F601 - Basic Training - EN

The document provides instructions for setting up and using a Flexim Instruments Asia Fluxus F601 ultrasonic flow meter, including selecting transducer types, measuring wall thickness, inputting pipe parameters, configuring output options and electrical outputs, and starting measurement.

Uploaded by

THiz OCtavv
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 66

Flexim Instruments Asia

Fluxus F601 Training


Seite 1
FLUXUS F601

Seite 2
FLUXUS F601 with transport case

Seite 3
Wall Thickness Measurement

Seite 4
FLUXUS F/G601 Power Pack

Seite 5
F601 typical setup

Seite 6
Preparation before starting

> Before starting:

• make sure that the pipe is completely filled with the fluid.

• Choose the right place for measuring point!


> observing inlet / outlet distances to disturbances (elbows, pumps, valves etc.)
according to manual recommendations

• Control pipe surfaces:


> surface should be reasonably clean and smooth

• Transducers should not be mounted in vicinity of welds or rough surface


• > see: recommendations in the manual! <

Seite 7
Selection of the measuring point

Seite 8
Pipe straight length recommendations

Please see further recommendations in the user manual!


Seite 9
Transducer selection for liquids

Shear wave transducers:

• Typically used for liquids


• High Pressure / thick wall gas applications
• High amplitude signals work well on high pressure gas
• Plastic pipes, hoses, other metals than steel

Seite 10
Transducer selection for liquids

Select a transducer for your current application based on diameter (ID):

Seite 11
Transducer selection for liquids

Minimum wall thickness requirement for Shearwave Transducers

For some special Liquid Applications, Lamb Wave Transducers could be an option!
e.g. highly viscous liquids, high flow velocities, changing liquids (e.g. concentration)

Seite 12
F601 Keyboard layout

Seite 13
Keyboard Details

: confirmation of selection or of entered value

: interruption of measurement / go to main menu / press 3x to power off

: Power on / delete or clear values

… : Input of the digit shown on the key

: Scroll upwards through a scroll list


: Activate totaliser measuring (3x totaliser reset)

: Scroll downwards through a scroll list


: Deactivate totaliser measuring (3x totaliser off)

: Scroll to the right through a scroll list


: Start totaliser measuring (positive flow)

: Scroll to the left through a scroll list


: Start totaliser measuring ( negative flow)

Seite 14
Keyboard Details

- Continuation:

: Switch LCD display upper row

: Switch LCD display lower row

: MUX : multiplexer on/off (manual/automatic switching)

: NEXT : switch between channels A/B/Y/Z

: sign minus

: decimal point / turn on LCD backlight

Seite 15
Reset to factory settings (INIT)

Initialization (reset to factory settings)


During an initialization (INIT) of the transmitter, the settings in the program branches Parameter and Output Options and some of
the settings in the program branch Special Funct. are reset to the default settings of the manufacturer.

2 methods:

1. While switching on the transmitter: keep keys BRK and C pressed (long press).

2. During the operation of the transmitter: press keys BRK, C and ENTER at the same time. A RESET is executed. Release
key ENTER only. Keep keys BRK and C pressed.

Download video here → https://flexim.sharefile.eu/d-see22546d28b40fe8


Seite 16
HOTCODES

A HotCode is a digit sequence that activates certain functions or changes settings.

A HotCode can only be entered in the main menu immediately after switching on the
transmitter. The HotCode is not displayed during the input.

Seite 17
Connectors Overview

Seite 18
Connector Adapters Overview

Seite 19
Transducer Connection

- Switch the Flowmeter on: press key

- Connect the transducers to the Flowmeter:

- Shown on FLUXUS:

Seite 20
Menu Overview

Seite 21
Parameters

- Following values are necessary:

• Outer Diameter
• Wall thickness - Input of the Parameters:
- Pipe material
- Liner no / yes
• Medium
• Medium / Temperature
leads to this menu point:

- Hint: unknown Medium?


choose „other medium“ and estimate the speed of sound
(see chart various medium / supplement in manual)
Seite 22
Measure the pipe
circumference and convert to
diameter

Use Vernier caliper on smaller pipes


to measure pipe outer diameter
Seite 23
Wall Thickness

Measuring the pipe wall thickness

1. Connect wall thickness probe to transmitter. Message of probe detected


will be confirmed.

2. Select “Wall Thickness” in “Output Options\Physic. Quant.” for Channel


where probe is connected

3. Select the correct pipe material in PAR menu and press “Enter”.

4. The pipe material sound speed is shown. Press “Enter”


24
Seite 24
Wall Thickness

5. Select “Measuring”. Press “Enter”

6. Apply thin film of acoustic coupling compound on contact surface and


place on pipe surface

The acoustic partition boundary of the probe must be perpendicular to the longitudinal axis of the pipe
25
Seite 25
Wall Thickness

Seite 26
Meter Formula

27
Seite 27
Meter Formula

28
Seite 28
Pipe nominal sizes and schedules

Always use the values in the pipe chart for OD and wall thickness when setting the parameters of the flowmeter!!!

Seite 29 Download here: https://flexim.sharefile.eu/d-s111af1e95fc4689a


Parameters

- just follow these steps, which are automatically displayed:

Seite 30
Output Options

Following settings are necessary:

- Physical quantity (displayed while measuring)


- Units of measurement (displayed while measuring)
- Damping factor (time / average measured values)
- Storing measured values
- Storage rate (storage interval of measured values)

- Input of the output options use:

to go to this menu point

Seite 31
Output Options

- just follow these steps, which are automatically displayed:

Seite 32
Installing Current output

Special Functions > System Settings > Proc. outputs

I1, I2
B1, B2

Seite 33
Installing Binary output

I1, I2
B1, B2

Pulse Alarm

Seite 34
Activation of Current output

4 mA

20 mA

12 mA

Seite 35
Activation of Binary output

As “pulse” output As “alarm” output


if “Impuls” is selected if “Limit” is selected

Do not set a high


Pulse rate!

Check with
Customer!

Seite 36
Output Options

Notes:

- Measured values could only be stored if activated under


Output option > store measured values is done!

- Storing amplitude, measured soundspeed and diagnostics activate under


> Special Functions > System settings > Storing

- Electrical outputs are only available / setable after activating under


Special Functions > System settings > Process outputs!

- An electrical output (e.g. current I1) can only be linked to one flow channel!

- One flow channel can have more than one output!

Seite 37
Special Functions in Short

-Main menu Special Functions

- Settings and choosing functions

- Device info / firmware version

- Store current parameter settings

- Delete parameter records

Seite 38
Special Functions in Short

-Print all measured values

- Delete all measured values


(will be asked: „really delete?“)

- Condition of the Battery Pack and Charging Cycles

- Additional materials can be added

- Additional Medium can be added

Seite 39
Special Functions in short

-Under Menu System Settings following settings are placed:

- Miscellaneous (e.g. contrast)

- Set clock and date

- Library (User Area for Media)

- Dialogs / Menus ( „How the meter talks“ e.g. Units)

Seite 40
Special Functions in short
Under Menu System Settings following settings are placed:

- Activating and Settings of Process Inputs

- Measuring Mode (e.g. Changing gas mode to liquid mode)


- important! Turbulence Mode = OFF

- Activating and Settings of Process Outputs

- Storing (e.g activating „Store Diagnostics“)

- Settings for the Serial Inteface RS232

- Installing/activating snapshots
Signal Snap: only visible under Firmware Version V6.0 or higher!

Seite 41
Start measuring

- Now we are going to start measuring use:

to go to this menu point

Seite 42
Start measuring

- just follow these steps, which are again automatically displayed:

- Now you should mount the transducers on the pipe!

Seite 43
Preparation of the Fastening Shoes

- apply the fastening shoes to the ruler:

- slot for cable left! → ← slot for cable right!

- apply the chains:

Seite 44
Preparation of the Fastening Shoes

- apply the transducers to the fastening shoes (compress the spring!):

- lock:

Seite 45
Preparation of Transducers

- fix the transducers in the recommended distance with the screws:

- apply coupling compound on the contact surface of the transducers:

Seite 46
Coupling Compound (Grease/Gel)

For short duration or temporary measurement set-up only!

coupling compound type N (-30...+130C)

coupling compound type E (-30...+200C)

coupling compound type H (-30...+250C)

Seite 47
Coupling Pads (for fixed installations)

Type: VT - Viton (-10C up to 200C)

Type: SI - Silicone (- 40C up to -10C )

Type: TF - Teflon (200C up to 240C)

Important: do not use coupling gel/grease!

Seite 48
Preparation of the Portable Variofix Rail

Spring clip
Chain tensioner

Ball chain

Variofix rail – 40 cm (L) x 11cm (W)

Attach chain and tensioner to the Position transducer on rail Secure transducer with spring clip
rail around the pipe

Seite 49
Preparation of the Portable Variofix Rail

Seite 50
Preparation of the ratcheting belts

Seite 51
Transducer mounting

- fix the transducers tight to the pipe with a proper alignment:

Seite 52
Transducer mounting

For horizontal pipes: avoid top or bottom mounting

> because of gas bubbles on the top or sedimentation/particles on the


bottom….
Seite 53
Transducer mounting

- Recommended position for horizontal pipes:


> on left or right side, beetween an angle of 45° up to 125° (shown: 90°)

Seite 54
Transducer mounting

For vertical pipes: the position is not really relevant, but care about the alignment!

Seite 55
Please do not do this....

Seite 56
Start measuring

- press:

- Following value should be controlled:


- Amplitude S > II
- Quality Q >IIIIIIII (Signal beam constant in the „rear third“)
- SCNR >> 20 dB / 30 dB (optimum)
- SNR >> 20 dB / 30 dB (optimum)
- Gain <<94dB GAIN
- Sound speed: does the measured value confirm to the real
Sound speed of the fluid
Seite 57
Transducer distance Consistency Check

Seite 58
Assessment of Measurement

How to be sure that the measuring results are trustable ?

1. Sound speed: no strong deviation from the theorietical sound speed of the
fluid
• c- range indication in FLUXUS display

2. SCNR > 30dB for full accuracy


• Available on display at measurement startup
• „green“ arrow in FluxDiag

Seite 59
Diagnostic Display during measuring

- Diagnostic while measuring:

- Sample:

Seite 60
Data logger configuration and activation

Configure data logger options: Enable data logger:

Seite 61
Activate Snapshots

Necessary for special Diagnostics > snap is a reading method of ultrasonic signals

For devices with Firmware V6.0 or higher available under Special Functions!

10

Seite 62
Taking Snapsots
- Start measuring procedure as usual

- Press Button for devices with keypad

- SNAPSHOT (number is shown)

Seite 63
Noisetrek (Hybrid Doppler)

Measuring in NoiseTrek mode allows to measure if the standard Transit Time mode can not measure
due to too much particles’ in the fluid (i.e. gas bubbles and/or solid particles).

To Enable:

Noisetrek switch during measurement: Press “MODE”


button during
measurement

Seite 64
Typically facts of a fail measurement

-Following points should be checked:

- all parameters set correct?


- customer information correct and believable?
- wrong transducer distances? (exact distance absolutely necessary!)
- wrong number of sound paths? (reflection or direct mode?)
- correct alignment of the transducers? (axial and horizontal?)
- pipe system completely filled?
- no transducer coupling (gel/viton rubber) to pipe?
- bad pipe surface? (corrosion, scale or loose paint removed?)
- wrong choice for the measuring place? (Observing inlet / outlet distances?)
- wrong transducer for the pipe size, or the medium?
- values for SNR / SCNR / Gain at optimum values?
- values (bargraph) for quality (Q) and signal (S) high enough?
- values steady shown?

Seite 65
Thank You!

…Questions???

Seite 66