Service Manual FMG60 SH01145FEN - 0221
Service Manual FMG60 SH01145FEN - 0221
Service Manual FMG60 SH01145FEN - 0221
Service Manual
Gammapilot FMG50
Radiometric measurement
Gammapilot FMG50
Serial number
2. www.endress.com/deviceviewer Endress+Hauser
Operations App
3.
A0023555
2 Endress+Hauser
Gammapilot FMG50 Table of contents
Table of contents
4 Conversions . . . . . . . . . . . . . . . . . . . . . . . . 8
4.1 Important information . . . . . . . . . . . . . . . . . . . 8
4.2 General guidelines for conversion . . . . . . . . . . . 8
4.3 Conversions that are not possible . . . . . . . . . . . 8
4.4 Conversions that are possible . . . . . . . . . . . . . . 8
5 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.1 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.2 Testing Ex d devices . . . . . . . . . . . . . . . . . . . . 9
5.3 Testing Ex i devices . . . . . . . . . . . . . . . . . . . . . 9
5.4 Documentation . . . . . . . . . . . . . . . . . . . . . . . . 9
5.5 Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.5.1 Tools for repairs on transmitter side . . 10
5.5.2 Tools for repairs on sensor side . . . . . 10
6 Final inspection . . . . . . . . . . . . . . . . . . . 11
6.1 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Endress+Hauser 3
About this document Gammapilot FMG50
1.1 Symbols
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
WARNING
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
CAUTION
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
NOTICE
This symbol contains information on procedures and other facts which do not result in
personal injury.
Result of a step
4 Endress+Hauser
Gammapilot FMG50 About this document
Endress+Hauser 5
Basic safety instructions Gammapilot FMG50
6 Endress+Hauser
Gammapilot FMG50 Notes for service staff
3.2 Contact
Inquiries and returns for repair and conversion must only be sent to the following address:
Calibration and Service Center
Endress+Hauser Messtechnik GmbH+Co. KG
Colmarer Straße 6
79576 Weil am Rhein
Germany
E-mail: [email protected]
Endress+Hauser 7
Conversions Gammapilot FMG50
4 Conversions
8 Endress+Hauser
Gammapilot FMG50 Repair
5 Repair
• The device may only be repaired and serviced by specialist staff.
• Pay attention to the associated device documentation and certificates, applicable
standards and legal regulations!
• For detailed information, see the associated Installation Instructions → SFT.
NOTICE
Loss of the factory dataset. There are no official HistoROM .dat files available in the
CER for devices older than 03/2021.
‣ Do not replace all parts at once.
‣ Contact the Calibration and Service Center.
Spare parts on transmitter side
• Display unit
• Main electronics
• Cover
• Dummy plug and cable glands
• Terminal strip
Spare parts on sensor side
• Housing adapter screws
• Sensor electronics
• Photomultiplier
• Coupling paste
• Scintillator
5.4 Documentation
In the case of certified devices, device repairs must be documented → CER (Common
Equipment Recorder)! This includes the serial number of the device, repair date, type of
repair and the technician performing the repair → SFT.
Endress+Hauser 9
Repair Gammapilot FMG50
5.5 Tools
For detailed information, see the relevant Installation Instructions → SFT and the
associated Operating Instructions.
Special tools for repairs on sensor side: Service case FMG60/FMG50 (order number:
71457678)
3 mm T10
Locking
6 mm 2 mm T20 Coupling Ø75 mm,
compound, Ø70 mm
paste 30 Nm
red
10 Endress+Hauser
Gammapilot FMG50 Final inspection
6 Final inspection
Once the spare part has been replaced, carry out a high-voltage test.
This requires the use of a high-voltage testing device with the following characteristics:
• Output voltage depending on the device version:
• Devices without overvoltage protection: 500 VAC or 707 VDC
• Devices with overvoltage protection: 410 VDC (lower testing voltage, otherwise
discharge tubes will ignite)
• Tripping current: 2.5 mA; discharge time 1.5 s
• Rise time of testing voltage: 2 s
• Hold time of testing voltage: 60 s
6.1 Preparation
Prior to performing the high-voltage test, use an ohmmeter to verify that there is
minimum contact resistance (0.1 Ω) between the PAL connectors and the housing. At the
terminals of the FMG50, use a wire jumper to short-circuit the signal circuits as illustrated
in the diagrams.
LDANGER
Danger! High voltage!
‣ During the test, place the device on an insulated surface, without ground connection.
Test steps
• 2-wire electronics module: step 1
• 4-wire electronics module: steps 1 - 3
1→ 2→ 3
V V V
Endress+Hauser 11
Diagnostics and troubleshooting Gammapilot FMG50
Pulse rate too low with empty Radiation source switched off Switch on source at the source
vessel container
Fittings in the vessel have not been Recalculate activity and change
considered in the activity calculation radiation source if required
Pressure in the vessel has not been Recalculate activity and change
considered in the activity calculation radiation source if required
Pulse rate too high with empty Activity too high Attenuate radiation, e.g. by mounting
vessel a steel plate in front of the source
container; or replace radiation source
Pulse rate too high with full External radiation sources present Shield off if possible; repeat
vessel (e.g. from gammagraphy) calibration without external radiation
source
Login via SmartBlue not No Bluetooth Enable Bluetooth function on smartphone or tablet
possible connection
Bluetooth function of sensor disabled, perform recovery
sequence
Device is being put into Enter initial password (device serial number) and
operation for the first change. Pay attention to lower/upper case when
time entering the serial number.
12 Endress+Hauser
Gammapilot FMG50 Diagnostics and troubleshooting
Device is visible in the live Android end device Is the location function permitted for the app, was it
list but cannot be accessed approved the first time?
via SmartBlue
GPS or positioning function must be activated for
certain Android versions in conjunction with Bluetooth
TAG in SmartBlue and HART System-related The device ID (TAG) is transferred to the live list via
do not match Bluetooth® to facilitate device identification. The tag is
abbreviated in the middle since the HART tag can be up
to 32 characters long but Bluetooth® can only use 29
characters as the device name.
Endress+Hauser 13
Diagnostics and troubleshooting Gammapilot FMG50
Diagnostic of sensor
007 Sensor defective 957 Sensor Defective Replace sensor electronics F Alarm
062 Sensor connection faulty 34 Sensor Connection Check sensor connection F Alarm
064 Pulse rate out of range 982 Count Values Not Vaild 1. Check process conditions C Warning
2. Check environmental conditions
3. Replace device
082 Data storage inconsistent 231 Memory Hardware Error 1. Check Data Unit F Alarm
Sensor 2. Contact service
Diagnostic of electronic
252 Module incompatible 107 Incompatible HW 1. Check if correct electronic modul is F Alarm
Configuration plugged
2. Replace electronic module
270 Main electronics defective 120 Sequence Counter MB Is Replace main electronics F Alarm
Invalid
14 Endress+Hauser
Gammapilot FMG50 Diagnostics and troubleshooting
272 Main electronics faulty 121 Internal Critical Error 1. Restart device F Alarm
2. Contact service
125 RAM Defect On MB F Alarm
273 Main electronics defective 229 Memory Hardware Error 1. Emergency operation via display F Alarm
Mainboard 2. Change main electronics
283 Memory content inconsistent 158 Memory Content Error FT 1. Transfer data or reset device F Alarm
2. Contact service
225 Memory Content Error F Alarm
PWD
287 Memory content inconsistent 108 Recovered From Memory 1. Restart device M Warning
0 Mirror 2. Contact service
311 Electronic failure 226 Memory Content Error Maintenance required! M Warning
Reset 1. Do not perform reset
2. Contact service
Diagnostic of configuration
410 Data transfer failed 139 Last Para Download Not 1. Check connection F Alarm
Successful 2. Retry data transfer
412 Processing download 516 Para Download Processing Download active, please wait C Warning
434 Real time clock defective 664 Real Time Clock Defect Replace sensor electronics C Alarm
435 Linearization faulty 117 LE Linearization Table Check linearization table F Alarm
Config Error
436 Date/time incorrect 75 Date Time Incorrect Check date and time settings. M Alarm
438 Dataset different 106 Last Para Download With 1. Check data set file M Warning
Warnings 2. Check device configuration
3. Up- and download new configuration
440 Device not calibrated 958 Device Not Calibrated Calibrate device F Alarm
441 Current output out of range 153 Range Current Out 1. Check process S Warning
2. Check current output settings
484 Failure mode simulation 144 Simulation Failsafe Deactivate simulation C Alarm
active
491 Current output 1 simulation 14 Simulation Current Output Deactivate simulation C Warning
active
Endress+Hauser 15
Diagnostics and troubleshooting Gammapilot FMG50
495 Diagnostic event simulation 350 Diagnostic Simulation Deactivate simulation C Warning
active
538 Configuration Sensor Unit 975 Configuration SEU Not 1. Check sensor configuration M Alarm
invalid Valid 2. Check device configuration
544 Background not calibrated 946 Background Not Calibrated Background not calibrated C Warning
586 Calibration active 102 Calibration In Progress Recording pulse rate M Alarm
8
593 Simulation pulse rate active 983 Simulation Pulse Rate Deactivate simulation C Warning
Diagnostic of process
801 Supply voltage too low 43 Power Low Increase supply voltage F Alarm
802 Supply voltage too high 44 Power High Decrease supply voltage S Warning
803 Loop current faulty 131 Read Back Current Is Wrong 1. Check wiring M Warning
2. Replace electronics
805 Loop current faulty 173 Read Back Current Is Wrong 1. Check wiring F Alarm
Alarm 2. Replace electronics
825 Operating temperature 135 Electronic Temperature Out 1. Check ambient temperature S Warning
Of Range 2. Check process temperature
826 Sensor temperature out of 978 Sensor Temperature Out Of 1. Check ambient temperature S Warning
range Range 2. Check process temperature
927 Overexposure detected 102 Over Exposure Please check source C Alarm
4
955 Gammagraphy detected 979 Gamma Graphy Detected Gammagraphy detected C Warning 1)
956 Evaluation plateau curve 102 Plateau Curve Evaluation Evaluation plateau curve M Warning
0
16 Endress+Hauser
Gammapilot FMG50 Diagnostics and troubleshooting
Actual diagnostics
The menu contains the "Actual diagnostics" parameter with a time stamp.
Previous diagnostics
The menu contains the "Previous diagnostics" parameter with a time stamp.
Event logbook
The events are saved in the event logbook.
Endress+Hauser 17
Diagnostics and troubleshooting Gammapilot FMG50
18 Endress+Hauser
Gammapilot FMG50 Firmware update with FieldCare
Vendor Endress+Hauser
Prerequisite: Ensure that the latest Fieldcare Device DTM Libraries are installed.
FieldCare Download Manager: http://portal.endress.com/wa010/
FieldCareDownloadGui/
1. Search for the DTM Release Notes in the DTM Documentation Library.
2. Check which is the latest available DTM version in the latest edition of the DTM
Release Notes.
3. Select the DTM version currently available.
4. Download with the FieldCare Download Manager.
Endress+Hauser 19
Firmware update with FieldCare Gammapilot FMG50
1 Sample graphic
2 Sample graphic
20 Endress+Hauser
Gammapilot FMG50 Firmware update with FieldCare
3. In the top left of the Network view, right-click the Host PC.
The context menu appears.
3 Sample graphic
4 Sample graphic
Endress+Hauser 21
Firmware update with FieldCare Gammapilot FMG50
5. In the Add new device window, select the CDI Communication FXA291 line and
confirm with OK.
In the Network view, the new line CDI Comm…◂▸ appears below the Host PC
line.
5 Sample graphic
6 Sample graphic
7. In this view, check to ensure that the right COM port is selected in the Serial port
line.
7 Sample graphic
22 Endress+Hauser
Gammapilot FMG50 Firmware update with FieldCare
8 Sample graphic
Endress+Hauser 23
Firmware update with FieldCare Gammapilot FMG50
9 Sample graphic
24 Endress+Hauser
Gammapilot FMG50 Firmware update with FieldCare
10. Below the CDI Communication FXA291 (configuration) view is the Flash icon: a
blue flash symbol.
10 Sample graphic
Endress+Hauser 25
Firmware update with FieldCare Gammapilot FMG50
11 Sample graphic
12 Sample graphic
An area opens at the bottom of the screen with information on the firmware
versions.
2. Select an entry for a new firmware version from this list.
26 Endress+Hauser
Gammapilot FMG50 Firmware update with FieldCare
3. Click the Start update icon to install the selected firmware version on the device.
The first feedback message displayed is as follows:
13 Sample graphic
4. Press OK to confirm.
The second feedback message displayed is:
14 Sample graphic
Endress+Hauser 27
Firmware update with FieldCare Gammapilot FMG50
5. Press OK to confirm.
Downloading commences.
15 Sample graphic
16 Sample graphic
6. Press OK to confirm.
28 Endress+Hauser
Gammapilot FMG50 Firmware update with FieldCare
17 Sample graphic
18 Sample graphic
Endress+Hauser 29
Firmware update with FieldCare Gammapilot FMG50
4. Click the Start update icon to install the selected firmware version on the device.
19 Sample graphic
20 Sample graphic
5. Press OK to confirm.
The second feedback message displayed is:
21 Sample graphic
30 Endress+Hauser
Gammapilot FMG50 Firmware update with FieldCare
6. Press OK to confirm.
Downloading commences.
22 Sample graphic
23 Sample graphic
7. Press OK to confirm.
Endress+Hauser 31
Service parameters Gammapilot FMG50
9 Service parameters
Reset option - - -
- - To service defaults
32 Endress+Hauser
Gammapilot FMG50 Service parameters
Endress+Hauser 33
*unbekannt*