CATIONIC
Applications
• Power
CONDUCTIVITY
ANALYZER
Specifications*
Cell Constant k 0.01 cm-¹ Analog Output Linear, Logarithmic, Bi-linear, PID
Functional Mode
Range Specific Conductivity:
0.01 to 200 µS/cm Communication Five 4-20 mA Outputs, MODBUS
RS232/RS485, PROFIBUS DPV1,
Measurement Range 2 Specific Resistivity:
or HART Communications
5 to 100000 kΩ x cm
Electrical EMC
Accuracy ± 1% of displayed value
Certifications
pH Range 7 to 10 for Ammonia CE compliant for conducted and
radiated emissions:
7 to 10.7 for Sodium Hydroxide
- CISPR 11 (Class A limits)
Operating -20 to 60 °C at 0 to 95% RH
Temperature Range (non-condensing) - EMC Immunity EN 61326-1
(Industrial limits)
Sample Input 1/4 inch exterior diameter tubing
Sample Output 1/2 inch interior diameter tubing Safety
with barbed connector CAN/CSA C22.2 No. 61010-1
Temperature No, Automatic, and Manual cETLus safety mark for:
Compensation
- General Locations per ANSI/UL
Temperature Sensor PT 100 Accuracy: < ± 0.2°C 61010-1 & CAN/CSA C22.2. No.
Power Requirements 100 - 240 V AC, 24 V DC 61010-1
(Voltage) Enclosure Rating NEMA 4X/IP66
Power Requirements 50 - 60 Hz Flow 83 to 333 mL/min (5 to 20 L/hr)
(Hz)
*Subject to change without notice.
Relays Four electromechanical SPDT
(Form C) contacts, 1200 W, 5 A
Analog Outputs Five available 0/4 to 20 mA isolated
current outputs, max 550 Ω ,
Accuracy:± 0.1% of FS (20mA)
at 25°C, ± 0.5% of FS over -20°C
to 60°C range
hach.com
9523sc Cationic Conductivity Analyzer 3
Principle of Operation
Measurement of pH in environments of low conductivity using the standard potentiometric method (glass electrode + reference) is
extremely delicate and not very accurate because it is proportional to the concentration logarithm. It also requires a more frequent
calibration to compensate for variations in the measurement chain (junction potential, degradation of the glass membrane).
On the other hand, measurement of conductivity in these environments is a lot more reliable and more accurate as it is directly
proportional to the concentration in impurity, and in addition requires little or no maintenance.
Therefore, given the relationship between the pH and conductivity of a product, the conductivity measurement can be used to
determine a precise pH.
If the product contains impurities (generally in the form of salts), this calculation cannot be applied. The principle is then to transform
the salt into acid by passing it through a cationic resin and, given the relationship of the conductivity between the acid and the
corresponding salt (always around 3), to determine the conductivity originating only from the conditioner:
∆C = Conductivity before resin (C1) – Conductivity after resin (C2) / A
and
pH = f(∆C)
Note: The calculated pH is the pH of the sample at the analyzer inlet (channel 1). The 9523sc analyzer does not calculate the pH
downstream of the resin cartridge.
Dimensions
hach.com
4 9523sc Cationic Conductivity Analyzer
Ordering Information
Complete Analyzers
9523.99.09P2 9523sc Cationic Conductivity Analyzer, 5x 4-20mA Out, AC-DC
9523.99.01P2 9523sc Cationic Conductivity Analyzer, Modbus 232/485, AC-DC
9523.99.03P2 9523sc Cationic Conductivity Analyzer, Profibus, AC-DC
9523.99.05P2 9523sc Cationic Conductivity Analyzer, HART, AC-DC
Communication and Module Options
9013200 Modbus 232/485 Module
9173900 Profibus DP Module
9328100 HART Module
9525800 Analog Conductivity Module for Polymetron Sensors
Accessories and Consumables
08310=A=0000 8310 Conductivity Sensor, K= 0.01 cm-¹
09523=A=7000 Spare Resin Cartridge (includes resin inside)
09523=A=7010 Resin Kit (includes 2 filters, 2L of resin, funnel, and instructions)
09123=A=8001 Electrode Cable (1m)
LIT2819 Rev 1