2.patient Monitor Creative (K15)
2.patient Monitor Creative (K15)
2.patient Monitor Creative (K15)
ECG CO2
K15
Patient Monitor
Input dynamic range: ±(0.5mVp~5mVp) Technique: Infrared optical method
Differential input impedance: ≥10MΩ Sampling mode: Sidestream or Mainstream
Bandwidth: 0.05~150Hz (Diagnostic) Measuring range: 0~150mmHg
0.5~40Hz (Monitoring) Measuring accuracy: 0~40mmHg ±2mmHg
1~20Hz (Operation) 41~70mmHg ±5% of reading
CMRR: ≥90dB (Diagnostic) 71~100mmHg ±8% of reading
≥105dB (Monitoring & Operation) 101~150mmHg ±10% of reading
Sensitivity selection: ×1/4, ×1/2, ×1, ×2, ×4 and Auto Flow rate: 50ml/min ±10 ml/min (Sidestream)
Sweeping speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
HR measuring range: 15~350bpm Cerebral State Monitoring (CSM)
HR accuracy: ±1% or ±2bpm, whichever is greate EEG sensitivity: ±400µV
Pacemaker pulse detection and rejection function Noise level: <2µVp-p, <0.4µV rms (1~250Hz)
CMR R: >140dB
Input impedance: >50Mohm
RESP CSI and update: 0-100. filter: 6-42Hz, 1 sec. update
Measuring range: 0~120rpm EMG%: 0-100 (logarithmic)
Measuring accuracy: ±5% or ±2 rpm, whichever is greater filter: 75-85 Hz, 1 sec. update.
BS%: 0-100. filter: 2-42 Hz, 1 sec. update
TEMP
IBP
Measuring range: 21.0~50.0℃
Technique: Strain gauge transducer
Measuring accuracy: ±0.2℃from 25~45℃
Input sensitivity: 5µV/V/mmHg
Measuring range: -50~300mmHg
NIBP
Measuring accuracy: ±2% or ±4mmHg, whichever is greater
Technique: Oscillometric method Measuring positions: ART, RAP, PA, LAP, CVP
Typical measurement time: <30 seconds (adult cuff) ICP, AUXP1, AUXP2
SYS: 40~275mmHg (Adult) Calibration: zero calibrating
NIBP measuring range: 40~200mmHg (Pediatric)
40~135mmHg (Neonate) Cardiac Output (C.O.)
DIA: 10~210mmHg (Adult) Blood temperature
NIBP measuring range: 10~150mmHg (Pediatric) measuring: range: 23-43℃, accuracy: ±0.5℃
10~95mmHg (Neonate) Injecta temperature
MAP: 20~230mmHg (Adult) measuring: range: 0-20℃, accuracy: ±0.5℃
NIBP measuring range: 20~165mmHg (Pediatric) Measuring range: 0.2~20 L/min
20~110mmHg (Neonate) Measuring accuracy: ±0.2 L/min or ±10%, whichever is greater
NIBP measuring accuracy: Mean difference: ±5mmHg
Standard deviation: 8mmHg
NIBP measurement mode: Manual, Auto, STAT, Multi-cycle mode Other Specifications
Auto measuring intervals: 1-480min Power supply: AC 100V-240V, 50/60Hz, 60VA
Built-in lithium battery: 11.1V/4400mAh
Display: 15 inch TFT display
SpO2 Alarming method: 3 levels audible-visible alarm
Technique: Dual-wavelength optical method Networking: Ethernet
Measuring range: 0%~100%
Measuring accuracy: Arms is not greater than 2% for Standard configuration
SpO2 range 70~100%. ECG, Respiration, SpO2, PR, NIBP, Temperature
PR measuring range: 30~250bpm
PR measuring accuracy: ±2bpm or ±2%, whichever is greater
Options
Low perfusion performance: As low as 0.3%.
2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, 12-lead ECG, Cardiac Output
Cerebral State Monitoring, Central Monitor Station, Touch Screen
Patient Monitor
Accessory box for standard configurations
cycles up to 5-phase
Data export and software upgrade * Drug concentration calculation
* Renal function calculation
Software upgrade
SpO2 sensor NIBP cuff ECG cable Temperature probe HL7 protocol to connect
hospital system
Bed to bed view via
central monitor station