....................................................................................................................................... 279
14.4 CheckWeigher Menu Usage ............................................................................................................................ 280
14.5 Check Weigher Connections ........................................................................................................................... 288
14.5.1 Input Connections ..................................................................................................................................... 288
14.5.2 Output Connection .................................................................................................................................... 289
15 Annex 6 Modbus .................................................................................................................................................... 291
15.1 Device Menu Settings...................................................................................................................................... 291
15.2 Connection Point Settings ............................................................................................................................... 291
15.3 Modbus Commands ........................................................................................................................................ 292
15.4 Modbus Directly Accessible Memory Map ....................................................................................................... 292
15.4.1 Belt Scale Program ................................................................................................................................... 295
15.4.1 Dosage Belt Scale Program ...................................................................................................................... 296
15.4.2 CheckWeigher Program ............................................................................................................................ 297

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 14


15.5 Examples: ....................................................................................................................................................... 298
15.5.1 Reading more than one register. ............................................................................................................... 298
15.5.2 Writing to the register. ............................................................................................................................... 299
15.5.3 Removing the Security Lock.(Command Register =1, RegA =1357h, RegB=2192h) ................................. 300
15.5.4 Writing to EEPROM section ...................................................................................................................... 301
15.6 Command Register Functions ......................................................................................................................... 302
15.7 Examples ........................................................................................................................................................ 304
15.7.1 Taring ....................................................................................................................................................... 304
15.7.2 Setting the Relay 1 Minimum Value........................................................................................................... 305
15.7.3 Setting the Relay 2 Maximum Value.......................................................................................................... 306
16 Annex 7 Profibus.................................................................................................................................................... 308
16.1 BeltScale Program........................................................................................................................................... 308
16.2 Dosage Belt Scale Program ............................................................................................................................ 310
16.3 CheckWeigher Program .................................................................................................................................. 312

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 15


17 Annex 8 Ethernet IP Addresses ............................................................................................................................. 317
17.1 Belt Scale Program.......................................................................................................................................... 317
17.2 Dosage Belt Scale Program ............................................................................................................................ 319
17.3 ChechWeigher Program .................................................................................................................................. 321
18 Annex 9 Profinet Addresses ................................................................................................................................... 326
18.1 Profinet Commands (Belt Scale Program) ....................................................................................................... 327
18.2 Profinet Commands (Checkweigher Program) ................................................................................................. 328
19 Error Warning And Information Messages .............................................................................................................. 346

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 16


ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 17
2 Introduction
ECI is a smart signal converter and control device which is designed for weigh measuring systems. It is a fast,
sensitive device with a high accuracy class. It is designed suitably for using at industrial applications and standard weigh
measuring areas. It is used as weigh, force and pressure measuring areas. Many optional modules can be
added/modified easily thanks to the modular smart “plug and play” expansion structure.
This operating manual contains exemplary wording related to the installation, operation and field operation of
the ECI.
All written and visual materials contained in this operating manual are the property of ESİT Elektronik
Sistemleri İmalat ve Tic. Ltd. Sti.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 18


ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 19
3 Physical Dimensions

Standard Model
Panel Cut
Weight 360 g

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 20


ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 21
4 Specifications
4.1 Technical Specifications
Model ECI, E Series
Load Cell Input +/- 2.50–160mV/V
A/D Speed (/second) 1600
Screen Resolution 9 digit (999999.999)
Supply 12-24V DC
Display LCD (101x33 pixel graphic, 4.5 inch)
Communication Terminal (Standard) RS-232 (9600-115200bps)
Communication Remote Display (Standard) For Remote Display (Opto-Coupler)
Communication Option Module (Optional) RS-232-RS-485 / RS-422 (E-MDL-COM insulated)
Load Cell Feeding 10 VDC (280mA, maximum 10 load cells)
Operating Temperature -10C / +50C
Protection Class IP65 (Front Panel)

(Standard: Supplied as standard hardware with the device)


(Optional: Inserted to the device as additional hardware)

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 22


4.2 Use Properties
 Multi-language support
 5 touch keys
 Plain use and setting thanks to the indexed menu system structure
 Internal beltscale dosage beltscale and checkweigher program
 Smart “plug and play” expansion structure
 Gain setting adjustable according to the sensor output (capable of reading between ±20mV and ±1.6V)
 Connectivity up to 10 load cells (350R)
 Software update by using USB
 Calibration without installation on the basis of the load cell data
 Facility to carry the calibration information between different devices.
 Easy test opportunity with the service menus (Serial Output, Digital inputs/Outputs, LCD, memory, ADC input)
 3 sectional weighing range (Wide Range , Multi Partition options)
 Automatic scale interval (e) and dot space adjustment
 gResolver, Smart numerical filter technology
 Storing current weighing device values and calibration info
 Monitoring the Load Cell info real time as mV/V from the service menu
 16 recipe and double speed batching up to 8 silos
 Up to 24 optional digital output and up to 16 digital input
 Optional 2 x 0/10V – 4/20mA analogue output
 USB reporting (Last 64 calibration log, Last 256 Error Record and Last 256 Overload log)

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 23


ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 24
5 Installation and Connections
The number of each slot is shown as follows on the back cover of the ECI device

Slot Numbers

Slot6 Slot5 Slot4 Slot3 Slot2 Slot1

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 25


 Slot1: Used for the Terminal (Supply & Communication) Module. Following module(s) can be used;
E-ECI-PWR (Standard) module can be inserted.
 Slot2: Used for Weighing (Load Cell) Module. Following module(s) can be used;
E-ECI-LC (Standard) module can be inserted.
 Slot3: Used for optional Communication Module. Following module(s) can be used;
E-MDL-COM (Optional), RS232 + RS485/RS422 Module
E-MDL-PRF (Optional), Profibus Module
E-MDL-EIP (Optional), Ethernet IP Module
 Slot4: Used for Optional Input/Output modules. Following module(s) can be used;
E-MDL-RLY, 4in/4out Module (Optional)
E-MDL-DAC, 0-10V / 4-20mA Analogue +2in/1out Module (Optional)
 Slot5: Used for Optional Input/Output modules. Following module(s) can be used;
E-MDL-RLY, 4in/4out Module (Optional)
E-MDL-DAC, 0-10V / 4-20mA Analogue +2in/1out Module (Optional)
 Slot6: Used for Expansion modules. Following module(s) can be used;
E-MDL-EXP, 8in/16out expansion module (Optional)
E-MDL-BCD, 20out/2In Optical Insulated BCD or Binary Output Module (Optional)

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 26


5.1 Standard Modules
Standard modules are the units that are given in the base mode with the ECI E series devices. Respective modules
and explanations are given below.

Terminal Module

Weighing
Module

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 27


5.1.1 Slot1: Terminal (Supply and Communication Module)

RS232 Connection

Remote Display Connection

Supply Connection

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 28


5.1.1.1 Supply Connection

5.1.1.2 Communication Connections

Note: When the “R.Display” output is used with Esit brand Remote displays, red and White cables of the Display must
be short circuited
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 29
5.1.1.3 Printer Connection
ECI, supports EPM203-MRS, TOSHIBA B-EV4T and Zebra EPL model printer. Printer connection must be
as follows.

EPM203-MRS Connection

ECI Printer

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 30


ZEBRA EPL Connection

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 31


TOSHIBA B-EV4T

Printer

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 32


5.1.2 Slot2: Weighing (Load Cell) Module
5.1.2.1 Esit Load Cell Connection

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 33


5.1.2.2 Six (6) Cable Load Cell Connection

Sense +
Esit Colours
Sense - Sense + Blue
Sense - Orange
Out + Out + White
Out - Red
Out - Supply + Green
Supply - Black
Supply +

Supply -

Grounding

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 34


5.1.2.3 Four (4) Cable Load Cell Connection

Esit Colours
Sense - Green
Sense - Black
Out + Out + White
Out - Red
Out - Supply + Green
Supply - Black
Supply +
Sense-terminal is short
Supply -
circuit with Supply – terminal
and Sense+terminal is short
Grounding
circuit with Supply + terminal

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 35


ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 36
6 Screenshots
6.1 Weighing Main Screen
Main screen shot will be as shown at side

The screen will be displayed as shown at side when the tare is


removed

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 37


6.1.1 Screen Layout
1. Weighing Indicator
2. Keys 3 4
3. Weighing Range 5
1 6
4. Mobility
5. Zero Indicator 7
6. Tare Status
2
7. Unit

6.1.2 Symbols
Indicates current weighing range. It is displayed when the measuring type is
Weighing Range selected as “Multi Partition” or “Multi Range”. E1: Lower Zone, E2: Upper
Zone and E3 represents the final zone.
Shows that the change in the weight value is within the stagnancy limits for
Stagnancy a certain period
(less than a one scale interval).
Mobility Shows that the change in the weight value is out of the stagnancy limits.
Zero Indicator Indicates that deviation from zero is less than 1/4e (resolution).
Shows that the tare has been removed and the displayed value is the net
Net
value.
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 38
6.1.3 Keys
Reset Key Used for resetting the display.

Tare Key Used for taking the weight on the load cell as tare.

Tare Clean Key Used for cancelling the tare value and showing the gross value.

Menu Key Used for bringing the display menu where the device settings are made.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 39


6.2 Menu Screen
Menu screenshot is as shown at side.

6.2.1 Screen Layout


1
1. Menu Logo
2. Menu No 2 3 4
3. Lock
4. Menu Names
5. Keys 5

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 40


6.2.2 Symbols

Menu Logo Respective menu logo.

Menu No Respective menu number.


Shows that a parameter has to be changed in order to
Software Lock
Access the menu.
Shows that it is required to add or replace hardware in
Hardware Lock
order to Access the menu.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 41


6.2.3 Keys
Left Arrow Key Used for shifting to an upper menu on the menu screen.
Up Arrow Key Used for browsing among the options on the menu screen.
Down Arrow Key Used for browsing among the options on the menu screen.
Right Arrow Key Used for shifting to an lower menu on the menu screen.
Enter Key Used for entering to the menu next to the menu number.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 42


6.3 Change Parameter Screen (Numeric or Alpha-Numeric Entry
Type)
The change parameter screen will be displayed as shown at
side:

Note: The values entred in the screen must the multiplies of the
“Resolution” parameter. For example if the ”Resolution” is selected as 5 and the value entered in this screen is 101, an
information message appears and the number is automatically rounded to 100 and returned to the parameter changing
screen.

6.3.1 Screen Layout 1 2


1. Menu No
2. Menu Title 3
3. Value to be Adjusted
4. Keys 4

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 43


6.3.2 Symbols
Menu No Respective menu number.
Title Name of the current menu
Option Value to be adjusted

6.3.3 Keys
Exit Key. Used for returning to the menu screen from the parameter settings screen
Left Arrow Key Used for shifting the cursor to the left.
Used for decreasing the digit on which the cursor is located in the areas
Minus Key
where numerical value is entered in the parameter setting screen.
Used for increasing the digit on which the cursor is located in the areas
Plus Key
where numerical value is entered in the parameter setting screen.
Right Arrow
Used for shifting the cursor to the right.
Key
Confirm Key Used for confirming the selected screen.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 44


6.4 Change Parameter Screen (Selective Type)
The selective type change parameter screen will be displayed as
shown at side:

6.4.1 Screen Layout


1 2
1. Menu No
2. Menu Title 3 4 5
3. Option No
4. Option
5. Arrow 6
6. Keys

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 45


6.4.2 Symbols
Menu No Respective menu number.
Menu Title Name of the respective parameter.
Option No Number of the parameter option.
Option Explanation of the respective parameter option.
Shows if there is any option above or below the
Arrow
selection list

6.4.3 Keys
Used for returning to the menu screen from the parameter settings
Exit Key.
screen
Up Arrow Key Used for browsing among the options in the selection list.

Down Arrow Key Used for browsing among the options in the selection list.

Enter Key Used for conforming the value selected.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 46


6.5 Ordered Menu Screen
Ordered menu is made for providing convenience in the menus
with more than one sub-menu. For example, there are Unit,
Capacity, Resolution and Measuring Type sub-menus in the
Settings menu under the Weighing menu. When you enter this
menu, you can change every parameter related to the weighing
setting subsequently. The ordered menu screenshot will be as
sown at side

6.5.1 Screen Layout


1. Menu No
2. Menu Title 1 2
3. Sub-Menu Title 3 4 5
4. Sub-Menu No
5. Sub-Menu Number 6 7 8
6. Option No
7. Option 9
8. Arrow
9. Keys

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 47


6.5.2 Symbols
Menu No Respective menu number.
Menu Title Name of the current menu.
Sub-Menu Title Name of the current sub-menu.
Sub-Menu No Sub-Menu number.
Number of Sub-
Number of Sub-Menus of the respective menu
Menu
Option No Option number
Option Respective option
Shows if there is any option above or below
Arrow
the selection list

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 48


6.5.3 Keys

Return Key Used for returning to the previous menu in the ordered menu.

Up Arrow Key Used for browsing among the options in the selection list.

Down Arrow Key Used for browsing among the options in the selection list.

Forward Key Used for forwarding to the next menu in the ordered menu.
Used for returning to the menu screen from the ordered menu
Exit Key.
screen

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 49


6.6 Confirmation Screen
When key is pushed after entering to menu and changing
any parameter, the confirmation screen as shown at side will
appear before saving the new parameter.

6.6.1 Screen Layout


1
1. Screen Title
2. Confirmation Logo
3. Parameters Changed 2 3 4
4. New Parameter Values
5. Keys
5

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 50


6.6.2 Symbols
Confirmation Screen
Respective menu title.
Title
Confirmation Screen
Respective menu logo.
Logo
Parameters Changed Name of the parameter changed

Parameter Value New value of the parameter changed.

6.6.3 Keys
Cancel Key Used for returning to the previous menu without saving the parameters changed.
Confirm
Used for saving the parameters changed.
Key
Save and
It is used for saving the parameters changed and returning to the main screen.
Exit Key

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 51


6.7 Message Screen
The message screen shown at side will display when it is
necessary to inform the user. There are 3 types of message
These are information, warning and error messages.

6.7.1 Screen Layout


1 2
1. Message Screen Title
2. Message No 3 4
3. Message Logo
4. Message

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 52


6.7.2 Symbols

Message Screen Title Indicates the message type

Respective message
Message No
number.
The logo related to the
Message Screen Logo
message type.
Message The message content

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 53


ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 54
7 Using the Device
The device is turned on by supplying energy after making the Supply and Load Cell connections. Start-up logo and
version number will be displayed on ECI LCD screen during the start-up. Then the functional test will be performed and
the Optional card inserted will be recognized and tested. Lastly, the main screen will be displayed and the measuring
starts.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 55


7.1 Program Selection
ECI, supports more than one operating screens (programs) (such
as "Scale" Batching", "Filling" etc). Operating screen selection is
made when the first energy is given to the device. After giving the
first energy, when the following screenshot appears, the 1. and 3.
keys from the left are pushed together.

The desired program is selected from "Select Program" menu


appeared. Related screenshot is shown below.

The status of the outputs may change when the program is


changed. Thus, the respective warning message is displayed
before the program selection screen.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 56


7.1.1 Start-up Screen
Please follow the steps blow;
1. Energize the device
2. Wait for the initializing screen which is shown after
the “ESIT” logo.
3. Press 1 and 3 keys together.
4. If correct combination is done “Select Program”
screen appears.

According to device type (shown on screen) the


selectable device modes are listed below;

Automatic Mode Dynamic Mode


Scale Belt Scale Push buttons together when this screen
Batching Impact Scale apperars.
Filling Dosage Belt Scale
Hopper Scale Checkweigher
Peak Hold

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 57


7.1.2 Program Selection Screen
After selecting the desired program, selection is confirmed
with key and it is shifted to the desired operating
program.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 58


7.2 Menu List
Menu List is as follows:

1. Terminal 2. Weighing 3. Optional Communication Menu


1.1. Language Selection 2.1. Security Type 4. Optional In/Out Menu
1.2. Date 2.2. Settings 5. Optional In/Out Menu
1.3. Time 2.3. Filter 6. Optional Expansion Menu
1.4. Communication 2.4. Zero/Tare
1.5. USB Recording 2.5. Calibration
1.6. Keys 2.6. Service
1.7. Service

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 59


7.3 Using the Menu
7.3.1 Terminal Menu
The menus related to the Terminal module inserted to slot 1
of the device are displayed under this menu.

Terminal
It is the main menu that contains the device settings and the settings related to Slot1.

Language Selection
The language selection for the device. The options are;
1) English
2) Turkish

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 60


Date
Date settings are made. The date data can be changed
in dd.mm.yyyy format.

Time
Time settings are made. The time data can be changed
in hh.mm format.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 61


Communication
Serial Out
Contains the terminal serial out settings. This menu is an ordered type menu. Sub-menus are Mode
and Period. It functions differently in Indication, Batching and Filling programs. Explanation is made
below for each program
Serial Out (Belt Scale Program)

Mode
It is the communication type selection. The options
are;
1) Flow: Current flow rate data is sent once every
second. (Baudrate is 9600bps)

Example;
Flow value “214” is shown as below.

Character ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘ 2‘ ‘1‘ ‘4 ‘ CR
Ascii 32 32 32 32 32 32 32 50 49 52 13
Hex 20 20 20 20 20 20 20 32 31 34 0D

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 62


2) Speed: Current belt speed is sent once per second. (Baudrate is 9600bps) Format of the belt
speed data is same with the flow data.
3) Zero: Current AutoZero value is sent once per second. (Baudrate is 9600bps) Format of this data
is same with the flow data.
4) Load: Load value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow data.
5) Total 1: Total 1 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow data.
6) Total 2: Total 2 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow data.
7) Total: Total value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow data.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 63


Serial Out (Check Weigher Program)
Mod
Selection of communication type. Options are;
1) Last Weight When after the transition is
finished, “Last Weight” information is sent.
(Baudrate is 115200bps)
For instance;
“120.0” value is sent as shown in the table below.
Karakter ‘‘ ‘‘ ‘‘ ‘‘ ‘ 0‘ ‘1‘ ‘2 ‘ ‘0 ‘ ‘.’ ‘0 ‘ CR
Ascii 32 32 32 32 32 49 52 56 46 48 13
Hex 20 20 20 20 20 31 32 30 2E 30 0D

2) Total: When after the transition is finished, “Total” information is sent. (Baudrate is 115200bps)
The format of the information sent is the same as “Last Weight” string.
3) Mean: When after the transition is finished, Mean of Product weighing information is sent.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 64


4) Detail: When after the transition is finished ,information is sentas detailed.(Baudrate is
115200bps).Sending information string format is shown as below.

Product Last
Date T Time T T T Total CR
Qantity Product
16.04.2015 * 13:50:19 * 4 * 80 * 300 Chr 13
T : Tab
* : Chr 9

Date Transition date


Time Transition time
Product Quantity Total product number
Last Product Last product wwighing of last transition on band
Total Total product weighing of transition on band

5) Command: Functions as a query from a remote computer or similar device.


Not case sensitive.
6) Service: While product is on weighing band, weighing information is send continuously.
7) Off: Data sending over serial port is disabled

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 65


Remote Display (In Belt Scale Program)
It is the section that contains the remote display out (opto-coupler) settings.
Mode
The options are;
1) Flow: Current Flow value is sent once per
second. (Baudrate is 9600bps)

Example;
Flow value “214” is shown as below.

Character ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘ 2‘ ‘1‘ ‘4 ‘ CR
Ascii 32 32 32 32 32 32 32 50 49 52 13
Hex 20 20 20 20 20 20 20 32 31 34 0D

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 66


2) Speed: Current belt speed value is sent once per second. (Baudrate is 9600bps) Format of this
data is same with the flow data.
3) Zero: Current AutoZero value is sent once per second. (Baudrate is 9600bps) Format of this data
is same with the flow data.
4) Load: Load value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow rate data.
5) Total 1: Total1 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow data.
6) Total 2: Total2 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow data.
7) Total: Total value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow data.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 67


Serial Out (Dosage Belt Scale Program)

Mode
Selection of communication type. Options are;

1) Flow: Current Flow value is sent once per


second. (Baudrate is 9600bps)

Example;
Flow value “214” is shown as below.

Character ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘ 2‘ ‘1‘ ‘4 ‘ CR
Ascii 32 32 32 32 32 32 32 50 49 52 13
Hex 20 20 20 20 20 20 20 32 31 34 0D

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 68


2) Speed: Current belt speed value is sent once per second. (Baudrate is 9600bps) Format of this
data is same with the flow data.
3) Zero: Current AutoZero value is sent once per second. (Baudrate is 9600bps) Format of this data
is same with the flow data.
4) Load: Load value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow rate data.
5) Total 1: Total1 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow data.
6) Total 2: Total2 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow data.
7) Total: Total value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow data.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 69


Remote Display (Belt Scale Program)
It is the section that contains the remote display out (opto-coupler) settings.
Mode
The options are;
1) Flow: Current Flow value is sent once per
second. (Baudrate is 9600bps)

Example;
Flow value “214” is shown as below.

Character ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘ 2‘ ‘1‘ ‘4 ‘ CR
Ascii 32 32 32 32 32 32 32 50 49 52 13
Hex 20 20 20 20 20 20 20 32 31 34 0D

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 70


2) Speed: Current belt speed value is sent once per second. (Baudrate is 9600bps) Format of this
data is same with the flow data.
3) Zero: Current AutoZero value is sent once per second. (Baudrate is 9600bps) Format of this data
is same with the flow data.
4) Load: Load value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow rate data.
5) Total 1: Total1 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow data.
6) Total 2: Total2 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow data.
7) Total: Total value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow data.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 71


Remote Display (In Dosage Belt Scale Program)
It is the section that contains the remote display out (opto-coupler) settings.
Mode
The options are;
8) Flow: Current Flow value is sent once per
second. (Baudrate is 9600bps)

Example;
Flow value “214” is shown as below.

Character ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘ 2‘ ‘1‘ ‘4 ‘ CR
Ascii 32 32 32 32 32 32 32 50 49 52 13
Hex 20 20 20 20 20 20 20 32 31 34 0D

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 72


9) Speed: Current belt speed value is sent once per second. (Baudrate is 9600bps) Format of this
data is same with the flow data.
10) Zero: Current AutoZero value is sent once per second. (Baudrate is 9600bps) Format of this data
is same with the flow data.
11) Load: Load value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow rate data.
12) Total 1: Total1 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow data.
13) Total 2: Total2 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow data.
14) Total: Total value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow data.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 73


Remote Display (In CheckWeigher Program)
Remote Display output(opto-coupler) configuration is set by this parameter.
Mode
1) Last Product: When transition is finished,
Product Weighing is shown. By this parameter.

For instance;
“120.0”value of display is shown as below table.

Character ‘‘ ‘‘ ‘‘ ‘‘ ‘ 0‘ ‘1‘ ‘2 ‘ ‘0 ‘ ‘.’ ‘0 ‘ CR


Ascii 32 32 32 32 32 49 52 56 46 48 13
Hex 20 20 20 20 20 31 32 30 2E 30 0D

2) Total: When the transition is finished, Total product weighing information is sent.(Baudrate is
115200bps) Sending information string format is same as “Last Product” mode.
3) Mean: When transition is finished, Mean product weighing information is sent.
4) Close: Data transfer is not done to external indicator.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 74


USB Logging
It is the menu where the logging settings are made. This menu is an ordered type menu. Sub-menus are
Mode and Period. It is active under filling and Scale Programs.

Mode
It is the recording type selection. The options are;
1) Inactive: No recording is made.
2) Weight: Net weight value is recorded.
3) Detail: Weight data is recorded in detail.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 75


The format of the data recorded in USB is as follows.

Belt Scale Programs Format

Mode
It is the recording type selection. The options are;
1) Disabled: No recording.
2) Detail: Date, Time, Flow, Total1, Total2, Total,
Speed, Kg/m values are recorded.
3) Custom: In this mode, only one measurement data
is recorded. The record data defined after this
parameter.

USB record mode “Detail”

Flow C
Date T Time T T Sum 1 T Sum 2 T Sum T Speed T Kg/m
Rate R
31.12.201 25.45 C
* 08:50:19 * 185.2 * 968.8 * 565.8 * 1232.6 * 2 *
4 R
T : Tab
* : Chr 9

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 76


Date Date of the record taken
Time Time of the record taken
Flow Amount of product passes through belt, in one hour.(It’s unit is Ton)
Total amount of product passed through belt. You can reset Sum1 value by
Total 1
pressing Zero button on main screen.
Total amount of product passed through belt. You can reset Sum1 value by
Total 2
pressing Zero button on main screen.
Total amount of product passed through belt. You can reset this value only
Total
when the program is changed.
Speed Actual belt speed.
Kg/m Average amount of product per meter of the belt.

Name of the recorded file is in the format below.

“YearMonthDay”_DETAIL. Such as: 20150501_DETAIL

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 77


Dosage Belt Scale Program Format

Mode
It is the recording type selection. The options are;
1) Disabled: No recording.
2) Detail: Date, Time, Flow, Total1, Total2, Total,
Speed, Kg/m values are recorded.
3) Custom: In this mode, only one measurement data
is recorded. The record data defined after this
parameter.

USB record mode “Detail”

Flow C
Date T Time T T Sum 1 T Sum 2 T Sum T Speed T Kg/m
Rate R
31.12.201 25.45 C
* 08:50:19 * 185.2 * 968.8 * 565.8 * 1232.6 * 2 *
4 R
T : Tab
* : Chr 9
Date of the record taken
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 78
Date
Time Time of the record taken
Flow Amount of product passes through belt, in one hour.(It’s unit is Ton)
Total amount of product passed through belt. You can reset Sum1 value by
Total 1
pressing Zero button on main screen.
Total amount of product passed through belt. You can reset Sum1 value by
Total 2
pressing Zero button on main screen.
Total amount of product passed through belt. You can reset this value only
Total
when the program is changed.
Speed Actual belt speed.
Kg/m Average amount of product per meter of the belt.

Name of the recorded file is in the format below.

“YearMonthDay”_DETAIL. Such as: 20150501_DETAIL

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 79


Format of Check Weigher Program
Mode
It is selected to kind of registiration.
Options;
1) Disabled: It is not done to registiration
2) Detail:Date,hour,Counter, Last product and Total
data values are saved.
Mode of USB registiration “Detail”

Last
Date T Time T Counter T T Total CR
Product
16.04.2015 * 13:50:19 * 4 * 80 * 300 Chr 13
T : Tab
* : Chr 9
Date Transition date
Time Transition time
Counter Total product number
Last Weight Last weight value of transition on band
Total Total product weighing of transition on band.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 80


Saved file name format is like “YearDaynMonth”_DETAIL . For instance: 20150416_DETAIL

3) Custom: When this mode is selected, special mode menu is shown.


Custom Mode
Selection to kind of records.
Options;
1) Last Product: Product weighing is saved to last
transition on band
2) Total: When the transition is finished, Total product
weighing information is sent
3) Mean: When transition is finished, Mean product weighing information is sent.
Period
It is the selection for USB Recording frequency. The
options are;
1) 10/Sec
2) 1/Sec
3) 1/Min
4) 4/Hour
Note: The Period menu is not displayed when "Inactivate" mode is selected.
Period menu is not displayed in Batching and Filling programs. When "Detail" mode is selected, Period can not be selected as 10/Sec

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 81


Keys
It is the menu where the key settings are made. This menu is an ordered type menu. Sub-menus are
"Tone" and "Sensitivity".

Sound
It is the key tone selection. The options are;
1) On
2) Off

Sensitivity
It is the selection for the perception setting of the keys.
The options are;
1) Sensitive: It is the option where the perception is
the highest. Suitable for using with gloves.
2) Standard: It is the state of normal operation.
3) Stable: It is the option where the perception is the
lowest.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 82


Maintenance
LCDPixelTest
It is the screen test menu in LCD. Pixel and rear light
control is made.

CPU Temp.
Inside temperature of the device is displayed for 3
seconds.

Memory Test
It is used for understanding whether the memory unit
is functional or not.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 83


Serial Test
Terminal
It is used for understanding whether the
communication port is functional or not. The data
received from the serial port is displayed at the
receiving section. "ESIT" wording is sent to the
serial port at the same time.

Remote
It is used for understanding whether the
communication port is functional or not. "ESIT"
wording is sent to the remote display.

Serial No
Used for displaying the device serial number.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 84


Firmware No
Used for displaying the device version number.
vX.Y.Z
X: Represents the updates in the basic structure of
the software.
Y: Represents the regional updates in the structure
of the software.
Z: Represents the small changes in the software
such as changing the Word written on a button.

FactoryDef
Used for returning the device parameters to the
Factory settings.

Note: The device can also be returned to the factory settings by pressing to the calibration key for 10 seconds while the
device is at the main screen. The warning screen appearing later on will display that the factory settings will be returned
and sound warning is given. The factory settings will be returned when it is pressed for an additional period of 10
seconds.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 85


ResetDevice
Used for restarting the device.
Reports
It is used for transferring calibration, error and over
loading logs to USB.

Calibration report format is as follows:

Date Time Coefficient Zero Value Calibration Type


19.11.2013 09:02:15 0.03331 1254 Load

Error report format is as follows:

Date Time Error No


19.11.2013 09:02:15 0.03331

Overload report format is as follows:

Date Time Overload


19.11.2013 09:02:15 0.03331

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 86


7.3.2 Scale Menu
The menus related to the weighing module inserted to slot 2
of the device are displayed under this menu.

Scale
It is the menu that contains weighing device (Slot2-Load Cell Module) parameters.
Note: If the symbol presents in any of the sub-menus of the Weighing menu, then the menu cannot be accessible. In
such case, it is necessary to press the calibration key for a short time to activate in order to access the menu.

Calibration Key

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 87


Security Type
Contains the weighing device security settings. The
options are;
1) LegalForTrd:Weighing parameters and calibration
cannot be changed. (Locked)
2) Calibra.Lock: Weighing parameters can be changed
but the calibration cannot be changed. (Locked)
3) FullUnlocked: Weighing parameters and calibration
can be changed.

Note: If the security type is selected as "LegalForTrd" or "Calibra.Lock", primarily the calibration key should be pressed
and the lock should be removed for making any change.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 88


Setup
This menu is an ordered type menu where the weighing parameters are set forth. The sub-menus are
"Unit", "Sys.Capacity", "Resolution" and "Measuring Type".
Units
Determines the unit of the weighing device . The
options are;
1) (g)Gram
2) (kg)Kilogram
3) (t)Ton
4) (N)Newton
5) (lb)Pounds
6) (lt)Liter
7) (mV/V) Input
8) (m3)CubicM
9) None

Sys.Capacity
It is the capacity value of the device. The capacity of
the load cell must not be exceeded.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 89


Resolution
It is the scale interval (e) selection.
e: It is the smallest screen change quantity.
Scale Interval
Scale Interval is automatically calculated as the
multiplies of 1, 2 and 5 depending on the capacity
parameter and presented as a selection. The options
are;
1) Capacity / 6000: The value calculated as 1/6000 as
standard.
2) Capacity / 3000: The value calculated as 1/3000 as standard.
3) Capacity / 1500: The value calculated as 1/1500 as standard.
4) Custom: This option is selected if it is desired to select a resolution other than these first three options.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 90


When the custom resolution is selected Multiple and Scale Interval menus appear and the product of the
values selected in these menus are determined as the walking step.

Multiplier
The options are;
1) 1000
2) 100
3) 10
4) 1
5) 0.1
6) 0.01
7) 0.001

Step Value
The options are;
1) 1
2) 2
3) 5

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 91


ScaleType
Determines the operation of the weighing device. In weighing devices;
“e”: is referred as the scale interval It is the difference between two subsequent indications.
There may be a signle "e" division as well as more than one partitions. The options related to this
parameter are;

1) Single
2) Multi Int
3) Multi Rng

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 92


Single: The weighing device with a single scale interval and
capacity.
Multi Int: In this operation there are 3 different scale intervals
and capacities. There is a different scale interval in each
capacity range;
Weight < LowerZoneLimit + 9e1 e1 (the first zone scale
interval)
LowerZoneLimit + 9e1 < Weight < UpperZoneLimit + 9e2 e2
= 2*e1 (the second zone scale interval)
UpperZoneLimit + 9e1 < Weight < Capacity + 9e3  e3 = 2*e2
(the third zone scale interval)
"e1" scale interval is determined with Resolution parameter. e2
and e3 are determined automatically. If the respective zone
capacity +9e is observed when the weight is increased, it is proceeded to the next scale interval. When the
weight is decreasing, it is decreased to the previous scale interval when it is shifted from the upmost zone to
the lowest one or from the lower zone to the 1. zone. The scale interval can be the multiples of 1,2 and 5. For
example when e1=2, e2=2*e1=2*2=4 and calculated as 5 which is the next scale interval.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 93


Rng: In this operation there are 3 different scale intervals and
capacities. There is a different scale interval in each capacity
range;
e1 (first zone scale interval): is determined with Resolution
parameter
e2 = 2*e1 (the second zone scale interval): is determined
automatically
e3 = 2*e2 (the third zone scale interval): is determined
automatically
If the respective zone capacity +9e is observed when the weight
is increased, it is proceeded to the next scale interval. However,
the scale interval does not change when shifting from the 3.
zone to the 2. one or from the 2. zone to the 1. one when the
weight is decreasing. It is shifted to e1 when the weight is below 0,5 e(i).
The scale interval can be the multiples of 1,2 and 5. For example when e1=2, e2=2*e1=2*2=4 and calculated
as 5 which is the next scale interval. It is shifted to 3 partition operation mode when the measuring type is
selected as "Multi Interval" or "Multi Partition". In such case, following menus are displayed respectively in
order to determine the upper and lower Zone limits.In multi partition systems
Capacity>UpperZoneLimit>LowerZoneLimit must be entered.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 94


UpperZoneLimit:
It is activated when the measuring type is selected as
"Multi Interval" or "Multi Rng". It is the maximum value
that relates to the 2. scale interval (e2).

LowerZoneLimit:
It is activated when the measuring type is selected as
"Multi Interval" or "Multi Partition". It is the maximum
value that relates to the 1. scale interval (e1).

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 95


Filter
It is the menu where the filter settings are made. This menu is an ordered type menu. The sub-menus are
"Mode", "AverageQuantity", "Speed", "Vibration" and "Stability Period".
Mode
It is the filter mode selection. The options are;
1) gResolver: It is the advanced numerical filtering
option. The input signal is processed through advanced
mathematical functions and gives rapid and stable
results.
2) Mov. Average: It is the standard numerical filtering
option. Moving average method is used.
3) No Filter: No filter is applied.

Vibration
It is the selection for the vibration intensity in the
environment. The options are;
1) Low
2) Medium
3) High
Note: When Moving Average is selected no "Vibration" is displayed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 96


AverageCount
The selection for the number data to be used for the
moving average filter. The options are;
1) 8
2) 16
3) 32
Note: When gResolver mode is selected, "AverageQuantity" is not
displayed

Speed
It is the ADC measuring speed. The options are;
1) 6.25 Hz
2) 12.5 Hz
3) 50 Hz
4) 100 Hz
5) 400 Hz
6) 800 Hz

Note: When gResolver mode is selected, "Speed" is not displayed It is detected automatically.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 97


StbltyDelay
The options are;
1) Short (0.5 sec): Current measurement value is
monitored for 0,5 sec.
2) Mid (1.0 sec): Current measurement value is
monitored for 1 sec.
3) Long (2.0 sec): Current measurement value is
monitored for 2 sec.
4) Custom: This option is selected if it is desired to set a stability delay other than these first three
options.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 98


Zero/Tare
It is the menu where the zeroing and taring are made. This menu is an ordered type menu. The sub*menus
are "Mode", "Zero Limit", "Starting Zero" and "Stagnancy".

Mode
The options are;
1) Manual: Zeroing is only made by using key.
2) ZeroTracking: The shifts in the weight value are
zeroed as long as the gross value is zero.
3) PowerOnZero: The option of zeroing at start-up. It
operates with the % ratio selected in "Start-up Zero"
menu. Zero monitoring is also active.
4) FullAutoZero: It is the full automatic zeroing option. Start-up is also active. In case of stagnancy (
), the screen is resetted in the % ratio as selected in "Zero Limit" menu once in every 3 seconds.

Note: ZeroTracking operates only in Scale Program.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 99


ZeroingLimit
1) %1 Capacity: Zeroing is made if the weight value of
less than 1 % of the Capacity.
2) %2 Capacity: Zeroing is made if the weight value of
less than 2 % of the Capacity.
3) %10 Capacity: Zeroing is made if the weight value
of less than 10 % of the Capacity.

PowerOn Zero
Determines at which ratio of the "Capacity" value will the
zeroing be made at first start-up of the device. The
options are;
1) %5 Capacity: Zeroing is made if the weight value of
less than 5 % of the Capacity.
2) %10 Capacity: Zeroing is made if the weight value
of less than 10 % of the Capacity.
3) %25 Capacity: Zeroing is made if the weight value
of less than 25 % of the Capacity.

Note: The start menu is not displayed when "Manuel" mode is selected.
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 100
Stability
The options are;
1) Wait Stable: Stagnancy( ) is necessary for
taring and zeroing. If there is no stagnancy warning
message is displayed on the screen.
2) No Wait: Taring and zeroing is made regardless of
stagnancy.

Tare
Options;
1) On: Taring function is enabled.
2) Off: Taring function is disabled.

Note: Tare menu is displayed only in “Indicator” program.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 101
Calibration
It is explained in detail in Calibration section.
Maintenance
SystemDoctor
System doctor interprets the values obtained from the load cell and provides detailed information on
the ambient vibration, load status and drift.

Detailed information can be seen by pressing


key.

Noise: Provides the ambient vibration.


Drift: Shows that the load is not stable and has the
tendency to dislodge continuously.
Input Saturation: Provides the capacity percentage
of the load cell used.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 102
A/D Count
ADC internal count value is displayed on the screen.
Expresses the numerical equivalent of the signal
connected to the Load Cell module. Count Value
varies depending on the "Load Cell Type" selection.
Maximum values are shown in the table given below.

Load Cell Type Maximum Value


Esit 2.0mV/V 6 400 000
2.5 mV/V …160.0 mV/V 8 000 000

A/D mV/V
ADC mV/V value is displayed on the screen.
Expresses the mV/V value received from the sensor
connected to the Load Cell module. It is used from
controlling the load cell output.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 103
Backup
The Weighing settings and calibration information on ECI can be backed-up to the "Load Cell Module"
in Slot 2. Thus, the previous weighing settings can be recovered anytime by making Back-up >
Recovery operation. Furthermore the "Load Cell Module" which contains the back-up can be inserted
to another ECI and the weighing setting and calibration data can be transferred to another device by
making Back-up > Recovery operation.

Backup
The Weighing settings and calibration
information saved on ECI are backed-up to the
"Load Cell Module".

Restore
The Weighing settings and calibration
information saved on the "Load Cell Module"
are transferred to ECI.

Note: If not Back-up operation is made to the "Load Cell Module"


before, then no "Recovery" operation can be made and relevant message will be displayed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 104
OverLoad
Overload menu provides information on the number
of occurrence of the “Maximum Load” error, its
peak value and average value by interpreting the
values obtained from the load cell.

Counter: It givens the number of “Maximum Load” errors.


Peak Value: It gives the highest weight value causing the occurrence of “Maximum Load” error.
Average Value: It gives the average of the weight values causing the occurrence of “Maximum Load”
error.

Values can be reset by key.

It is possible to return to the Menu with key.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 105
7.3.3 Optional Communication Menu
The menus related to the communication module inserted to slot 3 of the device are displayed under this menu.
If there is not any communication module, this menu is displayed under the name "Empty Slot".
7.3.3.1 Optional Serial Communication Menu
Following menus will appear if COM module is inserted to slot 3.

COM Module
This is the menu where the parameters of the COM module
are set forth.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 106
Settings (In Belt Scale Program)
This menu is a sorted menu, which contains the communication parameters.
Sub menus are “Mode”, “Baudrate”, “Period”, “Modbus Mode” and “Modbus No”.

Mode
It is the communication type selection. The options
are;

1) Flowrate: Current flow rate value is sent once


per second. (Baudrate is 9600bps)

For example;
Flow rate value “214” is shown as below.

Character ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘ 2‘ ‘1‘ ‘4 ‘ CR
Ascii 32 32 32 32 32 32 32 50 49 52 13
Hex 20 20 20 20 20 20 20 32 31 34 0D

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 107
2) Speed: Current belt speed value is sent once per second. (Baudrate is 9600bps) Format of this
data is same with the flow rate data.
3) AutoZero: AutoZero value is sent once per second. (Baudrate is 9600bps) Format of this data is
same with the flow rate data.
4) Load: Load value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow rate data.
5) Total 1: Sum1 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow rate data.
6) Total 2: Sum2 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow rate data.
7) Total: Sum value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow rate data
8) Modbus: Modbus mode.
9) Request: Disables data transaction by serial communication.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 108
Settings (In Dosage Belt Scale Program)
This menu is a sorted menu, which contains the communication parameters.
Sub menus are “Mode”, “Baudrate”, “Period”, “Modbus Mode” and “Modbus No”.

Mode
It is the communication type selection. The options
are;

1) Flowrate: Current flow rate value is sent once


per second. (Baudrate is 9600bps)

For example;
Flow rate value “214” is shown as below.

Character ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘‘ ‘ 2‘ ‘1‘ ‘4 ‘ CR
Ascii 32 32 32 32 32 32 32 50 49 52 13
Hex 20 20 20 20 20 20 20 32 31 34 0D

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 109
2) Speed: Current belt speed value is sent once per second. (Baudrate is 9600bps) Format of this
data is same with the flow rate data.
3) AutoZero: AutoZero value is sent once per second. (Baudrate is 9600bps) Format of this data is
same with the flow rate data.
4) Load: Load value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow rate data.
5) Total 1: Sum1 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow rate data.
6) Total 2: Sum2 value is sent once per second. (Baudrate is 9600bps) Format of this data is same
with the flow rate data.
7) Total: Sum value is sent once per second. (Baudrate is 9600bps) Format of this data is same with
the flow rate data
8) Modbus: Modbus mode.
9) Request: Disables data transaction by serial communication

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 110
Settings (Check Weigher Program)
Mode
Kind of communication is selected by this
parameter.
Options;
1) Last Product: When transition is finished,
Product Weighing is shown. (Baudrate is
115200bps)
For instance;
“120.0”value of display is shown as below table.

Karakter ‘‘ ‘‘ ‘‘ ‘‘ ‘ 0‘ ‘1‘ ‘2 ‘ ‘0 ‘ ‘.’ ‘0 ‘ CR


Ascii 32 32 32 32 32 49 52 56 46 48 13
Hex 20 20 20 20 20 31 32 30 2E 30 0D

2) Total: When the transition is finished, Total product weighing information is sent. (Baudrate is
115200bps) Sending information string format is same as “Last Product” mode.
3) Mean: When transition is finished, Mean product weighing information is sent.
4) Detay: The weight information is sent in detail. (Baudrate is 115200bps). The format of the data
string sent is as follows.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 111
Product Last
Date T Time T T T Total CR
Quantity Product
16.04.2015 * 13:50:19 * 4 * 80 * 300 Chr 13
T : Tab
* : Chr 9

Date Date of transition


Time Time of transition
Product
Total product number
Qantity
Last Product Last product weighing of last transition on band
Total Total product weighing of transition on band

5) Command: Functions as a query from a remote computer or similar device (Baudrate is 9600bps)
chr 13 and chr 10 are sent neccesary where in the end of sending data.
6) Modbus: It is the Modbus operation mode selection.
7) Service: While product is on weighing band, weighing information is send continuously.
8) Off: Data sending over serial port is disabled

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 112
Maintenance
It is used for understanding whether the communication
port is functional or not. The data received from the serial
port is displayed at the receiving section. "ESIT" wording
is sent to the serial port at the same time.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 113
7.3.3.2 Optional Profibus Communication Menu
Following menus will appear if Profibus module is inserted to slot 3.
ProfiBus Modul
This is the menu where the parameters of the Profibus module are set forth.

ProfiBus Adr
The address number to be used by the device for the
profibus communication.

Maintenance
The screen where the profibus communication settings
are displayed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 114
7.3.3.3 Optional Ethernet IP Communication Module
Following menus will appear if Ethernet IP module is inserted to slot 3.
Ethernet IP
This is the menu where the parameters of the Ethernet IP module are set forth.

IP Address
The address number to be used by the device for the Ethernet IP communication.
IP.xx.xx.xx
1. part of the IP address. Value between 1-255 can be
entered.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 115
xx.IP.xx.xx
2. part of the IP address. Value between 1-255 can be
entered.

xx.xx.IP.xx
3. part of the IP address. Value between 1-255 can be
entered.

xx.xx.xx.IP
4. part of the IP address. Value between 1-255 can be
entered.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 116
Maintenance
The screen where the Ethernet IP communication
settings are displayed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 117
7.3.3.4 Optional Profinet Communication Module
Following menus will appear if Profinet module is inserted to slot 3.
Profinet
This is the menu where the parameters of the Profinet module are set forth.

IP Address
The address number to be used by the device for the Profinet communication.
IP.xx.xx.xx
1. part of the IP address. Value between 1-255 can be
entered.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 118
xx.IP.xx.xx
2. part of the IP address. Value between 1-255 can be
entered.

xx.xx.IP.xx
3. part of the IP address. Value between 1-255 can be
entered.

xx.xx.xx.IP
4. part of the IP address. Value between 1-255 can be
entered.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 119
Maintenance
The screen where the Ethernet IP communication
settings are displayed.
Station name of the Profinet module is created from 4th
part of IP adress. Screenshot on the side, as 4th part of
IP adress is 100 thus station name is “ECI100”.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 120
7.3.4 Optional In/Out Menu
The menus related to the In/Out module inserted to slot 4 of the device are displayed under this menu. If there
is not any In/Out module, this menu is displayed under the name "Empty Slot".
7.3.4.1 Optional Relay Menu
Following menus will appear if Relay module is inserted to slot 4.
Relay Module
Display Program
This is the menu where the parameters of the Relay module
are set forth.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 121
Outputs (BeltScale Program)
This menu contains the relay settings.
Relay 1
This menu contains Relay1 settings.
Pulse Out
Pulse outputs value for the mechanical counter is
determined in this menu. When the desired
amount of product passed through the belt,
relevant relay gets energized for 0.5 seconds and
than goes back to its previous state.
The value of the mechanical counter connected to
relay output is raised by one. If the mechanical
counter won’t be used, this value is set to zero in order to point out that the system is ready to work.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 122
Relay 1 Function

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 123
Relay2
This is the menu where the relay settings are made. Moreover, relay3 and relay4 settings are made
as shown below

Mode
It is the relay operation mode selection. Options;
1) Flowrate: Operates according to the flowrate
value.
2) Load [kg/m]: Operates according to the load
value.
3) Error Signal: The relay contacts are short-circuited when one of the following error conditions
occurs, otherwise it is open circuit.
Error conditions:
LoadCellError (error in the sensor connected to the load cell)
Maximum Load! (The status when measured value exceeds the defined Maximum Capacity)
ADC Data Error! (Load Cell Module Error)
No Load Cell Module! (No Load Cell Module Identified Error)
4) Ready Signal: The output is given as long as the device is in operation.
5) Speed: Operates according to the speed value.
6) Batch: Operates according to the quantity value. It is activated with Start input. It goes into passive
state when the specified quantity of products passes or when the stop input is activated.
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 124
Minimum
It is active if“Flowrate” ,” Load [kg/m]” or “Speed”
mode is selected. It is the lowest value which the
relay contact is active.

Maximum
It is active if“Flowrate” ,” Load [kg/m]” or “Speed”
mode is selected. It is the top value which the relay
contact is active.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 125
Delay
Determines how long the relay contact will change
state after the desired state has occurred.

Contact Type
Determines the start-up position of the relay
contacts. Options;
1) Norm. Open: The relay contact is open circuit at
start-up.
2) Norm. Closed: The relay contact is short circuit
at start-up.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 126
Contact Status
u
Mode: Flow, Load
Contact Type: Norm.Open
Closed Delay Delay

Open System Speed


ed Minimum Maximum

Contact Status

Closed Mode: Flow, Load


Contact Type: Norm.Closed
Delay Delay
Norm.Closed
Open Screen Value (Weight)
ned Minimum Maximum
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 127
Deafult mode in Relay 2 is selected as “Batch”. Relay2 works as shown in the following graph at this
mode.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 128
Relay3
This is the menu where Relay3 settings are done. Default mode is selected as “Speed”. Relay works as
shown in the following graph at this mode.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 129
Relay4
This menu contents Relay4 settings. Default mode is selected as “Error Sign”.
Outputs (Dosage Belt Scale Program)
This menu contains the relay settings.
Relay 1
This menu contains Relay1 settings. Below modes are applicable to Relay2, Relay3, Relay4
Mod
1) Flow: Relay is on according to flow rate.
2) Load [kg/m]: Relay is on according to load.
3) Error Signal: In case below, relay is shorted,
otherwise on;.

Loadcell Error
Maximum Load
ADC Data Error
Loadcell card error
4) Ready Signal: Always on while system is working.
5) Speed: Relay is on according to speed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 130
6) Batch Value: Relay is on according to batch value. Set On after Start. Passive only in case of
STOP or Target Value Reached.
7) Start: ON if 24V is supplied and no Belt Slide Error.
8) Pulse Out: Pulse out each maximum value reached and releases in 0,5 sec.
9) Motor Check: ON till motor stops.

Relay2
Preset is START
Relay3
Preset is ERROR SIGNAL
Relay
Preset is READY.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 131
Minimum
It is active if“Flowrate” ,” Load [kg/m]” or “Speed”
mode is selected. It is the lowest value which the
relay contact is active.

Maximum
It is active if“Flowrate” ,” Load [kg/m]” or “Speed”
mode is selected. It is the top value which the relay
contact is active.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 132
Delay
Determines how long the relay contact will change
state after the desired state has occurred.

Contact Type
Determines the start-up position of the relay
contacts. Options;
3) Norm. Open: The relay contact is open circuit at
start-up.
4) Norm. Closed: The relay contact is short circuit
at start-up.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 133
Contact Status
u
Mode: Flow, Load
Contact Type: Norm.Open
Closed Delay Delay

Open System Speed


ed Minimum Maximum

Contact Status

Closed Mode: Flow, Load


Contact Type: Norm.Closed
Delay Delay
Norm.Closed
Open Screen Value (Weight)
ned Minimum Maximum
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 134
Deafult mode in Relay 2 is selected as “Batch”. Relay2 works as shown in the following graph at this
mode.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 135
Pulse Mode Relay:

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 136
This is the menu where Relay3 settings are done. Default mode is selected as “Speed”. Relay works as
shown in the following graph at this mode.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 137
Maintenance
Inputs
On this menu, Input statements are hounded and
controlled as of momentary. If the applied input
signal is 12-24 V DC,”1” statement is shown that on
display otherwise “0” statement is shown that on
display

Outputs
On this menu, relays are pulled to in order when the
button is pressed. If the relay is pulled, shown
that”1”character as on picture otherwise shown
that”0” character.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 138
7.3.5 Optional Input Output Menu
On this menu,Input/Output module where connected to Slot.5, corresponding of menu is shown that on
display.If any of Input/output module is not exist, “Slot Empty” information is shown that on display
7.3.5.1 Optional DAC Menu
If the DAC module is mounted on Slot.5, This screen will be shown as below.

DAC Module

DAC module is setted by this parameters on


shown that display.

Note: This module doesn’t use at batching and filling program. (shown that locked)

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 139
Analogue Output (BeltScale Program)
This menu contains the Analogue Output module settings.
Voltage Output : between- 0,500V / 10.500V range. (in V1.2 cards: - 0,500V / 12V)
Current Output : between 0mA / 20mA range. (in V1.2 cards: 0mA / 24mA)

Vmax>Vmin, Wmax>Wmin or Vmax>Vmin, Wmax<Wmin or


Vmax<Vmin, Wmax<Wmin Vmax<Vmin, Wmax>Wmin

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 140
Mode
1) Flowrate: Analog output operates according to the
flowrate value.
2) Speed: Analog output operates according to the
speed.

Flowrate Low:
Sets the flow rate minimum value.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 141
Flowrate High:
Sets the flow rate maximum value.

Note: When Speed is selected in the Mode menu, the Flow Min and Flow Max menus will not appear.

Minimum Speed:
The speed value corresponding to the initial value of
the analogue output.
For example, if it is desired to alternate the analog
output voltage between 10 m/s and 20 m/s, 10 is
entered here.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 142
Maximum Speed:
The speed value corresponding to the finishing value of
the analogue output.
For example, if it is desired to alternate the analog
output voltage between 10 m/s and 20 m/s, 20 is
entered here.
.
Note: If Flowrate is selected on the Mode menu, the Minimum Speed and Maximum Speed menus will not appear.

Reference:
1) Voltage: Analogue output shall be adjusted
according to the voltage.
2) Current: Analogue output shall be adjusted
according to the current value.

MinimumVolt.:
It is the start value of the Analogue Output.
For example if it is desired that the Analogue output
voltage varies between 0 V and 10 V, the value 0 V is
entered here.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 143
MaximumVolt.:
It is the end value of the Analogue Output.
For example if it is desired that the Analogue output
voltage varies between 0 V and 10 V, the value 10 V is
entered here.

Note: If Current is selected in the reference menu, “MinimumVoltage” and “MaximumVoltage” menus are not displayed.
Minimum mA:
It is the start value of the Analogue Output.
For example if it is desired that the Analogue output
current varies between 4mA and 20mA, the value
20mA is entered here.

Maximum mA:
It is the end value of the Analogue Output.
For example if it is desired that the Analogue output
current varies between 4mA and 20mA, the value
20mA is entered here.
Note: If Voltage is selected in the reference menu, “Minimum mA”
and “Maximum mA” menus are not displayed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 144
Analogue Output (Dosage Belt Scale Program)
It is used to adjust belt speed.
Belt Speed is integrated to PID Analogue output is locked.
Output Voltage : 0V - 10V

Digital Output
The section of the DAC module which contains the settings related to the Digital Output. This menu is an
ordered type menu. Sub menus are “Mode”, “Minimum”, “Maximum, “Delay” and “Contact Type”.
Mode
It is the output operation mode selection. The
options are;
1) Disabled: The relay is not active.
2) Net Value: Output is given between the minimum
and maximum values of the net value, otherwise no
output is given.
3) Gros Value: Output is given between the minimum and maximum values of the gross value,
otherwise no output is given.
4) Ready Sign: Output is given as long as the device is in operation.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 145
5) Error Sign: Output is given when any of the error situations stated below emerges, otherwise no
output is given. Error situations:
LoadCellError LoadCellModule reading error!
Maximum Load! LoadCell Module Hardware error!
6) Stable Sign If the change in the measurement result is less than a scale interval (if symbol is
displayed on the screen), output is given, otherwise (if symbol is displayed on the screen) no
output is given.
7) Tare Sign: While taring (when symbol is on the scree) output is given. When tare is removed
(when symbol is on the scree) no output is given.

Minimum
It is active when Net Value or Gross Value mode is
selected. It is the lowest value of the weight range
where the output is active.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 146
Maximum
It is active when Net Value or Gross Value mode is
selected. It is the highest value of the weight range
where the output is active

Delay
Determines the period after which the output will be
activated after the desired situation occur.

Output Type
Determines the status of the digital output signal at
start-up. The options are;
1) Norm. Open: There is digital output signal at
start-up.
2) Norm. Closed: There is no digital output signal at
start-up.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 147
DigitalInput
Input1
1) Deactivated: The input is not active.
2) T1 Zero: When the input signal is detected, the
device performs a reset operation.

Input2
1) Deactivated: The input is not active.
2) .T2 Zero: When the input signal is detected, the
device performs a reset operation.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 148
Maintenance
Inputs
Real time status of the inputs can be monitored from
this menu. The inputs that are applied 12-24 V DC
are displayed as 1 and other are displayed as 0.

Digital Output
In this menu,If button is pressed,output is
activated as periodic to control when the output is
activated,”1” is shown on display otherwise “0” is
shown on display.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 149
Analog Output
In this menu,When the button is pressed,output
is changed to control as periodical between 0 to
10V.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 150
7.3.5.2 Optional Relay Menu
If the RELAY module is inserted in the slot 5, following menus will be displayed.

Relay Module
This is the menu where the parameters of Relay Module are
determined.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 151
Outputs(Belt Scale Program)
Relay1
Relay1 is locked.
Relay2
This is the menu where the Relay2 settings are made. It is the same as Slot 4 relay menu settings.
Relay3
This is the menu where the Relay3 settings are made. It is the same as Slot 4 relay menu settings.
Relay4
This is the menu where the Relay4 settings are made. It is the same as Slot 4 relay menu settings.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 152
Outputs(Dosage Belt Scale Program)
Relay1
Relay1 is locked.
Relay2
This is the menu where the Relay2 settings are made. It is the same as Slot 4 relay menu settings.
Relay3
This is the menu where the Relay3 settings are made. It is the same as Slot 4 relay menu settings.
Relay4
This is the menu where the Relay4 settings are made. It is the same as Slot 4 relay menu settings.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 153
Maintenance
Inputs
On this menu, Input statements are hounded and
controlled as of momentary. If the applied input
signal is 12-24 V DC,”1” statement is shown that on
display otherwise “0” statement is shown that on
display

Outputs
When button is pressed in this menu output is
activated periodically to ensure controlling. 1 is
displayed when the output is active and 0 is
displayed otherwise.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 154
7.3.6 Optional Expansion Menu
The menus related to the expansion module inserted to slot 6 of the device are displayed. EXP Module or BCD
Module can be inserted to this slot. If there is not any expansion module, this menu is displayed under the
name "Empty Slot".
7.3.6.1 Optional BCD Module Menu
Following menus will appear if BCD module is inserted to
slot 6.
BCD Module
This is the menu where the parameters of the BDC module
are set forth.
Mode
1) BCD Out: BCD (Binary Coded Decimal) is the binary
coding of the numbers in Decimal number system 41/2
digit (39999) output data is coded.
2) Binary 2’ s: BINARY'2 is the integrated output in
binary number system. 1 sign + 19 bit (±524.287) output data is provided.
3) Binary Abs.: It is the absolute output coded in binary number system. 1 sign + 19 bit ( 524.287) output
data is provided. Binary functions in the same way with the complement.
Note: This module is not used in batching and filling programs (displayed locked).

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 155
BCD Output:
12495 kg -1780 kg ERROR
SIGN 0 1 1
D18 0 0 0
D17 0 0 0
D16 1 0 0
D15 0 0 0
D14 0 0 0
D13 1 0 0
D12 0 1 0
D11 0 0 0
D10 1 1 0
D9 0 1 0
D8 0 1 0
D7 1 1 0
D6 0 0 0
D5 0 0 0
D4 1 0 0
D3 0 0 0
D2 1 0 0
D1 0 0 0
D0 1 0 0

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 156
Binary 2’ All
12495 kg -1780 kg ERROR
SIGN 0 1 1
D18 0 0 0
D17 0 0 0
D16 0 0 0
D15 0 1 0
D14 0 1 0
D13 1 1 0
D12 1 1 0
D11 0 1 0
D10 0 0 0
D9 0 0 0
D8 0 1 0
D7 1 0 0
D6 1 0 0
D5 0 0 0
D4 0 0 0
D3 1 1 0
D2 1 1 0
D1 1 0 0
D0 1 0 0

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 157
BinaryAbsolute
12495 kg -1780 kg ERROR
SIGN 0 1 1
D18 0 0 0
D17 0 1 0
D16 0 1 0
D15 0 0 0
D14 0 0 0
D13 1 0 0
D12 1 0 0
D11 0 0 0
D10 0 1 0
D9 0 1 0
D8 0 0 0
D7 1 1 0
D6 1 1 0
D5 0 1 0
D4 0 1 0
D3 1 0 0
D2 1 1 0
D1 1 0 0
D0 1 0 0

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 158
Maintenance
This menu is used for making the control and testing of the inputs and outputs of the "BCD Module".

Inputs
The status of the inputs can be monitored real time
from this menu. The inputs that are applied 12-24 V
DC are displayed as 1 and others are displayed as
0.

Outputs
When key is pressed in this menu, one of the
outputs activate while the other remain inactive. The
active out is displayed as 1 and others are displayed
as 0.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 159
7.3.6.2 Optional EXP Module Menu
Following menus will appear if EXP module is inserted to slot 6.
EXP Module
Outputs
It is the menu where the Digital Output settings are made. Same as the relay module output menu.
Output1
.
Output16
Inputs
Same as the relay module input menu.
Input1
.
Input8
Maintenance
Same as the relay service input menu.
Inputs
Outputs
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 160
7.4 Zeroing and Tare Operations
7.4.1 Zeroing
Used for zeroing the indication value.
7.4.1.1 Manuel Zeroing:
key is used in order to make zeroing operation within the limits in Zero/Tare menu.
7.4.1.2 Auto Zeroing
Zeroing is made at start-up. Furthermore in case of stagnancy ( ), the screen is resetted in the % ratio
as selected in "Zero Limit" menu once in every 3 seconds.Other resetting options are explained in Zero/Tare
menu.
7.4.2 Tare
7.4.2.1 Manuel Tare
key is used for taring operation. After taring, the load with the weight of at least 20*e (Resolution) is
left on the weighing platform and if any stagnancy is observed, the tare is automatically cancelled when the
gross value is 0.
7.4.2.2 Auto Tare
There is no auto Tare function in ECI device.
Note: key is not displayed if Tare “Off” is selected from Zero/Tare menu.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 161
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 162
8 Calibration
8.1 Load Calibration
It is the operation of identification of a mass with a known weight to the device. It must be made by using real
weights.

1) It is entered to menu 2.5 for the load calibration.

2) Load cell type is selected. The options are;


1) Esit (2 mV/V)
2) 2.5mV/V
3) 5.0mV/V
4) 10.0mV/V
5) 20.0mV/V
6) 160.0mV/V
mV/V selection is made depending on the load cell model.

3) For entering to Calibration Type menu is pressed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 163
4) Calibration Type is selected. The options are;
1) Load Calibration
2) mV/V Calibration
3) Coefficient Calibration
4) Zero Calibration

5) After selecting the Calibration Type as the "Load


Calibration, proceeded to the next screen by pressing first,
then key. The information screen shown at side is
displayed.

If you select the cell type wrong you can return the previous screen by
pressing key.

6) key is pressed for starting the zeroing operation.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 164
7) After the zeroing operation finishes, the result of the
zeroing will be displayed on the screen. If you want to repeat
the zeroing operation press key.Press key to
continue.

8) Zeroing operation is completed. Press key to


continue the operations.

9) In order to proceed to the screen where the calibration


weight value to be entered key is pressed.
For a proper calibration operation, the load used must not be less
than the amount recommended on the screen.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 165
10) The weight value to be used in calibration is stated in
this screen. For confirming the weight value entered
key is pressed and proceeded to the calibration screen.

Approved and certified weights must be used for a healthy calibration.

11) By placing the load equal to the calibration weight stated


in the message screen and pressing the key, the
calibration process is initiated.

The screen shown at side is displayed during the calibration.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 166
12) When the calibration operation is completed, the
calibration result is displayed. If you want to repeat the
calibration operation press key. Press key to
continue.

13) General information screen indicates that the calibration


is completed. Key is pressed for proceeding to the next
screen.

14) key is pressed to store the calibration values or


is pressed to exit from the menu without storing the
calibration values.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 167
8.2 mV/V Calibration
It is the calibration operation made by using the load cell values without any loading. Following operations are
carried out in order.
1) It is entered to menu 2.5 for the mV/V calibration.

2) Load cell type is selected. The options are;


1) Esit(2 mV/V)
2) 2.5mV/V
3) 5.0mV/V
4) 10.0mV/V
5) 20.0mV/V
6) 160.0mV/V

3) For entering to Calib. Type menu key is pressed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 168
4) It is the Calibration Type selection. The options are;
1) Load Calibr.
2) mV/V Calibr.
3) CoeffCalibr.
4) Zero Calibr.
“mV/V Calibration” is selected and key is pressed.

5) mV/V calibration is made according to the total load cell


capacity of the system. The information screen indicating this
is shown. Key is pressed to proceed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 169
6) Total load cell capacity is entered. If there are more than
one load cells in the system, the total value of the capacities
of the load cells are entered. To proceed press Key.

7) mV/V calibration is made according to the mV/V value of


the average load cell of the system. The information screen
indicating this is shown. Key is pressed to proceed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 170
8) The average mV/V value of the system is entered. If one
load cell is used in the system, then the certificate value is
valid. Otherwise the average of the load cell certificate
values must be taken. For confirming the value entered
press key.

9) An information screen indicating that the calibration has


been completed appears. key is pressed to proceed to
the next screen.

10) Key is pressed to store the calibration values or


is pressed to exit from the menu without storing the
calibration values.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 171
8.3 Coefficient Calibration
1) It is entered to menu 2.5 for the coefficient calibration.

2) Load cell type is selected. The options are;


1) Esit(2 mV/V)
2) 2.5mV/V
3) 5.0mV/V
4) 10.0mV/V
5) 20.0mV/V
6) 160.0mV/V
3) For entering to Calibration Type menu key is pressed.
4) Coefficient Calibration is selected as the Calibration Type.
The options are;
1) Load Calibr.
2) mV/V Calibr.
3) CoeffCalibr.
4) Zero Calibr.
5) To proceed press Key.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 172
For skipping the information screen appeared key is
pressed.

6) Coefficient value is entered. For confirming the value


entered press key.
For example;
Load Cell Type: is selected as 2.0 mV/V and if the load cell
of 10 kg with the output of 2.0 mV/V is fully loaded, the
screen value will be 10000 for the coefficient=1,562500
Other sample values will be as shown in the following table

Coefficient Scale Interval (e) Screen Value


1.000000 1 6400
1.000000 1.0 6400.0 𝑆𝑐𝑟𝑒𝑒𝑛 𝑉𝑎𝑙𝑢𝑒(𝐹𝑢𝑙𝑙𝑦 𝐿𝑜𝑎𝑑𝑒𝑑)
1.000000 1.000 6400.000 𝐶𝑜𝑒𝑓𝑓𝑖𝑐𝑖𝑒𝑛𝑡 =
6400
0.250000 1.0 1600.0
1.562500 1 10000

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 173
7) key is pressed to finish the calibration.

8) Key is pressed to store the calibration values or


is pressed to exit from the menu without storing the
calibration values.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 174
8.4 Zero Calibration
The following operations must be performed respectively for the zero calibration.
1) It is entered to menu 2.5 for the zero calibration.
2) Load cell type is selected. The options are;
1) Esit(2 mV/V)
2) 2.5mV/V
3) 5.0mV/V
4) 10.0mV/V
5) 20.0mV/V
6) 160.0mV/V
3) For entering to Calibration Type menu key is pressed.
“Zero Calibration” is selected as the calibration type.
The options are;
1) Load Calibr.
2) mV/V Calibr.
3) CoeffCalibr.
4) Zero Calibr.

4) For proceeding to the zeroing screen key is pressed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 175
5) Weighing platform is emptied and Key is pressed for
starting the zeroing operation.

The screen shown at side is displayed during the zeroing operation.

6) After the zeroing operation finishes, the result of the zeroing


will be displayed on the screen. If you want to repeat the zeroing
operation press key. Press key to continue.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 176
7) Key is pressed to proceed to the calibration confirmation
screen.

8) Key is pressed to store the calibration values or is


pressed to exit from the menu without storing the calibration
values.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 177
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 178
9 Software Update
EECI, provides software update support by using USB disk.

Warning: You may lose your existing settings after the device software update. Thus please review the changes
brought by the update carefully and store/note old settings if necessary. The outputs (Relay, input etc) connected to the
device may behave different as the routine operation of the device will be disabled during the software update. Thus it is
recommended to disconnect or shutdown the connectors connected to the device.

The following steps must be followed in order for the software update;

1) The "image.s19" file to be installed is transferred to an USB disk that


is formatted as FAT32.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 179
2) USB disk is inserted to the USB slot of ECI.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 180
3) When the device is supplied power by keep pressing to the calibration key by means of a clip or stable, the
software update operation will start.

USB Input

Calibration Key

The LCD screen will be blank when the update starts. LCD light will blink is short intervals and sound alarm will be
heard at the same time. Do not cut the power of the device and wait during the update. When the operation is
completed the device will start operating automatically.

Note: v1.2 and older versions are not suitable for current USB update operations.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 181
4) The version number is the start screen must be control in order to verify that the installation after the update
is performed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 182
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 183
10 Annex 1: Optional Modules
ECI is a device with Smart "Plug-Play" Module structure and optional modules can be changed easily thanks to this
feature.

10.1 Slot3 (Optional Communication Slot):


10.1.1 E-MDL-COM (Serial Communication Module):
This module is an galvanic insulated serial communication module. RS232/R485 or RS422 communication
mode can be selected with the selection switch on it. Maximum Baudrate is 115200bps.

“RS232 & RS485”-“RS422” Selection Switch


Note: Before making the connections check “RS232 & RS485”-“RS422” Selection Switch of the respective module.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 184
10.1.1.1 Connection Example (RS232)

9 pin female connector

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 185
10.1.1.2 Connection Example (RS485)

RS485 Device

Note: In RS485 connection, if device is the first or the last in the communication line, put a resistor(120 ohm) between A
and B ports.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 186
10.1.1.3 Connection Example (RS422)

RS422 Device

Note: In RS422 connection, if device is the first or the last in the communication line, put a resistor (120 ohm)between
A,B and between Rx, Tx ports.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 187
10.1.2 E-MDL-PRF (Profibus Module)
10.1.2.1 Connection Example

Female Connector

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 188
10.2 Slot4 (Optional Input/Output Menu):
E-MDL-RY or E-MDL-DAC modules can be connected to Slot4. Detailed information about these modules are given
below.
10.2.1 E-MDL-RLY (Relay Module):
This is an input -output module with 4 digital inputs (Optically insulated 12-24 V DC) and 4 relay outputs
(250V/2A).
10.2.1.1 Terminal Explanations

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 189
10.2.1.2 Connection Example

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 190
10.2.2 E-MDL-DAC (Analogue Output Module):
This is a module with 1 analogue output (4-20 mA or 0-10 V), 2 Digital inputs (12-24 V DC) and 1 opto-coupler
output (maximum 100 mA).

10.2.2.1 Terminal Explanations

Note: In v1.2 boards, no external feeding connection (pin no 9 and 10) is necessary.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 191
10.2.2.2 Connection Example (Digital Input-Output)

Load External Supply

External Supply

Digital Input Connection Digital Output Connection (Open Collector)

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 192
10.2.2.3 Connection Example (Analogue Output)

0-20 Ma Analogue Output

0-10 V Analogue
Output

External Supply

External Supply

Analogue Output Connection (0-10 V) Analogue Output Connection (0-20 mA)


Note: In v1.2 boards, no external feeding (pin no 9 and 10) is necessary
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 193
10.3 Slot5 (Optional Input/Output Menu)
E-MDL-RY or E-MDL-DAC modules can be connected to Slot5. Module explanations are as shown in Slot4.

10.4 Slot6 (Optional Expansion Slot)


E-MDL-EXP or E-MDL-BCD modules can be connected to Slot6. The connections are given below in detail.
10.4.1 E-MDL-BCD (Binary Output Module)
This is a BCD or Binary output (19 + 1 Bit) module.
The outputs are active when the pin no 11 (VCC_INN) and pin no 24 (OUT_EN) are short circuited. If the user
wants to receive data continuously, the pins 11,24 are short circuited. No data transfer occurs as long as the pin
no 11 is empty.
-0 (minus zero) value is sent from Binary and BCD outputs before the measurement and in cases of errors.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 194
10.4.1.1 Connector Explanations
Output Connector

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 195
10.4.1.2 Connections

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 196
10.4.2 E-MDL-EXP (Relay Expansion Module)
It is used for providing output to the external relay card. It is used with Es Relay card. 16 relay outputs (10 A 250
V) and 8 digital input signal control can be provided with this module.2 es relay cards are necessary for 16
outputs and 8 inputs.
10.4.2.1 Connector Explanations
Output Connector

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 197
10.4.2.2 Connections

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 198
10.4.2.3 Es Relay Card

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 199
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 200
11 Annex 2 Belt Scale Program
Electronic conveyor belt scale system enables the weighing of bulk materials carried by conveyor belts during the
conveying process. The flow amount (capacity) is proportional to the speed and the material on the belt.

11.1 Specifications
 High internal resolution (24 bits)
 Digital filter
 3 total values (Total1 and Total2 are resettable . Total is non-resettable)
 Actual Flowrate t/h, belt speed, kg per meter values can be displayed
 Internal RTC and battery circuitry to hold Totalizing data
 Belt working simulator
 Alibi Memory

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 201
11.2 Main Program Screen
Beltscale program main screen shot will be as shown at side;

Note: If ECI is not in “Beltscale” program (If the main screen above is not displayed), You can refer to section 7.1
Program Selection for shifting to the “Beltscale” program.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 202
11.2.1 Screen Layout

1) Totalized weight
2
2) Total indicator index 1
3) Unit 3
4) Belt Motion Simulator 5 1
7
5) Flow percentage, Min flow, Max flow indicator 4
6 8
6) Zero Zone indicator
9
7) Flowrate ,Auto zero or Weight
8) Speed, load per meter
9) Keys

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 203
11.2.2 Keys
Reset Total Key Used to reset Total1 and Total2 values (T1 ve T2)
Used to swich between Total1, Total2 and general Total
Change Total Index Key indications for totalization indications. And also the
measurement values shown on the 7 and 8 area are toggles.
x10 Resolution Key Used to activate x10 resolution for Totalize indications
It is used for bringing the menu where the settings related to
BeltScale Menu Key
the beltscale are made.
Used for bringing the menu screen where the device settings
Setting Menu Key
are made.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 204
11.2.3 Symbols
Belt Animation symbol It is used for simulating the belt conveyor is running.

Belt Animation symbol It is used for simulating the belt conveyor is stopped.
This symbol appears on the screen when the measured
Minimum Flowrate symbol
flowrate value is under the “Minimum Flowrate”.
This symbol appears on the screen when the measured
Maximum Flowrate symbol
flowrate value is above the “Maximum Flowrate”.
This symbol appears on the screen while AutoZero
process is in operation. The symbol is not shown, when
Auto Zero Operation symbol
Autozero is Off state or measured weight above 2% of the
“Maximum Capacity” or Interlock input is not valid.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 205
11.3 Beltscale Menu

1. Initial Setup 2. Coefficients 3. ZeroingPasswd 4.Report


Singed Total Min.Flowrate Total1 Records
Weight Length Coefficient1 Total2 Period.Report
Speed Sensor Max.Flowrate
Pulse Distance Coefficient2
Sensor Timeout
Belt Length
Auto Zero
Belt Oscillation

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 206
11.4 Beltscale Menu Usage
Initial Setup
It is an ordered type menu that contains the Initial Setup settings for the beltscale program. The sub-menus are
the Signed Total, Weight Length, Speed Sensor, Pulse Distance and Sensor Time Out.
Signed Total
The options are;
1) Standard: Negative loadings are added to totalization.
2) PositiveOnly: Negative loadings are not added to
totalization.
Note: This menu is not shown when Security Type is selected as
“LegalForTrade” (It is fixed Standart mode).

WeightLength
Platform measurement area. Length of the
measurement platform. Sum of the half distance between
the left roller and weighing roller (a) and half distance
between right roller and the weighing roller
(b).WeightLength= a/2 + b/2

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 207
Speed Sensor
Selection of the Input type selection for speed
measurement. The options are;
1) Encoder: Encoder is used for speed measurement.
2) Fixed Speed: Used for fixed speed systems.

Note: The valid signal has to be applied to the speed sensor input (Input1) to notify that belt is running.

Pulse Distance
Distance per pulse value. In this parameter the distance for
every pulse is defined in “mm”.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 208
Sensor TimeOut
Time out information for speed sensor pulse intervals. If the
belt turning starts and stops continuosly, itis hard to
estimate that the belt is stopped. This parameter
determines the accceptable time period of pulse
inputs.

(Note: If you are not sure about the value, please enter 1.00 second. If something wrong with speed please enter 2.50
second)

Design Speed
It contains belt speed info. System speed at a constant
speed system is determined by this parameter.

Note: This parameter is not shown when the SpeedSensor parameter is selected as “Encoder”.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 209
Belt Length
Total length of the belt.

Note: This parameter used for calibration and autozero calculations. Please enter correctly.

Auto Zero
It contains auto zero selection. The options are;
1) Off: Auto zero is disabled.
2) On: Auto zero enabled.

Note:If the “LegalForTrade” option is selected in the “Security Type” menu, this option is not shown and AutoZero option
is fixed On state.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 210
Belt Oscillation
It is the zero belt value. The weight is not added to total if it
is below this value.

Debi Filtre
Debi değeri için Standart sayısal filtreleme seçeneğidir.
Hareketli ortalama metodu kullanılır.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 211
Coefficients
Belt multipliers defined in the section. According to uncertanity of the Weight Length and Pulse Distance, the
multipliers has to be defined to correct the system.

Note: This submenus are locked when "LegalForTrd" is selected in the "Security Type" menu.

Min.Flowrate
Minimum flowrate value. symbol appears on the
screen when the flowrate is less then the minimum flow
rate.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 212
Coefficient1
Multiplier for minimum Flowrate condition.

Max.Flowrate
Maximum flowrate value. symbol appears on the
screen when the flowrate is more then the maximum flow
rate.

Coefficient 2
Multiplier for maximum Flowrate condition.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 213
ZeroingPasswd
Passwords settings menu.

Total1
Used to clear “Total1” value.
This value shown as 0 when entered to this parameter
screen. And user can change it everytime. When entered
on mainscreen correctly the totalized value is cleared.

Total2
Used to clear “Total2” value.
This value shown as 0 when entered to this parameter
screen. And user can change it everytime. When entered
on mainscreen correctly the totalized value is cleared.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 214
Report
This is the menu where you can see the ALIBI records or
take a new ALIBI record for the belt scale program.
Records
Record Number Date/Time Total and Status values can
be seen. Use and keys to move between
records.

To acces the desired record, use key.

For taking a record manually, use key.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 215
Report Period
In this menu, the period for the USB records is determined.
The options are;
1) Off: No USB record is taken.
2) 1 hour: USB records period.
3) 6 hours: USB records period.
4) 12 hours: USB records period.
5) 24 hours: USB records period.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 216
11.5 Alibi Memory
All information and other pertinent information is stored in the Real-Time Clock unit on the device. To keep this
information in case of a power failure, there is a battery circuit in the ECI device. Information regarding the device's
production date and calibration date is stored in the EEPROM memory. During calibration, the new calibration date
will be automatically saved to permanent memory.

ECI device can store the last 1000 totalization records (minimum flow, maximum flow and memory error conditions)
and error log. An Alibi record contains the following information:
Date + Time + Total + Status

Status
Manual Record: With button, a manual record can be taken.
Minimum Flow: When the measured Flow value is lower than the “Minimum Flow” value in the Coefficients menu,
that info gets recorded automatically.
Maximum Flow: When the measurd Flow value is higher than the “maximum Flow” value in the Coefficients menu,
that info gets recorded automatically.
Memory Error: When the system cant acces the memory, that info gets recorded automatically.
Period. Report: Info gets recorded periodically, according to the period time choosen by the user.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 217
11.6 Auto Zero
Autozero feature can be enabled by setting the
“Auto Zero” parameter in “Initial Setup” menu or
in “Security Type” menu, “LegalForTrade” option is choosen.
Additionally the “AutoZero Interlock input” has to be activated to
validate the zeroing process.
Zeroing process is explained below;
 System awaits until the measured weight value falls under the
%2 of the capacity. If the measured weight value stays under
%2 of capacity for a period of belt length and sign blinks. Then auto zero process starts.
 Auto zero process continues for 3 belt length period, meanwhile sign shown permanently, if the
measured weight value gets out of range, auto set proces gets cancelled.
 At the end of Auto zeroing process, device waits for a period of belt once more, it is expected to see that the
measured weight value to measured under the%2 of the capacity, in the meanwhile, sign blinks. When the
proces is done, zero value on screen is updated. If the measured weight value gets out of the range before a
period of belt is completed, auto zero proces gets cancelled.

Note: Auto zero value can be displayed on the screen. The Picture above shows the position.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 218
11.7 Calibration of BeltScale
11.7.1 Load Calibration
It is the operation of identification of a mass with a known weight to the device. It must be made by using real
weights. Calibration must be done while belt is running. Device automatically calculates belt movement
distance.

1) Enter the menu 2.5 for the calibration.

2) Load cell type is selected. The options are;


1) Esit (2 mV/V)
2) 2.5mV/V
3) 5.0mV/V
4) 10.0mV/V
5) 20.0mV/V
6) 160.0mV/V
mV/V selection is made depending on the load cell model.

3) For entering the Calibration Type menu is pressed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 219
4) Calibration Type is selected. The options are;
1) Load Calibr.
2) mV/V Calibr.
3) CoeffCalibr.
4) Zero Calibr.

5) After selecting the Calibration Type as the "Load


Calibration”, proceeded to the next screen by pressing first,
then key. The information screen shown at side is
displayed.

If you select the “Calib. Type” incorrect, you can return the previous screen by pressing key.

6) key is pressed for starting the zeroing operation.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 220
7) After Zeroing operation is completed. Press key to
continue the operations.

8) For a proper calibration operation, the load used must not


be less than the amount recommended on the screen.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 221
9) The weight value to be used in calibration is stated in this
screen. For confirming the weight value entered key is
pressed and proceeded to the calibration screen.

Approved and certified weights must be used for a healthy calibration.

10) By placing the load equal to the calibration weight


stated
in the message screen and pressing the key, the
calibration process is initiated.

The screen shown at side is displayed during the calibration.


ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 222
11) General information screen indicates that the calibration
is completed. Key is pressed for proceeding to the next
screen.

12) key is pressed to store the calibration values or


is pressed to exit from the menu without storing the
calibration values.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 223
11.7.2 Zero Calibration
The following operations must be performed respectively for the zero calibration.
It is entered to menu 2.5 for the zero calibration:
1) Load cell type is selected. The options are;
1) Esit(2 mV/V)
2) 2.5mV/V
3) 5.0mV/V
4) 10.0mV/V
5) 20.0mV/V
6) 160.0mV/V
2) For entering to Calibration Type menu key is
pressed. “Zero Calibration” is selected as the calibration
type. The options are;
1) Load Calibration
2) mV/V Calibration
3) Coefficient Calibration
4) Zero Calibration

3) For proceeding to the zeroing screen key is pressed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 224
4) Weighing platform is emptied and key is pressed for
starting the zeroing operation.

The screen shown at side is displayed during the zeroing operation.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 225
5) Key is pressed to proceed to the calibration
confirmation screen.

6) Key is pressed to store the calibration values or is


pressed to exit from the menu without storing the calibration
values.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 226
11.8 BeltScale Connections
Note: When the BeltScale program is selected, the settings of the Relay modules in Slot 4 and Slot5 and the EXP
Module in Slot 6 cannot be changed and related menus will be locked. The connections of these modules must be made
as follows.
11.8.1 Input Connections

Slot 4 Input Connections

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 227
11.8.2 Output Connections

Slot 4 Output Connections

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 228
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 229
12 Annex 3 Dosage Belt Scale Program
Electronic conveyor belt scale system enables the weighing of bulk materials carried by conveyor belts during the
conveying process. The flow amount (capacity) is proportional to the speed and the material on the belt.

12.1 Specifications
 High internal resolution (24 bits)
 Digital filter
 2 total values (Total1 is resettable . Total is non-resettable)
 Actual Flowrate t/h, belt speed, kg per meter values
 Internal RTC and battery circuitry to hold Totalizing data
 Belt working simulator
 Alibi Memory

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 230
12.2 Main Program Screen
Dosage Belt Scale program main screen shot will be as
shown at side;

Note: If ECI is not in “Dosage Belt Scale” program (If the main screen above is not displayed), You can refer to section
7.1 Program Selection for shifting to the “Dosage Belt Scale” program.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 231
12.2.1 Screen Layout

1) Flow
2) Unit
3) Logo
4) Flow percentage 1 2
5) Auto zero, total
6) Speed, load per meter 5
7) Keys 3 4
6
7

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 232
12.2.2 Keys
Reset Total Key Used to reset Total1 and Total2 values (T1 ve T2)
Used to swich between Total1, Total2 and general Total
Change Total Index Key indications for totalization indications. And also the
measurement values shown on the 7 and 8 area are toggles.
x10 Resolution Key Used to activate x10 resolution for Totalize indications
It is used for bringing the menu where the settings related to
BeltScale Menu Key
the beltscale are made.
Used for bringing the menu screen where the device settings
Setting Menu Key
are made.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 233
12.2.3 Symbols
Belt Animation symbol It is used for simulating the belt conveyor is running.

Belt Animation symbol It is used for simulating the belt conveyor is stopped.
This symbol appears on the screen when the measured
Minimum Flowrate symbol
flowrate value is under the “Minimum Flowrate”.
This symbol appears on the screen when the measured
Maximum Flowrate symbol
flowrate value is above the “Maximum Flowrate”.
This symbol appears on the screen while AutoZero
process is in operation. The symbol is not shown, when
Auto Zero Operation symbol
Autozero is Off state or measured weight above 2% of the
“Maximum Capacity” or Interlock input is not valid.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 234
12.3 Dosage Belt Scale Menu

1. Initial Setup 2. Coefficients 3. ZeroingPasswd 4.Report


Singed Total Min.Flowrate Total1 Records
Weight Length Coefficient1 Total2 Period.Report
Speed Sensor Max.Flowrate
Pulse Distance Coefficient2
Sensor Timeout
Belt Length
Auto Zero
Belt Oscillation

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 235
12.4 Dosage Belt Scale Menu Usage
Initial Setup
It is an ordered type menu that contains the Initial Setup settings for the beltscale program.The sub-menus are
the Signed Total, Weight Length, Speed Sensor, Pulse Distance and Sensor Time Out.
Signed Total
The options are;
1) Standard: Negative loadings are added to totalization.
2) PositiveOnly: Negative loadings are not added to
totalization.
Note: This menu is not shown when Security Type is selected as
“LegalForTrade” (It is fixed Standart mode).

WeightLength
Platform measurement area. Length of the
measurement platform. Sum of the half distance between
the left roller and weighing roller (a) and half distance
between right roller and the weighing roller
(b).WeightLength= a/2 + b/2

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 236
Speed Sensor
Selection of the Input type selection for speed
measurement. The options are;
1) Encoder: Encoder is used for speed measurement.
2) Fixed Speed: Used for fixed speed systems.

Note: The valid signal has to be applied to the speed sensor input (Input1) to notify that belt is running.

Pulse Distance
Distance per pulse value. In this parameter the distance for
every pulse is defined in “mm”.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 237
Sensor TimeOut
Time out information for speed sensor pulse intervals. If the
belt turning starts and stops continuosly, itis hard to
estimate that the belt is stopped. This parameter
determines the accceptable time period of pulse
inputs.

(Note: If you are not sure about the value, please enter 1.00 second. If something wrong with speed please enter 2.50
second)

Design Speed
It contains belt speed info. System speed at a constant
speed system is determined by this parameter.

Note: This parameter is not shown when the SpeedSensor parameter is selected as “Encoder”.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 238
Belt Length
Total length of the belt.

Note: This parameter used for calibration and autozero calculations. Please enter correctly.

Auto Zero
It contains auto zero selection. The options are;
1) Off: Auto zero is disabled.
2) On: Auto zero enabled.

Note:If the “LegalForTrade” option is selected in the “Security Type” menu, this option is not shown and AutoZero option
is fixed On state.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 239
Belt Oscillation
It is the zero belt value. The weight is not added to total if it
is below this value.

Flow Filter
Flow counts.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 240
Coefficients
Belt multipliers defined in the section. According to uncertanity of the Weight Length and Pulse Distance, the
multipliers has to be defined to correct the system.

Note: This submenus are locked when "LegalForTrd" is selected in the "Security Type" menu.

Min.Flowrate
Minimum flowrate value. symbol appears on the
screen when the flowrate is less then the minimum flow
rate.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 241
Coefficient1
Multiplier for minimum Flowrate condition.

Max.Flowrate
Maximum flowrate value. symbol appears on the
screen when the flowrate is more then the maximum flow
rate.

Coefficient 2
Multiplier for maximum Flowrate condition.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 242
ZeroingPassword
Passwords settings menu.

Total1
Used to clear “Total1” value.
This value shown as 0 when entered to this parameter
screen. And user can change it anytime. When entered on
mainscreen correctly the totalized value is cleared.

Total2
Used to clear “Total2” value.
This value shown as 0 when entered to this parameter
screen. And user can change it anytime. When entered on
mainscreen correctly the totalized value is cleared.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 243
Report
This is the menu where you can see the ALIBI records or
take a new ALIBI record for the belt scale program.
Records
Record Number Date/Time Total and Status values can
be seen. Use and keys to move between
records.

To acces the desired record, use key.

For taking a record manually, use key.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 244
Report Period
In this menu, the period for the USB records is determined.
The options are;
1) Off: No USB record is taken.
2) 1 hour: USB records period.
3) 6 hours: USB records period.
4) 12 hours: USB records period.
5) 24 hours: USB records period.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 245
Dosage Belt Scale Settings
This menu shows PID settings of the system.

Target Flow
Enter desired t/h rate .

Gain
To set gain. High values of gain causes unstabilized
weighing while the system reaches to target value faster.
Low values of gain drives more stabilized weighing while
the system reaches to target slower.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 246
Sampling Time
To achieve fast response, the sampling interval should be
be0.10-0.50 seconds (100-500 milliseconds)..

Integral Time
Correction is applied by analysing recorded values of the
system for the specified period of time. Large integration
time value causes small integral effect on output
correction value.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 247
12.5 Alibi Memory
All information and other pertinent information is stored in the Real-Time Clock unit on the device. To keep this
information in case of a power failure, there is a battery circuit in the ECI device. Information regarding the device's
production date and calibration date is stored in the EEPROM memory. During calibration, the new calibration date
will be automatically saved to permanent memory.

ECI device can store the last 1000 totalization records (minimum flow, maximum flow and memory error conditions)
and error log. An Alibi record contains the following information:
Date + Time + Total + Status

Status
Manual Record: With button, a manual record can be taken.
Minimum Flow: When the measured Flow value is lower than the “Minimum Flow” value in the Coefficients menu,
that info gets recorded automatically.
Maximum Flow: When the measurd Flow value is higher than the “maximum Flow” value in the Coefficients menu,
that info gets recorded automatically.
Memory Error: When the system cant acces the memory, that info gets recorded automatically.
Period. Report: Info gets recorded periodically, according to the period time choosen by the user.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 248
12.5 Auto Zero
Autozero feature can be enabled by setting the
“Auto Zero” parameter in “Initial Setup” menu or
in “Security Type” menu, “LegalForTrade” option is choosen.
Additionally the “AutoZero Interlock input” has to be activated to
validate the zeroing process.
Zeroing process is explained below;
 System awaits until the measured weight value falls under the
%2 of the capacity. If the measured weight value stays under
%2 of capacity for a period of belt length and sign blinks. Then auto zero process starts.
 Auto zero process continues for 3 belt length period, meanwhile sign shown permanently, if the
measured weight value gets out of range, auto set proces gets cancelled.
 At the end of Auto zeroing process, device waits for a period of belt once more, it is expected to see that the
measured weight value to measured under the%2 of the capacity, in the meanwhile, sign blinks. When the
proces is done, zero value on screen is updated. If the measured weight value gets out of the range before a
period of belt is completed, auto zero proces gets cancelled.

Note: Auto zero value can be displayed on the screen. The Picture above shows the position.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 249
12.6 Calibration of Dosage Belt Scale
12.6.1 Load Calibration
It is the operation of identification of a mass with a known weight to the device. It must be made by using real
weights. Calibration must be done while belt is running. Device automatically calculates belt movement
distance.

1) Enter the menu 2.5 for the calibration.

2) Load cell type is selected. The options are;


1) Esit (2 mV/V)
2) 2.5mV/V
3) 5.0mV/V
4) 10.0mV/V
5) 20.0mV/V
6) 160.0mV/V
mV/V selection is made depending on the load cell model.

3) For entering the Calibration Type menu is pressed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 250
4) Calibration Type is selected. The options are;
1) Load Calibr.
2) mV/V Calibr.
3) CoeffCalibr.
4) Zero Calibr.

5) After selecting the Calibration Type as the "Load


Calibration”, proceeded to the next screen by pressing first,
then key. The information screen shown at side is
displayed.

If you select the “Calib. Type” incorrect, you can return the previous screen by pressing key.

6) key is pressed for starting the zeroing operation.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 251
7) After Zeroing operation is completed. Press key to
continue the operations.

8) For a proper calibration operation, the load used must not


be less than the amount recommended on the screen.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 252
9) The weight value to be used in calibration is stated in this
screen. For confirming the weight value entered key is
pressed and proceeded to the calibration screen.

Approved and certified weights must be used for a healthy calibration.

10) By placing the load equal to the calibration weight


stated
in the message screen and pressing the key, the
calibration process is initiated.

The screen shown at side is displayed during the calibration.


ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 253
11) General information screen indicates that the calibration
is completed. Key is pressed for proceeding to the next
screen.

12) key is pressed to store the calibration values or


is pressed to exit from the menu without storing the
calibration values.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 254
12.6.2 Zero Calibration
The following operations must be performed respectively for the zero calibration.
1) It is entered to menu 2.5 for the zero calibration:
2) Load cell type is selected. The options are;
1) Esit(2 mV/V)
2) 2.5mV/V
3) 5.0mV/V
4) 10.0mV/V
5) 20.0mV/V
6) 160.0mV/V
3) For entering to Calibration Type menu key is
pressed. “Zero Calibration” is selected as the calibration
type. The options are;
1) Load Calibration
2) mV/V Calibration
3) Coefficient Calibration
4) Zero Calibration

4) For proceeding to the zeroing screen key is pressed.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 255
5) Weighing platform is emptied and key is pressed for
starting the zeroing operation.

The screen shown at side is displayed during the zeroing operation.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 256
6) Key is pressed to proceed to the calibration
confirmation screen.

7) Key is pressed to store the calibration values or is


pressed to exit from the menu without storing the calibration
values.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 257
12.7 Dosage Belt Scale Connections
Note: When the BeltScale program is selected, the settings of the Relay modules in Slot 4 and Slot5 and the EXP
Module in Slot 6 cannot be changed and related menus will be locked. The connections of these modules must be made
as follows.
12.7.1 Input Connections

Slot 4 Input Connections

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 258
Slot 5 Input Connection (E-MDL-DAC)

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 259
12.7.2 Output Connections

Slot 4 Output Connections

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 260
13 Annex 4 Impact Scale Program
It is a program used for generally measuring flow rates of flowable and materials (coal dust, flour, cereals, sugar etc.) by
means of the free fall principle.

13.1 Impact Scale Program Specifications


1) High internal resolution (24bit)
2) Digital Filter
3) 3 total values (Total1 and Total2 are resettable . Total is non-resettable)
4) Current flow rate (t / h), belt speed, load per meter indicator
5) The RTC unit used for time operations and storing the total information
6) Flow Symbol
7) Alibi Memory

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 261
13.2 Home Screen View
The home screen of the impact scale program is as shown on
the right

Note: If the ECI is not in the "Impact Scale" program (the home screen view is not the same as above), please see 7.1
Program Selection section to switch to the "Imapct Scale" program.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 262
13.2.1 Screen Layout
1. Total passing product quantity
2. Total indicator index
3. Unit
2
4. Logo 1
5. Flowrate Percentage 3
6. Zero Zone symbol 5 1
7
7. Flow rate, Auto zero value, Weight 4
6 8
8. Instant Speed, Load per meter
9
9. Keys

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 263
13.2.2 Keys

Zero Key It is used to reset the total values. (T1 and T2)
It is used to toggle between Total1, Total2 and Total. It is also
Change Total Key
used to toggle between the values shown in field 7 and 8.
Increase Resolution Key It is used for 10 times more accurate display.
It is used to call the menu where the adjustments related to
Impact Scale Menu Key
the impact scale are made.
It is used to call the menu screen where the device settings
Setting Menu Key
are made.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 264
13.2.3 Symbols
Flow Symbol There is material flow.

Flow Symbol No material flow.


Indicates that the flow rate is below the value set in the
Minimum Flowrate Symbol
Minimum flow rate menu.
Maximum Flowrate Indicates that the flow rate is above the value set in the
Symbol Maximum flow rate menu.
Zero Symbol Appears when auto reset is performed.

Note: When the ECI program is changed, all total values are reset.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 265
13.3 Impact Scale Menu

1. Initial Setup 2. Coefficients 3. 4.Report


Signed Total Minimum Flowrate Zero.password Records
Flow Rate Coefficient1 Total1 PeriodicalReporting
AutoZero MaximumFlowrate Total2
Belt Oscillation Coefficient2

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 266
13.4 Use of the Impact Scale Menu
Initial Setup
It is the input settings menu related to the impact scale program. This menu is an ordered type menu. The
submenus are SignedTot., Flow Rate, AutoZero and
BeltOscillation.

Signed Total.
Options;
3) Standard: Negative loads are added to the total
4) OnlyPositive: Negative loads are not added to the
total

Flow Rate
The flow rate of the material is determined.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 267
AutoZero
It is the Auto Zero selection. Options;
3) Off: Auto zero is not active.
4) On: Auto Zero is active.

Note: The speed sensor input indicating the rotation of belt must be
activated when the "Constant Speed" option is active.

Belt Oscillation
It is the zero belt value. The weight is not added to total if it
is below this value.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 268
Coefficients
It is the menu where the impact scale coefficient settings are made. It is the same as the belt scale coefficient
menu.

Zero.Password
This is the menu where zero password settings are made. It is the same as the belt scale zero.password
menu.

Report
This is the menu where impact scale report settings are made. It is the same as the belt scale report menu.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 269
13.5 Alibi Memory
It is the same as the belt scale menu. Please see section "11.5".

13.6 Auto Zero


It is the same as the belt scale menu. Please see section "11.6".

13.7 Impact Scale Calibration


It is the same as the belt scale menu. Please see section "11.7".
13.7.1 Load Calibration
13.7.2 Zero Calibration
13.8 Impact Scale Connections
It is the same as the belt scale menu. Please see section "11.8".

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 270
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 271
14 Annex 5: CheckWeigher Program
14.1 The Operation Sequence
14.1.1 Dynamic Mode
1) System waits till the product is sensed by the input sensor.
2) After the product is sensed by the entrance sensor, "PreWaitTime" starts.(If "PreWaitTime" is selected)
3) Weight values are recorded at the end of the "PreWaitTime".
4) Product weight value is calculated when the product is sensed by the outlet sensor.
5) After the product passed through the outlet sensor, “AZInhibitTime” starts.(Auto zeroing inhibition time).
6) At the end of the auto inhibition time, if there are no products on the belt, auto zeroing process is completed.
(If AutoZero is “On”)

14.1.2 Static Mode


Value of the weight on the belt is updated contuniously. Tare function is activated.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 272
14.2 CheckWeigher Program Screen
Screenshot for the dynamic mode is shown on the right side of
this page.

Screenshot for the information screen is shown on the right side


of this page.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 273
Screenshot for the static mode is shown on the right side of this
page.

Note: If ECI is not in “CheckWeigher” program (If the main screen above is not displayed), You can refer to section 7.1
Program Selection for shifting to the “CheckWeigher” program.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 274
14.2.1 Screen Layout (Dynamic Mode)
1. Weight value of the product 8
2. Unit 1 2
3. Minimum product weight
4. Count 3 4
5. Maximum product weight 1
5 6
6. Average product weight 7
7. Keys
8. Product name

14.2.2 Screen Layout (Static Mode)


1. Weighing indicator 3
2. Keys 4
1 5
3. Motion Status
4. Zero Indicator 6
5. Tare Status
2
6. Unit

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 275
14.2.3 Screen Layout (Information Screen)
1. Weighing indicator 4
2. Keys
3. Unit 1 3
4. Product name

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 276
14.2.4 Symbols
Shows that the change in the weight value is within the stagnancy limits
Stagnancy
for a certain period (less than one scale interval).
Motion Shows that the change in the weight value is out of the stagnancy limits.
Zero Indicator Shows that the wander off zero less than 1/4 (resoulation)..
Net Indicates that deviation from zero is less than 1/4e (resolution).
Indicates that the product has passed through the input sensor and
Weighing
weighing has been performed.
Auto Zero Indicates that auto zero is on.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 277
14.2.5 Keys
Static Mode Key Used for selecting the static mode.
Reset Key Used for resetting display on static mode.
CW Menu Key Used for entering the check weigher settings menu.
Settings Menu Key Used to call the system settings menu.
Dynamic Mode Key Used for selecting the dynamic mode.
Tare Key Used for taking the weight on loadcell as tare in static mode.
Tare Clean Key Used for cancelling the tare value and showing the gross value.
Used for resetting istatistical data while operating in dynamic
Delete Key
mode.
Info key Used to switch to the information screen.
Product Selection Key Used to change the product.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 278
14.3 CheckWeigher Menu

1. Product Setting 2. Settings 3. Report


Name Minimum Load
Minimum PreWaitPeriod
Maximum AutoZero
Reject delay AZDataCount
Reject Period AZInhibitTime
Contact Type Coefficient
Band Speed

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 279
14.4 CheckWeigher Menu Usage

Product Setting
It is an ordered menu type menu accommodating the checkweigher program menu settings. The submenus are
"Name", "Minimum", "Maximum", "RejectDelay", "RejectTime", "Contact Type" and "Band Speed".

Name
Product name is set.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 280
Minimum
Minimum sets the upper limit of the weight value
where the eject relay is active.

Maximum
Maximum sets the lower limit of the weight value
where the eject relay is active.

Note: The acceptance relay is active when the weight of the product passing through is between the minimum and
maximum values.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 281
Reject delay
Indicates the time period required for the activation
of the reject relay and acceptance relay after the
passage of the product through the output sensor.

Reject Time
Indicates how long the reject relay and acceptance
relays will remain active.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 282
Contact Type
Determines the start-up position of the relay
contacts. Options;
1) Norm.Open: The relay contact is open circuit at
start-up.
2) Norm.Close: The relay contact is short circuit at
start-up.

Band Speed
Band speed setting is done. The value is entered
as a percentage (%).
When 100% is entered, 5 V appears at the analog
output terminals of the E-MDL-DAC module in slot
5.

NOTE: In static mode, the belt is stopped.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 283
Setup
This is a sequential menu where CheckWeigher programs settings are done. Submenus are “Minimum Load”,
“PreWaitTime”, “AutoZero”,”AZDataCount”, “AZInhibitTime” and “Coefficient”.

Minimum Load
Minimum load value is set by this parameter. Error
message is received when load is lower than minimum
load.

PreWaitTime
Weighing values take into account to how long thereafter
time, when then the product accrossed enterance sensor.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 284
AutoZero
Auto zero selection is set by this parameter.

Options
Off: Not to auto zero.
On: Auto zero is set when none of product on the band.

AZ Data Count
Auto zero data counting is set by this parameter.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 285
AZ Inhibit Time
Auto zero inhibit time is set by this parameter. This
paramater calculating to starting time when after the
product get through the output sensor.

Coefficient
Auto correction coefficient is set by this parameter.
Measurement value which that obtained value end of the
process, pounded by coefficient.After that way the product
weigh is obtained.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 286
Report
COUNTER, TOTAL and MEAN values are shown on screen.
button is used to set zeroing when on request.

COUNTER: All of product number.


TOTAL: Total weigh of product.
MEAN: Average weigh of product.
MIN REJECT: The quantity of products passing with the
weight below the minimum weight
MAX REJECT: The quantity of products passing with the
weight above the maximum weight

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 287
14.5 Check Weigher Connections
Note: When checkweigher program is chosen, Slot4 is used as Relay Module,Slot5 is used as DAC Module and Slot6
Module is not used. DAC Module wire connections are shown in section “10.2.2” .Relay Module wire connections are
shown in below.

14.5.1 Input Connections

Slot 4 Input Connection

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 288
14.5.2 Output Connection

Slot 4 Output Connection

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 289
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 290
15 Annex 6 Modbus
15.1 Device Menu Settings
Related settings are under COM Module Menu.
Mode: Modbus must be selected.
Following parameters must be set as the same on the device to be connected.
Baudrate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200
Modbus Mode: RTU-ASCII
Modbus No: 1-255

15.2 Connection Point Settings


Bits per Second 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200
Parity None
Data Bits 8
Stop Bits 1

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 291
15.3 Modbus Commands
3 Multiple Record Reading
6 Single Entry Writing
16 Multiple Entry Writing

15.4 Modbus Directly Accessible Memory Map


Status A (16 Bit, LSB to MSB)
BIT(s) Function
0-3 (/R) Program Mode: 0: Indicator, 1: Batching, 2: Filling, 3: PLC, 4-15:Out of use
4-8 (/R) (5bits) Out of use
9 (/R) 1:No load cell module inserted or it is not working, 0:load cell module is working.
10 (/R) 1:ADC input voltage over detected. It must be cleaned by the user.
11 (/R) Out of use
12 (/R) 1:ADC input voltage is very high. It is 0 when the value returns to normal.
Abnormal system operation flag. Set as 1 when wrong operation occurs. It means that the
13(R/W)
system has been restarted due to the electricity problems.
14-15 (2bits) Out of use

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 292
Status B (16 Bit, LSB to MSB)

BIT(s) Function
Scale Interval. (1, 2, 5 x10^n )
0-6 (/R) (7bits) 1:1, 2:2, 5: 5, 10:10, 20:20, 50:50, 100:100
101:200, 102:500, 103:1000, 104:2000, 105:5000
7 (/R) Slot 4 Relay 1 On/Rff Status. 1:On, 0:Off
8-10 (/R) (3bits) This value shows on which digit from right will the point be. It can be 3 maximum
11-14 (/R) (4bits) Unit (0:g, 1:Kg, 2:t, 3:N, 4:lb, 5:lt, 6:mV, 7:m3 , 8:None)
15 (/R) Slot 4 Relay 2 On/Off Status. 1: On, 0: Off

(R/W) Writable and Readable


(/R) Read Only

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 293
Status C (16 Bit, LSB to MSB)

BIT(s) Function
0-2 (RW) (3bits) Load Cell Type. 0: 2mV/V, 1: 2.5mV/V, 2: 5mV/V, 3: 10mV/V, 4: 20mV/V, 5: 160mV/V
3-4 (RW) (2bits) Out of use
5 (/R) 1:Error, 0: No Error
6 (/R) 1: Calibration is possible, 0: No calibration is possible.
7 (/R) Communication lock. 1:On 0:Off
8 (/R) 1:On device menu screen, 0: On Device main screen
9 (/R) 1: Absolute Zero yes, 0: Absolute Zero no
10 (/R) 1: Stagnancy, 0:Mobility
11 (/R) 1:Maximum load yes, 0: Maximum load no
12 (/R) 1:e2 zone
13 (/R) 1:e3 zone
14 (/R) Out of use
15 (RW) Measuring Type. 1:Multi Partition or Multi Interval, 0:Single Partition

(R/W) Writable and Readable


(/R) Read Only

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 294
15.4.1 Belt Scale Program
Adress Word Number Function
0 (R/W) 1 Command register: it is used for the special functions together with the Reg A and Reg B.
1 (R/W) 1 Reg A: General Purpose Data Register.
2 (R/W) 1 Reg B: General Purpose Data Register.
3(/R) 2 Total
5 1 Status A (Explanations are stated above in detail)
6 1 Status B (Explanations are stated above in detail)
7 1 Status C (Explanations are stated above in detail)
8 (/R) 2 Total 1
10 (/R) 2 Total 2
12 (/R) 2 Flowrate
14 (/R) 2 Belt Speed
16(/R) 2 Average kg value per meter
18(/R) 2 Weight
20(/R) 2 Auto Zero Value

(R/W) Writable and Readable


(/R) Read Only

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 295
15.4.1 Dosage Belt Scale Program
Adress Word Number Function
Command register: it is used for the special functions together with the Reg A
0 (R/W) 1
and Reg B.
1 (R/W) 1 Reg A: General Purpose Data Register.
2 (R/W) 1 Reg B: General Purpose Data Register.
3(/R) 2 Total
5 1 Status A (Explanations are stated above in detail)
6 1 Status B (Explanations are stated above in detail)
7 1 Status C (Explanations are stated above in detail)
8 (/R) 2 Total 1
10 (/R) 2 Total 2
12 (/R) 2 Flowrate
14 (/R) 2 Belt Speed
16(/R) 2 Average kg value per meter
18(/R) 2 Weight
20(/R) 2 Auto Zero Value
22(/R) 2 Status
(R/W) Writable and Readable
(/R) Read Only
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 296
15.4.2 CheckWeigher Program
Register Address(16Bit) Function
0(/O) Status A
1 (/O) Status B
2 (/O) Status C
3 (/O) Upper Product Count
4 (/O) Lower Product Count
5 (/O) Upper Last Product Weight
6 (/O) Lower Last Product Weight
7 (/O) Upper Total Product Weight
8 (/O) Lower Total Product Weight
9 (/O) Upper Average Product Weight
10 (/O) Lower Average Product Weight
11(/O) Upper Minimum Product Weight
12(/O) Lower Minimum Product Weight
13(/O) Upper Maximum Product Weight
14(/O) Lower Maximum Product Weight
15(/O) Upper Weight
16(/O) Lower Weight
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 297
15.5 Examples:
15.5.1 Reading more than one register.
Such that Modbus ID:1, in order to read the data from register no 8 to register no 14
following data array is sent.
Modbus Function Start Register CRC
ID Code Address Number Data
01 03 00 08 00 06 44 0A

The response received is as follows.


Byte
Modbus Function Screen Screen Tare Tare Gross Gross CRC
Numb
ID Code Top Bottom Upper Lower Upper Lower Data
er
01 03 0C 00 00 01 F4 00 00 00 64 00 00 02 58 85 23
Accordingly
Screen Value=0x000001F4=500
Tare Value=0x00000064=100
Gross Value=0x00000258=600

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 298
15.5.2 Writing to the register.
Such that Modbus ID: :5, in order to write 0x0018 data to register no 1 (Reg A: General Purpose Register)
following data array is sent.

Modbus Function CRC


Address Data
ID Code Data
05 06 00 01 00 18 D9 84

The response received is as follows.


Modbus Function CRC
Address Data
ID Code Data
05 06 00 01 00 18 D9 84

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 299
15.5.3 Removing the Security Lock.(Command Register =1, RegA
=1357h, RegB=2192h)
In order to write to the EEPROM section, first of all a particular password is sent and EEPROM
Communication lock is unlocked. This lock will unlock automatically after 3 seconds. Thus, this command must
be executed before each EEPROM writing operation. Such that the Modbus ID:1, the communication lock
opening command is;

Byte
Modbus Function Start Register Command
Numb Reg A Reg B CRC Data
ID Code Address Number Register
er
01 10 00 00 00 03 06 00 01 13 57 21 92 F7 E9

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 300
15.5.4 Writing to EEPROM section
Such that the Modbus ID:1, following data array is sent in order to change the Unit information at address no 47
(0x2F) on EEPROM as kilogram.

Byte
Modbus Function Start Register Command
Numbe Reg A Reg B CRC Data
ID Code Address Number Register
r
01 10 00 00 00 03 06 00 02 00 2F 00 01 6F 49

NOTE: As a matter of course, when a value is written to the EEPROM section, only the EEPROM
data changes. There are two ways to update the RAM values with the EEPROM values;
1-) It is necessary to disconnect, then connect the power.
2-) 0x0A Command must be executed over Modbus. (0x0A: update parameters from EEPROM)

NOTE: Please request the EEPROM list to change the device parameters via communication.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 301
15.6 Command Register Functions
Address no 0 is the command register. Following table explains the values written to this address and operations
performed in consideration

Value Function
No operation It is used for feedback. Command denied. (This operation cannot be performed under the
-2
current status.)
-1 No operation It is used for feedback. It means that last operation is not successful.
0 No operation It is used for feedback. It means that last operation is successful.
1 Unlocking Communication lock It is used for writing data on EEPROM.
2 Writing to EEPROM, data should be written to RegB and address should be written to RegA.
3 Reading from EEPROM:Address should be written to RegA. Data is place to RegB.
5 Reset: Used for resetting the screen value.
6 Tare: Tare the screen value. (If the screen value is not 0 or negative)
7 Learning the Communication Lock: To Both RegA and RegB
8 Device Type: Data is written to RegA. ECI Family: (0x3000)
9 Version No: Data is written to RegA.
10 Transfer device parameters from EEPROM to RAM.
12 Clean the error bits

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 302
13 Restart the device.
41 Total1: It is used to reset Total1 value in the belt scale program.
42 Total2: It is used to reset Total2 value in the belt scale program.
Select Recipe: It is used for selecting recipe in batching program. Recipe number should be written to
RegA.
20
Select Product: It is used for selecting product in filling program. Product number should be written to
RegA.
Start Batching: It is used for start batching in batching program.
21
Start Filling: It is used for start batching in filling program.
Start Batching: It is used for stop batching in batching program.
22
Start Filling: It is used for stop batching in filling program.
Relay 1 Minimum: It is used for setting the minimum value of the 1.relay of the relay card inserted to
128*
Slot 4 in the Scale Program.
Relay 1 Maximum: It is used for setting the maximum value of the 1.relay of the relay card inserted to
129*
Slot 4 in the Scale Program.
Relay 2 Minimum: It is used for setting the minimum value of the 2.relay of the relay card inserted to
130*
Slot 4 in the Scale Program.
Relay 2 Maximum: It is used for setting the maximum value of the 2.relay of the relay card inserted to
131*
Slot 4 in the Scale Program.
*When the device is restarted or any parameter is changed from the menus, these values return to their value displayed
in the menu.
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 303
15.7 Examples
15.7.1 Taring
0x06 is written to the Command register for Taring. (Modbus ID:2 was used)

Modbus Function CRC


Address Data
ID Code Data
02 06 00 00 00 06 09 FB
When the operation is performed, the same of the data string sent is returned.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 304
15.7.2 Setting the Relay 1 Minimum Value
Following data must be sent in order to set the minimum value of the 1. relay in Slot 4 as 500.
(Modbus ID:3 was used)

Modbus Function Start Register Byte Command


Reg A Reg B CRC Data
ID Code Address Number Number Register
03 10 00 00 00 03 06 00 80 00 00 01 F4 E0 0B

The response received is as follows.

Modbus Function Start Register


CRC Data
ID Code Address Number
03 10 00 00 00 03 81 EA

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 305
15.7.3 Setting the Relay 2 Maximum Value
Following data must be sent in order to set the maximum value of the 2. relay in Slot 4 as 800.
(Modbus ID:3 was used)

Modbus Function Start Register Byte Command


Reg A Reg B CRC Data
ID Code Address Number Number Register
03 10 00 00 00 03 06 00 83 00 00 03 20 A5 34

The response received is as follows.

Modbus Function Start Register


CRC Data
ID Code Address Number
03 10 00 00 00 03 81 EA

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 306
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 307
16 Annex 7 Profibus
16.1 BeltScale Program
Register Adresi (16 Bit) Function
0 (/R) Status A
1 (/R) Status B
2 (/R) Status C
3 (/R) Total (Upper)
4 (/R) Total
5 (/R) Total 1 (Upper)
6 (/R) Total 1
7 (/R) Total 2 (Upper)
8 (/R) Total 2
9 (/R) Flowrate (Upper)
10(/R) Flowrate
11(/R) Belt Speed (Upper)
12(/R) Belt Speed
13(/R) Batch output current value (Upper)

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 308
14(/R) Batch output current value
15(/R) Average kg value per second (Upper)
16(/R) Average kg value per second
17(/R) Average kg value per meter(Upper)
18(/R) Average kg value per meter
19(/R) Weight (Upper)
20(/R) Weight
21(/R) Internal count (Upper)
22(/R) Internal count
23(/R) Pulse output current value (Upper)
24(/R) Pulse output current value
25(/R) Status (Upper)
26(/R) Status
27(/R) Upper Command Response
28(/R) Lower Command Response

(/R) Read only

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 309
16.2 Dosage Belt Scale Program
Register Adresi (16 Bit) Function
0 (/R) Status A
1 (/R) Status B
2 (/R) Status C
3 (/R) Total (Upper)
4 (/R) Total
5 (/R) Total 1 (Upper)
6 (/R) Total 1
7 (/R) Total 2 (Upper)
8 (/R) Total 2
9 (/R) Flowrate (Upper)
10(/R) Flowrate
11(/R) Belt Speed (Upper)
12(/R) Belt Speed
13(/R) Batch output current value (Upper)
14(/R) Batch output current value
15(/R) Average kg value per second (Upper)

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 310
16(/R) Average kg value per second
17(/R) Average kg value per meter(Upper)
18(/R) Average kg value per meter
19(/R) Weight (Upper)
20(/R) Weight
21(/R) Internal count (Upper)
22(/R) Internal count
23(/R) Pulse output current value (Upper)
24(/R) Pulse output current value
25(/R) Status (Upper)
26(/R) Status
27(/R) Upper Command Response
28(/R) Lower Command Response

(/R) Read Only

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 311
16.3 CheckWeigher Program
Register Address(16Bit) Function
0(/O) Status A
1 (/O) Status B
2 (/O) Status C
3 (/O) Upper Product Count
4 (/O) Lower Product Count
5 (/O) Upper Last Product Weight
6 (/O) Lower Last Product Weight
7 (/O) Upper Total Product Weight
8 (/O) Lower Total Product Weight
9 (/O) Upper Average Product Weight
10 (/O) Lower Average Product Weight
11(/O) Upper Minimum Product Weight
12(/O) Lower Minimum Product Weight
13(/O) Upper Maximum Product Weight
14(/O) Lower Maximum Product Weight
15(/O) Upper Weight
16(/O) Lower Weight

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 312
Status A (16 Bit, LSB to MSB)

BIT(s) Function
0-3 (/R) Program Mode: 0: Indicator, 1: Batching, 2: Filling, 3: PLC, 4-15:Out of use
4-8 (/R) (5bits) Out of use
9 (/R) 1:No load cell module inserted or it is not working, 0:load cell module is working.
10 (/R) 1:ADC input voltage over detected. It must be cleaned by the user.
11 (/R) Out of use
12 (/R) 1:ADC input voltage is very high. It is 0 when the value returns to normal.
Abnormal system operation flag. Set as 1 when wrong operation occurs. It means that the
13(R/W)
system has been restarted due to the electricity problems.
14-15 (2bits) Out of use

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 313
Status B (16 Bit, LSB to MSB)

BIT(s) Function
Scale Interval. (1, 2, 5 x10^n )
0-6 (/R) (7bits) 1:1, 2:2, 5: 5, 10:10, 20:20, 50:50, 100:100
101:200, 102:500, 103:1000, 104:2000, 105:5000
7 (/R) Slot 4 Relay 1 On/Rff Status. 1:On, 0:Off
8-10 (/R) (3bits) This value shows on which digit from right will the point be. It can be 3 maximum
11-14 (/R) (4bits) Unit (0:g, 1:Kg, 2:t, 3:N, 4:lb, 5:lt, 6:mV, 7:m3 , 8:None)
15 (/R) Slot 4 Relay 2 On/Rff Status. 1: On, 0: Off

(R/W) Writable and Readable


(/R) Read Only

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 314
Status C (16 Bit, LSB to MSB)

BIT(s) Function
0-2 (RW) (3bits) Load Cell Type. 0: 2mV/V, 1: 2.5mV/V, 2: 5mV/V, 3: 10mV/V, 4: 20mV/V, 5: 160mV/V
3-4 (RW) (2bits) Out of use
5 (/R) 1:Error, 0: No Error
6 (/R) 1: Calibration is possible, 0: No calibration is possible.
7 (/R) Communication lock. 1:On 0:Off
8 (/R) 1:On device menu screen, 0: On Device main screen
9 (/R) 1: Absolute Zero yes, 0: Absolute Zero no
10 (/R) 1: Stagnancy, 0:Mobility
11 (/R) 1:Maximum load yes, 0: Maximum load no
12 (/R) 1:e2 zone
13 (/R) 1:e3 zone
14 (/R) Out of use
15 (RW) Measuring Type. 1:Multi Partition or Multi Interval, 0:Single Partition

(R/W) Writable and Readable


(/R) Read Only

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 315
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 316
17 Annex 8 Ethernet IP Addresses
TCP-IP Port No:502 dir.

17.1 Belt Scale Program


Register Adress (16 Bit) Function
512 (/R) Status A
513 (/R) Status B
514 (/R) Status C
515 (/R) Upper Total
516 (/R) Lower Total
517 (/R) Upper Total 1
518 (/R) Lower Total 1
519 (/R) Upper Total2
520 (/R) Lower Total2
521 (/R) Flow
522 (/R) Flow
523 (/R) Belt Speed

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 317
524 (/R) Belt Speed
525 (/R) Quantity output active value
526 (/R) Quantity output active value
527 (/R) Average kg value per second
528 (/R) Average kg value per second
529 (/R) Average kg value per meter
530 (/R) Average kg value per meter
531 (/R) Weight
532 (/R) Weight
533 (/R) Internal count
534 (/R) Internal count
535 (/R) Current value of pulse output
536 (/R) Current value of pulse output
537 (/R) Status
538 (/R) Status
539 (/R) Upper Command Response
540 (/R) Lower Command Response

(/R) Read only

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 318
17.2 Dosage Belt Scale Program
Register Adress (16 Bit) Function
512 (/R) Status A
513 (/R) Status B
514 (/R) Status C
515 (/R) Upper Total
516 (/R) Lower Total
517 (/R) Upper Total 1
518 (/R) Lower Total 1
519 (/R) Upper Total2
520 (/R) Lower Total2
521 (/R) Flow
522 (/R) Flow
523 (/R) Belt Speed Upper
524 (/R) Belt Speed Lower
525 (/R) Quantity output active value Upper
526 (/R) Quantity output active value Lower
527 (/R) Average kg value per second Upper

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 319
528 (/R) Average kg value per second Lower
529 (/R) Average kg value per meter Upper
530 (/R) Average kg value per meter Lower
531 (/R) Weight Upper
532 (/R) Weight Lower
533 (/R) Internal count Upper
534 (/R) Internal count Lower
535 (/R) Current value of pulse output Upper
536 (/R) Current value of pulse output Lower
537 (/R) Status Upper
538 (/R) Status Lower
539 (/R) Command Response Upper
540 (/R) Command Response Lower

(/R) Read Only

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 320
17.3 ChechWeigher Program
Register Adresi (16 Bit) Function
512(/O) Status A
513 (/O) Status B
514 (/O) Status C
515 (/O) Upper Product Count
516 (/O) Lower Product Count
517 (/O) Upper Last Product Weight
518 (/O) Lower Last Product Weight
519 (/O) Upper Total Product Weight
520 (/O) Lower Total Product Weight
521 (/O) Upper Average Product Weight
522 (/O) Lower Average Product Weight
523(/O) Upper Minimum Product Weight
524(/O) Lower Minimum Product Weight
525(/O) Upper Maximum Product Weight
526(/O) Lower Maximum Product Weight
527(/O) Upper Weight
528(/O) Lower Weight

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 321
Status A (16 Bit, LSB to MSB)

BIT(s) Function
0-3 (/R) Program Mode: 0: Indicator, 1: Batching, 2: Filling, 3: PLC, 4-15:Out of use
4-8 (/R) (5bits) Out of use
9 (/R) 1:No load cell module inserted or it is not working, 0:load cell module is working.
10 (/R) 1:ADC input voltage over detected. It must be cleaned by the user.
11 (/R) Out of use
12 (/R) 1:ADC input voltage is very high. It is 0 when the value returns to normal.
Abnormal system operation flag. Set as 1 when wrong operation occurs. It means that the
13(R/W)
system has been restarted due to the electricity problems.
14-15 (2bits) Out of use

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 322
Input-Output (16 Bit, LSB to MSB)

Input 4 Input 3 Input 2 Input 1 Input 4 Input 3 Input 2 Input 1 Relay 4 Relay 3 Relay 2 Relay 1 Relay 4 Relay 3 Relay 2 Relay 1

Slot 5 Slot 4 Slot 5 Slot 4

Note: Input exist: 0 Input none : 1

Note: Relay Contact Closed: 1 Relay Contact Open : 0

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 323
Status C (16 Bit, LSB to MSB)

BIT(s) Function
0-2 (RW) (3bits) Load Cell Type. 0: 2mV/V, 1: 2.5mV/V, 2: 5mV/V, 3: 10mV/V, 4: 20mV/V, 5: 160mV/V
3-4 (RW) (2bits) Out of use
5 (/R) 1:Error, 0: No Error
6 (/R) 1: Calibration is possible, 0: No calibration is possible.
7 (/R) Communication lock. 1:On 0:Off
8 (/R) 1:On device menu screen, 0: On Device main screen
9 (/R) 1: Absolute Zero yes, 0: Absolute Zero no
10 (/R) 1: Stagnancy, 0:Mobility
11 (/R) 1:Maximum load yes, 0: Maximum load no
12 (/R) 1:e2 zone
13 (/R) 1:e3 zone
14 (/R) Out of use
15 (RW) Measuring Type. 1:Multi Partition or Multi Interval, 0:Single Partition

(R/W) Writable and Readable


(/R) Read Only

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 324
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 325
18 Annex 9 Profinet Addresses
Profinet address sequence is same as Ethernet IP. Please see Annex 10
Input/Output 064 bytes block is used to access via PLC.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 326
18.1 Profinet Commands (Belt Scale Program)
Command Description
41 Total1 Reset
42 Total1 Reset

Total1 Reset

Write Area 1
Command
41

Total2 Reset

Write Area 1
Command
42

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 327
18.2 Profinet Commands (Checkweigher Program)
Command Description
1 Information
2 Write to EEPROM
3 EEPROM Read
4 Program Mode
5 Reset
6 Taring
7 Relay Driving
8 Input Reading
9 Device Version
13 Device Reset
18 Reset statistics
21 Accessible Menu
22 Inaccessible Menu
28 Change Product
29 Change Product Name

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 328
Information

REQUEST
Write Area 1 Write Area 2 Write Area 3 Write Area ... Write Area 16
Command 1 X X X X

Response:

32 … … 10 9 8 7 6 5 4 3 2 1

Auto Zero calculating


Auto Zero changing

New Product Input

Weight Calculating

Passage Started
Dynamic Mode

Engine Failure
X X X X X X

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 329
Write to EEPROM
Write Area 1 Write Area 2 Write Area 3 Write Area 4 Write Area 5
Command
2 Start Address Byte count Data Low Data High

Command Response Low Command Response High

Response Data Low Data High Operation successful

Command Response Low Command Response High


Response FE FF Operation failed

Note: Values that are 4 bytes in size should be multiplied by 1000 and then sent.
Note: Values that are 2 bytes in size should be multiplied by 10 and then sent.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 330
EEPROM Read

Write Area 1 Write Area 2 Write Area 3 Write Area 4 Write Area 5
Command
3 Start Address Byte count X X

Command Response Low Command Response High

Response Data Low Data High Operation successful

Command Response Low Command Response High


Response FE FF Operation failed

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 331
Program Mode
Program Mode
0 Scale
Write Area 1 1 Batching
Command 2 Filling
4 3 Belt Scale
4 Impact Scale
5 Hopper Scale
6 Checkweigher

Command Response Low Command Response High

Response Program Mode High Program Mode Low Operation successful

Command Response Low Command Response High


Response FE FF Operation failed

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 332
Reset
Write Area 1
Command
5

Command Response Low Command Response High

Response Reset Value Low Reset Value High Operation successful

Command Response Low Command Response High


Response FE FF Operation failed

NOTE: Not active in dynamic mode. It can be used in static mode.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 333
Taring

Write Area 1
Command
6

Command Response Low Command Response High

Response Tare Value Low Tare Value High Operation successful

Command Response Low Command Response High


Response FE FF Operation failed

NOTE: Not active in dynamic mode. It can be used in static mode.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 334
Relay Driving

Checkweigher Mode
Write Write Area 2 (Slot 5) Write Area 3 (Slot 4) Write Area
Area 4...16
1
16 . 4 3 2 1 16 . 4 3 2 1
7 X X Relay Relay Relay Relay X X Relay Relay Relay Relay
4 3 2 1 4 3 2 1
0 0 0 1 1 1 0 0 0 1 0 1 X X X

Writing Area 2 (Slot 5) Writing Area 3 (Slot 4)


Relay 4 Relay 3 Relay 2 Relay 1 Relay 4 Relay 3 Relay 2 Relay 1
0 1 1 0 0 1 0 1
Writing Area 2 = 6 Writing Area 3 = 5

Write 1 for the relay desired to be pulled and 0 for the one to be released. Write the value in the green field in the writing
area.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 335
Response

Command Response Low


Slot 5 Slot 4
8 7 6 5 4 3 2 1
Relay 4 Relay 3 Relay 2 Relay 1 Relay 4 Relay 3 Relay 2 Relay 1
0 1 1 0 0 1 0 1

NOTE: Remote relay control can not be performed in dynamic mode.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 336
Input Reading

Write Area 1
Command
8

Command Response Low


Slot 5 Slot 4
8 7 6 5 4 3 2 1
Input4 Input3 Input2 Input1 Input4 Input3 Input2 Input1
0 1 1 0 0 1 0 1

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 337
Device Version

Write Area 1
Command
9

Command Response Low Command Response High

Response Version No xx Operation successful

Command Response Low Command Response High


Response FE FF Operation failed

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 338
Device reset

Write Area 1
Command
13

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 339
Reset statistics
Values to reset
Counter Total number of products
Write Area 1 Total Total weight
Command
18 Mean Mean weight
Min Red Total number of products rejected
Max Red Total number of products rejected
Minimum Minimum weight weighed
Maximum Maximum weight weighed

Command Response Low Command Response High

Response 18 xx Operation successful

Command Response Low Command Response High


Response FE FF Operation failed

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 340
Unlocking the Menu Lock

Write Area 1
Command
21

The , and on the screen are opened to give


access to menu and authorization to change product.

Command Response Low Command Response High

Response 1 xx Operation successful


Command Response Low Command Response High
Response FE FF Operation failed

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 341
Menu Lock

Write Area 1
Command
22

The , and on the screen are removed to


prevent access to menu and to change product.

Command Response Low Command Response High

Response 0 xx Operation successful


Command Response Low Command Response High
Response FE FF Operation failed

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 342
Change Product

Write Area 1 Write Area 2


Command
28 Product No. (1-16)

Command Response Low Command Response High


Response Product No. (1-16) xx Operation successful
Command Response Low Command Response High
Response FE FF Operation failed

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 343
Change Product Name

Command
Write Area 1 29
Write Area 2 Product No. (1-16)
Write Area 3 The first letter of the product name
Write Area 4 The second letter of the product name
Write Area 5 The third letter of the product name
Write Area 6 The fourth letter of the product name
Write Area 7 The fifth letter of the product name
Write Area 8 The sixth letter of the product name
Write Area 9 The seventh letter of the product name
Write Area 10 The eighth letter of the product name

Command Response Low Command Response High


Response Product No. (1-16) xx Operation successful
Command Response Low Command Response High
Response FE FF Operation failed

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 344
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 345
19 Error Warning And Information Messages
Type No Short Name Explanation
This message is displayed when the measured value is more than
Error 1 Maximum Load!
"Capacity" + 9e.
The load with this value cannot be zerod. It can be zeroed with zero
Error 3 Zero Limit!
calibration.
Error 5 LoadCellError Load cell may be defected or there is a problem with the connection.
Error 6 ADC Data Error! ADC Module is defective. Consult ESIT.
It shows that the coefficient calculated as the result of the calibration is
Error 10 Coeff.OutofLimit out of limit. The load used must be increased or the value to be
calibrated must be decreased.
It shows that the load used for the calibration is not sufficient. The load
Error 11 Insufficient Load!
used must be increased.
Error 12 2mV/V Only It can be used only for 2mV/V load cell.
No back-up has been made for the load cell. Thus it cannot be
Error 21 No record!
recovered.
Error 22 Memory 1 Error The memory unit could not read/write. Consult ESIT.
The parameter attempted to be recorded is out of the allowed limits.
Error 23 Wrong Value
Check the values
ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 346
Error 24 Memory 2 Error The memory unit could not read/write. Consult ESIT.
Error 31 PrinterTime.Out Printer Connection is wrong or there is no printer connection.
Error 32 ProfiModulError Profibus Module is defective. Consult ESIT.
Error 42 Battery Error Date/Time info is unable to be read. Change the battery.
Error 50 No USB USB is not inserted or detected.
Error 51 No File No File in USB
Error 52 Corrupted File Reload the file
Error 53 Restart! Shut-down and turn on the device.
Sensitive Quantity cannot be more than the Target Value. (Shot Value
Error 60 SensitiveQuant.Error
+ Sensitive Quantity value) also cannot be more than the Target Value.
Shot Value cannot be more than the Target Value. (Shot Value +
Error 61 ShotValueError
Sensitive Quantity value) also cannot be more than the Target Value.
The operation cannot be started as the Tare value is lower than the
Error 62 LowTareValue
value entered to the Minimum Tare menu.
The operation cannot be started as the Tare value is higher than the
Error 63 HighTareValue
value entered to the Minimum Tare menu.
Error 64 TargetValue.Excess Target Value cannot be more than the Capacity Value.
The filling operation could not be started as the target weight is less
Error 66 Scale is full
than the the weight on the weighing platform. (When Auto Tare is Off)
Error 70 InsufficientDigit The digit is not sufficient for showing the value measured.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 347
More than the designated number of errors occurred in any of the silos
Error 71 ToleranceError!
as the result of batching
It is displayed when no weight changes occurs during the
Error 72 Disc.Tim.Lap
“FillTimeLapse” period while discharging
Error 90 ADC ModulError ADC could not be read. Consult ESIT.
Error 91 Defective Sensor The temperature sensor could not be read. Consult ESIT.
Warning 101 Mobility Zeroing or taring could not be performed due to mobility.
Warning 102 Tare active! No zeroing can be made while the tare is active.
Warning 103 Negative Load No taring can be possible for the negative loads.
Warning 104 No taring possible No taring can be possible for the zero value.
Warning 108 RestoreFact.Settings Factory Settings will be restored.
Warning 109 Restarting The device will be restarted.
Filling or discharging valves can be opened by activating the units
Warning 110 Output will change
connected to the relay outputs.
It shows that no weight change occurs during the "FillingTimeLapse"
Warning 120 FillingTimeLapse
period while filling.
It is shown when the maximum batching value of 100.000 for silo is
Warning 121 SiloValue Max
exceeded.
Warning 122 SiloValue Min It is shown when the maximum batching value of 0 for silo is calculated.
Warning 123 InsufficientCapacity Batching value is more than the capacity.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 348
Warning 124 Total < Package Total value of the batching cannot be less than the package value.
The filling operation is unable to be started as the weight values for the
Warning 125 Recipe Empty!
silo related to the recipe intended to be batched is entered as zero.
Errors more than detected occurred at any of the silos as the result of
Error 126 ToleranceError!
batching.
It shows that no weight change occurs during the "FillingTimeLapse"
Error 127 Discharge.TimeLapse
period while discharging.
It is shown when the module required for performing the operation is
Warning 130 No Hardware!
not present.
Warning 131 Missing Module Additional module is required for the batching operation.
Product has
Information 224 Product to be filled has been changed.
Changed!
Information 200 Stored New parameter value has been stored to the memory.
Information 201 Please wait... It is displayed during the long operations.
It is displayed when it is intended to perform input and taring operation
Warning 202 Tare Off
when Tare “off” is selected from the menu.
Information 207 vX.Y It is the version number screen. X:Major, Y:Minor
Information 208 Cancelled! Parameter change has been cancelled.
Information 211 It is the terminal serial port test screen.
Information 212 It is the remote display output test screen.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 349
Information 213 It is the E-MDL-COM module serial port test screen.
Information 220 Stopped. Batching operation has been cancelled.
No product can be selected from the menu when one of the product
Information 223 ExternalSelecActive
selection attempts is active.
Information 230 Memory-1 OK Memory unit has been tested successfully.
Information 231 Memory-2 OK Memory unit has been tested successfully.
Information 232 Backing-up... Calibration parameters will be copies to E-ECI-LC card.
Calibration parameters in E-ECI-LC card will be uploaded to the main
Information 233 Receiving...
card.
Information 255 Setting... It is shown when key sensitivity is set.
Report Messages
- 1000 Unknown Error -
- 1001 Key Pad Failure Keypad module error
- 1002 Soft Reset It is reported when ECI is restarted for a reason other than blackout.

ECI Multi Functional Weighing Indicator User Manual (Dynamic) Page 350
A Nişantepe Mh. Gelin Çiçeği Sk. No:36 Çekmeköy 34794 İstanbul T (0216) 585 18 18 F (0216) 585 18 19 W esit.com.tr

ECI Dynamic User Manual - R03