0% found this document useful (0 votes)
42 views2 pages

Report Hasil Analisa Penetapan Total Potassium Menggunakan IC Kation

This document is a report on the analysis of total potassium in samples using IC cation. It details the standard calculations for a potassium standard, including average peak area, standard deviation, and percent recovery. Sample calculations are also reported for two samples of TP-FM 155A021123, listing peak area, weight, assay percentage, and average assay percentage for total potassium. The specifications for potassium content in several product variants are also specified in a table.

Uploaded by

angellapratiwi5
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
42 views2 pages

Report Hasil Analisa Penetapan Total Potassium Menggunakan IC Kation

This document is a report on the analysis of total potassium in samples using IC cation. It details the standard calculations for a potassium standard, including average peak area, standard deviation, and percent recovery. Sample calculations are also reported for two samples of TP-FM 155A021123, listing peak area, weight, assay percentage, and average assay percentage for total potassium. The specifications for potassium content in several product variants are also specified in a table.

Uploaded by

angellapratiwi5
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

TP FM, GW, CG, MA, Ori 1 of 2

TP FM 155A021123

No CCSR CCSR-018-19-00
Report Hasil Analisa Penetapan Total Potassium
Judul Spreadsheet
Menggunakan IC Kation

Tujuan Kalkulasi dan Pembuatan Laporan Analisa Total


Potassium
Instruksi Penggunaan Mengacu pada Metode WI-QA-MA-032-02

Report Hasil Analisa Penetapan Total Potassium Menggunakan IC Kation

Prepared by : Angella Report No : TP - CFM23 - 155A


Date of Analysis : 09/11/23
Sequence Name : 20231109_1 - TP FM 155A021123 (AP)

RSD = STDEV x 100

Standard Calculation

Injection Peak Area


Standard 20 ppm - 1 6.1959
Standard 20 ppm - 2 6.1870
Standard 20 ppm - 3 6.2032
Standard 20 ppm - 4 6.2078
Standard 20 ppm - 5 6.2653
Standard Bracket - 1 6.3538
Standard Bracket - 2 6.1217
Standard Bracket - 3 6.1467
Standard Check 20 ppm 6.1980
Average 6.2102
Stdev 0.0720
RSD 1.16
% Recovery 100.20

Keterangan Standard Standard Check


Brand Supelco Supelco
Batch Nomor HC28254830 HC28254830
Expired date 31/03/26 31/03/26
Preparation date 09/11/23 09/11/23
Quantity (mL or g) 2 2
Prepared by Angella Angella

11/12/2023
2:48 AM C:\Users\IDLFAQCLaboratory\Downloads\TP FM 155A021123
TP FM, GW, CG, MA, Ori 2 of 2
TP FM 155A021123

Spesifikasi Produk

Variant Spesifikasi (%w/w)


Fresh Mint 4.60 - 5.40
Gentle Whitening 4.60 - 5.40
Cool Gel 4.60 - 5.40
Multi action 4.60 - 5.40
Original 4.60 - 5.40

Sample Calculation

Assay of Total
Sample Name Peak Area Weight (g) Potassium Average (% w/w)
(% w/w)
TP - FM 155A021123 (Simplo) 5.7871 1.0000 4.82
4.92
TP - FM 155A021123 (Duplo) 6.0365 1.0000 5.03
Average 5.9118 1.0000 4.9235 NA
Stdev 0.1764 0.0000 0.1469 NA
RSD 2.98 0.00 2.98 NA

Analyzed by, Reviewed by,

Analyst QC Spv

11/12/2023
2:48 AM C:\Users\IDLFAQCLaboratory\Downloads\TP FM 155A021123

You might also like