Hollander - HF Oscillation Shunt Vs Pearson
Hollander - HF Oscillation Shunt Vs Pearson
Hollander - HF Oscillation Shunt Vs Pearson
Topic: High frequency oscillations measured with high bandwidth current sensors at low current
Tested: Pearson 2878 and SDN-414 shunts with different resistance values
Date: 2014 April 11th
Author: Martin Hollander
1 Summary
1.1 Introduction
A comparison was made between the responses of two types of current sensors when used in a fast
switching commutation circuit.
The wide bandwidth current transformer (type Pearson 2878) benefits from the fact that it does not
galvanically connect the power circuit to the oscilloscope, but it adds considerable leakage inductance
because of its geometry and its iron core.
The shunt (type T&M Research, SDN-414-50, 50 mΩ) benefits from a very low added leakage
inductance, but it galvanically connects the power circuit to the oscilloscope.
A simple patch test setup was made to compare the response of both sensor types. A commutation
circuit was made operating at low voltage for ease of experimenting. Not much attention was paid to
its construction to minimize parasitic elements since the patch test circuit was intended to make a
1-to-1 comparison of both current sensor types. So possible oscillations resulting from the test setup
would be measured by both sensors.
The waveform as measured by the 50 mΩΩ shunt showed a high frequency oscillation that was not
observed on the waveform as measured by the wide bandwidth current transformer.
1.2 Experiments
A number of experiments were performed to investigate if the oscillation was due to actual current
present in the circuit and running through the sensors. Or if the observed oscillation was due to other
effects in the measurement setup for which the shunt is more sensitive than the current transformer.
1.3 Conclusion
The conclusion from the experiments described in this memo is that the high frequency signal as
initially observed was mainly due to the radiated EMI-field generated by the load coil and its
connecting wires and to a minor extend due to other parts in the circuit. The immunity of the shunt, in
the way it was used in the patch test setup, for this external EMI-field was not as good as for the
current transformer that was less sensitive for this external EMI-field.
After some relative simple improvements to the EMI performance of the patch test setup, both sensor
types showed similar good performance for fast switching current signals at low amplitude (relative to
the maximum current rating of the sensors).
2 Measurement setup
The measurement setup is shown in Figure 2.1, which is a commutation circuit. The IGBT and diode are
rated for 1700 V, but operated only at a DC-bus voltage of only 30 V for ease of experimenting without
safety concerns. Currents are up to 5 A, devices are rated for 10 A.
SDN-414
Figure 2.1 Measurement setup for testing Pearson 2878 and 50 mΩ shunt at low currents.
Both current sensor types are put in series as shown in Figure 2.2. No effort was made to minimize the
parasitic capacitance in parallel to the diode and the leakage inductance in series with the IGBT. The
gate driver is a ±15 V IGBT gate driver with optical signal input.
Figure 2.2 Measurement setup for testing Pearson 2878 and 50 mΩ shunt at low currents.
The shunt shows an oscillation with a frequency of about 20 MHz on the tail current after the
switching. This oscillation is not present on the response of the Pearson 2878.
The 20 MHz oscillation is within the specified measurement bandwidth of the Pearson 2878 which has
the -3dB point at 70 MHz. The attenuation at 20 MHz is A = 0.95 for a single order 70 MHz system.
Question statements:
Do the recorded waveforms represent the actual current in the circuit?
Why do the responses of both sensor types differ from each other?
The current at turn-off the IGBT has been measured with the SDN-414-50 shunt (Pearson removed
from system). Only measurement signal to the oscilloscope is from the shunt, other inputs to the
oscilloscope (LeCroy WaveAce 2024) are unused. Input impedance oscilloscope at 50 Ω.
Figure 4.1.1 compares two measurements, one at 1.2 A (10 mV/div) and the other at 2.4 A (20 mV/div)
with comparable results.
The oscilloscope is connected to the mains by a mains cable with a ferrite core, forming a common
mode filter. Another experiment with a complete disconnect of the DC-power supply feeding the DC-
bus capacitor during the pulse shows the same result.
Figure 4.1.2 shows identical oscillation for two BNC-cables to two channels of the oscilloscope. The
shunt is connected to a BNC T-split with two channels measuring the same shunt signal with two
different BNC-cables to the oscilloscope.
In a simplified setup the current of the Pearson 2878 and the 50 mΩ shunt is compared. An isolated
IGBT gate driver with optic input acts as an isolated current source. Uout = ±15 V, Rg = 4.7 Ω into
capacitor Cge of 10 nF. The resulting current to charge Cge is measured with the Pearson 2878 and the
shunt (type SDN-414-50).
Figure 4.1.3 shows the recoded waveforms of the Pearson and the shunt for the charge current of Cge.
The frequency of this signal is somewhat lower than the frequency of the turn-off current waveform
measured in Figure 3.1.
Conclusion:
• No trivial explanation found for oscillation based on measurements related to connection with
oscilloscope.
• Result of Figure 4.1.3 with floating gate driver proves that observed response difference (oscillation)
between current transformer and shunt is not present in a more simplified measurement setup.
The 50 mΩ shunt was connected with the 1m Radiall RG223 double screened cable, length 1 m
(R284C0351012), to the oscilloscope with input impedance set to 50 Ω by external 50 Ω termination on
BNC T-split.
Figure 4.2.1. shows the measured response for the Pearson and the 50 mΩ shunt for a resistive
switching load. With the fine gain adjustment of the oscilloscope, the waveform of the shunt is scaled
to visually fit the waveform of the current transformer.
Figure 4.2.1. Identical response for Pearson 2878 and shunt SDN-414-50 for switching a resistive load.
Conclusion:
• Oscillation observed in Figure 3.1 is not due to the high frequency current slope.
• Oscillation is not present in circuit with resistive load.
By removing the clamp capacitor on the DC-bus and adding an inductive loop in series with the
emitter of the IGBT and adding a capacitor in parallel to the diode, a high frequency oscillation is
provoked which should be recorded by the current sensors.
Figure 4.3 shows a resonance with a frequency of about 31 MHz during the slope. This resonance is
due to a 470 pF capacitor that is placed in parallel to the diode and additional leakage inductance.
Even more inductance is added by introducing a round loop with a diameter of about 5 cm between
the emitter of the IGBT and the current sensors. Figure 4.3.2 shows the response of the Pearson and
the shunt with this increased inductance (without the parallel capacitance to the diode). The
resonance frequency is about 20 MHz.
Figure 4.3.3 shows the same signals at same conditions at higher current. Itop = 2.3 A, tfall = 31 ns.
Conclusion:
• The Pearson and shunt show identical responses for the provoked high frequency oscillation during
and after the slope.
• The lower frequency oscillation shown about 100 ns after the slope in Figure 4.3.2 does not
significantly scale in amplitude when increasing the turn-off current a factor 4 in this test setup with
additional leakage inductance.
The circuit of Figure 2.1 with an additional inductive loop between the emitter of the IGBT and the
current sensors is used to test the response at low current for the Pearson 2878 and SDN-414 shunts
with different resistance values. Udc = 30 V, Itop = 612 mA, Rg,driver = 0 Ω, tfall = 38 ns.
Figure 4.4.1 shows the response of the Pearson and the 50 mΩ shunt. The oscillation directly after the
slope has a frequency of about 18.5 MHz.
Figure 4.4.2 shows the responses for the 100 mΩ shunt (left) and the 25 mΩ shunt (right).
yellow Pearson 2878 (100 mA/div) yellow Pearson 2878 (100 mA/div)
cyan SDN-414-100mΩΩ (100 mA/div) cyan SDN-414-25mΩΩ (100 mA/div)
Figure 4.4.2 Responses for Pearson 2878 and 100 mΩ shunt (left) and 25 mΩ shunt (right)
Figure 4.4.3 shows the response of the Pearson and the 1.84 mΩ shunt. The low shunt resistance in
combination with low current levels results in low voltage signal levels to be measured.
Figure 4.4.4 and 4.4.5 show the responses of the Pearson 2878 compared to SDN-414 shunts with
different resistance values for the test setup with additional leakage inductance loop at a turn-off
current of 0.6 A. Figure 4.4.4 shows the waveforms as recorded by the oscilloscope in Volts. Figure 4.4.5
shows the measured current, which is determined by same voltage waveforms divided by the sensor
sensitivity.
1"#&$2+(3$(')4"&"+5($6%+5(3#,%"$(7(!"#$8(32,5#*"(
"#"+%
"#"*%
"#")%
%%./01234%
!"#$%&"'($)*+#,(-./(
"#"(%
%%$""56%
"#"'% %%)"56%
%%&)56%
"#"&%
%%$"56%
%%$#,(56%
"#"$%
"%
"% $""% &""% '""% (""% )""% *""% +""% ,""% -""%
!"#"$%
0!"(-+$/(
Figure 4.4.4 Recorded voltage by oscilloscope for Pearson 2878 different SDN-414 shunts at 0.6 A.
.$/#-0%'1-'234$#$%&'-5"%&'1/6"$-'7'#$6/&$2'!"##$%&'
"#+%
"#*%
"#)%
"#(% %%,-./012%
%%&""34%
!"##$%&'()*'
"#'%
%%)"34%
"#$% %%$)34%
%%&"34%
"#&%
%%(34%
"%
&""% &)"% $""% $)"% '""% ')"% (""% ()"% )""%
!"#&%
!"#$%
+,$'(%-*'
Figure 4.4.5 Measured current by Pearson 2878 different SDN-414 shunts at 0.6 A.
Conclusion:
• The resistance of the shunt does not significantly affect the behavior of the oscillation.
• Signal fidelity is best at higher resistance values because the signal level has higher amplitude at
higher shunt resistance values.
The setup is in its original configuration without the additional inductive loop and without additional
parasitic capacitance. The Pearson 2878 is part of the setup, but the shunt is removed from the circuit
and replaced by a short-circuit.
Then, only the midpoint of the shunt is connected to the circuit as indicated in Figure 4.5.1.
14
SDN-4
Figure 4.5.1 Test setup for measuring disturbance with shunt only connected with one lead to the circuit.
Figure 4.5.2 shows the measured signals of the Pearson and the 25 mΩ shunt for the setup shown
above in Figure 4.5.1.
With the original setup as in Figure 2.1, the response of the Pearson and the 50 mΩ shunt was
measured while positioning a coil close to the Pearson and the shunt. This coil is formed by 11 turns
with a diameter of about 5 cm of the output leads of the midpoint to the coil. In Figure 4.5.3 it can be
seen that both the response of the Pearson and the shunt show a lower frequency oscillation on top of
the tail current of the IGBT. The connecting leads to the shunt with a length of about 4.5 cm are spread
apart and form a triangle shape with approximate distance at the end of the leads of about 2.5 cm.
With the original setup without the air coil, different experiments were done with near field probes
(H and E probes) of different sizes at various locations in the setup. Figure 4.5.4 shows the response of
the 50 mΩ shunt, the output voltage of the commutation cell and the signal of a big (10 cm) H-probe
close to the commutation cell. The response of the H-probe was found to be relative clean at the
power supply and coil. The output voltage is measured by a voltage probe that is connected to the
collector of the IGBT with the reference lead connected to the ground (flange) of the shunt.
Figure 4.5.5 shows the measured signals of the 50 mΩ shunt and the E-probe near the shunt leads. The
measured frequency of the E-field is approximately 9.6 MHz.
Similar frequencies were found with the E-probe at the leads from the DC-bus capacitor to the power
supply and near the core of the load coil, see Figure 4.5.6 and Figure 4.5.7 respectively.
Conclusion:
• The EM-field close to the commutation cell has a frequency in the range of 20 – 25 MHz. This
frequency correlates with the oscillation recorded by the Pearson 2878 at the end of the slope as
seen for example in Figure 4.5.2. (yellow waveform).
• At the load coil the E-field has a frequency of about 10 MHz, this frequency is also seen at the
location of the shunt and at the leads connecting the DC-bus capacitors to the power supply. The
Pearson 2878 is less sensitive for this field than the SDN-414 shunts (in the setup used), as can be
seen in Figure 4.5.2 where the shunt signal shows this lower frequency oscillation while no current
is routed through its leads as indicated in the circuit in Figure 4.5.1.
Figure 5.2.3 shows the response of the Pearson 2878 and the 1.84 mΩΩ shunt. This low resistive shunt
has limited bandwidth. The Pearson 2878 as well as the shunts with higher resistance have sufficient
bandwidth to measure the fast turn-off current waveform with tfall = 26 ns. The low resistance shunt
signal shows some remaining 18 MHz oscillation since the generated EMI field is relative high at 5 A
and the measurement signal is relative low (2 mV/div).
Figure 5.2.4 and 5.2.5 show the responses of the Pearson 2878 compared to SDN-414 shunts with
different resistance values for the test setup with improved EMI at a turn-off current of 0.5 A. Figure
5.2.4 shows the waveforms as recorded by the oscilloscope in Volts. Figure 5.2.5 shows the measured
current, which is determined by same voltage waveforms divided by the sensor sensitivity.
1"#&$2+(3$(')4"&"+5($6%+5(3#,%"$(7(!"#$8(32,5#*"(
!#!)"
!#!("
!#!'"
!"#$%&"'($)*+#,(-./(
""-./0123"
""$!!45"
!#!&"
""(!45"
""%(45"
!#!%"
""$!45"
""$#+'45"
!#!$"
!"
!" $!!" %!!" &!!" '!!" (!!" )!!" *!!" +!!" ,!!"
0!"(-+$/(
Figure 5.2.4 Recorded voltage by oscilloscope for Pearson 2878 different SDN-414 shunts at 0.5 A.
.$/#-0%'1-'234$#$%&'-5"%&'1/6"$-'7'#$6/&$2'!"##$%&'
!#)"
!#("
!#'"
""*+,-./0"
!"##$%&'()*'
""$!!12"
!#&"
""(!12"
""%(12"
!#%"
""$!12"
""$#3'12"
!#$"
!"
$!!" $(!" %!!" %(!" &!!" &(!" '!!" '(!" (!!"
+,$'(%-*'
Figure 5.2.5 Measured current by Pearson 2878 different SDN-414 shunts at 0.5 A.
1"#&$2+(3$(')4"&"+5($6%+5(3#,%"$(7(!"#$8(32,5#*"(
!#)"
!#("
!#'"
!"#$%&"'($)*+#,(-./(
""-./0123"
""$!!45"
!#&"
""(!45"
""%(45"
!#%"
""$!45"
""$#+'45"
!#$"
!"
!" $!!" %!!" &!!" '!!" (!!" )!!" *!!" +!!" ,!!"
0!"(-+$/(
Figure 5.2.6 Recorded voltage by oscilloscope for Pearson 2878 different SDN-414 shunts at 5 A.
.$/#-0%'1-'234$#$%&'-5"%&'1/6"$-'7'#$6/&$2'!"##$%&'
!#)"
!#("
!#'"
""*+,-./0"
!"##$%&'()*'
""$!!12"
!#&"
""(!12"
""%(12"
!#%"
""$!12"
""$#3'12"
!#$"
!"
$!!" $(!" %!!" %(!" &!!" &(!" '!!" '(!" (!!"
+,$'(%-*'