MTS Sector1 Summary 2015

Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

Specifications for Sector 1 MTS hydraulic system

Summary: This is an MTS model 858 system, with a customized baseplate to allow the unit to be
placed on a manipulator system, such that samples can be arbitrarily moved relative to the x-ray beam
for in situ studies. Samples can be loaded in monotonic tension/compression and fatigue. Various
sample grips exist, others are under consideration as detailed herein, and users are encouraged to
consult/collaborate with beamline staff if further custom designs are needed. The system is designed
with a heat shield for elevated temperature applications. Sample heating can be provided with an
beamline-owned infrared heater (from Research Incorporated) which is nominally rated to ~1100C
(actual max temperature is geometry-dependent), or with customized heaters developed by users. The
system can be operated in 2 locations, either inside the x-ray hutch (1-ID-E) or downstream of the hutch.

1. Details of model 858 system


see http://www.mts.com/stellent/groups/public/documents/library/dev_002083.pdf for further info

1.1 Load frame, model 359S.


- Nominal dynamic load rating: ±25 kN (±5,500 lbs).
- Overall Weight: 1,979.46 N (445lbs.)
- Overall height maximum: 1,686.6 mm (48.758 inches).
- Base dimensions:
- Width: 625 mm (24.6 inches).
- Depth: 254 mm (10.0 inches).
- Custom designed T-slotted base plate to attach grips and bend fixture.

1.2 15 kN Actuator, 4 in. Stroke, M12x1.25 mm Mounting Threads


- This actuator provides 15 kN, 10 kN or 5 kN force capacity at 21 MPa, 14 MPa or 7 MPa pump pressure
respectively.
- Position Transducer, 100mm

1.3 661.19F-01 Axial Load Cell


- Force capacity 15 kN, fatigue rated

1.4 661.18F-01 Axial Load Cell


- Force capacity 1 kN, fatigue rated
- Can be installed in series with 661.19F-01 axial load cell for low-load applications

1.5 Manifold Assembly-Proportional Standard, Axial


- Close-coupled accumulators allowing maximum high frequency response of servovalve(s).
- Maximum operating pressure: 21 MPa (3000 psi).
- Off/Low/High pressure control.
- Slow pressure turn on.
- Fast emergency unload for system depressurization.
- Optional pilot pressure available.

1.6 252.23G-04, 5 gpm (19 lpm) 5 port High Response Servovalve

1.7 Model 505.07 Hydraulic Power Unit


- Silent Flo Series.
- 60 db(A) noise fully compensated.
- 6 GPM ( 22.7 LPM) nominal flow rating.
- 3000 PSI (21 MPa)
- 220 VAC, 3 PH, 50 HZ incoming power.
- 24 VDC control.
- Water cooled.
- 46 gallon reservoir.
- 3 micron return filtration.

1.8 MTS control software and hardware:


- FlexTest SE Controller - PLUS
1.8.1 MTS Model 493.02 FlexTest SE 1-Channel Chassis
- The MTS FlexTest SE digital controller is a flexible and easy-to-use servo-controller for general testing
applications.
- FlexTest SE hardware is VME-based including an MVME processor to provide real-time closed-loop control. The
controller includes one analog input for high level data signal or external program, three analog outputs, and four
digital inputs and four digital outputs (one pair dedicated for interlock).
- The Basic model can operate stand-alone (without PC) or with PC automation. A dedicated PC is included as a
standard feature with the Plus model. The Plus model also features a higher performance controller processor,
capable of update rates up to 6KHz
- The higher performance processor is available as an option on the Basic model controller. If a stand-alone controller
is desired for low-cost initial use, and TestWorks 4 or other application software is anticipated in the future, the
Basic model should be ordered with the 6KHz processor.
- Capabilities include:
- Function generation of monotonic ramps and cyclic waveforms using sine, square and triangular shapes can be
generated.
- Graphic color display with optional general purpose scope; meters, which provide digital displays of specified
measurements.
- Auto zero, bumpless start, hydraulics on mode-switching.
- Ability to save and restore PID tuning settings.
- Adaptive controls compensation; Peak-Valley and Null Pacing.

1.8.2 Multi-position Handle


1.8.3 MTS Model 493.25 Universal Conditioner
- Provides transducer conditioning for AC or DC transducers.
- Normalized with an on board reference for easy portability of calibrations.
- Low noise, low drift, high accuracy signal conditioning.
- Computer control of range, transducer zero, excitation, filter.
- Designed for full range calibrations.
- Shunt calibration for 100% range.
- Supports strain gage bridge completion.
- Excitation loss detection.
- Computer controlled limit detection.
- Excitation frequency: 9.83 kHz.
1.8.4 MTS Model 493.47 Encoder Interface Board
1.8.5 493.14 Two Stage Valve Driver
- Provides servovalve drive signal for 2-stage servovalves.
- Supports single or dual MTS Model 252 style servovalves.
- Digitally controlled fault protection logic.
- 16 bit resolution on valve D/A.
1.8.6 Acceleration Compensation Module
- Provides signal conditioning with line current source for compensated load cell.
1.8.7 Remote Station Control Panel
- Provides signal control away from the personal controller. This is ideal for controllers enclosed in X-ray regions.
1.8.8 Remote E-Stop
- Provides emergency shutdown of system away from the personal controller. This is ideal for controllers enclosed in
X-ray regions.
1.8.9 FlexTest SE PC Automation Kit
- This kit includes the MTS computer referenced elsewhere in this document. Installed on the computer is Model
793.00 Software and 100MBPS Ethernet network card / cable.
- Enables use of Basic TestWare to setup and run cyclic and monotonic tests, and acquire and store test data.
1.8.10 Compaq Pentium IV / 2.66 GHz
- Compaq/HP Pentium IV/2.66 GHz
- 512MB SDRAM Memory, 512KB L2 CACHE Memory.
- 40GB SMART III Ultra ATA/100 Hard Drive, 1.44MB Diskette Drive.
- CD/RW, X48
- Windows XP.
1.8.11 793.10 MultiPurpose TestWare (FTSE/TSIIs)
- Multi-Purpose TestWare enables integration of function generation, data acquisition, and test sequencing with data
stored to disk in ASCII format for analysis with third party tools such as Excel (TM).
- Multi-Purpose TestWare functions include:
- Data acquisition (Timed, Peak/Valley, Level Crossing, Max/Min, Cyclic/Logarithmic).
- Function generation up to 100 Hz on all channels.
- Sine, square, triangle, ramp, hold, processes, and ability to play digitized profiles.
- Soft start/stop available on all channels.
- Limit sensing, sequencing triggers, and interface to 32 Digital I/O (16 in/16 out).
1.8.12 MTS Model 493.25 Universal Conditioner (Strain)
- Provides transducer conditioning for AC or DC transducers.
- Normalized with an on board reference for easy portability of calibrations.
- Low noise, low drift, high accuracy signal conditioning.
- Computer control of range, transducer zero, excitation, filter.
- Designed for full range calibrations.
- Shunt calibration for 100% range.
- Supports strain gage bridge completion.
- Excitation loss detection.
- Computer controlled limit detection.
- Excitation frequency: 9.83 kHz.
1.8.13 Cable To Epsilon Extensometer and Factory Calibration
1.8.14 Model 493.45 Analog Input with BNC adapter leads
- Provides set of 6 auxiliary input channels for high level (+/- 10V) analog signals which can be used for control and
data acquisition.
- Includes adapter supporting BNC connectors on six 'pig-tail' leads.
- Limit of 1 module per controller.
1.9. Grips
1.9.1 One Set of Custom Fatigue Grips for Threaded samples, M10-1.5mm threads.
- Unimet construction
- Rated to 1000C
- See section 4 for sketch
1.9.2 Model 642.001A-02 3 & 4 Point Bend Fixture
- 170 to 880 N Dynamic Force Capacity, 570 N to 2.89 kN Static Force Capacity, depending on roller diameter.
- 25 mm specimen width.
- ASTM D-790, ASTM D-2344 Compliant.
- Assembly includes 1, 2, 3, 4 and 5 mm rollers.
- Lower Span: 4 to 50 mm (1 mm roller), 13.7 to 59.7 mm (5 mm roller).
- Upper Span: 4 to 15.4 mm (1 mm roller), 13.7 to 25 mm (5 mm roller).
- Temperature Range: -130 to 177 deg C.
- Includes M6x1 Cap Screws for Mounting.
- More info: http://www.mts.com/downloads/642.001_low.pdf
2. Theoretical performance specs of MTS
- NOTE: MTS does not guarantee these specs for the custom configuration used here. Rather, these can be treated as
‘best case’ scenario.
- The following information can be found at:
- http://www.mts.com/stellent/groups/public/documents/library/dev_002083.pdf
From MTS:
4. Additional schematics and photos
4.1. Details of grip assembly for custom threaded samples
4.2 Front and side views of unit, with bend grips shown
4.3 Photos of unit & manipulator, April 2007