HI991300 Multiparameter

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

HI991300 · HI991301

7 pH/EC/TDS/
Temperature
Meters
• Simultaneous, pH, EC/TDS and
temperature measurements on a
large three-line LCD display;
• User-friendly Design • Optional shockproof
· With only two buttons, meter operation silicon rubber boot
could not be simpler. Two buttons ·· Specially designed to
allow you to quickly adjust settings, protect your instrument
Multiparameter

select the measurement range, and from damage or impact


choose calibration buffer sets. HI710028 Orange
HI710029 Blue

• Watertight Connection
· A Quick Connect DIN connector makes
attaching and removing the probe
simple and easy. The rubber coating
protects the cable and creates a sealed
connection for added reliability.
• Probe Condition
· An on-screen indicator provides
visual confirmation that your
portable

probe is working at its best.


• Large LCD
· A multilevel display provides at-
a-glance readings of your most The HI991300 and HI991301 are light weight, portable pH, conductivity (or total dissolved solids)
important numbers from any angle. and temperature meters for portable applications requiring both a pH and conductivity (or TDS)
• Durable IP67 waterproofcCasing measurement. Applications include measurements for greenhouses irrigation, hydroponics and
· Designed to withstand the knocks, groundwater monitoring from agriculture nutrient pollution.
drops, and spills of real life, the new
The HI991300 and HI991301 meters feature 2 button operation and are simple to use. All
IP67 body ensures top performance
operations and settings, including calibration buffers and temperature scale selections, are made
in any environment. These meters
through these 2 buttons. They have a waterproof and compact casing rated for IP67 conditions
are totally protected against dust and
and a large Tri-line display. The meters have automatic pH calibration at one or two points and a
water intrusion from any direction.
single conductivity calibration. Other user selectable features include different TDS factors from
• On-screen calibration tags 0.45 to 1.00, and a range of temperature coefficients (β) from 0.0 to 2.4% for better conductivity
• mV of pH measurement or TDS solution temperature compensation. These meters are supplied with a multi-parameter
for electrode check probe specifically designed for these meters. To increase conductivity accuracy, two meter
models are available, with different conductivity ranges, for applications from purified to brackish
• Selectable temperature unit (°C or °F) waters.
• Battery life indication and
low battery detection The HI12883 multi-parameter probe, incorporates a domed shaped pH bulb rated from 0-13
pH, a single junction Ag/AgCl reference electrode with gelled electrolyte and a retractable
cloth wick junction, a graphite EC/TDS cell, and a temperature sensor in one convenient, rugged
polypropylene body. In addition, to ensure against interference from transient electrical noise to
pH, a solidstate preamplifier is integrated into the probe. The probe is rated from 0 to 50°C.

7.42 | www.hannainst.com
HI1288 amplified pH electrode which acts as a barrier between the inside

7
reference cell to the outside sample. This
• 3 sensors in a single probe barrier allows for a diffusion electrolyte that
• Pre-amplified pH electrode for is necessary for the pH measurement. Any
resistance to electrical noise clogging of the junction will result in a reduced
diffusion and as a result the readings will
• Extractable cloth junction to clear
become erratic. Most probes will have to have
any clogging this junction cleaned and if not possible then
• Graphite EC/TDS sensor the probe has to be replaced. The extractable
cloth junction of the HI1288 allows for the
renewing of the junction. Simply extract ⅛”
The HI991301 and HI991300 are supplied
of the junction by pulling on the junction will
with an amplified polypropylene body pH/EC/ expose a new portion. Any clogging that was
TDS/temperature probe. The built in amplifier present will be cleared and the response time
will reduce the effects of electrical noise will be back to normal extending the life of the

Multiparameter
on the high impedance pH measurement. pH electrode.
Examples of sources of electrical noise
include rectifiers, motors and ballasts. The EC/TDS sensor is made of graphite. A in the solution migrate to one of the negative
common problem with amperometric sensors or positive poles. When the charges build up
The HI1288 pH electrode also features an is a polarization effect. With amperometric on one of these poles a polarization effect
extractable cloth junction. Every pH electrode sensors there are two poles in which a voltage occurs. Having a conductivity sensor made of
has a junction. Many use a single ceramic frit is alternated. The positive and negative ions graphite reduces the polarization effect.

Specifications HI991300 HI991301


Range* -2.00 to 16.00 pH / -2.0 to 16.0 pH -2.00 to 16.00 pH / -2.0 to 16.0 pH

Resolution 0.01 pH / 0.1 pH 0.01 pH / 0.1 pH


pH Accuracy (@25°C/77°F) ±0.02 pH / ±0.1 pH ±0.02 pH / ±0.1 pH

automatic, 1 or 2 points choose between 2 sets of buffers automatic, 1 or 2 points choose between 2 sets of buffers
Calibration
(standard: 4.01; 7.01; 10.01 or NIST: 4.01; 6.86; 9.18) (standard: 4.01; 7.01; 10.01 or NIST: 4.01; 6.86; 9.18)

Range ±825 mV ±825 mV

pH-mV Resolution 1 mV 1 mV

Accuracy (@25°C/77°F) ±1 mV ±1 mV

Range 0 to 3999 μS/cm** 0.00 to 20.00 mS/cm**

EC Resolution 1 μS/cm 0.01 mS/cm

Accuracy (@25°C/77°F) ±2% F.S. ±2% F.S.

Range 0 to 2000 ppm (mg/L) 0.00 to 10.00 ppt (g/L)

TDS Resolution 1 ppm (mg/L) 0.01 ppt (g/L)

Accuracy (@25°C/77°F) ±2% F.S. ±2% F.S.

portable
Range* -5.0 to 105.0°C / 23.0 to 221.0°F -5.0 to 105.0°C / 23.0 to 221.0°F

Temperature Resolution 0.1°C/0.1°F 0.1°C/0.1°F

Accuracy (@25°C/77°F) ±0.5°C / ±1.0°F ±0.5°C / ±1.0°F

automatic, one point at: 1413 µS/cm or 1382 ppm automatic, one point at: 12880 µS/cm or 6.44 ppt
EC/TDS Calibration
(CONV=0.5) or 1500 ppm(CONV=0.7) (CONV=0.5) or 9.02 ppt (CONV=0.7)

pH Temp. Compensaiton automatic automatic

EC/TDS Temperature
automatic with β selectable from 0.0-2.4%/°C with 0.1 increments
Compenation
TDS Conversion Factor selectable from 0.45 to 1.00 with 0.01 increments
Probe (included) HI12883 pH/EC/TDS/temperature sensor, DIN connector and 1 m (3.3’) cable

Battery Type/Life 1.5V AAA (3) /approx. 600 hours of continuous use
Auto-off user selectable: after 8 min, 60 min or disabled
Environment 0 to 50°C (32 to 122°F); RH max. 100%
Meter Dimensions 154 x 63 x 30 mm (6.1 x 2.5 x 1.2”)
Meter Mass (with batteries) 196 g (6.91 oz.)

Casing Ingress Protection Rating IP67

HI991300 is supplied with HI12883 pH/EC/TDS probe with built-in temperature sensor, DIN connector and 1m (3.3’) cable, pH 4.01 and 7.01 buffer
sachets, HI70031 1413 μS/cm and HI70032 1382 ppm calibration solution sachets, HI700601 Electrode cleaning solution sachets (2), 100 mL beaker,
Ordering 1.5V AAA batteries (3), calibration certificate of meter, calibration certificate of probe, instruction manual and HI710142 rugged carrying case.
Information HI991301 is supplied with HI12883 pH/EC/TDS probe with built-in temperature sensor, DIN connector and 1m (3.3’) cable, pH 4.01 and 7.01 buffer
sachets, HI70030 12880 μS/cm and HI70038 6.44 ppt calibration solution sachets, HI700601 electrode cleaning solution sachets (2), 100 mL beaker,
1.5V AAA batteries (3), calibration certificate of meter, calibration certificate of probe, instruction manual and HI710142 rugged carrying case.

* the pH range is limited from 0 to 13 pH and the temperature range from 0 to 50°C (32 to 122°F) using HI12883 probe
** displays μS for μS/cm
** displays mS for mS/cm

pH solutions begin on page 2.144; EC and TDS solutions begin on page 5.34;
See page 7.50 for probe specifications www.hannainst.com | 7.43