Lecture 3
Lecture 3
Lecture-3
3.1 Introduction
1. Accuracy
| − |
= − × %
Where,
A= Accuracy
Xm= the measured value
XT= the true value.
Inaccuracy or measurement uncertainty is the extent to which a reading might be
wrong and is often quoted as a percentage of the full-scale reading of an
instrument. Inaccuracy is given as:
| − |
= × %
2. Precision
| − |
= − × %
Where,
Pi= the precision of the ith measured value.
Xi= the ith measured value.
= the mean of the measured values.
3. Repeatability
4. Reproducibility
5. Sensitivity
Sensitivity is the ratio of the change in the instrument output to the change
in the measured quantity. i.e,
∆ ∆
= =
∆ ∆
The sensitivity of measurement is therefore the slope of the straight line drawn to
represent the relationship between the measured quantity and output reading.
Nominal Characteristic
Nominal Characteristic
The effect when the zero reading of the instrument is modified by the change in
the ambient conditions is called the zero drift. The description of the zero drift is
illustrated in Fig.(3.3).
Nominal Characteristic
Zero drift
0 Input (measured quantity)
If the instrument suffers both the sensitivity drift and zero drift at the same time,
output (Instrument Reading)
Nominal Characteristic
Zero drift
6. Resolution
=
.
7. Tolerance
Are the maximum and minimum values of the quantity that the instrument
is designed to measure.
-20 oC 0V
Thermometer
60 oC 50 V
9. Significant Figure
10. Backlash
Backlash is the maximum distance or angle through which any part of the
mechanical system of instrument may be moved in one direction without applying
appreciable force or motion to the next part in the mechanical sequence.
11. Threshold
Threshold Point
Dead band
13. Hysteresis
Is the non-coincidence between the loading and the unloading curves of the
output characteristics of the measuring instrument.
Therefore, to minimize the values of the threshold and the dead band, it is
important should be minimized the effect of hysteresis. The hysteresis effect can
be minimized by making the iron parts of instruments small so that they
demagnetize themselves quickly. Also, the hysteresis effect may be minimized by
using an iron parts with low flux density such as the nickel-iron alloys.
Illustration of the
Examples – L3
Ex3.1 By using a mass scale, five readings were taken to measure a mass of 50 Kg
as follows: (50.8, 50.9, 50.7, 50.6 and 51) Kg
(ii)
| − |
= − × %
The mean value for the measured values is given as follows:
+ + + +
= =
Or
. + . + . + . +
=
Or
= .
| − . |
= − × %
.
Or
= 99.61 %
Ex3.2 A pressure gauge with a measurement range of 0-10 bar has a quoted
inaccuracy of 1% of the full-scale reading.
(i) What is the maximum measurement error expected for this
instrument?
(ii) What is the likely measurement error expressed as a percentage of the
output reading if this pressure gauge is measuring a pressure of 1 bar?
Sol.
(i)
= × = .
(10- Lecture-3) Assist. Prof. Dr. Haider K. Latif
Instrumentations and Measurements Static Characteristics of Measuring Instruments
(ii)
= . × = .
This example carries a very important message. Because the maximum measurement
error in an instrument is usually related to the full-scale reading of the instrument,
measuring quantities that are substantially less than the full-scale reading means that
the possible measurement error is amplified. For this reason, it is an important that
instruments are chosen such that their range is appropriate to the values being
measured, so that the best possible accuracy is maintained in instrument readings.
Clearly, if we are measuring pressures with expected values between 0 and 1 bar, we
would not use an instrument with a measurement range of 0-10 bar.
o
Ex3.3 An electrical mass measuring instrument is calibrated at 20 C and has the
following mass/voltage characteristics:
(ii)
=
At 20 oC
−
= = =
−
At 60 oC
−
= = =
−
Therefore,
= − =
And the sensitivity drift per oC change is given as:
= = . . /
Ex3.4 Two resistors, R1 and R2 are connected in series. The individual resistance
measurements, using digital ohmmeter, give R1=20.35 Ω and R2= 15.735 Ω.
Calculate the total resistance to the appropriate number of significant
figure.
Sol.
R1 = 20.35 Ω (four significant figures).
R2 = 15.735 Ω (five significant figures).
Now,
= + = . + . = . Ω
Thus the total resistance to the appropriate number of significant figure is given as:
= . Ω (four significant figures).
= − = Ω
= + = Ω
Ex3.5 A moving coil ammeter has a uniform scale with 50 divisions and a full
Sol.
= = = .
.