Model: HSY-H-III
1
Notice!
2. Please clean and disinfect the dry chemical urine analyzer according to the
I. Overview
1.1 Introduction
The dry chemical urine analyzer is used in conjunction with Beijing Huashengyuan Medical
Technology Co., Ltd. Urine-H, T series urine analysis test strips (dry chemical method). Through the
color development of the test strip, the urine glucose, pH, ketone bodies, nitrite, bilirubin, protein,
occult blood, urobilinogen, specific gravity, white blood cells, ascorbic acid, creatinine, urine calcium,
microalbumin in urine Fourteen indicators are tested.
The dry chemical urine analyzer adopts ultra-high brightness cold light source testing technology,
which has the characteristics of anti-environmental interference and long service life. It can complete
the determination of 14 biochemical components in urine samples within one minute, can correct the
influence of ambient temperature and ambient light, pH, hematuria and abnormally colored samples
on the test results, and can be connected to the urine sediment analyzer.
The dry chemical urine analyzer is an in vitro diagnostic medical device for professionals.
2
Relative humidity <80%
Communication interface RS-232 interface
Transmission rate 9600bps
Fuse F2AL 250V
Weight 6.5kg±0.2kg
Dimensions 340 ㎜×330 ㎜×290 ㎜
Printer Built-in thermal printer
Rated power supply AC220V, 50HZ
Power 60VA
Noise less than 70dB
Applicable test paper Urine-T/Urine-H urine analysis test strips
The dry chemical urine analyzer uses the principle of photoelectric colorimetry to determine the
content of the biochemical components in the urine according to the color changes produced by the
reaction between the reagent area on the test strip and the biochemical components in the urine.
1.4.1 Accuracy: The difference between the test result and the label value of the corresponding
reference solution is not more than one order of magnitude in the same direction, and there must be no
reverse difference. A positive reference result must not have a negative result, and a negative reference
result must not have a positive result.
1.4.2 Repeatability: The coefficient of variation (CV, %) of the reflectance test result of the analyzer is
≤1.0.
1.4.3 Stability: Within 8 hours of starting the analyzer, the coefficient of variation (CV, %) of the
reflectance test results is ≤1.0.
1.4.4 Main functions:
(1) After starting up, the system software can perform self-check on each part of the system, and enter
the system standby screen after the self-check is completed.
(2) When each test record needs to be displayed on one screen (without turning the screen), the screen
displays the plus sign system, and the printed results are "plus sign system + conventional unit
system" and "plus sign system + international unit system". .
(3) The system can display and print the results in Chinese and English, and the result unit should
have at least the international system of units.
(4) It has an output device to output data and information.
(5) The instrument can transmit data through the serial port connection with the computer;
3
(6) The instrument can store 2000 test results.
(7) Test data can be stored and memorized after power off.
(8) There is a fault prompt function:
a) It can prompt that the position of the test strip is incorrect;
b) For hardware failure, please contact the product seller or manufacturer;
c) The built-in printer is faulty.
(9) The stored inspection records can be browsed and printed by serial number.
(10) Can choose the type of test paper, Urine-H, T series urine analysis test strips.
(11) The system has a real-time clock, which can modify the date.
(12) The instrument has batch printing function.
(13) The instrument has a unit conversion function.
(14) The instrument has a Chinese-English conversion function.
(15) The instrument has a calibration function.
Upper and lower case, main support, transmission box, switching power supply, LCD screen,
workbench, built-in thermal printer, fan, main control board, test head control board, printing control
board, infrared scanning control board, LCD control board.
1.6 Mark
4
II. Functions and setting
2.1 Overview
20 12 - 01 - 01 08 : 08
(Manu) →
International unit
Conventional unit
(ID setting)
(Transport)
(No transporting)
(Print)
(No printing)
(Scan)
(No scanning)
Strip setting T11
T11
T14
(Advanced setting)
(Return)
5
2.2 Installation Environment
2.2.1 Please place the instrument on a stable and level table, and do not put it together with a vibration
source such as a centrifuge.
2.2.2 Do not place the instrument where it can be affected by chemicals, corrosive gases or strong
electromagnetic interference.
2.2.3 Please do not put the instrument in direct sunlight, humid or high temperature places.
2.2.4 The power connection should be well grounded, and the power supply should be completely
disconnected when the instrument is not in use or stored.
Figure 1 Figure 2
6
Figure 3
2.3.4 Install printing paper
⑴Choose a thermal printing paper roll with a width of 57 mm and a diameter of less than 45 mm.
(2) Put the printing paper roll into the paper roll box, insert one end of the printing paper under the
printer roll, the printer will automatically load the paper.(As shown in Figure 4)
⑶Put the printing paper through the paper outlet of the printer cover and close the printer cover. (As
shown in Figure 5)
Figure 4 Figure 5
2.3.5 Install the power cord
Plug one end of the power cord into the power jack on the back of the urine analyzer, and plug the
other end into the AC power source. (As shown in Figure 6)
2.3.6 Connect to computer
The urine analyzer can be connected to the RS232 interface of the computer through a communication
cable. Insert one end of the communication cable into the communication interface on the back of the
instrument, and insert the other end into the RS232 communication interface of the host computer. (As
shown in Figure 6)
Figure 6
7
2.3.7 Connect Printer
The urine analyzer can be connected to the interface of the printer through a communication cable.
Insert one end of the communication cable into the parallel communication interface on the back of
the instrument, and insert the other end into the communication interface of the printer.
2.3.8 External scanner
The urine analyzer can be connected to an ID scanning gun. Insert one end of the communication
cable of the scanning gun into the serial communication interface on the back of the urine analyzer.
(Search)
(Manu)
Input the NO. of 4 numbers, click ,The instrument automatically stores the modified content.
Return, Delete
8
Exactly search Search in batch
Total record:XXX
20 12 - 01 - 01 ------
At this point of time, the date is the date of the final result. Input the date: 20 12 - 01 01
and enter the sequence number in ------- After that, click on the button on the right
bottom corner, the corresponding sequence number will be shown. By clicking on the number, the
detail will be given and the information can be transmitted and printed by clicking . If the
sequence number is not entered, all the sequence number on the selected date will be given. After
selecting one of them, all the result will be shown, transmitted and printed by clicking
20 12 - 01 - 01 000000
<------->
20 12 - 01 - 01 009999
Input the date period in: 20 12 - 01 - 01 , input number range in ________ , to click
,it will show ‘XX records matched’, could be transmitted and printed by clicking . If not
enter the range of consultation numbers to be queried, all records within the entered date range can be
9
searched.
2.4.1.3 ID search:
The instrument has a buffer that can store 200 ID numbers. When the ID number is entered in the
cache, during the test, the ID number is assigned to the test result according to the principle of input
first, use first. After the ID number in the cache is used up, if you continue to test, the ID number
assigned by the new test result is the new ID number generated after the last ID number is increased
by 1.
(1)Directly input
In the main interface, press to enter the ID query interface. The screen will display a list of IDs
that already exist but have not been used (if no such ID exists, shows "empty"). Press to see all
the IDs, press to clear all the ID cache, press to enter ID number. After pressing
and enter the ID through keyboard, two options will appear: "Add ID" and "Overwrite the current
ID". If "Add ID" is selected, the newly inputed ID will be stored as a new ID; if "Overwrite the
current ID" is selected, the last ID will be replaced by new ID, and store in cache.
(2) ID Scanning Input:
First, make sure that the ID scanning function is enabled. Touch to enter the setting interface.
shows that the ID scanning function is enabled, and shows the otherwise. Touch it to
change the status of ID scanning function. After that, go back to the main interface, touch to
enter the ID number view interface, the scanner connected with the instrument, ID bar code will be
scaned and the ID number will be input ID cache.
20 12 - 01 - 01 08 : 08
10
H11
20 12 - 01 - 01 08 : 08
To input the content, click , the instrument automatically stores the modified content.
is Return, is Backspace
Click on , the screen will become . By doing so, the unit of the device is
set to conventional unit.
2.4.2.4 ID setting:
ID setting
11
Input the required number of ID digits on the keyboard, the input range is 6-20 digits. Click on ,
Click on , the screen will become . By doing so, the device is switched for allow
transmission to prohibit transmission.
2.4.2.6 Printing Settings:
Click on , the screen will become . By doing so, the device is switched for allow
printing to prohibit printing.
2.4.2.7 Scanning Gun Settings:
Click on , the screen will become . By doing so, the device is switched for allow scanning
to prohibit scanning.
2.4.2.8 Test Strip Settings:
Click on H11, to choose from Type H10 and H11 test strips.
12
self-check. The screen displays "systems self-test". After the self-test is over, the screen displays the
main screen:
2012-01-01 08:08
(Search)
(ID Search)
(Menu)
3.2.2 After passing the self-test, it will directly enter the automatic test mode, and then the test can be
performed continuously in this mode.
3.2.3 The operation method is as follows:
a., after passing the self-check, the thruster is in the initial position.
b. to completely invade the reagent area of the test strip into the fresh, well-mixed, uncentrifuged urine
sample and take it out immediately, and scrape the side edge of the test strip along the tube wall of the
urine sample container to remove excess urine.
c. to place the test paper flat on the workbench, and ensure that the front end of the test paper is in
contact with the front wall of the workbench. The instrument automatically detects the test strip, the
pusher pushes the test strip into the test area, and the instrument starts testing.
d. when the thruster returns to the initial position, put down a piece of test paper to achieve continuous
testing.
Note: Do not place test strips when the thruster is moving.
e: The discarded test strips after use will automatically fall into the discard box for disposal.
13
must be kept clean.
4.2.4 Remove the propeller, workbench, and stepping board from the instrument in sequence, rinse
with clean water, and wipe with a soft cloth to dry.
4.3 Regular cleaning
If there is dirt on the propeller, workbench, and stepping board, you can clean it according to the
following steps
4.3.1 Prepare a small amount of 0.1mol/L NaOH solution
4.3.2 Remove the thruster, workbench, and stepping board in sequence from the instrument.
4.3.3 Dip a cotton swab with 0.1mol/L NaOH solution, scrub thoroughly and rinse off the remaining
NaOH solution with clean water.
4.3.4 Dry the propeller, workbench, and stepping board with a soft cloth and install according to the
content of Chapter 2.
4.3.5 Restart self-test.
4.4 Disinfection
Since the pusher, workbench, and step board are in contact with urine samples, they should be
disinfected regularly.
4.4.1 One of the following solutions can be used for disinfection:
-----2% glutaraldehyde solution.
-----5% sodium hypochlorite solution.
4.4.2 Choose an appropriate container to inject the disinfectant.
4.4.3 Invade the workbench, pusher, and step board into the disinfectant, and ensure that the white
reference does not touch the disinfectant.
4.4.4 Soak for 10 minutes.
4.4.5 After taking it out, rinse the disinfectant on the above items with water.
4.4.6 Use a soft cloth to dry the workbench, propeller and stepping board.
4.4.7 Restart self-test.
4.5 Use period and production date
The design life of the dry chemical urine analyzer is 10 years. If it is used for more than 10 years, it
may affect the correct judgment of the instrument. When the instrument is used for more than 10 years,
it is recommended to scrap it or notify the manufacturer for disposal according to the relevant
regulations of the medical institution.
See product label for production date.
4.6 Contraindications
No contraindications for dry chemical urine analyzers have been found.
14
V. Fault information table
This instrument is equipped with a replaceable fuse. When replacing the fuse, be sure to turn off the
power switch and unplug the power cord plugged into the power socket. Use a screwdriver to buckle
off the small tongue on the power socket to remove the damaged fuse. Take out and install the fuse of
the same specification (do not use fuses of other specifications)
The dry chemical urine analyzer has rainproof and soft liners in the outer packaging box to ensure that
the product is not damaged by nature. The product should be protected from strong mechanical
collisions during transportation, or choose a reasonable transportation method according to customer
requirements.
Dry chemical urine analyzer should be stored in a clean, ventilated and dry environment, away from
direct sunlight and away from chemical reagents and corrosive gases.
16
Manual revision date: August 10, 2020