Peak Integration2
Peak Integration2
A b s tra c t:
The time required to analyze large quantities of data obtained from the increasingly quick analyses and increasingly prevalent simultaneous
multicomponent analyses in recent years has become a challenging issue. In addition, the strengthening of regulations associated with data
integrity has given rise to demands for automated and simplified manual methods for integrating the areas of problematic chromatogram
peaks. Herein, we introduce a new peak integration algorithm for LabSolutions, i-PeakFinder, which is designed to resolve these issues.
1 . I n tro d u c tio n der and main peaks using traditional peak integration methods, i-Peak-
Finder can automatically detect shoulder peaks while maintaining con-
i-PeakFinder, a new peak integration algorithm for LabSolutions, is a sistent peak detection sensitivity throughout the entire chromatogram.
completely automated integration function that can detect peaks with In general, it is difficult to automatically detect considerably small shoul-
high levels of accuracy without requiring special parameter adjust- der peaks, as shown in Fig. 2; however, i-PeakFinder can even detect
ments. In addition, this algorithm has adjustable parameters that such peaks automatically based on the threshold judgment.
allow the integration function to be applied to a wide range of com-
plex chromatogram patterns. Furthermore, the adjustable parameters
allow to output highly accurate peak integration results from large
quantities of data, even for batch analysis.
2 . F e a tu re s o f i-Pe a k Fin d e r
With Shimadzu's traditional peak integration methods or competitor
methods, some complex chromatograms require time programming in
addition to the typical parameter adjustments for peak integration. Highly accurate detection of shoulder peaks
However, i-PeakFinder can perform peak integration of these chro-
matograms just with simple parameter adjustments. This peak inte-
gration function has the following features:
• Highly accurate detection of shoulder peaks
• Simple adjustment of peak-baseline processing
• Improved peak-baseline positioning that results in improved reproducibility
• Performing accurate peak integration even with variabilities caused
by baseline drift
Peak detection at low S/N levels Automatic determination of
Shimadzu also emphasizes on compatibility, so LabSolutions can also
baseline undulation and drift
be used with the traditional peak integration methods (Chromatopac
mode). Switching between the traditional peak integration and
i-PeakFinder methods during analysis is easy, allowing the user to
select an appropriate peak integration method for the circumstances.
This includes selecting a traditional method for compatibility with
past data. Fig. 1 shows examples of using the completely automated
integration function to analyze typical peaks.
2-1. Highly Accurate Detection of Automatic tracking of peak Automatic tracking of noise
Shoulder Peaks width variation intensity changes
i-PeakFinder can accurately detect shoulder peaks. In cases wherein Fig. 1 i-PeakFinder's Completely Automated Integration Function
manual peak integration is required to distinguish and detect the shoul-
Main peak
Impurity
[Detection Threshold] = 10
1.5 3
2
1.0
1
0.5
0
0.0
12.5 15.0 17.5 min
17.5 20.0 min [Baseline Type] Not configured
Normal shoulder peaks can be detected with no special parameter adjustment mV
mAU 4
Impurity
Main peak
2.0
[Detection Threshold] = 10
3
[Detection Threshold of
1.5 Unseparated Small Peak] = 0.3
2
1.0 1
0
0.5
Main peak
Impurity
3
Fig. 2 Examples of Shoulder Peak Detection
2
2-2. Simple Adjustment of Peak-baseline
Processing 1
0
Accurate detection of impurity peaks is essential for pharmaceutical
quality control and other applications. Impurity peaks are often fused at
12.5 15.0 17.5 min
the base of a major peak, and quantitative results obtained via area nor-
[Baseline Type] Base to Base
malization can vary depending on the method used for peak-baseline
processing. The method used for peak-baseline processing also differs Note: Other than peak integration range and peak-baseline processing type, all settings are at default.
0
(1) To Detect Peaks Not Affected by Noise [Minimum Half Width]
1.00 1.05 1.10 min Smoothing is sometimes performed on chromatograms obtained via
Fig. 5 Chromatogram with Peak Tailing (N=5)
LCMS. If the noise frequency is close to the peak frequency, then it be-
comes difficult to determine peaks automatically and single peaks may
be recognized as multiple peaks. In this situation, configuring the mini-
Table 2 Comparison of Area Reproducibility Between the Traditional
mum half width setting ignores noise smaller than that value and en-
Method and i-PeakFinder sures that peaks with a FWHM value above the set minimum value are
i-PeakFinder Traditional method detected among the peaks with a wide wave profile due to smoothing.
%RSD 0.106 0.275
Fig. 8 shows the difference made by increasing the minimum FWHM
value. This feature is useful when noise can be observed in a peak.
Peak-baseline Type: Base to Base Width: 1 s
Processing
Peak-Baseline Height* 2 Slope: 2000 uV/min
* Described in more detail later in the article
3 . S e tti n g P a ra me te rs
i-PeakFinder not only performs highly accurate peak integration with de-
14.00 14.50 15.00 14.00 14.50 15.00
fault settings but also includes more detailed settings for adjusting the peak [Minimum Half Width] Not configured [Minimum Half Width] = 2
detection conditions (peak width and peak-baseline start and end heights)
and peak-unifying conditions that determine how to combine multiple
peaks. The most important of these settings will be explained here.
10 10
5 5
0 0
1.0 2.0 min 1.0 2.0 min
[Detection Threshold] = 5 (default value) [Detection Threshold] = 2000