Ilovepdf Merged
Ilovepdf Merged
Medical imaging techniques are the special non-invasive procedure for obtaining images of
the internal structures or organs of the body. The image is not exactly like a real photo but
one which highlights certain required regions of the organ. The images of heart, brain, blood
vessel, bone, muscle, etc are obtained either using x-rays, ultrasonic waves, radio frequency,
infrared or radioisotopes substances. Medical imaging techniques are classified as:
Radiography
Computed Tomography
Digital Subtraction Angiography
Ultrasonography
Positron Emission Tomography
Medical Thermography
Magnetic Resonance Imaging
Nuclear Medicine
Others
The main objectives are:
1. To produce / reconstructed images of the internal structure or organs of the human
body
2. Helps to interpret and facilitates diagnosis
X-RAYS / RADIOLOGY
X-Rays were discovered accidentally by Rontgen on November 8, 1895 when he was
experimenting with a partly evacuated cathode ray tube. The tube was completely wrapped in
black paper to keep the room in total darkness. So he noticed the glow of a plate covered with
fluorescent material lying nearby. Further investigations showed that this “X-light” was able
to penetrate solid material such as wood, aluminium and even hand, showing the bones on a
photographic plate.
Roentgen's remarkable discovery was one of the most important medical advancements in
human history. X-ray technology lets doctors see straight through human tissue to examine
broken bones, cavities and swallowed objects with extraordinary ease. Modified X-ray
procedures can be used to examine softer tissue, such as the lungs, blood vessels or the
intestines.
Roentgen placed various objects between the tube and the screen, and the screen still glowed.
Finally, he put his hand in front of the tube, and saw the shadow of his bones projected onto
the fluorescent screen. So he not only discovered X-rays but also, their most beneficial
application.
X-rays from about 0.12 to 12 keV are classified as soft X-rays, and from about 12 to 150 keV
as hard X-rays, due to their penetrating abilities. X-rays are generated by an X-ray tube, a
vacuum tube that uses a high voltage to accelerate electrons released by a hot cathode to a
high velocity. The high velocity electrons collide with a metal target, the anode, creating the
X-rays. In medical X-ray tubes the target is usually tungsten or a more crack-resistant alloy of
rhenium (5%) and tungsten (95%), but sometimes molybdenum for more specialized
applications, such as when soft X-rays are needed as in mammography. In crystallography, a
copper target is most common, with cobalt often being used when fluorescence from iron
content in the sample might otherwise present a problem. The maximum energy of the
produced X-ray photon is limited by the energy of the incident electron, which is equal to the
voltage on the tube, so an 80 kV tube cannot create X-rays with energy greater than 80 keV.
According to the quantum theory, electromagnetic radiation consists of “photons” which are
conceived as “packet of energy” and expressed as
E = h ν = h. c/λ
Electromagnetic radiation
X-rays are produced when an accelerated electron with tremendous velocity with tremendous
velocity bombard (strike) suddenly on a metal target. The kinetic energy of the electrons is
converted into X-rays (1%) and into heat (99%).
Kinetic energy = charge of electron x voltage
eV = hf = h c/λ
h. c/λ
The hole so created in the K-shell is most likely to be filled by an electron falling in from the
L-shell with the emission of a single X-ray photon of energy equal to the difference in the
binding energies of the two shells, E k-EL. The photon is referred to as Kα radiation.
Alternatively, but less likely, the hole may be filled by electron from M-shell with the
emission of a single X-ray photon to energy, EK-EM and referred to as Kβ radiation.
There is also L-radiation produced when a hole is created in the L-shell is filled by an
electron from further out. In case of tungsten, these photons energy have only 10 KeV of
energy insufficient to leave the X-ray tube assembly, and so they play no part in radiation.
The X-ray photons produced in an X-ray tube in this way has a few discrete or separate
photon energies and constitutes a line spectrum.
PROPERTIES OF X-RAY:
X-ray photons carry enough energy to ionize atoms and disrupt molecular bonds. This makes
it a type of ionizing radiation, and therefore harmful to living tissue. A very high radiation
dose over a short amount of time causes radiation sickness, while lower doses can give an
increased risk of radiation-induced cancer. In medical imaging this increased cancer risk is
generally greatly outweighed by the benefits of the examination. The ionizing capability of
X-rays can be utilized in cancer treatment to kill malignant cells using radiation therapy. It is
also used for material characterization using X-ray spectroscopy.
Hard X-rays can traverse relatively thick objects without being much absorbed or scattered.
For this reason, X-rays are widely used to image the inside of visually opaque objects. The
most often seen applications are in medical radiography and airport security scanners, but
similar techniques are also important in industry (e.g. industrial radiography and industrial
CT scanning) and research (e.g. small animal CT). The penetration depth varies with several
orders of magnitude over the X-ray spectrum. This allows the photon energy to be adjusted
for the application so as to give sufficient transmission through the object and at the same
time good contrast in the image. X-rays have much shorter wavelength than visible light,
which makes it possible to probe structures much smaller than what can be seen using a
normal microscope. This can be used in X-ray microscopy to acquire high resolution images,
but also in X-ray crystallography to determine the positions of atoms in crystals.
After ejection of the electron, the neutral atom becomes a positively charged ion with a
vacancy in an inner shell that must be filled with a nearby less tightly bound electron. The
kinetic energy of the ejected photoelectron (Ee) is equal to the incident photon energy (E 0)
minus the binding energy of the orbital electron (Eb). Ee = E0-Eb.
Photoelectric Absorption
These two particles form the pair referred to in the name of the process. At least 1.022 MeV
of photon energy is necessary as the resting mass (using E=MC2) of the electron and positron
expressed in units of energy 0.511 MeV each (0.511 MeVx2 = 1.022 MeV). Photon energy in
excess of 1.02 MeV appears as kinetic energy which may be distributed in any proportion
between the electron and the positron.
The electron and positron lose their kinetic energy via excitation and ionization, when the
positron comes to rest it interacts with an electron. Then both particles undergo mutual
annihilation, with the appearance of two annihilation photons each with an energy of
0.511MeV travelling in opposite directions. Pair production becomes more likely with
increasing atomic number and increasing photon energy. Pair production has no importance
in diagnostic X-ray imaging because of the extremely high energies required for this
interaction to occur.
Pair Production
X-rays have several fates as they traverse tissue. These fates fall into 3 main categories.
No interaction: X-ray passes completely through tissue and into the image recording device.
Complete absorption: X-ray energy is completely absorbed by the tissue. This produces
radiation dose to the patient.
Partial absorption with scatter: Scattering involves a partial transfer of energy to tissue, with
the resulting scattered X-ray having less energy and a different trajectory. This interaction
The probability of X-ray interaction is a function of tissue electron density, tissue thickness,
and X-ray energy (kVp). Electron dense material like bone and contrast dye attenuates more
X-rays from the X-ray beam than less dense material (muscle, fat, air). The differential rate of
interaction provides the contrast that forms the image.
In this example, 900 X-rays are capable of penetrating the soft tissue, while only 400
penetrate the bone (Higher electron density compared with soft tissue). The contrast between
the bone and soft tissue is (900-400)/900 = 0.56.
Tissue Thickness:
As tissue thickness increases, the probability of X-ray interaction increases. Thicker body
portions remove more X-rays from the useful beam compared to thinner portions. In
fluoroscopy, this effect must be compensated for while panning across variable tissue
thickness to provide consistent information to the image-recording device.
Note that on average, only 1 percent of the X-rays reach the image-recording device (e.g., image intensifier,
film), yielding useful information. Thus, 99 percent of the X-rays generated are either absorbed within the
patient (patient radiation exposure) or are scattered throughout the examination room (staff radiation exposure).
Energy:
Higher kVp X-rays are less likely to interact with tissue and are described as more
"penetrating." Increasing kVp, thereby generating more penetrating radiation, reduces the
relative image contrast (or visible difference) between dense and less dense tissue.
Conversely, less radiation dose results to the patient since less X-rays are absorbed. The X-
rays that do not reach the image recording device are either absorbed in the patient (patient
radiation dose) or are scattered throughout the exam room (staff radiation dose)
RADIOGRAPHS
A radiograph is an X-ray image obtained by placing a part of the patient in front of an X-ray
detector and then illuminating it with a short X-ray pulse. Bones contain much calcium,
which due to its relatively high atomic number absorbs x-rays efficiently. This reduces the
amount of X-rays reaching the detector in the shadow of the bones, making them clearly
visible on the radiograph. The lungs and trapped gas also show up clearly because of lower
absorption compared to tissue, while differences between tissue types are harder to see.
In medical diagnostic applications, the low energy (soft) X-rays are unwanted, since they are
totally absorbed by the body, increasing the radiation dose without contributing to the image.
Hence, a thin metal sheet, often of aluminium, called an X-ray filter, is usually placed over
the window of the X-ray tube, absorbing the low energy part in the spectrum. This is called
hardening the beam since it shifts the centre of the spectrum towards higher energy (or
harder) X-rays. To generate an image of the cardiovascular system, including the arteries and
veins (angiography) an initial image is taken of the anatomical region of interest. A second
image is then taken of the same region after an iodinated contrast agent has been injected into
the blood vessels within this area. These two images are then digitally subtracted, leaving an
image of only the iodinated contrast outlining the blood vessels. The radiologist or surgeon
then compares the image obtained to normal anatomical images to determine if there is any
damage or blockage of the vessel.
X-rays are generated when accelerating electrons interact with matter. In a radiographic X-ray
tube, energetic electrons interact with a target material and a portion of the kinetic energy of the
electrons is converted to electromagnetic radiation, or X-rays. X-rays are produced in a specially
constructed high vacuum diode tube (gas pressure near about 10 -2 mm of Hg). The vacuum tube
consists of a filament, a cathode and an anode i.e. a heavy metal target. The tungsten wire
filament is heated by current supplied by a step-down transformer and ultimately heated the
(nickel) cathode. The heated cathode emits copious thermions which is directed towards the
target anode by a high potential difference setup between the cathode and anode by means of a
step-up transformers i.e. electrostatically focused on a target anode. The excited accelerated
electrons with tremendous velocity strike the target anode (tungsten metal). Part of this energy is
converted into X-rays. The anode is kept at 450 so that photons will come out and collect at 900.
The electron beam intensity and hence the X-ray intensity can be controlled by varying the
filament heating current. The quality of X-ray output is also controlled by altering the potential
difference across the tube. This radiation is filtered and collimated to produce optical contrast
relative to the patient dose.
[The interactions of the accelerated electrons with the tungsten of the target, two possibilities for
interaction are common. In the first, the accelerated electron interacts directly with an electron in
an orbital shell of a tungsten atom. The orbital electron is displaced but the orbital gap is rapidly
filled by an electron from a more distant orbit. The difference in the energies of the two electron
orbits is radiated as an X-photon the energy of which is characteristic for the specific element
and the specific orbital shell. The other and more frequent interaction is termed Bremsstrahlung.
This German word means "braking" radiation and that describes what happens to the incoming
electron. In this interaction, the accelerated electron passes relatively close to the nucleus of a
tungsten atom in the anode. The path of the accelerated electron is affected by the nucleus with a
resulting change in direction (decelerated) and the kinetic energy of the electron is dissipated.
The difference in the kinetic energy before and after interacting with the nucleus is radiated as an
X-photon. This type of interaction (Bremsstrahlung) can result in X-photons of almost any
energy, limited only by the tube potential. A typical radiation spectrum from a medical X-ray
tube is illustrated where, the most frequent photon energy will be approximately one third (1/3)
of the tube potential voltage. There will be superimposed "characteristic" photon spikes which
correspond in energy to differences of the specific electron orbits.]
The second control of the output of the X-ray tube is called the mA (milliamperage) control.
This control determines how much current is allowed to flow through the filament which is the
cathode side of the tube. If more current (and therefore more heating) is allowed to pass through
the filament, more electrons will be available in the "space charge" for acceleration to the target
and this will result in a greater flux of photons when the high voltage circuit is energized. The
effect of the mA circuit is quite linear.
The third control of the X-ray tube which is used for medical imaging is the exposure timer.
This is usually denoted as an "S" (exposure time in seconds) and is combined with the mA
control. The combined function is usually referred to as mAS or milliampere seconds, so to give
an exposure using 10 milliampere seconds may use a 10 mA current with a 1.0 second exposure
or a 20 mA current for a 0.5 second exposure or any combination of the two which would result
in the number 10. Both of these factors and their combination affect the film in a linear way.
Basically, there are two parts of the circuit. One of them is for producing high voltage, which is
applied to the tube’s anode and cathode and comprises a high voltage step-up transformer
followed by rectification. The current through the tube follows the HT pathway and is measured
by a mA meter. A KV selector switch facilitates change in voltage between exposures. The
The second part of the circuit concerns the control of heating X-ray tube filament. The filament
is heated with 6–12V of AC supply at a current of 3–5 amperes. The filament temperature
determines the tube current or mA and therefore, the filament temperature control has an
attached mA selector. The filament current is controlled by using in the primary side of the
filament transformer, a variable choke or a rheostat. The rheostat provides a stepwise control of
mA and is most commonly used in modern machines.
A preferred method of providing high voltage DC to the anode of the X-ray tube is by use a
bridge rectifier using four valve tubes or solid state rectifiers. This results in a much more
efficient system than the half wave of self-rectification methods.
Housing:
X-ray tube itself is enclosed in metal (aluminum) sealed housing. Circulating air is sufficient to
cool (remove heat) mobile and mammography units, but insulating oil (fixed / circulate) is
necessary within the enclosure for cooling conventional and DSA equipment. The total heat
capacity of the tube enclosure is largely dependent on the volume of oil. The metal housing is
also properly grounded to prevent shock.
Radiation Shielding:
X-rays are emitted from the tube in all directions and overall lead shielding must cover the
complete housing assembly except window region. The lead shielding prevents any leakage
radiation [˂1mGyh-1 (100mR) at one meter]. A fixed lead diaphragm in the window region
restricts beam dimensions yielding the useful beam width. Other adjustable diaphragm /
collimators are used for varying the overall beam size for different field of views.
Copper being an excellent thermal conductor and carrying the heat rapidly away from the
tungsten target. Finally, the heat is removed from the total anode surface to the outside of the
tube by convection. Generally, an oil environment is provided for convection current cooling. In
addition, the electrodes have open high voltages on them and must be shielded. The tube will
emits X-rays in all directions and a lead lining over the glass envelop is provided except (where
the useful beam emerges from the tube) window area. A metal housing surrounding the tube and
is called “shield”. Every joint on a shield are hermetically sealed and must be made shockproof
by an efficient grounding (earthing) arrangement.
Application: Stationary x-ray tubes are only suitable in dentistry and small mobile X-ray units.
The electron source (cathode) is placed so that it will impinge (hit) on only a small spot on the
rim of the disc. This keeps the focal area down to a few square millimeters, yet the structure is
large enough to permit the dissipation of a large amount of heat. Rotating anode increases the
heat-loading capacity of the tube and allows higher power level for the generation of more
intense X-radiation. In CT or DSA, graphite is brazed onto the back of the anode which increases
heat radiating efficiency in tubes. Graphite is a very low mass material with high melting point.
STATOR-ROTOR ASSEMLY
The electric induction motor that turns the rotating anode consists of a rotor which rotates on a
set of bearings within the glass X-ray tube envelope and the stator consists of wire windings
external to the envelope. An alternating current is applied to the stator windings, inducing a
changing magnetic field within the region of the rotor. The metal rotor turns in response to the
changing magnetic field causing the attached anode to turn. Anode rotation speeds range from
3,000 rpm (revolutions per minute) to 10,000 rpm for a high-speed anode. The rotor is connected
to the anode disk by the anode stem which is typically made of molybdenum or stainless steel.
The stem is designed to protect the bearings from heat damage, a common cause of x-ray tube
failure.
Induction Motor
The Rotor is made of copper which is actually part of a motor that is made to rotate by the device
called the stator. The stator is a set of electric coils that produces a very strong magnetic field
outside of the glass envelope that in turn “pulls along” the copper rotor. The stator causes the
rotor to rotate very fast just like any electric motor. The stator is also called an induction motor.
The bearings which permit a very smooth rotation of the anode are very sensitive and if the X-
ray tube heats up abnormally high, they can be damaged. If this occurs, the target surface can be
damaged or “pitted”
X-ray tube rating is the total workload that can be placed on a tube. The combining effects of
kilovolts (kVp), tube current (mA) and exposure time (S) for a certain focal spot size is called
electrical rating, whereas anode heat gain and heat loss determines the thermal rating,
commonly referred to as loadability. The workload or rating of X-ray tube depends on factors:
(a) Variable: tube kVp, tube current, filament current, focal spot size and exposure time, (b)
Non-variable: anode diameter and anode rotational speed. By creating tube ratings the operator
can ensure that the parameters set are appropriate for the examination whilst minimizing the risk
of damage to the X-ray tube. Typical X-ray tube ratings are between 5-100kW and are dependent
on focal spot size.
Overheating a Cold Anode: Warming up an X-ray tube is an important step in prolonging the
useful life of the tube, if heat is generated in a cold target too quickly the target will not expand
uniformly and in many cases it will crack.
Exceeding Tube Heat Capacity: The importance of X-ray tube heat rating charts has been
taught by radiologic technology programs for many years but in actual practice a few
technologists ever refer to them. To overcome the problem of exceeding instantaneous tube
ratings, most modem generators are designed with overload relays that will come into play if the
combination of milliampere, time and kilovoltage exceeds the kilowatts allowed for an
instantaneous exposure on that particular tube.
Misuse of High Speed Rotation: The question of high speed rotation of the X-ray tube anode is
a perplexing one. In normal tube operation, the anode rotates approximately 3,000 times per
minute (rpm); at high speed, 9000-10,000 rpm. High speed rotation is three times faster than
standard speed (and, thus provides for more rapid dissipation of heat) whereas, the wear factor
on the X-ray tube bearings is nine times that experienced at standard speed. This leads to a great
decrease in tube life.
The most common causes of X-ray tube failure are:
Extremely high voltage
Excessive heat generation
Poor cooling system
Rough handling / careless
Aging
Essentially, three basic controls exist on X-ray machines to control patient X-ray dose
(penetrating quality, quantity and timing). These are inter-related and must be properly chosen to
suit the slim or obese patient. Good photographic results are sometimes difficult to obtain. These
controls are filament heat control (mA) for exposure strength, not depth; kilovolt control (kV) for
penetration depth and contrast; and timing devices for time exposure length.
The multitap ac line autotransformer (3) is a single-coil transformer that serves three functions:
it provides the means for kVp selection, it provides compensation for fluctuations in
the incoming line voltage (9), and it supplies power to other parts of the x-ray circuit. The
autotransformer’s primary purpose is to vary the voltage to the primary side of the step-up
Prepared by: Dr. S. Roy, Prof.-Dept. of Biomedical Engineering, NSEC Page 1
transformer. This is accomplished by the major and minor kVp selectors (4), which are on the
secondary (output) side of the autotransformer. The autotransformer varies the kVp to the tube
by controlling the input to the step-up transformer.
The low voltage circuit is the sub-circuit between the alternating current (AC) power supply (1)
and the primary (input) side of the high-voltage (step-up) transformer (7). With the exception of
the step-up transformer, all of the devices in this sub-circuit are actually located within
the control console. The control console is the unit where the operator sets all of the exposure
techniques, such as kilovolts peak (kVp), milliamperes (mA), and exposure time. They include
the main switch (2), autotransformer (3), kVp selectors (4), exposure switch (5), and exposure
timer (6).
Filament Circuit is divided into two parts by the step-down transformer (11 and 12). The
primary purpose of the filament circuit is to supply a low current to heat the x-ray tube filament
for thermionic emission of electrons. The filament circuit is activated any time the operator
adjusts the mA on the generator. The primary side of this circuit begins and ends with the
contacts on the autotransformer (9). Current in this circuit flows from the autotransformer,
through the mA selector (10) and the primary side of the step-down transformer (11), and back to
the autotransformer. The secondary side begins and ends with the secondary side of the step-
down transformer (12) conducting current through the x-ray tube filament (13). The step-down
transformer reduces the voltage on the secondary side, providing an appropriate current to heat
the filament. The mA selector (10) controls amperage in the filament circuit. Since the current
through this circuit controls filament heat, this setting determines the number of available
electrons at the x-ray tube filament and thus determines the mA in the high-voltage circuit that
includes the x-ray tube.
High-Voltage Circuit begins and ends with the secondary side of the step-up transformer (8). It
includes the x-ray tube (14) and the rectifier unit (15). Current flows in this circuit only during
an exposure. This is a dangerous circuit due to the very high voltage. The high-voltage cables
going to the x-ray tube are very thick due to their high insulation requirement. The step-up
transformer is also referred to as the high-voltage or high-tension transformer. The primary
purpose of the high-voltage circuit is to supply the x-ray tube with voltage high enough to create
x-rays.
The exposure switch (5) closes the circuit, allowing electric current to flow through the primary
side of the step-up transformer. When this occurs, current is induced to flow through the
secondary side of the transformer, creating voltage across the x-ray tube. This voltage causes the
electron stream to flow across the tube, producing x-rays. The exposure timer (6) is a device
that terminates the exposure and is set by the operator on the control console.
The core of the transformer is very important to increase the efficiency or to minimize the heat
loss. Generally silicone steel or sometimes cold roll steel is used to improve efficiency. A full
wave rectifier bridge to output terminals of secondary coil is connected for converting AC to DC
voltage output.
Rectifiers use diodes to convert the circuit from AC to DC. A diode is an electronic device that
permits current to flow in one direction only. Each diode can withstand 1 kV. Both primary and
secondary coils and bridge rectifier are immersed in high grade insulating oil for isolation of the
part from other parts of the instruments.
In the Medium/ HF generators, higher frequency square wave voltage is sent to primary of the
high-frequency transformer. Higher frequency of pulses to the high tension supply of an X-ray
generator results in the spectrum of the X-ray beam being narrowed, potentially lowering of
patient dose. The exposure timing can be more precise with and short with near constant
potential wave form. Further, when an X-ray transformer operates in the high range of 2000 to
over 8000Hz, instead of the conventional line frequency, it becomes more efficient. HF
generators significantly decrease in transformer size for the same power output. It can be reduced
in weight to perhaps one-tenth and even more in bulk. As a result of the diminished size,
transformers up to 30kW can be mounted within the tube head obviating high-tension cables.
The low frequency (50/60Hz) main supply is first rectified to give a DC voltage which then
supplies a high frequency generator (2 to 20 KHz). The output of high frequency generator is
rectified and supplied the X-ray tube circuit. The generator design can utilize either single or 3-
phase (3-sine waves with a 1200 phase difference) main supply. These are full wave rectified and
smoothed to give a steady DC voltage supplies a thyristor converter which switches the DC
voltage producing a medium (2 KHz) to high (20 KHz) square waveform. A high frequency
transformer then converts this low voltage high frequency AC to KV levels. Rectification and
smoothing gives high voltage DC for the x-ray tube. The thyristor converter can be controlled by
low voltage signals to regulate its high voltage output.
The tube KV can be electronically switch on and off at any point in time. Tube voltage is present
according to reference control value Vref. For power rating up to 50 KW, generator is small
enough to be built into a single housing with X-ray tube but for 150 KW, generator and tube
must be separated.
CONTROL CIRCUITS
Controlling high frequency generator output is achieved by altering frequency by a feedback
control signal. In this way, KV can be finely adjusted within a close tolerance (˂1%).
Exposure Control:
Control of X-ray exposure was obtained by using mechanical timers either clockwork or
electrical. They are unable to offer fast switching rates necessary for effective X-ray exposure
(0.01 to 0.05s) and have been replaced by either electronic, digital timers, frequency control or
exposure control timers (microcontroller based).
A. Electronic Timer:
Various types of electronic timers are developed as exposure time is too short to be
controlled by mechanical timers.
RC Timing Circuit
The exposure time is decided by the time constant of the RC circuit. Switch “S” is closed to
initiate exposure and “S” is opened to stop exposure and the relay in primary of the high
voltage transformer circuit is closed beginning exposure capacitor charging.
𝑡
𝑉𝑡 = 𝑉𝑠 (1 − 𝑒 ⁄𝑅𝐶 ) Where C is constant and R is variable
B. Digital Timer: Frequency or Pulse Counting Timer
Modern x-ray machine use digital timers. These are either use dedicated timer ICs or circuit
using a reference oscillator, a counter and associated logic. The generation of time period
‘T’ is based on counting ‘N’ cycles of a precise frequency ‘F’ by a relation- F x T = N
Actually, there are two sandwich design detectors separated by a metal filter. The output signals
from D1 and D2 are electronically balanced; detector D2 behind its beam hardening filter acts as
a reference. Any magnitude of differential signal controls the film exposure (mAS value).
Information concerning screen and film sensitivity, kV and tube current influence this difference
signal to give an optimum exposure value for all variables.
Ionization Chamber:
Ionization chamber is placed between the patient and the cassette. The signal from the chamber
is amplified and used to control a high speed relay which terminates the exposure when a preset
density is reached.
The X-ray beam is filtered by absorbers at three different levels: Inherent filtration, Added
filtration & Patient. Inherent filtration: Glass envelope, the insulating oil surrounding the tube
aand window in the tube housing. Added filters: either Aluminiu, copper or compound (Al +
Cu). Copper will absorb high energy photons (also cut down the thickness), aluminum will
absorp characteristics radiation (10 KeV)
Collimators:
Collimators adjust the X-ray beam width and focusing the X-ray beam (also called X-ray beam
restrictors). It is placed between the X-ray tube and the patient, normally attached at the window
of X-ray tube. Collimator consists of tube having an achromatic lens is fitted at one end. The
other end is fitted with a second brass-tube which can slide inside the 1st tube. The outer end of
the sliding tube has vertical slit (circular / rectangular) whose width is adjustable with the help of
four lead strips which can be moved relative to each other. In practice, it is advisable to use the
smallest possible field size. This results in a low dose to the patient and simultaneously increases
the image contrast, because less scattered radiation reaches the image plane. The scattered
radiation produces diffuse illumination and fogging of the image without increasing its
Grids:
One of the problems in getting a sharply defined image in clinical radiology is the presence of
scattered or secondary radiation. These photons are created in the body of the patient or closely
surrounding objects by the interaction of that material. In the typical clinical imaging, the most
common method of reducing scatter is to use a radiographic grid. The separation of scatter from
primary radiation was first addressed by Bucky (in 1913) who introduced a grating or grid of thin
lead strips. Grids are placed between the patient and the film cassette in order to reduce the loss
of contrast due to scattered radiation. It is actually composed of radiopaque (lead) and
radiolucent (low attenuating material such as carbon fiber, aluminium or organic spacer) strips.
These are arranged on edge and the edge of these strips is turned towards the source of X-ray.
This arrangement of the radiographic grid will give the highest possibility for primary X-ray
photons passing between the lead strips and reaching the film, while the off-focus or secondary
photons are likely to interact in the lead strips and never reach the film. The use of this
radiographic grid will greatly improve image sharpness when a relatively thick body part is being
imaged.
Mammographic X-ray equipment can either be used with special film/screen cassette or as xero-
radiographic units. The units intended for film/screen use have a molybdenum target X-ray tubes
with a beryllium window and a 0.03 mm molybdenum filter. Radiographs are usually taken at
25–35 kV and tube current 5-10 mA. Xero-radiographic systems use X-ray tubes with tungsten
targets and about 1 mm aluminum filter. Radiographs with this technique are taken at 40–50 kV.
Hence, both types of mammographic units operate at low peak voltages.
The primary winding of the transformer is fed with mains voltage via an exposure timer and the
high voltage developed in the secondary windings is fed to the self-rectifying X-ray tube. The
complete assembly is contained in a metal case filled with special insulating oil. The X-ray tube
is of special design and employs a third electrode, called a ‘grid’, between the anode and the
cathode electrodes. The grid restricts electrons from leaving the cathode until the high voltage
reaches its peak value, whereupon all electrons are released and impinge on the anode at a very
high velocity. Consequently, the x-radiation generated contains fewer useless soft X-rays and
more hard rays. The total radiation is, therefore, more effective and can be compared
mathematically to a much higher output resulting in shorter exposure times.
Because of the high current requirements of the mobile units, the mains resistance becomes a
problem, especially if the mobile is to be used in different parts of the hospital. In order to ensure
consistent results from one power supply socket to another, a means for mains resistance
calibration is provided and must be adjusted to suit each location before making an exposure.
Where there are limitations on the electrical supply, mobile units make use of stored energy. This
may be from the capacitor discharge or battery-powered invertor circuits. The former releases
stored energy from the capacitor during exposure, while the latter converts energy stored in the
battery.
Silver halides: silver chloride, bromide and iodide are used in various mixtures to alter the
sensitivity of the film. The order of light sensitivity is AgBr > AgCl > AgI. They have a cuboid
crystal structure and are formed by reacting silver nitrate with alkaline halides.
AgNO3 + (KBr or NaCl or KI) AgBr or AgCl or AgI
Silver Nitrate and Potassium Bromide are combined Silver Bromide precipitates out. Potassium
Nitrate is washed away as a waste product. Silver halides are flat and roughly triangular in shape.
They create the density on the film. Normally, 95 to 98% are silver bromide (AgBr) with the
remaining being silver iodide (Agl). Each crystal is a cubic lattice (or matrix) of silver, bromide and
iodide atoms. Negative bromine and iodine halides ions remain on the surface of positive silver
ions inside.
Sensitivity Speck:
Silver halide crystal (SHC's) have an impurities added (usually silver sulfide, AgS) to form
sensitivity specks. The sensitivity speck is what makes the SHC photosensitive. During latent
image formation sensitivity specks attract the silver ions inside the SHC.
Conventional screen film has a thin emulsion layer designed for automatic processing. The thinner
emulsion gives – (i) shorter development and fixing times, (ii) more efficient washing (removing
practically all of the resident chemicals) and (iii) rapid drying. Non-screen film has thicker
emulsion which is more sensitive to X-ray radiation and useful for bone imaging.
Types of films:
A. Screen type films: faster when used with intensifying screen
- Conventional / Monochromatic (without optical sensitizer) sensitive to UV and
blue light.
- Orthochromatic film (with optical sensitizer) sensitive to UV, blue and green
light.
B. Direct exposure type: used for dental exposures
It is important to note that the colour of the light emitted (wavelength) must match the light
sensitivity of the film used. This is known as spectral matching.
Film Storage:
Ideal storage conditions for unexposed film are 200C (max.) with a relative humidity of 50%. Cold
storage can significantly lengthen the useful life of a film. X-ray film is always kept in a cassette
(made up of metal cover).
Cassettes are rigid holders used in conventional and computed radiography (CR) for the screen
film system and imaging plate respectively. The back side of the cassette has a rubber or felt for
adequate contact between screen film systems or with the imaging plate. The front is made of low
atomic number material (e.g. plastic or carbon) and the back is made of high atomic number
material (e.g. lead) to reduce backscatter.
In case of conventional radiography, two screens are mounted on each side of the cassette, except
in mammography, where a single screen is mounted on the back side. These cassettes have to load
with film in the darkroom unlike the cassettes used in CR which can be loaded with imaging plate
in the light. The formats of the cassettes used in conventional radiology are: 13 x 18 cm; 18 x 24
cm; 24 x 30 cm; 20 x 40 cm; 30 x 40 cm; 35 x 35 cm; 35 x 43 cm; 30 x 90 cm.
Intensifying screens are used in the X-ray cassette to intensify the effect of the X-ray photon by
producing a larger number of light photons. It decreases the mAs required to produce a particular
density and hence decreases the patient dose significantly. It also reduces motion blur and X-ray
tube loading by reducing exposure time. In cassettes, which use double emulsion films, two screens
are used, mounted on both sides of the cassette. In mammography, however, a single screen on the
back side and a single emulsion film is used.
The thickness of an intensifying screen is about 0.4 mm. The thickness of the screen affects the
screen speed and spatial resolution: thicker screen improves speed but reduces spatial resolution
(increased diffusion of light before image formation). The intensity factor is a measure of efficiency
of intensifying screen and is the ratio photon exposures with and without intensifying screen to
achieve a designated film density. The absorption and conversion efficiency of the screen affect the
intensity factor.
Generally, doctors keep the film image as a negative. That is, the areas that are exposed to more
light appear darker and the areas that are exposed to less light appear lighter. Hard material, such as
bone, appears white, and softer material appears black or gray. Doctors can bring different
materials into focus by varying the intensity of the X-ray beam.
FILM SENSITOMETRY
Film is originally constructed for recording the changes in photon intensity. The film is more
sensitive to light than X-ray energy. At least 2.4 eV energy is required to free electron from
bromide and is corresponds to 520 nm i.e. blue light or shorter. However, it does have a useful
linear response to X-ray and γ-ray radiation which is utilized for measuring radiation dose as a film
dosimeter. Film sensitometry is the quantitative evaluation of how a film emulsion is responds to
radiation and processing. This study is important in order to produce images with optimum contrast
that reveal high details of the object examined. Film sensitometry actually measures the response of
photosensitive film to different levels of exposures. The response of the film to exposures is
manifested as a degree of blackening produced after chemical processing.
In radiography, the degree of blackening is quantitatively indicated by the term optical density. The
optical density describes how much a certain area of the film is opaque to light incident upon it. A
film contains number of areas with different optical densities from white (fully transparent) to black
(fully opaque). A film can be considered as a contrast converter. A contrast is the variation in film
density (shades of gray) that actually forms the image. One of the functions of film is to convert
differences in exposure (subject contrast) into film contrast (differences in density). The amount of
film contrast resulting from a specific exposure difference can vary considerably. The exposure
range in which a film can produce useful contrast is known as its latitude. The latitude of a specific
film is determined primarily by the composition of the emulsion and, to a lesser extent, by
processing conditions. The significance of film latitude is that it represents the limitations of the
exposure range that will yield useful image contrast.
Film characteristics are best obtained by illuminating the film with a calibrated gray scale from a
densitometer. The gray scale densities (resultant optical densities) are plotted against exposure (log
relative exposure) then a film characteristic curve is obtained. The curve describes the film
quantitative response to photon and reveals many important film qualities.
The film characteristic curve has three distinct regions with different contrast transfer
characteristics. The part of the curve associated with relatively low exposures is designated the toe
and also corresponds to the light or low-density portions of an image. When an image is exposed so
that areas fall within the toe region, little or no contrast is transferred to the image
A film also has a reduced ability to transfer contrast in areas that receive relatively high exposures.
This condition corresponds to the upper portion of the characteristic curve in which the slope
decreases with increasing exposure. This portion of the curve is traditionally referred to as the
shoulder.
The highest level of contrast is produced within a range of exposures falling between the toe and
the shoulder. This portion of the curve is characterized by a relatively straight and very steep slope
(γ) in comparison to the toe and shoulder regions. In most imaging applications, it is desirable to
expose the film within this range so as to obtain maximum contrast.
The fog level is a density reading of the background noise caused mainly by thermal effects. The
threshold value (Dmin) marks the lowest exposure that fives a density value above the fog level. The
shoulder (Dmax) region (non-linear response) indicates that all the film grains have been saturated.
The precise shape of the curve depends on the characteristics of the emulsion and the processing
conditions. The primary use of a characteristic curve is to describe the contrast characteristics of the
film throughout a wide exposure range. At any exposure value, the contrast characteristic of the
film is represented by the slope of the curve. At any particular point, the slope represents the
density difference (contrast) produced by a specific exposure difference.
The number of grain per unit area and size can vary with type of emulsion and developer, but the
amount of photon energy required per grain is constant. Thus fine-grain films require more photon
energy per unit area than the coarse-grain films. The energy required to free the electron from the
bromide is at least 2.4 eV (1 eV = 1.602 x 10-19 J).
The image formed prior to chemical development is called “latent image”. It is still possible for
some of the metallic silver atoms to return to their lattice coupling because of slow thermal effects
and is called “latent image fading”. All films have some range of exposure times such that a
constant photon intensity time product will bring the film to a particular value of latent image
𝟏
density (D). 𝐃 = 𝐥𝐨𝐠 𝐓
Where, T = Intensity of the transmitted X-ray beam that expose the film
D = Density of latent image
The areas of the emulsion most exposed to the photons end up with greatest concentration of
metallic sliver. These are the darkest areas when project light through them. Areas with little light
end up with less metallic silver and are more transparent. This is negative film. On the negative
film all the brightness values of the original scene are reversed. Light becomes dark and dark
becomes light.
Upon the completion of fixation, the film should be immersed in fresh, cool, circulating water for
at least 20 minutes to ensure complete removal of the fixing solution. If not washed properly, the
radiograph will turn yellow and fade with time if any of the fixing chemical remains on its surface.
Developers:
Developers usually contain ring compounds of hydroquinone, amitol or metol.
Silver halide crystals reduced completely to silver atoms representing amplified latent image
information.
Fixers:
The fixer compound is ammonium thiosulfate (NH4)2S2O3 which removes the unexposed, unaltered
silver halides as a soluble complex.
Ag+Br- + (S2O3)2- Ag (S2O3)- + Br-
The silver thiosulfate complex is carried out of the emulsion when film is thoroughly washed with
clean water.
A. Contrast: Contrast quantifies the differences between image characteristics (eg. Shades of
gray or colour) of an object or feature within an object and surrounding objects or
background. It describes the amount of additional darkening caused by an additional
amount of light when working near the center of a film exposure range. The ability of
medical professional to discriminate among anatomical or functional features in a given
image strongly contrast. It is influenced by system performance but also by the physical
attributes of the different tissues, KV, detectors, etc.
B.
C. Resolution: It refers to the degree to which the measuand can be broken into identifiable
adjacent parts. Resolution can be thought of as the ability of a medical imaging system to
accurately depict two distinct events in space, time or frequency as separate i.e., spatial,
temporal or spectral resolution. Resolution is influenced by the machine itself (X-ray tube),
movement (patient) and recording medium (film, image plate, etc.)
Image quality in the brighter areas of the images is limited by blurring or poor spatial
resolution. Spatial resolution is the ability to detect a small structure against its background
or to distinguish as separate two structures close together. Any objects or points is
distinguishable by two times, therefore, spatial resolution is measured in line pairs per mm
(LPmm-1).
The ratio of image brightness of the two phosphors is called brightness gain (range 2500 to 7000 in
practice) of the intensifier tube. The brightness gain is the product of geometric gain and electronic
gain. A lens mounted on the image intensifier has the purpose of collimating or focusing the image.
Geometric gain: Ratio of the areas of input and output phosphor.
Electronic gain: Product of input quantum efficiency, photocathode efficiency, potential
difference between input and output phosphors and phosphor output efficiency.
Digital X-rays
Digital radiography is currently practiced through the use of three commercial approaches, two of
which also depend on phosphor screens. The first phosphor based approach is to digitize the signal
from a video camera that is optically coupled to an X-ray image intensifier to provide an instant
readout The second phosphor based approach is a stimulable phosphor system, in which the
Digital X-ray imaging system consists of: (1) X-ray machine, (2) X-ray imaging transducer or data
collector & (3) computer for processing, storage and data display. Two major variant of detectors
used for digital image capture are – (A) Flat Panel Detectors (FPDs) & (B) High density line scan
solid state detectors.
(a) Absorbed Dose (DT): It is the ratio of energy absorbed by a unit mass measured in J/kg.
This is fundamental quantity in radiation protection and is called gray (Gy) normally
describe the average dose to an organ or tissue Radiation dose is the energy (joules)
imparted per unit mass of tissue and has the US units of rad (radiation absorbed dose).
Patient exposures, particularly in Radiation Oncology are described in units of radiation
dose. The international (SI) unit for dose termed the Gray (Gy). The conversion between
the units is: 100 rad = 1 Gy. One rad is the radiation dose that will result in energy
absorption of 1.0x10-2J/kg of irradiated material.
The relation between absorbed dose (D) and radiation exposure (R) is as follows:
DT = f x R
Where, f is a proportionality constant which depends on the composition of the irradiated
material and quality of the radiation beam. Generally, f is 0.87 rad/R for air, 1 rad/R for soft
tissue and f is >1 rad/R for hard tissue. The value of f is significantly decreases with
increases in KV.
(b) Equivalent Dose (HT): It is a measure of the radiation dose to tissue where an attempt has
been made to allow for relative biological effects of different types of ionizing radiation.
This is related to biological damage and is denoted as rem (radiation equivalent man, US
unit). The probability of tissue radiation damage depends not only on the absorbed dose but
also on type and energy of radiation. The unit rem is used to compare dose received by
different types of radiations (e.g. alpha particles) which have a different capacity for causing
harm than X-ray radiation. This unit is properly termed dose equivalent. The dose
equivalent is the product of the dose times a quality factor. In quantitative terms, equivalent
dose is less fundamental than absorbed dose, but it is more biologically significant.
Equivalent dose is measured using Sievert (SI unit) but rem is still commonly used (1 Sv =
100 rem). The relation between equivalent dose and absorbed dose is as follows:
Sv = Q x DT (rad)
Q is the quality factor and in diagnostic X-ray, Q=1 and therefore, 1 rad = 1 rem.
Q has been applied to the absorbed dose at a point and not over the organ as a whole. So,
this dose equivalent (HT) can be rewrite as HT = ∑DTWR.
HT is the equivalent dose in tissue for an absorbed dose DT from a radiation having a
weighting factor WR. The equivalent dose is measured in Sievert (Sv) and forms the
fundamental unit for other parameters in radiation protection. [The radiation weighting
factor, WR is dependent on the type and energy of the incident radiation. The value of WR is
1 for X-rays].
Effective Dose Equivalent (EDE): The effective dose is used to compare the stochastic
risk of non-uniform exposure to radiation. Body tissues react differently to radiation and
cancer-induction occurs at different rate of dose in different tissues. Hence, the effective
dose is the risk of developing fatal cancer in the tissue in question. If the body is uniformly
irradiated, the summed effective doses are equal to 1.
The effective dose is calculated by multiplying the equivalent dose (HT) by a tissue
weighting factor (WT).
[Radiation absorbed dose and effective dose in the international system of units (SI system) for
radiation measurement uses "gray" (Gy) and "Sievert" (Sv), respectively.
In the United States, radiation absorbed dose, effective dose and exposure are sometimes
measured and stated in units called rad, rem, or roentgen (R).]
Fundamental to radiation protection is the avoidance or reduction of dose using the simple
protective measures of time, distance and shielding. The duration of exposure should be limited to
that necessary, the distance from the source of radiation should be maximized and the source
shielded wherever possible. To measure personal dose uptake in occupational or emergency
exposure for external radiation personal dosimeters are used and for internal dose to due to
ingestion of radioactive contamination bioassay techniques are applied.
According to the International Atomic Energy Agency (IAEA), radiation protection can be divided
into three groups:
occupational radiation protection, which is the protection of workers in situations where
their exposure is directly related to or required by their work
medical radiation protection, which is the protection of patients exposed to radiation as part
of their diagnosis or treatment
public radiation protection, which is the protection of individual members of the public and
of the population in general
RADIOTHERAPY
Radiation therapy has a pivotal role in the treatment of cancer. Radiation therapy uses high-energy
radiation to shrink tumors and kill cancer cells. X-rays, gamma rays and charged particles are types
of radiation used for cancer treatment. The types of radiation therapy are:
A. External Beam Radiation Therapy (EBRT) or Teletherapy: This technique involve the
delivery of electromagnetic radiation (X-rays, γ-rays) or particulate radiation (electrons,
protons) from a linear accelerator or radionuclide source (Co-60) from outside of the body.
It accounts for almost 90% of radiation treatment.
B. Internal Radiation Therapy (IRT) or Brachytherapy or Implant Radiation Therapy:
Radioactive material sealed in needles, seeds, wires or catheters is placed directly into or
near a tumor.
C. Systemic Radiation Therapy: Radioactive substances (radioactive iodine-131, strontium-89,
samarium-153) are injected in the blood stream ad are travel in the blood to kill cancer cells.
Radiation therapy may be curative in a number of types of cancer if they are localized to one area
of the body. It may also be used as part of adjuvant therapy to prevent tumor recurrence after
surgery to remove a primary malignant tumor (eg. early stages of breast cancer). Radiation therapy
is synergistic with chemotherapy and has been used before, during and after chemotherapy in
susceptible cancers.
RADIOTHERAPY PRINCIPLES
Radiation therapy is commonly applied to the cancerous tumor because of its ability to control cell
growth. The underlying principle of radiation therapy is the destruction of malignant tissues while
minimizing damage to normal tissues within a treatment field. Radiation therapy kills cancer cells
by damaging their DNA (the molecules inside cells that carry genetic information and pass it from
one generation to the next). Radiation therapy can either damage DNA directly or create charged
particles (free radicals*) within the cells that can in turn damage the DNA. Cancer cells whose
DNA is damaged beyond repair stop dividing or die. When the damaged cells die, they are broken
down and eliminated by the body’s natural processes.
*A type of unstable molecule that is made during normal cell metabolism (chemical changes that
take place in a cell). Free radicals can build up in cells and cause damage to other molecules, such
as DNA, lipids, and proteins. This damage may increase the risk of cancer and other diseases.
RADIOTHERAPY DOSE
The amount of radiation used in photon radiation therapy is measured in gray (Gy) and varies
depending on the type and stage of cancer being treated. For curative cases, the typical dose for a
solid epithelial tumor ranges from 60 to 80 Gy, while lymphomas are treated with 20 to 40 Gy.
Preventative (adjuvant) doses are typically around 45-60 Gy in 1.8-2 Gy fractions (for breast, head
and neck cancers.)
Radiotherapy is an effective treatment method for cancers. During radiation treatment, a patient
must be in the same position from the start to the end of radiation treatment. Patient movements are
usually monitored by the radiation technologists through the closed circuit television (CCTV)
during treatment. The success of radiotherapy is to deliver maximum dose of radiation to the tumor
tissue. At the same time, the minimum dose of radiation has to be ensured to the surrounding
normal tissues to reduce the incidence of radiation of acute and late effects. To achieve this, proper
treatment planning and careful execution is mandatory. Patient immobilization is an important
parameter during the radiotherapy treatment.
Cobalt-60 Therapy:
Most common RT can be used anywhere on the body and particularly useful in brain tumor. Under
local anesthesia, the patient is positioned on a table with a special rigid frame covering the head.
Based on the results of an image study conducted just prior to treatment, Co-60 therapy unit directs
approximately 200 beams of 200 beams of gamma radiation at the patient’s tumour. Treatment
takes anywhere from several minutes to a few hours to complete. Following treatment, the head
frame is removed and the patient may return to normal activity.
As used in radiotherapy, cobalt units produce stable, dichromatic beams of 1.17 and 1.33 MeV,
resulting in average beam energy of 1.25 MeV. The cobalt-60 isotope has a half-life of 5.3 years so
the cobalt-60 needs to be replaced occasionally.
The X-ray image is resolution limited by the dimension of X-ray source and noise limited by beam
intensity. The X-rays emerging from the anatomy are detected to for a one-dimensional image
A digital image is a numerical representation of an image via a set of picture elements known as
pixels. This simplified article lists three parameters of a digital image that moderate spatial and
contrast resolution.
Image matrix
The image matrix is comprised of columns (M) and rows (N) that define the elements or pixels
within an image. The size of an image is:
Matrix = M x N x k bits
The field of view (FOV) is the size of the displayed image. For same FOV and increase the matrix
size, the pixels will be smaller and hence spatial resolution is improved.
Pixel
The pixel is the discrete picture element (2D information) that makes up the image matrix; each
pixel is a respective value that will represent a brightness level. The size is determined via: pixel
size = FOV/matrix. A decreased FOV means that the pixel is smaller and results in an improvement
in spatial resolution.
Voxel
The voxel is a pixel that represents information that is contained in a volume (3D information).
Bit depth (k bit)
The k bit is the number of bits per pixel, the grey scale of an image is equal to 2 k-bit, for example:
k bit of 2 = 4 shades of grey
k bit of 8 = 256 shades of grey
The higher the bit depth means more grey scale and therefore the higher the contrast resolution.
Introduction
Digital subtraction angiography (DSA) is a type of fluoroscopic technique used extensively in
interventional radiology to clearly visualize blood vessels in a bony or dense soft tissue
environment. Radiopaque structures such as bones are eliminated ("subtracted") digitally from
the image, thus allowing for accurate depiction of the blood vessels. “Angio” means blood vessel
and angiography is the radiological study of blood vessel in the body after the introduction of
(iodinated) contrast media. Subtraction is a simple technique by which bone structures images
are subtracted or cancelled out from a film of bones plus opacified vessels, leaving an
unobscured image of the blood vessels. The acquisition of digital fluoroscopic images combined
with injection of contrast material and real-time subtraction of pre- and post-contrast images to
perform angiography is referred to as digital subtraction angiography.
The second operation occurs after subtraction and is expansion of the dynamic range of the
subtracted image which results in enhancement of the final image. This is necessary because the
range of contrast within the initial subtracted image is very small. Subtraction and enhancement
are performed in real time, which means that the processing of data is sufficiently rapid that the
results are available in time to influence the clinical examination. The speed and apparent
simplicity of computerized subtraction are two major advantages of DSA over standard film
subtraction angiography.
Several further steps are a routine part of the process. Amplification of the output of the image
intensifier is one. This amplification may occur before or after digitization of the data and may
be fixed or selectable, depending on the individual system. Choices for amplification include
linear and logarithmic modes. In linear amplification, the unsubtracted signal is amplified
linearly, independent of its numerical value. This is appropriate if there is uniform tissue density
/
DSA Principles
DSA Equipment
The fluoroscopy unit consists of a C-arm unit that can be rotated axially and sagittally around the
floating-top table. The distance between the X-ray tube and the image intensifier can be adjusted,
as can collimation and several other parameters. In dedicated angiography units, there is a second
set of controls for the angiographer (radiographer). A modern angiography unit has all of the
following features:
Collimators (including oblique) and filters for dose reduction
Pulsed fluoroscopy with a variety of frame rates for dose reduction
Ability to change and display collimator position without fluoroscopy
Road mapping and land marking
Last image hold and frame-grab
Display of images side-by-side
It should be noted that since image subtraction causes a decrease in signal-to-noise ratio, the
subtraction images appear noisier than the source images. The inevitable solution to this is to
increase mA. There are also algorithms in place for reducing scatter.
Procedural Technique
For every purpose, there is at least one technique, but common to them all is the application of
DSA for visualization:
Patient lies on the angiography table
Local anesthesia is administered at the intended puncture site (usually lidocaine
hydrochloride 1% or 2% w/v)
In certain procedures (e.g. a child undergoing cerebral angiography) general anesthesia is
performed
The Seldinger technique is used to gain access to a blood vessel
ultrasound is often used for visualizing the vessel in real-time for puncturing
a standard access kit includes a straight 18 gauge needle and .035" guidewires, on
which the diagnostic and therapeutic catheters are threaded
in many cases, a micro-introducer access kit (.018" guidewire threaded through a
21 gauge initial access needle) is used for access, either for the entire procedure or
to be replaced with the standard kit. Using a micro-introducer facilitates less
traumatic entry and can be retrieved without massive bleeding should there be a
need for re-puncturing
On procedure completion, hemostasis is applied to the puncture site
Introduction
Fluoroscopy is a type of medical imaging to obtain a continuous X-ray image on a monitor,
much like an X-ray movie. During a fluoroscopy procedure, an X-ray beam is passed through
the body. The transmitted beam then strikes a fluorescent plate that is coupled to an image
intensifier. The image intensifier itself is coupled to a video camera that captures and display
the image on a monitor. The image is transmitted to a monitor so that the movement of a body
part or of an instrument or contrast agent (“X-ray dye”) through the body can be seen in
detail. Fluoroscopy provides real-time X-ray images that are especially useful for guiding a
variety of diagnostic and interventional procedures. The ability of fluoroscopy to display
motion is provided by a continuous series of images produced at a maximum rate of 25-30
complete images per second. Modern fluoroscopes couple the screen to an X-ray image
intensifier and charge coupled device (CCD) video camera allowing the images to be recorded
and played on a monitor.
The X-ray exposure needed to produce one fluoroscopic image is low (compared to
radiography), whereas long exposures to patients can result from the large series of images
that are encountered in fluoroscopic procedures. Therefore, the total fluoroscopic time is one
of the major factors that determine the exposure to the patient from fluoroscopy. The X-ray
beam is usually moved over different areas of the body during a procedure, there are two very
different aspects that must be considered. One is the area most exposed by the beam, which
results in the highest absorbed dose to that specific part of the skin and to specific organs. The
other is the total radiation energy imparted to the patient’s body, which is related to the Kerma
Area Product (KAP or PAK), a quantity that is easily measurable.
The absorbed dose to a specific part of the skin and other tissues is of concern in fluoroscopy
for two reasons: one is the need for minimizing the dose to sensitive organs, such as the
gonads and breast, by careful positioning of the X ray beam and using shielding when
appropriate. The second is the possible incidence of the radiation beam to an area of the skin
for a long time that can result in radiation injuries in cases of very high exposure.
The fluoroscope is a type of X-ray machine that can use either a continuous or a pulsing X-ray
beam. The fluoroscope uses X-rays and as they pass through a patient's body the tissues and
bone will absorb different amount of X-rays. The transmitted beam / rays hits an image
intensifier that increases the brightness of the image many times (e.g. x1000 to x5000) so that
it can be viewed on a display screen. The image intensifier itself is coupled to a video camera
that captures and encodes the two-dimensional patterns of light as a video signal from the X-
ray machine. The signal is converted back into a pattern of light seen as the image on the
monitor.
The fluoroscope consists of X-ray generator, X-ray tube, collimator, filters, patient table, grid,
image intensifier, optical coupling and television system. Fluoroscopy is normally performed
using 2-6 mA and an accelerating voltage of 75 to 125 kVp. The rate of X-ray production is
directly proportional to the tube current, but is more sensitive to increasing kVp than mA. For
example, increasing the kVp by 15% is equivalent to doubling the mA. [The flow of electrons
from the filament to the target is called the tube current and is described in units of
milliamperes (mA)].
Fluoroscopy units are usually operated in an automatic brightness control (ABC) model in
which a sensor in the image intensifier monitors the image brightness. When there is
inadequate brightness, the ABC increases the kVp first, which increases the X-ray penetration
through the patient, and then adjusts the mA to increase the brightness. The thicker tissue or
larger cross-sectional area, such as the abdomen, will require greater exposure.
A major change from conventional fluoroscopy to digital fluoroscopy (DF) is the use of a
charge-coupled device (CCD) instead of a TV camera pickup tube. The CCD has greater
sensitivity to light (detector quantum efficiency [DQE]) and a lower level of electronic noise
than a television camera tube. The result is a higher signal-to-noise ratio and better contrast
resolution (1000 lines than the 525 lines for conventional fluoroscope). These characteristics
also result in substantially lower patient dose. The response of the CCD to light is very stable
and warm-up is not required. It has essentially an unlimited lifetime and requires no
maintenance. Perhaps the single most important feature of CCD imaging is its linear response.
The speed of acquisition is very high and the usual pixel numbers in an image are 1024 x
1024 than the 512 x 512 for conventional. The digital image allows for post processing and
electronic storage and distribution.
Introduction
Medical Thermography (MT) is the science of visualizing the temperature (heat) distribution
pattern over the surface of the human skin. The term comes from the Greek words "therme,"
meaning heat and "graphos," meaning writing or drawing. This technique involves the
detection of infrared (IR) radiation (heat energy) that can be directly correlated with the
temperature distribution of a defined body region. Infrared camera generates images based on
the amount of heat dissipated at the surface. This is a non-invasive, non-contact method that
uses heat from the body to aid in making diagnostic tool for analysing physiological functions
related to the control of skin-temperature. This rapidly developing technology is used to
detect and locate thermal abnormalities characterized by an increase or decrease found at the
skin surface. Infrared cameras generate images based on the amount of heat dissipated at the
surface by infrared radiation. The technology is a sophisticated way of receiving
electromagnetic radiation and converting it into electrical signals. These signals are finally
displayed in gray shades or colors which represents temperature values. The image is actually
the map of body’s temperature / heat distribution pattern and is referred as “Thermogram”. It
identifies physical changes in the body at the cellular level. It detect skin temperature
variances caused by new blood vessel growth reflecting possible predisposition to cancer,
hormonal changes, cardiovascular diseases, diabetes, musculo-skeletal changes and soft
tissue injury.
Thermography is a sophisticated clinical imaging tool which records thermal changes in skin
temperature. Medical thermography is so sensitive that it can detect slight and dynamic
temperature changes (0.050C) on the surface of the skin. Human body has higher degree of
thermal symmetry in normal condition and any subtle (abnormal) temperature asymmetry can
be easily identified. It helps to detect the pathological changes in some circumstances and
used as an early diagnostic tool. It is safe for women with mastectomies, implants, and can be
performed on all ages. The examination of the female breast as a reliable aid for diagnosing
breast cancer is probably the best known application of thermography.
Physics of Thermography
Any object having a temperature above absolute zero (00K) or higher, radiates heat in the
form of infrared radiation (wavelength, 0.75-100 µm) and sometimes called heat wave.
1
Therefore, infrared is an indicator of temperature of an object. The total energy “W” emitted
by the object and its temperature are related to Stefan Boltzman formula,
W = ϬЄT4
Where, W = Radiation flux density and is expressed in W/Cm2
Є = Emissivity factor
Ϭ = Stefan-Boltzman constant = 5.67 x 10-12 W/ Cm2K4
T = Absolute temperature
Emissivity refers to an object’s ability to emit radiation. Regarding the spectral region,
human skin is a black body radiator with an emissivity factor of 0.98 and is therefore a
perfect emitter of infrared radiation at room temperature. Human skin emits infrared radiation
mainly in the wavelength range of 2–20 μm with an average peak of 9–10 μm. Based on
Plank’s Law roughly 90% of the emitted infrared radiation in humans is of longer wavelength
(8–15 μm).
IR Imaging System
An infrared scanning device convert IR radiation emitted from the skin surface into a visual
image that depicts temperature variations on a monitor. Most systems have a wide range of
absolute temperature sensitivity ranging from 1 to 50 0C. The IR equipment usually has two
parts, the IR camera and a standard PC or laptop computer. These systems have only a few
controls and relatively easy to use. The main IR camera components are a lens, a detector in
the form of a focal plane array (FPA), possibly a cooler for the detector. Most detectors have
a response curve that is narrower than the full IR range (~ 900–14,000 nm or 0.9–14µm). The
optical system of IR camera scans the field of view at a very high speed and focuses the IR
radiation onto the detector which covert the radiation into an electrical signal. The signal
from the camera is amplified, processed and stored in the computer. The image is finally
displayed on a monitor with adjustable controls for contrast (temperature range) and
brightness (temperature level) on the display unit.
The PC or laptop computer can able to store tens of thousands of images (and these images
may be retrieved for later analysis). The ability to statistically analyse the thermogram at a
later date is very important in clinical work. Copies of images can easily be sent (via e-mail,
floppy disk, etc.) to referring doctors or other healthcare professionals. Monitors are high-
resolution full colour, isotherm or grey scale, and usually include image manipulation,
isothermal temperature mapping, and point-by-point temperature measurement with a cursor
or statistical region of interest. The systems measure temperatures ranging from 10° C - 55°
C to an accuracy of 0.1° C. Focus adjustment should cover small areas down to 75 x 75mm.
2
IR Imaging System
IR Detectors
IR detectors are used to convert IR energy into electrical signals. The two main types of
detectors: Thermal and Photo detectors. Thermal detectors respond to temperature changes
from incident IR radiation through changes in physical and electrical properties. Photo
detectors generate free electrical carriers through interaction of photons and bound electrons.
Thermal detectors include thermocouple and thermistor bolemeter. Photo detectors are
semiconductors with narrow band gaps, eg. Indium Gallium Arsenide (InGaAs), Germanium
(Ge), Lead Selenide (PbSe), Indium Antimonide (InSb), Indium Arsenide (InAs), Mercury
Cadmium Telluride (MCT, HgCdTe), etc. The response time and sensitivity of photo detector
can be higher but they generally have to be cooled in order to cut thermal noise.
3
Most IR cameras use InSb detector which detects IR radiation in the range of 2-6 µm (2.4%
of total IR). They are highly sensitive and are capable of detecting small temperature
variations as compared to a thermistor.
MCT also find application which detect IR rays in the range of 0.8-25 µm but need to cool
with liquid nitrogen.
Medical thermography is a reliable aid for successful evaluation of breast cancer. It is also
find application in the assessment and monitoring of inflammatory joints diseases, diagnosis
of deep vein thrombosis, peripheral vascular diseases, neurological disorders, musculo
skeletal disorders and mass fever screening.
4
OLD TECHNIQUE NEW TECHNIQUE
Insufficient software and tools User-friendly image processing software
5
ULTRASOUND IMAGING SYSTEM
Ultrasonography is a useful and flexible modality in medical imaging, and often provides an
additional or unique characterization of tissues, compared with other modalities such as
conventional radiography or computer tomography (CT). Ultrasound relies on properties of
acoustic physics (compression/rarefaction, reflection, impedance, etc.) to localize and
characterize different tissue types. Ultrasonography can be used to examine many parts of the
body, such as the abdomen (liver, kidney, spleen, pancreas, bladder, etc.), female reproductive
system, prostate, heart, major blood vessels, foetus and even possible detection of cysts, tumors
in these organs. It has also been useful to image thyroid glands, eyes, breasts, and a variety of
other superficial structures. Ultrasound studies have also been extensively used to investigate the
dynamics of blood flow in the cardiovascular system.
Ultrasonography is increasingly being used in the detection and diagnosis of heart disease, heart
attack, and vascular diseases that can lead to stroke. It is also used to guide a fine needle, for
tissue biopsy and amniocentesis, and to assist in taking a sample of cells from an organ or
structure for lab testing (for example, a test for cancer in breast tissue and fetal growth in uterus).
Superficial structures such as muscles, tendons, testes, breast and the neonatal brain are imaged
at a higher frequency (7-15 MHz), which provides better axial and lateral resolution. Deeper
structures such as liver and kidney are imaged at a lower frequency 2-6 MHz with lower axial
and lateral resolution but greater penetration.
An ultrasound transducer sends an ultrasound pulse into tissue and then receives echoes back.
The echoes contain spatial and contrast information. The concept is analogous to sonar used in
nautical applications, but the technique in medical ultrasound is more sophisticated, gathering
enough data to form a rapidly moving two-dimensional gray scale image. Some characteristics of
returning echoes from tissue can be selected out to provide additional information beyond a gray
scale image. Doppler ultrasound, for instance, can detect a frequency shift in echoes, and
Ultrasound Machine
Medical Applications
Ultrasound examinations can help to diagnose a variety of conditions and to assess organ
damage following illness. Ultrasound is used to help physicians diagnose symptoms such as:
pain
swelling
infection
Ultrasound is a useful way of examining many of the body's internal organs, including but not
limited to the:
heart and blood vessels, including the abdominal aorta and its major branches
liver
gallbladder
spleen
pancreas
kidneys
bladder
uterus, ovaries, and unborn child (fetus) in pregnant patients
eyes
thyroid and parathyroid glands
scrotum (testicles)
Ultrasound is also used to:
guide procedures such as needle biopsies, in which needles are used to extract sample
cells from an abnormal area for laboratory testing.
image the breasts and to guide biopsy of breast cancer
diagnose a variety of heart conditions and to assess damage after a heart attack or other
illness.
Doppler ultrasound images can help the physician to see and evaluate:
blockages to blood flow (such as clots)
narrowing of vessels (which may be caused by plaque)
tumors and congenital malformation
With knowledge about the speed and volume of blood flow gained from a Doppler ultrasound
image, the physician can often determine whether a patient is a good candidate for a procedure
like angioplasty.
Physics of Ultrasound
Ultrasound is simply sound waves, like audible sound. Although some physical properties are
dependent on the frequency, the basic principles are the same. Sound consists of waves of
compression and decompression of the transmitting medium (e.g. air or water), travelling at a
fixed velocity. Sound is an example of a longitudinal wave oscillating back and forth in the
direction the sound wave travels, thus consisting of successive zones of compression and
rarefaction. Transverse waves are oscillations in the transverse direction of the propagation. (For
instance surface waves on water or electromagnetic radiation.)
Schematic illustration of a longitudinal compression wave (top) and transverse wave (bottom). The bottom figure
can also represent the pressure amplitude of the sound wave.
Ultrasound obeys the law of reflection and refraction same as light. Ultrasound waves may be
longitudinal, transverse or shear but diagnostic applications, only longitudinal waves are used.
Ultrasound waves are vibration or disturbances causing of alternating zones of compression and
rarefaction in a medium. A single sound wave (one compression and one rarefaction) has a
temporal (duration / period, Sec) and a spatial (λ, mm) dimension. It causes particles in the
medium to oscillate to and fro (back & forth) at a particular frequency so that mechanical energy
is transported across the medium. This can be treated as sine wave having wavelength (λ),
frequency (f) and amplitude (A) depending on the sound pressure characteristics and transmitting
medium.
Sound of different frequencies travel at same speed in the same medium i.e. it is the medium that
determines the velocity of the sound waves. Velocity is the speed at which sound waves travel
through a particular medium and is equal to the product of frequency and wavelength.
𝑉=𝑓𝑥𝜆
Sound propagation velocity is greater in stiffer (high molecular bond strength) material with
smaller mass. Propagation speed increases from gas to liquid and highest in solids.
Solid ˃ Liquid ˃ Gas
Ultrasonic energy is transmitted through a medium as a wave motion and, therefore, no net
movement of the medium is expected to occur. The velocity of propagation of the wave motion
is determined by the density of the medium it is travelling through and the stiffness of the
medium. At a given temperature and pressure, the density and stiffness of the biological
substances are relatively constant, and, therefore, the sound velocity in them is also constant. The
velocities of ultrasound through human soft tissue are 1540 m/s and for bone 4080 m/s [air 330
m/s].
The knowledge of velocity of sound in a particular medium is important in calculating the depth
to which the sound wave has penetrated before being reflected. If the time taken by the ultrasonic
wave to move from its source through a medium, reflect from an interface and return to the
source can be measured, then the depth of penetration is given by:
Velocity of sound in the medium x time
Depeth of peneration =
2
Sound and ultrasonic waves consist of a mechanical disturbance of a medium such as air. The
disturbance passes through the medium at a fixed speed causing vibration. The rate at which the
particles vibrate is the frequency, measured in cycles per second or Hertz (Hz). The pressure of
sound is reported on a logarithmic scale called sound-pressure level, expressed in decibel (dB).
The effect is demonstrated by crystals of materials like quartz, tourmaline and Rochelle salt. This
phenomenon offers an excellent method for converting electrical energy into mechanical energy
and vice versa.
Quartz Crystal
While working with natural crystals, it is difficult to establish the appropriate axis and cut the
crystal in the required form. Therefore, quartz has generally been replaced by synthetic
piezoelectric materials namely barium nitrate and lead zirconate titanate (PbZrTiO4). They offer
several advantages because they are far cheaper to produce and are much easier to construct
transducers of complex shape and large areas. They can be moulded to any shape to obtain a
better focusing action for producing high intensity ultrasonic waves.
The choice of piezo-electric material for a particular transducer depends upon its applications.
Materials with high mechanical Q factor are suitable as transmitters whereas those with low
mechanical Q and high sensitivity are preferred as receivers and in case of non-resonance
applications. There are three parameters that are important in optimizing transducers for various
types of applications. These are frequency, active element diameter and focusing. Their effects
on the performance are as follows:
Frequency also influences lateral resolution by affecting beam divergence. The following rule
applies, assuming all other factors remain constant.
Active Element Diameter (AED): As the transducer face diameter increases, the beam width
decreases and therefore, lateral resolution improves.
Focusing: Focusing a transducer is a means of minimizing the beam width and adjusting the
focal zone to give optimum results for a particular examination. Acoustic lenses can be used to
shape the ultrasonic beam pattern. The width of the beam can be made narrow with the result that
better lateral resolution can be obtained. The focal point can be selected at different depths from
the face of the transducer. The ability to select different focal points allows for the optimization
of transducers for a particular type of studies. Modern transducers are internally focused and
externally are of flat face.
For production of ultrasound, piezoelectric crystal is placed in between two metal plates and
connected to an AC voltage. The crystal vibrates mechanically and the amplitude becomes max.
when the frequency of the applied AC voltage is equal to the natural frequency of the crystal
(called PZ resonator / crystal resonator). If the thickness of the crystal is small, the frequency is
high.
In diagnostic purposes, sufficient amount of signal has to be reflected from the interface for
electronic processing, no additional energy is necessary. Therefore, considerably lower
ultrasonic power levels are employed for diagnostic applications. A cross sectional images can
be formed from a mapping of echo intensities. The intensity of the echo is used to determine the
brightness of the image at the reflecting tissue surface. Diagnostic ultrasounds are used either as
continuous waves or in the pulsed wave mode. Applications making use of continuous waves
depend for their action on Doppler’s effect. Among the important commercially available
instruments based on this effect are the foetal heart detector and blood flow measuring
instruments.
There are, however, many applications where only pulsed waves can be employed. In fact, the
majority of modern ultrasonic diagnostic instrumentation is based on the pulse-technique. Pulse
echo based equipment is used for the detection and location of defects or abnormalities in the
structures at various depths in the body. The pulse-echo technique, basically, consists in
transmitting a train of short duration ultrasonic pulses into the body and detecting the energy
reflected by a surface or boundary separating two media of different specific acoustic
impedances.
Extracorporeal shock wave lithotripsy (ESWL) is a special use of kidney ultrasound, where high
intensity focused ultrasound pulses are used to break up calcified stones in the kidney, bladder,
or urethra. Pulses of sonic waves pulverize dense renal stones, which are then more easily passed
through the ureter and out of the body in the urine. The ultrasound energy at high acoustic power
levels is focused to a point exactly on the stone for maximum acoustic transmission. The acoustic
power of sound and ultrasound is the energy delivered per unit of time. The power is measured in
Watt (W) and is proportional to the square of the amplitude. 1 W = 1 joule/second. The intensity
(I) of a wave is the rate of power through a unit region perpendicular to the direction of
propagation. The unit of intensity is watts per square meter (W/cm2).
The therapeutic uses of ultrasound fall into two categories: inducing nondestructive heating or
other, mechanical effects to stimulate or accelerate normal physiological response to injury; and
the production of controlled, selective destruction of tissue. The first category includes
physiotherapeutic applications; focused ultrasound surgery (FUS) falls into the second.
Ultrasound is used by physiotherapists for a number of purposes, including relieving pain,
Acoustic Impedance
Ultrasound travels in a very straight line. An ultrasonic wave is reflected when it strikes an
interface between materials with different speeds of sound (acoustic impedance). Furthermore,
an interface between materials with a larger difference in acoustic impedance reflects ultrasonic
waves more strongly and that with a smaller difference in acoustic impedance reflects them less
strongly and lets part of them travel through. For example, the human body consists of a variety
of cells and tissues with their acoustic impedance different from each other. Ultrasonic waves
striking these cells or tissues are reflected differently. Using the difference in energy level of the
echoes makes it possible to image the internal tissues of the body. Ultrasonic waves are gradually
attenuated to become weaker while travelling through the medium. Those with a higher
frequency show a higher attenuation factor.
Acoustic impedance (Z) is a physical property of tissue. It describes how much resistance
ultrasound beam encounters as it passes through a tissue. Acoustic impedance of a material is
actually a measure of its opposition to the propagation of sound waves. On the other hand, it is
characterized as the ability of tissues to transmit sound waves. Acoustic impedance is a constant
for a specific material and is analogous to electrical impedance. It is defined as the product of the
density of the medium with the velocity of sound wave in the same medium.
Z = ρ0 x c
Where, Z = Acoustic impedance (kg/m2s), ρo = density of the tissue (kg/m3), c = speed of the
sound wave (m/s)
So, if the density of a tissue increases, impedance increases. Similarly, but less intuitively, if the
speed of sound increases, then impedance also increases. The effect of acoustic impedance in
medical ultrasound becomes noticeable at interfaces between different tissue types. The ability of
an ultrasound wave to transfer from one tissue type to another depends on the difference in
impedance of the two tissues. If the difference is large, then the sound is reflected. Similarly,
when an ultrasound beam passes through muscle tissue and encounters bone, it reflects off of it
due to the difference in density between the tissues.
Biological tissues are not homogenous acoustically i.e. ρ & v are not constant within the
medium. The acoustic impedance determines the degree of reflection and refraction between two
The SI unit for acoustic impedance is the Rayl, kg/(m2s), after J W Strutt, 3rd Baron Rayleigh.
In ultrasound, absorption is the reduction in intensity of the sound waves as it passes through
tissue. An absorption result in heat due to frictional forces as the particles in the medium opposes
the motion. The amount of absorption is determined by the viscosity of the medium, its
relaxation time and frequency of ultrasound. Quantitatively, the average value of sound
absorption in soft tissues is of the order 0.5-1 dB/cm/MHz.
Ultrasound pulses lose energy continuously as they move through matter. This is unlike x-ray
photons, which lose energy in "one-shot" photoelectric or Compton interactions. Scattering and
refraction interactions also remove some of the energy from the pulse and contribute to its
overall attenuation, but absorption is the most significant. Overall, the beam energy decreases
more or less exponentially as it penetrate the tissue (similar to the attenuation of x-ray).
𝐼𝑥 = 𝐼0 𝑒 −𝜇𝑥
Where, I0 = Initial intensity of US
Ix = Intensity of US at a distance x
μ = Attenuation coefficient (dB/cm)
1 𝐼𝑥
𝜇 = − ( ) 10 log( )
𝑥 𝐼0
Conversion of transmitted energy to another form of energy such as heat (absorption) is the
primary means by which attenuation of ultrasound occurs in biologic tissue, with scatter
comprising the other significant contributing factor. The intrinsic property of a medium to
attenuate sound waves at a given frequency may be represented by its attenuation coefficient.
The rate at which an ultrasound pulse is absorbed generally depends on two factors: (1) the
material through which it is passing, and (2) the frequency of the ultrasound. The attenuation
(absorption) rate is specified in terms of an attenuation coefficient in the units of decibels per
centimeter. Since the attenuation in tissue increases with frequency, it is necessary to specify the
frequency when an attenuation rate is given. The attenuation through a thickness of material, x, is
given by:
Attenuation (dB) = (μ) (f) (x)
Scattering occurs when a sound wave strikes a structure with different acoustic impedance to the
surrounding tissue and which is smaller than the wavelength of the incident sound wave. Such
structures are known as “diffuse reflectors,” with examples being red blood cells and non-smooth
surfaces of visceral organs. In contrast to “specular reflectors”, tissues with smooth interfaces
from which ultrasound waves are reflected in a specular fashion. Diffuse reflectors cause
ultrasound waves to scatter in all directions thus resulting in multiple echoes propagating from
the numerous tiny structures. Not only does this scattering result in echoes with smaller
amplitudes (compared to specular reflection) but the scattered echoes also interact with each
other. This interaction causes constructive and destructive interference of the waves. The
resultant image is termed “speckle” due to the various intensities of the echoes received by the
transducer, and this is seen as an irregularity in the grey scale of the image. Most echoes from
ultrasound imaging arise from scattering, rather than the reflection from specular reflectors. The
speckle arising from this scatter results in the grainy appearance of the parenchyma of organs
and also the signal in Doppler ultrasound.
Reflection of a sound wave occurs when the wave passes between two tissues of different
acoustic speeds and a fraction of the wave 'bounces' back. This forms one of the major principles
of ultrasound imaging as the ultrasound probe detects these reflected waves to form the desired
image.
Types of Echoes
Specular - echoes originating from relatively large regularly shaped objects with smooth
surfaces. These echoes are relatively intense and angle dependent (i.e. IVUS, valves).
Scattered - echoes originating from relatively small, weakly reflective, irregularly shaped objects
are less angle dependent and less intense (ie. blood cells).
There are three parameters that are important in optimizing transducer for various types of
applications: (a) frequency, (b) active element diameter, and (c) focusing.
US Probe Design:
Ultrasound probe (called a transducer) is a hand-held device that is placed directly on and moved
over the patient. A single transducer is now rarely used instead a multiple crystals (multi-
element) transducers are most common for probe design. In multi-element probe, each crystal
has its own circuit. The advantage is that the ultrasound beam can be controlled by changing the
timing in which each element gets pulsed. Probes are formed in many shapes and sizes. The
shape of the probe determines its field of view. Transducers are described in megahertz (MHz)
indicating their sound wave frequency. The frequency of emitted sound waves determines how
deep the sound beam penetrates and the resolution of the image. The field of view (FOV) is the
plane or area depicted by the ultrasound transducer.
4. Matching layer
interface between the transducer element and the tissue.
plastic polymer layer (Perspex/Plexiglas) reduces the impedance mismatch between
crystal and soft tissues.
allows close to 100% transmission of the ultrasound from the element into the tissues
by minimizing reflection due to traversing different mediums (Z M = 6.7 x 106 kg/m2s)
achieves this by consisting of layers of material with acoustic impedances that are
between soft tissue and transducer material.
- may consist of one or multiple layers
- each layer is one-quarter wavelength thick
𝑍𝑀 = √𝑍𝑇 𝑥𝑍𝑇𝑖
5. Housing
electrical insulation and protection of the element
includes a plastic case, metal shield and acoustic insulator
Under normal conditions, the SCR is non-conducting. The capacitor C1 can charge through the
resistance R to the +V potential. If a short triggering positive pulse is applied to the gate of the
SCR, it will fire and conduct for a short time. Consequently, the voltage at ‘A’ will fall rapidly
resulting in a short duration, high voltage pulse at ‘B’. This pulse appears across the crystal
which generates short duration ultrasonic pulse. For producing a pulse with a very short duration
it is necessary to use an SCR with a fast turn ‘on’ time and high switching current capability,
which can be able to withstand the required supply voltage. SCR 2N4203 can be used because of
its high peak forward blocking voltage (700 V), high switching current capability (100 A) and
fast turn-on time (100 ns).
Receiver: The function of the receiver is to obtain the signal from the transducer and to extract
from it the best possible representation of an echo pattern. To avoid significant worsening of the
axial resolution, the receiver bandwidth is about twice the effective transducer bandwidth.
Types of Probes
Product for 2D imaging
Radius
Central
Probe type Feature Application Element of
frequency
curvature
Vasucular application
•Arteria carotis
•IMT,FMD measurement
for early detection of 2.5MHz 64ch
arterial sclerosis | | -
venipuncture,blood vessel 12MHz 256ch
wide footprint and keep visualization
Linear type same field of view at deep Breast,Thyroid
part. Tendon,arthrogenous
Abdominal application.
2.5MHz 64ch 5R
Transvaginal and Transrectal
| | |
application.
7.5MHz 192ch 80R
Convex(Curved) wide footprint, field of Diagnoses organs.
type view will be spreaded at
deep part.
Cardiac application.
2MHz 32ch
Transesophageal application.
| | -
Abdominal application.
7.5MHz 128ch
small footprint, field of Brain diagnosis
Phased(Sector) view will be spreaded
type widely at deep part.
Ophthalmology
Transesophageal application.
Transvaginal and Transrectal 2MHz
application. | - -
Transurethral application. 22MHz
The blood flow measurement with
Single type A) The measurement of Doppler.
distance.
B) B mode scanning by
mechanical rotation.
Transducer Focusing
Transducers can be designed to produce either a focused or non-focused beam, as shown in the
following figure. A focused beam is desirable for most imaging applications because it produces
pulses with a small diameter which in turn gives better visibility of detail in the image. The best
detail will be obtained for structures within the focal zone. The distance between the transducer
and the focal zone is the focal depth.
An unfocused transducer produces a beam with two distinct regions, as shown in the previous
figure. One is the so-called near field or Fresnel zone and the other is the far field or Fraunhofer
zone. In the near field, the ultrasound pulse maintains a relatively constant diameter that can be
used for imaging.
In the near field, the beam has a constant diameter that is determined by the diameter of
the transducer. The length of the near field is related to the diameter, D, of the transducer and the
wavelength, l, of the ultrasound by
Near field length = D2/4l.
The ultrasound beam intensity decreases with time, therefore, intensity of returning echoes are
compensated with time gain. As a result, exact amplitude of the echoes is reproduced from the
same impedance difference irrespective of the depth. The reflected or echo pulses provide two
types of information: (a) amplitude indicates the relative density / thickness of the tissue layers
and represented in the form of brightness, and (b) time between transmitting and receiving a
pulse reveals the tissue depth or position within a structure or organs.
Basically, all ultrasound imaging is performed by emitting a pulse, which is partly reflected from
a boundary between two tissue structures, and partially transmitted. The reflection depends on
the difference in impedance of the two tissues. The pulse is thus emitted, and the system is set to
await the reflected signals, calculating the depth on the basis of the time from emission to
reception of the signal. The total time for awaiting the reflected ultrasound is determined by the
preset depth desired in the image.
D = (v x t)/2 Where, D is the depth / position of the tissue
Schematic illustration of the reflection of an ultrasound pulse emitted from the probe P, being reflected at a, b and
c. Part of the pulse energy is transmitted from the scatterer a, the rest is transmitted, part from b and the rest from c.
When the pulse returns to P, the reflected pulse gives information of two measurements: the amplitude of the
reflected signal, and the time it takes returning, which is dependent on the distance from the probe
Real time imaging system is an automatic scanning and is generally grouped into (a) Mechanical
and (b) Electronic Scanner. Mechanical scanner represents a low cost technique extending the
performance of a manually scanned system to achieve 2D field of view. The probe consists of
one or more transducers depending upon the need and types. The image is produced by rapidly
rotating the crystal to generate fan-shaped ultrasound beam, resulting a V-shaped sector image.
Depending upon the movement of transducer, mechanical scanner classified as (1) Rotating
wheel transducer, and (2) Oscillatory motion transducer.
Electronic Scanners
Electronic scanner is a multi-element (i.e. multiple piezo electric crystal element) array
transducers. Mechanical scanning methods have almost entirely been replaced with electronic
scanner. Electronic scanner differs from mechanical scanner both in construction and beam
deflecting system. There are two basic types of electronic scanners: (a) Linear sequence /switch
transducer arrays (bilinear or curvilinear) and (b) Phase transducer arrays (linear or annular).
a. Linear Sequential Array
The linear array is formed from a large number of (up to 256) individual transducer
elements covering 4 to 10 cm length. Each parallel rectangular transducer has dimension
2mmx10mm or 4mmx18mm. A group of transducer called aperture which typically
consists of 16 to 32 elements. During each transmission / receive cycle only one
transducer group is active. The 1st aperture produces one scan line and then electronically
shifted over next one element with exclusion of 1 st one and forms the 2nd scan line. In this
way, it produces 100-115 scan lines for a complete sectional image.
Transmit pulse are applied to each element via a delay which gives a swiveled (angled)
wave front. The degree of swivel requires very narrow elements of about ½ λ diameter.
The effective aperture A' = A Cosθ. Annular phase array consist of a series of concentric
transducers. This produces a conical beam section.
Linear Scan:
This linear type uses the world's first "linear scan" technology that allows the built-in probe to
directly move in a lateral direction (reciprocal scan) by a motor. Compared to the traditional arc
scanning, this linear scanning provides higher lateral resolution of diagnostic images. The probe
also has a flat surface that touches the target part of the human body (the surface that sends and
receives ultrasonic waves). This geometry allows the probe to ensure more uniform contact with
the irregular surface of breasts and other parts of the human body.
Speckle artifact may be encountered in ultrasound. It is caused by the scattering of waves from
the surface of small structures within a certain tissue. The artifact produces a textured
appearance.
Ultrasound beam width artifact occurs when a reflective object located beyond the widened
ultrasound beam, after the focal zone, creates false detectable echoes that are displayed as
overlapping the structure of interest.
A multipath artifact is an ultrasound beam artifact in which the primary beam reflects off
anatomy at an angle, resulting in a portion of the beam returning to the transducer, whilst another
portion takes a longer duration as it reflects a second structure. This phenomenon results in a
propagation path error in which the transducer will interpret a structured to be deeper than it is.
A Doppler ultrasound is a noninvasive test that can be used to evaluate blood flow and pressure
by bouncing high-frequency sound waves (ultrasound) off red blood cells. A Doppler ultrasound
can estimate how fast blood flows by measuring the rate of change in its pitch (frequency). It
helps doctors evaluate blood flow through the major arteries and veins of the arms, legs, and
Colour Flow Imaging (CFI) or Colour Velocity Imaging (CVI) or Triplex Ultrasound
The most recent technology change in diagnostic ultrasound is color flow imaging (CFI), which
is a merging of gray-scale and motion-detection processing to produce an image that depicts soft
tissue in gray scale and blood flow in color. Color-flow imaging (also called triplex ultrasound)
is an enhanced form of Doppler ultrasound technology. This technique is similar to duplex
ultrasound and use third parameter i.e. color to highlight the direction of blood flow. Vessels in
which blood is flowing are colored: red for flow in one direction and blue for flow in the other,
with a color scale that reflects the speed of the flow.
CFI is a US imaging modality that processes echo signals for amplitude and motion information
to produce a composite image that depicts soft tissues in gray scale and blood flow in color in a
real-time format. The motion information comes from either Doppler signal processing on a
direct measurement of movement oven time. The system combines the signal processing to form
anatomic (soft tissue) and physiologic (blood-flow) information in the same image at the same
time.
This technique is used to depict the peripheral arterial flow and blood cells geometry. The echo
image is used to locate the artery and determine the position of the walls. The echo image is then
used as a guide to map out the flow field with the Doppler shift. The frequency shifts are
encoded as colour bar. In a typical display, blood flowing towards the transducer (positive
Doppler shift) is coded as red and away (negative shift) as blue. Various intermediate colours
represent the different velocities of the flow.
Color flow Doppler ultrasound produces a color-coded map of Doppler shifts superimposed onto
a B-mode ultrasound image (color flow maps). Although color flow imaging uses pulsed wave
ultrasound, its processing differs from that used to provide the Doppler sonogram. Color flow
imaging may have to produce several thousand color points of flow information for each frame
superimposed on the B-mode image. Color flow imaging uses fewer, shorter pulses along each
color scan line of the image to give a mean frequency shift and a variance at each small area of
measurement. This frequency shift is displayed as a color pixel. The scanner then repeats this for
several lines to build up the color image, which is superimposed onto the B-mode image.
An arterial duplex ultrasound showing blood flow A venous duplex ultrasound showing blood flow
in the right femoral artery in the leg in right Sapheno-femoral junction in the leg
Colour velocity imaging (CVI) identifies a group of blood cells in a vessel and times their shift.
The grout which is identified at a timeT 1 has moved to a new position at T2 and the separation is
measured. A group of blood cells gives a unique signal which is followed for a fixed time in a
“window”. The blood velocity is computed from the time. When utilizing this equipment for
ultrasonography, two displays are available. These are color-flow Doppler and gray-scale B-
mode. The color-flow Doppler displays a flow velocity distribution while the gray-scale B-
mode provides an anatomic image.
Duplex ultrasound produce images show how blood is moving through arteries and veins. It
enables the physician to see the speed and direction of blood flow and where the blood flow may
be blocked. This type of Doppler examination provides a 2-dimensional (2-D) image of the
arteries so that the structure of the arteries and location of an occlusion can be determined, as
well as the degree of blood flow. This system is used in the evaluation of the extracranial carotid
arteries, iliofemoral arteries, and hemodialysis shunts. The system obtains information regarding
blood velocity, lumen area, plaque composition and ulceration, and hemodynamic turbulence.
Necessary equipment consists of an ultrasound machine with duplex capabilities. This will
include a probe with piezoelectric crystals. The machine and transducers enable varying
focusing, direction, and adjustment of the gain, resolution, and depth of the ultrasound waves.
Typical depth adjustments range from less than 1 cm to 20 cm. Equipment is also generally
compatible with multiple transducers which provide differing advantages. The linear array
transducers are beneficial for anatomic mapping of arterial and venous systems. Lower
frequency transducers, which are typically curved linear or phased array, are better for visceral
vessels or abdominal imaging studies. Two-dimensional transducers are also available which
provide the construction of three-dimensional imaging from ultrasound data.
A duplex ultrasound may help diagnose many conditions, including:
Blood clots
Poorly functioning valves in leg veins, which can cause blood or other fluids to pool in
legs (venous insufficiency)
Heart valve defects and congenital heart disease
Various arterial diseases are identified and monitored via duplex ultrasonography. These include
cerebrovascular, renal, mesenteric, and aortoiliac disease. Additional diagnostic testing should
follow a screening study and positive results. In the case of cerebrovascular arterial analysis,
duplex ultrasound is particularly useful for carotid arteries evaluation. This type of Doppler
examination provides a 2-dimensional (2-D) image of the arteries so that the structure of the
arteries and location of an occlusion can be determined, as well as the degree of blood flow. It is
highly sensitive and specific for the identification of plaques and can be used to characterize and
identify plaques which are more likely to increase the risk of future strokes.
Power Doppler sonography is a new technique that displays the strength of the Doppler signal in
color, omitting the other parameters of speed and direction. It has three to five times the
sensitivity of conventional color Doppler for detection of flow and is particularly useful for small
vessels and those with low-velocity flow. The frequency is determined by the velocity of the red
blood cells, while the power depends on the amount of blood present. All moving blood cells
(and, alas, all moving particles) yield a color signal, regardless of their speed or direction. The
more cells in motion there are, the stronger the color signal. In addition to the number of moving
cells in the sample, the diameter of the vessel studied influences the strength of the power
Doppler signal.
Power Doppler can get some images that are hard or impossible to get using standard color
Doppler. Power Doppler is most commonly used to evaluate blood flow through vessels within
solid organs. Power Doppler is also referred to as energy Doppler, amplitude Doppler and
Doppler angiography. Power Doppler sonography finds potential for detecting areas of ischemia
in the kidney, brain, and prepubertal testis and for demonstrating hyperemia in areas of
inflammation. In inflammatory conditions, power Doppler is valuable in depicting increased
flow in vessels that are dilated because of inflammatory response.
Safety Levels:
In low MHz frequency range there have been no confirmed biological effects in mammalian
tissues exposed in-vivo to unfocussed ultrasound with ISPTA (spatial peak temporal average
intensity) below 100mW/cm2 for less than 500 sec or for focused ultrasound with I SPTA below
1W/cm2 for less than 50 sec. furthermore, for exposure times more than one second and less than
500sec (unfocussed) or 50sec focused such effect have not been demonstrated at even high
intensities. The zone of safety was valid for 0.5 to 15 MHz and for all anatomic sites except the
eye. Re-exposures are limited to 10% and 30% per year. Biologically sensitive areas would be:
1. 1st trimester embryo
2. Fetus skeletal heating
3. Transcranial Doppler
4. Eye
5. Intracavity
A. Pulser
It produces pulses for the transducer probe. The pulse repetition frequency (PRF) of pulser is
controlled by the Master Clock which is the heart of the system. The master clock also
controls the time gain compensation (TGC).
B. Receiver
The same transducer (probe) will receive all the echoes from the object. A single ultrasound
wave may produce many echoes in the path. A series of echoes will be received during
scanning and the amplitude will vary depending upon the depth of tissue and its nature.
C. Signal Pre-processing
The signal pre-processing unit consists of the following sub-units:
1. Amplification:
Echo signals are small with a dynamic range from 100 & 150 dB (1μV to 316μV) and are
linearly amplified to millivolts range. The amplification is achieved in a wide band
amplifier with a desirable gain of 80-100dB. In modern instruments, the input amplifier is
dual gate MOSFET.
2. Time-Gain Compensation (swept gain):
It is known that the emitted ultrasound wave amplitude gets smaller as it penetrates
tissue, a phenomenon called attenuation. So, one might expect late echoes (from deep
layers) to have smaller amplitudes than early (superficial) echoes even if those layers
have the same echogenicity. A way to overcome ultrasound attenuation is time gain
compensation (TGC), in which signal gain is increased as time passes from the emitted
wave pulse. This correction makes equally echogenic tissues look the same even if they
are located in different depths.
Types of scanning arrangement: (a) Linear, (b) Sector & (c) Compound
The first is the A-mode, in which the amplitude of the reflections is plotted against time (or
distance assuming that the velocity of sound in the body is 1540 m·s-1). A-mode ultrasound is
used to judge the depth of an organ, or otherwise assess an organ’s dimensions. A-mode is has
been used extensively in midline echoencephalography and ophthalmologic scanning. Echo-
encephaloscope uses 1-3 MHz whereas echo-ophthalmoscope employs a 7.5–15 MHz pencil
type transducer.
The details of the arrival times of echoes from the interior tissue boundaries are illustrated in the
figure below. The boundary depths can be related to echo arrival times through the
proportionality constant 2/c.
The electronics of the receiver system tailors the received signal for maximum readability. This ‘tailoring” includes
a time gain control (TGC) that helps compensate for the ever-decreasing signal strengths from deeper tissues due to
the greater attenuation over longer paths. The rate of TGC increase should be approximately 1 dB/MHz for each
interval of time that corresponds to 1 cm of travel. For example, assuming c=1.54x105 cm s-1, the gain should
increase at a rate of approximately 154 dB ms-1.
B-mode ultrasound produces the sector (or pie segment-shaped) scans that one normally sees. B-
model machines represent the vast majority of machines in existence. These machines are used
in echocardiology, obstetrical scans, abdominal scans, gynecological scans, etc. These scans
require either a mechanical scanner transducer (the transducer moves to produce the sector
scan), or a transducer with a linear array operated as a phased array (as discussed in class).
A mechanical scanned B-mode ultrasound system, the transducer is rocked back and forth to
produce a corresponding sector scan. Instead of using a mechanical scanned probe, modern
machines use an electronically scanned phased array of piezoelectric probes. This is illustrated
in in which the direction of the ultrasound beam is altered by phasing the excitation of the array
elements.
Other derived modes are possible. One particularly useful mode is the C-mode in which an
image is derived by either placing a receiver opposite a transmitter (ignoring reflections), or by
deriving the cross-section C-model data from three-dimensional imaging.
Most modern machines can generate multiple modes, and the choice is left to the user. Most
modern machines can also display Doppler echo pictures in which the data being imaged are
due to Doppler shifts resulting from moving parts. This is particularly useful for imaging blood
flow.