Acid Number of Insulating Oil: Acid Base Titration (Non Aqueous) by Automatic Potentiometric Titrator
Acid Number of Insulating Oil: Acid Base Titration (Non Aqueous) by Automatic Potentiometric Titrator
1.Abstract
To maintain the acid number within specified level is the key of quality control of
the insulating oil. Acid number is expressed by potassium hydroxide in mg required to
neutralize free fatty acid in 1g of oil.
The oil sample is titrated with 0.05mol/L potassium hydroxide ethanol as reagent
and with toluene ethanol (3+2 ) as solvent and with alkali blue 6B ethanol, and the
endpoint is detected on titration curve. The acid number is calculated from the
titration volume of reagent reaching the endpoint.
2.Reference
1) JIS-C2101-1999 Test Method for Electrical Insulating Oil
2) JPI-5S-57-99 Japan Petroleum Institute - Neutralization Test Method for
Electrical Insulating Oil – Total Acid Number Indicator Photometric Titration
3) ASTM D974-06 Standard Test Method for Acid and Base Number by
Color-Indicator Titration
4) BS148-1998 Specification for unused and reclaimed mineral insulating oils for
transformers and switchgear
3.Cautions in measurement
1) The indicator of alkali blue 6B is hard to dissolve in ethanol, and needs to be
stirred more than 2 hours before titration.
2) The indicator is liable to change, and recommended to use within 2 days after
prepared.
3) Be aware the accuracy of measurement depends on the correct amount of
indicator when added during the process of titration.
4.Post-measurement care
No special note
6.Reagent
Reagent : 0.05mo1/L potassium hydroxide・ethanol solution ( f=0.6823 )
Solvent : Toluene + ethanol (3+2) mixture
Indicator : Alkali blue 6B ethanol solution
7.Test procedure
―Preparation―
1) Obtain the factor of reagent using amide sulfuric acid.
2) Dissolve 0.1~0.2g alkali blue 6B in 100mL ethanol, and filtrate as indicator.
3) Calibrate transmissivity with 100mL solvent added with 3mL indicator and 2mL
reagent as 100%T, and with 100mL solvent added with 3mL indicator as 0%T
respectively.
―Measurement―
1) Transfer 20g sample to a 200mL beaker, and weigh it on a balance.
2) Add 100mL solvent and 3mL indicator.
3) Titrate with 0.05mol/L potassium hydroxide ethanol.
4) Perform a blank test in advance to obtain blank level.
8.Formula
Acid number ( mgKOH / g ) = ( EPl - BLl ) × TF × Cl × K1 / S
EPl : Titration volume ( mL )
BLl : Blank level ( 0.2999mL )
TF : Reagent factor ( 0.6823 )
Cl : Concentration conversion coefficient ( 56.l g/mL )
(equivalent to potassium hydroxide in 1mo1/L potassium hydroxide ethanol)
Kl : Unit conversion coefficient ( 0.05 )
S : Sample size ( g )
The following parameters and titration curve were obtained by AT-610 titration system:
-Titration parameter- -Titration curve-
Model : AT-610
Method No. : 01
Titr.mode : Intermit
Titr.form : EP Stop
10.Summary
Acid number is expressed by potassium hydroxide in mg required to neutralize free
fatty acid in 1g of oil. To maintain the acid number within specified level is the key of
quality control of the insulating oil.
The measurement results of the sample this time show a good repeatability.
The automatic potentiometric titration system provides stable and reliable
measurements. The acid number of insulating oil is measured by any of the models
shown below.
Awarded Product of Supreme • Compact and cost performance • Low cost and high performance
Technology from Kyoto City model • Easy view with back light LCD
• Easy key entry by touch panel of • PC card expands data memory for • GLP/GMP conformed model
large color LCD (8-inch wide) convenience and versatility.
• Simultaneous titration in parallel
• Both potentiometric and Karl
Fischer moisture titration
(coulometric・volumetric) can be
performed at a time.