DelWorks Calibration 2018
DelWorks Calibration 2018
3. If it is a multi detector system, select the detector you would like to calibrate, if it is a single detector system it will not
matter which receptor is selected.
NOTE:
Detector needs to be setup at 72inch SID with collimation wide open covering the entire detector
and nothing in the field, with NO grid. Place detector on top of a lead apron if detector is calibrated
on the floor.
9. If the technique/dose is not accepted a message like the image below will
display. Based on whether the dose is higher or lower than the allowable
range you may need to adjust for a higher or lower technique. Click the OK
and adjust the technique. Click the Continue button to retry the next
exposure. Again, when the indicator bar in the button right corner turns
green and starts counting down, click and hold all the way down on the hand
switch until the exposure is terminated.
Please see the Reference Chart Below for Exposures and Dose Levels for each of the
Exposure Sets for the calibration:
Please make sure the X-Ray generator is calibrated before the detector is to be
calibrated.
If you have any issue feel free to give UMG Technical
Support a call 1-800-261-9808